PHOTONIC SENSORS / Vol. 7, No. 2, 2017: 97104
Optofluidic Refractive Index Sensor Based on Partial Reflection
Lei ZHANG1*, Zhang ZHANG1, Yichuan WANG1, Meiying YE2, Wei FANG1*, and Limin TONG1
1State Key Laboratory of Modern Optical Instrumentation, College of Optical Science and Engineering, Zhejiang
University, Hangzhou, 310027, China
2College of Material, Chemistry and Chemical Engineering, Hangzhou Normal University, Hangzhou, 310036, China
*Corresponding author: Lei ZHANG and Wei FANG E-mail: [email protected] and [email protected]
Abstract: We demonstrate a novel optofluidic refractive index (RI) sensor with high sensitivity and wide dynamic range based on partial reflection. Benefited from the divergent incident light and the output fibers with different tilting angles, we have achieved highly sensitive RI sensing in a wide range from 1.33 to 1.37. To investigate the effectiveness of this sensor, we perform a measurement of RI with a resolution of ca. 5.0105 refractive index unit (RIU) for ethylene glycol solutions. Also, we have measured a series of liquid solutions by using different output fibers, achieving a resolution of ca. 0.52 mg/mL for cane surge. The optofluidic RI sensor takes advantage of the high sensitivity, wide dynamic range, small footprint, and low sample consumption, as well as the efficient fluidic sample delivery, making it useful for applications in the food industry.
Keywords: Optofluidic sensor; refractive index; partial reflection
Citation: Lei ZHANG, Zhang ZHANG, Yichuan WANG, Meiying YE, Wei FANG, and Limin TONG, Optofluidic Refractive Index Sensor Based on Partial Reflection, Photonic Sensors, 2017, 7(2): 97104.
1. Introduction
Miniaturized optofluidic devices offer great potential for realizing more functional and more compact optical systems through the fusion of microfluidics and optics [1, 2]. The past decade has witnessed the realizations of optofluidic devices including tunable waveguides [3, 4], lenses [5, 6], switches [7, 8], apertures [9, 10], reconfigurable lasers [11, 12], interferometers [13, 14], gratings [15, 16], and sensitive sensors with fast response and low sample consumption [1720], as well as demonstrations of their applications in chemical and biological analysis [21], energy [22], and photonics [23]. For many biological and chemical applications,
the measurement of refractive index (RI) is extremely useful for detecting compounds that are: nonionic, transparent in the UV/vis range, or with no fluorescence, moreover, it has great potential for none invasive and label-free biosensing [2426].
To date, a variety of optofluidic RI sensors have been reported by using integrated interferometers [27, 28], Fabry-Prot cavities [29, 30], microring resonators [31, 32] gratings [33, 34], microstructured optical fibers [35, 36], and surface plasmon resonance [37, 38]. Although the sensitivity of the aforementioned sensors can be as high as 106
RIU by measuring the wavelength shift, these sensors require expensive instruments, e.g., optical spectrum analyzers. Alternatively, one can design a
Received: 12 August 2016/ Revised: 22 September 2016 The Author(s) 2017. This article is published with open access at Springerlink.com DOI: 10.1007/s13320-017-0369-5Article type: Regular
Photonic Sensors
98
highly sensitive RI sensor based on partial reflection or refraction by measuring the intensity of the reflected or transmitted light or both [39, 40]. It is well known that the reflectivity of liquid-solid interface can be approximated by combining Snells law with the Fresnel equations of reflection. When the RI of the liquid increases to a value where total internal reflection (TIR) at the liquid-solid interface is no longer satisfied, the reflectivity will decrease sharply, leading to a strong increase of the transmitted light. This phenomenon provides an attractive feature for assembling tunable optofluidic device and RI sensors. For example, Lapsley and co-authors [39] reported a variable optical attenuator where the light attenuation was achieved by adjusting the RI of the liquid in the microfluidic channel and thus altering the reflectivity of the light at the sidewall of the microchannel. Recently, Weber and Vellekoop [40] reported an optofluidic sensor, which was consisted of one input fiber and two out fibers with fixed angles for guiding the incident light, reflected light, and refracted light, respectively. Also, integrated microlenses were used for collimating the divergent light. Note that an inherent disadvantage of this sensor is the narrow dynamic range, typically, 0.01 refractive index unit (RIU) for a fixed incident angle. Thus, one has to prepare many chips with different incident angles to determine samples within a wide range of RI.
In this work, we report an optofluidic RI sensor with one straight detection channel, two input fibers, and ten output fibers for RI sensing with high sensitivity and wide dynamic range. Compared with the existing counterparts that require microlenses for light collimation, our design takes advantage of the divergent incident light and the optimized position of the output fibers to realize high sensitive sensing in a dynamic range from 1.33 to 1.37. In this case, we can measure the RI of most aqueous samples with high sensitivity by choosing an output fiber with a specific tilting angle. Since microlenses are removed from our design, the fabrication of the sensor becomes much simpler and cost-effective. To
investigate the effectiveness of the RI sensor, we perform a measurement of ethylene glycol (EG) solutions with a RI range from 1.33 to 1.37, achieving a resolution of ca. 5.0105 RIU. Furthermore, we present measurements of liquid concentrations with a resolution of ca. 0.52 mg/mL for cane surge. The optofluidic RI sensor shown here may provide a compact and versatile sensing platform for sensitive and fast detection of low-volume samples.
2. Experimental section
2.1 Reagents and instruments
All reagents and standards were of analytical grade and purchased from Sinopharm Chemical Reagent (Shanghai, China) unless otherwise stated. SU-8 photoresist and SU-8 developer were purchased from MicroChem Corp. (Newton, MA, USA). Polydimethylsiloxane (PDMS) (Sylgard 184) was purchased from Dow Corning (Midland, MI, USA). Ultrapure water (Siemens Labostar2-UV) was used throughout. Sample solutions were prepared before use. Optical micrographs were obtained using a charge coupled device (CCD) camera (DS-Fi1, Nikon, Japan) mounted on a microscope (Eclipse 90i, Nikon, Japan).
2.2 Fabrication of the optofluidic device
Figure 1 shows an optical image of a PDMS microfluidic chip for multi-fiber RI sensing, in which it contains one detection channel, two fiber channels with a tilting angle of 71, and ten output fibers channels with tilting angles range from 67 to 76, respectively. The tilting angle of 71 was calculated by using Snells law [40], and the tilting angles of the output fibers were defined according to the divergent incident light. The microfluidic chip was fabricated by using standard soft lithography process [41]. Briefly, uncured PDMS was poured onto a SU-8 master, followed by curing at 80 for 20 min. The cured PDMS slab was then peeled from the SU-8 master, and punched holes at the ends of
Lei ZHANG et al.: Optofluidic Refractive Index Sensor Based on Partial Reflection 99
the detection channel and the PDMS channels to introduce sample and uncured PDMS, respectively. The PDMS slab was bonded with a clean glass slideby using a plasma cleaner (PDC-32G-2, Harrick, USA). Typically, the detection channel was 2 cm in length, 125 m in width, and 150 m in depth with a rectangular cross section. In this case, optical fibers could be inserted into the fiber channels and well aligned by fiber channels. To remove the scattering light, the gap between the fiber and fiber channel was sealed by introducing uncured PDMS from the PDMS inlets, followed by curing at 80 for 20 min.
Fig. 1 Optical image of a microfluidic chip for multi-fiber RI sensing, which contains one detection channel, two input fibers channels, four PDMS channels, and ten output fiber channels. The dot and dash lines indicate the divergent incident light and reflected light, respectively.
2.3 Procedures
In the experiment, 5 L of sample solution was added into the sample inlet and was introduced into the detection channel by negative pressure generated by a syringe. After each measurement, the channel was flushed with ultrapure water. A 473 nm laser and a broadband light from a tungsten halide lamp (PHILIPS 7748XHP) were coupled into input fibers and served as incident light for visualization and RI measurements, respectively. The reflected light was collected by one of the ten output fibers and recorded by a spectrometer (Maya2000 Pro, Ocean optics, Dunedin, FL, USA).
3. Working principle
Figure 2(a) schematically shows a typical structure for illustrating the operating mechanism of the RI sensor, where and denote the tilting
angles of the input and output fibers, respectively, and d denotes the distance between the end face of an output fiber and the reflection zone.
Input fiber PDMS inlet Output fiber
Sample outlet
R
Sample inlet
Detection channel
I
d
T
(a)
Reflection region
Detection channel
(b)
(c)
Fig. 2 Illustration of a microfluidic RI sensor: (a) a schematic illustration of a microfluidic RI sensor with one pair of fibers for optimizing the tilting angles of the input fiber and the distance between the end face of the output fiber to the reflection zone and (b) optical micrograph of a microfluidic RI sensor with ==71, and d=0.6 mm, and (c) cross section of the detection channel. Scale bars: 125 m.
Figure 2(b) shows an optical micrograph of an optofluidic RI sensor with ==71 and d=0.6 mm. As shown in Fig. 2(c), the vertical sidewalls of the detection channel serve as reflective interfaces. When the liquid sample is introduced into the detection channel, a RI mismatch between PDMS and the liquid sample in the detection channel will cause the incident optical beam (I) to be reflected and refracted at the PDMS-liquid interface, resulting in a reflected beam (R) and a transmitted beam (T). The angle of R() is the same as the angle of I, allowing the reflected beam to be captured by the
Photonic Sensors
100
output optical fiber. The intensity of the reflected light is a function of RI of the liquid sample, and thus, we can determine the concentration of solutes by measuring the intensity of the reflected light. Note that we did not fabricate microlenses to collimate incident light, it offers a possibility to arrange several output fibers with different tilting angles to record the divergent reflected beam. In this case, each output fiber corresponds to an angle of reflection, providing an effective method to achieve a wide dynamic range. On the other hand, the insertion loss of the device is relatively greater than its counterparts with microlenses, and approximately 5% incident light could be collected by the output fiber for the microfluidic chip as shown in Fig. 2(b).
4. Results and discussion
4.1 Visualization of the reflected and refracted beam
To visualize the reflected and refracted beams, a 473-nm-wavelength laser was used as incident light and 0.01 mM fluorescein solutions with different RIs were used to visualize the refracted light. Figure 3
shows four typical images of an optofluidic RI sensor (==71) with different fluids in the detection channel. When air (n=1) was in the detection channel, TIR occurred at the PDMS-air interface, and the divergent reflected light could be clearly seen in Fig. 3(a). When the tilting angle ofthe input fiber was 71, the estimated angle of reflection was from 67 to 76. When fluorescein solutions were injected into the detection channel,the fluorescence intensity increased obviously withan increase in RI from 1.33, 1.36, to 1.42 [Figs. 3(b)(d)], indicating an increase in the refracted light intensity and a decrease in the reflected one. When the fluorescein solution with aRI of 1.42 was in the detection channel, the sidewallof the detection channel could hardly be identified,and the incident beam transmitted through the detection channel without obvious refraction and reflection [Fig. 3(d)].
(a)
(b)
(c)
(d)
Fig. 3 Optical images of reflected and refracted light with different fluids within the detection channel: (a) air (n=1) and (b)(d) 0.01 mM fluorescein was introduced into the detection channel as an indicator of the refracted light. The RI of the solutions were 1.33, 1.36, and 1.42, respectively. Scale bar: 125 m.
4.2 Theoretical calculation
To achieve a fundamental understanding of the sensors RI response, theoretical analysis was conducted using Fresnel equations. We calculated the reflection coefficients rs and rp for s- and ppolarizations by (1) and (2), respectively, with the total reflectivity calculated by (3)
1 2
1 2
n n r
n n
cos
cos
(1)
s
n n r
n n
cos
cos
2 1
2 1
p
(2)
2
2
r r
R
s p
(3)
where n1=1.41 is the RI of PDMS, and n2 is the RI of the liquid in the detection channel. Thus, the reflectivity is a function of n2. Figure 4(a) shows a theoretical analysis of the reflected light intensity of an RI sensor with different angles of reflection. For a fixed angle , we found that the intensity kept as unity for small n2 because of the TIR at the PDMS-liquid interface. When n2 increased to a value where TIR was no longer satisfied, a very sharp decrease in reflectivity was observed with very small change in n2. Based on theoretical prediction, the slope can be infinity right at the critical point, thus a very sensitive RI sensor could be realized. For example, when n2 changed 0.001
2
Lei ZHANG et al.: Optofluidic Refractive Index Sensor Based on Partial Reflection 101
RIU, the maximum change in the intensity could be 30%, however, the dynamic range of the sensor was ca. 0.01 RIU, which was relatively narrow for real application. As shown in Fig. 4(a), when increases from 69 to 74, the critical RI of n2 shifts from 1.33 to 1.37. Thus, we could achieve a wide dynamic range by choosing an output fiber with different .
3.3 Effect of d on the measurement of RI
To optimize d of the RI sensor, we fabricated a series of microfluidic chips with one input fiber and one output fiber for characterization. Owing to the large aperture and core diameter, a multi-mode fiber could collect light from a wide angle, and thus smearing out the sharp decrease of the light coming from a single collection angle as indicated by the theoretical calculations [Fig. 4(a)]. In this work, we adopted SMFs with a core diameter around 10 m as input and output optical fibers, thus, the reflection light within a narrow range of angle could be collected by the output fiber. We found that d had a great impact on the performance of the sensor. Figures 4(b) and 4(c) show normalized intensity of the light collected by the output fiber with different
tilting angles as a function of RI for d of 0.6 mm, and 1.5 mm, respectively. The results did not match with the theoretical analysis [Fig. 4(a)], especially, the response curves for =69 and =71 can be hardly resolved in the RI range of 1.331.34. This mismatch could be attributed to the output fiber collecting the light with different angles of reflection. Increasing d is an effective method to narrow the range of the angle of reflection. When we increased d to 5 mm, the flare angle of the reflected light that could be collected by a SMF with a core diameter of 10 m decreased dramatically. To obtain the estimated flare angle of the reflected light, we considered the reflection area as a point, thus, the sides of the isosceles triangle were 5 mm and 10 m, respectively, resulting in a flare angle of approximately 0.1. As shown in Fig. 4(d), all of the four response curves for the different tilting angles are well separated, and the slopes after the critical points become steeper, which matched well with the theoretical calculations [Fig. 4(a)]. It is worth mentioning that we can arrange more output fiber on one side the detection to collect reflected light with different angles when d is 5 mm.
6 9
7 1
7 2 7 4
69
71
72
74
Normalized intensity
100
75
50
25
Normalized intensity
100
75 50 25
1.32 1.34 1.36 1.38 1.40 1.42
Refractive index
d=0.6 mm
0
0
1.32 1.34 1.36 1.38 1.40 1.42
R efractive index
(a)
(b)
Normalized intensity
100
75
50
25
1.32 1.34 1.36 1.38 1.40 1.42 Refractive index
69 71 72 74
69
71
72
74
Normalized intensity
100
75 50 25
1.32 1.34 1.36 1.38 1.40 1.42
Refractive index
d=1.5 mm
d=5 mm
0
0
(c)
(d)
Fig. 4 Reflection response of the RI sensor for different : (a) theoretical analysis of the reflection response of the RI sensor for different and (b)(d) normalized intensity of the reflected light collected by four output fibers with reflected angles of 69, 71, 72, and 74, respectively, except for the variations to the distance between the endface of an output fiber and the reflection zone as indicated in the figures.
Photonic Sensors
102
4.4 Performance of the multi-fiber RI sensor
Figure 5 shows four typical curves recorded by the output fibers with a tilting angle of 69, 71, 72, and 74, respectively, which covers a RI range from 1.33 to 1.37. The result confirmed that the dynamic range could be extended by using different output fibers for sensing. When RI increased from 1.35 to 1.36, the normalized intensity for the output fiber with a tilting angle of 72 decreased from 103 to 23. Because the stability of the light source was measured with an outstanding reproducibility of 0.13% RSD, the RI resolution could be estimated as 5.0105 RIU based on 3 times the standard deviation of the light source. Note that the signals collected by the output fibers with a tilting angle less than 68 or greater than 74 were too weak for sensing owing to the Gaussian intensity distribution of the divergent incident beam.
Normalized intensity
100
80
60
40
20
69 71 72 74
1.33 1.34 1.35 1.36 1.37 Refractive index
Fig. 5 Normalized intensity of the reflected light collected by four output fibers with tilting angles of 69, 71, 72, and 74, respectively.
Table 1 Measurement results of the concentration of cane surge obtained by using three output fibers with different tilting angles.
Referenced concentration
(g/mL)
Referenced RI
Experimental value (g/mL)
Tilting
Angle
RE (%)
0.100 1.345 9 0.105 71 5.0 0.150 1.352 5 0.154 72 2.7
0.200 1.358 6 0.192 72 4.0 0.250 1.364 2 0.246 74 1.6 0.300 1.369 2 0.307 74 2.3
To further evaluate the performance of the RI sensor, aqueous cane surge solutions with different concentrations (0.1 g/mL0.3 g/mL) were prepared and analyzed. Because the RI of the cane surge solution is in a range from 1.34 to 1.37, we adopted
three output fibers with tilting angle of 71, 72, and 74 for collecting the signals, respectively. Table 1 gives obtained values and the relative errors (RE) for the sensor, showing a wide dynamic range and high accuracy with an estimated resolution of 0.52 mg/mL.
5. Conclusions
In summary, we have developed a sensitive, robust, and inexpensive optofluidic RI sensor. We have realized highly sensitive RI sensing in a wide dynamic range of 1.331.37 with a RI resolution of 5.0105 RIU by choosing an output fiber with different tilting angles for signal recording. Also, we have measured a series of liquid solutions by using different output fibers, achieving a resolution of ca. 0.52 mg/mL for cane surge. To further enhance the sensing performance, one might use a higher resolution lithography instrument to fabricate a SU-8 master with a smoother sidewall. We believe this optofluidic RI sensor can be used for real-time, low-cost, and multifunctional measurements in a wide range of applications.
Acknowledgment
This work has been supported in part by National Basic Research Program of China (Nos. 2013CB328703 and 2014CB921303), and National Natural Science Foundation of China (61275217, 21407039).
Open Access This article is distributed under the terms of the Creative Commons Attribution 4.0 International License (http://creativecommons.org/ licenses/by/4.0/), which permits unrestricted use, distribution, and reproduction in any medium, provided you give appropriate credit to the original author(s) and the source, provide a link to the Creative Commons license, and indicate if changes were made.
References
[1] D. Psaltis, S. R. Quake, and C. H. Yang, Insight review: developing optofluidic technology through the fusion of microfluidics and optics, Nature, 2006, 442(7101): 381386.
Lei ZHANG et al.: Optofluidic Refractive Index Sensor Based on Partial Reflection 103
[2] C. Monat, P. Domachuk, and B. J. Eggleton, Integrated optofluidics: a new river of light, Nature Photonics, 2007, 1(2): 106114.
[3] D. B. Wolfe, R. S. Conroy, P. Garstecki, B. T. Mayers, M. A. Fischbach, K. E. Paul, et al., Dynamic control of liquid-core/liquid-cladding optical waveguides,Proceedings of the National Academy of Sciences, 2004, 101(34): 1243412438. [4] B. T. Mayers, D. V. Vezenov, V. I. Vullev, and G. M.
Whitesides, Arrays and cascades of fluorescent liquid-liquid waveguides: broadband light sources for spectroscopy in microchannels, Analytical Chemistry, 2005, 77(5): 13101316.
[5] X. L. Mao, J. R. Waldeisen, B. K. Juluri, and T. J.
Huang, Hydrodynamically tunable optofluidic cylindrical microlens, Lab on a Chip, 2007, 7(10): 13031308.
[6] P. Fei, Z. He, C. Zheng, T. Chen, Y. Men, and Y.
Huang, Discretely tunable optofluidic compound microlenses, Lab on a Chip, 2011, 11(17): 28352841.
[7] A. Groisman, S. Zamek, K. Campbell, L. Pang, U.
Levy, and Y. Fainman, Optofluidic 14 switch,Optics Express, 2008, 16(18): 1349913508.
[8] W. Z. Song and D. Psaltis, Electrically tunable optofluidic light switch for reconfigurable solar lighting, Lab on a Chip, 2013, 13(14): 27082713.
[9] Y. Hongbin, Z. Guangya, C. F. Siong, and L. Feiwen, Optofluidic variable aperture, Optics Letters, 2008, 33(6): 548550.
[10] C. L. Song, N. T. Nguyen, A. K. Asundi, and C. L. N.
Low, Tunable optofluidic aperture configured by a liquid-core/liquid-cladding structure, Optics Letters, 2011, 36(10): 17671769.
[11] Z. Li, Z. Zhang, T. Emery, A. Scherer, and D. Psaltis, Single mode optofluidic distributed feedback dye laser, Optics Express, 2006, 14(2): 696701.
[12] M. Aas, A. Jonas, A. Kiraz, O. Brzobohaty, J. Jezek, Z. Pilat, and P. Zemanek, Spectral tuning of lasing emission from optofluidic droplet microlasers using optical stretching, Optics Express, 2013, 21(18): 2138021394. [13] C. Grillet, P. Domachuk, V. Ta'eed, E. Mgi, J.
Bolger, B. Eggleton, et al., Compact tunable microfluidic interferometer, Optics Express, 2004, 12(22): 54405447.
[14] Y. Zou, Z. Hen, X. Hen, Z. Di, and X. Chen, An integrated tunable interferometer controlled by liquid diffusion in polydimethylsiloxane, Optics Express, 2012, 20(17): 1893118936.
[15] M. Hashimoto, B. Mayers, P. Garstecki, and G. M.
Whitesides, Flowing lattices of bubbles as tunable, self-assembled diffraction gratings, Small, 2006, 2(11): 12921298.
[16] J. Q. Yu, Y. Yang, A. Q. Liu, L. K. Chin, and X. M. Zhang, Microfluidic droplet grating for reconfigurable optical diffraction, Optics Letters, 2010, 35(11): 18901892.
[17] P. Liu, H. Huang, T. Cao, X. Liu, Z. Qi, Z. Tang, et al., An ultra-low detection-limit optofluidic biosensor with integrated dual-channel Fabry-Prot cavity, Applied Physics Letters, 2013, 102(16): 163701-1163701-4.
[18] A. A. P. Trichet, J. Foster, N. E. Omori, D. James, P. R. Dolan, G. M. Hughes, et al., Open-access optical microcavities for lab-on-a-chip refractive index sensing, Lab on a Chip, 2014, 14(21): 42444249. [19] Z. Xu, K. Han, I. Khan, X. Wang, and G. L. Liu, Liquid refractive index sensing independent of opacity using an optofluidic diffraction sensor, Optics Letters, 2014, 39(20): 60826085.
[20] C. Wu, M. L. V. Tse, Z. Liu, B. O. Guan, A. P. Zhang, C. Lu, et al., In-line microfluidic integration of photonic crystal fibers as a highly sensitive refractometer, Analyst, 2014, 139(21): 54225429. [21] X. D. Fan and I. M. White, Optofluidic microsystems for chemical and biological analysis, Nature Photonics, 2011, 5(10): 591297.
[22] D. Erickson, D. Sinton, and D. Psaltis, Optofluidics for energy applications, Nature Photonics, 2011, 5(10): 583590.
[23] L. Pang, H. M. Chen, L. M. Freeman, and Y. Fainman, Optofluidic devices and applications in photonics, sensing and imaging, Lab on a Chip, 2012, 12(19): 35433551.
[24] P. J. Viskari and J. P. Landers, Unconventional detection methods for microfluidic devices, Electrophoresis, 2006, 27(9): 17971810.
[25] Y. Wang, S. Meng, Y. Liang, L. Li, and W. Peng, Fiber-optic surface plasmon resonance sensor with multi-alternating metal layers for biological measurement, Photonic Sensors, 2013, 3(3): 202207.
[26] J. Zhu, L. Qin, S. Song, J. Zhong, and S. Lin, Design of a surface plasmon resonance sensor based on grating connection, Photonic Sensors, 2015, 5(2): 159165.
[27] L. K. Chin, A. Q. Liu, Y. C. Soh, C. S. Lim, and C. L. Lin, A reconfigurable optofluidic Michelson interferometer using tunable droplet grating, Lab on a Chip, 2010, 10(8): 10721078.
[28] M. I. Lapsley, I. K. Chiang, Y. B. Zheng, X. Y. Ding, X. L. Mao, and T. J. Huang, A single-layer, planar, optofluidic Mach-Zehnder interferometer for label-free detection, Lab on a Chip, 2011, 11(10): 17951800. [29] W. Z. Song, X. M. Zhang, A. Q. Liu, C. S. Lim, P. H. Yap, and H. M. M. Hosseini, Refractive index
Photonic Sensors
104
measurement of single living cells using on-chip Fabry-Prot cavity, Applied Physics Letters, 2006, 89(20): 203901-1203901-3. [30] Y. Guo, H. Li, K. Reddy, H. S. Shelar, V. R. Nittoor, and X. Fan, Optofluidic Fabry-Prot cavity biosensor with integrated flow-through micro-/ nanochannels, Applied Physics Letters, 2011, 98(4): 041104-1041104-3.
[31] L. Q. Ren, X. Wu, M. Li, X. W. Zhang, L. Y. Liu, and L. Xu, Ultrasensitive label-free coupled optofluidic ring laser sensor, Optics Letters, 2012, 37(18): 38733875.
[32] M. Li, X. Wu, L. Y. Liu, X. D. Fan, and L. Xu, Self-Referencing optofluidic ring resonator sensor for highly sensitive biomolecular detection,Analytical Chemistry, 2013, 85(19): 93289332.
[33] W. Liang, Y. Y. Huang, Y. Xu, R. K. Lee, and A.
Yariv, Highly sensitive fiber Bragg grating refractive index sensors, Applied Physics Letters, 2005, 86(15): 151122-1151122-3.
[34] F. Xu, G. Brambilla, and Y. Q. Lu, A microfluidic refractometric sensor based on gratings in optical fibre microwires, Optics Express, 2009, 17(23): 2086620871.
[35] J. Wu, D. Day, and M. Gu, A microfluidic refractive index sensor based on an integrated three
dimensional photonic crystal, Applied Physics Letters, 2008, 92(7): 071108-1071107-3. [36] A. P. Zhang, G. Yan, S. Gao, S. He, B. Kim, J. Im, et al., Microfluidic refractive-index sensors based on small-hole microstructured optical fiber Bragg gratings, Applied Physics Letters, 2011, 98(22): 221109-1221109-3.
[37] Y. Wang, S. Meng, Y. Liang, L. Li, and W. Peng, Fiber-optic surface plasmon resonance sensor with multi-alternating metal layers for biological measurement, Photonic Sensors, 2013, 3(3): 202207.
[38] J. Zhu, L. Qin, S. Song, J. Zhong, and S. Lin, Design of a surface plasmon resonance sensor based on grating connection, Photonic Sensors, 2015, 5(2): 159165.
[39] T. J. Huang, M. I. Lapsley, S. C. S. Lin, and X. L. Mao, An in-plane, variable optical attenuator using a fluid-based tunable reflective interface, Applied Physics Letters, 2009, 95(8): 083507-1083507-3.
[40] E. Weber and M. J. Vellekoop, Optofluidic micro-sensors for the determination of liquid concentrations, Lab on a Chip, 2012, 12(19): 37543759.
[41] D. Qin, Y. Xia, and G. M. Whitesides, Soft lithography for micro- and nanoscale patterning, Nature Protocols, 2010, 5(3): 491502.
You have requested "on-the-fly" machine translation of selected content from our databases. This functionality is provided solely for your convenience and is in no way intended to replace human translation. Show full disclaimer
Neither ProQuest nor its licensors make any representations or warranties with respect to the translations. The translations are automatically generated "AS IS" and "AS AVAILABLE" and are not retained in our systems. PROQUEST AND ITS LICENSORS SPECIFICALLY DISCLAIM ANY AND ALL EXPRESS OR IMPLIED WARRANTIES, INCLUDING WITHOUT LIMITATION, ANY WARRANTIES FOR AVAILABILITY, ACCURACY, TIMELINESS, COMPLETENESS, NON-INFRINGMENT, MERCHANTABILITY OR FITNESS FOR A PARTICULAR PURPOSE. Your use of the translations is subject to all use restrictions contained in your Electronic Products License Agreement and by using the translation functionality you agree to forgo any and all claims against ProQuest or its licensors for your use of the translation functionality and any output derived there from. Hide full disclaimer
Photonic Sensors is a copyright of Springer, 2017.
Abstract
We demonstrate a novel optofluidic refractive index (RI) sensor with high sensitivity and wide dynamic range based on partial reflection. Benefited from the divergent incident light and the output fibers with different tilting angles, we have achieved highly sensitive RI sensing in a wide range from 1.33 to 1.37. To investigate the effectiveness of this sensor, we perform a measurement of RI with a resolution of ca. 5.0×10 -5 refractive index unit (RIU) for ethylene glycol solutions. Also, we have measured a series of liquid solutions by using different output fibers, achieving a resolution of ca. 0.52 mg/mL for cane surge. The optofluidic RI sensor takes advantage of the high sensitivity, wide dynamic range, small footprint, and low sample consumption, as well as the efficient fluidic sample delivery, making it useful for applications in the food industry.
You have requested "on-the-fly" machine translation of selected content from our databases. This functionality is provided solely for your convenience and is in no way intended to replace human translation. Show full disclaimer
Neither ProQuest nor its licensors make any representations or warranties with respect to the translations. The translations are automatically generated "AS IS" and "AS AVAILABLE" and are not retained in our systems. PROQUEST AND ITS LICENSORS SPECIFICALLY DISCLAIM ANY AND ALL EXPRESS OR IMPLIED WARRANTIES, INCLUDING WITHOUT LIMITATION, ANY WARRANTIES FOR AVAILABILITY, ACCURACY, TIMELINESS, COMPLETENESS, NON-INFRINGMENT, MERCHANTABILITY OR FITNESS FOR A PARTICULAR PURPOSE. Your use of the translations is subject to all use restrictions contained in your Electronic Products License Agreement and by using the translation functionality you agree to forgo any and all claims against ProQuest or its licensors for your use of the translation functionality and any output derived there from. Hide full disclaimer