1. Introduction
Objective refraction remains an integral part of the refraction process, as it allows a nonsubjective way of estimating the magnitude of refractive errors. Over the years, there is evidence and consensus that objective refraction still does not replace subjective refraction [1–8]. There are situations wherein subjective refraction is not applicable such as in young children and toddlers [8] and patients with neurological deficits whose subjective responses are not reliable.
There are also situations where the ocular media have obstructions such as cataracts, vitreous hemorrhage, and hyphema which prevent adequate objective refraction results. In such cases, obtaining adequate total astigmatism is problematic. There may be many of such instances where the cornea may be normal and for that reason keratometry readings may be helpful in estimating the total astigmatism. Javal’s rule attempts this possibility of estimating total astigmatism from keratometry reading [9, 10]. However, this rule has been modified by Grosvenor et al. [10] based on evidence from research into a simplified version [10]. Several versions of Javal’s rule also exist including Sutcliff’s rule and O’Shea’s rule [10]. A common weakness of these modified versions is their relative complexity lending them to less usage in clinical practice [10].
The proponents for the use of Javal’s rule or the simplified version by Grosvenor et al. [10] highlight its benefits in estimating total astigmatism in patients with problems with communication and senile lens changes wherein retinoscopy and autorefraction are unreliable [9, 10]. For instance, some researchers studied the lens as a potential source of astigmatism but found that it was not [9]. However, others found a very small correlation between lenticular astigmatism and cataracts [9].
Elliott et al. [9] reported that Javal’s rule is of limited clinical value compared to autorefraction and retinoscopy in estimating total or refractive spectacle astigmatism [4]. However, they came to this conclusion after comparing Javal’s original rule and its Grosvenor et al. [10] version in estimating spectacle astigmatism with subjective refraction refined by the Jackson cross cylinder in their study with findings of other studies which found much more accurate prediction of spectacle astigmatism with retinoscopy and autorefraction [9, 11, 12]. One could undoubtedly argue that a different conclusion could have been reached if autorefraction and retinoscopy were done in the Elliott et al. [9] study on the same subjects rather than comparing their findings with those of other studies.
In this study, autorefraction and retinoscopy are compared with each other and their agreement with subjective refraction spherical equivalent and total astigmatism. Also Javal’s original rule, Grosvenor et al. version, retinoscopy, and autorefraction estimation of total astigmatism were compared with subjective refraction to determine which method was the best in estimating spectacle total astigmatism. Furthermore, autorefraction and retinoscopy prediction of the cylinder axis were compared with subjective refraction to determine which method is the best in predicting the axis of spectacle astigmatism.
2. Methods
This was a hospital based cross-sectional study. Participants were purposely sampled. All participants were patients visiting the Komfo Anokye Teaching Hospital Eye Center. Participants were examined at the eye center’s general clinic by a specialist ophthalmologist or consultant ophthalmologist after which they were referred to the refraction clinic. Participants were included in the study if they had no significant ocular pathology. All subjects had healthy corneas. Informed consent was sought from the participants and each participant underwent thorough refractive assessment.
In order to avoid bias, noncycloplegic retinoscopy was performed by an optometrist who was masked to the results of the autorefraction and autokeratometry findings. Noncycloplegic autorefraction and autokeratometry were performed using the same Humphrey Zeiss 599 autorefractor/keratometer. Humphrey Zeiss 599 autorefractor/keratometer is a reliable and accurate instrument which is commercially available, often utilized in clinical practice and in several clinical studies [13, 14]. All subjects underwent subjective refraction with refinement of the cylinder power and axis with a ±0.25 Jackson cross cylinder. Participants who were finally included in the study had their two meridians within 15° of 90 and 180.
After autokeratometry reading, Javal’s original rule (1.2 (keratometric astigmatism) −0.50 × 90) and Grosvenor et al. modified Javal’s rule (keratometric astigmatism −0.50 × 90) were used to determine refractive or total astigmatism. Ethical clearance for this study was obtained from the Department of Optometry Ethics Review Committee. All procedures and protocol for the study were conducted in accordance with the tenets of the declaration of Helsinki.
2.1. Statistical Analysis
All statistical analyses were performed using SPSS V.21.0 (SPSS, Chicago, IL, USA) statistical package and
Hence, in this study if the intraclass correlation between the left and right eyes was ≥0.75 then the right eye was arbitrarily selected for the analysis. The agreement between techniques or methods was determined using the intraclass correlation coefficient and Bland-Altman analysis. If there existed proportional bias, the natural logarithm of the values of the techniques was used in the Bland-Altman analysis. All comparisons between techniques with
3. Results
The study included 36 subjects of whom 14 were males and 22 were females. The mean age (SD) for the entire sample was
Table 1
Intraclass correlation between right and left eyes for various parameters.
Parameter | Intraclass correlation coefficient | Confidence interval | |
Autorefraction spherical equivalent | 0.934 | 0.871–0.966 | |
Autorefraction total astigmatism | 0.921 | 0.845–9.960 | |
Retinoscopy spherical equivalent | 0.929 | 0.861–0.964 | |
Retinoscopy total astigmatism | 0.962 | 0.926–0.984 | |
Subjective refraction spherical equivalent | 0.968 | 0.937–0.977 | |
Subjective refraction total astigmatism | 0.952 | 0.913–0.977 | |
Total astigmatism by Javal’s rule | 0.783 | 0.562–0.892 | |
Total astigmatism by Grosvenor’s version of Javal’s rule | 0.783 | 0.562–0.892 | |
Table 2
The Bland-Altman 95% limits agreement and intraclass correlation coefficient for spherical equivalent.
Parameter | Mean difference (standard deviation) | Bland-Altman | | ICC (95% confidence interval) |
Autorefraction versus retinoscopy | 0.26 (1.67) | −3.01 to 3.53 | 0.348 | 0.902 (0.809–0.950) |
Subjective versus retinoscopy | 0.12 (0.51) | −0.88 to 1.12 | 0.157 | 0.989 (0.978–0.994) |
Subjective versus autorefraction | 0.37 (1.64) | −2.84 to 3.58 | 0.167 | 0.895 (0.796–0.947) |
ICC = intraclass correlation coefficient.
Table 3
The Bland-Altman 95% limits agreement and intraclass correlation coefficient for total astigmatism.
Parameter | Mean difference (standard deviation) | Bland-Altman | | ICC |
Autorefraction versus subjective | 0.19 (0.54) | −0.87 to 1.25 | 0.039 | 0.92 (0.84–0.96) |
Retinoscopy versus subjective | 0.25 (0.89) | −1.49 to 1.99 | 0.1 | 0.85 (0.699–0.921) |
Javal’s rule versus subjective | 0.6 (0.68) | −0.73 to 1.93 | <0.001 | 0.82 (0.23–0.93) |
Grosvenor’s modification versus subjective | 0.4 (0.66) | −0.89 to 1.7 | 0.01 | 0.85 (0.6–0.933) |
Autorefraction versus retinoscopy | 0.06 (0.62) | −0.6 to 1.2 | 0.5 | 0.92 (0.857–0.963) |
Table 4
Agreement between cylinder axis of subjective refraction and that of retinoscopy or autorefraction.
Parameter | Kappa | |
Autorefraction versus subjective (exact) | 0.051 | 0.38 |
Autorefraction versus subjective (within 5 degrees) | 0.107 | 0.26 |
Retinoscopy versus subjective (exact) | 0.64 | <0.001 |
Retinoscopy versus subjective (within 5 degrees) | 0.64 | <0.001 |
4. Discussion
In this study, Humphrey Zeiss 599 autorefractor/keratometer was utilized in obtaining the autorefraction and keratometry readings. This particular instrument has been utilized in several clinical studies and also shown to provide accurate and reliable readings [13, 14]. Its test retest reliability (ICC) exceeds 0.95 which makes it a suitable instrument for this study [13].
There was agreement between subjective refraction, retinoscopy, and autorefraction for the spherical equivalent as the mean differences between subjective refraction and either technique were not statistically different from zero. This implies there was agreement between the techniques for the spherical equivalent as further evidenced by the high intraclass coefficient between techniques. The ICC values are traditionally interpreted as follows: ICC of 0.4 indicates poor agreement; ICC from 0.4 to 0.75 indicates fair to good agreement; and ICC greater than 0.75 indicates good to exceptional agreement [16, 17]. Also, lower 95% limits of agreement indicate stronger agreement, provided the one-sample
There was also agreement between retinoscopy and autorefraction for the total astigmatism as evidenced by the statistically insignificant mean difference between the techniques and the high ICC between techniques. This indicates that autorefraction might be a suitable substitute objective refractive technique for a starting point in subjective refraction.
For estimating, total astigmatism no technique had a mean difference that was statistically insignificant except retinoscopy. This indicates only retinoscopy produced total astigmatism that substantially agreed with subjective refraction total astigmatism. All other techniques including autorefraction, Javal’s rule, and Grosvenor modification did not exceptionally agree with subjective refraction in terms of the total astigmatism. Notwithstanding, all other techniques had ICC greater than 0.75 indicating some fair agreement with subjective refraction. Hence, they may be suitable alternative techniques in estimating the crude magnitude of astigmatism as a starting point in subjective refraction.
Autorefraction seems to have a higher intraclass correlation coefficient with subjective refraction than retinoscopy in estimating total astigmatism. Notwithstanding, the mean differences between total astigmatism on subjective refraction and retinoscopy were not statistically different from zero but those of autorefraction and subjective refraction were statistically different from zero. Though autorefraction had a lower mean difference than retinoscopy, the one-sample
The current study findings are consistent with those of previous studies that found agreement between autorefraction and retinoscopy [18–23]. This study showed that even though autorefraction predicted the magnitude of refraction well, it was not particularly good in estimating the axis of the cylinder. However, retinoscopy had excellent agreement with subjective refraction cylinder axis.
One of the limitations of this study is the relatively small sample size utilized in the study. Notwithstanding, the sample size utilized in this study can be considered adequate since several studies [24–27] of comparison between refraction techniques utilized smaller sample sizes compared to that of the current study.
In summary, the study showed moderate agreement between Javal’s rule and subjective refraction total astigmatism. Autorefraction may be an appropriate starting point for subjective refraction but only retinoscopy may satisfactorily replace subjective refraction when subjective refraction is not applicable.
Competing Interests
The authors have no proprietary interests or conflict of interests related to this submission.
Acknowledgments
The authors are grateful to Mrs. Agnes Asiedu and staff of the Refraction and Low Vision Clinic of the Komfo Anokye Teaching Hospital for their administrative support.
[1] M. Farook, J. Venkatramani, G. Gazzard, A. Cheng, D. Tan, S. M. Saw, "Comparisons of the handheld autorefractor, table-mounted autorefractor, and subjective refraction in Singapore adults," Optometry and Vision Science, vol. 82 no. 12, pp. 1066-1070, DOI: 10.1097/01.opx.0000192344.72997.7c, 2005.
[2] D. J. Salchow, M. E. Zirm, C. Stieldorf, A. Parisi, "Comparison of objective and subjective refraction before and after laser in situ keratomileusis," Journal of Cataract and Refractive Surgery, vol. 25 no. 6, pp. 827-835, DOI: 10.1016/S0886-3350(99)00033-4, 1999.
[3] M. A. Bullimore, R. E. Fusaro, C. W. Adams, "The repeatability of automated and clinician refraction," Optometry and Vision Science, vol. 75 no. 8, pp. 617-622, DOI: 10.1097/00006324-199808000-00028, 1998.
[4] M. Zhang, R. Zhang, M. He, W. Liang, X. Li, L. She, Y. Yang, G. MacKenzie, J. D. Silver, L. Ellwein, B. Moore, N. Congdon, "Self correction of refractive error among young people in rural China: results of cross sectional investigation," British Medical Journal, vol. 343 no. 7820,DOI: 10.1136/bmj.d4767, 2011.
[5] M. He, N. Congdon, G. MacKenzie, Y. Zeng, J. D. Silver, L. Ellwein, "The child self-refraction study: results from urban Chinese children in Guangzhou," Ophthalmology, vol. 118 no. 6, pp. 1162-1169, DOI: 10.1016/j.ophtha.2010.10.003, 2011.
[6] J. Jorge, A. Queirós, J. B. Almeida, M. A. Parafita, "Retinoscopy/autorefraction: which is the best starting point for a noncycloplegic refraction?," Optometry and Vision Science, vol. 82 no. 1, pp. 64-68, 2005.
[7] Y.-F. Choong, A.-H. Chen, P.-P. Goh, "A comparison of autorefraction and subjective refraction with and without cycloplegia in primary school children," American Journal of Ophthalmology, vol. 142 no. 1, pp. 68-74, DOI: 10.1016/j.ajo.2006.01.084, 2006.
[8] S. Prabakaran, M. Dirani, A. Chia, G. Gazzard, Q. Fan, S.-W. Leo, Y. Ling, K.-G. Au Eong, T.-Y. Wong, S.-M. Saw, "Cycloplegic refraction in preschool children: comparisons between the hand-held autorefractor, table-mounted autorefractor and retinoscopy," Ophthalmic and Physiological Optics, vol. 29 no. 4, pp. 422-426, DOI: 10.1111/j.1475-1313.2008.00616.x, 2009.
[9] M. Elliott, M. G. Callender, D. B. Elliott, "Accuracy of javal’s rule in the determination of spectacle astigmatism," Optometry and Vision Science, vol. 71 no. 1, pp. 23-26, DOI: 10.1097/00006324-199401000-00005, 1994.
[10] T. Grosvenor, S. Quintero, D. M. Perrigin, "Predicting refractive astigmatism: a suggested simplification of Javal's rule," American Journal of Optometry and Physiological Optics, vol. 65 no. 4, pp. 292-297, DOI: 10.1097/00006324-198804000-00009, 1988.
[11] L. N. Davies, E. A. H. Mallen, J. S. Wolffsohn, B. Gilmartin, "Clinical evaluation of the Shin-Nippon NVision-K 5001/Grand Seiko WR-5100K autorefractor," Optometry and Vision Science, vol. 80 no. 4, pp. 320-324, DOI: 10.1097/00006324-200304000-00011, 2003.
[12] G. Cleary, D. J. Spalton, P. M. Patel, P.-F. Lin, J. Marshall, "Diagnostic accuracy and variability of autorefraction by the tracey visual function analyzer and the shin-nippon N Vision-K 5001 in relation to subjective refraction," Ophthalmic and Physiological Optics, vol. 29 no. 2, pp. 173-181, DOI: 10.1111/j.1475-1313.2008.00627.x, 2009.
[13] S. J. Isenberg, M. D. Signore, G. Madani-Becker, "Use of the HARK autorefractor in children," American Journal of Ophthalmology, vol. 131 no. 4, pp. 438-441, DOI: 10.1016/S0002-9394(00)00861-8, 2001.
[14] J. Cooper, K. Citek, J. M. Feldman, "Comparison of refractive error measurements in adults with Z-View aberrometer, Humphrey autorefractor, and subjective refraction," Optometry, vol. 82 no. 4, pp. 231-240, DOI: 10.1016/j.optm.2010.09.013, 2011.
[15] R. A. Armstrong, "Statistical guidelines for the analysis of data obtained from one or both eyes," Ophthalmic and Physiological Optics, vol. 33 no. 1,DOI: 10.1111/opo.12009, 2013.
[16] M. W. Rouse, E. Borsting, P. N. Deland, L. Hyman, M. Hussein, H. Solan, S. Cotter, D. Grisham, L. Press, M. Scheiman, "Reliability of binocular vision measurements used in the classification of convergence insufficiency," Optometry and Vision Science, vol. 79 no. 4, pp. 254-264, 2002.
[17] J. L. Fleiss, The Design and Analysis of Clinical Experiments, 1986.
[18] P. M. Allen, H. Radhakrishnan, D. J. O'Leary, "Repeatability and validity of the PowerRefractor and the Nidek AR600-A in an adult population with healthy eyes," Optometry and Vision Science, vol. 80 no. 3, pp. 245-251, DOI: 10.1097/00006324-200303000-00014, 2003.
[19] R. Du Toit, K. Soong, G. Brian, J. Ramke, "Quantification of refractive error: comparison of autorefractor and focometer," Optometry and Vision Science, vol. 83 no. 8, pp. 582-588, DOI: 10.1097/01.opx.0000230270.39804.54, 2006.
[20] K. Smith, E. Weissberg, T. G. Travison, "Alternative methods of refraction: a comparison of three techniques," Optometry and Vision Science, vol. 87 no. 3, pp. E176-E182, DOI: 10.1097/opx.0b013e3181cf86d6, 2010.
[21] L. N. Davies, E. A. H. Mallen, J. S. Wolffsohn, B. Gilmartin, "Clinical evaluation of the Shin-Nippon NVision-K 5001/Grand Seiko WR-5100 K autorefractor," Optometry and Vision Science, vol. 80 no. 4, pp. 320-324, DOI: 10.1097/00006324-200304000-00011, 2003.
[22] A. L. Sheppard, L. N. Davies, "Clinical evaluation of the Grand Seiko Auto Ref/Keratometer WAM-5500," Ophthalmic and Physiological Optics, vol. 30 no. 2, pp. 143-151, DOI: 10.1111/j.1475-1313.2009.00701.x, 2010.
[23] J. P. Lowery, A. Joachim, R. Olson, J. Peel, N. N. Pearce, "Autorefraction vs. retinoscopy: a comparison of non-cycloplegic measures in a pediatric sample," Journal of Behavioral Optometry, vol. 16 no. 1, article 38, 2005.
[24] J. D. Twelker, D. O. Mutti, "Retinoscopy in infants using a near noncycloplegic technique, cycloplegia with tropicamide 1%, and cycloplegia with cyclopentolate 1%," Optometry and Vision Science, vol. 78 no. 4, pp. 215-222, DOI: 10.1097/00006324-200104000-00010, 2001.
[25] S. Hopkins, G. P. Sampson, P. Hendicott, P. Lacherez, J. M. Wood, "Refraction in children: a comparison of two methods of accommodation control," Optometry and Vision Science, vol. 89 no. 12, pp. 1734-1739, DOI: 10.1097/opx.0b013e318277182c, 2012.
[26] G. Steele, D. Ireland, S. Block, "Cycloplegic autorefraction results in pre-school children using the Nikon Retinomax Plus and the Welch Allyn Suresight," Optometry and Vision Science, vol. 80 no. 8, pp. 573-577, DOI: 10.1097/00006324-200308000-00010, 2003.
[27] B. Vasudevan, K. J. Ciuffreda, K. Meehan, D. Grk, M. Cox, "Comparison of objective refraction in darkness to cycloplegic refraction: a pilot study," Clinical and Experimental Optometry, vol. 99 no. 2, pp. 168-172, DOI: 10.1111/cxo.12367, 2016.
You have requested "on-the-fly" machine translation of selected content from our databases. This functionality is provided solely for your convenience and is in no way intended to replace human translation. Show full disclaimer
Neither ProQuest nor its licensors make any representations or warranties with respect to the translations. The translations are automatically generated "AS IS" and "AS AVAILABLE" and are not retained in our systems. PROQUEST AND ITS LICENSORS SPECIFICALLY DISCLAIM ANY AND ALL EXPRESS OR IMPLIED WARRANTIES, INCLUDING WITHOUT LIMITATION, ANY WARRANTIES FOR AVAILABILITY, ACCURACY, TIMELINESS, COMPLETENESS, NON-INFRINGMENT, MERCHANTABILITY OR FITNESS FOR A PARTICULAR PURPOSE. Your use of the translations is subject to all use restrictions contained in your Electronic Products License Agreement and by using the translation functionality you agree to forgo any and all claims against ProQuest or its licensors for your use of the translation functionality and any output derived there from. Hide full disclaimer
Copyright © 2016 Kofi Asiedu et al. This work is licensed under http://creativecommons.org/licenses/by/4.0/ (the “License”). Notwithstanding the ProQuest Terms and Conditions, you may use this content in accordance with the terms of the License.
Abstract
The aim of the study was to determine the level of agreement between Javal’s rule, autorefraction, retinoscopy, and refractive astigmatism and to determine which technique is the most suitable substitute when subjective refraction is not applicable using a clinical sample. A total of 36 subjects, 14 males and 22 females, were involved in this study. The intraclass correlation coefficients between subjective refraction, autorefraction, and retinoscopy were 0.895 and 0.989, respectively, for the spherical equivalent. The Bland-Altman 95% limits of agreement between subjective refraction and autorefraction; subjective refraction and retinoscopy; and autorefraction and retinoscopy were −2.84 to 3.58, −0.88 to 1.12, and −3.01 to 3.53, respectively, for the spherical equivalent. The intraclass correlation coefficients between spectacle total astigmatism and the following techniques were as follows: retinoscopy (0.85); autorefraction (0.92); Javal’s rule (0.82); and Grosvenor et al. version (0.85). The Bland-Altman 95% limits of agreement between subjective refraction and autorefraction; subjective refraction and retinoscopy; subjective refraction and Javal’s rule; and subjective refraction and Grosvenor et al. version were −0.87 to 1.25, −1.49 to 1.99, −0.73 to 1.93, and −0.89 to 1.7, respectively, for the total astigmatism. The study showed that autorefraction and Javal’s rule may provide a starting point for subjective refraction cylinder power determination but only retinoscopy may satisfactorily replace subjective refraction total astigmatism when subjective refraction is not applicable.
You have requested "on-the-fly" machine translation of selected content from our databases. This functionality is provided solely for your convenience and is in no way intended to replace human translation. Show full disclaimer
Neither ProQuest nor its licensors make any representations or warranties with respect to the translations. The translations are automatically generated "AS IS" and "AS AVAILABLE" and are not retained in our systems. PROQUEST AND ITS LICENSORS SPECIFICALLY DISCLAIM ANY AND ALL EXPRESS OR IMPLIED WARRANTIES, INCLUDING WITHOUT LIMITATION, ANY WARRANTIES FOR AVAILABILITY, ACCURACY, TIMELINESS, COMPLETENESS, NON-INFRINGMENT, MERCHANTABILITY OR FITNESS FOR A PARTICULAR PURPOSE. Your use of the translations is subject to all use restrictions contained in your Electronic Products License Agreement and by using the translation functionality you agree to forgo any and all claims against ProQuest or its licensors for your use of the translation functionality and any output derived there from. Hide full disclaimer
Details

1 Refraction and Low Vision Clinic, Komfo Anokye Teaching Hospital Eye Center, P.O. Box 1934, Kumasi, Ghana; Department of Optometry, School of Allied Health Sciences, College of Health and Allied Science, University of Cape Coast, Cape Coast, Ghana
2 Department of Optometry, School of Allied Health Sciences, College of Health and Allied Science, University of Cape Coast, Cape Coast, Ghana
3 Refraction and Low Vision Clinic, Komfo Anokye Teaching Hospital Eye Center, P.O. Box 1934, Kumasi, Ghana; Ophthalmology Unit, Cape Coast Teaching Hospital, Cape Coast, Ghana