© 2019 Hsiao et al. This is an open access article distributed under the terms of the Creative Commons Attribution License: http://creativecommons.org/licenses/by/4.0/ (the “License”), which permits unrestricted use, distribution, and reproduction in any medium, provided the original author and source are credited. Notwithstanding the ProQuest Terms and Conditions, you may use this content in accordance with the terms of the License.

Abstract

Background

Most manufacturer manuals do not verify the use of dual energy X-ray absorptiometry for body composition analysis in subjects with a metal implant. This study aimed to quantify the effects of a spinal implant on body composition, and to determine whether unadjusted lean mass estimates are valid for patients with a spinal implant.

Methods

A total of 30 healthy subjects were recruited. Three consecutive scans were performed for each participant, one with and two without extraneous spinal implant, without repositioning between scans. Lean, fat and bone estimates in the total body, trunk and limb were measured.

Results

Precision errors for all total and regional body compositions were within the recommended ranges. Bone masses in the trunk and total body were significantly increased with spinal implant, and the increases exceeded the least significant change. For total and regional lean and fat estimates, the measurements between subjects with and without metal implants were in substantial to almost perfect agreement and the differences were not significant and did not exceed the least significant change.

Conclusions

Spinal metal artifacts significantly increased the total body and trunk bone mass but the differences in lean- and fat-related estimates at total and regional body levels and all estimates in the extremity remained within the clinical acceptable range. Thus, a spinal implant may not compromise screening of patients for fat and lean masses using dual energy X-ray absorptiometry. Application of image reconstruction or a filtering algorithm may help reduce the effect of metallic artifacts and further study is needed.

Details

Title
Does a spinal implant alter dual energy X-ray absorptiometry body composition measurements?
Author
Pei-Lin, Hsiao; Shu-Feng, Hsu; Po-Han, Chen; Hsiao-Wei, Tsai; Hsin-Ying, Lu; Yue-Sheng, Wang; Li-Wen, Lee
First page
e0222758
Section
Research Article
Publication year
2019
Publication date
Sep 2019
Publisher
Public Library of Science
e-ISSN
19326203
Source type
Scholarly Journal
Language of publication
English
ProQuest document ID
2294170324
Copyright
© 2019 Hsiao et al. This is an open access article distributed under the terms of the Creative Commons Attribution License: http://creativecommons.org/licenses/by/4.0/ (the “License”), which permits unrestricted use, distribution, and reproduction in any medium, provided the original author and source are credited. Notwithstanding the ProQuest Terms and Conditions, you may use this content in accordance with the terms of the License.