Doc number: 77
Abstract
Background: Implantation of implantable cardioverter-defibrillators (ICD) from the left pectoral region is the standard therapeutical method. Increasing numbers of system revisions due to lead dysfunction and infections will consecutively increase the numbers of right-sided implantations. The reliability of devices implanted on the right pectoral side remains controversially discussed, and the question of testing these devices remains unanswered.
Methods: In a prospectively designed study all 870 patients (60.0±14 years, 689 male) who were treated with a first ICD from July 2005 until May 2012 and tested intraoperatively according to the testing protocol were analyzed. The indication for implantation was primary prophylactic in 71.5%. Underlying diseases included ischemic cardiomyopathy (50%), dilative cardiomyopathy (37%), and others (13%). Mean ejection faction was 27±12%. Implantation site was right in 4.5% and left in 95.5%.
Results: Five patients supplied with right-sided ICD (13%, p = 0.02 as compared to left-sided) failed initial intraoperative testing with 21 J. 3 patients were male. The age of the patients failing intraoperative testing with right-sided devices appeared higher than of patients with left-sided devices (p = 0.07). The ejection fraction was 28±8%. All patients reached a sufficient DFT ≤ 21 J after corrective procedures.
Conclusion: Implantation of ICDs on the right side results in significantly higher failure rate of successful termination of intraoperatively induced ventricular fibrillation. The data of our study suggest the necessity of intraoperative ICD testing in right-sided implanted ICDs.
You have requested "on-the-fly" machine translation of selected content from our databases. This functionality is provided solely for your convenience and is in no way intended to replace human translation. Show full disclaimer
Neither ProQuest nor its licensors make any representations or warranties with respect to the translations. The translations are automatically generated "AS IS" and "AS AVAILABLE" and are not retained in our systems. PROQUEST AND ITS LICENSORS SPECIFICALLY DISCLAIM ANY AND ALL EXPRESS OR IMPLIED WARRANTIES, INCLUDING WITHOUT LIMITATION, ANY WARRANTIES FOR AVAILABILITY, ACCURACY, TIMELINESS, COMPLETENESS, NON-INFRINGMENT, MERCHANTABILITY OR FITNESS FOR A PARTICULAR PURPOSE. Your use of the translations is subject to all use restrictions contained in your Electronic Products License Agreement and by using the translation functionality you agree to forgo any and all claims against ProQuest or its licensors for your use of the translation functionality and any output derived there from. Hide full disclaimer