1 Introduction
Aerosol particles play crucial roles in various atmospheric processes and the Earth's climate system
Scanning transmission X-ray microscopy with near-edge X-ray absorption fine structure analysis (STXM-NEXAFS) in the soft X-ray regime (270–2000 eV) has become a widely used and powerful technique to resolve the micromorphology and chemistry of laboratory and ambient aerosol particles on submicron scales
Here, we present the development of a gas flow system coupled with a microreactor as an accessory for STXM instruments for in situ studies of particles in a controlled gas-phase environment; we emphasize its analytical capabilities and show initial results. The instrument's design and construction was inspired by previous developments of environmental chambers for X-ray microscopes, namely by , , , and . The microreactor system has been developed according to the following requirements:
-
Compactness and portability. facilitating compatibility of the system with different STXM instruments and application at different synchrotron sites.
-
Minimal optical path length. accounting for short focal lengths in STXM optics, e.g., to allow measurements at the carbon (C) K-edge and at even lower energies.
-
Maximum sample compatibility. suitable for standard silicon nitride membrane windows ( m) operated at up to 1000 hPa pressure difference between the inside of the microreactor and the surrounding STXM enclosure, quick and safe sample (un-)mounting.
-
Reliable and stable parameter control. environmental parameters , , and RH tunable over a wide value range; particularly, humidity control in high-RH regimes (i.e., 80 % RH up to saturation) and control over down to 250 for kinetic studies and freezing experiments.
-
Extension options. interfaces to the gas supply circuit for the introduction of reactive atmospheres to study particle–gas phase reactions (e.g., ozonolysis).
2 Technical description
2.1 Control system design
The control system is of compact size: the gas mixing and cooling circuits, along with the power converters and electronics are integrated into a 19 in. enclosure with a height of four rack units (total dimensions: ). The relevant parts of these circuits and the positions of the environmental sensors (BME280 by ) therein, which were used throughout the system to trace changes in the , , and RH values, are shown in Fig. . The only external supplies needed, besides mains voltage, are a vacuum pump and a source of pressurized process gas, such as nitrogen or synthetic air, but preferably helium to achieve the best signal-to-noise ratio across all reachable absorption edges. However, provision was made for operation under reactive (e.g., ozone-enriched) atmospheres by using external gas supplies, indicated by the orange dashed line in Fig. , which can be attached to the system via designated ports.
Figure 1
Gas flow and cooling system schematics of the entire system. Left: X-ray microscope chamber with sketched microreactor. Refer to Fig. for a detailed view of the microreactor. Right: 19 in. enclosure that includes the control system. S1–S4: , , and RH environmental sensors (Bosch BME280); M: solenoid-operated shut-off valves (Bronkhorst® EV-02-NC-V); TC: thermocouple (Omega™ 5TC-TT-TI-4)
[Figure omitted. See PDF]
The tasks of control and data acquisition are performed by the so-called “VBUS system”, which has been developed at the Max Planck Institute for Chemistry (MPIC). This miniature measurement system consists of microcontroller-based electronic modules and a flexible software environment including scripts and a graphical user interface (GUI). An example GUI screenshot is provided in the Supplement Fig. S1. The gas humidification system is similar to the one used by and mixes wet and dry gas flows to provide a process gas of desired humidity. Upstream, two Bronkhorst® IQFLOW® IQFD-200C mass flow controllers (MFCs) assure that the combined flows equal 20 mL min. By running one of the gas streams through a Permapure Nafion humidifier (MH-110-24F-4, 60 cm length), the RH directly after the humidifier can be increased to up to 70 % RH at ambient system temperature and should not exceed this value to avoid condensation inside the control system. The RH values are measured, along with and , at three locations within the circuit by Bosch BME280 environmental sensors , indicated by circles in Fig. , labeled with “, , RH” and named accordingly: (S1) is located directly downstream of the humidifier. Its RH value is taken as the control variable, which is continuously checked at a rate of 4 Hz against a desired set point. The RH reading typically fluctuates by about 0.05 % RH, which can be taken as the relative accuracy of the software-implemented proportional-integral (PI) controller that steers the MFCs' flows; (S2) measures , , and RH values inside the microreactor close to the sample; (S3) sits symmetrical to S1 in the return flow. This sensor is particularly useful to detect losses due to leaks or condensation inside the flow system. S1 and S3 are interfaced via the Inter-Integrated Circuit (IC) bus, S2 is mounted onto a separate printed circuit board (PCB) with Serial Peripheral Interface (SPI) capabilities (compare with Fig. ).
Stable temperature, , is critical for accurate RH control. Therefore, our design includes a closed circuit active cooling system, which combines a Twinbird Corp. SC-UB04 Free Piston Stirling Cooler with a HNP Mikrosysteme mzr®-2921 micro annular gear pump. As a coolant the hydrofluoroether was used, commercially available as 3M™ Novec™ 7200 Engineered Fluid. By controlling the pump speed between 0.3 and 18 mL min via a second software PI-regulator that takes the value from inside the microreactor as the control variable, temperatures between the ambient and can be set. The control system stability, the sensor characteristics, and possible applications for the cooling capabilities will be detailed in Sect. .
Figure 2
Rendered views of the microreactor assembly and the sensor PCB with the most important parts labeled. The four-way microfluidic connector (Dolomite Ltd.) connects to the internal channel structure in the brass metal body of the microreactor via a compression seal at the top face. The signals from the BME280 (, , RH) sensor (S2 in Fig. ) and the ADS1118 16 bit analog-to-digital converter (ADC) with internal reference sensor for the thermocouple readout are available through a Serial Peripheral Interface (SPI) at the six-pin ZIF connector.
[Figure omitted. See PDF]
Figure 3
(a) The microchamber adapted to the MAXYMUS instrument on a customized holder, thermally insulated on MACOR® posts. This photo emphasizes the spatial restrictions due to the integral parts of the microscope (zone plate, OSA, detector). (b) Sectional view through the microchamber, revealing the internal channel structures, the positions of O-rings, the sensor cavity and the groove into which the thermocouple wires are firmly glued. (c) Cross-sectional view at a larger scale, emphasizing the physical thickness of the microreactor along the optical axis (700 m), given by the thicknesses of the silicon nitride window frames, 200 m on each side, and the gap width of 300 m, in which the gas stream flows. The microreactor can be freely driven by 1 mm in every direction in the focal plane.
[Figure omitted. See PDF]
The system's pressure, , can be controlled between and 1000 hPa, by using a Bronkhorst® IQFLOW® IQPD-700C Back Pressure Controller downstream of the flow circuit. This controller is also active during the evacuation of the microscope chamber to keep the differential pressure between the microreactor and the surrounding, monitored by S2 and S4 (refer to Fig. ), respectively, at a minimum, if not desired otherwise. Not shown in the flow scheme in Fig. are ports for venting the STXM vacuum chamber for the reverse case. Three Bronkhorst® solenoid-operated shut off valves (EV-02-NC-V) can be used to bypass the microreactor, e.g., for leak testing. If not stated otherwise, the gas and fluid system components, fittings, filters, and connectors used in this system were purchased from Swagelok. The essential parts of the flow circuit are surface mounted to an aluminum circuit board, which interconnects them via internal channel structures as displayed in Fig. S2.
2.2 Microreactor designThe microreactor, displayed in Fig. , serves multiple purposes: it holds the sample in place at a defined , and exposes it to a process gas previously mixed inside the control system. All custom-designed parts were manufactured on computer numerical control (CNC) machines in the mechanical workshop of the MPIC. The individual components were conceptualized in Autodesk® Inventor 2014 as computer-aided design (CAD) models, which were used for the renderings in Figs. and .
The fluidic connections between the front panel of the control box and the microreactor were realized via four Upchurch Scientific (IDEX Health & Science) PEEK tubings (1.6 mm outer diameter mm inner diameter m length) for the gas and coolant flows, guided through a custom-made vacuum feed-through. A four-way microfluidic connector (Dolomite Ltd., Part number 3000024) then seals the microreactor body (Fig. ) from the interior of the STXM chamber. The reactor body itself consists of two CNC-machined brass metal parts with internal channel structures, glued together with LOCTITE® EA 9497 two-component epoxy adhesive. The microreactor features a quick-change mechanism via twist-lockable sample mounting disks, which are 1.5 mm thick, manufactured from an aluminum alloy. The usage of these disks allow a tension-free sample exchange and minimizes the number of separate parts, such as screws. Standard silicon nitride membrane windows ( m membranes with mm outer silicon frame dimensions) act as the sample substrate and are mounted to these plates, which simultaneously act as the microreactor's front cover. For exchanging the sample, the mounting plate needs to be pressed downwards against the brass metal body to compress the underlying O-ring seal. A 45 rotation then locks or unlocks the sample, respectively. For instance, this can be done with sharp tweezers. After mounting the sample, the expanding O-ring in its groove presses the sample holder against its counterpart, the front steel plate, which keeps the sample at a defined position and well-sealed.
Sectional views in Fig. b and c illustrate how the parts are assembled. A rendered image sequence, composed to a video clip by using the video editing software DaVinci Resolve 16, shows the assembly of the microreactor; emphasizes its internal structures, the locations of the sensors, and the O-ring seals; and visualizes the sample mounting process. It is provided as a Video supplement below.
With a typical focal length of 1.36 mm at 280 eV, the space between the focal plane and the zone plate (compare Fig. a and Fig. 1 in ), with the order sorting aperture (OSA) in between, is very limited. The OSA (60 m diameter) is located approximately 320 m away from the sample plane to prevent unfocused X-rays and stray light from passing through the sample and from entering the detector. Subtracting the silicon frame thickness of the front window leaves just 120 m between the OSA and the microreactor. We therefore avoided a sample holder design that adds more material to this side of the microreactor. Due to inherent geometric restrictions it is not possible to reach energies below 200 eV with the current setup. However, silicon nitride windows with just 100 m thick frames are commonly available and their use together with a modified mounting disk would in principle bring the sulfur L-edge at 170 eV into range, neglecting any limitation by X-ray optics and insertion devices. It has been shown that it is feasible to image the sulfur distribution in aerosol particles at the STXM beamline 11.0.2 at the Advanced Light Source (Berkeley, CA, USA) and at MAXYMUS
For a gas-tight seal, the silicon nitride windows must be glued into the sample holder disks. We either used IMI 7031, also known as GE varnish, and let it cure at room temperature, or Apiezon Wax W, which melts at about 373 and therefore is only suitable for empty windows prior to sampling or temperature-insensitive samples.
Behind the front window, the process gas flows in a 300 m wide gap, as emphasized in Fig. b. This specific gap width was chosen for maintaining sufficient X-ray transparency while having enough room for a thermocouple (Omega™ 5TC-TT-TI-40), which is in loose contact with the silicon frame of the front window to provide a reading from as close to the sample as possible. The pressurized gap is closed towards the detector side by a second window with a circular frame. The detector-side window, 3 mm in diameter with 0.2 mm frame thickness, was glued to the tapered aluminum insert, which surrounds the detector tip, with the same adhesive as the front window. Consequently, the physical thickness of the microchamber along the optical axis only measures 700 m (compare Fig. c). With a back-window membrane size of m, unobstructed views of the front membrane at the carbon K-edge with differential pressures to up to 500 hPa are possible. Usually membranes with m were used, as vignetting in the outer regions of the sample window was less important to us compared to increased burst resistance at atmospheric . Please note that the inward facing of the silicon nitride windows is crucial for assuring the best resistance, i.e., a safe operation at 1000 hPa differential pressure. An outward facing front window as shown in is prone to delamination of the silicon nitride film and subsequent burst of the membrane.
The rearmost part of the microreactor is a PCB, which is screwed to the brass metal body. It includes read-out electronics (ADS1118 ADC) for the thermocouple, an electrical connector as well as the environmental sensor S2, which reaches into the gas stream through an O-ring-sealed recess in the metal body (Fig. b). A total number of three O-rings seal the microreactor from the microscope. The O-rings were slightly lubricated with Apiezon N Cryogenic High Vacuum Grease to assure a tight seal and a smooth rotation of the sample locking mechanism. This solved the initial problem we had with a different high vacuum grease, which introduced an organic contamination into the samples (Fig. S3). We regularly conduct carbon K-edge spectroscopy to identify beam damage and potential sources of contamination in our analysis but do not detect any impurities originating from microreactor components, even in studies where particles were processed over many hours
3 Performance evaluation
3.1 Parameter control and stability
A wide range of atmospheric conditions present in the troposphere can be reproduced by the microreactor system. More precisely, the system uses pressures ranging from 180 to 1000 hPa and has cooling capabilities for highly stable controlling between room temperature and approximately 250 , depending on the insulation quality in the individual setup. All parameters are actively regulated via feedback control systems. The (controllable) working range for the RH spans from dry conditions to above 98% RH, depending on the residual moisture of the process gas, and the temperature difference between the humidifier and the microreactor body. Due to heat dissipation by electronic components, a control system of 301.2–302.2 , measured inside the aluminum circuit board, is typical. A 65 % RH at sensor S1 directly after the humidifier therefore translates into saturation conditions at a microreactor of 295.2 . Steep and RH gradients (e.g., from dry conditions to above 80 % RH within less than a minute) can be achieved in a reliable and reproducible manner. Besides a stepwise increase in the humidity, it is possible to run preprogrammed ramps or periodically repeated hydration–dehydration cycles to mimic dynamic atmospheric processes.
In Fig. a the RH trend upon a stepwise increase in the setpoint over a time period of min is shown. Noticeable is the discrepancy between the RH measured at the exit of the humidifier (sensor S1) and inside the microreactor body (sensor S2) as a result of the temperature difference and a slowed response of the S2 reading due to diffusion inside the gas stream and the wetting of exposed surfaces.
Figure 4
(a) Response characteristic of the humidifier at different setpoints. Blue: humidity measured at S1 (control variable); orange: humidity measured at S2 (compare with Fig. ) at an ambient temperature between 301.2 and 302.2 and a microreactor of 283.2 . (b) Microreactor stability over min at a setpoint of 283.2 . The gray shading emphasizes a margin.
[Figure omitted. See PDF]
In the particular example shown here, the microreactor was held at 283.2 and the ambient system varied between 301.2 and 302.2 . This experiment can be seen as a typical example for sensing the deliquescence point of ammonium sulfate (AS) at 82 % RH at 283.2 , as reported by .
A control with the least possible fluctuation margin is crucial, especially at high RH close to saturation conditions to avoid undesired water condensation inside the microreactor. Therefore, the control system was designed for high-precision regulation and not for high cooling ( min) or warming rates. In Fig. b, the high stability of the regulator set to a target of 283.2 is shown over a 4 h time span.
The as well as the RH regulation is based on software PI controllers running as scripts inside the “VBUS” software environment. In a test case, as low as 250.2 could be reached inside the microreactor. For details, refer to Fig. S1. Besides kinetic studies for the measurement of reaction rates, with this capability, freezing experiments, and the study of diffusion-controlled reactions at a single-particle level come into reach.
Note that the thermocouple in principle is redundant for the temperature measurement inside the microreactor body, as the BME280 sensor (compare Fig. ) usually gives the same temperature readings, but with a higher resolution. Any heat transfer mechanisms besides radiant heat transfer are minimized within the microreactor. Convective heat transfer is suppressed by the surrounding vacuum and the microreactor body is clamped to the microscope's stage, thermally well insulated by MACOR® ceramic posts, which can be seen in Fig. a. However, for very low temperatures, a discrepancy of a few tenths of a Kelvin at 283.2 between both sensors was present, increasing when temperatures are lowered. We attribute this to the radiant heat transfer between the OSA and the sample window.
3.2 Sensor calibration and initial testsThe BME280 environmental sensors used in this setup were chosen because of their very small dimensions of . The BME280 sensors' and the thermocouple's temperature readings were calibrated in a cooling bath between 263.2 and 300.2 against a reference thermometer (Fluke 2180A, Fluke Deutschland GmbH with 0.01 resolution and a minimum uncertainty of . The humidity readings were calibrated using the deliquescence-relative humidities (DRHs) of salt standards documented in the literature, e.g., , , and with DRH values of % RH , 80 % RH , and % RH , respectively. For test measurements, the setup was successfully adapted to two STXM instruments, (i) MAXYMUS, located at beamline UE46-PGM-2 of the synchrotron BESSY II, Berlin, Germany , and (ii) PolLux, located at beamline X07DA of the Swiss Light Source in the Paul Scherrer Institute, Villigen, Switzerland .
Figure 5
Image sequence showing the hydration (ac) and dehydration (cd) of . The RH values here represent raw sensor data. Panel (a) represents dry conditions, (b) shows the situation shortly before deliquescence. An aqueous shell has already formed around the particles and crystal edges are rounded off. In panel (c) full deliquescence has already occurred. The deliquescence point was reached between 99 % RH and 100 % RH at 294.2 K. This is in good agreement with the literature value of % RH , taking into account the sensor accuracy. Efflorescence was observed during the scan of the image in panel (d) with a sharp transition (red arrow) at %RH, in contrast to the literature value of 60 % RH, reported by . Note that white arrows indicate the direction of the raster scan pattern during the measurements. The images in the left column (a, d) represent raw X-ray absorption data at 528 eV and in the right column (b, c) optical density (OD) maps are shown to emphasize the oxygen distribution. The maps were generated with the Multivariate ANalysis Tool for Spectromicroscopy software (MANTiS v.3.0.0.1) in commit version no. 5847171 .
[Figure omitted. See PDF]
A hydration–dehydration experiment with particles was measured at PolLux as a first proof-of-performance application for the microreactor system. The results shown in Fig. illustrate the operation of the system in the high RH regime as reliable humidity control was possible even beyond the high-deliquescence RH of . The humidity values in the individual panels of the figure are uncorrected and represent averaged raw sensor data. The water uptake can be easily recognized from the changes in particle morphology and by the optical density (OD) increase at the oxygen K-edge. (For a definition of the OD, please refer to .) Full deliquescence occurred between 99 % RH and 100 % RH (between Fig. b and c), which agrees well with the aforementioned literature value taking into account the sensor's uncertainty of at least % RH (specified only up to 80 % RH by ). The efflorescence happened suddenly during dehydration at % RH (see red arrow in Fig. d). In the literature, a value of 60 % RH was reported for the efflorescence relative humidity (ERH) by . This deviation might be attributed to the presence of the silicon nitride substrate but likely is a result of the hysteresis of the sensor, which was operated under fairly extreme conditions here and needs up to 10 min to equilibrate after operation close to saturation conditions. In general, it is therefore recommended to take temperature readings, and the RH values of the two other environmental sensors into consideration when evaluating the humidity inside the microreactor, instead of relying on just one sensor.
A pressure calibration was not done for the measurements presented here, as the RH is independent from . Furthermore, the absolute accuracy of hPa, as reported by , was considered sufficiently accurate. However, note in this context that using helium as a process gas can cause a gradual drift of the measured pressure values, due to helium permeating the sensor, as was reported by .
Figure 6
Hygroscopic growth curve of ammonium sulfate (AS, ) based on two datasets with four AS particles each, ranging from 500 nm to 2 m in size. The image sequence above at different humidities, taken from , shows similar AS particles at the oxygen pre-edge (528 eV) and illustrates the water uptake and the Ostwald ripening. The oxygen mass uptakes of individual particles were calculated from OD maps (527 vs. 560 eV) at the oxygen K-edge. All data were normalized to the dry particles (encircled data points). Only masses from oxygen-containing species contributed to the AIM model II curves . Error bars represent the sensor accuracy ( % RH below and % RH above 80 % RH) and an estimated measurement error ( oxygen mass growth units).
[Figure omitted. See PDF]
As a second proof-of-performance application, presented in Fig. , the hygroscopic growth curve of AS was recorded using fine-pitched RH steps at the MAXYMUS instrument. The figure contains data from two independent measurements on four AS particles each. Again, the increase in optical density at the oxygen K-edge served as a measure for the water uptake, similar to the measurements by and the water uptake study by . A good agreement with the AIM model II by for the deliquescence point and the overall trend was found. In terms of representative statistics of the number of analyzed particles, methods based on STXM-NEXAFS are inherently limited by comparatively long scan times.
Figure 7
Tentative observation of ice at in an aqueous isoprene SOA particle. (a) Optical density map at the oxygen K-edge composed from five single images at different energies: 528.2 and 532.0 eV (pre-edge) and 544.2, 546.2, and 553.2 eV (post-edge). Results from the non-negative matrix approximation (NNMA) analysis are displayed in panels (b1), (b2), (c1), and (c2). The cluster thickness maps (weights for hyperspectral pixels) are shown in panels (b1) for the matrix component (aqueous phase) and in (b2) (ice-enriched phase). a.u. stands for arbitrary units. Panels (c1) and (c2) show the corresponding spectra along with reference spectra. Accordingly, the three main spectral features are labeled I, II, and III and emphasized by a gray background similar to and . All spectra were shifted by 3.2 eV to larger energy values with respect to the reference spectra. The OD map, the thickness maps and the spectral components were extracted from raw data by using the NNMA analysis feature included in MANTiS v.3.0.0.1 , committed version no. 5847171 . (NNMA parameters chosen: ; spectra similarity ; spectra smoothness ; sparseness ; number of iterations ; delta error threshold ).
[Figure omitted. See PDF]
As a rough measurement time estimate, the recording of a full hydration–dehydration cycle with 22 RH steps took 2 h in the case of dataset 1. The scan time itself was about 2 min at each RH step for recording images at two different energies (65 nm pixel size, 1 ms dwell time per pixel on a px area). The remaining time was spent on waiting for the RH to stabilize, which is particularly important during dehydration experiments and at high relative humidities, as was mentioned above.
Accordingly, some of the results on the standard compounds shown here can be obtained more efficiently and probably more precisely with other techniques, such as hygroscopicity tandem differential mobility analyzer (HTDMA) systems , and differential mobility analyzers coupled to a humidified centrifugal particle mass analyzer (DMA-HCPMA) . Single-particle traps, e.g., the electrodynamic balance (EDB) , and the aerosol optical tweezers are also highly accurate, well-established techniques and do not require a substrate . Therefore, the purpose of showing these results here is to illustrate the analytical capabilities of the system, particularly its operation at high RH.
As a third proof-of-performance application, a water freezing experiment with isoprene SOA particles at high RH was conducted. These results illustrate that the system can be used for controlled freezing experiments as well as for kinetic deceleration of fast processes. The particles were produced using a potential aerosol mass (PAM) chamber in the presence of AS seed particles to increase the SOA yield, as described by , and subsequently impacted on a silicon nitride membrane window. In the course of the experiment, starting from , the target temperature of was reached within an hour with an initial average cooling rate of about 1 K min between and and with the humidified gas flow fully enabled to initiate a fast hygroscopic particle growth. The maximum cooling rate achieved with this system was 2.5 K min. At the desired temperature the gas flow through the microreactor was stopped by closing the two shut-off valves (compare Fig. ) to prevent more water from condensing and to prevent ice from nucleating inside the microreactor body and potentially blocking the gas channel. As the RH reading of sensor S2 quickly went into saturation during the experiment, we can only estimate the relative humidity to be likely above 95 % at the sample location. Subsequently, in-depth X-ray microspectroscopic analysis at the oxygen K-edge of one exemplary SOA particle revealed an embedded crystalline structure (Fig. a). The recorded hyperspectral NEXAFS data obtained from a subregion of the entire particle were separated by the NNMA analysis feature of the MANTiS software into two main spectral components, which are clearly localized, either at the matrix or the embedded structure, visualized by the cluster thickness maps (Fig. b1 and b2). A comparison of the corresponding spectra (Fig. c1 and c2) with reference spectra of liquid water and ice suggests that the spectral components represent coexisting water-enriched and ice-enriched phases. Their spectra differ significantly from the spectrum of isoprene SOA under dry conditions (orange spectrum in Fig. c1), which was obtained from the same sample prior to wetting and also does not match the AS reference spectrum (orange spectrum in Fig. c2) taken from . We therefore assume that the observed droplet is highly diluted with water and that the characteristic reversal of the intensity ratios of the spectral features II and III (compare blueish spectra in Fig. c1 with c2), as a result of a change in coordination geometry and bond strength as observed by, for example, , and , can to a large extent be attributed to molecules only. The crystalline structure could be observed over many hours and remained stable in shape even though the temperature fluctuated by 0.2 K. During warming up from 261 to 269 K at a rate of 2.7 K min a restructuring of the crystalline structure could be observed. (The warming rate largely depends on the quality of the surrounding vacuum since no active heating is involved.)
It should be noted that it is unlikely to find immersion freezing in aqueous isoprene SOA (seeded with AS), as it occurred at an unusually high temperature of , compared to model predictions
Furthermore, MIMiX was used in a study on the diffusion-limited oxidation of in particles composed of xanthan gum and by ozone supplied by an external ozone generator (compare Fig. ). In this study, a direct comparison between in-situ-observed and modeled chemical changes in micron- and submicron-sized particles was achieved. This study shows the potential of STXM-NEXAFS in combination with MIMiX towards the in situ investigation of atmospherically relevant multiphase reactions and may inspire more studies such as this on the topic of diffusion limitation and surface reactions.
One of the real analytical strengths of STXM-NEXAFS analysis in combination with MIMiX emerges in the analysis of ambient aerosol particles, since detailed single-particle studies can typically not be conducted on site, particularly at remote locations. Thus, sampling of ambient particles onto suitable substrates for a subsequent investigation by offline techniques is required. Such samples are well-suited for in-depth studies with the MIMiX system. Another analytical strength of the system relates to the unique combination of microstructural, hygroscopic, and chemical information, which can be obtained on the level of individual particles in the submicron particle size range, while being relatively damage-free through the use of soft X-rays in a dose-efficient scanning system.
4 ConclusionsThis study presents the design, construction, and initial testing of a microreactor system for in situ STXM-NEXAFS analyses of aerosol particles under controlled environmental conditions. Its compact size ensures high portability of the setup, without sacrificing functionality. The operating ranges cover a wide spectrum of , , and RH conditions, representing large parts of the troposphere. Due to the integrated cooling system, the accessible and RH are wider than the corresponding ranges in previously reported setups. Moreover, through a compact design of the microreactor, a measurement of the essential environmental parameters very close to the sample has been realized. The microreactor can be operated safely at atmospheric pressure inside the sample chamber, i.e., at a differential pressure of at least 1000 hPa relative to the microscope chamber. Despite significant spatial limitations in the STXM optics, the microreactor has been kept compatible to a variety of STXM instruments. For initial measurements, it has been installed at the Helmholtz-Zentrum Berlin (BESSY/MAXYMUS) and at the Swiss Light Source (SLS/PolLux). The sample exchange mechanism allows quick and convenient substrate changes and minimizes mechanical stress on the fragile samples.
The results from initial experiments (i.e., hygroscopic growth of , deliquescence and efflorescence of , as well as the observation of water ice in an aqueous isoprene SOA droplet) confirm that the microreactor is a promising and flexible tool for a variety of in situ particle processing studies in environmental STXM experiments. In particular, the system allows controlled studies under high RH and/or low conditions, which are relevant for in-depth investigation of various atmospheric processes.
We intend for the future development of the MIMiX system to include an extension of the cooling capabilities for in situ ice nucleation observations and the introduction of an optical fiber to study photochemically driven multiphase reactions. On the software side, integration with the Experimental Physics and Industrial Control System (EPICS) and the Pixelator software is planned in order to store environmental parameters in parallel with the X-ray microscopic data at per-pixel resolution.
Data availability
The STXM-NEXAFS data used for Figs. , , and ; the NNMA analysis results; and the corresponding spectra have been deposited in Edmond, the Max Planck Society's open-access data repository under 10.17617/3.39 (). For specific data requests beyond the deposited data, please contact the corresponding authors.
Video supplement
A video of the microreactor assembly can be found in the same repository as the scientific data and is available in 720p and 1080p resolution under 10.17617/3.39 ().
The supplement related to this article is available online at:
Author contributions
JDF was responsible for the mechanical and electrical design of the microreactor, the conceptual design of the control system, and its assembly. CP supervised the construction work. CG and ML developed the microcontroller-based electronic system and helped JDF with the programming of scripts for the graphical user interface. MA, SSS, MW, and BW were consulted in an early design stage and influenced the final design of the microreactor. HT, CP, and JDF prepared the samples. JDF led the writing of the paper. CP, MOA, and UP supervised the paper writing. The adaptation of the microreactor to the STXM instruments were conducted by JDF, CP, and MOA, with the technical assistance of MW, BW, and JR. The measurements were led by JDF, CP, and MOA and supported by FD and PAA. All authors contributed to the paper finalization.
Competing interests
The authors declare that they have no conflict of interest.
Acknowledgements
The authors gratefully acknowledge the support provided by the Max Planck Society (MPG). The authors thank Thomas Kennter, Frank Kunz, and the MPIC's mechanical workshop team for their excellent work. We acknowledge the Helmholtz-Zentrum Berlin, Germany, for the allocation of the synchrotron radiation beamtime at BESSY II and the Paul Scherrer Institute, Villigen, Switzerland, for provision of synchrotron radiation beamtime at the PolLux beamline of the SLS. The PolLux endstation was financed by the Federal Ministry of Education and Research (BMBF) through contracts 05KS4WE1/6 and 05KS7WE1. We thank Michael Bechtel and Blagoj Sarafimov for technical assistance during the beamtimes. We further thank David Walter and Nina Löbs for being part of our experiment team and Frank Helleis, Ralf Wittkowski, Mario Birrer, Thomas Berkemeier, Stefan Blanckart, and Berthold Kreuzburg for their support and stimulating discussions.
Financial support
The article processing charges for this open-access publication were covered by the Max Planck Society.
Review statement
This paper was edited by Mingjin Tang and reviewed by two anonymous referees.
You have requested "on-the-fly" machine translation of selected content from our databases. This functionality is provided solely for your convenience and is in no way intended to replace human translation. Show full disclaimer
Neither ProQuest nor its licensors make any representations or warranties with respect to the translations. The translations are automatically generated "AS IS" and "AS AVAILABLE" and are not retained in our systems. PROQUEST AND ITS LICENSORS SPECIFICALLY DISCLAIM ANY AND ALL EXPRESS OR IMPLIED WARRANTIES, INCLUDING WITHOUT LIMITATION, ANY WARRANTIES FOR AVAILABILITY, ACCURACY, TIMELINESS, COMPLETENESS, NON-INFRINGMENT, MERCHANTABILITY OR FITNESS FOR A PARTICULAR PURPOSE. Your use of the translations is subject to all use restrictions contained in your Electronic Products License Agreement and by using the translation functionality you agree to forgo any and all claims against ProQuest or its licensors for your use of the translation functionality and any output derived there from. Hide full disclaimer
© 2020. This work is published under https://creativecommons.org/licenses/by/4.0/ (the “License”). Notwithstanding the ProQuest Terms and Conditions, you may use this content in accordance with the terms of the License.
Abstract
The dynamic processing of aerosols in the atmosphere is difficult to mimic under laboratory conditions, particularly on a single-particle level with high spatial and chemical resolution. Our new microreactor system for X-ray microscopy facilitates observations under in situ conditions and extends the accessible parameter ranges of existing setups to very high humidities and low temperatures. With the parameter margins for pressure (180–1000 hPa), temperature (
You have requested "on-the-fly" machine translation of selected content from our databases. This functionality is provided solely for your convenience and is in no way intended to replace human translation. Show full disclaimer
Neither ProQuest nor its licensors make any representations or warranties with respect to the translations. The translations are automatically generated "AS IS" and "AS AVAILABLE" and are not retained in our systems. PROQUEST AND ITS LICENSORS SPECIFICALLY DISCLAIM ANY AND ALL EXPRESS OR IMPLIED WARRANTIES, INCLUDING WITHOUT LIMITATION, ANY WARRANTIES FOR AVAILABILITY, ACCURACY, TIMELINESS, COMPLETENESS, NON-INFRINGMENT, MERCHANTABILITY OR FITNESS FOR A PARTICULAR PURPOSE. Your use of the translations is subject to all use restrictions contained in your Electronic Products License Agreement and by using the translation functionality you agree to forgo any and all claims against ProQuest or its licensors for your use of the translation functionality and any output derived there from. Hide full disclaimer
Details










1 Multiphase Chemistry Department, Max Planck Institute for Chemistry, 55128 Mainz, Germany; Biogeochemistry Department, Max Planck Institute for Chemistry, 55128 Mainz, Germany
2 Instrument Development Group, Max Planck Institute for Chemistry, 55128 Mainz, Germany
3 Multiphase Chemistry Department, Max Planck Institute for Chemistry, 55128 Mainz, Germany
4 Department of Environmental Sciences, University of Basel, 4001 Basel, Switzerland; now at: Department of Environmental Science, Stockholm University, 106 91 Stockholm, Sweden
5 Laboratory of Environmental Chemistry, Paul Scherrer Institute, 5232 Villigen PSI, Switzerland
6 Laboratory Condensed Matter Physics, Paul Scherrer Institute, 5232 Villigen PSI, Switzerland
7 Institute for Nanospectroscopy, Helmholtz-Zentrum Berlin für Materialien und Energie GmbH, 12489 Berlin, Germany
8 Biogeochemistry Department, Max Planck Institute for Chemistry, 55128 Mainz, Germany; Scripps Institution of Oceanography, University of California San Diego, La Jolla, CA 92037, USA