Abstract
Background
Hypoxia is common during daily nursing procedures (DNPs) routinely performed on mechanically ventilated patients. The impact of automated ventilation on the incidence and severity of blood oxygen desaturation during DNPs remains unknown.
Methods
A prospective randomized controlled crossover trial was carried out in a French intensive care unit to compare blood oxygen pulse saturation (SpO2) during DNPs performed on patients mechanically ventilated in automated and conventional ventilation modes (AV and CV, respectively). All patients with FiO2 ≤ 60% and without prone positioning or neuromuscular blocking agents were included. Patients underwent two DNPs on the same day using AV (INTELLiVENT-ASV®) and CV (volume control, biphasic positive airway pressure, or pressure support ventilation) in a randomized order. The primary outcome was the percentage of time spent with SpO2 in the acceptable range of 90–95% during the DNP.
Results
Of the 265 included patients, 93% had been admitted for a medical pathology, the majority for acute respiratory failure (52%). There was no difference between the two periods in terms of DNP duration, sedation requirements, or ventilation parameters, but patients had more spontaneous breaths and lower peak airway pressures during the AV period (p < 0.001). The percentage of time spent with SpO2 in the acceptable range during DNPs was longer in the AV period than in the CV period (48 ± 37 vs. 43 ± 37, percentage of DNP period; p = 0.03). After adjustment, AV was associated with a higher number of DNPs carried out with SpO2 in the acceptable range (odds ratio, 1.82; 95% CI, 1.28 to 2.6; p = 0.001) and a lower incidence of blood oxygen desaturation ≤ 85% (adjusted odds ratio, 0.50; 95% CI, 0.30 to 0.85; p = 0.01).
Conclusion
AV appears to reduce the incidence and severity of blood oxygen desaturation during daily nursing procedures (DNPs) in comparison to CV.
Trial registration
This study was registered in clinical-trial.gov (NCT03176329) in June 2017.
You have requested "on-the-fly" machine translation of selected content from our databases. This functionality is provided solely for your convenience and is in no way intended to replace human translation. Show full disclaimer
Neither ProQuest nor its licensors make any representations or warranties with respect to the translations. The translations are automatically generated "AS IS" and "AS AVAILABLE" and are not retained in our systems. PROQUEST AND ITS LICENSORS SPECIFICALLY DISCLAIM ANY AND ALL EXPRESS OR IMPLIED WARRANTIES, INCLUDING WITHOUT LIMITATION, ANY WARRANTIES FOR AVAILABILITY, ACCURACY, TIMELINESS, COMPLETENESS, NON-INFRINGMENT, MERCHANTABILITY OR FITNESS FOR A PARTICULAR PURPOSE. Your use of the translations is subject to all use restrictions contained in your Electronic Products License Agreement and by using the translation functionality you agree to forgo any and all claims against ProQuest or its licensors for your use of the translation functionality and any output derived there from. Hide full disclaimer