This is an open access article distributed under the Creative Commons Attribution License, which permits unrestricted use, distribution, and reproduction in any medium, provided the original work is properly cited.
1. Introduction
Tracheal sounds have received much attention in recent years and are often used on respiration-related occasions, such as detecting apnea in sedated volunteers [1] and postanesthesia care unit patients [2], screening obstructive sleep apnea (OSA) during wakefulness [3–6] and detecting OSA during sleep [7–10]. OSA, which increases risk of cardiocerebral vascular disease [11], depression [12], diabetes [13], and even traffic accidents [14], is characterized by repetitive upper airway obstruction occurring during sleep [15]. However, the foremost problem is that talking, machine alarms in the postanesthesia care unit, and ambient noises during wakefulness or sleep are still interferences when tracheal sounds are acquired. A quiet environment on these occasions is frequently highly difficult to maintain. Hence, a simple and effective denoising algorithm is needed.
Adaptive filtering (AF), which can deal with various kinds of signals in unknown statistical environment or in nonstationary environment, is a promising method of signal processing in adaptive noise cancellation [16]. It is usually better than a fixed filter designed through conventional methods and has been widely used [17], such as removing artifacts in electroencephalography (EEG) [18, 19], electrocardiography (ECG) [20], impedance cardiography [21], and photoplethysmography (PPG) [22]. However, the accuracy and reliability of the apnea detection algorithm using tracheal sounds after AF have not been examined. In this study, our hypothesis was that AF could effectively improve the quality of recorded tracheal sounds and result in better performance of apnea detection.
2. Materials and Methods
2.1. Experiment Procedure and Data Collection
Forty-six healthy male (
The acquisition procedure of tracheal sounds and nasal flow pressure signals was as follows: The subjects in the supine position were asked to breathe normally under resting states for 2 minutes. Then, the subjects were asked to hold their breath for about 20 sec to simulate an event of apnea. This type of apnea simulation was repeated every 40 sec for 10 times. For every subject, the total duration was about 15 minutes. The procedure above was performed in a quiet environment and noisy environment where a speaker randomly played some TV dramas, respectively. Our pilot study determined the volume of the speaker, which produced missed events of apnea detected by tracheal sounds. The tracheal sounds and polysomnography (PSG) signals from the subjects were recorded simultaneously during the whole experiment.
The tracheal sounds from the subjects were acquired in the supine position, using a microphone (HC4015G-02L25-423, Hong Chang Electronics, Shenzhen, China) encased in a plastic bell. The bell was attached to the subject’s neck by a double-stick disc just below the larynx and above the suprasternal notch, as is shown in Figure 1. The primary microphone encased in the bell provided the tracheal sounds mixed with noise as the primary input for AF. The secondary microphone resting outside the bell provided the ambient noise as the reference input for AF. A diagram of this setup is illustrated in Figure 2. The audio signals were continuously recorded on the computer using the software Adobe Audition CC 2018 (Adobe Inc., San Jose, California, USA) at 22050 Hz.
[figure omitted; refer to PDF][figure omitted; refer to PDF]Each subject was fitted with a series of sensors connected to a PSG (Alice PDx, Amsterdam, Holland) to measure nasal flow pressure, thoracic movement, abdominal movement, etc. PSG is the gold standard for diagnosing sleep apnea but has not been widely used because of the high cost. In the experiment, flow pressure signals and thoracic and abdominal movement were used as the gold standard signals (PSG deriving signals) to determine apnea events. The recorded PSG signals were transferred to the computer. The acoustic and PSG data on the computer were processed and analyzed in Matlab R2017a (MathWorks Inc., Natick, MA). In our study, the absence of nasal flow pressure and thoracic and abdominal movement signals for more than 15 sec was considered an event of apnea.
2.2. The Implementation of Adaptive Filtering
AF is a signal processing method that uses multiple signal sources to produce desired signals. The block diagram of AF is shown in Figure 3. As is shown in Figure 3, there are two inputs: the primary input
In our study, signal
2.3. Evaluation Metrics of the Performance of Adaptive Filtering Algorithm
2.3.1. The Synthetic Signal Simulation and Signal-to-Noise Ratio (SNR) Value Calculation
In this section, a simulation experiment was performed on the synthetic acoustic signal using the NLMS AF algorithm. In the simulation, we mixed the two signals: the 20 sec ambient noise signal
2.3.2. The Performance Presentation of Apnea Detection Using Tracheal Sounds
The detection of apnea was performed using two kinds of tracheal sounds, each in two different ambient noise conditions: the tracheal sounds before and after being processed by the AF algorithm in the quiet environment and the tracheal sounds before and after being processed by AF algorithm in the noisy environment. The apnea detection algorithm has been described in our previous study [2]. The tracheal sounds were initially filtered using a 5th-order Butterworth filter with a passband between 150 and 800 Hz to minimize heart sounds, muscle interference, and high-frequency noise. The processed signal was segmented into windows of 20 ms with 75% overlap between adjacent windows. The logarithm of the tracheal sound variance (log-var) in each window was calculated. Then, using a calculated log-var threshold, an inspiration or expiration was marked when the log-var signal crossed the threshold and lasted for at least 0.5 sec [2]. The absence of inspiration or expiration for more than 15 sec was considered an event of apnea. Figure 4 showed the identified apneic periods based on PSG deriving signal (nasal flow pressure signal) and tracheal sounds.
[figures omitted; refer to PDF]
Apnea detection performance using tracheal sounds was assessed by comparing it with apnea events detected by the PSG deriving signals. A true positive (TP) was defined if apnea was detected from the tracheal sounds and the PSG deriving signals. A false negative (FN) was defined if apnea was detected from the PSG deriving signals but not from tracheal sounds. False positive (FP) meant that apnea was detected from the tracheal sounds but not from the PSG deriving signals, and true negative (TN) meant that apnea was detected from neither PSG deriving signals nor tracheal sounds. The amount of time in which neither the acoustic apnea detection method nor the gold standard apnea detection method detected apnea was divided by 15 sec to calculate a value for the number of TN. Sensitivity (
3. Results
Forty-six subjects were included in the data analysis. In the quiet environment, the total monitoring time was 13.3 hours, including 499 events of apnea. In the noisy environment, the total monitoring time was 13.2 hours, including 492 events of apnea. Table 1 showed the demographics of the 46 subjects.
Table 1
Demographic data of 46 subjects.
Min | Max | ||
Male/female | 24/22 | ||
Smoking/no smoking | 2/44 | ||
Age (year) | 18 | 35 | |
Height (cm) | 153 | 186 | |
Weight (kg) | 42 | 105 | |
BMI (kg/m2) | 17.1 | 31.4 | |
Neck circumference | 31 | 47 |
The mixed signal
Table 2
The calculated SNR before and after filtering.
Gi | SNRo (dB) | SNRf (dB) | The improvement of SNR |
1 | -1.35 | 0.55 | 1.9 |
1.5 | -4.87 | 0.86 | 5.73 |
2 | -7.37 | 1.32 | 8.69 |
Table 3 shows that in the quiet environment, the tracheal sounds without and with AF correctly detected apnea 485 times (average length was
Table 3
The number of times apnea was detected by tracheal sounds and the PSG deriving signal in the quiet environment.
PSG deriving signal | ||
Apnea | Normal | |
Tracheal sounds without AF | ||
Apnea | 485 times (TP) | 1 time (FP) |
Normal | 14 times (FN) | 2487 times (TN) |
Tracheal sounds with AF | ||
Apnea | 490 times (TP) | 3 times (FP) |
Normal | 9 times (FN) | 2485 times (TN) |
TP = true positive; FP = false positive; FN = false negative; TN = true negative.
[figure omitted; refer to PDF]Table 4 showed that in the noisy environment, the tracheal sounds without and with AF correctly detected apnea 399 times (average length was
Table 4
The number of times apnea was detected by tracheal sounds and the PSG deriving signal in the noisy environment.
PSG deriving signal | ||
Apnea | Normal | |
Tracheal sounds without AF | ||
Apnea | 399 times (TP) | 70 times (FP) |
Normal | 93 times (FN) | 2230 times (TN) |
Tracheal sounds with AF | ||
Apnea | 450 times (TP) | 59 times (FP) |
Normal | 42 times (FN) | 2241 times (TN) |
TP = true positive; FP = false positive; FN = false negative; TN = true negative.
[figure omitted; refer to PDF]4. Discussion
In this study, tracheal sounds collected from the subjects in the quiet and noisy environments were used to detect apnea. The final results showed that in the quiet environment, tracheal sounds without and with AF had similar apnea detection performance, and in the noisy environment, the NLMS AF algorithm significantly improved the performance of apnea detection using tracheal sounds. The validity of the AF algorithm in apnea detection using tracheal sounds was verified.
According to the difference of the AF algorithm optimization criterion, the AF algorithm can be divided into least mean square (LMS), Recursive Least Square (RLS), NLMS, etc. Compared with the LMS algorithm, the NLMS algorithm maintains a higher rate of convergence no matter whether the input signals are correlative or not, and it provides a better trade-off between convergence speed and computational complexity. Meanwhile, the RLS algorithm was used in our pilot study, and it showed similar apnea detection performance in comparison with the NLMS algorithm. However, the real-time processing for tracheal sounds of the RLS algorithm can be limited because of the complicated calculation, the large storage space, and the long program running time. Therefore, the NLMS algorithm was chosen to process tracheal sounds in our study.
When the tracheal sounds were recorded in the noisy environment, the speaker played the TV show using a constant volume. However, the amplitude of noise is uncontrollable in practice. Hence, we introduced Gi to adjust the proportion of the noise in the synthetic signal
In the quiet environment, the unfiltered and filtered tracheal sounds detected apnea with similar sensitivity, specificity, PPV, and NPV. This was because when we applied the NLMS AF algorithm on the tracheal sounds in the quiet environment, the reference input signal was almost zero which resulted in the desired signals (the filtered tracheal sounds) being basically the same with the unfiltered tracheal sounds. In the noisy environment, the apnea detection sensitivity of the filtered tracheal sounds increased from 81.1% to 91.5% compared with the unfiltered tracheal sounds. The proposed filtering algorithm suppressed the noise in tracheal sounds which might be detected as an inspiration or expiration when apnea occurred, and accordingly, the fewer missed events of apnea (from 93 to 42) resulted in the improvement of sensitivity. Meanwhile, the filtered tracheal sounds obtained higher Cohen’s kappa coefficient (0.877 versus 0.795) in apnea detection compared with unfiltered tracheal sounds. These results suggested that the proposed filtering algorithm provided a better performance in apnea detection using tracheal sounds.
The volunteers in our study held their breath with their own ways to simulate the events of apnea. The PSG data review showed that some volunteers simulated “central” apnea with an absence of respiratory effort, and the other volunteers simulated “obstructive” apnea using the pattern similar with the Mueller maneuver [23] where participants’ attempt at inspiration was made with closed mouth and nose after a forced expiration. This meant that the simulated events of apnea were approximately “obstructive” and “central.” Therefore, we employed our previous tracheal sound apnea detection algorithm. Although it has some limitations in detecting hypopnea and severe airway obstruction and in diagnosing the severity of apnea, the algorithm proved to be reliable and accurate in detecting complete obstructive and central apnea during anesthesia [1, 2].
In these previous studies [1, 2], noise in tracheal sounds was detected as a normal breath when actual apnea appeared. Persistent noise in tracheal sounds raised the log-var threshold and resulted in false apnea alarms. According to the results in this study, although the secondary analysis of these previous studies cannot be conducted because of the lack of a reference signal, we feel that the apnea detection performance in these previous studies would further improve if the proposed AF algorithm was performed. Apnea was defined as a cessation of breath longer than 15 sec in our study. The criterion for apnea is also from our previous studies [1, 2]. Although the sleep-related breathing disorder studies usually used 10 sec as the threshold, the conclusion of our study would not change, especially for the validity of the AF algorithm.
Although many approaches have been investigated in screening OSA during wakefulness using tracheal sounds [3–6], none of these studies developed its algorithm in the noisy environment. In these studies, a considerable proportion (1/3 to 1/2) of tracheal sound data from the subjects was excluded from analysis because of noise contamination, which narrowed the application of these algorithms. Hence, we feel that these studies would achieve more application scenarios and less application restrictions if the proposed AF algorithm was introduced.
In our experiment, the quiet environment was just defined by the researchers subjectively and may contain some inherent noise actually. We did not provide a quantitative description of a quiet environment. Therefore, SNRo and SNRf may not be absolutely accurate. However, we feel the conclusion from SNR comparison will not change because the NLMS AF algorithm could suppress the noise in tracheal sounds and improve the tracheal sounds quality under different noise levels. A quantitative description of noisy environment was also not provided. The noise was generated using a speaker which could lead to missed events of apnea detected by tracheal sounds. In practice, the type and level of noise might be different from our experiment. Hence, validating and optimizing the AF algorithm in real-life medical settings may be necessary to improve the quality of the recorded tracheal sounds and result in better performance of apnea detection.
5. Conclusions
The feasibility and validity of the NLMS AF algorithm which was applied on tracheal sounds have been verified under different noise levels. This filtering algorithm can effectively improve the quality of the recorded tracheal sounds in the noisy environment and result in better performance of apnea detection. Thus, the AF technology should be integrated into the application of real-time respiratory monitoring using tracheal sounds.
Authors’ Contributions
Yanan Wu and Jing Liu contributed equally to the work.
Acknowledgments
This work was supported by grant No.81401485 from the National Natural Science Foundation of China, Beijing, China, and grant No.20180550912 from the Department of Science and Technology of Liaoning Province, Shenyang, China.
[1] L. Yu, C. K. Ting, B. E. Hill, J. A. Orr, L. M. Brewer, K. B. Johnson, T. D. Egan, D. R. Westenskow, "Using the entropy of tracheal sounds to detect apnea during sedation in healthy nonobese volunteers," Anesthesiology, vol. 118 no. 6, pp. 1341-1349, DOI: 10.1097/ALN.0b013e318289bb30, 2013.
[2] J. Liu, C. Ai, B. Zhang, Y. Wang, L. M. Brewer, C. K. Ting, D. Huang, L. Yu, "Tracheal sounds accurately detect apnea in patients recovering from anesthesia," Journal of Clinical Monitoring and Computing, vol. 33 no. 3, pp. 437-444, DOI: 10.1007/s10877-018-0192-6, 2019.
[3] M. Aman, G. Eleni, M. Zahra, "Assessment of obstructive sleep apnea and its severity during wakefulness," Annals of Biomedical Engineering, vol. 40 no. 4, pp. 916-924, DOI: 10.1007/s10439-011-0456-5, 2012.
[4] A. Elwali, Z. Moussavi, "Obstructive Sleep Apnea Screening and Airway Structure Characterization During Wakefulness Using Tracheal Breathing Sounds," Annals of Biomedical Engineering, vol. 45 no. 3, pp. 839-850, DOI: 10.1007/s10439-016-1720-5, 2017.
[5] A. Elwali, Z. Moussavi, "A novel decision making procedure during wakefulness for screening obstructive sleep apnea using anthropometric information and tracheal breathing sounds," Scientific Reports, vol. 9 no. 1, 2019.
[6] F. Hajipour, M. J. Jozani, A. Elwali, Z. Moussavi, "Regularized logistic regression for obstructive sleep apnea screening during wakefulness using daytime tracheal breathing sounds and anthropometric information," Medical & Biological Engineering & Computing, vol. 57 no. 12, pp. 2641-2655, DOI: 10.1007/s11517-019-02052-4, 2019.
[7] A. Yadollahi, E. Giannouli, Z. Moussavi, "Sleep apnea monitoring and diagnosis based on pulse oximetery and tracheal sound signals," Medical & Biological Engineering & Computing, vol. 48 no. 11, pp. 1087-1097, DOI: 10.1007/s11517-010-0674-2, 2010.
[8] K. Christoph, E. Manuel, R. Stefan, K. Cornelia, R. Wolfgang, B. Rainer, "Apnea and heart rate detection from tracheal body sounds for the diagnosis of sleep-related breathing disorders," Medical & Biological Engineering & Computing, vol. 56 no. 4, pp. 671-681, DOI: 10.1007/s11517-017-1706-y, 2018.
[9] M. Glos, A. Sabil, K. S. Jelavic, G. Baffet, C. Schöbel, I. Fietze, T. Penzel, "Tracheal sound analysis for detection of sleep disordered breathing," Somnologie, vol. 23 no. 2, pp. 80-85, DOI: 10.1007/s11818-019-0200-1, 2019.
[10] A. Sabil, M. Glos, A. Günther, C. Schöbel, C. Veauthier, I. Fietze, T. Penzel, "Comparison of Apnea Detection Using Thermal Airflow Sensor, Nasal Pressure Transducer, Respiratory Inductanceplethysmography and Tracheal Sound Sensor," Journal of Clinical Sleep Medicine, vol. 15 no. 2, pp. 285-292, DOI: 10.5664/jcsm.7634, 2019.
[11] G. R. Geovanini, R. Wang, J. Weng, N. S. Jenny, S. Shea, M. Allison, P. Libby, S. Redline, "Association between Obstructive Sleep Apnea and Cardiovascular Risk Factors: Variation by Age, Sex, and Race. The Multi-Ethnic Study of Atherosclerosis," Annals of the American Thoracic Society, vol. 15 no. 8, pp. 970-977, DOI: 10.1513/AnnalsATS.201802-121OC, 2018.
[12] A. S. Tubbs, W. Khader, F. Fernandez, M. A. Grandner, "The common denominators of sleep, obesity, and psychopathology," Current Opinion in Psychology, vol. 34, pp. 84-88, DOI: 10.1016/j.copsyc.2019.11.003, 2019.
[13] T. Huang, B. M. Lin, M. J. Stampfer, S. S. Tworoger, F. B. Hu, S. Redline, "A Population-Based Study of the Bidirectional Association Between Obstructive Sleep Apnea and Type 2 Diabetes in Three Prospective U.S. Cohorts," Diabetes Care, vol. 41 no. 10, pp. 2111-2119, DOI: 10.2337/dc18-0675, 2018.
[14] M. W. Purtle, C. H. Renner, D. A. McCann, J. C. Mallen, S. K. Spilman, S. M. Sahr, "Driving with undiagnosed obstructive sleep apnea (OSA): High prevalence of OSA risk in drivers who experienced a motor vehicle crash," Traffic Injury Prevention, vol. 21 no. 1, pp. 38-41, DOI: 10.1080/15389588.2019.1709175, 2020.
[15] R. B. Berry, R. Budhiraja, D. J. Gottlieb, D. Gozal, C. Iber, V. K. Kapur, C. L. Marcus, R. Mehra, S. Parthasarathy, S. F. Quan, S. Redline, K. P. Strohl, S. L. Davidson Ward, M. M. Tangredi, American Academy of Sleep Medicine, "Rules for scoring respiratory events in sleep: update of the 2007 AASM Manual for the Scoring of Sleep and Associated Events. Deliberations of the Sleep Apnea Definitions Task Force of the American Academy of Sleep Medicine," Journal of Clinical Sleep Medicine, vol. 8 no. 5, pp. 597-619, DOI: 10.5664/jcsm.2172, 2012.
[16] J. Yadav, M. Kumar, R. Saxena, A. K. Jaiswal, "Performanceanalysis of LMS FIRfilter and RLS FIRfilter for Cancellation," Signal & Image Processing : An International Journal, vol. 4 no. 3, pp. 45-56, DOI: 10.5121/sipij.2013.4304, 2013.
[17] L. Dos Santos, J. J. Barroso, E. E. Macau, M. F. de Godoy, "Application of an automatic adaptive filter for Heart Rate Variability analysis," Medical Engineering & Physics, vol. 35 no. 12, pp. 1778-1785, DOI: 10.1016/j.medengphy.2013.07.009, 2013.
[18] V. Mihajlovic, S. Patki, B. Grundlehner, "The impact of head movements on EEG and contact impedance: an adaptive filtering solution for motion artifact reduction," Conference Proceedings: Annual International Conference of the IEEE Engineering in Medicine and Biology Society, vol. 2014, pp. 5064-5467, 2014.
[19] X. Navarro, F. Porée, A. Beuchée, G. Carrault, "Denoising preterm EEG by signal decomposition and adaptive filtering: a comparative study," Medical Engineering & Physics, vol. 37 no. 3, pp. 315-320, DOI: 10.1016/j.medengphy.2015.01.006, 2015.
[20] Y. Gong, T. Yu, B. Chen, M. He, Y. Li, "Removal of cardiopulmonary resuscitation artifacts with an enhanced adaptive filtering method: an experimental trial," BioMed Research International, vol. 2014 no. 1,DOI: 10.1155/2014/140438, 2014.
[21] V. K. Pandey, P. C. Pandey, N. J. Burkule, L. R. Subramanyan, "Adaptive filtering for suppression of respiratory artifact in impedance cardiography," Conference Proceedings: Annual International Conference of the IEEE Engineering in Medicine and Biology Society, vol. 2011, pp. 7932-7936, 2011.
[22] F. Peng, Z. Zhang, X. Gou, H. Liu, W. Wang, "Motion artifact removal from photoplethysmographic signals by combining temporally constrained independent component analysis and adaptive filter," Biomedical Engineering Online, vol. 13 no. 1,DOI: 10.1186/1475-925X-13-50, 2014.
[23] B. D. Borowiecki, J. F. Sassin, "Surgical Treatment of Sleep Apnea," Archives of Otolaryngology, vol. 109 no. 8, pp. 508-512, DOI: 10.1001/archotol.1983.00800220014004, 1983.
You have requested "on-the-fly" machine translation of selected content from our databases. This functionality is provided solely for your convenience and is in no way intended to replace human translation. Show full disclaimer
Neither ProQuest nor its licensors make any representations or warranties with respect to the translations. The translations are automatically generated "AS IS" and "AS AVAILABLE" and are not retained in our systems. PROQUEST AND ITS LICENSORS SPECIFICALLY DISCLAIM ANY AND ALL EXPRESS OR IMPLIED WARRANTIES, INCLUDING WITHOUT LIMITATION, ANY WARRANTIES FOR AVAILABILITY, ACCURACY, TIMELINESS, COMPLETENESS, NON-INFRINGMENT, MERCHANTABILITY OR FITNESS FOR A PARTICULAR PURPOSE. Your use of the translations is subject to all use restrictions contained in your Electronic Products License Agreement and by using the translation functionality you agree to forgo any and all claims against ProQuest or its licensors for your use of the translation functionality and any output derived there from. Hide full disclaimer
Copyright © 2020 Yanan Wu et al. This is an open access article distributed under the Creative Commons Attribution License (the “License”), which permits unrestricted use, distribution, and reproduction in any medium, provided the original work is properly cited. Notwithstanding the ProQuest Terms and Conditions, you may use this content in accordance with the terms of the License. http://creativecommons.org/licenses/by/4.0/
Abstract
Objective. Tracheal sounds were used to detect apnea on various occasions. However, ambient noises can contaminate tracheal sounds which result in poor performance of apnea detection. The objective of this paper was to apply the adaptive filtering (AF) algorithm to improve the quality of tracheal sounds and examine the accuracy of the apnea detection algorithm using tracheal sounds after AF. Method. Tracheal sounds were acquired using a primary microphone encased in a plastic bell, and the ambient noises were collected using a reference microphone resting outside the plastic bell in quiet and noisy environments, respectively. Simultaneously, the flow pressure signals and thoracic and abdominal movement were obtained as the standard signals to determine apnea events. Then, the normalized least mean square (NLMS) AF algorithm was applied to the tracheal sounds mixed with noises. Finally, the algorithm of apnea detection was used to the tracheal sounds with AF and the tracheal sounds without AF. Sensitivity, specificity, positive predictive value (PPV), negative predictive value (NPV), accuracy, and Cohen’s kappa coefficient of apnea detection were calculated. Results. Forty-six healthy subjects, aged 18-35 years and with
You have requested "on-the-fly" machine translation of selected content from our databases. This functionality is provided solely for your convenience and is in no way intended to replace human translation. Show full disclaimer
Neither ProQuest nor its licensors make any representations or warranties with respect to the translations. The translations are automatically generated "AS IS" and "AS AVAILABLE" and are not retained in our systems. PROQUEST AND ITS LICENSORS SPECIFICALLY DISCLAIM ANY AND ALL EXPRESS OR IMPLIED WARRANTIES, INCLUDING WITHOUT LIMITATION, ANY WARRANTIES FOR AVAILABILITY, ACCURACY, TIMELINESS, COMPLETENESS, NON-INFRINGMENT, MERCHANTABILITY OR FITNESS FOR A PARTICULAR PURPOSE. Your use of the translations is subject to all use restrictions contained in your Electronic Products License Agreement and by using the translation functionality you agree to forgo any and all claims against ProQuest or its licensors for your use of the translation functionality and any output derived there from. Hide full disclaimer