Background: Four oscillometric devices, including the Omron M6 Comfort, Omron HEM-7420, Withings BP-800, and Polygreen KP-7670, designed for self-blood pressure measurement (SBPM) were evaluated according to the European Society of Hypertension (ESH) International Protocol Revision 2010 in four separate studies.

Methods: The four devices measure brachial blood pressure (BP) using the oscillometric method. The Withings BP-800 has to be connected to an Apple® iOS device such as an iPhone®, iPad®, or iPod®. The ESH International Protocol Revision 2010 includes a total number of 33 subjects. The difference between observer and device BP values was calculated for each measure. Ninety-nine pairs of BP differences were classified into three categories (≤5 mmHg, ≤10 mmHg, ≤15 mmHg). The protocol procedures were followed precisely in each of the four studies.

Results: All four tested devices passed the validation process. The mean differences between the device and mercury readings were: −1.8±5.1 mmHg and −0.4±2.8 mmHg for systolic and diastolic BP, respectively, using the Omron M6 Comfort device; 2.5±4.6 mmHg and −1.2±4.3 mmHg for the Omron HEM-7420 device; −0.2±5.0 mmHg and 0.4±4.2 mmHg for the Withings BP-800 device; and 3.0±5.3 mmHg and 0.3±5.2 mmHg for the Polygreen KP-7670 device.

Conclusion: Omron M6 Comfort, Omron HEM-7420, Withings BP-800, and Polygreen KP-7670 readings differing by less than 5 mmHg, 10 mmHg, and 15 mmHg fulfill the ESH International Protocol Revision 2010 requirements, and therefore are suitable for use by patients for SBPM, if used correctly.

Details

Title
Validation of four devices: Omron M6 Comfort, Omron HEM-7420, Withings BP-800, and Polygreen KP-7670 for home blood pressure measurement according to the European Society of Hypertension International Protocol
Author
Topouchian, Jirar; Agnoletti, Davide; Blacher, Jacques; Youssef, Ahmed; Chahine, Mirna N; Ibanez, Isabel; Assemani, Nathalie; Asmar, Roland
Pages
33-44
Section
Original Research
Publication year
2014
Publication date
2014
Publisher
Taylor & Francis Ltd.
ISSN
1176-6344
e-ISSN
1178-2048
Source type
Scholarly Journal
Language of publication
English
ProQuest document ID
2229682226
Copyright
© 2014. This work is licensed under https://creativecommons.org/licenses/by-nc/3.0/ (the “License”). Notwithstanding the ProQuest Terms and Conditions, you may use this content in accordance with the terms of the License.