© 2017 de Oliveira et al. This is an open access article distributed under the terms of the Creative Commons Attribution License: http://creativecommons.org/licenses/by/4.0/ (the “License”), which permits unrestricted use, distribution, and reproduction in any medium, provided the original author and source are credited. Notwithstanding the ProQuest Terms and Conditions, you may use this content in accordance with the terms of the License.

Abstract

Background

During anesthesia, as compared with intensive care, the time of the tracheal intubation is much shorter. An inhaled gas minimum humidity of 20 mgH2O.L-1 is recommended to reduce the deleterious effects of dry gas on the airways during anesthesia with tracheal intubation. The Fabius GS Premium® anesthesia workstation (Dräger Medical, Lübeck, Germany) has a built-in hotplate to heat gases in the breathing circuit. A heat and moisture exchanger (HME) is used to further heat and humidify the inhaled gas. The humidity of the gases in the breathing circuit is influenced by the ambient temperature. We compared the humidity of the inhaled gases from a low-flow Fabius anesthesia workstation with or without thermal insulation (TI) of the breathing circuit and with or without an HME.

Methods

We conducted a prospective randomized trial in 41 adult female patients who underwent elective abdominal surgery. The patients were allocated into four groups according to the devices used to ventilate their lungs using a Dräger Fabius anesthesia workstation with a low gas flow (1 L.min-1): control, with TI, with an HME or with TI and an HME (TIHME). The mean temperature and humidity of the inhaled gases were measured during 2-h after connecting the patients to the breathing circuit.

Results

The mean inhaled gas temperature and absolute humidity were higher in the HME (29.2±1.3°C; 28.1±2.3 mgH2O·L-1) and TIHME (30.1±1.2°C; 29.4±2.0 mgH2O·L-1) groups compared with the control (27.5±1.0°C; 25.0±1.8 mgH2O·L-1) and TI (27.2±1.1°C; 24.9±1.8 mgH2O·L-1) groups (P = 0.003 and P<0.001, respectively).

Conclusions

The low-flow Fabius GS Premium breathing circuit provides the minimum humidity level of inhaled gases to avoid damage to the tracheobronchial epithelia during anesthesia. TI of the breathing circuit does not increase the humidity of the inhaled gases, whereas inserting an HME increases the moisture of the inhaled gases closer to physiological values.

Details

Title
The Humidity in a Low-Flow Dräger Fabius Anesthesia Workstation with or without Thermal Insulation or a Heat and Moisture Exchanger: A Prospective Randomized Clinical Trial
Author
Sergius A R de Oliveira; Lucio, Lorena M C; Modolo, Norma S P; Hayashi, Yoko; Braz, Mariana G; de Carvalho, Lídia R; Braz, Leandro G; José Reinaldo C Braz
First page
e0170723
Section
Research Article
Publication year
2017
Publication date
Jan 2017
Publisher
Public Library of Science
e-ISSN
19326203
Source type
Scholarly Journal
Language of publication
English
ProQuest document ID
1862248564
Copyright
© 2017 de Oliveira et al. This is an open access article distributed under the terms of the Creative Commons Attribution License: http://creativecommons.org/licenses/by/4.0/ (the “License”), which permits unrestricted use, distribution, and reproduction in any medium, provided the original author and source are credited. Notwithstanding the ProQuest Terms and Conditions, you may use this content in accordance with the terms of the License.