Abstract

Background

To compare the measurement of intraocular pressure (IOP) among the three different non-contact tonometers (NCT) and the Goldmann applanation tonometer (GAT) for non-glaucomatous subjects.

Methods

In 52 eyes of 52 non-glaucomatous subjects, IOP was measured sequentially with the Canon TX-20P, the Nidek NT-530P, the Topcon CT-1P, and the GAT at the same time. We evaluated the IOP-measurement agreement among the tonometers as well as the factors affecting the measurements.

Results

A significant positive correlation was shown between the IOP values obtained with GAT and each NCT. The Canon TX-20P showed statistically the most significant agreement with the GAT (ICC 0.906, 95% CI 0.837–0.946). In an analysis of the Bland-Altman plots, the Canon TX-20P also showed the largest mean bias (1.38 mmHg) but the narrowest limits of agreement (LoA) (95% LoA; ± 3.43 mmHg). The Topcon CT-1P showed the smallest mean bias (0.48 mmHg) but the widest LoA (95% LoA; ± 4.16 mmHg). The Topcon CT-1P and Nidek NT-530P both showed a significantly positive correlation between increasing central corneal thickness (CCT) and increasing IOP.

Conclusion

There was a statistically significant correlation between each of the three different NCT and the GAT measurements. IOP measured with the Canon TX-20P and Topcon CT-1P tended to be higher, and with the Nidek NT-530P lower, than with the GAT. Practitioners need to know the properties of their own NCTs and their respective measurement tendencies.

Details

Title
Comparison of intraocular pressure as measured by three different non-contact tonometers and goldmann applanation tonometer for non-glaucomatous subjects
Author
Bang, Seung Pil; Chong Eun Lee; Kim, Yu Cheol
Publication year
2017
Publication date
2017
Publisher
Springer Nature B.V.
e-ISSN
14712415
Source type
Scholarly Journal
Language of publication
English
ProQuest document ID
1959025617
Copyright
Copyright BioMed Central 2017