© 2017 Lin et al. This is an open access article distributed under the terms of the Creative Commons Attribution License: http://creativecommons.org/licenses/by/4.0/ (the “License”), which permits unrestricted use, distribution, and reproduction in any medium, provided the original author and source are credited. Notwithstanding the ProQuest Terms and Conditions, you may use this content in accordance with the terms of the License.

Abstract

Despite measures to reduce disease transmission, a risk can occur when blood glucose meters (BGMs) are used on multiple individuals or by caregivers assisting a patient. The laboratory and in-clinic performance of a BGM system before and after disinfection should be demonstrated to guarantee accurate readings and reliable control of blood glucose (BG) for patients. In this study, an effective disinfection procedure, conducting wiping 10 times to assure a one minute contact time of the disinfectant on contaminated surface, was first demonstrated using test samples of the meter housing materials, including acrylonitrile butadiene styrene (ABS), polymethyl methacrylate (PMMA), and polycarbonate (PC), in accordance with ISO 15197:2013. After bench studies comprising 10,000 disinfection cycles, the elemental compositions of the disinfected ABS, PMMA, and PC samples were almost the same as in the original samples, as indicated by electron spectroscopy for chemical analysis. Subsequently, the validated disinfection procedure was then directly applied to disinfect 5 commercial BGM systems composed of ABS, PMMA, or PC to observe the effect of the validated disinfection procedure on meter accuracy. The results of HBsAg values after treatment with HBV sera and disinfectant wipes for each material were less than the LoD of each material of 0.020 IU/mL. Before and after the multiple disinfection cycles, 900 of 900 samples (100%) were within the system accuracy requirements of ISO 15197:2013. All of the systems showed high performance before and after the series of disinfection cycles and met the ISO 15197:2013 requirements. In addition, our results demonstrated multiple cleaning and disinfection cycles that represented normal use over the lifetime of a meter of 3–5 years. Our validated cleaning and disinfection procedure can be directly applied to other registered disinfectants for cleaning commercial BGM products in the future.

Details

Title
Demonstration of disinfection procedure for the development of accurate blood glucose meters in accordance with ISO 15197:2013
Author
Shu-Ping, Lin; Wen-Ye, Lin; Jung-Tzu Chang; Chun-Feng, Chu
First page
e0180617
Section
Research Article
Publication year
2017
Publication date
Jul 2017
Publisher
Public Library of Science
e-ISSN
19326203
Source type
Scholarly Journal
Language of publication
English
ProQuest document ID
1916681564
Copyright
© 2017 Lin et al. This is an open access article distributed under the terms of the Creative Commons Attribution License: http://creativecommons.org/licenses/by/4.0/ (the “License”), which permits unrestricted use, distribution, and reproduction in any medium, provided the original author and source are credited. Notwithstanding the ProQuest Terms and Conditions, you may use this content in accordance with the terms of the License.