Abstract
The standard treatment of cardiac arrest is cardiopulmonary resuscitation (CPR), performed with effective manual chest compressions. Although current CPR was developed 50 years ago, cardiac arrest still has a high mortality rate and manual chest compressions have some potential limitations. Because of these limitations, mechanical chest compression devices were developed to improve the efficiency of CPR. This CPR technology includes devices such as the mechanical piston load-distributing band, active compression-decompression CPR, simultaneous sterno-thoracic CPR, impedance threshold valve, phased thoracic-abdominal active compression-decompression CPR and active compression-decompression CPR with enhanced external counterpulsation, and the impedance threshold valve. The purpose of this manuscript was to draw attention to developments in this medical area and to examine studies on the effectiveness of these devices. (Eurasian J Emerg Med 2016; 15: 94-104)
Keywords: Cardiac arrest, cardiopulmonary resuscitation, mechanical chest compression devices
Introduction
Standard cardiopulmonary resuscitation (S-CPR) refers to the entire body of techniques of external chest compression and securing positive pressure ventilation for the purpose of achieving adequate blood and oxygen flow into vital organs such as the heart and brain following cardiac arrest (1). The current application of S-CPR is based on the technique of "external chest compression" that was defined by Kouwenhoven in 1960 and comprises the phases of active compression and passive decompression. Despite the evolution of resuscitation medicine, the limited improvement in survival rates following cardiac arrest has led researchers to explore the possibility of different CPR techniques and also to develop devices that support ventilation and circulation (2-6). This manuscript was prepared to review the experimental and clinical studies conducted on the historical progress and effectiveness of mechanical chest compression devices (MCCD).
Limitations of S-CPR
The fundamental goal of effective CPR applications is to achieve return of spontaneous circulation (ROSC) and a good neurological outcome and the return of the patient to the patient's previous quality of life and functional level of health. Guidelines emphasize the importance of effective chest compression for successful CPR. The effectiveness of chest compressions depends on a couple of parameters (such as application of compressions to the right place, at an adequate depth and rate, on a regular and uninterrupted basis; letting the chest to fully recoil after each compression; avoiding overventilation; and maintaining a balance between compression and ventilation) (1, 5, 7, 8).
One of the basic problems related to S-CPR techniques is that even in the most effective chest compressions, a physiologically adequate amount of cardiac output may not be reached and because the quality of compression may change over time, this may cause cerebral and coronary blood flow to reduce even further as a result of the interruptions (9-11). If the chest compressions could reach the needed depth, as it does in infants and children, a higher intrathoracic pressure and cardiac output would be possible (12, 13). Another problem is that the quality of CPR is limited to the degree of knowledge, experience, and endurance of the rescuer (9, 14-17). Transferring the patient into an ambulance, discontinuing CPR prior to defibrillation, the difficulty of effectively applying the technique in a moving ambulance, failure to maintain the relationship between compression and ventilation, and reduced elastic recoil of the chest wall in prolonged CPR applications could all have an adverse effect on the success of CPR (5, 7, 13-17).
When the importance of chest compression is considered and in the light of the issues experienced in currently employed techniques and the low survival rates following cardiac arrest, the necessity of further developing S-CPR techniques and increasing the effectiveness of chest compression becomes clear.
Mechanical Chest Compression Devices: Definition, History, and Classification
Definition
MCCDs are noninvasive circulation support devices that function manually, pneumatically, or electrically and in accordance with CPR guidelines, provide uninterrupted and effective external chest compression to achieve an adequate blood flow to the heart and other vital organs during non-traumatic adult cardiac arrest. MCCDs can be used as an alternative to S-CPR in cases that may hinder effective compressions such as prolonged CPR during the transport of the patients or in the shortage of personnel (2, 6, 10, 18, 19).
These devices are included in the guidelines of the American Heart Association (AHA) under the heading "circulatory support devices" (6). They are described by the U.S. Food and Drug Administration (FDA) as "External cardiac compressors" (19). In the literature, the nomenclature varies, and "external cardiac compression devices," "automatic chest-compression devices," and "mechanical CPR devices" are some of the terms of reference (2, 10, 13, 18).
History
The advent of MCCDs is not new. These devices began to be developed in the beginning of the 1960s, when resuscitation medicine was merely in its infancy. Chronologically speaking, the "electro- pneumatic machine" developed by Harkins and Bramson (20) in 1961; the "portable pneumatic pump" developed by Nachlas and Siedband (21) in 1962; the "Beck-Rand external cardiac compression machine" developed by Safar et al. (22) in 1963; the "cardiac massage machine" developed by Bailey and Browse in 1964 (23); and the "hospital mechanical pump" developed by Nachlas and Siedband (21) in 1965 were the first MCCDs to be invented. In later years, experiments were conducted with many manually operated devices such as the "cardio-massager," "cardio-pulser," pneumatical "iron heart," and "Travenol LR50-90" (13, 24).
Many of the first developed and tested of these devices were very complex, too heavy, or ineffective for use in CPR; therefore, they were found to be nonfunctional and unacceptable for the clinical setting. On the other hand, since the 2000s, many devices have begun to be developed and have found a clinical area of use, and the literature of the effects of these devices on CPR outcomes are steadily increasing.
Classification
These devices are different from one another in terms of their working principles, the energy they consume, and their electronic features. MCCDs used currently can be classified as follows:
1. Piston-driven CPR devices (PD-CPR),
2. Load-distributing band CPR devices (LDB-CPR).
In addition to these two fundamental groups, third-generation devices, which combine different working mechanisms and different CPR techniques, have been used in recent years, aiming to increase the hemodynamic effects of S-CPR. These are as follows:
3. Active compression-decompression CPR devices (ACD-CPR),
4. Simultaneous sterno-thoracic cardiopulmonary resuscitation devices (SST-CPR/S-CPR/X -CPR),
5. Inspiratory impedance threshold valve/devices and ResQCPR (ACD + ITD CPR),
6. Phased thoracic-abdominal active compression-decompression CPR devices (PTACD-CPR), and
7. Active compression-decompression CPR with enhanced external counterpulsation and the inspiratory impedance threshold valve (AEI-CPR) (2, 6, 13, 25-29).
The devices in these groups and their working principles have been discussed below.
1. Piston-driven devices (PD-CPR)
These are based on the "cardiac pump theory" and are first-generation mechanical devices that use a piston to exert "single-point compression" on the sternum. One of the first examples in this group was the Pneumatically Run Thumper. A more developed model of this device is the Thumper Mechanical CPR Device Model 1007 and its updated model the Life-Stat. The Life-Stat consists of a backboard attached to a column and operates pneumatically with a piston. It has a ventilator that is meant to be used in conjunction with chest compression (Figure 1a) (Michigan Instruments, USA). Mechanical piston-driven devices that are operated manually work with a lever system and are marketed under brand names such as the "Animax Mono" (Figure 1b) (AAT Alber Antriebstechnik GmbH, Albstadt, Germany) and the "CPR RsQAssist," which employs an audio-visual metronome (10, 13, 24, 26, 29).
2. Load-distributing band devices (LDB-CPR)
"Load-distributing band devices" are based on the "thoracic pump theory" and represent second-generation mechanical chest compression technology. These devices exert thoracic compression on the anterior-anterolateral thorax using a wide pneumatic band that wraps around the chest, inflating and deflating at cyclically. The basic equipment in these devices consists of a backboard, a chest compression band (load-distributing Life Band), and a power system. The first example of this type of device was the "Vest-CPR." Currently, devices that work with this mechanism are marketed under commercial names such as the pneumatic Automated CPR Vest (Reax resuscitation device) and the pneumatic or electrical AutoPulse (Figure 2). Studies show that the chest compression achieved all around the chest with the AutoPulse creates higher coronary perfusion pressure than sternal pressure (13, 24, 25, 29-33).
3. Active compression-decompression CPR devices (ACD-CPR)
ACD-CPR devices are third-generation devices that work on the piston principle. These devices were developed based on a news article published in 1990 about a successful resuscitation attempt of a lay person performed with a toilet plunger to his father (34, 35). As is known, in S-CPR, the return of blood to the heart is dependent only on the passive recoil of the chest wall. The principle behind this technique may be summarized as the pumping of blood outside of the thorax through positive pressure in the active compression phase and then exerting an external negative vacu- um, creating an intrathoracic negative pressure during the active decompression phase to increase the venous return of blood to the heart. Thus, in the next compression phase, an increase is achieved in cardiac output, coronary and cerebral blood flow, and in arterial blood pressure. The most widely known and commonly used of the ACD-CPR devices is the LUCAS (Lund University Cardiac Assist System). The original LUCAS 1 was a pneumatic device that was developed in Sweden by Steen in 2002. The new model of the device, introduced in 2010 under its new name LUCAS 2 (LUCAS(TM) 2 Chest Compression System), runs on electricity and consists of a piston for compression, a silicone suction cup for decompression, a rechargeable battery, a backboard, and connecting legs (Figure 3a). The device allows defibrillation without interrupting compression, and its X-ray translucent capability makes cardiac catheterization possible. The ACDC Thumper is another pneumatic device. The manual devices that operate with the ACD-CPR technique are marketed under trade names such as CardioPump, ResQPump, and Ambu® Cardio Pump (Figure 3b) (6, 13, 24, 29, 36-40).
4. Simultaneous sterno-thoracic CPR devices (SST-CPR / X-CPR)
These devices were designed to benefit from both the cardiac pump and thoracic pump theories. These devices have two components: a piston (which depresses the sternum in the compression phase) and a circumferential band (which constricts the thorax simultaneously compressions).The "Life Belt" is a device that is operated manually using the SST-CPR principle. Another such device is the pneumatic "Weil Mini Chest Compressor" (Figure 4) (Resuscitation International, USA) (13, 41-44).
5. Inspiratory Impedance Threshold Device (ITD) and ResQCPR (ACD+ITD CPR)
The inspiratory ITD is a pressure-sensitive one-way valve system that can be connected to a face mask or to any developed airways equipment such as endotracheal tubing. The valve closes in the decompression phase of CPR, temporarily blocking the more than necessary passage of passive air through the open airway into the patient's lungs, thus decreasing intrathoracic pressure and creating a small vacuum. This increases the flow of venous blood to the heart, and the increased venous return increases cardiac output in the next compression. ITD are marketed under the trademark "ResQPOD® ITD 16." ITD can be used alone during S-CPR as well as it may be used in combination with manual ACD-CPR devices such as the CardioPump and the ResQPump. This system is known as ResQCPR. ResQCPR=ACD-CPR (ResQPUMP)+ITD (ResQPOD) (Figure 5a) (6, 27, 45-47).
6. Phased thoracic-abdominal active compression- decompression CPR devices (PTACD-CPR)
Interposed abdominal compression CPR (IAC-CPR) activates the abdominal venous reservoir by increasing abdominal pressure; this CPR technique is based on forcing venous return, thereby increasing venous return to the heart. Abdominal compression is applied to the area midpoint between the xiphoid and umbilicus in the relaxation phase of chest compression. Phased thoracic-abdominal active compression-decompression CPR constitutes the working principle behind ACD-CPR and IAC-CPR and is a new method that combines the two techniques. PTACD-CPR is applied by simultaneous chest compression (positive intrathoracic pressure) and active abdominal decompression and then following this phase, simultaneous active chest decompression (negative intrathoracic pressure) and abdominal compression. The Lifestick(TM) was developed for use in this technique; it is a manually controlled device. The device consists of a rigid central bar and two arms with adhesive pads that are connected to this rigid bar. The larger adhesive pad is placed over the abdomen and the smaller over the anterior chest wall. An implementer compresses the two sides of the device just like a seesaw, applying pressure both on the chest and the abdomen alternately (Figure 5b) (11, 27, 48-50).
7. Active compression-decompression CPR with Enhanced External Counterpulsation and the Inspiratory Impedance Threshold Valve (AEI-CPR)
Enhanced external counterpulsation (EECP) is a circulatory support system that achieves increased cardiac output using a method whereby cuffs applied to the lower extremities are inflated during diastole to increase coronary blood circulation and deflated at the early systole to reduce afterload and increase venous return. AEI-CPR is another experimental technique, which is a combination of active compression-decompression CPR, EECP, and Inspiratory impedance threshold valve, aiming to improve CPR hemodynamics and increase survival rates. This technique, which is still in its theoretical and experimental stage, is simply expressed as AEI-CPR=ACD-CPR+EECP+ITV (51, 52).
Studies Conducted on the Effectiveness of MCCDs
Theoretically, MCCDs appear to provide many practical advantages, such as the mechanical devices deliver compressions at the same frequency and depth which are recommended in the guidelines, as opposed to the inter-rescuer variations and fatigue factors that affect the quality of chest compression; these devices allow the rescuers to perform other tasks (cannulation, airway, etc.) and defibrillation without the need of interruption in CPR; and they provide consistent rate and depth of chest compressions during transport of the patient.
However, the main issue is to what degree these devices have an impact on survival in cardiac arrest, on hemodynamic parameters, and on the survival neurologically intact and whether they produce a significant difference in in-hospital and out-of-hospital cardiac arrests (IHCA and OHCA, respectively) compared to S-CPR. Experimental studies conducted with some mechanical chest compression devices developed in recent years present strong evidence that these devices increase the effectiveness and quality of CPR. Moreover, although they were first developed to achieve uninterrupted CPR, particularly in cases of OHCA, the studies on the use of these devices in IHCAs help to expand their clinical usage area. This section will review some of the results of some MCCD-related experimental or clinical studies on in-hospital and out-of-hospital cases.
Experimental studies with LUCAS-CPR have shown that the device enables significantly higher cerebral blood circulation than S-CPR as well as higher rates of cardiac output, carotid artery blood flow, end tidal CO2 (ETCO2), aortic and coronary perfusion pressure, and ROSC (9, 36, 53). The LUCAS device has been used in IHCA situations, in cardiac catheterization laboratories, and intensive care units. Studies show that the LUCAS device is functional during percutaneous coronary interventions (54-58). The findings reported in the mentioned studies have been summarized in Table 1. The use and effectiveness of mechanical chest compression devices such as LUCAS in organ transplants from non-heart-beating donors, in situations where a decision to terminate life and execute an organ transplant has been made, and where CPR is continued until the start of extracorporeal oxygenation (ECMO) are the subjects of ongoing studies (59).
Details of LUCAS-CPR studies with nontraumatic adult OHCA patients are shown in Table 1. Axelsson et al. (60) and Smekal et al. (61) have reported no significant differences in their studies when compared with S-CPR, whereas the same team in another study (62) revealed that ETCO2 values, which are a prognostic value for cardiac output and survival, were significantly higher in the LUCAS group than in the S-CPR group. In two large randomized studies ("LINC-LUCAS in Cardiac Arrest" and "PARAMEDIC-The Prehospital Randomised Assessment of a Mechanical Compression Device In Cardiac Arrest"), no significant difference was found between LUCAS-CPR and S-CPR in terms of ROSC and survival with good neurological outcomes (63, 64). In another study in which the skills of healthcare personnel in using LUCAS were evaluated, it was determined that the rate and depth of compressions applied to a manikin using LUCAS were inadequate compared with S-CPR (65).
Table 2 displays the findings of studies conducted on the effectiveness of AutoPulse in CPR (A-CPR). A study conducted with IHCA patients where AutoPulse was used reported that A-CPR produced higher coronary perfusion pressure than S-CPR (66). In two non-randomized studies using AutoPulse in OHCA patients, it was shown that A-CPR produced better results than S-CPR (32, 67). In a multi-center randomized study, the results of the ASPIRE Trial (Assisted Prehospital International Resuscitation Research) indicated that survival to hospital discharge and good neurological outcomes were lower in A-CPR than in S-CPR (68). The results of two nonrandomized, small sample size studies support the effectiveness of A-CPR (69, 70). Another large, randomized study (30) reports that survival rates with ROSC and satisfactory neurological outcomes were better with A-CPR than with S-CPR. A review of the results of the "Circulation-Improving Resuscitation Care (CIRC) Trial," another multicenter randomized study conducted with AutoPulse, revealed that A-CPR is equal to S-CPR in terms of ROSC and survival rates (71).
Table 3 presents the results of some studies that have reported on other MCCDs and techniques. Some trials on ResQCPR (ACD+ITD CPR) have reported short- and long-term survival rates to be higher than those for S-CPR (47, 72, 73). The results of trials with simultaneous sterno-thoracic CPR devices (SST-CPR/X-CPR) reveal hope for the future of these devices. In a study that consisted of a small series of cases, X-CPR produced higher coronary perfusion pressure than S-CPR (42). In a study with the Lifestick, a phased thoracic-abdominal active compression-decompression CPR device, no difference was detected compared with S-CPR in terms of ROSC, but it was reported that this technique could be advantageous for patients with asystole or pulseless electrical activity (48). Another study conducted with a small sample (49) did not report any statistically significant difference between using the Lifestick and applying S-CPR.
In the literature, there are also simulation studies where CPR devices and techniques were compared. Zhang et al. (11) used a circulation computer model in an experimental study to compare five ITD-supported techniques (S-CPR, ACD-CPR, IAC-CPR, Lifestick-CPR, and EECP-CPR) in terms of their hemodynamic effects, and they found Lifestick-CPR to be the most effective. A similar simulation study of five CPR techniques [S-CPR, ACD-CPR, IAC-CPR, ACD-CPR+External counterpulsation (ECP), and S-CPR+ECP] had made a comparison and found that cardiac output, cerebral blood flow, coronary blood flow, and mean coronary perfusion pressure to be the lowest in S-CPR and highest in IAC-CPR, with ACD-CPR+ECP exhibiting values close to this (51).
MCCDs in the Guidelines
An evaluation was made of the recommendations for use and the levels of evidence cited in the AHA 2010 and 2015 guidelines based on large randomized trials. In the case of automatic ACD-CPR devices such as LUCAS and LDB-CPR devices such as Autopulse, the guidelines state that the evidence to support or reject the routine use of these devices in the treatment of cardiac arrest is not sufficient and that manual chest compressions remain as the standard treatment of cardiac arrest; however, these devices may be a reasonable alternative for use by properly trained personnel (AHA 2015: Class IIb, LOE B-R). Furthermore, the guidelines state that the use of mechanical piston devices may be considered in specific settings where the delivery of high-quality manual compressions may be challenging or dangerous for the provider [e.g., limited rescuers available and prolonged CPR during hypothermic cardiac arrest, in a moving ambulance, in the angiography suite, and during preparation for extracorporeal CPR (ECPR)], provided that rescuers strictly limit interruptions in CPR during deployment and removal of the devices (AHA 2015: Class IIb, LOE C-EO) (6, 74).
The AHA 2010 guideline states with regard to manual ACD-CPR devices that there is no adequate evidence to either recommend or reject the routine use of these devices and that the use of the devices may be considered in the event of properly trained personnel. The AHA 2015 guideline has not made any revision with regard to these devices, maintaining the same recommendations and evidence level specified in the 2010 guideline (Class IIb, LOE B) (6, 74).
With respect to the sole use of the ITD-CPR device, the AHA 2010 guideline's recommendation and evidence level places the device in Class IIb, LOE B, whereas the AHA 2015 guideline has changed the recommendation and evidence level, placing it in the category of "Not recommended for routine use S-CPR" (Class III: No benefit, LOE A) (6, 74).
There appears to be no evaluation in the AHA 2010 guideline for ITD+ACD-CPR (RESQCPR). In the 2015 guideline, however, it is stated that this combination is not recommended for routine use as an alternative for S-CPR but may be considered as an alternative only in the presence of available equipment and trained personnel (Class IIb, LOE C-LD) (6, 74).
It can be seen that the AHA 2010 and 2015 guidelines do not include information and data on CPR devices and combinations such as the SST-CPR, PTACD-CPR, and AEI-CPR because these are still in the experimental stage and are not supported by adequate clinical research (6, 74).
Conclusion
In a general assessment, it may be stated that although large randomized trials have as yet highlighted the superiority of these devices over S-CPR in OHCA, they have at the same time not produced any evidence to prove their failure or harm.
The possibilities to be created by the harmonious cooperation of the disciplines of mathematics, biology, medicine, engineering, and the physical sciences in "the process of developing biomedical equipment technologies" and the role technology will play in constructing the future cannot be ignored. Therefore, an increase in the number of experimental and clinical research on CPR technologies and the evaluations & revisions performed according to the results of these studies will pave the way for the development of changes with respect to the application methods and areas of these devices. Ensuring that these devices become more functional, effective, and reliable will improve the effectiveness of CPR and may reduce the incidence of morbidity and mortality accompanying cardiac arrest.
Peer-review: Externally peer-reviewed.
Conflict of Interest: No conflict of interest was declared by the authors.
Financial Disclosure: The authors declared that this study has received no financial support.
References
1. Travers AH, Rea TD, Bobrow BJ, Edelson DP, Berg RA, Sayre MR, et al. 2010 American Heart Association Guidelines for Cardiopulmonary Resuscitation and Emergency Cardiovascular Care: Part 4: CPR Overview. Circulation 2010; 122: 676-84. [CrossRef]
2. Cooper JA, Cooper JD, Cooper JM. Contemporary reviews in cardiovascular medicine: cardiopulmonary resuscitation; history, current practice, and future direction. Circulation 2006; 114: 2839-49. [CrossRef]
3. Sasson C, Rogers MA, Dahl J, Kellermann AL. Predictors of survival from out-of-hospital cardiac arrest: a systematic review and meta-analysis. Circ Cardiovasc Qual Outcomes 2010; 3: 63-81. [CrossRef]
4. Kozaci N, Ay MO, Içme F, Aktürk A, Satar S. Are we successful in cardiopulmonary resuscitation? Cukurova Med J 2013; 38: 601-9.
5. Kleinman ME, Brennan EE, Goldberger ZD, Swor RA, Terry M, Bobrow BJ, et al. 2015 American Heart Association Guidelines Update for Cardiopulmonary Resuscitation and Emergency Cardiovascular Care: Part 5: Adult Basic Life Support and Cardiopulmonary Resuscitation Quality. Circulation 2015; 132: 414-35. [CrossRef]
6. Cave DM, Gazmuri RJ, Otto CW, Nadkarni VM, Cheng A, Brooks SC, et al. 2010 American Heart Association Guidelines for Cardiopulmonary Resuscitation and Emergency Cardiovascular Care: Part 7: CPR Techniques and Devices. Circulation 2010; 122: 720-8. [CrossRef]
7. Berg RA, Hemphill R, Abella BS, Aufderheide DC, Hazinski MF, Lerner B, et al. Part 5: adult basic life support: 2010 American Heart Association Guidelines for Cardiopulmonary Resuscitation and Emergency Cardiovascular Care. Circulation 2010; 122(18 Suppl 3): S685-705. [CrossRef]
8. Nolan JP, Soar J, Zideman DA, Biarent D, Bossaert LL, Deakin C, et al. European Resuscitation Council Guidelines for Resuscitation 2010 Section 1. Executive summary. Resuscitation 2010; 81: 1219-76. [CrossRef]
9. Rubertsson S, Karlsten R. Increased cortical cerebral blood flow with LUCAS; a new device for mechanical chest compressions compared to standard external compressions during experimental cardiopulmonary resuscitation. Resuscitation 2005; 65: 357-63. [CrossRef]
10. Keseg DP. The Merits of Mechanical CPR: Do mechanical devices improve compression consistency and resuscitation outcomes. JEMS 2012; 37: 24-9.
11. Zhang Y, Wu X, Yuan H, Xu L, Karemaker J. Computer models in bedside physiology. Chapter 2: Optimal cardiopulmonary resuscitation as identified by computer modeling. 2013: 12-32. Downloaded from UvA-DARE. Available at http://dare.uva.nl.
12. Topjian AA, Nadkarni AA, Berg VM. Cardiopulmonary resuscitation in children. Curr Opin Crit Care 2009; 15: 203-8. [CrossRef]
13. Gaxiola A, Varon J. Evolution and new perspective of chest compression mechanical devices. Am J Emerg Med 2008; 26: 923-31. [CrossRef]
14. Hightower D, Thomas SH, Stone CK, Dunn K, March JA. Decay in quality of closed-chest compressions over time. Ann Emerg Med 1995; 26: 300- 3. [CrossRef]
15. Ochoa FJ, Ramalle-Gomara E, Lisa V, Saralegui I. The effect of rescuer fatigue on the quality of chest compressions. Resuscitation 1998; 37: 149- 52. [CrossRef]
16. Wik L, Kramer-Johansen J, Myklebust H, Sorebo H, Svensson L, Fellows B, et al. Quality of cardiopulmonary resuscitation during out-of-hospital cardiac arrest. JAMA 2005; 293: 299-304. [CrossRef]
17. Perkins GD, Boyle W, Bridgestock H, Davies S, Oliver Z, Bradburn S, et al. Quality of CPR during advanced resuscitation training. Resuscitation 2008; 77: 69-74. [CrossRef]
18. Wigginton JG, Miller AH, Benitez FL, Pepe PE. Mechanical devices for cardiopulmonary resuscitation. Curr Opin Crit Care 2005; 11: 219-23. [CrossRef]
19. Classification of the External Cardiac Compressor (including CPR aid devices). FDA Executive Summary. Available at: http://www.fda.gov/downloads/ AdvisoryCommittees/CommitteesMeetingMaterials/Medical Devices. September 11, 2013.
20. Harkins GA, Bramson ML. Mechanized external cardiac massage for cardiac arrest and for support of the failing heart. A preliminary communication. J Surg Res 1961; 1: 197-200. [CrossRef]
21. Nachlas MM, Siedband MP. A simple portable pneumatic pump for external cardiac massage. Am J Cardiol 1962; 10: 107-9. [CrossRef]
22. Safar P, Harris LC. The Beck-Rand external cardiac compression machine. Anesthesiology 1963; 24: 586-8. [CrossRef]
23. Bailey, Browsw NL, Keating VJ. Automatic external cardiac massage: a portable pneumatic external cardiac compression machine. Brit Heart J 1964; 26: 481-9. [CrossRef]
24. Harrison-Poul R. A history of mechanical devices for providing external chest compressions. Resuscitation 2007; 73: 330-6. [CrossRef]
25. Jennings PA, Harris L, Bernard S, Bray J, Walker T, Spelman T, et al. An automated CPR device compared with standard chest compressions for out-of-hospital resuscitation. BMC Emerg Med 2012; 12: 2-5. [CrossRef]
26. Xu L, Wu X, Zhang Y, Yuan H. The Optimization Study on Time Sequence of Enhanced External Counter-Pulsation in AEI-CPR. Journal of Computers 2009; 4: 1243-8. [CrossRef]
27. Lurie KG. Mechanical devices for cardiopulmonary resuscitation: an update. Emerg Med Clin N Am 2002; 20: 771-84. [CrossRef]
28. Perkins GD, Brace S, Gates S. Mechanical chest-compression devices: current and future roles. Curr Opin Crit Care 2010; 16: 203-10. [CrossRef]
29. Halperin H, Carver DJ. Mechanical CPR devices. SIGNA VITAE 2010; 5: 69-73.
30. Ong ME, Fook-Chong S, Annathurai A, Hu Ang S, Tiah L, Yong KL, et al. Improved neurologically intact survival with the use of an automated, load-distributing band chest compression device for cardiac arrest presenting to the emergency department. Crit Care 2012; 16: 14-21.
31. Eroglu SE. Türkiye'de satisa sunulan gögüs kompresyon cihazlari: Autopulse. Available at: http://www.acilci.net/eroglu-turkiyede-satisa-sunulan-gogus- kompresyon-cihazlari- Autopulse. Erisim tarihi: 10 Agustos 2013.
32. Ong ME, Ornato JP, Edwards DP, Dhindsa HS, Best AM, Ines CS, et al. Use of an automated, load-distributing band chest compression device for out-of-hospital cardiac arrest resuscitation. JAMA 2006; 295: 2629-37. [CrossRef]
33. Haydin S, Ündar A. Yasam destek sistemlerinin Dünya'daki gelismeleri ve Türkiye'deki son durum. Anadolu Kardiyol Derg 2013; 13: 580-8.
34. Lurie KG, Lindo C, Chin J. CPR: the P stands for plumber's helper. JAMA 1990; 264: 1661. [CrossRef]
35. Altman LK. Toilet Plunger Is the Model For Device to Restart Hearts. The New York Times. Published: June 3, 1992.
36. Steen S, Liao Q, Pierre L, Paskevicius A, Sjoberg T. Evaluation of LUCAS, a new device for automatic mechanical compression and active decompression resuscitation. Resuscitation 2002; 55: 285-99. [CrossRef]
37. Eroglu SE. Türkiye'de satisa sunulan gögüs kompresyon cihazlari: LUCAS-2. Available at: http://www.acilci.net/eroglu-turkiyede-satisa-sunulan-gogus- kompresyon-cihazlari-LUCAS-2. Erisim tarihi: 16 Agustos 2013.
38. Ju H. Technology Brief Update: Mechanical chest compression devices for patients experiencing cardiac arrest. HealthPACT. Available at: https:// www.health.qld.gov.au/healthpact/docs/updates/WP153_update. pdf. May 2013.
39. Smekal D. Safety with mechanical chest Compressions in CPR; clinical studies with the LUCAS(TM) device. Available at: http://www.diva-portal. org/smash/get/diva2:639365/FULLTEXT01.pdf. 2013
40. Cohen TJ, Goldner BG, Maccaro PC, Ardito AP, Trazzera S, Cohen MB, et al. A comparison of active compression-decompression cardiopulmonary resuscitation with standard cardiopulmonary resuscitation for cardiac arrests occurring in the hospital. N Engl J Med 1993; 329: 1918-21. [CrossRef]
41. Hwang SO, Lee KH, Lee JW, Lee SY, Yoo BS, Yoon J, et al. Simultaneous sterno-thoracic cardiopulmonary resuscitation improves short-term survival rate in canine cardiac arrests. Resuscitation 2002; 53: 209-16. [CrossRef]
42. Cha KC, Kim H, Lee DG, Cha YS, Kim OH, Kim TH, et al. Hemodynamic effects of an automatic simultaneous sternothoracic CPR device in patients with cardiac arrest. Circulation 2014; 130: A108.
43. Yang Z, Gong P, Wu X, Qian J, Zhao S, Duggal N, et al. Similar hemodynamic efficacy between 30 mm and 50 mm compression depth during mechanical chest compression with weil mini chest compressor. Circulation 2014; 130: A306.
44. Chen W, Yinlun W, Xiaobo W, Shijie S, Joe B. The effects of a newly developed miniaturized mechanical chest compressor on outcomes of cardiopulmonary resuscitation in a porcine model. Crit Care Med 2012; 40: 3007-12. [CrossRef]
45. Lurie KG, Voelckel WG, Zielinski T, Mc Knite S, Lindstrom P, Colleen P, et al. Improving standard cardiopulmonary resuscitation with an inspiratory impedance threshold valve in a porcine model of cardiac arrest. Anesth Analg 2001; 93: 649-55. [CrossRef]
46. Thayne RC, Thomasa DC, Neville JD, Van Dellen A. Use of an impedance threshold device improves short-term outcomes following out-of-hospital cardiac arrest. Resuscitation 2005; 67: 103-8. [CrossRef]
47. Frascone RJ, Wayneb MA, Sworc RA, Mahoneyd BD, Domeiere RM, Olinger ML, et al. Treatment of non-traumatic out-of-hospital cardiac arrest with active compression decompression cardiopulmonary resuscitation plus an impedance threshold device. Resuscitation 2013; 84: 1214-22. [CrossRef]
48. Arntz HR, Agrawal R, Richter H, Schmidt S, Rescheleit T, Menges M, et al. Phased chest and abdominal compression-decompression versus conventional cardiopulmonary resuscitation in out-of-hospital cardiac arrest. Circulation 2001; 104: 768-72. [CrossRef]
49. Havel C, Berzlanovich A, Sterz F, Domanovits H, Herkner H, Zeiner A, et al. Safety, feasibility, and hemodynamic and blood flow effects of active compression-decompression of thorax and abdomen in patients with cardiac arrest. Crit Care Med 2008; 36: 1832-37. [CrossRef]
50. Kern KB, Hilwig RW, Berg RA, Schock RB, Ewy GA. Optimizing ventilation in conjunction with phased chest and abdominal compression-decompression (Lifestick(TM)) resuscitation. Resuscitation 2002; 52: 91-100. [CrossRef]
51. Luo J, Zeng H. Modeling and Simulation of Hemodynamic effects during different cardiopulmonary resuscitation (CPR) technologies. PeerJ PrePrints. Available at: http://dx.doi.org/10.7287/peerj.preprints.334v1. 31 Mar 2014.
52. Yannopoulos D, Halperin HR, Hopkins J. Lower extremity counterpulsation during the decompression phase of CPR improves hemodynamics and provides continuous forward carotid blood flow. Circulation 2007; 116: II-485.
53. Liao Q, Sjöberg T, Paskevicius A, Wohlfart B, Steen S. Manual versus mechanical cardiopulmonary resuscitation. An experimental study in pigs. BMC Cardiovasc Disord 2010; 10: 2-8. [CrossRef]
54. Larsen AI, Hjornevik AS, Ellingsen CL, Dennis WT. Cardiac arrest with continuous mechanical chest compression during percutaneous coronary intervention. A report on the use of the LUCAS device. Resuscitation 2007; 75: 454-9. [CrossRef]
55. Bonnemeier H, Olivecron G, Simonis G, Götberg M, Weitz G, Iblher P, et al. Automated continuous chest compression for in-hospital cardiopulmonary resuscitation of patients with pulseless electrical activity: A report of five cases. Int J Cardiol 2009; 136: 39-50. [CrossRef]
56. Bonnemeier H, Simonis G, Olivecron G. Continuous mechanical chest compression during in hospital cardiopulmonary resuscitation of patients with pulseless electrical activity. Resuscitation 2011; 82: 155-9. [CrossRef]
57. Wagner H, Terkelsen CJ, Friberg H, Harnek J, Kern K. Cardiac arrest in the catheterisation laboratory: a 5-year experience of using mechanical chest compressions to facilitate PCI during prolonged resuscitation efforts. Resuscitation 2010; 81: 383-7. [CrossRef]
58. Fidler R, Hirsch J, Strechert M, Johnson M. Three modes of cardiac compressions in a single patient: A comparison of usual manual compressions, automated compressions, and open cardiac massage. Resuscitation 2014; 85: 75-6. [CrossRef]
59. Rodríguez AM, Ferrara LP, Pascuala JMN, Alonsoa CB, Maldonadoa ME, Belmonte AA, et al. Kidney transplant function using organs from non-heart-beating donors maintained by mechanical chest compressions. Resuscitation 2010; 81: 904-7. [CrossRef]
60. Axelsson C, Nestin J, Svensson L, Axelsson AB, Herlitz J. Clinical consequences of the introduction of mechanical chest compression in the EMS system for treatment of out-of-hospital cardiac arrest- A pilot study. Resuscitation 2006; 71: 47-55. [CrossRef]
61. Smekal D, Johansson J, Huzevka T, Rubertsson S. A. Pilot study of mechanical chest compressions with the LUCAS(TM) device in cardiopulmonary resuscitation. Resuscitation 2011; 82: 702-6. [CrossRef]
62. Axelsson C, Karlsson T, Axelsson AB, Herlitzb J. Mechanical active compression decompression cardiopulmonary resuscitation (ACD-CPR) versus manual CPR according to pressure of end tidal carbon dioxide (PETCO2) during CPR in out-of-hospital cardiac arrest (OHCA). Resuscitation 2009; 80: 1099-103. [CrossRef]
63. Rubertsson S, Lindgren E, Smekal D, Östlund O, Silfverstolpe J, Lichtveld RA, et al. Mechanical chest compressions and simultaneous defibrillation vs conventional cardiopulmonary resuscitation in out-of-hospital cardiac arrest: the LINC randomized trial. JAMA 2014; 311: 53-61. [CrossRef]
64. Perkins GD, Lall R, Quinn T, Deakin CD, Cooke MW, Horton J, et al. Mechanical versus manual chest compression for out-of-hospital cardiac arrest (PARAMEDIC): a pragmatic, cluster randomized controlled trial. Lancet 2015; 385: 947-55. [CrossRef]
65. Blomberg H, Gedeborg R, Berglund L, Karlsten R, Johansson J. Poor chest compression quality with mechanical compressions in simulated cardiopulmonary resuscitation: A randomized, cross-over manikin study. Resuscitation 2011; 82: 1332-7. [CrossRef]
66. Timerman S, Cardoso LF, Ramires JA, Halperin H. Improved hemodynamic performance with a novel chest compression device during treatment of in-hospital cardiac arrest. Resuscitation 2004; 61: 273-80. [CrossRef]
67. Casner M, Andersen D, Isaacs SM. The impact of a new CPR device on rate of spontaneous circulation in out-of-hospital cardiac arrest. Prehosp Emerg Care 2005; 9: 61-7. [CrossRef]
68. Hallstrom A, Rea TD, Sayre MR, Christenson J, Anton AR. Manual chest compression vs use of an automated chest compression device during resuscitation following out-of-hospital cardiac arrest: a randomized trial. JAMA 2006; 295: 2620-8. [CrossRef]
69. Krep H, Mamier M, Breil M, Heister U, Fischer M, HoeftA, et al. Out-of-hospital cardiopulmonary resuscitation with the AutoPulseTM system: A prospective observational study with a new load-distributing band chest compression device. Resuscitation 2007; 73: 86-95. [CrossRef]
70. Duchateau FX, Gueye P, Curac S, Tubach F, Broche C, Plaisance P, et al. Effect of the AutoPulse(TM) automated band chest compression device on hemodynamics in out-of-hospital cardiac arrest resuscitation. Intensive Care Med 2010; 36: 1256-60. [CrossRef]
71. Wik L, Olsen JA, Presse D, Sterz F, Lozano M, Brouwer MA, et al. Manual vs. integrated automatic load-distributing band CPR with equal survival after out of hospital cardiac arrest. The randomized CIRC trial. Resuscitation 2014; 85: 741-8. [CrossRef]
72. Plaisance P, Lurie KG, Vicaut E, Martin D, Gueugniaud PY, Petit JL, et al. Evaluation of an impedance threshold device in patients receiving active compression decompression cardiopulmonary resuscitation for out of hospital cardiac arrest. Resuscitation 2004; 61: 265. [CrossRef]
73. Wolcke BB, Mauer DK, Schoefmann MF, Teichmann H, Provo TA, Lindner KH, et al. Comparison of standard cardiopulmonary resuscitation versus the combination of active compression-decompression cardiopulmonary resuscitation and an inspiratory impedance threshold device for out-of-hospital cardiac arrest. Circulation 2003; 108: 2201-5. [CrossRef]
74. Brooks SC, Anderson ML, Bruder E, Daya MR, Gaffney A, Otto CW, et al. 2015 American Heart Association Guidelines Update for Cardiopulmonary Resuscitation and Emergency Cardiovascular Care: Part 6: Alternative Techniques and Ancillary Devices for Cardiopulmonary Resuscitation. Circulation 2015; 132: 436-43.[CrossRef]
Mahmure Aygün1, Hacer Erten Yaman2, Aslý Genç1, Funda Karadaðlý1, Nadiye Barýþ Eren1
1Department of Nursing, Ýstanbul Geliþim University College of Health Science, Ýstanbul, Turkey
2Department of Nursing, Ýstanbul Maltepe University College of Health Science, Ýstanbul, Turkey
Correspondence to: Mahmure Aygün e-mail: [email protected]
Received: 18.11.2015 Accepted: 02.03.2016
You have requested "on-the-fly" machine translation of selected content from our databases. This functionality is provided solely for your convenience and is in no way intended to replace human translation. Show full disclaimer
Neither ProQuest nor its licensors make any representations or warranties with respect to the translations. The translations are automatically generated "AS IS" and "AS AVAILABLE" and are not retained in our systems. PROQUEST AND ITS LICENSORS SPECIFICALLY DISCLAIM ANY AND ALL EXPRESS OR IMPLIED WARRANTIES, INCLUDING WITHOUT LIMITATION, ANY WARRANTIES FOR AVAILABILITY, ACCURACY, TIMELINESS, COMPLETENESS, NON-INFRINGMENT, MERCHANTABILITY OR FITNESS FOR A PARTICULAR PURPOSE. Your use of the translations is subject to all use restrictions contained in your Electronic Products License Agreement and by using the translation functionality you agree to forgo any and all claims against ProQuest or its licensors for your use of the translation functionality and any output derived there from. Hide full disclaimer
Copyright Aves Yayincilik Ltd. STI. Jun 2016
Abstract
The standard treatment of cardiac arrest is cardiopulmonary resuscitation (CPR), performed with effective manual chest compressions. Although current CPR was developed 50 years ago, cardiac arrest still has a high mortality rate and manual chest compressions have some potential limitations. Because of these limitations, mechanical chest compression devices were developed to improve the efficiency of CPR. This CPR technology includes devices such as the mechanical piston load-distributing band, active compression-decompression CPR, simultaneous sterno-thoracic CPR, impedance threshold valve, phased thoracic-abdominal active compression-decompression CPR and active compression-decompression CPR with enhanced external counterpulsation, and the impedance threshold valve. The purpose of this manuscript was to draw attention to developments in this medical area and to examine studies on the effectiveness of these devices.
You have requested "on-the-fly" machine translation of selected content from our databases. This functionality is provided solely for your convenience and is in no way intended to replace human translation. Show full disclaimer
Neither ProQuest nor its licensors make any representations or warranties with respect to the translations. The translations are automatically generated "AS IS" and "AS AVAILABLE" and are not retained in our systems. PROQUEST AND ITS LICENSORS SPECIFICALLY DISCLAIM ANY AND ALL EXPRESS OR IMPLIED WARRANTIES, INCLUDING WITHOUT LIMITATION, ANY WARRANTIES FOR AVAILABILITY, ACCURACY, TIMELINESS, COMPLETENESS, NON-INFRINGMENT, MERCHANTABILITY OR FITNESS FOR A PARTICULAR PURPOSE. Your use of the translations is subject to all use restrictions contained in your Electronic Products License Agreement and by using the translation functionality you agree to forgo any and all claims against ProQuest or its licensors for your use of the translation functionality and any output derived there from. Hide full disclaimer