A measurement concept has been realized for the detection of carbon dioxide, where the CO2 induced pressure generation by an enclosed pH-sensitive hydrogel is measured with a micro pressure sensor. The application of the sensor is the quantification of the partial pressure of CO2 (Pco2) in the stomach as diagnosis for gastrointestinal ischemia. The principle is put to the proof by examining the sensor response to changes in Pco2. Furthermore, the response time, temperature-sensitivity and resolution are determined. The sensor responds well to changes in Pco2 with a maximum pressure generation of 0.29× 105 Pa at 20 kPa CO2. The 90% response time varies between 1.5 and 4.5 minutes at 37[composite function]C. The sensor shows a linear temperature-sensitivity which can easily be compensated for, and enables detection of Pco2 changes as small as 0.5 kPa CO2.

Details

Title
A Miniaturized Carbon Dioxide Gas Sensor Based on Sensing of pH-Sensitive Hydrogel Swelling with a Pressure Sensor
Author
Herber, S; Bomer, J; Olthuis, W; Bergveld, P; van den Berg, A
Pages
197-204
Publication year
2005
Publication date
Sep 2005
Publisher
Springer Nature B.V.
ISSN
13872176
e-ISSN
1572-8781
Source type
Scholarly Journal
Language of publication
English