About the Authors:
Yanguang Shen
Affiliations Department of Radiology, Beijing Anzhen Hospital, Capital Medical University—Beijing Institute of Heart Lung and Blood Vessel Diseases, Beijing, China, Department of Radiology, Hospital Affiliated to Hainan Medical College, Haikou, City of Hainan Province, China
Zhonghua Sun
Affiliation: Discipline of Medical Imaging, Department of Imaging and Applied Physics, Curtin University, Perth, Australia
Lei Xu
Affiliation: Department of Radiology, Beijing Anzhen Hospital, Capital Medical University—Beijing Institute of Heart Lung and Blood Vessel Diseases, Beijing, China
Yu Li
Affiliation: Department of Radiology, Beijing Anzhen Hospital, Capital Medical University—Beijing Institute of Heart Lung and Blood Vessel Diseases, Beijing, China
Nan Zhang
Affiliation: Department of Radiology, Beijing Anzhen Hospital, Capital Medical University—Beijing Institute of Heart Lung and Blood Vessel Diseases, Beijing, China
Zixu Yan
Affiliation: Department of Radiology, Beijing Anzhen Hospital, Capital Medical University—Beijing Institute of Heart Lung and Blood Vessel Diseases, Beijing, China
Zhanming Fan
* E-mail: [email protected]
Affiliation: Department of Radiology, Beijing Anzhen Hospital, Capital Medical University—Beijing Institute of Heart Lung and Blood Vessel Diseases, Beijing, China
Introduction
Multislice computed tomography angiography (CTA) has become the preferred method to assess aortic diseases [1–6]. However, X-ray radiation and iodine hazards are the major concern associated with CTA, as repeat CTA scans are commonly performed during pre- and postoperative assessments of aortic disease [7–9]. To address these concerns, clinical studies have recently reported the use of the second-generation dual-source computed tomography (DSCT) Flash protocol for dose reduction. This protocol uses a high-pitch acquisition mode, reducing the scan time for the entire aorta to about 2 s and with resultant very low radiation dose [10–11]. High-pitch CTA of the aorta was shown to reduce radiation exposure by 45–50% and to allow the use of less contrast medium (CM) while maintaining vessel attenuation at a diagnostic level [12]. Liu et al. reported that DSCT can provide motion artifact-free imaging of the ascending aorta at a low radiation dose compared to the conventional protocol [7]. However, that study used a CM with a high concentration of iodine (370 mg I/mL).
Reducing the iodine concentration would help to avoid contrast-induced acute kidney injury (CI-AKI) in at-risk patients because the probability of CI-AKI is mainly determined by the amount of delivered iodine [13–14]. Cademartiri et al. reported that under the same injection volume and flow rate, a CM with a low iodine concentration (hereinafter, low-iodine CM) can reduce the iodine burden to patients [15]. However, previous studies found that low-iodine CM was associated with poorer outcomes for vascular attenuation, image quality, and diagnostic accuracy [15–16]. Using a low-tube-voltage technique may help to improve contrast conspicuity in CTA [16–17], but this technique still results in degraded image quality. On the other hand, iterative reconstruction (IR) techniques can be used to reduce image noise and increase the signal- and contrast-to-noise ratios (SNR and CNR, respectively) [18–20]. Zhang et al. reported with the help of IR algorithm techniques, the head-and-neck CTA with diagnostic quality can be adequately achieved with low tube voltage (80 kVp) and low concentration contrast media (270 mg I/mL). This method could be potentially extended to include any part of the body to reduce the ionizing radiation related risks [21].
To the best of our knowledge, no previous report has studied the image quality of the aorta that is obtained by using low-iodine CM and IR during aortic CTA in the Flash Spiral scan mode. Therefore, the goal of this study was to assess the quality of images of the whole aorta obtained by CTA when using a combination of low-iodine CM and scanning with high pitch, low tube voltage, and IR techniques.
Methods
Patient population
One-hundred patients (67 males, 33 females) who were referred for noninvasive whole-aortic DSCT angiography were included in this study. Patients had a mean age of 54.3 ± 15.7 years (range: 18–88 years) and a mean body mass index (BMI) of 24.4 ± 2.6 kg/m2 (range: 17.7–29.6). Reasons for referral included suspected aortic disease (n = 20), postoperative follow-up after thoraco-abdominal vascular surgery (n = 25), endovascular aneurysm repair (n = 15), endovascular aortic dissection repair (n = 30), and follow-up examination of conservatively treated aortic aneurysm (n = 10). General exclusion criteria for contrast-enhanced CT were patients with renal insufficiency (serum creatinine > 1.5 mg/dL), BMI > 30 kg/m2, history of allergic reaction to CM, untreated hyperthyroidism, and women who were pregnant or nursing. Age, sex, height, and body weight of all patients were recorded for further analysis. The study was approved by Beijing Anzhen Hospital Ethics Committee, and written informed consent was obtained from all patients.
Patients were randomly assigned to one of two groups, according to the iodine concentration of the CM. The 100 patients was first divided into two groups with respect to the range of BMI (BMI≤25 kg/m2 and BMI between 25 kg/m2 and 30 kg/m2), and each group was further divided into subgroups (low-iodine group and high-iodine group). The low-iodine group (n = 50) received Iodixanol 270 as CM (270 mg I/mL, GE Healthcare). The high-iodine group (n = 50) received Iopamidol 370 as CM (370 mg I/mL, Shanghai Bracco Sine Pharmaceutical Co., Ltd., China).
CT image acquisition
All studies were performed on a second-generation 128-slice dual-source computed tomography system (SOMATOM Definition Flash, Siemens Healthcare, Forchheim, Germany). All scans were performed in a cranio-caudal direction with a prospective electrocardiogram (ECG)-triggered Flash protocol. Contrast-enhanced scans were performed from the thoracic inlet to the pubic symphysis. All CT imaging data were acquired while the patient held his or her breath in deep inspiration, to eliminate respiratory motion artifacts. Scanning parameters for both groups were as follows: slice collimation of 128 × 0.6 mm with a z-flying focal spot, gantry rotation time of 280 ms, pitch of 3.2, and tube voltage of 100 kV (low-iodine group) or 120 kV (high-iodine group).
CM was injected with an 18- gauge needle through the right antecubital vein by a dual-syringe power injector. A test bolus of 15 mL of CM followed by 30 mL of saline was used to evaluate the scan delay during the acquisition of a series of dynamic low-dose monitoring scans (100 kV, 20 mA for the low-iodine group; 120 kV, 20 mA for the high-iodine group) at the middle of the descending aorta. Regions of interest (ROIs) were placed within the descending aorta to calculate enhancement over time. Monitoring scans (with a temporal resolution of 1 s) began to be acquired 5 s after the start of the injection. The optimal scan delay time was calculated by adding the peak enhancement time from the monitoring scan to 10 s. The actual image was acquired by using 1 mL of CM per kg body weight, followed by 30 mL of saline solution. The injection rate of CM and saline solution was 4 mL/s for all subjects.
Image reconstruction
In both the low- and high-iodine groups, CTA images were reconstructed by a conventional filtered back projection (FBP) algorithm with a medium smooth kernel designed for cardiac imaging (B26f). In the low-iodine group, images were also reconstructed by a sinogram-affirmed IR algorithm (SAFIRE, Siemens Healthcare) with the corresponding vascular kernel (I26f). With the IR algorithm, five adjustable strength settings (strength 1–5) were available for adaptation of the noise model (SAFIRE). As recommended by manufacturer, a medium strength of 3 was used.
In both groups, transverse images were reconstructed with a slice thickness of 1 mm in 1-mm increments. Patient information was removed from all images, which were transferred to an external workstation (Syngo Multi-Modality Work Place, CT 2011A, Siemens Healthcare) for further image analysis. Using the axial data, two cardiac radiologists with more than 8 years of experience in cardiac imaging reconstructed the images by volume rendering technique, maximum intensity projection, and multiplanar reconstruction (Fig. 1).
[Figure omitted. See PDF.]
Fig 1. 2D and 3D reconstructions with images generated using low-iodine and high-iodine groups.
Volume rendering, maximum-intensity projection and multiplanar reformation images (A-F) show endovascular repair of aortic dissection (A, C, E) and aneurysm (B, D, F) with stent graft placed just below the left subclavian artery. A, C and E represent images acquired with the low-iodine protocol, while B, D and F are images generated with the high-iodine protocol. There is no difference in the visualization of stent graft and aortic branches between the two groups.
https://doi.org/10.1371/journal.pone.0117469.g001
Evaluation of objective image quality in the aorta
Axial slices were selected to ensure that the aortic enhancement values, aortic image noise, and the same paraspinal muscle noise were measured in identical ROIs in the ascending aortic root, aortic arch, descending aorta at the first lumbar (L1), and iliac artery bifurcation (Fig. 2). Furthermore image noise was also measured in the aortic main branches, including regions in the proximal segment of the brachiocephalic trunk, left common carotid artery, left subclavian artery, celiac trunk, superior mesenteric artery, renal arteries, and the distal segment of the common iliac arteries (Fig. 3). In each patient and for each structure, three ROIs, each measuring 100 mm2, were drawn on three consecutive transverse sections. When the vessel area was less than 100 mm2, the ROI was drawn to encompass the entire contrast-enhanced aortic lumen. Care was taken to avoid including vessel walls, emboli, calcified plaques, areas of stenosis, or motion artifacts in the ROIs. The mean attenuation (in Hounsfield Unit [HU]) and standard deviation (SD) in each ROI on three consecutive sections were calculated for each target structure (Figs. 2 and 3).
[Figure omitted. See PDF.]
Fig 2. Axial images of aortic segments with paraspinal muscle.
A) Lumen of the ascending aortic root, B) the aortic arch, C) the descending aorta at the first lumbar (L1), and D) common iliac artery bifurcation. Mean attenuation values with standard deviations are shown in the images.
https://doi.org/10.1371/journal.pone.0117469.g002
[Figure omitted. See PDF.]
Fig 3. Axial images of aortic branches and segments with paraspinal muscle.
A) Lumen of the Brachiocephalic trunk (2D1), the Left common carotid artery (2D2) and the Left subclavian artery (2D3), B) Celiac trunk, C) the superior mesenteric artery, D) the right renal artery (2D1) and the left renal artery (2D2), E) the right common iliac artery, F) the left common iliac artery. Mean attenuation values with standard deviations are shown in the images.
https://doi.org/10.1371/journal.pone.0117469.g003
A single reader independently calculated the SNR and CNR in the aortic images. Means and SDs of attenuation of the contrast-enhanced vessel lumen (SIaorta and SDaorta, respectively) and the paraspinal muscle tissue (at the same level of the spine in compared images; SImuscle and SDmuscle, respectively) were recorded. Each parameter was calculated three times, and the mean value was used in the SNR and CNR calculations. Noise ratios were determined by the following equations: SNR = SIaorta – SImuscle/SDaorta and CNR = SIaorta – SImuscle/SDmuscle [10–11,16].
Assessment of subjective image quality in the aorta
Subjective image quality was independently rated by two radiologists, who had 15 and 10 years of experience in CTA, respectively, and who were blinded to all patient data and to the CM, scanning protocols, and reconstruction algorithms that were used. Images were rated on a 3-point Likert scale, on the basis of contour delineation, presence of motion artifacts, and general image quality (1 = poor, 2 = moderate, and 3 = good) [10]. Grades 2 and 3 were considered diagnostic images.
Measurement of radiation and CM doses
The CT dose index volume (CTDIvol) and dose length product (DLP) were recorded during the scans. The estimated effective dose was derived from the DLP by using the equation Effective dose = DLP × k, where k is the conversion coefficient for the thoraco-abdominal region (k = 0.017 mSvmGy-1cm-1) [10, 22]. The iodine delivery rate (g I/s) was calculated as follows: Iodine delivery rate = iodine concentration (mg I/mL) × injection rate of CM (mL/s) / 1000 mg/g. The iodine weight (g I) was calculated as follows: Iodine weight = iodine concentration (g I/mL) × injection dose of CM (mL).
Evaluation of adverse effects due to CM
Patients were asked to rate their discomfort, in terms of pain at the injection site, sensations of cold or heat in the injected vein, and other discomfort, immediately and 15 to 20 min after injections. Scores were reported verbally on a scale from 1 (severe discomfort) to 10 (no discomfort at all) [23]. All adverse effects to both kinds of CM were recorded.
Statistical analysis
Statistical analysis was performed with the SPSS software package (SPSS V 19.0, Chicago, ILL). Continuous variables were expressed as mean ± SD and were analyzed by independent t-tests for normally distributed data or Mann-Whitney U tests for non-normally distributed data. Independent t-tests were performed to analyze differences between two groups regarding attenuation values, image noise, and radiation doses. The Mann-Whitney U test was used to detect differences in the subjective evaluation of image quality between the two groups. Cohen’s kappa statistic (κ) was calculated to determine the interreader agreement in the assessment of image quality, where κ > 0.81 indicated excellent, κ = 0.61–0.80 indicated good, κ = 0.41–0.60 indicated moderate, κ = 0.21–0.40 indicated fair, and κ < 0.20 indicated poor agreement. A P-value < 0.05 was considered statistically significant for all data analyses.
Results
Demographic data and scanning parameters
High-pitch CTA was successfully performed in the 100 patients with suspected aortic disorders. Patient demographics and CTA acquisition characteristics are shown in Table 1. No significant differences were found between groups for any demographic characteristics, scan time, or scan length.
[Figure omitted. See PDF.]
Table 1. Patient demographics and characteristics.
https://doi.org/10.1371/journal.pone.0117469.t001
Assessment of image quality
There were no significant differences in mean aortic attenuation, SNR, or CNR between the low- and high-iodine groups of the aorta and aortic branches. However, image noise differed significantly between the groups of the aorta and aortic branches (Tables 2, 3 and Fig. 4). In the low-iodine group, images reconstructed by the FBP and IR algorithms demonstrated significant differences in terms of image noise, SNR, and CNR (P < 0.05), but no significant difference in the mean aortic attenuation (Table 4). Of the subgroups by BMI≤25 kg/m2, image noise demonstrated significant differences (P < 0.05) in the mean aortic attenuation of ascending aorta root and aorta arch, but no significant difference in SNR, CNR and the mean aortic attenuation of descending aorta at L1and iliac artery bifurcation. No significant difference was found in the subgroups by BMI between 25 kg/m2 and 30 kg/m2 except the SNR of the ascending aorta root (Table 5). Subjective image quality scores did not differ significantly between the low- and high-iodine groups (2.92 ± 0.27 vs. 2.92 ± 0.40; P = 0.917). Interreader agreement was excellent for both groups (κ = 0.904).
[Figure omitted. See PDF.]
Table 2. Attenuation, image noise, SNR, and CNR in anatomic regions of interest of aorta.
https://doi.org/10.1371/journal.pone.0117469.t002
[Figure omitted. See PDF.]
Table 3. Attenuation, image noise, SNR, and CNR in anatomic regions of interest of aortic branches.
https://doi.org/10.1371/journal.pone.0117469.t003
[Figure omitted. See PDF.]
Fig 4. Image noise in aortic segments in the low-iodine and high-iodine groups.
Segments include the aorta root, arch, descending aorta at the first lumbar (L1), and the iliac artery bifurcation.
https://doi.org/10.1371/journal.pone.0117469.g004
[Figure omitted. See PDF.]
Table 4. Attenuation, image noise, SNR, and CNR in low-iodine group.
https://doi.org/10.1371/journal.pone.0117469.t004
[Figure omitted. See PDF.]
Table 5. Attenuation, image noise, SNR, and CNR in various regions of aorta in Low-iodine and High-iodine group in patients stratified by BMI.
https://doi.org/10.1371/journal.pone.0117469.t005
Estimation of radiation and CM doses
The CTDIvol and DLP were significantly higher in the high-iodine group compared to the low-iodine group (Table 6). A comparison of the effective radiation doses revealed that the low-iodine group received 34.3% less radiation than the high-iodine group (P < 0.001). The iodine weight and iodine delivery rate were lower in the low-iodine compared to the high-iodine group (Table 6, P < 0.001).
[Figure omitted. See PDF.]
Table 6. Radiation dose and contrast medium dose measurements.
https://doi.org/10.1371/journal.pone.0117469.t006
Adverse effects with CM injection
There were no cases of contrast extravasation in any patient. In the low-iodine group, severe, moderate, and mild sensations of heat during the scan were reported by 0 (0%), 7 (14%), and 40 (80%) patients, respectively. In the high-iodine group, these values were 20 (40%), 18 (36%), and 12 (24%) patients, respectively. Three patients (6%) in the low-iodine group experienced sensations of cold, compared to no patients in the high-iodine group.
In the low-iodine group, 5 (10%) patients experienced pain. In this group, 2 patients had moderate adverse reactions (rash and vomit), which occurred 1 min and 6–8 h, respectively, after CM injection. All adverse symptoms were resolved appropriately. No delayed adverse reactions occurred in the high-iodine group.
Discussion
Since the introduction of multidetector CT technology, CTA has become a commonly performed and routine tool to evaluate diseases of the aorta and its major branches [24]. Various techniques and patient-based strategies have focused on reducing the radiation dose that is delivered during aortic CTA. Given that the radiation dose varies as a function of the tube voltage squared, lowering the tube voltage is an important approach to reducing the radiation dose [25]. The results of this study show that high-pitch low-dose, low iodine CTA reconstructed with IR provides diagnostically adequate image quality for evaluating aortic diseases, which is consistent with findings from others using IR reconstruction algorithms [2, 26].
Lowering the tube voltage has the additional advantage of offering higher attenuation levels for iodinated CM at lower X-ray tube voltages because of a greater photoelectric effect and decreased Compton scattering [27]. The inherent attenuation of iodinated CM increases as the tube voltage approaches the K-edge of iodine (33.2 keV) [28–30]. Therefore, lowering the tube voltage may help to improve vascular attenuation when using the same concentration of CM, or when using lower concentrations without increasing the injection rate. This approach offers the possibility of improving vascular attenuation while using low-iodine CM [15–16].
It is generally accepted that improved vascular visualization can be achieved by increasing the injection rate or iodine concentration of the CM [10, 31–32]. However, the results of our study showed that aortic attenuation could be retained by using a low-iodine CM and reduced tube voltage with the same intravenous CM dose and injection rate (4 mL/s). These findings further demonstrate the potential utility of lowering the tube voltage for increasing the vascular CT values. Using low-iodine CM may reduce the iodine flux and dose. In at-risk patients, using low-iodine CM may help to prevent CI-AKI, which is the third most common cause of acute renal failure among hospitalized patients [13–14, 33–34]. Low-iodine CM can be easily delivered to patients, with less patient discomfort. In our study, the mean iodine dose (iodine weight, 20.36 ± 2.65 g in the low-iodine group) was lower than that in a previous study, indicating the feasibility of using low iodine CM in CTA scans of the aorta. Apparently, it is important to obtain high and homogenous enhancement of the arterial tree in aortic CTA and to synchronize the acquisition with the enhancement. The optimization of acquisition timing and contrast medium delivery is essential for vascular assessment and image postprocessing. Due to the inter-individual hemodynamic variability in patients undergoing aortic CTA, the appropriate scan delay may become more critical since the chances for timing errors are increasingly related to the increased speed of z-axis coverage with the use of multi-detector row CT technology. The test bolus technique provides more information on the individual hemodynamic situation of the study subject. The second-generation dual-source CT allows performance of thoraco-abdominal CTA within less than 2 s and therefore the required enhancement plateau phase can be considerably shortened. Moreover, with short data acquisition time, recirculation effects, which contributed to the prolonged enhancement plateau, can be ignored. Due to these factors, the demand on bolus geometry shifted from a long balanced to a compact bolus [35]. In this study, data acquisition began 25–30 s following injection of contrast medium in all patients, and this is based on the calculation of timing of about 15s by test bolus to reach peak enhancement, plus 10 s of further delay after initiation of the bolus injection and 1.5–2 s of scan time. We chose 25–30 seconds as the scan delay in this study on the basis of our prior experience and other results in the literature [36]. CM volume was acquired by using 1 mL of CM per kg body weight. According to reports available in the literature, adequate vascular enhancement is suggested to be equal or above 200 HU [31,35]. In our experience, the cut off level of 200 HU can be achieved in almost every patient, indicating that our results are consistent with others.
Low-tube-voltage CT scanning has some limitations, including increased image noise due to low photon flux [25] and degraded image quality due to the higher susceptibility to beam-hardening artifacts. There is also a noise penalty due to the simplicity of reconstruction by conventional FBP [37]. An IR algorithm can be used to counteract this problem, as IR minimizes the noise effects while maintaining spatial resolution and other image quality properties. Unlike FBP, the IR technique reconstructs CT datasets by fully modeling the system. The reconstruction process is iterative in nature, to overcome the mathematical complexity introduced by the added modeling [19, 37–38]. The IR technique does not assume that the measured signal is free of noise (X-ray or electronic), but rather uses more accurate statistical modeling during the reconstruction process [31, 38–40].
The results of this study showed that it is possible to achieve diagnostic image quality by using a strength level of 3 in SAFIRE with a low tube voltage (100 kVp). In the low-iodine group, using the IR algorithm reduced image noise and improved the SNR and CNR compared to the FBP algorithm. Image noise was slightly higher in the high-iodine group compared to the low-iodine group, although SNR and CNR did not differ significantly between the two groups. These findings may suggest that a SAFIRE strength level of 3 reduces the noise and increases the SNR or CNR. When combined with a low-iodine CM and low tube voltage, IR can effectively suppress image noise and retain image quality. Moreover, the increased image noise does not necessarily result in diminished subjective image quality, because the increased attenuation of the iodine-containing arterial system and the high attenuation difference between the arterial system and surrounding tissues can partially offset the higher image noise.
The major advantage of lowering the tube voltage is to reduce the radiation dose. However, high pitch and IR are also effective approaches that are increasingly used to reduce the radiation dose. Liu et al. and Bolen et al. [7, 39] showed that imaging of the thoraco-abdominal aorta with ECG-triggered high-pitch CTA provided higher quality images of the aortic root and ascending aorta, with sufficient contrast enhancement and decreased estimated radiation dose compared to standard-pitch helical CT. Our results are consistent with these findings, as both qualitative and quantitative assessments of image quality confirmed that the diagnostic images were acquired with the use of this low-dose protocol.
Studies have demonstrated that IR could increase image quality and reduce the effective radiation dose compared to FBP [16, 22–23]. Winklehner et al. reported that raw data-based IR allowed for a dose reduction of more than 50%, while maintaining the quality of body CTA images [40]. For long z-axis ECG-gated CTA of the whole aorta in patients with aortic diseases, the high radiation dose remains a concern. The results of our study show that using the combination of lower tube voltage, high pitch, and IR offers one strategy for maintaining image quality while minimizing radiation exposure during aortic CTA. This approach may be particularly useful in female patients and in patients who require follow-up for aortic aneurysm, aortic stent graft implantation, and other operations [34,39]. In our study, 70% of patients (70/100) required frequent postoperative follow-up examinations of aortic disease, thus, a low-dose CTA protocol is suitable for these patients, particularly for those treated with endovascular repair. Hansen et al in their recent study showed that low-dose CTA with model-based IR in patients undergoing endovascular aneurysm repair results in up to 73% dose reduction compared to standard CTA protocol (mean dose: 4.4 and 2.4 mSv vs 16.2 and 6.7 mSv, respectively corresponding to the arterial and delayed phases) [26]. The mean effective dose of our low-dose protocol is 4.4 mSv, which is the same as that reported in Hansen’s study as only arterial phase was involved in our scans. Koike et al. showed that low-dose dynamic volumetric CTA is feasible after endovascular aneurysm repair [6]. However, the mean effective dose of their study was 13.1 mSv, while our low-dose protocol resulted in much low dose than that study. It has been reported that excessive dependence on CT is expensive and exposes the patients to nephrotoxic contrast media and ionizing radiation [41–43], making dose reduction desirable. Thus, it is of paramount importance to implement low-dose CTA protocol in aortic imaging. Although ultrasound is increasingly used to follow-up endovascular repair of aortic aneurysm as it does not involve ionizing radiation [44–46], CTA still remains the preferred method in the current clinical practice.
Limitations
This study has several limitations. First, this study involved a relatively small number of patients (only 50 per group). Although no significant differences in objective or subjective image quality were detected, these results should be confirmed in a larger clinical cohort. Second, the diagnostic accuracy to detect aortic disease was not evaluated. The aim of the study was to assess the performance of CMs with different iodine doses in terms of image quality alone, not diagnostic accuracy. Third, only a single IR level was used in the low-iodine group. Further evaluation of different IR strength levels within each group should be performed. Lastly, we did not compare the reconstruction time between IR and conventional FBP. IR requires more time than a standard FBP reconstruction, which may influence the clinical utility.
Conclusions
High-pitch, low-kilovoltage dual-source CT angiography is feasible in patients with suspected aortic diseases. Iterative reconstruction with SAFIRE appears to complement low-iodine CTA protocol acquired using a low voltage and high pitch technique with resultant 34% reduction in radiation dose, but with significant improvements in image quality.
Author Contributions
Conceived and designed the experiments: ZF LX YL ZS. Performed the experiments: YS NZ ZY. Analyzed the data: YS. Contributed reagents/materials/analysis tools: LX YL NZ. Wrote the paper: YS ZS LX.
Citation: Shen Y, Sun Z, Xu L, Li Y, Zhang N, Yan Z, et al. (2015) High-Pitch, Low-Voltage and Low-Iodine-Concentration CT Angiography of Aorta: Assessment of Image Quality and Radiation Dose with Iterative Reconstruction. PLoS ONE 10(2): e0117469. https://doi.org/10.1371/journal.pone.0117469
1. Fujioka C, Horiguchi J, Kiguchi M, Yamamoto H, Kitagawa T (2009) Ito K: Survey of aorta and coronary arteries with prospective ECG-triggered 100-kV 64-MDCT angiography. AJR Am J Roentgenol 193:227–233. pmid:19542418
2. Buffa V, Solazzo A, D’Auria V, Del Prete A, Vallone A, et al. (2014) Dual-source dual-energy CT: dose reduction after endovascular abdominal aortic aneurysm repair. Radiol Med.Jul 2. . pmid:25520297
3. Nayeemuddin M, Pherwani AD, Asquith JR (2012) Imaging and management of complications of open surgical repair of abdominal aortic aneurysms. Clin Radiol 67: 802–814. pmid:22341185
4. Duquette AA, Jodoin PM, Bouchot O, Lalande A (2012) 3D segmentation of abdominal aorta from CT-scan and MR images. Comput Med Imaging Graph 36:294–303. pmid:22257909
5. Khan NU, Yonan N (2009) Does preoperative computed tomography reduce the risks associated with re-do cardiac surgery? Interact Cardiovasc Thorac Surg 9:119–123. pmid:19339275
6. Koike Y, Ishida K, Hase S, Kobayashi Y, Nishimura J, et al. (2014) Dynamic volumetric CT angiography for the detection and classification of endoleaks: application of cine imaging using a 320-row CT scanner with 16-cm detectors. J Vasc Interv Radiol 25:1172–1180. pmid:24837981
7. Liu Y, Xu J, Li J, Ren J, Liu H, et al. (2013) The ascending aortic image quality and the whole aortic radiation dose of high-pitch dual-source CT angiography. J Cardiothorac Surg 8:228. pmid:24330784
8. Sodickson A, Baeyens PF, Andriole KP, Prevedello LM, Nawfel RD, et al. (2009) Recurrent CT, cumulative radiation exposure, and associated radiation-induced cancer risks from CT of adults. Radiology 251:175–184. pmid:19332852
9. Katzberg RW, Barrett BJ (2007) Risk of iodinated contrast material-induced nephropathy with intravenous administration. Radiology 243(3):622–8. pmid:17446526
10. Goetti R, Baumüller S, Feuchtner G, Stolzmann P, Karlo C, et al. (2010) High-pitch dual-source CT angiography of the thoracic and abdominal aorta: is simultaneous coronary artery assessment possible? AJR Am J Roentgenol 194:938–944. pmid:20308495
11. Karlo C, Leschka S, Goetti RP, Feuchtner G, Desbiolles L, et al. (2011) High-pitch dual-source CT angiography of the aortic valve-aortic root complex without ECG-synchronization. Eur Radiol 21:205–212. pmid:20677006
12. Apfaltrer P, Hanna EL, Schoepf UJ, Spears JR, Schoenberg SO, et al. (2012) Radiation dose and image quality at high-pitch CT angiography of the aorta: intraindividual and interindividual comparisons with conventional CT angiography. AJR Am J Roentgenol 199: 1402–9. pmid:23169737
13. Worasuwannarak S, Pornratanarangsi S (2010) Prediction of contrast-induced nephropathy in diabetic patients undergoing elective cardiac catheterization or PCI: role of volume-to-creatinine clearance ratio and iodine dose-to-creatinine clearance ratio. J Med Assoc Thai 93 Suppl 1:S29–34. pmid:20364554
14. Nyman U, Almen T, Aspelin P, Hellstrom M, Kristiansson M, et al. (2005) Contrast medium-induced nephropathy correlated to the ratio between dose in gram iodine and estimated GFR in ml/min. Acta Radiol 46:830–42. pmid:16392608
15. Cademartiri F, Mollet NR, van der Lugt A, McFadden EP, Stijnen T, et al. (2005) Intravenous contrast material administration at helical 16-detector row CT coronary angiography: effect of iodine concentration on vascular attenuation. Radiology 236:661–5. pmid:16040923
16. Kim EY, Yeh DW, Choe YH, Lee WJ, Lim HK (2010) Image quality and attenuation values of multidetector CT coronary angiography using high iodine-concentration contrast material: a comparison of the use of iopromide 370 and iomeprol 400. Acta Radiol 51: 982–9. pmid:20849317
17. Qi L, Zhao Y, Zhou CS, Spearman JV, Renker M, Renker M, et al. (2014) Image quality and radiation dose of lower extremity CT angiography at 70 kVp on an integrated circuit detector dual-source computed tomography. Acta Radiol Jun 11. pii: 0284185114535391.
18. Mitsumori LM, Shuman WP, Busey JM, Kolokythas O, Koprowicz KM (2012) Adaptive statistical iterative reconstruction versus filtered back projection in the same patient: 64 channel liver CT image quality and patient radiation dose. Eur Radiol 22:138–43. pmid:21688003
19. Prakash P, Kalra MK, Kambadakone AK, Pien H, Hsieh J, et al. (2010) Reducing abdominal CT radiation dose with adaptive statistical iterative reconstruction technique. Invest Radiol 45202–10.
20. Marin D, Nelson RC, Barnhart H, Schindera ST, Ho LM, et al. (2010) Detection of pancreatic tumors, image quality, and radiation dose during the pancreatic parenchymal phase: effect of a low-tube-voltage, high-tube-current CT technique—preliminary results. Radiology 256: 450–9. pmid:20656835
21. Zhang WL, Li M, Zhang B, Geng HY, Liang YQ, et al. (2013) CT angiography of the head-and-neck vessels acquired with low tube voltage, low iodine, and iterative image reconstruction: clinical evaluation of radiation dose and image quality. PLoS One 8:e81486. pmid:24339936
22. McCollough C, Edyvean S, Cody D, Geise R, Gould B, et al. (2008) AAPM report no. 96: the measurement, reporting, and management of radiation dose in CT—report of AAPM Task Group 23 of the Diagnostic Imaging Council CT Committee. American Association of Physicists in Medicine Website: http://www.aapm.Org/pubs/reports/RPT96.pdf.
23. Rutten A, Meijs MF, de Vos AM, Seidensticker PR, Prokop M (2010) Biphasic contrast medium injection in cardiac CT: moderate versus high concentration contrast material at identical iodine flux and iodine dose. Eur Radiol 20:1917–25. pmid:20306079
24. Nakayama Y, Awai K, Funama Y, Liu D, Nakaura T, et al. (2006) Lower Tube Voltage Reduces Contrast Material and Radiation Doses on 16-MDCT Aortography, AJR Am J Roentgenol 187:W490–7. pmid:17056879
25. Rubin GD, Shiau MC, Leung AN, Kee ST, Logan LJ, et al. (2000) Aorta and iliac arteries: single versus multiple detector-row helical CT angiography. Radiology 215:670–676. pmid:10831682
26. Hansen NJ, Kaza RK, Maturen KE, Liu PS, Platt JF (2014) Evaluation of low-dose CT angiography with model-based iterative reconstruction after endovascular aneurysm repair of a thoracic or abdominal aortic aneurysm. AJR Am J Roentgenol 202:648–55. pmid:24555604
27. Rubin GD, Schmidt AJ, Logan LJ, Sofilos MC (2001) Multi-detector row CT angiography of lower extremity arterial inflow and runoff: initial experience. Radiology 221:146–158. pmid:11568333
28. Bazalova M, Carrier JF, Beaulieu L, Verhaegen F (2008) Dual-energy CT-based material extraction for tissue segmentation in Monte Carlo dose calculations. Phys Med Biol 5: 2439–56.
29. Prakash P, Kalra MK, Kambadakone AK, Pien H, Hsieh J, et al. (2010) Reducing abdominal CT radiation dose with adaptive statistical iterative reconstruction technique. Invest Radiol 45:202–10. pmid:20177389
30. Hara AK, Paden RG, Silva AC, Kujak JL, Lawder HJ, et al. (2009) Iterative reconstruction technique for reducing body radiation dose at CT: feasibility study. AJR Am J Roentgenol 193:764–71. pmid:19696291
31. Fleischmann D (2010) CT angiography: injection and acquisition technique. Radiol Clin North Am 48:237–47. pmid:20609872
32. Herman S (2004) Computed tomography contrast enhancement principles and the use of high-concentration contrast media. J Comput Assist Tomogr 28(Suppl.1):S7–11. pmid:15258487
33. Li Y, Fan Z, Xu L, Yang L, Xin H, et al. (2012) Prospective ECG-gated 320-row CT angiography of the whole aorta and coronary arteries. Eur Radiol 22:2432–40. pmid:22661055
34. Zheng M, Liu Y, Wei M, Wu Y, Zhao H, et al. (2014) Low concentration contrast medium for dual-source computed tomography coronary angiography by a combination of iterative reconstruction and low-tube-voltage technique: Feasibility study. Eur J Radiol 83:e92–e99. pmid:24332352
35. Puippe GD, Winklehner A, Hasenclever P, Plass A, Frauenfelder T, et al. (2012) Thoraco-abdominal high-pitch dual-source CT angiography: experimental evaluation of injection protocols with an anatomical human vascular phantom. Eur J Radiol 81:2592–6. pmid:22226854
36. Macari M, Israel GM, Berman P, Lisi M, Tolia AJ, et al. (2001) Infrarenal abdominal aortic aneurysms at multi-detector row CT angiography: intravascular enhancement without a timing acquisition. Radiology 220:519–23. pmid:11477263
37. Wang R, Schoepf UJ, Wu R, Reddy RP, Zhang C, et al. (2012) Image quality and radiation dose of low dose coronary CT angiography in obese patients: sinogram affirmed iterative reconstruction versus filtered back projection. Eur J Radiol 81:3141–5. pmid:22578834
38. Hou Y, Liu X, Xv S, Guo W, Guo Q (2012) Comparisons of image quality and radiation dose between iterative reconstruction and filtered back projection reconstruction algorithms in 256-MDCT coronary angiography. AJR Am J Roentgenol 199:588–94. pmid:22915398
39. Bolen MA, Popovic ZB, Tandon N, Flamm SD, Schoenhagen P, et al. (2012) Image quality, contrast enhancement, and radiation dose of ECG-triggered high-pitch CT versus non-ECG-triggered standard-pitch CT of the thoracoabdominal aorta. AJR Am J Roentgenol 198:931–8. pmid:22451563
40. Winklehner A, Karlo C, Puippe G, Schmidt B, Flohr T, et al. (2011) Raw data-based iterative reconstruction in body CTA: evaluation of radiation dose saving potential.Eur Radiol 21:2521–6. pmid:21822785
41. Michaels JA, Drury D, Thomas SM (2005) Cost effectiveness of endovascular abdominal aortic aneurysm repair. Br J Surg 92:960–967. pmid:16034841
42. Weerakkody RA, Walsh SR, Cousins C, Goldstone KE, Tang TY, et al. (2008) Radiation exposure during endovascular aneurysm repair. Br J Surg 95:699–702. pmid:18446782
43. Sun Z (2012) Evidence for contrast-enhanced ultrasound in fenestrated EVAR surveillance. J Endovasc Ther 19: 656–660. pmid:23046332
44. Sun Z (2006) Diagnostic value of color duplex ultrasonography in the follow-up of endovascular repair of abdominal aortic aneurysm. J Vasc Interv Radiol 17:759–764. pmid:16687740
45. Lu MH, Wu D, Lin WH, Li WF, Zhang HY, et al. (2014) A stochastic filtering approach to recover strain images from quasi-static ultrasound elastography. BioMedical Engineering OnLine 13:15. pmid:24521481
46. Hu ZH, Zhang HY, Yuan JW, Lu MH, Chen SP, et al. (2013) An H∞ strategy for strain estimation in ultrasound elastography using biomechanical modeling constraint. PLoS One 8(9):e73093. pmid:24058460
You have requested "on-the-fly" machine translation of selected content from our databases. This functionality is provided solely for your convenience and is in no way intended to replace human translation. Show full disclaimer
Neither ProQuest nor its licensors make any representations or warranties with respect to the translations. The translations are automatically generated "AS IS" and "AS AVAILABLE" and are not retained in our systems. PROQUEST AND ITS LICENSORS SPECIFICALLY DISCLAIM ANY AND ALL EXPRESS OR IMPLIED WARRANTIES, INCLUDING WITHOUT LIMITATION, ANY WARRANTIES FOR AVAILABILITY, ACCURACY, TIMELINESS, COMPLETENESS, NON-INFRINGMENT, MERCHANTABILITY OR FITNESS FOR A PARTICULAR PURPOSE. Your use of the translations is subject to all use restrictions contained in your Electronic Products License Agreement and by using the translation functionality you agree to forgo any and all claims against ProQuest or its licensors for your use of the translation functionality and any output derived there from. Hide full disclaimer
© 2015 Shen et al. This is an open access article distributed under the terms of the Creative Commons Attribution License: http://creativecommons.org/licenses/by/4.0/ (the “License”), which permits unrestricted use, distribution, and reproduction in any medium, provided the original author and source are credited Notwithstanding the ProQuest Terms and Conditions, you may use this content in accordance with the terms of the License.
Abstract
Objective
To assess the image quality of aorta obtained by dual-source computed tomography angiography (DSCTA), performed with high pitch, low tube voltage, and low iodine concentration contrast medium (CM) with images reconstructed using iterative reconstruction (IR).
Methods
One hundred patients randomly allocated to receive one of two types of CM underwent DSCTA with the electrocardiogram-triggered Flash protocol. In the low-iodine group, 50 patients received CM containing 270 mg I/mL and were scanned at low tube voltage (100 kVp). In the high-iodine CM group, 50 patients received CM containing 370 mg I/mL and were scanned at the tube voltage (120 kVp). The filtered back projection (FBP) algorithm was used for reconstruction in both groups. In addition, the IR algorithm was used in the low-iodine group. Image quality of the aorta was analyzed subjectively by a 3-point grading scale and objectively by measuring the CT attenuation in terms of the signal- and contrast-to-noise ratios (SNR and CNR, respectively). Radiation and CM doses were compared.
Results
The CT attenuation, subjective image quality assessment, SNR, and CNR of various aortic regions of interest did not differ significantly between two groups. In the low-iodine group, images reconstructed by FBP and IR demonstrated significant differences in image noise, SNR, and CNR (p<0.05). The low-iodine group resulted in 34.3% less radiation (4.4 ± 0.5 mSv) than the high-iodine group (6.7 ± 0.6 mSv), and 27.3% less iodine weight (20.36 ± 2.65 g) than the high-iodine group (28 ± 1.98 g). Observers exhibited excellent agreement on the aortic image quality scores (κ = 0.904).
Conclusions
CT images of aorta could be obtained within 2 s by using a DSCT Flash protocol with low tube voltage, IR, and low-iodine-concentration CM. Appropriate contrast enhancement was achieved while maintaining good image quality and decreasing the radiation and iodine doses.
You have requested "on-the-fly" machine translation of selected content from our databases. This functionality is provided solely for your convenience and is in no way intended to replace human translation. Show full disclaimer
Neither ProQuest nor its licensors make any representations or warranties with respect to the translations. The translations are automatically generated "AS IS" and "AS AVAILABLE" and are not retained in our systems. PROQUEST AND ITS LICENSORS SPECIFICALLY DISCLAIM ANY AND ALL EXPRESS OR IMPLIED WARRANTIES, INCLUDING WITHOUT LIMITATION, ANY WARRANTIES FOR AVAILABILITY, ACCURACY, TIMELINESS, COMPLETENESS, NON-INFRINGMENT, MERCHANTABILITY OR FITNESS FOR A PARTICULAR PURPOSE. Your use of the translations is subject to all use restrictions contained in your Electronic Products License Agreement and by using the translation functionality you agree to forgo any and all claims against ProQuest or its licensors for your use of the translation functionality and any output derived there from. Hide full disclaimer