ARTICLE
Received 22 Oct 2013 | Accepted 5 Feb 2014 | Published 4 Mar 2014
Gabriel S. Perrone1,2, Gary G. Leisk2, Tim J. Lo1, Jodie E. Moreau1, Dylan S. Haas1, Bernke J. Papenburg1, Ethan B. Golden1, Benjamin P. Partlow1, Sharon E. Fox3, Ahmed M.S. Ibrahim4, Samuel J. Lin4 &David L. Kaplan1
Metallic xation systems are currently the gold standard for fracture xation but have problems including stress shielding, palpability and temperature sensitivity. Recently, resorbable systems have gained interest because they avoid removal and may improve bone remodelling due to the lack of stress shielding. However, their use is limited to paediatric craniofacial procedures mainly due to the laborious implantation requirements. Here we prepare and characterize a new family of resorbable screws prepared from silk broin for craniofacial fracture repair. In vivo assessment in rat femurs shows the screws to be self-tapping, remain xed in the bone for 4 and 8 weeks, exhibit biocompatibility and promote bone remodelling. The silk-based devices compare favourably with current poly-lactic-co-glycolic acid xation systems, however, silk-based devices offer numerous advantages including ease of implantation, conformal t to the repair site, sterilization by autoclaving and minimal inammatory response.
DOI: 10.1038/ncomms4385
The use of silk-based devices for fracture xation
1 Department of Biomedical Engineering, Tufts University, Medford, Massachusetts 02155, USA. 2 Department of Mechanical Engineering, Tufts University, Medford, Massachusetts 02155, USA. 3 Beth Israel Deaconess Medical Center, Department of Pathology, Harvard Medical School, Boston, Massachusetts 02215, USA. 4 Divisions of Plastic Surgery and Otolaryngology, Beth Israel Deaconess Medical Center, Harvard Medical School, Boston, Massachusetts 02215, USA. Correspondence and requests for materials should be addressed to S.J.L. (email: mailto:[email protected]
Web End [email protected] ) or to D.L.K. (email: mailto:[email protected]
Web End [email protected] ).
NATURE COMMUNICATIONS | 5:3385 | DOI: 10.1038/ncomms4385 | http://www.nature.com/naturecommunications
Web End =www.nature.com/naturecommunications 1
& 2014 Macmillan Publishers Limited. All rights reserved.
ARTICLE NATURE COMMUNICATIONS | DOI: 10.1038/ncomms4385
The advent of reliable resorbable device systems has altered the landscape for conditions requiring osseous xation1. Approximately three million facial injuries occur annually,
and the American Society of Plastic Surgeons reported over 200,000 craniomaxillofacial surgical procedures in 2012, which is a 164% increase from 2000 and a 7% increase from 2011 (ref. 2). High-impact fractures have a mortality rate as high as 12%3. Currently, metal alloys are the gold standard for fracture-xation devices, despite their numerous disadvantages, which include extreme stiffness causing stress shielding of the underlying bone, potentially poor wound healing and plate exposure, infection and in some cases the need for removal of metallic implants after fracture healing is complete as in the lower extremity4,5. Stress shielding may result in a reduction of the density of bone affecting bone remodelling and healing and impacting the mixed dentition when used for paediatric mandible fractures. Metal devices used in paediatric patients may also result in implant migration inciting an inammatory response in the dura and brain tissue5. In addition, metal is incapable of degrading in vivo, may corrode in the body causing pain and inammation, can create temperature sensitivity and tactile sensation resulting in discomfort for the patient, and can produce problems with visibility59.
Recently, the use of resorbable plates and screws composed of poly-L-lactic acid, polyglycolic acid (PGA) and poly-lactic-coglycolic acid (PLGA) in fracture xation has reduced the need for hardware removal and resulted in improved bone remodelling after injury by permitting a gradual transfer of stress to the healing bone. These potential benets have been supported by reports demonstrating favourable results in treating fractures with these resorbable systems4,1015. However, conventional resorbable materials are associated with inammatory reaction, can generate sterile sinus formation, result in incomplete remodelling of the bone and exhibit improper dynamics between device strength retention and wound healing11,12,16. They are primarily limited to specic applications such as paediatric craniofacial reconstruction10, xation of metacarpal fractures1315 and long-term soft-tissue moudling17 due to their lower strength and stiffness. In addition, the fragility of conventional resorbable plates and screws requires careful pre-tapping of the screw hole and insertion while the plates need to be pre-shaped to the exact curvature of the fracture location14. Thus, new options for resorbable devices that overcome these limitations would impact surgical practice and expand the use of such devices in a broader range of xation needs.
In the present study, we assess silk-based screws in vitro and in vivo for xation devices. Silk plates were prepared from the same material and explored from a manufacturing and implantation standpoint and showed feasibility for internal xation but were not assessed in functional models. Silk is selected due to its robust mechanical features, biocompatibility and FDA-approved (Allergan Medical, Inc.) uses in medical devices. It is degradable in vivo with controllable rates from days to years and has the capability of being functionalized with bioactive compounds1824. Moreover, silk materials maintain structural stability to 4170 C, are stable at extremes of pH and can be autoclaved for sterilization, all features that enhance potential utility in new medical devices when added to the mechanical features and biocompatibility25,26. The current study proposes silk screws that are self-tapping, remain xed in the bone for 4 and 8 weeks, exhibit biocompatibility and promote bone remodelling. The properties of these new silk-based devices compare favourably with those of current PLGA resorbable xation systems, however, the advantages of the silk systems include an easier implantation technique, conformal t to the repair site, autoclave sterilization and minimal inammatory response.
ResultsManufacturing and feasibility of silk screws and plates. Using a solvent-based process (Fig. 1, see Methods), silk blanks were moulded to an average length of 2.5 cm and a diameter of 0.40 cm after drying (Fig. 2a). The blanks were then machined into screws (172 machine screws, English designation) (Fig. 2c) which was chosen as the maximum screw size that could be safely implanted into a rat femur. The major diameter of the screws used for in vitro and in vivo testing was B1.70 mm. For shear and in vitro enzymatic degradation testing, simple cylindrical pins with a diameter of B1.50 mm were made to simplify testing and analysis. Silk plates were moulded and machined into samples of various thicknesses and shapes. Once dry, the plates were malleable on hydration and could be formed in situ to t a patients unique anatomy. This is shown in Fig. 2e, where a full skull models (Sawbones, USA) orbital fracture was xed with a 4-hole 100% silk screw and plate system. Fixation was achieved by hydrating the plate with water to allow shaping, pre-drilling holes into the bone using the plate as a guide and inserting the screws without pre-tapping of the hole. While an in vivo fracture xation was not explored in this study, this preliminary trial exhibits the potential for fracture xation with silk-based devices. Screws were also implanted into rat femurs and did not require pre-tapping of the bone.
1 2 3 4
5
6
7
8
Bone plates Bone screws
Figure 1 | Processing diagram of the silk screws and plates. (1) Fill test tube with silk solution then freeze and lyophilize. (2) Use scissors or a blender to cut lyophilized silk into small pieces. (3) Combine pieces with 1,1,1,3,3,3 hexauoro-2-propanol (HFIP) in a syringe. (4) Dissolve silk in HFIP. (5) Inject dissolved silk into bone plate or screw blank moulds.(6) Place moulds in methanol for 34 days to induce b-sheet conversion.(7) Remove and allow to dry (fume hood for 1 week then 60 C oven for 5 days), then autoclave for stability. (8) Machine using a mill, lathe or die to obtain desired geometry.
2 NATURE COMMUNICATIONS | 5:3385 | DOI: 10.1038/ncomms4385 | http://www.nature.com/naturecommunications
Web End =www.nature.com/naturecommunications
& 2014 Macmillan Publishers Limited. All rights reserved.
NATURE COMMUNICATIONS | DOI: 10.1038/ncomms4385 ARTICLE
a b
c d
1,700 1,680 1,660 1,640 1,620 1,600
Wavenumber (cm1)
Figure 3 | FTIR spectroscopy of the silk material. Measurements represent the average of 64 scans taken at a resolution of 4 cm 1 (n 6).
Peaks at 1,621 cm 1 indicate b-sheet conformation while peaks at 1,640 cm 1 indicate random coil conformation. FTIR, Fourier transform infrared spectroscopy.
e f
Figure 2 | General images of the silk devices. (a) Silk blanks prior to machining (scale 6 mm). The left portion of the blank is considered
uniform and used for machining while the right portion was discarded due to bubbles and formation of weld lines. (b) SEM image of the surface of silk material after being turned on a lathe (scale 100 mm). (c) Pure silk screws
turned on a lathe and then with a 172 (English designation) thread-cutting die (scale 6 mm). The major diameter of the screws was B1.70 mm.
(d) SEM image of 172 silk screw (scale 1 mm). (e) Combination of silk
screws and plates on model skull orbital bone to show xation potential (scale 25 mm). (f) SEM image of milled surface of silk plate
(scale 1 mm).
Absorbance (a.u.)
Due to the viscosity of the high-concentration silk solutions, the moudling of silk screw blanks was challenging; the high viscosity did not permit uniform ow and settling of solution into the moulds, resulting in weld lines where different ows met (Fig. 2a, red circle). Because 1,1,1,3,3,3 hexauoro-2-propanol (HFIP) is volatile and evaporates quickly, the moulds could not be left in ambient conditions for extended periods of time to allow bubbles to migrate out of the solution. Therefore, it was necessary to place the moulds in methanol within minutes after lling, which prevented the air bubbles from escaping or the solution from completely settling. As a result, half of the blank could be used for machining while the other half was discarded due to non-cylindrical geometry or the presence of bubbles. As indicated in Fig. 2a, it is clear which part of the blank was uniform and which was not. If there was a bubble in the visually uniform portion, the blank would typically break during machining due to mechanical instability. This should not be a problem for making larger screws, as the sizes of the moulds can be increased to increase the amount of usable material for machining without added risk for non-uniformity. Using the uniform portion of the blanks, silk screws were manufactured using standard machining practices. The pure silk blanks were mounted in a lathe, turned down to the desired major diameter and threads were cut using a die.
Structure and morphology. Fourier transform infrared spectroscopy showed signicant b-sheet crystalline fractions in the screw blanks. These fractions dene the mechanical properties of silk,
thus making it important to have high b-sheet content27. As shown in Fig. 3 and Table 1, the spectra exhibited a distinct peak at 1,621 cm 1, which is characteristic of b-sheet-rich silk materials, and deconvolution showed a b-sheet content of51.21.45% (n 6) in the silk screws. Scanning electron
microscope (SEM) images showed a smooth surface nish (Fig. 2b,f) after machining with the lathe or mill, respectively, and the ability to make reproducible geometries. The screw threads (Fig. 2d), while slightly rougher than the machined surfaces, were geometrically reproducible between and within different batches of silk and the increased roughness did not exhibit any deleterious reactions in vivo. The surface roughness could be attributed to the detached silk material that was not properly removed from the screw threads or due to the inconsistency of cutting threads by hand.
Swelling properties. The swelling ratio and water uptake (%) were determined for the silk screws in vitro. The screws had a swell ratio of 0.220.03 and a water uptake of 17.91.99% (n 6) after 72 h of hydration in PBS buffer, indicating that this
silk format absorbs a signicant amount of water. There were no signicant differences in weight of the samples between 24, 48 and 72 h (P40.05) using a repeated measures ANOVA with
Tukeys multiple comparison test, but to be completely sure the values at 72 h of hydration were used as the equilibrium values. The swelling properties of the screws were also assessed in vivo after 4 weeks in the subcutaneous space of rats. The screws yielded a swell ratio of 0.240.04 and a water uptake of19.42.45% (n 12). One-tailed unpaired t-tests for both the
swell ratio (P40.09) and the water uptake (P40.11) of the silk screws in vitro versus in vivo showed there were no signicant differences between the samples, signifying that the silk material swells comparably in vitro as in vivo.
As described below, the mechanical properties decrease signicantly when hydrated. Thus, the swelling timeline was characterized to determine how quickly the material absorbs uid. This is important with regard to implantation because if the material swells too quickly, resulting in a reduction in mechanical strength, the surgeon may not be able to implant the screws. Samples (n 6) were placed in 37 C PBS for various amounts of
time (0 min, 1 min, 2 min, 4 min, 8 min, 15 min, 30 min, 60 min,
NATURE COMMUNICATIONS | 5:3385 | DOI: 10.1038/ncomms4385 | http://www.nature.com/naturecommunications
Web End =www.nature.com/naturecommunications 3
& 2014 Macmillan Publishers Limited. All rights reserved.
ARTICLE NATURE COMMUNICATIONS | DOI: 10.1038/ncomms4385
Table 1 | Quantitation of the spectra using deconvolution.
Structural conformation of silk screwsConformation Average (%) Standard deviation (%) Side chain 9.34 2.24b-sheet 51.2 1.45
Random coil 17.6 2.36 a-helix 7.75 1.71
Turns 14.1 1.91
Deconvolution of the spectra shows a b-sheet content of 51.21.45% (n 6) in the silk screws.
20
15
10
5
Table 2 | Mechanical property comparison table of relevant materials.
Mechanical properties and resorption time of relevant materials Material Maximum shear strength
(MPa)
Resorption time (months)
Solid HFIP-derived silk
90 (dry) 19 (hydrated)
NA
PLGA29,3840 100 1218 Ti-6Al-4V41 550 Never
HFIP, 1,1,1,3,3,3 hexauoro-2-propanol; PLGA, poly-lactic-co-glycolic acid.
a
Water uptake (%)
NS
0 0 min 1 min 2 min 4 min 8 min 15 min 30 min 72 h(Equilibrium)
(Equilibrium)
Time
Time
b
10
9876543210 0 min 1 min 2 min 4 min 8 min 15 min 30 min 72 h
NS
Increase in diameter (%)
Figure 4 | In vitro swelling properties. (a) Swelling timeline of the average of six samples with error bars (s.d.) of the water uptake (%) versus time (min). The silk material did not gain a signicant amount of weight until after 15 min of incubation in PBS (P40.05 using Dunnetts multiple comparison test). The blue bar indicates the water uptake (%) at equilibrium (72 h). (b) Swelling timeline of the average of six samples with error bars (s.d.) showing increase in diameter (%) versus time (min). The silk material did not change diameter signicantly until after 8 min of incubation in PBS (P40.05 using Dunnetts multiple comparison test). The blue bar indicates the increase in diameter (%) at equilibrium (72 h).
2 h, 4 h, 24 h, 48 h and 72 h) and observed for changes in weight and diameter to signify uid uptake and swelling. Figure 4a shows the water uptake (%) versus time for the rst 30 min of testing. A repeated measures ANOVA with Dunnetts multiple comparison test was performed and showed that samples did not gain a signicant amount of weight due to hydration between 0 and 15 min (P40.05). After 30 min, the difference in weight was considered signicant (Po0.05). Figure 4b shows the increase in diameter (%) versus time for the rst 30 min of testing. For changes in diameter, there were no signicant changes between 0 and 8 min (P40.05). After 15 min, the change in diameter was considered signicant (Po0.05). We conclude that at least until 8 min, the material does not absorb a signicant amount of uid even if it does come into contact with surrounding tissue or
bodily uids. This time frame is sufcient for the surgeon to implant the screw without concern for hydration and weakening or swelling of the material. Furthermore, the hydration rate will most likely be faster in the in vitro model where the material is in a 100% aqueous environment as compared with brief contact with bodily uids during a surgery.
Mechanical properties. Shear forces are one of the most prominent forces experienced in osseous-xation devices. Silk pins with a diameter of 1.50 mm, not screws, were used to determine the maximum shear stress, as the simplied geometry allows more straightforward testing and analysis. The pins were tested in shear in dry conditions. The silk pins achieved a maximum shear stress of 90.011.8 MPa (n 8). The pure silk pins were con
sistent in terms of mechanical properties during dry double lap shear testing and achieved a comparable maximum shear stress to PLGA (Table 2). PLGA screws composed of 82:18 or 80:20 PLA:PGA, commonly used in craniomaxillofacial surgeries, achieve a maximum shear stress of roughly 100 MPa28,29. Further, improvements to the silk material can be made with the addition of silk micro-bres or silk micro-particles that have been shown to enhance mechanical properties30,31. However, the limitation of the current silk format is the hydrated mechanical strength, which is signicantly decreased when compared with the dry samples. In the hydrated condition, the pins achieved a maximum shear stress of 19.30.51 MPa (n 6), which was signicantly different at the
Po0.01 level, using an unpaired t-test with Welchs correction to account for differences in standard deviations of the dry and hydrated samples.
Pull-out forces are another common load experienced by bone screws. Silk screws (172 designation, 1.70 mm major diameter) were assessed in vitro and in vivo for the pull-out strength (N) or the maximum load before failure (Fig. 5). For in vitro testing, the screws were xed in polyurethane articial bone blocks, a metallic plate was attached to the screw head and the xture was extended in the tensile direction until failure. There were two observed failure modes of the screws. First, the screw would pull completely out of the bone with minimal visible damage to the screw. Second, the screw would fracture mid-shaft leaving the bottom of the screw still xed in the bone. In dry conditions, the screws achieved a maximum load of 80.79.38N (n 6). All
six samples pulled completely out of the bone block with minimal visible damage to the screw, implying that the failure was in the polyurethane articial bone. In hydrated conditions, the screws had a maximum load of 29.05.71N (n 6). Four of the six
samples broke along the shank of the screw while two samples pulled completely out of the bone block signifying that the failure was mostly due to the silk material. For in vivo testing, screws were implanted and animals killed after 4 and 8 week time points. Samples were immediately excised with surrounding tissue, bone and screw intact and frozen until testing to minimize tissue
4 NATURE COMMUNICATIONS | 5:3385 | DOI: 10.1038/ncomms4385 | http://www.nature.com/naturecommunications
Web End =www.nature.com/naturecommunications
& 2014 Macmillan Publishers Limited. All rights reserved.
NATURE COMMUNICATIONS | DOI: 10.1038/ncomms4385 ARTICLE
40 35 30 25 20 15 10
5 0
a
b
Pull-out strength (N)
c
d
Hydrated in vitro 4 weeks in vivo 8 weeks in vivo Test condition
Figure 5 | Pull-out strength summary. The gure compares hydratedin vitro, 4 week in vivo and 8 week in vivo data showing the average of six samples with error bars (s.d). No differences between groups are considered signicant with P40.05 using one-tailed ANOVA with Tukeys multiple comparison test.
e 110 105 100
9590858075 0 7 14 21 28 35 42 49 56
Time (days)
Protease degradation PBS control
Mass remaining (%)
*
*
R2 = 0.094
R2 = 0.999 *
degradation and dehydration effects. The samples were thawed and the tissue removed immediately prior to testing to ensure a hydration state similar to in vivo conditions. After 4 weeks in vivo, the screws had a maximum pull-out load of 22.45.21N (n 6). Four of the six samples pulled completely out of the bone
while two samples broke along the shank of the screw. After 8 weeks in vivo, the screws achieved a maximum load of23.35.49N (n 6). Four of the six samples broke along the
shank of the screw while two samples pulled completely out of the bone. While not quantiable or statistically signicant, the fact that four of the six 8 week in vivo samples broke along the screw shank, as compared with two out of six 4 week in vivo samples, suggests that there may be more bone ingrowth after 8 weeks. An increased number of samples at longer time points need to be examined to fully quantify the relationship between implant residing time, bone integration and mechanical properties.
There are many factors that inuence the in vivo pull-out strength tests results, such as pilot hole size and depth of insertion that are much harder to control as compared with in vitro models. A one-tailed ANOVA with Tukeys multiple comparison test was performed and showed no signicant differences between hydrated in vitro samples, 4 week and 8 week in vivo samples, signifying that the hydrated in vitro test is a good model for pull-out strength testing. All samples were considered signicant as compared with the dry in vitro pull-out strength (Po0.05).
Enzymatic degradation. In vitro degradation was observed in the silk samples incubated in 5 U ml 1 of protease XIV at 1, 4 and 8 week time points (n 3). The samples were 1.50 mm cylindrical
geometries of similar sizes to the shanks of silk screws (1.70 mm) with the goal of correlating in vitro enzymatic degradation to in vivo resorption of the screw shank in the bone. As shown in Fig. 6b,d, the in vitro enzymatic degradation indicates degradation by surface erosion as compared with the control samples incubated in PBS (Fig. 6a,c). As shown in Fig. 6e, there was 17% mass loss after 8 weeks (56 days) of incubation. All differences between mass loss (%) of the PBS control and protease-incubated samples were considered signicant for all time points using a one-tailed unpaired t-test (Po0.05). A one-way ANOVA with
Tukeys multiple comparison test showed no signicant difference (P40.05) in mass loss (%) for 1 and 4 week samples incubated in protease but signicant differences (Po0.05)
between 1 and 8 week samples and 4 and 8 week samples. The mass loss was consistent with a second-order polynomial
(R2 0.999) (Fig. 6e). From the in vitro degradation data alone,
the projected in vitro degradation to complete mass loss of the samples is B7 months. However, this cannot be directly correlated to in vivo degradation, as more than one enzyme will play a role in the degradation in vivo and other factors such as the screw breaking down into small pieces may expose more surface area.
In vivo biocompatibility of the silk screws. A total of 28 screws were inserted unilaterally into the left hind limb of the rats. No screws failed during implantation and all screws were able to be inserted identically to current metallic devices where a pilot hole is drilled and the screw inserted without pre-tapping of the hole. During implantation, the screws came into contact with blood and surrounding soft tissue but still maintained their mechanical integrity for implantation. This result substantiates prior claims that the silk material does not swell quickly enough to create complications during surgery despite the presence of uids.
Immediately after surgery, after the effects of anaesthesia diminished, the rats were mobile on all four legs and were able to rear up on the leg containing the screw. While this could be attributed to the lingering effects of anaesthesia or the pain
*
*
Figure 6 | Representative images of the enzymatic degradation. (a) Eight-week PBS control at low magnication (scale 200 mm).
(b) Eight-week protease XIV at low magnication (scale 200 mm). (c)
Eight-week PBS control at high magnication (scale 100 mm). (d) Eight-
week protease XIV at high magnication (scale 100 mm). (e) Percent
mass remaining of silk screw material as a function of time in PBS and in5.0 U ml 1 of protease XIV with curve ts and error bars (s.d.) (n 3 for
each time point). A one-way ANOVA with Tukeys multiple comparison test showed no signicant difference (P40.05) in mass loss for 1 and 4 week samples incubated in protease but signicant differences (Po0.05)
between 1 and 8 week samples and 4 and 8 week samples. All differences between mass loss of the PBS control and protease-incubated samples were considered signicant for all time points using one-tailed unpaired t-test (Po0.05). The mass loss due to protease XIV was consistent with a second-order polynomial (y 0.0012x2 0.229x 100.05; R2 0.999).
NATURE COMMUNICATIONS | 5:3385 | DOI: 10.1038/ncomms4385 | http://www.nature.com/naturecommunications
Web End =www.nature.com/naturecommunications 5
& 2014 Macmillan Publishers Limited. All rights reserved.
ARTICLE NATURE COMMUNICATIONS | DOI: 10.1038/ncomms4385
medication, the similar ndings were observed when the rats were examined at 34 days post surgery. Throughout the 4 and 8 week periods, the rats continued to exhibit mobility on all four extremities without showing any visible signs of pain, discomfort or distress. Two animals experienced wound dehiscence but were treated with the antibiotic Baytril: the wound closed by itself and did not need further attention. Explants were isolated and characterized with photographs (Fig. 7a) to show that the screws did not back out of the bone and there were no visible negative reactions.
To conrm biocompatibility and initiation of the remodelling process, the samples were prepared in parafn blocks for histological staining with hematoxylin and eosin and Masson trichrome stains. As described previously, screws were also characterized by performing pull-out strength testing. As shown in Fig. 8, there was recruitment of multinucleated, macrophage-like cells, possibly representing osteoclasts and early resorption of the screw at the periphery, paired with early osteoblast deposition
of osteoid. Furthermore, there were multinucleated, macrophage-like cells with adjacent resorption of screw material and active osteoblasts with osteoid deposition. Although not quantiable at this point, the early resorption of the screw material at 4 and 8 weeks in vivo is consistent with the enzymatic degradation in vitro in terms of method of degradation as both appear to degrade by surface erosion.
Figures 8a and 7be are representative perspective images of a silk screw in a rat femur. The conical screw head can be seen above the cortical bone and the screw threads are xed in the bone. These images show that there are boney units in between the threads of the screw. After 4 (Fig. 7b,c) and 8 (Fig. 7d,e) weeks in vivo, it is a good sign that the screw does not simply look forced into the pilot hole since the bone mirrors the thread prole of the screw. The screw most likely cut the threads as it is unlikely to see ingrowth of bone in the cortical area after 4 weeks. It should be noted that the tissue in histological images is often folded or pushed during sectioning. It can be stated with condence that there are no large voids in the bone in these images but simply artefacts of histology.
DiscussionConventional resorbable xation devices provide benets over metallic variants; however, they present laborious implantation requirements and signicant limitations remain that preclude their use in broader clinical applications. Silk screws were implanted identically to current metallic xation procedures where a pilot hole is drilled and a self-tapping screw inserted. The silk-based screws were well tolerated and initiated the typical bone-remodelling processes. After 4 and 8 week time points in vivo, there was early resorption of the silk screw. Future material iterations may involve the incorporation of bioactive compounds such as BMP-2 to promote bone ingrowth or antibiotics to prevent infection1923,32,33. Such factors can be used to modulate healing and promote remodelling of bone. Silk plates also showed utility in implantation feasibility in vitro, as the application of water allowed malleability of the plates. Rather than having to pre-shape the plates to the exact curvature of the bone, as is the case with current resorbable devices, the plates conform to the shape of the bone and held in place after the insertion of the screws. This trait eliminates unnecessary effort by the surgeon, as the silk plates are able to actively adapt to the shape of the bone without heat treatment. Nevertheless, future studies are needed to characterize whether the hydrated silk plates maintain the required mechanical support for implantation and xation. While swelling and hydration of the material are an obvious concern, the swelling timeline shows that there should not be a concern for losing mechanical integrity during implantation, as the material did not take up a signicant amount of water or change diameter until after 8 min in vitro. Furthermore, 28 screws were inserted into rat femurs for characterization, of which none failed during implantation.
With these successful outcomes with silk plates and screws, further enhancements can be envisioned in terms of mechanical properties. In some cases, the screws and plates may also need to be tuned to degrade in a shorter time frame to allow for dynamic transfer of the load back to the healing bone. Coating or incorporating the silk devices with bioactive compounds such as BMP-2 and other pharmaceuticals could provide benets not successfully utilized in bone xation systems at present.
A limitation of the current study is the lack of a PLGA control group. Due to the manufacturing process, the silk screws had machine screw threads rather than bone screw threads and thus do not have the same geometry as PLGA screws as presented in the literature. Thus, geometrically dependent tests such as
a
b c
S
S
ST
B
ST
B
d e
S
ST S
B
ST
B
Figure 7 | Histological comparison of the silk screws. Analysis was performed after 4 (n 10) and 8 (n 6) weeks in vivo. (a) Silk screw and
bone interface after 4 weeks in vivo. Screw did not back out and remained xed in the femur (scale 3 mm). Cross-sections of in vivo silk screws
within bone at 4 weeks (b,c) and 8 weeks (d,e) (scale 800 mm). The
screws (S) retain their position at both 4 and 8 weeks following placement, and sections show continued remodelling of adjacent bone (B), with no evidence of inammatory reaction in bone or soft tissue (ST). Hematoxylin and eosin stains (b and d) and Masson trichrome stains (c and e) at 20
power.
6 NATURE COMMUNICATIONS | 5:3385 | DOI: 10.1038/ncomms4385 | http://www.nature.com/naturecommunications
Web End =www.nature.com/naturecommunications
& 2014 Macmillan Publishers Limited. All rights reserved.
NATURE COMMUNICATIONS | DOI: 10.1038/ncomms4385 ARTICLE
b
ST
S
a
S
B
ST
c
B
S B
S
d
Figure 8 | Histological characterization of the silk screws. Analysis was performed after 4 weeks in vivo (n 10). (a) Hematoxylin and eosin stain at 20 power: a cross-section of the screw (S) can be seen within surrounding bone (B) and soft tissue (ST) (scale 400 mm). Higher magnication
(scale 60 mm) (b) demonstrates recruitment of multinucleated, macrophage-like cells, possibly representing osteoclasts for early resorption of the screw
( 200), and (c) osteoblast deposition of osteoid to form early bone-modelling units ( 100). Masson trichrome stain at 400 (d) highlights the
resorption of screw material and active bone formation. No signicant accumulation of inammatory cells is present.
pull-out strength cannot be compared with literature values for PLGA screws. While we do not expect to see any major differences, the histological characterization of the in vivo implants may be altered as well with a different geometric prole interacting with the bone.
In the present study, we demonstrated the feasibility of silk as a moudling and machinable biomaterial system for craniofacial devices. Silk has been utilized in formats for biomedical applications including sponges, bres, foams, but not previously as a solid, hard, compact medical device. The silk material was easily machinable to different geometries and sizes allowing for exibility in future designs. The silk screws in this study exhibited comparative shear properties to resorbable PLGA materials and were biocompatible and remained xed in the bone up to 4 and 8 weeks. The screws were self-tapping eliminating the need for a two-step implantation process as is the case with current resorbable screws. The plates adapted to the curvature of the bone with the application of water and did not require a time-dependent heating step and pre-shaping. The versatility of silk protein will act as a benet in future studies with future material iterations.
Methods
Preparation of vacuum-dried silk. Silk was prepared using our established purication protocols34,35. Taiwanese Bombyx mori cocoons were boiled for 30 min in aqueous 0.02 M Na2CO3 (Sigma-Aldrich, USA) and then rinsed for 3 30 min in
distilled water to remove the Na2CO3 and sericin. The degummed cocoons were allowed to dry for more than 12 h and then subsequently dissolved in 9.3 M LiBr (Sigma-Aldrich, USA) solution at 60 C for 34 h. The solution was dialysed for 2 days in distilled water using Slide-a-Lyzer dialysis cassettes (MWCO 3,500, Pierce, USA). The solution was centrifuged for 2 20 min at 18,000 r.p.m. The
concentration was determined by measuring a volume of solution and the nal dried weight. The nal solution, 68 w/v%, was frozen for 25 days and vacuum dried at 20 or 0 C until complete sublimation and stored in ambient conditions
covered with tin foil to prevent any rehydration of the lyophilized solution until used in the processing steps below. The drying temperature was chosen on the basis of the speed of freeze drying, as higher temperature allowed quicker drying.
Preparation of HFIP silk solutions. The vacuum-dried silk was resolubilized in HFIP (Fluka Analytical, Sigma-Aldrich, USA) to generate solutions for preparation of the devices. Before dissolving in HFIP, the vacuum-dried silk was processed in a Waring blender so that the pieces could be packed into a syringe. For 25 w/v% silk in HFIP, 3 g of vacuum-dried silk was added to 30 ml syringes, then 9 ml of HFIP was added. The solubilized HFIP silk solutions were stored at ambient temperatures until used in the moudling steps below.
Preparation of HFIP silk screw and plate blanks for machining. Silk screw blank moulds were prepared using wax from MachinableWax.com (USA) and machined to 1.3 in (3.30 cm) height with 6 through holes of 0.30 in (0.76 cm) diameter. A thin wafer of 0.10 in (0.25 cm) was also machined to act as a base for the mould. The thin wafer was attached to the base of the mould by melting both edges of the wax and then allowing it to dry. Silk plate blank moulds were prepared using wax that was machined into a rectangular mould with inner dimensions of 1.75 1.75 1.3
in3 (4.45 4.45 3.30 cm3). Once the screw and plate moulds were lled with
HFIP silk solution, they were placed in methanol (Fisher Scientic, USA) for 34 days to induce the b-sheet conversion and also to displace the HFIP. Subsequently, a methanol-to-water gradient was used and water was added at 4 1 h intervals to
slowly transition the solution from 100% methanol and HFIP to 100% water. Post-gradient water changes were used to remove excess HFIP. The materials were then dried in a fume hood for at least 1 week, baked in a 60 C oven for at least 5 days and then autoclaved prior to machining for stability.
Machining of silk screws and plates. The silk blanks were machined using a MicroLux True-Inch 7 16 Variable Speed Lathe (Micromark, USA). For the
screws, the blanks were cut to the outside diameter and head diameter of the desired screw size. While the screw remained attached to the blank on the lathe, a 172 NF Die 13/16 OD (McMaster-Carr, USA) was used to machine the threads into the screw shank. The screw was cutoff behind the head. A diamond cutter was mounted to the lathe and a slot was cut in the screw heads. For the plates, the diamond cutter or sand paper was used to reduce the oversized blank into desirable plate thicknesses and a razor blade was used to cut the plates into various geometries. Alternatively, machine tools including a mill and band saw were used to cut the plates to the desired geometries. The silk bone screw/plate processing is depicted in Fig. 1.
Structural and morphological analysis. Conformation of the silk screws were analysed with a JASCO FTIR 6200 spectrometer (Jasco, USA) using a MIRacle-attenuated total reection accessory with Ge crystal (Pike Technologies, USA) and analysis was performed as previously described36. Fourier self-deconvolution of the
NATURE COMMUNICATIONS | 5:3385 | DOI: 10.1038/ncomms4385 | http://www.nature.com/naturecommunications
Web End =www.nature.com/naturecommunications 7
& 2014 Macmillan Publishers Limited. All rights reserved.
ARTICLE NATURE COMMUNICATIONS | DOI: 10.1038/ncomms4385
amide I region was performed using OPUS 5.0 software (Bruker Optics, USA) with a half bandwidth of 27 cm 1 and a noise reduction factor of 0.3. For surface morphological analysis, samples were sputter coated for 90120 s and then imaged using a ZEISS EVO MA10 SEM (Carl Zeiss, Germany).
Mechanical properties. The double lap shear test consisted of three consecutive stainless steel plates with an aligned 1.50 mm hole for dry shear testing and a1.65 mm hole for hydrated shear testing. The 1.50 mm silk pins were produced to t tightly through the xture. The xture was mounted into an Instron 3366 (Instron, USA) test frame. The bottom xture remained stationary while the top xture was compressed at a rate of 5.0 mm min 1 until fracture of the silk pin28.
The maximum stress (MPa) was reported. Tests were performed in dry and hydrated environments by immersion in Dulbeccos PBS (Sigma-Aldrich, USA) solution at 37 C for greater than 72 h. The tests were then performed in a 37 C bath of PBS.
Pull-out strength consisted of an articial bone block (Sawbones, USA) in accordance with ASTM standard F1839-08 (Standard Specication for Rigid Polyurethane Foam for Use as a Standard Material for Testing Orthopaedic Devices and Instruments) to x the threads and a stainless steel screw head holder. The articial bone block was drilled with a #53 drill bit (1.5 mm diameter) but was not tapped. The 172 machine style silk screws were generated, placed in the head holder and then screwed into the articial bone block. The head holder was then fastened to an Instron 3366 test frame and extended in the tensile direction at a rate of 5.0 mm min 1 (ASTM standard F2502-11). Tests were performed in dry and hydrated environments by immersion in PBS solution at 37 C for greater than 72 h. The tests were then performed in a 37 C bath of PBS.
In vitro enzymatic degradation. The degradation of the silk screws was studied using protease XIV (Sigma-Aldrich, USA) with an activity of 5 U ml 1 following our published procedures32,37. Silk plugs (B0.0065 g, B1.50 mm major diameter, B3.00 mm overall length) were placed in well plates with 2 ml of PBS (pH 7.4) containing protease and PBS alone for control at 37 C. The enzyme solution was replaced every 2448 h with freshly prepared solution. Samples were killed at 1, 4 and 8 weeks, and dried, weighed and imaged using an SEM to determine degradation.
In vitro swelling properties. For in vitro testing, the silk material was machined, weighed to determine the dry weight of the sample (Wd) and then immersed in 37 C PBS for various amounts of time (0 min, 1 min, 2 min, 4 min, 8 min, 15 min, 30 min, 60 min, 2 h, 4 h, 24 h, 48 h and 72 h) and observed for changes in weight and diameter to signify uid uptake and swelling (n 6 per time point). Surface
moisture was removed from the sample by wiping with a Kimwipe (Kimberly-Clark, USA), the wet weight of the sample taken (Ws) and the major diameter was measured. The swelling ratio and water uptake (%) were calculated.
Swelling ratio Ws Wd=Wd 1
Water uptake % Ws Wd=Ws 100: 2
In vivo swelling properties. Animal studies were conducted under approved protocol #M2010-121 and amendments were approved on June 18, 2012 (#M2010-121A-1) at Tufts University. For the studies, 280 g, 20-week-old female Sprague Dawley Rats (Charles River Laboratories International, Inc., USA) were anaesthetized with isourane at 4% concentration in oxygen for induction and 2.53% for maintenance. All samples were sterilized by autoclaving prior to implantation with a dry cycle at 25 min of steam cycle and 15 min of drying. Silk screws were implanted bilaterally, subcutaneously on the back of the rats. Rats were killed at 4 weeks (n 12). The screws were removed, cleaned of excess tissue and weighed
(Ws). The screws were then allowed to dry in excess of 2 weeks and then weighed (Wd). As described previously, the swelling ratio and water uptake (%) were calculated based on the dry weight (Wd) and wet weight (Ws).
In vivo characterization and implantation technique. All samples were sterilized by autoclaving prior to implantation with a dry cycle at 25 min of steam cycle and 15 min of drying. Animal studies were conducted as described in the above section, In vivo swelling properties. The incision area was shaved and disinfected with a povidone-iodine scrub. A 510 mm longitudinal incision was made laterally along the left hind limb. The fascia was separated between the vastus lateralis and biceps femoris and followed the length of the femur to expose the distal end. A hole with a diameter of 1.5 mm and length of 3 mm was drilled through one of the cortices with a Synthes Electric Pen Drive (Synthes, USA) and drill bit with drill stop. The screw was inserted into the femur manually with a screw driver to rotate the screw and forceps to hold the screw in the desired location. The screws were 172 machine screw designations with a diameter of 1.70 mm. The incision was closed using a simple interrupted pattern with 5-0 Vicryl suture and overlaid with tissue glue. Screws were inserted unilaterally to prevent complete immobilization of the animal. Buprenorphine was administered at a dosage of 0.01 mg kg 1 pre-op and pro re nata (as needed with visible signs of distress). The animals were killed at 4
(n 10) and 8 (n 6) weeks and assessed histologically to determine host response
and signs of bone remodelling and screw resorption. There are a higher number of samples for the 4 week study due to initial pilot studies focusing on feasibility of silk screw implantation. The femur was removed and the bone sectioned B4 mm from either side of the screw head. The samples were xed in 10% formalin solution (Sigma-Aldrich, USA) immediately after retrieval, then sent to Beth Israel Deaconess Medical Center for histological preparation and characterization.
Histological preparation. The bone specimens were placed in Surgipath Decalcier (Leica, USA) at room temperature for 8 h or until the bone became exible. Following decalcication, tissue was processed with ethanol and xylene, and then embedded in parafn on an automatic processor. Parafn blocks of each specimen were then sectioned along the long axis of the bone and through the screw at 5 m intervals. Sections were baked on glass slides at 60 C and then stained on an autostainer. Sections through the center of each specimen were stained with hematoxylin and eosin (Fisher Scientic, USA) or Masson trichrome stain.
In vivo pull-out strength tests. Pull-out strength tests were performed at 4 (n 6)
and 8 (n 6) weeks. Screws were inserted identically as described above but
inserted so that there was a minimum distance between the bottom of the screw head and femur of 1 mm. This was performed by placing forceps (tip width 1
mm) in between the screw head and femur to allow for proper spacing. The spacing was necessary to ensure the head could be attached to the pull-out strength testing apparatus. After the animals were killed, the femur was dislocated from the hip and the entire leg without removing any tissue was wrapped in a damp cloth and frozen at 20 C until testing. Prior to testing, the femurs were placed in a
4 C refrigerator for 2 days and then allowed to equilibrate to room temperature for 3 h. The femur was isolated and all soft tissues were removed. Immediately after removal of soft tissue, the femur and screw were mounted to a testing block with zip ties and mounted on the Instron 3366 testing frame as described previously. The samples were tested immediately to ensure a hydration state similar to in vivo.
Statistics. All experiments were run with a minimum of n 3 for each data point.
Statistical analysis was performed using GraphPad Prism software. Statistical tests included one-way ANOVA with Tukeys multiple comparison test or Dunnetts multiple comparison test, unpaired t-test and unpaired t-test with Welchs Correction. Differences were considered signicant when Pr0.05. Regression analysis was performed using Microsoft Excel.
References
1. Fairley, J. D., Sackerer, D., Zeilhofer, H. F. & Sturtz, G. Preliminary experience with a dynamic resorbable xation device for craniosynostosis surgery.J. Craniofac. Surg. 23, e98-100 (2012).2. American Society of Plastic Surgeons. 2012 Reconstructive Plastic Surgery Statistics http://www.plasticsurgery.org/Documents/news-resources/statistics/2012-Plastic-Surgery-Statistics/Reconstructive-Surgery-Procedure-Trends-2012.pdf
Web End =http://www.plasticsurgery.org/Documents/news-resources/statistics/ http://www.plasticsurgery.org/Documents/news-resources/statistics/2012-Plastic-Surgery-Statistics/Reconstructive-Surgery-Procedure-Trends-2012.pdf
Web End =2012-Plastic-Surgery-Statistics/Reconstructive-Surgery-Procedure-Trends- http://www.plasticsurgery.org/Documents/news-resources/statistics/2012-Plastic-Surgery-Statistics/Reconstructive-Surgery-Procedure-Trends-2012.pdf
Web End =2012.pdf (2013).
3. Widell, T. Face Fracture. http://emedicine.medscape.com/article/824743-overview#showall
Web End =http://emedicine.medscape.com/article/824743- http://emedicine.medscape.com/article/824743-overview#showall
Web End =overview#showall (2012).
4. Zhang, J., Ebraheim, N., Lause, G. E., Xiao, B. & Xu, R. A comparison of absorbable screws and metallic plates in treating calcaneal fractures: a prospective randomized trial. J. Trauma Acute Care Surg. 72, E106E110 (2012).
5. Imola, M. J., Hamlar, D. D., Shao, W., Chowdhury, K. & Tatum, S. Resorbable plate xation in pediatric craniofacial surgery: long-term outcome. Arch. Facial Plast. Surg. 3, 7990 (2001).
6. Pietrzak, W. S., Sarver, D. R. & Kohn, D. H. in Mechanical Testing of Bone and the Bone-Implant Interface. (eds An, Y. H. & Draughn, R. A.) 581591 (CRC Press, 1999).
7. Larsen, M. W., Pietrzak, W. S. & DeLee, J. C. Fixation of osteochondritis dissecans lesions using poly(l-lactic acid)/ poly(glycolic acid) copolymer bioabsorbable screws. Am. J. Sports Med. 33, 6876 (2005).
8. Iizuka, T. & Lindqvist, C. Rigid internal xation of mandibular fractures. An analysis of 270 fractures treated using the AO/ASIF method. Int. J. Oral Maxillofac. Surg. 21, 6569 (1992).
9. Bos, R. R., Boering, G., Rozema, F. R. & Leenslag, J. W. Resorbable poly (L-lactide) plates and screws for the xation of zygomatic fractures. J. Oral Maxillofac. Surg. 45, 751753 (1987).
10. Eppley, B. L. Use of resorbable plates and screws in pediatric facial fractures.J. Oral Maxillofac. Surg. 63, 385391 (2005).11. Bostman, O. M. Absorbable implants for the xation of fractures. J. Bone Joint Surg. Am. 73, 148153 (1991).
12. Bozic, K. J., Perez, L. E., Wilson, D. R., Fitzgibbons, P. G. & Jupiter, J. B. Mechanical testing of bioresorbable implants for use in metacarpal fracture xation. J. Hand Surg. Am. 26, 755761 (2001).
13. Waris, E. et al. Bioabsorbable miniplating versus metallic xation for metacarpal fractures. Clin. Orthop. Relat. Res. 310319 (2003).
8 NATURE COMMUNICATIONS | 5:3385 | DOI: 10.1038/ncomms4385 | http://www.nature.com/naturecommunications
Web End =www.nature.com/naturecommunications
& 2014 Macmillan Publishers Limited. All rights reserved.
NATURE COMMUNICATIONS | DOI: 10.1038/ncomms4385 ARTICLE
14. Dumont, C. et al. Clinical results of absorbable plates for displaced metacarpal fractures. J. Hand Surg. Am. 32, 491496 (2007).
15. Lionelli, G. T. & Korentager, R. A. Biomechanical failure of metacarpal fracture resorbable plate xation. Ann. Plast. Surg. 49, 202206 (2002).16. Bostman, O. M., Laitinen, O. M., Tynninen, O., Salminen, S. T. & Pihlajamaki, H. K. Tissue restoration after resorption of polyglycolide and polylaevo-lactic acid screws. J. Bone Joint Surg. Br. 87, 15751580 (2005).
17. Varghai, D., Chim, H. & Gosain, A. K. Mechanical analysis of resorbable plates for long-term soft-tissue molding. J. Craniofac. Surg. 22, 543545 (2011).
18. Pritchard, E. M., Valentin, T., Boison, D. & Kaplan, D. L. Incorporation of proteinase inhibitors into silk-based delivery devices for enhanced control of degradation and drug release. Biomaterials 32, 909918 (2011).
19. Karageorgiou, V. et al. Bone morphogenetic protein-2 decorated silk broin lms induce osteogenic differentiation of human bone marrow stromal cells.J. Biomed. Mater. Res. A 71, 528537 (2004).20. Karageorgiou, V. et al. Porous silk broin 3-D scaffolds for delivery ofbone morphogenetic protein-2 in vitro and in vivo. J. Biomed. Mater. Res. A 78, 324334 (2006).
21. Li, C., Vepari, C., Jin, H. J., Kim, H. J. & Kaplan, D. L. Electrospun silk-BMP-2 scaffolds for bone tissue engineering. Biomaterials 27, 31153124 (2006).
22. Wang, X. et al. Growth factor gradients via microsphere delivery in biopolymer scaffolds for osteochondral tissue engineering. J. Control. Release 134, 8190 (2009).
23. Lu, S. et al. Stabilization of enzymes in silk lms. Biomacromolecules 10, 10321042 (2009).
24. Lu, Q. et al. Stabilization and release of enzymes from silk lms. Macromol. Biosci. 10, 359368 (2010).
25. Altman, G. H. et al. Silk-based biomaterials. Biomaterials 24, 401416 (2003).26. Vepari, C. & Kaplan, D. L. Silk as a Biomaterial. Prog. Polym. Sci. 32, 9911007 (2007).
27. Qin, Z. & Buehler, M. J. Cooperative deformation of hydrogen bonds in beta-strands and beta-sheet nanocrystals. Phys. Rev. E 82, 061906 (2010).
28. Pietrzak, W. S., Caminear, D. S. & Perns, S. V. Mechanical characteristics of an absorbable copolymer internal xation pin. J. Foot Ankle Surg. 41, 379388 (2002).
29. Makinen, T. J. et al. Efcacy of bioabsorbable antibiotic containing bone screw in the prevention of biomaterial-related infection due to Staphylococcus aureus. Bone 36, 292299 (2005).
30. Mandal, B. B., Grinberg, A., Gil, E. S., Panilaitis, B. & Kaplan, D. L. High-strength silk protein scaffolds for bone repair. Proc. Natl Acad. Sci. USA 109, 76997704 (2012).
31. Rajkhowa, R. et al. Reinforcing silk scaffolds with silk particles. Macromol. Biosci. 10, 599611 (2010).
32. Numata, K., Cebe, P. & Kaplan, D. L. Mechanism of enzymatic degradation of beta-sheet crystals. Biomaterials 31, 29262933 (2010).
33. Zhang, J. et al. Stabilization of vaccines and antibiotics in silk and eliminating the cold chain. Proc. Natl Acad. Sci. USA 109, 1198111986 (2012).
34. Rockwood, D. N. et al. Materials fabrication from Bombyx mori silk broin. Nat. Protoc. 6, 16121631 (2011).
35. Rockwood, D. N. et al. Ingrowth of human mesenchymal stem cells into porous silk particle reinforced silk composite scaffolds: An in vitro study. Acta Biomater. 7, 144151 (2011).
36. Hu, X., Kaplan, D. & Cebe, P. Determining beta-sheet crystallinity in brous proteins by thermal analysis and infrared spectroscopy. Macromolecules 39, 61616170 (2006).
37. Horan, R. L. et al. In vitro degradation of silk broin. Biomaterials 26, 33853393 (2005).
38. Pietrzak, W. S., Caminear, D. S. & Perns, S. V. Mechanical characteristics of an absorbable copolymer internal xation pin. J. Foot Ankle Surg. 41, 379388 (2002).
39. Tiainen, J., Soini, Y., Trmala, P., Waris, T. & Ashammakhi, N. Self-reinforced polylactide/polyglycolide 80/20 screws take more than 1 years to resorb in rabbit cranial bone. J. Biomed. Mater. Res. B Appl. Biomater. 70, 4955 (2004).
40. Landes, C., Ballon, A. & Roth, C. In-patient versus in vitro degradation of P (L/DL) LA and PLGA. J. Biomed. Mater. Res. B Appl. Biomater. 76, 403411 (2006).
41. MatWeb, L. L. C. Titanium Ti-6Al-4V (Grade 5), Annealed. http://www.matweb.com/search/DataSheet.aspx?MatGUID=a0655d261898456b958e5f825ae85390&ckck=1
Web End =http:// http://www.matweb.com/search/DataSheet.aspx?MatGUID=a0655d261898456b958e5f825ae85390&ckck=1
Web End =www.matweb.com/search/DataSheet.aspx?MatGUID=a0655d261898456b958e http://www.matweb.com/search/DataSheet.aspx?MatGUID=a0655d261898456b958e5f825ae85390&ckck=1
Web End =5f825ae85390& http://www.matweb.com/search/DataSheet.aspx?MatGUID=a0655d261898456b958e5f825ae85390&ckck=1
Web End =ckck=1 (2014).
Acknowledgements
We thank Mike Ryan, Patrick Hughes and Roberto Elia for helping with training, Scott
MacCorkle and Denis Dupuis for their expertise in the machine shop, Bruce Panilaitis for
assistance with the animal studies and Rosario Friedman, Ata Kiapour, and Carla
Haslauer for help with graphics and gures. We thank the NIH (EB002520) for support
of this project. B.J.P was supported by the Department of Defense (DoD) through the
National Defense Science & Engineering Graduate Fellowship (NDSEG) Program.
Author contributions
G.S.P., G.G.L., T.J.L., J.E.M., D.S.H., E.B.G., S.E.F. and B.P.P. performed experimental
work and data analysis. D.L.K., G.S.P., G.G.L., B.J.P. and S.J.L. planned the study project.
G.S.P., A.M.S.I., G.G.L., S.J.L. and D.L.K. wrote the manuscript.
Additional information
Competing nancial interests: The authors declare no competing nancial interests.
Reprints and permission information is available online at http://npg.nature.com/reprintsand
Web End =http://npg.nature.com/
http://npg.nature.com/reprintsand
Web End =reprintsand permissions/
How to cite this article: Perrone, G.S. et al. The use of silk-based devices for fracture
xation. Nat. Commun. 5:3385 doi: 10.1038/ncomms4385 (2014).
NATURE COMMUNICATIONS | 5:3385 | DOI: 10.1038/ncomms4385 | http://www.nature.com/naturecommunications
Web End =www.nature.com/naturecommunications 9
& 2014 Macmillan Publishers Limited. All rights reserved.
You have requested "on-the-fly" machine translation of selected content from our databases. This functionality is provided solely for your convenience and is in no way intended to replace human translation. Show full disclaimer
Neither ProQuest nor its licensors make any representations or warranties with respect to the translations. The translations are automatically generated "AS IS" and "AS AVAILABLE" and are not retained in our systems. PROQUEST AND ITS LICENSORS SPECIFICALLY DISCLAIM ANY AND ALL EXPRESS OR IMPLIED WARRANTIES, INCLUDING WITHOUT LIMITATION, ANY WARRANTIES FOR AVAILABILITY, ACCURACY, TIMELINESS, COMPLETENESS, NON-INFRINGMENT, MERCHANTABILITY OR FITNESS FOR A PARTICULAR PURPOSE. Your use of the translations is subject to all use restrictions contained in your Electronic Products License Agreement and by using the translation functionality you agree to forgo any and all claims against ProQuest or its licensors for your use of the translation functionality and any output derived there from. Hide full disclaimer
Copyright Nature Publishing Group Mar 2014
Abstract
Metallic fixation systems are currently the gold standard for fracture fixation but have problems including stress shielding, palpability and temperature sensitivity. Recently, resorbable systems have gained interest because they avoid removal and may improve bone remodelling due to the lack of stress shielding. However, their use is limited to paediatric craniofacial procedures mainly due to the laborious implantation requirements. Here we prepare and characterize a new family of resorbable screws prepared from silk fibroin for craniofacial fracture repair. In vivo assessment in rat femurs shows the screws to be self-tapping, remain fixed in the bone for 4 and 8 weeks, exhibit biocompatibility and promote bone remodelling. The silk-based devices compare favourably with current poly-lactic-co-glycolic acid fixation systems, however, silk-based devices offer numerous advantages including ease of implantation, conformal fit to the repair site, sterilization by autoclaving and minimal inflammatory response.
You have requested "on-the-fly" machine translation of selected content from our databases. This functionality is provided solely for your convenience and is in no way intended to replace human translation. Show full disclaimer
Neither ProQuest nor its licensors make any representations or warranties with respect to the translations. The translations are automatically generated "AS IS" and "AS AVAILABLE" and are not retained in our systems. PROQUEST AND ITS LICENSORS SPECIFICALLY DISCLAIM ANY AND ALL EXPRESS OR IMPLIED WARRANTIES, INCLUDING WITHOUT LIMITATION, ANY WARRANTIES FOR AVAILABILITY, ACCURACY, TIMELINESS, COMPLETENESS, NON-INFRINGMENT, MERCHANTABILITY OR FITNESS FOR A PARTICULAR PURPOSE. Your use of the translations is subject to all use restrictions contained in your Electronic Products License Agreement and by using the translation functionality you agree to forgo any and all claims against ProQuest or its licensors for your use of the translation functionality and any output derived there from. Hide full disclaimer