The aim of this study was to investigate the safety and efficacy of self-retaining barbed sutures in comparison with monofilament clip-fixated sutures for rectal wall closure in transanal endoscopic microsurgery.
Horizontal full-thickness wall defects (3.5 cm) of cattle rectal specimens were closed via transanal endoscopic microsurgery using a monofilament suture with clips at the end (Surgipro^sup ^ 2/0; Covidien, Mansfield, MA, USA, n = 25) or a self-retaining barbed suture (V-Loc(TM) 180 3/0; Covidien, Mansfield, MA, USA, n = 25). The primary endpoint was the pneumatic leakage pressure of the suture line. As a secondary endpoint, suture time was evaluated.
The median pneumatic leakage pressure for barbed sutures was 45.5 mbar (range 17-106 mbar) and 33.5 mbar (range 19-106 mbar) for monofilament sutures (p = 0.58). A pneumatic leak at a critical pressure below 25 mbar occurred in 3 cases with barbed sutures and in 7 cases with monofilament sutures (p = 0.29). Median suturing time [19:25 min:s (range 12:00-33:30) vs. 20:41 (17:00-28:33), p = 0.23] did not differ between the two groups.
Barbed sutures display the same bursting pressure as monofilament sutures and their use for rectal wall closure seems feasible.[PUBLICATION ABSTRACT]