Introduction
The quantification of trace elements in airborne particulate matter (PM) can be achieved with various techniques, such as inductively coupled plasma mass spectrometry (ICP-MS), inductively coupled plasma optical emission spectrometry (ICP-OES), X-ray fluorescence spectrometry (XRF), and particle-induced X-ray emission spectrometry (PIXE). These methods require a two-step procedure, i.e. sample collection in the field followed by laboratory analysis. Ambient pollutants are conventionally collected on filter substrate for large time duration such as 8 or 24 h sampling time to ensure that sufficient elemental mass is available for analytical analysis (ICP). For high time resolution, impactors are used where the sample is collected on a foil (e.g. a rotating drum impactor; Lundgren, 1967) or on a combination of an impactor plate and a filter, such as in a streaker sampler (e.g. Annegarn et al., 1992). These samples are then exposed to X-rays or a particle beam without further treatment, which provide quantitative data with low detection limits. In contrast to the non-destructive XRF method, sample preparation for ICP analysis is very laborious, and the samples are destroyed during this process. The XRF method has been successfully applied to aerosol characterization in the last decades. Measurement of low sample mass typically requires high sensitivity (or a low minimum detection limit, MDL) and hence access to a synchrotron or accelerator facilities (Bukowiecki et al., 2005, 2008; Calzolai et al., 2010, 2015; Lucarelli et al., 2011; Richard et al., 2010; Visser et al., 2015b; Yatkin et al., 2016), which is notoriously difficult due to the demand for analysis time at such facilities. Access restrictions limit the number of collected samples to be analysed, and hence field campaigns are predominantly episodic. Technical advances in X-ray sources and detectors have recently resulted in the development of commercial systems capable of sampling and analysing ambient PM samples in sub-hourly or hourly resolution in quasi-real time. Instruments of this type can be used for continuous (months, years) monitoring at a site, but their cost may restrict the simultaneous deployment of a larger number of devices for different size fractions or at different sites. The benefit of long-term, quasi-real-time data access, favourable, for example, for air quality monitoring, contrasts with the possibilities of relatively low-cost, multi-site, and multi-size samplers used so far in episodic field studies.
Sampling with high time resolution generates large quantities of data capable of capturing source emission patterns occurring at shorter time duration. For source apportionment of PM components like elements, a high time resolution of the order of 1 h or less is advantageous, as temporally variable environmental factors such as wind direction and speed or insolation may affect transport and source processes, e.g. resuspension (Annegarn et al., 1992; D'Alessandro et al., 2003; Sánchez-Rodas et al., 2007; Sarmiento et al., 2007; Visser et al., 2015b; Yadav and Turner, 2014). One such instrument, Cooper Environmental's Xact® 625 Ambient Metals Monitor, performs in situ automated measurements of ambient PM or PM elemental concentrations for a user-defined set of 24 or more elements with a user-selected sampling time resolution of 15 to 240 min. The instrument is transportable and can be deployed in field campaigns where a suitable shelter with electric power and an appropriate sampling line connecting the outdoor with the indoor is available. Remote access to the data is possible during operation, which allows for a continuous, quasi-real-time (i.e. with a delay of one sampling interval) monitoring of the operation status and the ambient metal content. An in-depth evaluation of the forerunner instrument Xact 620 was previously published by Park et al. (2014).
An Xact 625 monitor was deployed for a month to test the handling and data production of the instrument. A small field campaign was organized at a monitoring station of the Swiss Air Pollution Monitoring Network (NABEL), where quality-controlled air pollution measurements are performed continuously. The NABEL network provided the reference for previous data intercomparisons (Hueglin et al., 2005; Lanz et al., 2010), as well as for the intercomparisons of this study. Comparisons between SR-XRF and filter samples analysed with ICP-OES and ICP-MS have been performed previously (Richard et al., 2010). Comparisons of XRF on samples collected on different substrates were performed, e.g. by Yatkin et al. (2012). A recent interlaboratory comparison of PM filter analysis methods is presented in Yatkin et al. (2016), where XRF/PIXE and ICP methods were compared for several metrics. Some of these metrics are also applied in this study.
The goals of this article are (1) to characterize the measurements of the test period in Härkingen and compare them with previous studies in Switzerland; (2) to examine the achieved data quality for the selected elements with respect to their MDLs; (3) to quantify the measurement quality based on intercomparisons between the Xact and NABEL PM data (1 h tapered element oscillating microbalance (TEOM) data and 24 h filter samples) for Härkingen; (4) to evaluate the applicability of the instrument at high time resolution in typical summer conditions and concentration ranges at a traffic-influenced rural site in Switzerland; and (5) to gauge the advantages of high-time-resolution sampling for a preliminary investigation of sources based on enhancement ratios and diurnal variability of elements. A pollution episode captured during the campaign resulted in high ambient concentrations, widening the range of studied concentrations. The selected elements represented a typical mix of elements at the selected site. In addition, a few elements notoriously difficult to measure in Switzerland due to their generally low ambient concentration were included, namely Ni, As, Pt, and Hg.
Experimental set-up
Site characteristics
The field campaign was performed at the permanent station Härkingen
(47.311877 N, 7.820453 E) of NABEL (
The first week of the campaign was characterized by moderate summer temperatures (maximum temperature < 25 C, except 23 July) with some occasional rain cleansing the atmosphere. The remaining 2 weeks were part of a summer heat wave, with temperatures reaching 36.4 C at the maximum and values above 30 C on 7 days of this period. Only one precipitation event occurred during the hot period. The Swiss National Day (1 August) fell on a Saturday, and the weekend weather promoted outdoor barbecues and fireworks. The bulk of the fireworks were burnt on 1 August after 22:00 LT (local time UTC 2 h), but some individual fireworks were also burnt on 31 July and 2 and 3 August.
For this study, the Xact 625 monitor and a Q-ACSM (quadrupole aerosol chemical speciation monitor, Aerodyne Inc.) were installed in an air-conditioned trailer parked next to the NABEL station. This trailer was placed to the north of the freeway at 23 m away from the centre of the freeway. This placed the trailer on the orthogonal transect between the freeway and NABEL shelter, which is located 27 m from the centre of the freeway. The trailer's instruments were connected to the NABEL station's power grid and Ethernet.
NABEL instrumentation
The NABEL station is equipped with a broad range of air quality instrumentation and standard meteorological sensors. The relevant instruments for this field test were the Digitel DA-80H HiVol sampler with a DPM 10/30/00 inlet for 24 h PM sample collection and a TEOM Filter Dynamics Measurement System (FDMS 8500, Thermo Scientific) for continuous (10 min) PM mass concentration measurements. The time constant used for noise reduction in the TEOM and the averaging procedure caused a significant time delay in peak concentrations. For 1 h values, the random error of a TEOM as derived from parallel operation of two identical instruments is about 2 g m. Standard meteorological measurements such as temperature, wind speed, and direction, and precipitation records are also monitored at this station. Furthermore, the station also provided hourly traffic counts for the freeway in the form of total number of vehicles, number of heavy-duty vehicles (HDV), and number of light-duty vehicles (LDV).
Xact 625
The Xact 625 Ambient Metals Monitor (Cooper Environmental Services (CES), Beaverton, OR, USA) is a sampling and analysing X-ray fluorescence spectrometer designed for online, semi-continuous measurements of elements in aerosols. In this study, ambient air was sampled with a flow rate of 16.7 actual lpm through a PM flow separator (Tisch Environmental, TE-PM10-D) and the PM collected onto a Teflon filter tape. The flow is maintained to within about 1 %. After each sampling interval the filter tape is moved into the analysis area of the spectrometer, where it is illuminated with an X-ray tube with three excitation conditions, and the excited X-ray fluorescence is measured with a silicon drift detector (SDD). During this XRF analysis, the next sample is collected on a clean spot of the filter tape. This cycle is repeated during each sampling interval, which was configured to be 60 min for this study. While this approach is non-destructive, the samples collected are typically not amenable to offline analysis post-sampling due to potential for cross-contamination from sampled filter tape wound upon itself onto the filter wheel. After each analysis interval, raw and calibrated (for the actual volume in units of ng m concentration data were stored on the hard disk of the control unit. Daily advanced quality assurance checks (QA energy calibration test, QA upscale test) were performed during 30 min after midnight to monitor shifts in the calibration. Thus, the sampling interval following midnight was limited to 30 min only.
The instrument was configured to quantify 24 elements (Si, S, Cl, K, Ca, Ti, V, Cr, Mn, Fe, Co, Ni, Cu, Zn, As, Se, Cd, Sn, Sb, Ba, Pt, Hg, Pb, Bi, plus Pd for QA). Each of these elements was calibrated individually with a reference sample. MDLs for 1 h sampling for each element are listed in Table 1. CES calculates MDLs using the sensitivity of the element and the counts in the region of interest of a blank unsampled section of tape, from where 1 interference-free detection limits are reported. XRF-based MDLs are inversely proportional to the square root of the X-ray analysis time (Currie, 1977), which in the case of Xact is the same as the sampling duration. Hence, Xact MDLs are lower for longer sampling durations. Interference-free MDLs, while true are idealized lower limits of detection of one single element. As with most analytical methods, matrix effects in ambient samples from interferences between different elements and analyte concentrations could potentially result in MDLs of ambient samples to be higher and vary across samples, which makes them difficult to generalize and report. It is therefore often preferred to report measurement uncertainties to characterize a measurement. An uncertainty of 5 % or less due to fitting errors and uncertainties in the standards has been derived from laboratory experiments with NIST standards (benchtop XRF, filter analyses). Uncertainties are expected to be higher for concentrations close to the MDL, for elements with potential for line interferences in multi-element samples, and, from self absorption effects, for the lightest elements (Si, S, Cl, K, Ca). Line interference is well-known for element couples like Fe–Co, Pb–As, and Ba–Ti and makes detection of one element difficult if the other is abundant in the sample. The linear least-squares reference deconvolution algorithm implemented in the Xact fits the measured sample spectrum with the library of pure element reference spectra to resolve concentrations of each calibrated element. The Xact reports purely elemental mass concentrations, which are the focus of discussion and unless otherwise noted.
Regression coefficients for the comparison of Xact 625 and offline data. The 1 h values of the Xact 625 were averaged to 24 h values. Primed quantities are uncertainties.
Regression coefficients | Xact | ICP (24 h) | |||||||||||
---|---|---|---|---|---|---|---|---|---|---|---|---|---|
Group | Element | Analysis | fitX | Average | Ratio intercept/ | MDL | Pts > MDL | MDL | Pts > MDL | ||||
method | XRF conc. | average | (60 min) | (24 h) | |||||||||
ng m | ng m | % | ng m | % | |||||||||
A | S | ICP-OES | 170 | 30.6 | 1.37 | 0.03 | 1.00 | 956 | 0.18 | 7.7 | 100 | ||
K | ICP-OES | 52.4 | 19.2 | 1.15 | 0.02 | 1.00 | 703 | 0.07 | 4.20 | 100 | 37.8 | 100 | |
Ca | ICP-OES | 13.9 | 17.9 | 1.45 | 0.06 | 0.99 | 365 | 0.04 | 1.60 | 100 | 49.2 | 90 | |
Ti | ICP-MS | 5.58 | 0.57 | 1.13 | 0.06 | 0.98 | 14.9 | 0.37 | 0.68 | 100 | 1.04 | 90 | |
Mn | ICP-MS | 1.72 | 0.28 | 1.31 | 0.06 | 0.99 | 7.6 | 0.23 | 0.51 | 100 | 0.26 | 100 | |
Fe | ICP-OES | 93.1 | 35.8 | 1.34 | 0.08 | 0.97 | 656 | 0.14 | 1.40 | 100 | 3.4 | 100 | |
Cu | ICP-MS | 4.93 | 1.27 | 1.27 | 0.05 | 0.99 | 33.5 | 0.15 | 0.48 | 100 | 0.055 | 100 | |
Zn | ICP-MS | 5.56 | 2.18 | 1.80 | 0.13 | 0.96 | 23.0 | 0.24 | 0.41 | 100 | 0.96 | 100 | |
Ba | ICP-MS | 1.62 | 1.73 | 1.01 | 0.03 | 0.99 | 31.9 | 0.05 | 1.70 | 94 | 0.82 | 100 | |
Pb | ICP-MS | 0.32 | 0.17 | 0.97 | 0.04 | 0.99 | 3.8 | 0.08 | 0.39 | 98 | 0.22 | 100 | |
average | 1.28 | ||||||||||||
standard deviation | 0.24 | ||||||||||||
B | V | ICP-MS | 0.03 | 0.03 | 0.15 | 0.05 | 0.57 | 0.05 | 0.62 | 0.52 | 2 | 0.026 | 100 |
Cr | ICP-MS | 2.13 | 0.26 | 1.02 | 0.28 | 0.63 | 2.5 | 0.86 | 0.52 | 75 | 0.61 | 40 | |
Co | ICP-MS | 0.01 | 0.02 | 0.18 | 0.29 | 0.05 | 0.02 | 0.55 | 0.57 | 0 | 0.018 | 70 | |
Ni | ICP-MS | 0.82 | 0.14 | 0.16 | 0.11 | 0.22 | 0.63 | 1.31 | 0.40 | 68 | 0.58 | 100 | |
As | ICP-MS | 0.09 | 0.06 | 0.56 | 0.20 | 0.50 | 0.06 | 1.48 | 0.20 | 4 | 0.026 | 100 | |
Se | ICP-MS | 0.00 | 0.15 | 0.76 | 0.41 | 0.30 | 0.26 | 0.01 | 0.25 | 38 | 0.015 | 100 | |
Cd | ICP-MS | 6.28 | 0.28 | 10.4 | 7.9 | 0.18 | 6.5 | 0.96 | 10.3 | 13 | 0.028 | 20 | |
Sn | ICP-MS | 19.7 | 0.96 | 0.36 | 0.21 | 0.27 | 21.3 | 0.93 | 13.3 | 85 | 0.028 | 100 | |
Sb | ICP-MS | 22.2 | 3.58 | 3.9 | 1.63 | 0.42 | 30.6 | 0.72 | 16.0 | 94 | 0.026 | 100 | |
Hg | Au AAA | 0.64 | 0.06 | 24.9 | 24.1 | 0.12 | 0.69 | 0.94 | 0.34 | 87 | 0.001 | 40 | |
Bi | ICP-MS | 0.16 | 0.06 | 0.79 | 0.04 | 0.98 | 0.49 | 0.32 | 0.43 | 7 | 0.015 | 100 | |
C | Si | ||||||||||||
Cl | |||||||||||||
Pt | 0.41 | 2 |
Q-ACSM
A Q-ACSM (Aerodyne Inc., Billerica, MA, USA) was operated in the trailer housing the Xact 625 during the campaign. The Q-ACSM determines quantitative mass spectra of non-refractory (NR) particles up to mass to charge ratios () of 150 (Crenn et al., 2015; Ng et al., 2011). Ion fragments were attributed mainly to organic aerosols, nitrate, sulfate, ammonium, and chloride, which comprise the reported data used in this study. The collection efficiency (CE) was determined for each spectrum according to Middlebrook et al. (2012), and its distribution peaked at the mode of 0.62 (0.11). 293 out of 1055 CE values of the full ACSM dataset were equal to 0.45. The Q-ACSM collected sub-micron (PM particles and chemically analysed them in 30 min intervals, which were aggregated to 1 h averages for comparison with the Xact 625 data. All concentrations used in this study were CE corrected.
24 h PM filter samples
The 24 h PM samples collected by the HiVol sampler on quartz filters were weighed at Empa laboratory in Dübendorf, Switzerland, to determine the gravimetric daily PM concentrations. These values were then used to correct the TEOM PM concentrations on a daily basis. Therefore the 24 h TEOM values correspond to 24 h gravimetric PM. Ten 24 h PM samples were analysed for their elemental composition at IDAEA-CSIC laboratory in Barcelona. A quarter of each filter was acid digested using a mix of HF : HNO (2.5 : 1.25 mL); the solution was kept in a Teflon reactor at 90 C for at least 6 h, and after cooling 2.5 mL of HClO were added. The acid solution was brought to evaporation and the dry residual was re-dissolved with HNO and diluted with Milli-Q water for subsequent ICP-OES and ICP-MS analysis. This method has been validated and used in many studies and is discussed in detail elsewhere (Minguillón et al., 2012; Querol et al., 2001, 2008). A total of 41 elements from Li to U were analysed: the major elements Na, Mg, Al, P, S, K, Ca, Ti, and Fe with ICP-OES and the trace elements with ICP-MS. Si, Cl, and Pt were not analysed on the filters. Analyses of the reference material NIST 1633b (constituent elements in coal fly ash) using the same methodology as that for the samples yielded satisfactory results, with approximately 100 % recoveries for the elements under study. Tests of the used methodology with respect to other ICP sample preparation and analysis methods, and applications of the methodology to NIST standards indicated the reliability of the method, exerting a maximum scatter of 10 % for any of the elements, with most uncertainty values clearly below this limit. Relative uncertainties (precision) of the ICP measurements are less than 5 % for the elements with concentrations well above their respective detection limits, whereas the overall uncertainty reflecting the entire sampling procedure, the digestion and the ICP analysis is on the order of 25 %. MDLs for ICP were determined according to Escrig et al. (2009), and the values for the elements relevant for our intercomparison are listed in Table 1. Hg was analysed with a Hg gold amalgamation atomic absorption analyser (AMA-254, LECO instruments, Botasini et al., 2013; Diez et al., 2007). The three methods are referred to as the offline or ICP methods (ICP-OES, ICP-MS), and the Hg gold amalgamation atomic absorption spectrometry is abbreviated as AuAAA in this paper.
Three of the 10 filters were also analysed with a benchtop XRF system by CES, and with ICP-MS by an independent lab (Eastern Research Group, ERG, Research Triangle Park, NC, USA) to investigate inter-laboratory scatter. ERG used a different digestion method than IDAEA-CSIC. In addition, three filters were prepared with a reference aerosol of known concentration for Fe, Cu, Zn, Sr, and Pb, which then were analysed by CES, IDAEA-CSIC, and ERG, again to gain insight into the inter-laboratory scatter. Details on these data and the methods are given in the supplement of this article.
Main panel: relative amount of analysed elements by Xact 625 during the field campaign. Top panel: absolute concentrations, stacked. The grey shaded area denotes the fireworks period. The red squares mark the days when 24 h filters were analysed and used for comparisons in this study. Bottom panel: relative cumulative elemental concentrations, stacked. Right panel: average relative contributions (in %) of elements for the fireworks period, the non-fireworks period, and the south and north sectors during the non-fireworks period.
[Figure omitted. See PDF]
Data coverage and synchronization
The Xact 625 measurements started on 23 July 2015 12:00 LT and ended on 13 August 2015 06:00 LT. The sampling interval was set to 1 h. Two interruptions occurred during the sampling period: one due to an Xact 625 computer problem (33 h) and the other one due to a delayed filter tape change (10 h). The Xact dataset consists of 456 valid 1 h samples out of 499 possible samples, attaining a coverage of 91.4 %. The NABEL data were tailored to coincide with the Xact data. The 10 min TEOM PM values were aggregated to 1 h values to synchronize them with hourly Xact 625 measurements. Additionally, the TEOM data were also adjusted to the gravimetrically determined PM masses from the HiVol filters to provide an independent reference for intercomparisons. The data used here contained some gaps, which were only partly synchronous for the selected parameters. Wind speed and direction missed 12 data points (2.4 %), precipitation 26 (5.2 %), and PM 53 (10.6 %) at hourly time resolution. The ACSM data contained a gap of 14.5 h due to software malfunction, which caused the data to be very noisy for that short period. These values were rejected from the analyses, and only the remaining 972 data points were averaged to 465 1 h values, which then were resampled to the 456 Xact data points.
For the comparisons of the different instruments and sampling intervals, all data were resampled to the corresponding times of the Xact 625, according to the sub-classifications of the dataset (e.g. according to wind sectors). For the intercomparisons with the 10 filter samples, the hourly Xact data of the corresponding days were averaged to the 24 h of the filter samples. During each 24 h period, Xact generated 23 1 h values and 1 30 min value, which were aggregated to 24 h daily averages. This procedure implicitly assumes that the half-hour sample of the first sampling hour is representative for the whole hour. Tests with a 23.5 h weighted average yielded differences of less than 3 % between the two calculation methods. Comparisons of hourly Xact data were only possible for S with the ACSM data (in the form of PM sulfate, assuming that all S occurs in PM and between the total Xact element mass and PM of the NABEL TEOM instrument; see Sects. 3.2 and 3.3.
Results and discussion
Data validity derived from general statistics and MDLs
The complete Xact dataset is visualized in Fig. 1, and general statistics are given in Table S1 in the Supplement. The salient feature of the concentration time series is the huge peak late on 1 August, caused by the National Day's fireworks episode. Further peaks before and after that day warranted dividing the full dataset into a fireworks period and a non-fireworks period. The fireworks period started on 31 July 2015 22:00 LT and lasted until 4 August 2015 11:00 LT, as will be discussed in more detail in Sect. 3.3. The remaining non-fireworks period is representative for the typical background concentrations at Härkingen and can be compared to literature values.
MDLs for the Xact 625 and for ICP-OES/MS and the Hg AuAAA method are listed in Table 1. Note that MDLs of elements measured by Xact 625 are based on 1 h sampling time while MDLs of filter-based elemental concentrations are based on 24 h. Generally, values below 3 MDL are expected to have much higher uncertainty. Hence, elements with more than 80 % of the data below 3 MDL were rejected from further examination. Xact 625 MDLs have not been determined by the manufacturer for Si, S, and Cl, because self-absorption effects for elements lighter than Ca become more important with decreasing atomic number (Formenti et al., 2010). However, these three elements are abundant, and we assume that they are well above their Xact detection limit. For these elements, an ICP MDL is only given for S, because Si cannot be determined in the filter samples, as it is a main constituent of the quartz filters and is also digested during sample preparation, and Cl cannot be determined by ICP. The table indicates the amount of data points > MDL in percent for the different analysis methods. The elements K, Ca, Ti, Mn, Fe, Cu, Zn, Sn, Sb, Ba, and Pb have most values above the MDL, and their measurement should thus be reliable. Seven Xact elements have > 50 % of their data points below MDL and more than 90 % below 3*MDL: V, Co, As, Se, Cd, Pt, and Bi. Cr and Cd show the same behaviour for ICP. Ni revealed variable blank concentrations in the filters and could therefore not be reliably measured with ICP. Hg is also mostly below MDL in the AuAAA measurement, and Pt was not measured with ICP at all. In summary, 11 elements (K, Ca, Ti, Mn, Fe, Cu, Zn, Sn, Sb, Ba, Pb) are above their MDLs for both the XRF and the offline methods, 7 elements (V, Co, As, Se, Cd, Pt, Bi) are below MDL for the XRF, and 3 elements (Cr, Cd, Hg) are below MDL for the offline methods (of which only Cd is below MDL for both XRF and ICP).
The comparisons between online Xact 625 and offline 24 h PM elemental concentrations for 21 elements are shown in Figs. 2, 3, and Table 1. Only the 21 elements analysed by both methods are compared by dividing them into two groups based on data characteristics.
Scatter plots and regression lines of Xact 625 (ordinate) vs. ICP-OES/MS (abscissa) data for groups A and B. Elements with an asterisk () were analysed with ICP-OES. The axes have been scaled by the maximum concentration indicated in the legends for each element ( for Hg). The Levenberg–Marquardt linear least-squares fitting method was applied, taking the ICP measurements as the independent data. Regression equation is a . See Table 1 for data.
[Figure omitted. See PDF]
(a) Slopes and intercept ratios (intercepts divided by the average concentrations measured with XRF) with standard deviations for all elements measured with the Xact. (b) Minimum detection limits (MDL), interference free for the Xact 625 (XRF) and for the ICP analyses. Hg was analysed with AuAAA spectrometry.
[Figure omitted. See PDF]
Group A shows excellent correlations between the two measurement methods ( values > 0.95) and intercepts < 40 % of mean concentration and consists of the elements S, K, Ca, Fe (analysed by ICP-OES), Ti, Mn, Cu, Zn, Ba, and Pb (analysed by ICP-MS). The regression intercepts were not forced to be zero to enable examination of potential differences in the measurement accuracy of each of the compared methods, e.g. blank subtraction. The slopes are more relevant and indicate biases between the methods. Orthogonal least-squares regressions metrics were calculated, which incorporates measurement errors in both quantities being compared. The slopes differed by less than 3.5 % between the two regression methods for the Group A elements. Ba and Pb achieved an almost perfect match with slopes around 1 and negligible intercepts. The other extreme is Zn with a slope of 1.8. Ti is another peculiar case with a slope of 1.13 and the largest intercept/average concentration ratio of 0.37. On average, the Xact 625 yielded approximately 28 % higher elemental concentrations than ICP for the Group A elements.
The high linearity and little scatter in the regressions testify for the precision of both the Xact and the ICP methods, but the differences in the slopes (range 0.97 to 1.8) for different elements require further investigation. No systematic deviations based on elemental molecular weight or X-ray excitation conditions were observed in these slopes. The deviations of the slopes from unity may be partially attributed to the different inlets for the Xact and the HiVol samplers (Panteliadis et al., 2012), which may produce a difference in collected mass on the order of 10 %. The inlets were not explicitly tested for their cut-off characteristics in this study. A slightly different cut-off value for the particle size may lead to differences in the collected mass, especially for the largest and heaviest particles in PM, and hence to an underestimation or overestimation of the total mass collected with a particular inlet. This may be of special relevance in a near-road setting with lots of re-suspended dust (ACES, 2012). Potential line interference between Ti and Ba can be excluded, because the element couple reveals two different regressions for fireworks and non-fireworks cases, as well as distinct diurnal variations in the non-fireworks cases.
The results of additional investigations of a few selected filters by independent labs and analytical methods for understanding these differences are discussed in the supplementary material. Examination of reference samples indicated a high precision in XRF measurements despite consistently underestimated absolute concentration (underestimated by 6 to 14 % depending on the element). In contrast, ICP measurements indicated greater variability (range of 30 % underestimation to 60 % overestimation, depending on the element) and hence higher uncertainty in estimated ambient concentrations. Examination of three filter samples collected during the campaign by an offline XRF instrument (by CES) and by ICP at an external laboratory (ERG) indicated a variability of about 30 % for most elements, an almost perfect match (0.1 %) between ICP labs for the average concentrations when Ca and Se are excluded, and deviations of less than 5 % between benchtop XRF and ICP (Table S2). These results are comparable to those in Gerboles et al. (2011). Comparisons between the Xact and benchtop XRF in previous tests by the manufacturer with better control of the sampler inlet conditions almost always were within 5 % of each other. The relative mean difference of 28 % between Xact and filter data (analysed with ICP) for samples collected during the field campaign appears to be systematic. Such differences may result from a difference in location of the Xact and filter sampling inlets ( 5 m) and their relative distance from the freeway. Ultrafine particle number concentrations from dust resuspension due to vehicle traffic are known to decrease with increasing distance from the road, with the sharpest decline observed within the first 50 m (Hagler et al., 2009). Crustal elements, which dominate in the PM size fraction, are expected to settle faster due to larger aerosol size. Hence the difference in Xact- and ICP-reported PM elemental concentrations may be indicative of a gradient in PM occurring for some elements in close proximity to roadways. To quantify such near-source PM gradient, a field campaign with a different design would be needed, e.g. an array of samplers along a line perpendicular to the freeway. However, since the difference is also observed for S, which is typically found in the fine mode, does not have a major traffic related source apart from coarse re-suspended dust, and is not expected to suffer from incomplete digestion we assign part of the differences also to calibration issues with Xact.
Comparison of Xact data with published ICP data of other campaigns.
Group | Xact 625 averages | Filter samples | |||||||||
---|---|---|---|---|---|---|---|---|---|---|---|
Non-fireworks | Belp | Geneva | Basel, | Payerne | Zürich | Zürich, | Payerne, | Payerne, | Härkingen | ||
days | summer | summer | summer | summer | NABEL | ||||||
Reference | 1 | 2 | 3 | 4 | 5 | 6 | 6 | 7 | 8 | ||
No. cases, size | 17 | PM | PM | PM | PM | PM | PM | PM | PM | PM | |
Sampling period | 2015 | 1985/86 | 1996/97 | 1997/98 | 1998/99 | 2008/09 | 2009 | 2009 | 2012 | 2015 | |
Unit | ng m | ng m | ng m | ng m | ng m | ng m | ng m | ng m | ng m | ng m | |
A | S | 790 | 2394 | 625 | 637 | 360 | |||||
K | 167 | 630 | 98 | 1318 | 187 | 188 | 120 | ||||
Ca | 398 | 720 | 100 | 137 | 451 | 355 | 180 | ||||
Ti | 11.3 | 38 | 6.5 | 13.9 | 14.3 | 9.9 | |||||
Mn | 7.2 | 31.7 | 16 | 2.8 | 5.8 | 7.2 | 5 | 2.8 | |||
Fe | 594 | 760 | 89 | 390 | 455 | 202 | 130 | ||||
Cu | 24.4 | 7.9 | 35 | 75 | 6 | 28.1 | 17.4 | 4.3 | 2.8 | 19.7 | |
Zn | 19.0 | 65.0 | 120 | 73 | 20.3 | 16.1 | 9.4 | 7.5 | |||
Ba | 7.1 | 110 | 6.7 | 6.5 | 3.9 | 1.8 | |||||
Pb | 3.0 | 134 | 95 | 51 | 10 | 14.2 | 3.6 | 3.1 | 1.2 | 4.9 | |
B | V | 0.06 | 3.9 | 0.7 | 0.4 | 1.0 | 1.2 | 0.7 | |||
Cr | 2.4 | 8 | 0.9 | 2.3 | 1.6 | 1.0 | |||||
Co | 0.02 | 2.6 | 0.1 | 0.1 | |||||||
Ni | 0.63 | 8 | 1.2 | 0.6 | 0.9 | 1.0 | 0.5 | 0.9 | |||
As | 0.02 | 2.2 | 2 | 1 | 0.53 | 0.3 | 0.5 | 0.2 | 0.31 | ||
Se | 0.30 | 6 | 0.16 | 0.3 | 0.3 | 0.2 | |||||
Cd | 6.7 | 0.88 | 0.4 | 0 | 0.32 | 0.1 | 0.1 | 0.07 | |||
Sn | 20.9 | 2.6 | 2.6 | 1.0 | |||||||
Sb | 31.6 | 29 | 0.26 | 2.5 | 2.4 | 0.5 | |||||
Hg | 0.64 | 0.5 | |||||||||
Bi | 0.07 | 0.2 | 0.1 | ||||||||
C | Si | 829 | 211 | 571 | 634 | 370 | |||||
Cl | 109 | 41 | 657 | 66 | 190 | 30 | |||||
Pt | 0.05 |
1: Gälli Purghart et al. (1990). 2: Chiaradia and Cupelin 2000 – fortnight averages. 3: Röösli et al. (2001). 4: Hueglin et al. (2005). 5: Richard et al. (2011). 6: Minguillón et al. (2012). 7: Alastuey et al. (2016). 8: BAFU and Empa (2015) – annual mean values.
Group B, the remainder, consists of the elements V, Cr, Co, As, Se, Cd, and Bi, i.e. of those elements that are close to or below their Xact MDL, plus Ni, Cd, Sn, Sb, and Hg. Ni, Cd, and Hg were below MDL for the offline methods (Cd for Xact and offline methods). Although an intercomparison of these elements may not be justified, we observed some features in the regressions of the Group B elements in Fig. 2 that are worth commenting. Cr is below the ICP-MS MDL for 60 % and below 3*MDL for all filter values, but 75 % are above the Xact MDL. The ICP measurement is below MDL because of the high and variable blank concentrations, which make a meaningful blank subtraction difficult and increase the Cr MDL in these samples. Although the slope of Cr is 1.23 and thus comparable to the other Group A slopes, a comparison with ICP values is statistically not robust. However, Cr seems to be quantifiable with the Xact. The regression plot of Bi shows two extreme values on 31 July and 1 August corresponding to the fireworks days. These two points are above MDL for both methods and suggest good quantitative agreement between XRF and ICP for these two high-concentration cases. Sb shows a moderate correlation ( 0.47) and a large intercept. Sn behaves similarly as Sb, with an value of 0.15. The large intercepts hint toward a problem in processing the Xact blanks. In addition, when Ca is abundant, as in our case in Härkingen, the Sb L line interferes with the Ca K line, producing low signal-to-noise ratios for Sb, and similarly for the pair K–Sn. Hence, the reported Xact concentrations of Sn and Sb reflect mainly spectral noise. Of the filter Hg data, 60 % were below MDL and thus cannot be compared well with the Xact Hg data. Inspection of the Xact raw Hg spectra showed a possible interference from Br causing the fitting routine to attribute some Br mass to non-existent Hg peaks in the spectra. Br was not calibrated in the fitting routine. Thus the Hg concentrations reported by the Xact seem to be due to this interference and are not realistic, even though 86 % of the measured values are above the Xact MDL. Values < 1.5 ng m are in the same order of magnitude as the fortnight values of Chiaradia and Cupelin (2000) for the city of Geneva (Table 2). The element couples of Fe–Co and Pb–As do not show correlations within the couples because most of their data points are below their respective MDLs, and no conclusion about the deconvolution of interfering lines can be drawn for these elements. Comparing the Xact values with the NABEL annual mean values (Table 2) shows differences smaller than 40 % for Cu, Pb, and Ni, while the differences are much larger for As and Cd. The latter two elements are below their respective MDL (Fig. 2). To summarize, the Group B elements show issues with the MDLs of at least one of the analysis methods, which restricts thorough interpretation. Individual data points above MDL reveal nevertheless a usable quantification by the Xact in these particular cases. Determination of Sb, Sn, and Hg by Xact is potentially impacted by XRF line interferences and warrants further improvements for better quantification by Xact.
The Group C elements Si, Cl, and Pt were not measured on the filters. An Xact MDL for S has not yet been determined. Data quality of S measurements was inferred by comparing its Xact concentrations with the concentrations of another element originating from the same source or belonging to the same chemical compound. Xact-reported S (with unknown MDL) and K (with known MDL) concentrations were highly correlated ( 0.99) during the fireworks period, with a slope of 2.30 0.05, which agrees with the stoichiometric relation between K and S when forming KSO. For the non-fireworks periods the correlation was weak ( 0.16), which hints towards a completely different, more random relationship between the two elements, as expected. Hence, high-concentration K reported by the Xact provided reliable S during the fireworks episode, despite the lack of an established MDL for S. Xact Pt measurements were always below MDL, and no conclusion about the Pt accuracy can be drawn.
Comparisons with other data
Figure 1 shows that roughly 95 % of the total analysed elemental mass by Xact is comprised of six elements: Si, S, Cl, K, Ca, and Fe. These major elements all show average concentrations > 100 ng m. Si, S, Ca, and Fe are observed throughout the study, although with high variability. Cl and K are abundant only episodically: Cl is strongly tied to westerly winds during the last week of July and is practically absent after 2 August. K is prominent during the fireworks period. Ti, Cu, and Zn show daily average concentrations between 11 and 34 ng m. The other analysed elements were found in daily average concentrations < 10 ng m. The concentrations are of the same order of magnitude as those recently measured at other places in Switzerland, e.g. at an urban background site in Zurich (Minguillón et al., 2012; Richard et al., 2011, Table 2), but are generally lower than the measurements in older studies (Chiaradia and Cupelin, 2000; Gälli Purghart et al., 1990; Röösli et al., 2001, Table 2). The decreasing trends in PM and trace element concentrations have been documented in numerous NABEL reports on the air quality in Switzerland (e.g. BAFU and Empa, 2015; Gianini et al., 2012). These trends make it preferable to use modern studies for comparisons. Furthermore, the episodic nature of the 2015 campaign also demands for some generosity when comparing the measured values with annual or seasonal mean values.
A time series of the Xact 625 total element concentrations together with the NABEL TEOM PM data and the total ACSM NR-PM concentrations with 1 h resolution is presented in Fig. 4. The total ACSM NR-PM concentration is the sum of sulfates, nitrates, ammonia, chlorides. and organic aerosols. Total PM shows a generally increasing trend over the whole campaign, with a strong peak superposed on 1 August 2015, whose increase coincides with the peak in the Xact data, whereas the maximum is delayed by 1 h relative to the Xact data. The peak is due to the fireworks burnt on that evening. On average the Xact 625 elements make up about 20 % of the total PM mass (regression slope 0.2, 0.63). A complete mass closure cannot be achieved because the NABEL station only reports the total gravimetric PM mass and PM EC concentrations with diurnal or better time resolution.
Time series of Xact 625 total elemental concentration, ACSM PM1 data, NABEL TEOM PM data, and wind speed (WSpd) and direction (WDir) measurements in Härkingen. Numbers at the peaks indicate 1 h concentration maxima.
[Figure omitted. See PDF]
Comparison of Xact PM SO vs. ACSM PM SO. Data were split into fireworks (red) and non-fireworks (blue) periods.
[Figure omitted. See PDF]
Xact's measurement accuracy for S was tested by comparison with ACSM sulfate measurements (Fig. 5) performed at 1 h resolution. The S concentrations of Xact 625 were multiplied with a factor of 3, assuming that all S occurred in the form of SO. The slope of the regression line for the non-fireworks case is 1.34, with 0.85, in line with the Group A elements and in agreement with the comparison of S from Xact and from ICP; hence the slope between ACSM SO and ICP SO would be 1. The high linear correlation suggests a high precision of the Xact 625 data but does not allow a definitive answer on the accuracy because of expected self-absorption effects, which would increase the slope further if corrected. During the fireworks period, the scatter is large, and the correlation coefficient is only 0.1. We hypothesize that fireworks produce larger and non-refractory particles (e.g. KSO not measured by the ACSM.
In summary, the on average 25 to 30 % difference between the Xact and ICP data can probably be explained by differences in the sampling inlets, the distance between the inlets, and uncertainties of the different analysis methods. The correlation coefficients close to 1 for many elements demonstrate the high precision of the Xact and ICP methods. The obtained time series of those elements can thus reliably be used for source apportionment. The subsequent analyses (e.g. elemental concentration ratios, enhancement ratios) were done with the unmodified Xact data. The only exception is estimation of a mass budget in the discussion of the extreme concentrations in Sect. 3.3.
Extreme concentrations: the fireworks period
As mentioned in Sect. 3.1 the measurement campaign can be divided into a fireworks and a non-fireworks period. A K concentration > 220 ng m served as the criterion to distinguish between these periods, and we required the fireworks period to be contiguous from the first increase in K on 31 July 2015 22:00 LT to the final decay to background values on 4 August 2015 11:00 LT. The average K concentration during the fireworks period was 2 g m, but this period showed extremely high hourly PM concentrations and an element mix different from the remainder of our test campaign.
Investigation of the highest peaks reveals the performance of the Xact under high load conditions, when sample thickness may become critical for XRF analysis. A comparison of the two instruments' peaks could demonstrate how closely the Xact mass represents the total measured PM mass. Inspection of the different time series indicates that the TEOM peak is broader (3 h) and higher (60 g m, and its maximum concentration is reached 1 h later (at 2 August 2015 00:00 LT), but its increase in concentration starts at the same time as the Xact (at 22:00 LT). The delay in the maximum can be attributed to the time constant of the TEOM used for reducing measurement noise and to the averaging procedure. For a comparison of the two peaks their measured masses were integrated over the duration of the peaks, i.e. over 2 h for the Xact data and the ACSM data and 3 h for the NABEL data. The integrated NABEL PM mass reached a concentration of 122 g m. The Xact 625 monitor reported a total of 86 g m for the sum of all analysed elements (except Pd, which was used only as an internal standard).
The bulk of this concentration (84 g m was made up of a few elements (in parentheses: concentrations in g m, and abundances relative to total analysed element mass, PM: K (49, 57 %), S (21.5, 25 %), Cl (7.4, 8.6 %), Fe (3.2, 3.7 %), Ba (2.1, 2.4 %), and Si (1.7, 1.9 %). Absolute K and S concentrations are in good agreement with the values in Drewnick et al. (2006). K likely originated from KNO, a basic constituent of black powder (Drewnick et al., 2006; Kong et al., 2015).
The ACSM PM contributed 20.5 g m (17 %) to the aerosol concentration in the fireworks peak.
The comparison of the Xact concentration and the TEOM PM concentration for the 1 August peak requires taking into account the systematic difference between the Xact and TEOM measurements discussed above and the fact that the elements are typically not present in elemental but rather in their oxidized form, such that the mass of the latter needs to be included for a quantitative comparison. We therefore estimate a mass budget for the fireworks peak, when the six elements K, S, Cl, Fe, Ba, and Si comprise the bulk of the total mass. We calculate the ion balance for the positive ions K, Fe, Ba, Si, and submicron NH and negative ions Cl, SO, and submicron NO. We further add all available components from the NABEL station (PM equivalent black carbon, eBC) and the ACSM (NO, NH, and organic aerosols, but not SO, as this is already considered with the S in the Xact data) to the mass balance. Using the measured values of the Xact 625 yields an excess of negative ions of 2.7 % and a total mass of 141 g m, which overestimates the TEOM value of 122 g m by 16 %. If Xact concentrations were scaled towards the ICP concentrations with the corresponding regression slopes from Table 1, using for Cl and S the average slope of 1.22 from the other four elements, then the calculation yields an excess of 11 % in positive ions, which are then assumed to be neutralized by oxygen. This yields a total mass of 116 g m, which is 4.6 % lower than the TEOM value. Our values are lower limits of the total mass because the balance is incomplete with respect to relevant elements in the fireworks (e.g. Sr) and other chemical species like carbonaceous and nitrogen containing molecules in the coarse fraction. The result shows that the bulk of fireworks PM aerosols are a few metals oxidized to sulfates, chlorides, and oxides. Overall, these results further demonstrate the advantage of Xact's high-time-resolution sampling for associating high-metal-concentration episodes with source emission activities.
Investigation of sources
Trace elements can be excellent tracers for specific aerosol sources (e.g. Hopke, 2016; Park et al., 2014; Querol et al., 2007; Visser et al., 2015a). A simple approach for characterizing a common source for a group of elements is to study the temporal covariation of the elements in this group. For our Härkingen data, the time series indicate the strong influence of the fireworks on the concentrations of K, S, Ti, Cu, and Ba (Figs. 6 and 7), which are important constituents of fireworks along with Sr as fireworks tracer (Kong et al., 2015; Moreno et al., 2007). Examination of a few raw spectra from the fireworks and non-fireworks periods indeed identified an enhancement of Sr during the fireworks duration, while the peak was definitively absent during the non-fireworks period. Sr was, however, not quantified in our configuration, as we put emphasis on crustal elements and some special trace elements difficult to detect in Switzerland (Hg, Pt).
Mean diurnal variations of (a) Ca and (b) Ba, stratified for fireworks (red) and non-fireworks (blue) periods. Error bars denote 1 standard deviation of the averaging period. Diurnal variations for the other elements are shown in the Supplement (Fig. S4).
[Figure omitted. See PDF]
(a) Difference between average concentrations of fireworks and non-fireworks periods. Numbers indicate values outside the axis range for S, K, and Si (negative). (b) Enhancement ratios for all analysed elements for fireworks/non-fireworks classification, sorted by group. Numbers indicate values outside the axis range for S, K, and Si.
[Figure omitted. See PDF]
Mean diurnal variations of (a) S, (b) Mn, (c) Pb, and (d) NO for the non-fireworks period stratified for south and north wind directions. Error bars denote 1 standard deviation of the averaging period. Diurnal variations for the other elements are shown in Fig. S5.
[Figure omitted. See PDF]
(a) “Background”-subtracted mean concentrations of the south sector for the non-fireworks period. Number indicates value outside the axis range for Fe. (b) Enhancement ratios for all analysed elements for south–north sector classification for the non-fireworks period, sorted by group.
[Figure omitted. See PDF]
The 1 h sampling interval further enabled examination of diurnal variations of the elements. Ca and Ba are presented in Fig. 6 and the other elements in Fig. S4, which shows the classification of the data according to fireworks and non-fireworks periods. It can be seen for the time from 23:00 to 06:00 that the elements Ba, Bi, Cu, K, S, and Ti show a clear distinction between the two periods. The fireworks elements show a maximum concentration at 23:00 LT and a gradual decay over the next 6 to 12 h into the morning hours of the (following) day. Mn, Fe, and Zn also show an increased and then decaying concentration after midnight, but the difference compared to non-fireworks days is within the data variability. Si is depleted during the fireworks period relative to the non-fireworks period. This is probably a weekday vs. weekend effect, when fewer HDV circulate (Switzerland does not permit HDV use on Sundays) and less road dust is re-suspended (Bukowiecki et al., 2009, 2010). For the non-fireworks cases the transition elements Mn, Fe, Zn, and the element Pb are characterized by a broad morning peak with a maximum around 10:00 LT, correlating well with the increasing traffic in the morning hours and the breakup of the temperature inversion.
To identify the fireworks tracers, an enhancement ratio (ER) was defined as the ratio between the mean concentrations of an element for the fireworks period to the concentrations in the non-fireworks period (Fig. 7). For K, Cu, and Ba the ER is larger than 2 (Cu) and goes up to 10.6 (Ba). S, Cl, Ti, Zn, and Pb show an intermediate ER between 1 and 2. Cr, Mn, and Fe ER are close to 1. Si and Ca are depleted with an ER around 0.5, both of which are probably related to the above weekend effect. The elements with the high ER are clearly identified as elements of fireworks: S, K, Ti, Cu, Zn, Ba, and Bi.
Further refinement of sources can be obtained when classifying the non-fireworks data by wind direction into a north (270–0–90) and a south (90–180–270) sector (Fig. 8), with the south sector more strongly influenced by highway emissions (Hueglin et al., 2006). The north sector characterizes the (rural) background concentrations of the central Swiss plateau. Table S1 summarizes the mean element concentrations for the campaign divided into the different periods and wind sectors. Ba, Cr, Cu, Fe, and Mn show the signature of continuous freeway traffic emissions during the day. Pb and Zn show a morning peak only and are well correlated in both sectors. Si, K, Ca, and Ti show another pattern that could reflect the local and regional transport of crustal material partly re-suspended by traffic (south sector), partly originating from the agricultural area north of the freeway.
Figure 9 shows the enhancement ratios south–north. Apart from Cl, all south–north differences are positive, and Si, S, Ca, and Fe concentration differences are larger than 80 ng m. These are mainly crustal elements (although Fe is also emitted from vehicles). The enhancement ratios of the transition elements Cr, Mn, Fe, Cu, Zn, Ba, and Pb are larger than 1.2 and related to traffic emissions (engine abrasion, tyre wear, brake wear).
Conclusions
A 3-week test of a Cooper Environmental Xact 625 Ambient Metals Monitor was performed at the Swiss NABEL station Härkingen. The instrument was configured to measure 24 elements continuously with 1 h time resolution. The selection of elements ranged from Si to Bi, thus covering a range of environmentally relevant elements. Besides the “standard” elements from K to Pb, which have been well characterized by the manufacturer with respect to their accuracies and detection limits, we included several abundant light elements (Si, S, Cl) and – more for curiosity - some low-concentration elements (As, Pt, Hg) in our selection to test the behaviour of the instrument in a typical Swiss environment. We tested the measurement quality of the Xact 625 by intercomparison with well-established methodologies (ICP-OES and ICP-MS analyses on 24 h PM samples for major and trace elements and AuAAA for Hg), ACSM, and TEOM and used additional meteorological data for the interpretation of the results.
The general findings are as follows:
-
The total of elements analysed with the Xact comprised approximately 20 % of the PM mass.
-
The Xact 625 produced element concentration time series that were highly correlated with the ICP analyses of 24 h filter samples ( 0.95), even though the slopes deviated from 1.
-
Element concentrations ranged from ng m (in background conditions) to tens of g m (during the fireworks), and no instability in operation due to sample overload or else could be observed.
-
The measured concentrations agreed reasonably well with other recent field measurements in Switzerland.
-
Excellent correlation between Xact 625 and ICP-OES/ICP-MS was observed for 24 h averages of the elements S, K, Ca, Ti, Mn, Fe, Cu, Zn, Ba, and Pb (Group A). The daily averages calculated from hourly measurements by Xact were on average 30 % higher (range 3 to 80 %, depending on the element) than 24 h integrated filter measurements by ICP. Systematic differences of on average 25 to 30 % could be attributed to physical reasons in the experiment settings, such as the different characteristics of the two inlet systems, the distance between the inlets and to the main source (freeway), and uncertainties in the various analysis methods. For XRF this includes particle-size-dependent self-absorption effects for the lighter elements and line interferences between different elements. For ICP this includes the entire sampling, digestion, and the analysis procedure, as indicated by limited inter-laboratory and inter-method comparisons, as well as the impurities in blank filters. Further research on these issues is needed.
-
The accuracy of hourly values has only been tested for the case of the fireworks peak late on 1 August 2015, where the sum of all elements has been compared to the total mass of the NABEL TEOM. Good agreement between the Xact and TEOM mass was found when corrections derived from the 24 h filter analyses were applied. This was a special case dominated by just three elements, S, K, and Cl, and a generalization of all measured elements is not recommended.
-
The remaining elements (Group B) of the filter intercomparison, V, Cr, Co, Ni, As, Se, Cd, Sn, Sb, Hg, and Bi (11 elements), were mostly below detection limit of at least one method or showed issues with the analysis procedures (Sn, Sb, Hg). A general quantitative statement on their quality could not be achieved. Notice here that a longer sampling time, e.g. 2 or 4 h, would have lowered the Xact MDLs and therefore increased the number of good measurements, but at the cost of a reduced time resolution.
-
Si and Cl were not analysed on the filters, and their Xact detection limits have not yet been determined. Hence their accuracies could not be quantified directly. Some indirect approaches were calculated.
-
The Pt concentrations reported by the Xact 625 were below MDL, and Pt was not analysed on the filters. No conclusion about the accuracy of this difficult to measure element can be drawn.
Compared to rotating drum impactor sampling with synchrotron radiation induced XRF or streaker sampling with PIXE analysis, the Xact 625 measures ambient concentrations of the most relevant elements in near real time and provides data with a delay of only one sampling/analysis cycle. The continuous operation capability of Xact also circumvents the sample number limitations due to restricted beam-time assignments at synchrotrons. This is a major advantage compared to the usual time delay of a couple months caused by the restricted access to synchrotron or accelerator facilities. Of course, the high time resolution of the Xact 625 comes at the cost of sensitivity, visible in the MDLs, which are higher than the MDLs for the offline methods. Xact can be set for longer sampling intervals to extend the number of samples with analysed elements above their MDLs depending on the objective of monitoring campaigns. Xact streamlines near-real-time monitoring of multi-metals despite not being as cost effective relative to conventional samplers that could be deployed in larger numbers at many sites simultaneously or that could sample several size fractions at once, although their actual analysis costs (laboratories, accelerator facilities and staffing needs) are not considered here and they may surmount the instrument costs manifold. Overall, high-time-resolution sampling of metals provides a rich dataset for associating high-metal-concentration episodes with source emission activities. Useful extensions of the present capabilities of the Xact could be the addition of more elements to be analysed (especially under the circumstance that the full mix of observed elements cannot always be known in advance), improved quantification of the lightest elements (especially their MDLs), a vacuum or helium device for analysing light elements like Na and Mg, and an inlet switch to alternately measure PM and PM with one single instrument.
The datasets are available upon request to the corresponding author.
The Supplement related to this article is available online at
Krag Petterson and Varun Yadav are employed by Cooper Environmental Services, the manufacturer of the Xact® 625.
Acknowledgements
This study has been partly funded by the Swiss Federal Office for the Environment (FOEN). M. C. Minguillón acknowledges the Ramón y Cajal Fellowship awarded by the Spanish Ministry of Economy, Industry and Competitiveness. We thank René Richter and Roland Scheidegger of PSI for their support during the field campaign. We are grateful to Chris Koch and John Cooper of Cooper Environmental Services for instructions on instrument operation and numerous discussions on the results. Andrés Alastuey, Xavier Querol, and laboratory personnel from IDAEA-CSIC are also acknowledged. We also thank Julie Swift and Randy Mercurio of ERG for the ICP-MS analyses. Edited by: W. Maenhaut Reviewed by: three anonymous referees
You have requested "on-the-fly" machine translation of selected content from our databases. This functionality is provided solely for your convenience and is in no way intended to replace human translation. Show full disclaimer
Neither ProQuest nor its licensors make any representations or warranties with respect to the translations. The translations are automatically generated "AS IS" and "AS AVAILABLE" and are not retained in our systems. PROQUEST AND ITS LICENSORS SPECIFICALLY DISCLAIM ANY AND ALL EXPRESS OR IMPLIED WARRANTIES, INCLUDING WITHOUT LIMITATION, ANY WARRANTIES FOR AVAILABILITY, ACCURACY, TIMELINESS, COMPLETENESS, NON-INFRINGMENT, MERCHANTABILITY OR FITNESS FOR A PARTICULAR PURPOSE. Your use of the translations is subject to all use restrictions contained in your Electronic Products License Agreement and by using the translation functionality you agree to forgo any and all claims against ProQuest or its licensors for your use of the translation functionality and any output derived there from. Hide full disclaimer
© 2017. This work is published under https://creativecommons.org/licenses/by/3.0/ (the “License”). Notwithstanding the ProQuest Terms and Conditions, you may use this content in accordance with the terms of the License.
Abstract
The Xact 625 Ambient Metals Monitor was tested during a 3-week field campaign at the rural, traffic-influenced site Härkingen in Switzerland during the summer of 2015. The field campaign encompassed the Swiss National Day fireworks event, providing increased concentrations and unique chemical signatures compared to non-fireworks (or background) periods. The objective was to evaluate the data quality by intercomparison with other independent measurements and test its applicability for aerosol source quantification. The Xact was configured to measure 24 elements in PM
You have requested "on-the-fly" machine translation of selected content from our databases. This functionality is provided solely for your convenience and is in no way intended to replace human translation. Show full disclaimer
Neither ProQuest nor its licensors make any representations or warranties with respect to the translations. The translations are automatically generated "AS IS" and "AS AVAILABLE" and are not retained in our systems. PROQUEST AND ITS LICENSORS SPECIFICALLY DISCLAIM ANY AND ALL EXPRESS OR IMPLIED WARRANTIES, INCLUDING WITHOUT LIMITATION, ANY WARRANTIES FOR AVAILABILITY, ACCURACY, TIMELINESS, COMPLETENESS, NON-INFRINGMENT, MERCHANTABILITY OR FITNESS FOR A PARTICULAR PURPOSE. Your use of the translations is subject to all use restrictions contained in your Electronic Products License Agreement and by using the translation functionality you agree to forgo any and all claims against ProQuest or its licensors for your use of the translation functionality and any output derived there from. Hide full disclaimer
Details



1 Laboratory of Atmospheric Chemistry, Paul Scherrer Institute, 5232 Villigen PSI, Switzerland
2 Institute of Environmental Assessment and Water Research (IDAEA), Consejo Superior de Investigaciones Científicas (CSIC), Jordi Girona 18-26, 08034 Barcelona, Spain
3 Cooper Environmental Services (CES), 9403 SW Nimbus Avenue, Beaverton, OR 97008, USA
4 Laboratory for Air Pollution/Environmental Technology, Empa, Überlandstrasse 129, 8600 Dübendorf, Switzerland