MR image‐guided radiotherapy has the potential to improve patient care, but integration of an MRI scanner with a linear accelerator adds complexity to the commissioning process. This work describes a single institution experience of commissioning an Elekta Unity MR‐linac, including mechanical testing, MRI scanner commissioning, and dosimetric validation. Mechanical testing included multileaf collimator (MLC) positional accuracy, measurement of radiation isocenter diameter, and MR‐to‐MV coincidence. Key MRI tests included magnetic field homogeneity, geometric accuracy, image quality, and the accuracy of navigator‐triggered imaging for motion management. Dosimetric validation consisted of comparison between measured and calculated PDDs and profiles, IMRT measurements, and end‐to‐end testing. Multileaf collimator positional accuracy was within 1.0 mm, the measured radiation isocenter walkout was 0.20 mm, and the coincidence between MR and MV isocenter was 1.06 mm, which is accounted for in the treatment planning system (TPS). For a 350‐mm‐diameter spherical volume, the peak‐to‐peak deviation of the magnetic field homogeneity was 4.44 ppm and the geometric distortion was 0.8 mm. All image quality metrics were within ACR recommendations. Navigator‐triggered images showed a maximum deviation of 0.42, 0.75, and 3.0 mm in the target centroid location compared to the stationary target for a 20 mm motion at 10, 15, and 20 breaths per minute, respectively. TPS‐calculated PDDs and profiles showed excellent agreement with measurement. The gamma passing rate for IMRT plans was 98.4 ± 1.1% (3%/ 2 mm) and end‐to‐end testing of adapted plans showed agreement within 0.4% between ion‐chamber measurement and TPS calculation. All credentialing criteria were satisfied in an independent end‐to‐end test using an IROC MRgRT phantom.

Details

Title
Commissioning of a 1.5T Elekta Unity MR‐linac: A single institution experience
Author
Snyder, Jeffrey E 1 ; Joël St‐Aubin 1 ; Sridhar Yaddanapudi 1 ; Boczkowski, Amanda 1 ; David A.P. Dunkerley 1 ; Graves, Stephen A 2 ; Hyer, Daniel E 1 

 Department of Radiation Oncology, University of Iowa, Iowa City, IA, USA 
 Department of Radiology, University of Iowa, Iowa City, IA, USA 
Pages
160-172
Section
MEDICAL IMAGING
Publication year
2020
Publication date
Jul 2020
Publisher
John Wiley & Sons, Inc.
e-ISSN
15269914
Source type
Scholarly Journal
Language of publication
English
ProQuest document ID
2427513996
Copyright
© 2020. This work is published under http://creativecommons.org/licenses/by/4.0/ (the “License”). Notwithstanding the ProQuest Terms and Conditions, you may use this content in accordance with the terms of the License.