© 2018 Guidetti et al. This is an open access article distributed under the terms of the Creative Commons Attribution License: http://creativecommons.org/licenses/by/4.0/ (the “License”), which permits unrestricted use, distribution, and reproduction in any medium, provided the original author and source are credited. Notwithstanding the ProQuest Terms and Conditions, you may use this content in accordance with the terms of the License.

Abstract

Purpose

This study aimed to examine the validity, reliability and minimum detectable change (MDC) of the Cosmed K5 in breath by breath (BxB) mode, against VacuMed metabolic simulator. Intra and inter-units reliability was also assessed.

Methods

Fourteen metabolic rates (from 0.9 to 4 L.min-1) were reproduced by a VacuMed system and pulmonary ventilation (VE), oxygen consumption (VO2) and carbon dioxide production (VCO2) were measured by two different K5 units. Validity was assessed by ordinary least products (OLP) regression analysis, Bland-Altman plots, intraclass correlation coefficients (ICC), mean percentage differences, technical errors (TE) and MDC for VE, VO2, and VCO2. Intra- and inter-K5 reliability was evaluated by absolute percentage differences between measurements (MAPE), ICCs, TE, and MDC.

Results

Validity analysis from OLP regression data and Bland- Altman plots indicated high agreement between K5 and simulator. ICC values were excellent for all variables (>0.99). Mean percentage differences in VE (-0.50%, p = 0.11), VO2 (-0.04%, p = 0.80), and VCO2 (-1.03%, p = 0.09) showed no significant bias. The technical error (TE) ranged from 0.73% to 1.34% (VE and VCO2 respectively). MDC were lower than 4% (VE = 2.0%, VO2 = 3.8%, VCO2 = 3.7%). The intra and inter K5 reliability assessment reveled excellent ICCs (>0.99), MAPE <2% (no significant differences between trials), TE < or around 1%, MDC <or around 3%.

Conclusions

K5 in BxB mode is a valid and reliable system for metabolic measurements. This is the first study assessing the MDC accounting only for technical variability reporting intra- and inter-units MDCs <3.3%.

Details

Title
Validity, reliability and minimum detectable change of COSMED K5 portable gas exchange system in breath-by-breath mode
Author
Guidetti, Laura; ⨯ Marco Meucci; Bolletta, Francesco; ⨯ Gian Pietro Emerenziani; Gallotta, Maria Chiara; ⨯ Carlo Baldari ⨯
First page
e0209925
Section
Research Article
Publication year
2018
Publication date
Dec 2018
Publisher
Public Library of Science
e-ISSN
19326203
Source type
Scholarly Journal
Language of publication
English
ProQuest document ID
2161934878
Copyright
© 2018 Guidetti et al. This is an open access article distributed under the terms of the Creative Commons Attribution License: http://creativecommons.org/licenses/by/4.0/ (the “License”), which permits unrestricted use, distribution, and reproduction in any medium, provided the original author and source are credited. Notwithstanding the ProQuest Terms and Conditions, you may use this content in accordance with the terms of the License.