Lizhi Xu1,*, Sarah R. Gutbrod2,*, Andrew P. Bonifas1, Yewang Su3,4, Matthew S. Sulkin2, Nanshu Lu5, Hyun-Joong Chung6, Kyung-In Jang1, Zhuangjian Liu7, Ming Ying1, Chi Lu1, R. Chad Webb1, Jong-Seon Kim1,8, Jacob I. Laughner2, Huanyu Cheng3, Yuhao Liu1, Abid Ameen1, Jae-Woong Jeong1, Gwang-Tae Kim1, Yonggang Huang3, Igor R. Emov2 & John A. Rogers1
Means for high-density multiparametric physiological mapping and stimulation are critically important in both basic and clinical cardiology. Current conformal electronic systems are essentially 2D sheets, which cannot cover the full epicardial surface or maintain reliable contact for chronic use without sutures or adhesives. Here we create 3D elastic membranes shaped precisely to match the epicardium of the heart via the use of 3D printing, as a platform for deformable arrays of multifunctional sensors, electronic and optoelectronic components. Such integumentary devices completely envelop the heart, in a form-tting manner, and possess inherent elasticity, providing a mechanically stable biotic/abiotic interface during normal cardiac cycles. Component examples range from actuators for electrical, thermal and optical stimulation, to sensors for pH, temperature and mechanical strain. The semiconductor materials include silicon, gallium arsenide and gallium nitride, co-integrated with metals, metal oxides and polymers, to provide these and other operational capabilities. Ex vivo physiological experiments demonstrate various functions and methodological possibilities for cardiac research and therapy.
1 Department of Materials Science and Engineering, Frederick Seitz Materials Research Laboratory, University of Illinois at Urbana-Champaign, Urbana, Illinois 61801, USA. 2 Department of Biomedical Engineering, Washington University in Saint Louis, Saint Louis, Missouri 63130, USA. 3 Department of Civil and Environmental Engineering, Department of Mechanical Engineering, Center for Engineering and Health and Skin Disease Research Center, Northwestern University, Evanston, Illinois 60208, USA. 4 Center for Mechanics and Materials, Tsinghua University, Beijing 100084, China. 5 Department of Aerospace Engineering and Engineering Mechanics, University of Texas at Austin, Austin, Texas 78712, USA. 6 Department of Chemical and Materials Engineering, University of Alberta, Edmonton, Alberta, Canada T6G 2V4. 7 Institute of High Performance Computing, Agency for Science, Technology and Research,1 Fusionopolis Way, #16-16 Connexis, Singapore 138632, Singapore. 8 Department of Chemical and Biomolecular Engineering (BK21 Program), Korea Advanced Institute of Science and Technology, Daejeon 305-701, Republic of Korea. * These authors contributed equally to this work. Correspondence and requests for materials should be addressed to I.R.E. (email: mailto:[email protected]
Web End [email protected] ) or to J.A.R. (email: mailto:[email protected]
Web End [email protected] ).
NATURE COMMUNICATIONS | 5:3329 | DOI: 10.1038/ncomms4329 | http://www.nature.com/naturecommunications
Web End =www.nature.com/naturecommunications 1
& 2014 Macmillan Publishers Limited. All rights reserved.
DOI: 10.1038/ncomms4329
3D multifunctional integumentary membranes for spatiotemporal cardiac measurements and stimulation across the entire epicardium
ARTICLE NATURE COMMUNICATIONS | DOI: 10.1038/ncomms4329
Tools for cardiac physiological mapping are indispensable for the clinical identication and mechanistic understanding of excitationcontraction coupling, metabolic
dysfunction, arrhythmia and other conditions. Devices developed in the 1980s attempted to address this need by using synthetic fabrics sewn to loosely resemble the shape of the ventricle, with bulk electrodes manually assembled and woven into this platform14. Although such schemes provide some utility, they do not enable uniform quality of contact across the heart high-density mapping capabilities, provision for multifunctional recordings, precision measurement/stimulation or practical deployment as chronic implants. As a result, alternative strategies based on serial mapping with point-contact catheters or on imaging techniques that use uorescence, nuclear magnetic resonance or ultrasound have emerged, even though each has signicant shortcomings58.
The ideal scenario remains one in which device functionality integrates directly and non-invasively with the heart, suitable for chronic use. The essential challenge is that the heart is a complex electromechanical syncytium with numerous elements working in synchrony to reliably pump blood and respond to changing metabolic demands. Although much has been gained from isolated cellular studies, the integral functional behaviour on the organ level and the interaction between the electrical, metabolic and mechanical remodelling in disease states, especially in vivo, remain poorly explored due to paucity of adequate tools. Thus, there is an unmet need for multiparametric mapping capabilities inclusive but far beyond electrical sensing in a conformal, high-resolution manner, which cannot be realized using conventional materials, device technologies or imaging modalities.
Recent developments in materials and mechanics concepts for stretchable electronics914 create an opportunity to meet this challenge of direct, full three-dimensional (3D) integration of devices with the epicardial surface. Here we expand on previously reported small-scale electronic devices as two-dimensional (2D) exible sheets15, to build multifunctional semiconductor systems in lithographically dened congurations on 3D, thin elastic membranes, custom-formed to match the shape of the heart. The physical format resembles that of the naturally occurring membrane that surrounds the heart, known as the pericardium. These systems, which we refer to as 3D multifunctional integumentary membranes (3D-MIMs) provide conformal interfaces to all points on the heart, with robust but noninvasive contacts enabled by the soft elasticity of the membrane itself, throughout dynamic cardiac cycles, even when completely immersed in uid media. Measurements on isolated perfused rabbit hearts demonstrate the utility of these ideas as a general platform for multifunctional, high-density epicardial mapping/ stimulation. The results provide advanced methodological possibilities for basic and clinical cardiology.
ResultsDesign and fabrication. The fabrication begins with the creation of a thin, 3D elastic membrane shaped to the heart. As shown in Fig. 1a, optical segmentation techniques rst capture the full 3D geometry of a heart of interest16. A commercial 3D printer (ZPrinter 450, Z-Corporation) then renders a solid model of the heart in a proportionally scaled form, as described later, to serve as a substrate for mounting ultrathin electronic/optoelectronic and sensor systems, separately prefabricated on planar substrates. Casting and curing a thin layer of silicone elastomer on top of the heart model with these multifunctional devices on its surface denes the overall format. The front faces of the device components contact the model while the back faces bond to the
elastomer. Removing the system (that is, 3D membrane with integrated device components) from the model prepares it for installation around a living heart, as a type of instrumented, articial pericardium.
Figure 1b shows a representative 3D-MIM that includes microscale, inorganic light-emitting diodes (m-ILEDs) based on indium gallium nitride (InGaN) for optical mapping, silicon (Si) nanomembranes for strain gauges, gold (Au) electrodes for electrical sensing/stimulation, iridium oxide (IrOx) pads for pH sensors and Au serpentine resistors for temperature sensors/ heaters. The methods for creating these components exploit modern integrated circuit technologies and achieve spatial resolution far beyond that possible with manually assembled arrays. A thin, exible heat-seal conductive cable (Elform, HST-9805-210) serves as a connection to external hardware for data acquisition, power supply and control. The 3D-MIM is engineered with overall dimensions slightly smaller than those of the real heart, to provide adequate elasticity and mechanical support for robust contact with the epicardium during diastole and systole, but with sufciently small pressures to avoid disruption of natural behaviours of the cardiac tissue. The serpentine mesh that interconnects the device components covers the ventricle and conforms to the contours of the epicardium. Although this example is designed for research applications on rabbit hearts, the same strategies are applicable to human hearts, or even other organ systems. Here the 3D geometries can be obtained using similar 3D-printed substrates with patient-specic MRI or CT organ segmentation.
Mechanical analysis. A critical feature of this type of device is that it can be designed to maintain a stable mechanical interface to the tissue while exerting minimal force on the contracting and relaxing heart muscle. In the cardiac anatomy of humans and other vertebrates, the myocardium is enclosed in a space sealed by the pericardium, which allows reversible volume change within a certain range. When a pathophysiological condition leads to inammation, the pericardium exerts pressure to constrain the motions of the heart chambers. Quantitative analysis allows a comparative assessment of the pressure associated with our 3D device membrane on the epicardium, as well as the dependence of this pressure on material properties and design parameters. Figure 2a shows results for a 3D-MIM with a membrane thickness of 150 mm and effective Youngs modulus of B60 kPa (Ecoex, Smooth-on) at various states of volume expansion (1 e)3 of a heart geometry, calculated using 3D nite element
methods (FEM), where e is the linear expansion factor. The thickness of the membrane is uniform in the analysis; the nonuniformity due to the electronic devices results in local increase of the effective Youngs modulus to B80 kPa and adds o50% of the approximate pressure, as discussed in details in Methods section and Supplementary Fig. 1. The form of the undeformed membrane follows that of a 3D model, proportionally size-reduced (B30% volume reduction compared with the diastolic state of the real heart) to ensure a baseline level of pressure upon application on the real heart. Computations correspond to the heart at its contracted volume (3D model), and at systolic (120% of the contracted volume) and diastolic (145% of the contracted volume) conditions. The calculated average pressures are similar to those of pericardium under normal physiological conditions, and only B20% of these pressures are under conditions of pericardial constraint1722. The results suggest that the device is unlikely to cause restrictive impact, as conrmed by ex vivo studies described subsequently. FEM and analytic modelling also establish general relationships between the pressure and the design parameters. Figure 2b shows the average pressure as a
2 NATURE COMMUNICATIONS | 5:3329 | DOI: 10.1038/ncomms4329 | http://www.nature.com/naturecommunications
Web End =www.nature.com/naturecommunications
& 2014 Macmillan Publishers Limited. All rights reserved.
NATURE COMMUNICATIONS | DOI: 10.1038/ncomms4329 ARTICLE
Geometrical information 3D printed model 3D-MIM Integrate with heart
Si strain gauge ECG sensor
pH sensor
-ILED
Temperature sensor
ECG sensor Si strain gauge -ILED Temperature sensor
pH sensor
Figure 1 | 3D-MIMs for spatiotemporal measurement and stimulation across the entire epicardial surface. (a) Graphical depiction of the key steps in device design and fabrication. Scale bar, 2 cm. (b) Images of a representative 3D multifunctional integumentary membrane (3D-MIM) integrated on a Langendorff-perfused rabbit heart. The white arrows highlight various function elements in this system. The electronics can cover both anterior and posterior surfaces of the heart (inset). Scale bars, 6 mm. (c) Magnied views of the functional elements in conformal contact with the epicardium.
The images are recorded from the back side of the devices. Scale bars, 500 mm.
function of the volume expansion, the thickness of the membrane and its Youngs modulus. The analytic model uses a partial ellipsoid to approximate the geometry of the heart. Details appear in Methods section and Supplementary Fig. 1. The following expression connects the average pressure, the membrane geometry, mechanical properties and expansion factor:
Paverage C
Ete 1 n
1 e
1
where t is the thickness of the membrane, E and n are the effective Youngs modulus and the Poissons ratio, respectively. The constant C decreases as the heart size increases, and C also depends on the shape of the heart (B0.2 mm 1 for a rabbit heart). Decreases in membrane thicknesses and Youngs moduli both linearly reduce the pressure. This scaling allows designs that provide pressures sufciently large to maintain good contact between the sensor/actuator network and the epicardial surface, but sufciently small to avoid impact on the intrinsic physiology. Monitoring the time course of several electrophysiological parameters that indicate ischaemia in an isolated pressure-loaded, working rabbit heart model23 with and without a
3D-MIM reveals the effects. The results, based on control (N 3) and experimental (N 3) hearts (Supplementary Fig. 2),
suggest that there is no additional ischaemia caused by the devices, as measured by ST elevation and the amplitude of the LV pressure waveform.
Spatiotemporal cardiac measurements and stimulation. To demonstrate the various functional modes of operation we begin with high-precision mapping of epicardial electrical activity. These experiments, and all of those that follow, used explanted Langendorff-perfused rabbit hearts. The 3D geometrical information was obtained from a representative rabbit heart. A single 3D-MIM can accommodate some range in specic sizes and shapes associated with a single type of animal model, due to its soft, elastic construction. The device here incorporates 68 Au electrodes (1 mm2 surface area and spacing of 3.5 mm), distributed across both the anterior and posterior surfaces of the epicardium (Fig. 3a,d and Supplementary Fig. 3). The electro-chemical impedances of individual electrodes are B2 kO at a frequency of 1 kHz, measured in phosphate-buffered saline
2
NATURE COMMUNICATIONS | 5:3329 | DOI: 10.1038/ncomms4329 | http://www.nature.com/naturecommunications
Web End =www.nature.com/naturecommunications 3
& 2014 Macmillan Publishers Limited. All rights reserved.
ARTICLE NATURE COMMUNICATIONS | DOI: 10.1038/ncomms4329
+400.0 +333.3 +266.7 +200.0 +133.3 +66.67 +0.000
Pressure (Pa)
Unexpanded (100% volume) Average pressure=0 Pa
Analytic model
Expanded (120% volume) Average pressure=164 Pa
Expanded (145% volume) Average pressure=296 Pa
1,000
Average pressure (Pa)
1,000
Average pressure (Pa)
500
FEM Analytic
Expanded (145% volume)
Average pressure (Pa)
Expanded (145% volume)
800
800
400
600
600
300
200
400
400
Unexpanded (100% volume)
100
200
200
Unexpanded (100% volume)
0 100 120 140 160 180 200
0 200 300
400 500
0 60 90 120 150
Volume after expansion (%)
Thickness (m)
Youngs modulus (kPa)
Figure 2 | Analysis of pressures on the epicardium associated with integration of a 3D-MIM. (a) Calculated pressure distribution induced by a device with total thickness of 150 mm and effective Youngs modulus of 60 kPa under various conditions of volume expansion of a heart geometry. (b) FEM and analytical results of average pressure as functions of volume expansion (left), thickness (middle) and Youngs modulus (right) of the membrane.
(Supplementary Fig. 4). The transparency of the membrane allows simultaneous optical mapping through voltage-dependent uorescence, as a means for validating the electrical measurements5. Experiments involved signals acquired from four hearts for a variety of conditions: normal sinus rhythms, and paced at a range of frequencies and from a range of electrode pairs to increase the variability of the propagation patterns in the spatial activation maps. The surface electrograms (EGs) captured various key morphologies associated with the QRS and T waves (Fig. 3a). Representative maps and correlations between electrical and optical activation times appear in Fig. 3c,b, respectively. The overall linear correlations between optical and electrical activation times were 0.957 for sinus data and 0.943 for paced data. These studies indicate that this conguration of measurement electrodes can replicate patterns of activation to a resolution that captures the spatial variations observed optically. Analyses for additional electrophysiological parameters are summarized in Supplementary Fig. 6. Figure 3d presents a 3D map derived from signals recorded from the anterior and posterior surfaces of the heart. Unlike optical mapping where motion artefacts dramatically impact the measurement quality requiring static heart geometries, electrophysiological mapping with 3D-MIMs can be applied under normal beating condition. As shown in Supplementary Video, the integrated sensors move synchronously with the underlying cardiac tissue. Although it is practically difcult to avoid relative lateral motion between the sensors and the epicardium during beating cycles, due to the engineered geometries of 3D-MIMs, the displacement can be minimized to be less than the characteristic sizes of the sensors and to have negligible impact to the signal quality (Supplementary Fig. 5). This feature is necessary for extending the mapping capabilities beyond laboratory studies and implementing in clinical electrophysiology.
Mapping of changes in pH provides useful information on the metabolic state of the heart. Here iridium oxide (IrOx), a well-established material for pH sensing, enables the measurement24. Electrodeposited IrOx on Au electrodes provides pH sensors with average open circuit potential responses of 68.9 mV per pH with
s.d. of 8.6 mV per pH for 32 sensors over the array at 37 C in Tyrodes solution (Supplementary Fig. 7). The temperature dependence of the pH sensors is nearly 1.60.02 mV C 1
(ms), which corresponds to B0.02 pH for a 1 C change in temperature. Temperature variations over a physiologically relevant range will, therefore, have a small effect on the pH measurement. For cases where large changes in temperature are externally introduced, the temperature dependence of the pH sensor must be accounted for explicitly. Such pH sensors, along with optical mapping techniques, enable acquisition of maps of pH, transmembrane potential (Vm) and calcium transient (CaT)
signals during global no-ow ischaemiareperfusion. The pH sensors cover the left anterior and posterior surface of the rabbit heart (Fig. 4a). At baseline, all pH sensors record values between7.34 and 7.40. The responses of two pH sensors (highlighted by grey and charcoal colours) are plotted (Fig. 4b) throughout the protocol. Complete spatial pH maps at time points t1 (baseline), t2 (10 min into ischaemia) and t3 (20 min into reperfusion) appear in Fig. 4df (left). Turning off the perfusion pump immediately reduced coronary pressure to 0 mm Hg and led to an approximately linear decrease in pH to minimum values of 6.40 (grey) and 6.22 (charcoal). Upon reperfusion, the pH rapidly increased until initiation of ventricular tachycardia where the pH stabilized at levels somewhat below baseline values. A sample fareld ECG of reperfusion-induced ventricular tachycardia appears in Fig. 4c. After spontaneous conversion back to sinus rhythm, the pH values increased again to pre-ischemic values. Figure 4df shows pH maps (left), representative optical signals (Vmblack and CaTgrey; middle) and side-by-side action potential duration at 70% repolarization (APD70) and calcium transient duration at 70% return to baseline (CaT70) maps. At the baseline, pH, APD70 and CaT70 maps highlight that the pH and electrophysiological parameters were initially uniform over the surface of the heart. After 10 min of ischaemia, pH, APD70 and CaT70 changed, though not in a spatially uniform manner. CaT alternans (short and long duration) were observed during ischaemia and here we show a short CaT70 in Fig. 4e. After 20 min of reperfusion, parameters returned to values close to
4 NATURE COMMUNICATIONS | 5:3329 | DOI: 10.1038/ncomms4329 | http://www.nature.com/naturecommunications
Web End =www.nature.com/naturecommunications
& 2014 Macmillan Publishers Limited. All rights reserved.
NATURE COMMUNICATIONS | DOI: 10.1038/ncomms4329 ARTICLE
Normalized voltage/fractional fluorescence (a.u.)
Pacing spikes
3
1
2 4 5
1
4
3
2
2
5
0 2 s 0 2 s
0 ms
1) Optical Electrical
Optical Electrical
2) 3) Optical Electrical
4) Optical Electrical
5) Optical Electrical
Electrode position Optical signals
Epicardial electrograms
25 ms
0
20
15
10
Act time
OAP
EG
5
0 4 8 16
12 20 Time (ms)
50
Electrogram activation time (ms)
45
Electrode position Optical signals
Epicardial electrograms
40
10 ms
35
30
8
25
6
20
15
4
10
5
Sinus
Paced
2
0 0 10 20 30 40 50
Optical activation time (ms)
Anterior
Posterior
25 ms
20
15
10
5
0
Figure 3 | High-density electrical mapping. (a) Representative optical and electrical signals acquired simultaneously from the corresponding numbered electrode locations on a Langendorff-perfused rabbit heart. Scale bar, 7 mm. (b) Top: schematic illustration of a representative optical action potential, unipolar EG and position of the activation time, dened as dV@tmax for the optical action potential and dV@tmax for the EG. Bottom: correlation of electrical and
optical activation times for hearts tested in a variety of states. (c) Interpolated spatial activation maps determined from the electrical and optical measurements. Top: heart paced by the red pair of electrodes on the membrane. Bottom: sinus rhythm. (d) 3D mapping of electrical signalling from both the anterior and posterior surfaces of the heart. Interpolated spatial maps of electrical activation time are projected on a representative rabbit heart geometry, for purposes of visualization. Scale bar, 7 mm.
NATURE COMMUNICATIONS | 5:3329 | DOI: 10.1038/ncomms4329 | http://www.nature.com/naturecommunications
Web End =www.nature.com/naturecommunications 5
& 2014 Macmillan Publishers Limited. All rights reserved.
ARTICLE NATURE COMMUNICATIONS | DOI: 10.1038/ncomms4329
Baseline
Reperfusion-induced VT
Time (ms)
t1
*
VT
1
7.45
6.00
t3
*
Voltage
t2
0
*
0
2,000
Time (ms)
pH
1
Pressure pH Sensor pH Sensor
Voltage
0
0
0
60
100
Perfusion pressure (mmHg)
0
2,000
Time (min)
t1 (baseline)
7.5
6.0
1
0
APD70 CaT70
Vm CaT
Fluorescence
Fluorescence
Fluorescence
pH
0 300
Time (ms)
50 ms 175 ms
t2 (10 minutes ischemia)
1
7.5
6.0
APD70 CaT70
Vm CaT
pH
50 ms 175 ms
0 0 300
Time (ms)
t3 (20 minutes reperfusion)
1
0
APD70 CaT70
Vm CaT
7.5
6.0
pH
0 300
Time (ms)
50 ms 175 ms
Figure 4 | High-density pH and optical mapping of transmembrane potential and calcium transients. (a) 3D-MIM with pH sensors array integrated on a Langendorff-perfused rabbit heart with 2 pH sensors highlighted and values displayed in (b). Scale bar, 7 mm. (b) Temporal change in pH during 30 min of no-ow ischaemia followed by 30 min of reperfusion. Three times starred as t1, t2 and t3 correspond to spatial pH maps in (df). (c) Representative far-eld ECG during baseline and reperfusion-induced ventricular tachycardia (VT). (df) pH map of 32 sensors (left), representative transmembrane potential and calcium transient signals (middle), and APD70CaT70 maps (right) at baseline (d), 10 min of no-ow ischaemia (e) and 20 min of reperfusion (f). The results provide maps of the anteriorposterior LV. The Vm and CaTare taken from the same pixel, but the location may vary slightly due to shrinkage of the heart during no-ow ischaemia. White circles denote pH sensors positions. VT, ventricular tachycardia; Vm, transmembrane potential;
CaT, calcium transient; APD70, action potential duration at 70% repolarization; CaT70, calcium transient duration at 70% relaxation.
the baseline levels. This experiment demonstrates possibilities in multiparametric mapping during ischaemia/reperfusion. The information establishes anatomical relationships between metabolism and excitationcontraction coupling.
A 3D-MIM with arrays of temperature sensors illustrates capabilities in monitoring spatial distributions of cardiac temperature. The temperature sensor elements use designs established previously, consisting of serpentine traces of gold
6 NATURE COMMUNICATIONS | 5:3329 | DOI: 10.1038/ncomms4329 | http://www.nature.com/naturecommunications
Web End =www.nature.com/naturecommunications
& 2014 Macmillan Publishers Limited. All rights reserved.
NATURE COMMUNICATIONS | DOI: 10.1038/ncomms4329 ARTICLE
(20 mm wide, 50 nm thick) (Supplementary Fig. 8) in which changes in resistance correlate to changes in temperature25. The temperature sensors exhibit linear responses over physiological range, with a measurement precision of B23 mK when sampled at 2 Hz in typical hospital settings. Figure 5a shows a 3D-MIM
with 16 integrated temperature sensors during use on a beating heart. The sensors are calibrated in temperature-controlled water bath before the animal experiments, exhibiting average responses of 1.23 O C 1 with s.d. of 0.05 O C 1 over 16 sensors across the array (Supplementary Fig. 9). In one experiment, the
30.0
Baseline
(122 b.p.m.)
40
40.0 C
Temperature (C)
38
36
34
Cold
(71 b.p.m.)
35.0
32
Warm
(116 b.p.m.)
30
0 200 400 600 800 1,000
Temperature sensors array Time (s)
58.0C
Sinus rhythm Pacing VF
0.5%
47.0
0.5%
36.0
65
Temperature (C)
60
55
50
45
40
Cautery pen
35 0 5 10 15
0
0.5 1.0 1.5
Time (s)
0.5%
2.0
20
Time (s)
(R-R0)/R ECG
Normalized fractional fluorescence
3D-MIM
3
External
3
Normalized fractional fluorescence
2
2
Voltage (V)
Voltage (V)
1
1
1
1
0
0
0
0
0 200 400 0 200 400
Time (ms) Time (ms)Pacing spike Optical action potential
Figure 5 | Temperature and strain sensing combined with imaging enabled by integrated l-ILEDs. (a) Application of a 3D-MIM for temperature monitoring during cold perfusion. Left: image of a 3D-MIM with 4 4 temperature sensors array integrated on a Langendorff-perfused rabbit heart. Middle:
temperature recordings from a representative sensor illustrated in the left inset. Right: temperature maps at representative time points in the middle inset with corresponding heart rate calculated from ECG. Each pixel in the colour map corresponds to recording from one temperature sensor. Scale bar, 1 cm. (b) Temperature measurements during an ablation experiment. Positions of the sensor array and cautery pen are shown in the left inset. Temperature map during ablation (upper right) and recordings from representative sensors (bottom right) are shown respectively. Scale bar, 7 mm. (c) Responsesof a Si strain sensor under representative physiological conditions, compared with simultaneous ECG recordings. (d) Left: image of a 3D-MIM with m-ILEDs array in optical mapping experiments. Inset shows a magnied view of area around a representative m-ILED. Right: comparison of optical signalsfrom a representative pixel (blue dot on the left inset) recorded during excitation using m-ILEDs on 3D-MIM and external optical excitation, respectively.
Scale bar, 3 mm.
NATURE COMMUNICATIONS | 5:3329 | DOI: 10.1038/ncomms4329 | http://www.nature.com/naturecommunications
Web End =www.nature.com/naturecommunications 7
& 2014 Macmillan Publishers Limited. All rights reserved.
ARTICLE NATURE COMMUNICATIONS | DOI: 10.1038/ncomms4329
temperature of the heart was changed by altering the temperature of the perfusion. As shown in Fig. 5a, the measured epicardial temperature gradually decreased by B7C during cooling of the perfusate, with a uniform distribution of temperature across the heart. The heart rate, determined from the far-eld EG, decreased with decreasing temperature and recovered to the original value as the temperature returned to physiological levels, indicating that temperature controlled the rate of myocardial metabolism. In a second experiment, a cautery pen was used to acutely burn a small region of the epicardium, simulating clinical ablation. The associated temperature map (Fig. 5b) shows localized elevation of temperature near the point of ablation. Such information can be used as feedback for clinical control of ablation time and size of affected area. In combination with electrical sensors, such a device could provide the real-time relation between temperature and excitation.
In addition to electrical and chemical evaluation, mechanical characteristics can be determined. Here strain sensors based on piezoresistive effects in nanomembranes of Si allow monitoring of the mechanics of contractions of the heart during a variety of propagation states. Careful mechanical design of the serpentine interconnect structures allows accurate measurement in spite of the fact that typical epicardial strains26 greatly exceed the fracture threshold of Si, as described in previously reported small-scale 2D devices15. In the present design, the 3D-MIM strain sensors include three p-doped Si piezoresistors in a rosette conguration (Supplementary Fig. 10). Two of the piezoresistors, with longitudinal axes perpendicular to each other, are aligned to the /110S crystalline directions of the Si, offering an effective longitudinal gauge factor of B0.33 and effective transverse gauge factor of nearly 0.06 for each piezoresistor (Supplementary
Fig. 11). The piezoresistor aligned to the /100S crystalline direction exhibits relatively small change in resistance under strain, due to the intrinsic sensitivity associated with the crystalline direction as well as the overall device geometry. The piezoresistors aligned to the /110S directions provide maximum sensitivity for characterization of mechanical rhythms of the heart while the piezoresistor aligned to the /100S direction can be used to calibrate for effects of temperature. Experiments revealed the mechanical behaviours during sinus rhythm, ventricular pacing and pharmacologically induced ventricular brillation (VF) with Pinacidil (30 mM bulk dose). The use of Pinacidil signicantly reduces the action potential duration and subsequently increases the vulnerability to reentrant arrhythmias when stimulated with 50 Hz A/C burst pacing. Bath electrodes simultaneously recorded a far-eld ECG to establish the temporal correlation between the electrical and mechanical behaviour. Figure 5c shows the response of a representative piezoresistor aligned to the /110S direction. The measurements reveal mechanical rhythms of the cardiac cycles, with consistent cycling with ECG recordings. During VF condition, both the strain gauges and ECG show that the waveform lost normal rhythm and displayed a random pattern typical for VF.
A nal demonstration exploits arrays of m-ILEDs27,28 to illustrate the capacity for advanced semiconductor integration and optical mapping/stimulation. Here nine ultrathin (3 mm), microscale (300 300 mm2) LEDs based on aluminium indium
gallium phosphide (AlInGaP) with peak emission wavelengths of 670 nm (Supplementary Figs 12 and 13) served as local light sources for the excitation of voltage-sensitive dyes. Changes in uorescence associated with these dyes allowed measurement of the cardiac action potential. Figure 5d compares signals obtained with an external light source (Prizmatix, 630 nm) and with the integrated m-ILEDs. In spite of their small sizes, the LEDs enable recording of clear action potentials, with waveform shapes consistent with external light. The signal-to-noise ratio of the
m-ILED-excited action potentials is lower than the externally excited action potentials due to a necessary decrease in light intensity to minimize the power delivered to the device. The results demonstrate the future possibility of an in vivo optical mapping using either externally applied dyes or internal uorescent indicators and/or stimulation system in a 3D integration format.
DiscussionThe results presented here suggest routes for integrating active electronic materials and sensors in 3D, organ-specic designs, with potential utility in both biomedical research and clinical applications. With attention to materials, engineering mechanics and functional devices, these systems can establish conformal interfaces with the epicardium, and perform a variety of high-density physiological multiparametric mapping and stimulation. The use of transfer printing and the reported scheme for integration onto the printed 3D heart structure allows diverse sensor/actuator devices on a single platform. Separate electrical connection, with a single trigger channel to synchronize the timing, eliminates effects of crosstalk. The devices can provide local information on the metabolic, excitable, ionic, contractile and thermal state for investigations of both the spatial and temporal responses to a variety of insults, diseases and therapies. The devices could be used to identify critical regions that indicate the origin of pathophysiological conditions such as arrhythmias, ischaemia or heart failure. These regions could then be used to guide therapeutic interventions. The techniques in microfabrication, transfer printing and 3D shape denition are scalable to larger sizes and smaller, denser arrays of sensors. To increase the resolution and numbers of sensors, it may be necessary to incorporate transistors into the device to allow multiplexed addressing. Remaining challenges for use as a chronic implant include means for power supply, control/communication and encapsulation. These approaches present a promising opportunity to design and implement high-denition implantable devices for diagnostics and therapy of lethal heart diseases.
Methods
Fabrication of 3D-MIMs. The process, detailed in Supplementary Methods, starts with standard planar processing of inorganic semiconductor materials (Si, InGaN or AlInGaP) followed by transfer printing onto substrates coated either with a bilayer of polyimide (PI) on poly(methyl methacrylate) (PMMA) or poly(ethylene terephthalate) (PET) on poly(dimethylsiloxane) (PDMS)15,2730. Dissolution of the PMMA or delamination from the PDMS allows release of the devices. Metal layers (Cr/Au) are vacuum-deposited and patterned to form interconnects, resistors and electrodes. Application and patterning of a polymer encapsulation layer (PI or a photosensitive epoxy, SU8) on top of the devices completes their fabrication. Transfer printing delivers the resulting structures to a thin lm of a low-modulus silicone elastomer (Ecoex, Smooth-on). A exible conductive cable (Elform) bonded to contact pads at the periphery provides an interface to external hardware. A lamination process attaches the devices to a desired 3D-printed model of the heart, with the sensors in direct contact with the surface. Casting and curing another layer of the same type of elastomer denes the overall 3D geometry of the system (Supplemental Fig. 16). In some cases, elastomer straps enhanced the manoeuvrability for use in Langendorff-perfused heart experiments (Supplementary Fig. 14). Additional openings in the membrane can be included to prevent uid build-up associated with ow in the supporting bath. Removal from the model allows electrodeposition of IrOx on to yield pristine surfaces for precision pH sensing.
Mechanical analysis. Numerical analysis by 3D FEM: The 3D FEM is used to study the pressure between the 3D-MIM and the heart for a wide range of device parameters and the expansion of the heart. The 3D geometric model of the heart is reconstructed from the data obtained with optical segmentation. The geometric model is imported into the pre-processor in the ABAQUS nite element program31. The heart and the 3D-MIM are modelled by the four-node, linear tetrahedron solid element C3D4 and the four-node quadrilateral membrane element M3D4 in ABAQUS, respectively. The total number of elements exceeds 60,000, and mesh renement ensures the accuracy of the numerical results. For the prescribed expansion of the heart, FEM gives the pressure distribution at the
8 NATURE COMMUNICATIONS | 5:3329 | DOI: 10.1038/ncomms4329 | http://www.nature.com/naturecommunications
Web End =www.nature.com/naturecommunications
& 2014 Macmillan Publishers Limited. All rights reserved.
NATURE COMMUNICATIONS | DOI: 10.1038/ncomms4329 ARTICLE
interface between the 3D-MIM and the heart. The average pressure is then obtained over the contact area between the 3D-MIM and the heart, that is, the ventricles of the heart as in the experiment, as shown in Fig. 2a.
Analytical calculations: The part of the heart that is covered by the 3D-MIM (Fig. 2a) is approximately a partial axisymmetric ellipsoid with the lengths a and b of semi-principal axes, as shown in Supplementary Fig. 1. The 3D-MIM on the heart surface is modelled as a membrane, which deforms from Z2/a2 R2/b2 1
when being fabricated on the model to z2/((1 e)2a2) r2/((1 e)2b2) 1 due to
the linear expansion e of the heart, where (Z, R) and (z,r) ((1 e)Z, (1 e)R) are
axial and radial coordinates of the 3D-MIM without and with expansion in the axisymmatric coordinates, respectively. The plane stress state of the 3D-MIM with biaxial linear strain e gives the biaxial stress as Ee/(1 v), where E and n are the
Youngs modulus and the Poissons ratio of the 3D-MIM, respectively. The membrane force, accounting for the change of length due to linear expansion, is
T
Ete 1 n
1 e
; 2
where t is the thickness of the 3D-MIM. For a planar curve r r(z), the principal
curvature along the meridional direction is (dr2/dz2)/(1 (dr/dz)2)3/2 at any
point (z, r) on the surface. The other principal curvature along the circumferential
direction is given by r
1 dr=dz
q
2
. For z (1 e)Z and r (1 e)R, the two
principal curvatures are given by
k1
(
3
a b 1 e
a a X b X
k2
a
1 e
b
p a a X b X
:
The pressure on the heart is obtained in terms of the membrane tension and curvatures as in ref. 32.
P T k1 k2
4 Its average over the part of Z0rZra (Supplementary Fig. 1a) of the ellipsoid surface that is in contact with the heart is given by
Paverage
R
aZ P 2pR
References
1. Harrison, L. et al. The sock electrode arraya tool for determining global epicardial activation during unstable arrhythmias. Pacing Clin. Electrophysiol. 3, 531540 (1980).
2. Worley, S. J. et al. A new sock electrode for recording epicardial activation from the human heart - one size ts all. Pacing Clin. Electrophysiol. 10, 2131 (1987).
3. Farid, T. A. et al. Role of K-ATP channels in the maintenance of ventricular brillation in cardiomyopathic human hearts. Circ. Res. 109, 13091318 (2011).
4. Faris, O. P. et al. Novel technique for cardiac electromechanical mapping with magnetic resonance imaging tagging and an epicardial electrode sock. Ann. Biomed. Eng. 31, 430440 (2003).
5. Emov, I. R., Nikolski, V. P. & Salama, G. Optical imaging of the heart. Circ. Res. 95, 2133 (2004).
6. Rieke, V. & Pauly, K. B. MR thermometry. J. Magn. Reson. Imaging 27, 376390 (2008).
7. Chan, K. W. Y. et al. MRI-detectable pH nanosensors incorporated into hydrogels for in vivo sensing of transplanted-cell viability. Nat. Mater. 12, 268275 (2013).
8. Dhooge, J. et al. Regional strain and strain rate measurements by cardiac ultrasound: principles, implementation and limitations. Eur. J. Echocardiogr. 1, 154170 (2000).
9. Sekitani, T. et al. A rubberlike stretchable active matrix using elastic conductors. Science 321, 14681472 (2008).
10. Kaltenbrunner, M. et al. An ultra-lightweight design for imperceptible plastic electronics. Nature 499, 458463 (2013).
11. Lipomi, D. J. et al. Skin-like pressure and strain sensors based on transparent elastic lms of carbon nanotubes. Nat. Nanotechnol. 6, 788792 (2011).
12. Yao, H. F., Shum, A. J., Cowan, M., Lahdesmaki, I. & Parviz, B. A. A contact lens with embedded sensor for monitoring tear glucose level. Biosens. Bioelectron. 26, 32903296 (2011).
13. Kim, D.-H. et al. Epidermal electronics. Science 333, 838843 (2011).14. Kim, D.-H. et al. Materials for multifunctional balloon catheters with capabilities in cardiac electrophysiological mapping and ablation therapy. Nat. Mater. 10, 316323 (2011).
15. Kim, D. H. et al. Electronic sensor and actuator webs for large-area complex geometry cardiac mapping and therapy. Proc. Natl Acad. Sci. USA 109, 1991019915 (2012).
16. Lou, Q., Li, W. W. & Emov, I. R. The role of dynamic instability and wavelength in arrhythmia maintenance as revealed by panoramic imaging with blebbistatin vs. 2,3-butanedione monoxime. Am. J. Physiol. Heart Circ. Physiol. 302, H262H269 (2012).
17. Holt, J. P., Rhode, E. A. & Kines, H. Pericardial and ventricular pressure. Circ. Res. 8, 11711181 (1960).
18. Hancock, E. W. Subacute effusive-constrictive pericarditis. Circulation 43, 183 (1971).
19. Smiseth, O. A., Frais, M. A., Kingma, I., Smith, E. R. & Tyberg, J. V. Assessment of pericardial constraint in dogs. Circulation 71, 158164 (1985).
20. Tyberg, J. V. et al. The relationship between pericardial pressure and right atrial pressure - an intraoperative study. Circulation 73, 428432 (1986).
q
1 dRdZ
q
2
dZ
R
aZ 2pR
1 dRdZ
2
dZ
: 5
This gives Equation (1) in the main text, and
C
R
a Z
a a b
a a b
Z
b a a Z b Z
R
q
1 dRdZ
2
dZ
R
aZ R
q
1 dRdZ
2
dZ
6
For a 15 mm, b 10 mm and X0 5.5 mm, which best t the geometric model
of the heart, the average pressure in Equation (1) agrees well with the 3D FEM results, as shown in Fig. 2b.
The analysis above does not account for the effect of electronic devices on the pressure between the 3D-MIM and heart. Such an effect can be estimated from Equation (1) by replacing the tensile stiffness Et of the 3D-MIM with the effective tensile stiffness of the 3D-MIM with the electronic devices. The inset in Supplementary Fig. 1b shows an electronic device on a sheet of the membrane material (62.8 24.3 0.15 mm3), consisting of the interconnects and electrodes.
All of the interconnects consist of the Au: 120 nm/Cr: 2 nm composite layer sandwiched by 1.2-mm-thick P) layers on each side. The cross section of the electrodes is similar to that of interconnects but without the top 1.2-mm-thick PI layer to expose Au. The sheet is modelled by the eight-node solid element C3D8R, and interconnects and electrodes are modelled by the four-node shell element S4R in ABAQUS, respectively. FEM gives its tensile stiffness to be B1.5 times that without the electronic devices, as shown in Supplementary Fig. 1b.
Animal experiments. Experiments were conducted in accordance with the ethical guidelines of the National Institutes of Health and with the approval of the Institutional Animal Care and Use Committee of Washington University inSt Louis. Briey, the heart is removed via a thoracotomy and a cannula is placed in the aorta to allow retrograde perfusion of oxygenated Tyrodes solution. The perfusion mimics the electrolyte balance within the animal and provides an energy substrate for the heart to continue to function normally from an electrical perspective. The heart is submerged in a perfusion chamber maintained at 37 C with a pH in the range of 7.357.45, controlled by oxygen ow. The optical signals of transmembrane potential (Vm) and calcium transients (CaT) rely on the collection of uorescent signals from a potentiometric dye (di-4 ANEPPS or RH-237) or calcium indicator (Rhod-2) added to the perfusate with a CMOS camera; when needed to avoid motion artifacts an excitationcontraction uncoupler (Blebbistatin) is also added to the perfusate. Supplementary Fig. 15 illustrates representative experimental settings.
Data acquisition and processing. Electrophysiology. The electrical signals are recorded from the Au electrodes on the 3D-MIMs with a 240-channel unipolar EG acquisition system (Astrocard, Boston) and a custom-built interface. Both the optical and electrical signals are collected at a sampling frequency of 1 kHz, aligned
with a trigger TTL pulse and post-processed separately in custom MATLAB software. Post-processing: the electrical signals acquired from the 3D-MIMs are rst ltered with a 60-Hz notch lter internal to the acquisition software, then the electrophysiological parameter of interest activation time is calculated (Fig. 3c) and aligned to the spatial coordinates of the electrodes based on the optical background le. The optical signals are binned, ltered and normalized as previously described33. A spatial 3 3 binning lter is applied, as well as a 100-Hz low-pass digital
lter, and the amplitude of the fractional uorescence is normalized to between 0 and 1 for each pixels temporal sequence. The electrophysiological parameters are calculated for the complete eld of view. To create the spatial maps, the activation times are interpolated using MATLABs internal function for cubic interpolation of scattered data. The optical map is also sampled at the coordinates of the electrodes and the same interpolation method is applied to compare the full-resolution optical pattern with the sampled optical map and the electrical map.
pH data. pH data were acquired using measurement of open circuit potential of each sensor reference to an Ag/AgCl electrode, using Gamry Reference 600 potentiostat. A custom-made relay system was used for multiplexing.
Temperature and strain. Data for temperature and strain sensors are acquired with measurements of resistance of each sensor using a custom-built system based on National Instruments PXI-6289 board. The resistance of each of the 16 sensors is recorded simultaneously with a 160-mA probe current, a 16-bit A/D converter and a 15-ms sampling period at each sensor.
NATURE COMMUNICATIONS | 5:3329 | DOI: 10.1038/ncomms4329 | http://www.nature.com/naturecommunications
Web End =www.nature.com/naturecommunications 9
& 2014 Macmillan Publishers Limited. All rights reserved.
ARTICLE NATURE COMMUNICATIONS | DOI: 10.1038/ncomms4329
21. deVries, G., Hamilton, D. R., Ter Keurs, H. E. D. J., Beyar, R. & Tyberg, J. V. A novel technique for measurement of pericardial pressure. Am. J. Physiol. Heart Circ. Physiol. 280, H2815H2822 (2001).
22. Shabetai, R. Pericardial effusion: haemodynamic spectrum. Heart 90, 255256 (2004).
23. Neely, J. R., Lieberme., H., Battersb., E. j. & Morgan, H. E. Effect of pressure development on oxygen consumption by isolated rat heart. Am. J. Physiol. 212, 804814 (1967).
24. Chung, H.-J. et al. Stretchable, multiplexed pH sensors with demonstrations on rabbit and human hearts undergoing ischemia. Adv. Healthc. Mater. 3, 5968.
25. Webb, R. C. et al. Ultrathin conformal devices for precise and continuous thermal characterization of human skin. Nat. Mater. 12, 938944 (2013).
26. Moore, C. C., Lugo-Olivieri, C. H., McVeigh, E. R. & Zerhouni, E. A. Three-dimensional systolic strain patterns in the normal human left ventricle: characterization with tagged MR imaging. Radiology 214, 453466 (2000).
27. Kim, R. H. et al. Waterproof AlInGaP optoelectronics on stretchable substrates with applications in biomedicine and robotics. Nat. Mater. 9, 929937 (2010).
28. Kim, R. H. et al. Flexible vertical light emitting diodes. Small 8, 31233128 (2012).29. Ahn, J. H. et al. Heterogeneous three-dimensional electronics by use of printed semiconductor nanomaterials. Science 314, 17541757 (2006).
30. Kim, H. S. et al. Unusual strategies for using indium gallium nitride grown on silicon (111) for solid-state lighting. Proc. Natl Acad. Sci. USA 108, 1007210077 (2011).
31. Dassault Systmes. Abaqus analysis users manual v.6.10 (Dassault Systmes Simulia Corp., 2010).
32. Timoshenko, S. & Woinowsky-Krieger, S. Theory of Plates and Shells (McGraw-Hill, 1959).
33. Laughner, J. I., Ng, F. S., Sulkin, M. S., Arthur, R. M. & Emov, I. R. Processing and analysis of cardiac optical mapping data obtained with potentiometric dyes. Am. J. Physiol. Heart Circ. Physiol 303, H753H765 (2012).
Acknowledgements
This material is based upon work supported by the NIH grants R01 HL115415, R01 HL114395 and R21 HL112278, and through the Frederick Seitz Materials Research Laboratory and Center for Microanalysis of Materials at the University of Illinois at Urbana-Champaign. L.X. acknowledges support from China Scholarship Council. R.C.W. acknowledges support from the National Science Foundation under grant no. DGE-1144245. H.C. is a Howard Hughes Medical Institute International Student Research fellow. We would like to thank Byung-Gik Kim and Hsin-Yun Chao for the help on device fabrication and data processing.
Author contributions
L.X., S.R.G., A.P.B., M.S.S., N.L., H.-J.C., I.R.E. and J.A.R. designed the experiments. L.X., A.P.B., N.L., H.-J.C., K.-I.J., M.Y., C.L., R.C.W., J.-S.K., A.A., J.-W.J., G.-T.K. and J.A.R. carried out the devices fabrication and characterization. L.X., S.R.G., A.P.B., M.S.S., N.L.,H.-J.C, K.-I.J., M.Y., C.L., R.C.W., J.-S.K., J.I.L., Y.L., J.-W.J. and I.R.E carried out the animal experiments and data analysis. L.X., Y.S., Z.L., H.C., Y.H. and J.A.R. contributed to the mechanical modelling of the system. L.X., S.R.G., Y.S., M.S.S., H.-J.C., Y.H., I.R.E. and J.A.R. co-wrote the paper.
Additional information
Supplementary Information accompanies this paper at http://www.nature.com/nature
Web End =http://www.nature.com/nature communications
Competing nancial interests: The authors declare no competing nancial interests.
Reprints and permission information is available online at http://npg.nature.com/reprintsand
Web End =http://npg.nature.com/ http://npg.nature.com/reprintsand
Web End =reprintsand permissions/
How to cite this article: Xu, L. et al. 3D multifunctional integumentary membranes for spatiotemporal cardiac measurements and stimulation across the entire epicardium. Nat. Commun. 5:3329 doi: 10.1038/ncomms4329 (2014).
10 NATURE COMMUNICATIONS | 5:3329 | DOI: 10.1038/ncomms4329 | http://www.nature.com/naturecommunications
Web End =www.nature.com/naturecommunications
& 2014 Macmillan Publishers Limited. All rights reserved.
You have requested "on-the-fly" machine translation of selected content from our databases. This functionality is provided solely for your convenience and is in no way intended to replace human translation. Show full disclaimer
Neither ProQuest nor its licensors make any representations or warranties with respect to the translations. The translations are automatically generated "AS IS" and "AS AVAILABLE" and are not retained in our systems. PROQUEST AND ITS LICENSORS SPECIFICALLY DISCLAIM ANY AND ALL EXPRESS OR IMPLIED WARRANTIES, INCLUDING WITHOUT LIMITATION, ANY WARRANTIES FOR AVAILABILITY, ACCURACY, TIMELINESS, COMPLETENESS, NON-INFRINGMENT, MERCHANTABILITY OR FITNESS FOR A PARTICULAR PURPOSE. Your use of the translations is subject to all use restrictions contained in your Electronic Products License Agreement and by using the translation functionality you agree to forgo any and all claims against ProQuest or its licensors for your use of the translation functionality and any output derived there from. Hide full disclaimer
Copyright Nature Publishing Group Feb 2014
Abstract
Means for high-density multiparametric physiological mapping and stimulation are critically important in both basic and clinical cardiology. Current conformal electronic systems are essentially 2D sheets, which cannot cover the full epicardial surface or maintain reliable contact for chronic use without sutures or adhesives. Here we create 3D elastic membranes shaped precisely to match the epicardium of the heart via the use of 3D printing, as a platform for deformable arrays of multifunctional sensors, electronic and optoelectronic components. Such integumentary devices completely envelop the heart, in a form-fitting manner, and possess inherent elasticity, providing a mechanically stable biotic/abiotic interface during normal cardiac cycles. Component examples range from actuators for electrical, thermal and optical stimulation, to sensors for pH, temperature and mechanical strain. The semiconductor materials include silicon, gallium arsenide and gallium nitride, co-integrated with metals, metal oxides and polymers, to provide these and other operational capabilities. Ex vivo physiological experiments demonstrate various functions and methodological possibilities for cardiac research and therapy.
You have requested "on-the-fly" machine translation of selected content from our databases. This functionality is provided solely for your convenience and is in no way intended to replace human translation. Show full disclaimer
Neither ProQuest nor its licensors make any representations or warranties with respect to the translations. The translations are automatically generated "AS IS" and "AS AVAILABLE" and are not retained in our systems. PROQUEST AND ITS LICENSORS SPECIFICALLY DISCLAIM ANY AND ALL EXPRESS OR IMPLIED WARRANTIES, INCLUDING WITHOUT LIMITATION, ANY WARRANTIES FOR AVAILABILITY, ACCURACY, TIMELINESS, COMPLETENESS, NON-INFRINGMENT, MERCHANTABILITY OR FITNESS FOR A PARTICULAR PURPOSE. Your use of the translations is subject to all use restrictions contained in your Electronic Products License Agreement and by using the translation functionality you agree to forgo any and all claims against ProQuest or its licensors for your use of the translation functionality and any output derived there from. Hide full disclaimer