© 2018 Dong et al. This is an open access article distributed under the terms of the Creative Commons Attribution License: http://creativecommons.org/licenses/by/4.0/ (the “License”), which permits unrestricted use, distribution, and reproduction in any medium, provided the original author and source are credited. Notwithstanding the ProQuest Terms and Conditions, you may use this content in accordance with the terms of the License.

Abstract

Purpose

To compare the axial length (AL), anterior chamber depth (ACD) and intraocular lens power (IOLP) of IOLMaster and Ultrasound in normal, long and short eyes.

Methods

Seventy-four normal eyes (≥ 22 mm and ≤ 25 mm), 74 long eyes (> 25 mm) and 78 short eyes (< 22 mm) underwent AL and ACD measurements with both devices in the order of IOLMaster followed by Ultrasound. The IOLP were calculated using a free online LADAS IOL formula calculator.

Results

The difference in AL and IOLP between IOLMaster and Ultrasound was statistically significant when all three groups were combined. The difference in ACD between IOLMaster and Ultrasound was statistically significant in the normal group (P<0.001) and short eye group (P<0.001) but not the long eye group (P = 0.465). For the IOLP difference between IOLMaster and Ultrasound in the normal group, the percentage of IOLP differences <|0.5|D, ≥|0.5|D<|0.75|D, ≥|0.75|D<|1.0|D, and ≥|1.0|D were 90.5%, 8.1%, 1.4% and 0%, respectively. For the long eye group, they were 90.5%, 5.4%, 4.1% and 0%, respectively. For the short eye group, they were 61.5%, 23.1%, 10.3%, and 5.1%, respectively.

Conclusions

IOLMaster and Ultrasound have statistically significant differences in AL measurements and IOLP (using LADAS formula) for normal, long eye and short eye. The two instruments agree regarding ACD measurements for the long eye group, but differ for the normal and short eye groups. Moreover, the high percentage of IOLP differences greater than |0.5|D in the short eye group is noteworthy.

Details

Title
Comparison of axial length, anterior chamber depth and intraocular lens power between IOLMaster and ultrasound in normal, long and short eyes
Author
Dong, Jing; Zhang, Yaqin; Zhang, Haining; Jia, Zhijie; Zhang, Suhua; Wang, Xiaogang
First page
e0194273
Section
Research Article
Publication year
2018
Publication date
Mar 2018
Publisher
Public Library of Science
e-ISSN
19326203
Source type
Scholarly Journal
Language of publication
English
ProQuest document ID
2014442346
Copyright
© 2018 Dong et al. This is an open access article distributed under the terms of the Creative Commons Attribution License: http://creativecommons.org/licenses/by/4.0/ (the “License”), which permits unrestricted use, distribution, and reproduction in any medium, provided the original author and source are credited. Notwithstanding the ProQuest Terms and Conditions, you may use this content in accordance with the terms of the License.