1 Introduction
The nighttime chemistry in polluted urban or suburban areas has been proved to be governed by NO radicals since its discovery in the 1980s (Naudet et al., 1981; Noxon et al., 1978, 1980; Platt et al., 1980). In particular, the NO radical has been shown to be an efficient oxidant for some organic compounds, or in some cases even the dominant one, thus impacting the budget of these species and their degradation products. Unsaturated volatile organic compounds (VOCs), including biogenic VOCs (BVOCs), are particularly reactive towards NO radicals (Wayne et al., 1991). Providing kinetic data for these reactions is essential for a better understanding of the role of NO radicals in their degradation. Nevertheless, due to the high reactivity of some unsaturated VOCs with NO (with rate constants which can reach 10 to 10 cm molecule s), an absolute rate determination for these reactions appears to be difficult as it requires the use of a highly sensitive method for NO monitoring. As a consequence, the number of absolute kinetic studies for the NO-initiated oxidation of terpenes is very limited, and this leads to large uncertainties on this chemistry as has been pointed out in the literature (Atkinson, 2000; Brown and Stutz, 2012; Ng et al., 2017). Calvert et al. (2015) gave recommendations for NO oxidation rate constants for 91 alkenes (ranging between 10 and 10 cm molecule s), and more than 98 % of the determinations on which these recommendations are based were conducted using the relative rate method. One of the reasons for this is still the challenging measurement of NO radicals at low mixing ratios ( 100 ppt) during such experiments. For these compounds, new absolute determinations are essential to better evaluate the role of NO radicals in their degradation.
Among the various experimental tools which are currently used to measure rate constants, atmospheric simulation chambers represent suitable tools for performing experiments under very realistic atmospheric conditions. This implies low concentrations of reactants in order to minimize possible secondary reactions. Another benefit of these facilities is the high analytical capabilities which allow for the in situ monitoring of reactants and products with a high sensitivity. Even though significant progress has been made in the last decades for NO radicals' measurement at low concentrations with the rise of cavity enhanced and cavity ring-down spectroscopy (CRDS) techniques (Ball et al., 2004; Bitter et al., 2005; Kennedy et al., 2011; Langridge et al., 2008) as well as laser-induced fluorescence techniques (Matsumoto et al., 2005b, c, a; Wood et al., 2003), it is observed that only few were coupled to simulation chambers (Dorn et al., 2013; Venables et al., 2006; Wu et al., 2014). In addition, to the best of our knowledge, none of these techniques has been used for kinetic applications involving NO radicals in simulation chambers.
For this purpose, the analytical capabilities of the CSA chamber available at LISA have been improved by developing a sensitive technique for measuring NO radicals at very low concentration. An incoherent broadband cavity enhanced absorption spectroscopy (IBBCEAS) technique has been coupled to the chamber with the objective of performing high-sensitivity in situ NO monitoring with an integration time of seconds.
In this paper, we describe the experimental setup and the characterization of the technique. Finally, the IBBCEAS technique has been validated thanks to an intercomparison of NO and NO measurement with the Fourier-transform infrared (FTIR) technique and an absolute rate determination for the well-known reaction trans-2-butene NO.
2 Experimental section
2.1 The CSA chamber
The CSA chamber is made of a large and evacuable Pyrex® reactor (6 m length, 45 cm diameter and 977 L volume) which has been previously presented in detail (Doussin et al., 1997). It is equipped with a homogenization system which is made of (i) an injection pipe (4 m long, 1 cm diameter and regularly drilled with 1 mm holes), which allows the reactants to be injected all along the chamber; (ii) two stainless steel fans, allowing for a homogenization of gas inside the chamber; and (iii) a closed-circuit Teflon pump connected at both ends, allowing for a recirculation of the gas mixing. This system allows a mixing time below 1 min. The chamber is also equipped with two in situ spectroscopic analytical devices coupled with White-type multiple reflection systems inside the reactor: (i) a FTIR spectrometer (Bruker Vertex 80), which allows spectra to be acquired in the range of 600–4000 cm with a maximal spectral resolution of 0.07 cm and an optical path length of 204 m; and (ii) an UV–Visible grating spectrometer, consisting of a high-pressure Xenon arc lamp (Osram XBO, 450 W Xe UV), an HR 320 monochromator (Jobin-Yvon) and a CCD camera (CCD 3000, 1024 58 pixel, Jobin-Yvon) as a detector. This spectrometer allows spectra to be acquired with a spectral resolution of 0.18 nm and an optical path length of 72 m.
This facility has intensively been used to investigate complex gas-phase chemistry involving organic compounds and to provide kinetic and mechanistic data. In particular, it has been used for absolute rate determination of reactions involving NO radicals with a series of VOCs like ethers, esters and aldehydes (Kerdouci et al., 2012; Picquet-Varrault et al., 2009; Scarfogliero et al., 2006). In these studies, NO was monitored at 662 nm with the in situ UV–Visible spectrometer. However, due to the poor detection limit (0.5 ppb for 1 min of acquisition), and taking into account the experimental conditions, the range of rate constants that can be investigated is limited ( 10 cm molecule s), preventing very reactive chemical systems from being studied, such as BVOC NO reactions.
2.2 The IBBCEAS setup
In order to improve the analytical capabilities of the CSA chamber, an IBBCEAS has been developed and coupled to the chamber for high-sensitivity in situ NO monitoring. A detailed description of the technique has been provided in previous works (Langridge et al., 2008; Romanini et al., 1997). IBBCEAS measurements are conducted by exciting, with an incoherent broad-band source, a high-finesse optical cavity formed by two mirrors with high reflectivity ( 99.98 %). Photons resonate between the two mirrors, increasing their lifetime in the cavity by a factor of . During this time, photons traverse an effective path length of kilometers inside the cavity, making observations of absorbing species at very low concentrations possible. The intensity transmitted by the optical cavity rapidly reaches a steady state. The optical intracavity absorption coefficient of the sample can then be calculated with the following expression if an accurate measurement of the cavity reflectivity and of the distance between the mirrors () is provided (Engel et al., 1998):
1 Here and are the transmitted intensities measured in the presence and in the absence of the absorbing species respectively.
The concentrations of the absorbing molecules can then be calculated using a least-squares algorithm to simultaneously fit the molecules' absorption cross section using the formula: 2 where are the absorption cross sections, are the species absorbing in the considered spectral region and is a cubic polynomial to correct baseline deformations due to potential variations of the source intensity (Venables et al., 2006) or to absorption and/or scattering of particles in the chamber (Varma et al., 2013).
The optical cavity is made of two high-reflectivity mirrors (Layertec, planar/concave mirrors with a 1 m radius of curvature, nominal reflectivity of 99.98 0.01 % between 630 and 690 nm). It has been transversally installed on the CSA chamber using two co-axial outputs of the reactor. A scheme of the IBBCEAS instrument interfaced to the chamber is shown in Fig. 1. The distance between the mirrors is 82 cm and includes 45 cm for the chamber diameter, 2 10.5 cm for the Pyrex outputs and 2 8 cm for the interface mounts between the chamber and the commercial CRD Optics mount support. In order to prevent the adsorption of semi-volatile species or deposition of particles on the mirrors, which would result in a significant decrease of the reflectivity, the mirrors can be protected thanks to a continuous nitrogen flush (flow rate: 300 mL min) using a in. (1.59 mm) input disposed close to the mirror surface. This flow rate has been optimized in order to efficiently protect the mirrors while limiting the dilution of the mixture in the measurement area. By comparing the absorption coefficients measured with and without the flush, for a known quantity of NO in the chamber, this effective length was estimated to be 62 3 cm (i.e., 24 % lower than the physical length of the cavity). A nitrogen flush was not used in this study but is available for further types of experiments. Thanks to the mixing system which ensures a fast homogenization of the mixture in the chamber, this effective length was observed to be constant during the whole duration of the experiment.
Figure 1
Scheme (transverse section) of the IBBCEAS instrument interfaced to the CSA chamber. Collimating lenses and curved mirrors are not shown on the beam injection side.
[Figure omitted. See PDF]
Figure 2
(a) NO and NO absorption cross sections (Orphal et al., 2003; Vandaele et al., 1997) between 640 and 680 nm (convolved with apparatus function of the spectrometer) and LED emission spectrum and (b) mirror reflectivity.
[Figure omitted. See PDF]
A light emitting diode (LED; Mouser Electronics – Starboard, Luminus SST-10-DR-B130 Deep Red, K, max. 430 mW, D5) with an approximative Gaussian-shaped emission of 19 nm full width at half maximum (FWHM) and centered at 662 nm was used in order to monitor NO at its maximum absorption wavelength. The LED emission spectrum is compared to the cross section spectra of NO (Orphal et al., 2003) and NO (Vandaele et al., 1997) in Fig. 2. The spectral range of the LED is large enough to allow both NO and NO to be monitored.
The LED is mounted on a thermoelectric controller (TEC) device (ThermoElectric Cooling, Laser Mount Arroyo Instruments) to ensure a very precise temperature regulation ( 0.01 C) and stabilize the spectral distribution of the LED. With this device, changes in the LED intensity have been observed to be below 0.3 %. A laser diode controller (Arroyo Instrument 6310) provides the electric power for both TEC and LED (LED current is fixed at 900 mA). Light emitted by the LED is spatially incoherent, and collimation is required for an effective coupling with the optical cavity. The light is hence focused with a convex lens (Thorlabs Aspheric Condenser Lens, 25.4 mm diameter, 16 mm, NA 0.79) and injected into the optical cavity with two concave mirrors (Thorlabs, protected silver, diameter 50.8 mm, 50 mm and diameter 75 mm, 500 mm respectively) in order to focus the beam in the middle of the cavity. Light transmitted through the cavity is directed thanks to a collimator (Thorlabs SMA Fiber Collimation Pkg, 635 nm, 35.41 mm, NA 0.25) and an optical fiber (Ocean Optics Vis–NIR, 200 m slit, 5 m long) to a miniature Ocean Optics QE65000 spectrometer. The spectrometer measures the cavity output wavelength-dependent intensity and comprises a spectrograph interfaced to a charged coupled device (CCD) thermally stabilized at 15 C to minimize the dark current. The spectral range covered by the spectrometer is 45 nm (640–685 nm) with a spectral resolution of 0.2 nm. In order to calculate the concentrations of absorbing species, a data treatment program has been developed in R (Ihaka and Gentleman, 1996) using a least-squares algorithm. A third-degree polynomial function is used in the fit to take into account baseline deformation due to small changes in the source intensity. The concentrations of the absorbing species and the polynomial function are fitted by minimizing the RMSE (root mean square error). In practice, the optimization was run following a bound optimization using the quadratic approximation (BOBYQA) method (Powell, 2009). The iterative process to minimize the RMSE (between absorption coefficients from Eqs. 1 and 2) stops when none of the parameters vary more than 0.2 % between two successive iterations. In the studied spectral range, absorbing species are HO, NO and NO. Absorption by water vapor may be very high under atmospheric conditions. Nevertheless, due to dry conditions used during the experiments (RH 1 %), HO absorption was negligible here. The absorption cross sections used are provided by the literature (Orphal et al., 2003; Vandaele et al., 1997) and have been convoluted with the apparatus function of the instrument. Because NO cross sections provided by Vandaele et al. (1997) are measured up to 666.5 nm, the treatment has systematically been conducted up to this value. The cross sections used for the data treatment are presented in Fig. 2.
3 Technique qualification and characterizationSeveral experiments have been carried out to assess the stability, the accuracy and the detection limit of the technique. First experiments have been conducted to test the optical stability and the influence of pressure variations on the device. These tests have shown that the instrument is very stable (variations 1 %). Two aspects have been shown to be particularly critical for measurement with IBBCEAS technique: measurement and the determination of mirror reflectivity.
3.1 Determination of the cavity reflectivity
Having a precise knowledge of the wavelength-dependent mirror reflectivity, , is one of the most critical points of the IBBCEAS technique (Venables et al., 2006). Two different methods have been proposed for accurate determination of : (i) measurement of a known concentration of an absorbing species (Ventrillard-Courtillot et al., 2010) and (ii) measurement of the ring-down time in the empty cavity using a pulsed laser CRDS technique (Ball et al., 2004). The first method has been employed here. The absorbing species which has been chosen for the experiments is NO as it absorbs in the whole spectral region of the LED emission, and its absorption cross sections are known with high accuracy. NO concentrations in the simulation chamber were obtained from in situ FTIR measurements using IBI (1530–1680 cm) (5.6 0.2) 10 cm molecule (base ). To retrieve the mirror reflectivity from NO concentration, the following equation is used:
3 where is the NO absorption cross section (Vandaele et al., 1997), and [NO] is the concentration of NO determined by FTIR. In order to reduce the uncertainty on the reflectivity determination and to compensate the weak cross sections of NO in the 660–670 nm region, high concentrations (up to 800 ppb) were used. A plot showing the variation of the reflectivity in function of the wavelength is presented as an example in Fig. 2. Due to the NO reference spectrum, the reflectivity is measured up to 666.5 nm. During this experiment, the reflectivity was observed to vary between 99.975 at 640 nm and 99.974 % at 667 nm, and this is in agreement with the reflectivity provided by the supplier (99.98 0.01 %). It was found to have a slight dependence on wavelengths (), which justifies that it is necessary to measure it in a wide wavelength range. At 662 nm, which corresponds to the maximum of NO absorption, the reflectivity was found to be 99.974 %. At this wavelength, the effective absorption path length estimated from Eq. (4) is found to be 3.15 km: 4 In addition, it has been observed that the reflectivity of the cavity can significantly vary with the mirror cleanliness. As an example, successive experiments showed that reflectivity can vary at 662 nm from 99.974 % to 99.971 % from one experiment to another, leading to variations of the effective absorption path length of almost 12 %. Therefore, it is crucial to precisely determine the reflectivity prior to each experiment.
3.2measurement
Previous studies (Fuchs et al., 2010; Kennedy et al., 2011; Ventrillard-Courtillot et al., 2010) have pointed out that the has to be periodically recorded during an experiment to ensure accurate measurement with the IBBCEAS technique. Indeed, changes in the lamp emission spectrum or poor optical stability may induce changes in the absorption coefficient and therefore generate errors in the quantification of the species. This fact may be an issue for experiments in simulation chambers as the can only be recorded before injecting the reactants, and experiments can then last for several hours.
In order to evaluate the stability of the signal during a typical experiment and the uncertainty generated by the use of a unique on the quantification of the species, two types of experiments have been performed: first, the stability of the optical system was verified. For this purpose, long-term measurements of the signal were performed, leading to variations lower than 0.3 % and to very small baseline distortions. Then, to verify the impact of these variations on the quantification of absorbing species, experiments were conducted by injecting NO, which was observed to be stable in the chamber, and by monitoring its concentration with the IBBCEAS technique for several hours: after the chamber was filled with a mixture of N O () at atmospheric pressure, the was measured. Then, a concentration of NO was introduced into the chamber (mixing ratios ranging between 100 ppb and 1 ppm, depending of the experiment), and the signal was measured. From these measurements, the absorption coefficient and the NO concentration were calculated and plotted as a function of time in Fig. 3, for a concentration of 1 ppm of NO. An increase up to 3 % in NO concentration was observed 2 h after recording the due to the deviation of the baseline which is no longer well corrected by the polynomial function. These results suggest that the accuracy of the measurement is significantly reduced after 2 h. The length of the experiments should therefore not exceed this duration. Above this limit the uncertainty due to a unique measurement of the can be considered as being negligible.
Figure 3
Evolution of fit quality for 1 ppm of NO as a function of time since was measured.
[Figure omitted. See PDF]
3.3 Detection limit and Allan varianceThe detection limit for NO radical was estimated by considering 3 times the peak-to-peak noise on the absorption coefficient at 662 nm, which corresponds to the maximum absorption of NO radical. For 10 s of acquisition time (which corresponds to 25 acquisitions of 400 ms) and an unpurged system, it has been found to be 1.2 10 cm. Considering that NO radical cross section at this wavelength is 2.2 10 cm molecule, the detection limit for NO was estimated to be 6 ppt. The same approach was used to estimate the detection limit for NO. Between 645 and 650 nm, which correspond to the two main absorption peaks of NO, noise has been found to be 2 10 cm. Considering the difference of maximum and minimum absorption of 2.2 10 cm molecule in this range, the detection limit has been found to be 11 ppb for 10 s of integration time. A spectrum measured with an acquisition time of 10 s, for 6 ppt of NO and 630 ppb of NO, is shown for illustration in Fig. 4. The fit range used was between 655 and 666.5 nm. Even at this low level of NO concentration, the absorption is clear and allows its quantification. The fit appears to be satisfying for both NO and NO, and the residual spectrum appears to be essentially made of noise, showing a good efficiency of the polynomial fit and a satisfying subtraction of species contributions. This figure shows that this wavelength range is efficient for a precise detection and quantification of both species and validates the detection limit.
Figure 4
(a) Measured absorption coefficient (between 655 and 669 nm) with an integration time of 10 s (in black); complete fit of NO and NO (in blue) with [NO] 6 ppt; [NO] 630 ppb; NO fit only (in green); (b) measured absorption coefficient (in black) without NO contribution and fitted with [NO] 6 ppt (in red); (c) residue of measured and fitted absorption coefficient.
[Figure omitted. See PDF]
Table 1Comparison of characteristics and performances of various IBBCEAS devices coupled to simulation chambers for the detection of NO radicals.
| In situ IBBCEAS | (cm) | DL/integration time | Reference |
|---|---|---|---|
| LISA, Créteil, France | 82 | 6 ppt/10 s | Current work |
| UCC, Cork, Ireland | 462 | 4 ppt/57 s | Venables et al. (2006) |
| UCC, Jülich, Germany | 1800 20 | 0.5–2 ppt/5 s | Dorn et al. (2013) |
| LPCA, Wimereux, France | 2000 | 7.9 ppt/1 min | Wu et al. (2014) |
is the effective length of the cavity, calculated by taking into account the dilution generated by the mirror protective flush. UCC's IBBCEAS device was used in the SAPHIR chamber during an intercomparison campaign.
The potential of the IBBCEAS technique for measuring NO radicals during simulation chamber experiments has already been explored in previous works. It has been coupled to the simulation chamber at UCC (Cork, Ireland), to the SAPHIR chamber at FZJ (Jülich, Germany) and to the CHARME chamber at LPCA (Wimereux, France). The characteristics and performances of the various instruments are compared in Table 1. Our instrument exhibits very good performance, with the detection limit similar to that of the other instruments but for shorter integration time and/or for a smaller effective length. This reflects the very good stability of the optical system. These results also prove the potential of this technique for measuring NO radicals at low level of concentrations with a good time resolution (10 s) and thus its suitability for kinetic measurements.
In order to evaluate if our detection limit can be improved by increasing the integration time, the Allan variance has been calculated for various integration times during an experiment in which NO concentration was monitored. The NO mixing ratio was approximatively 1300 ppb. The Allan variance is given by the equation: 5 with the number of measurements, the integration time and the concentration of NO measured. In this experiment, 30 000 measurements of 2 s were performed, and the Allan variance was then calculated for various integration times ranging between 2 and 4096 s. The standard deviation of Allan, defined as the square root of the Allan variation, provides an indication of the instrument stability in time. It is plotted as a function of the integration time in Fig. 5. For a very short integration time (a few seconds) the Allan deviation is very high due to the white noise of the instrument. The Allan deviation decreases with increasing integration time up to 100 s. For longer integration times, the deviation increases with increasing integration time. Nevertheless, the deviation is low, showing that the instrument is very stable. Due to this stability, we expect that the detection limits calculated before can be improved by increasing the integration time. From this test, it can also be concluded that the optimal integration time is around 100 s.
Figure 5
Allan variance vs. integration time calculated for the IBBCEAS technique.
[Figure omitted. See PDF]
The high stability of the Allan deviation after 100 s also suggests that the stability of the optical device is optimal to perform measurements without recording a new for at least 4096 s, in agreement with the result of the test presented in Sect. 3.2. In conclusion, the good stability of the optical device complies with the constraint of experiments in the simulation chamber for which can only be recorded at the beginning of the experiment.
3.4 Determination of the uncertaintyConsidering Eqs. (1) and (2), the overall relative error on NO concentration can be considered as the square root of the sum of the square relative errors, the reflectivity and the NO absorption cross sections. In the case of the use of the nitrogen flush, the uncertainty on also has to be taken into account (8 %; see Sect. 2). Considering Eq. (3), the uncertainty on should include the uncertainty on NO concentration measured by FTIR estimated to be 8 % as well as the uncertainty on NO absorption cross sections. The uncertainties on NO and NO cross sections are estimated to be 3 % in the spectral range of interest (Vandaele et al., 1997; Orphal et al., 2003). However, the uncertainty generated by the data treatment, fit, i.e., by the fit, which results mainly from the noise in the spectra, should also be taken into account. Because the nitrate radical is an unstable species, this uncertainty cannot be estimated by calculating the standard deviation of its concentrations measured for a long period of monitoring. It was therefore estimated by considering the noise of a NO concentration time profile and has been found to be 3 ppt for 10 s of integration time. The overall absolute error on NO concentration is then expressed by the following formula:
6 where , and are relative uncertainties on NO cross sections, NO concentration used for the reflectivity measurement and NO cross sections respectively, and is the concentration of NO. For 10 s of integration time, the uncertainty is thus 9 % with an absolute part of 3 ppt.
4 Intercomparison studyAfter having defined the optimal operation conditions of the IBBCEAS, the technique has been validated thanks to an intercomparison with another instrument. During a dedicated experiment, NO and NO concentrations were measured using the IBBCEAS technique, while NO and NO were monitored by in situ FTIR. The chamber was first filled with dry synthetic air (RH 1 %) at atmospheric pressure, and spectra were recorded for both FTIR and IBBCEAS. Several hundreds of parts per billion of NO (Air Liquide N20, purity 99 %, HO 3000 ppm) were then introduced into the chamber, and IBBCEAS spectra were recorded in order to determine the mirrors' reflectivity, (see Sect. 3.2). NO radicals were then formed in situ, using thermal dissociation of NO (Reaction R1), which was synthesized in a vacuum line following the reaction between O and NO (Reactions R2 and R3). This protocol was adapted from Atkinson et al. (1984) and Schott and Davidson (1958) and is detailed in Picquet-Varrault et al. (2009). Then, by considering the following equilibrium, and by assuming that this equilibrium is reached, NO concentration can be deduced from NO and NO concentrations measured by FTIR. This hypothesis appears justified as no other reactive species has been introduced into the chamber and may thus disturb the equilibrium. The equilibrium constant () is well known and has been shown to highly depend on temperature and pressure (Atkinson et al., 2004). These two parameters were therefore precisely monitored during the experiment, allowing an equilibrium constant of 2.17 10 cm.molecule at 298 K and at 1030 mbar to be calculated using IUPAC database parameters (Atkinson et al., 2004). IBBCEAS spectra were recorded every 30 s, while FTIR ones were recorded every 2 min. The integrated band intensities used to quantify NO and NO with FTIR were IBI (1530–1680 cm) (5.6 0.2) 10 cm molecule and IBI (1200–1285 cm) (4.05 0.4) 10 cm molecule (base ). The correlation plots between FTIR and IBBCEAS for NO and NO measurements are shown in Fig. 6. NO concentrations measured by the two techniques are in very good agreement (the maximum difference between the two techniques is 6 %), with a slope of 1.0 0.2. Here, the overall uncertainty was calculated as the sum of the statistical error on the slope (twice the standard deviation, 4 %) and systematic errors on FTIR (i.e., on IBI, 4 %) and IBBCEAS measurements (which includes uncertainties on NO cross sections and on the mirrors reflectivity, 9 %). The intercept of the linear regression ( 15.0 ppb) is not significantly different from zero as it is lower than its uncertainty calculated as twice the standard deviation ( 25.1 ppb). For NO radicals, the concentrations obtained by the two techniques are also in good agreement for the whole range of concentrations, from a few parts per trillion to several hundred parts per trillion. The slope of the NO correlation plot is 1.1 0.3, suggesting a bias of 10 % between the two techniques. The error is calculated with the same method as for NO. However, this difference is not significant in regards to the uncertainties which are represented by the dashed black lines. For FTIR measurement, the uncertainties are calculated as the error on NO and NO measurement and on the equilibrium constant (21 %). The calculation for uncertainty on NO IBBCEAS measurement is presented in Sect. 3.4. The intercept appears to be very low (around 3 ppt).
Figure 6
Correlation between FTIR and IBBCEAS measurements for NO (a) and NO (b). Uncertainties are shown by dashed straight lines. Dashed blue lines show the 1 : 1 ratio.
[Figure omitted. See PDF]
In conclusion, the IBBCEAS exhibits very good agreement with the FTIR, for both NO and NO monitoring, with good sensitivity. This agreement is very satisfactory considering the fact that the IBBCEAS samples across the reactor width, while the FTIR provides an integrated measurement on the whole reactor length. Finally, the intercomparison experiment shows that an eventual loss due to photolysis of NO by the beam is not significant because it would have led to lower concentrations measured by the IBBCEAS than by the FTIR. In addition, for wavelengths longer than 640 nm, which is the case here, Johnston et al. (1996) have shown that the photolysis quantum yield is close to 0. It is thus expected that the photolysis of NO does not occur in the wavelength range used. Furthermore, even though NO was slightly subject to photolysis locally, the volume enlightened by the probe beam is very small in comparison to the overall volume of the chamber ( 0.04 %), and the homogenization system allowed a mixing with the rest of the volume.
5Kinetic study: NO trans-2-butene
The last step of the validation consisted in a kinetic experiment in order to assess the potential of the technique for kinetic studies: the IBBCEAS has been used to measure the rate constant of a well-known reaction: trans-2-butene NO. This reaction has been chosen as it has been intensively studied in the literature. Six absolute rate determinations (Benter et al., 1992; Berndt et al., 1998; Dlugokencky and Howard, 1989; Kasyutich et al., 2002; Ravishankara and Mauldin, 1985; Rudich et al., 1996) and two relative ones (Atkinson et al., 1984; Japar and Niki, 1975) have been published, leading to a recommendation by IUPAC (Atkinson et al., 2006). This reaction is also particularly interesting because it does not produce secondary organic aerosol (SOA). This will allow us to test the performances of the instrument for monitoring NO concentrations with a high time resolution and to validate our kinetic determination by comparison with previous ones.
The rate constant was determined using the absolute rate technique and by measuring the consumption of trans-2-butene due to its reaction with NO. Because no other oxidant was present in the mixture, it was therefore assumed that trans-2-butene is consumed only by reaction with nitrate radical: R6 For this reaction, the kinetic equation can be established as 7 By making the hypothesis of small variations of time and [trans-2-butene], this relationship can be approximated to 8 where corresponds to the consumption of trans-2-butene during the time interval , and [trans-2-butene] and [NO] are averaged concentrations measured during this period. By plotting [trans-2-butene] vs. the product [trans-2-butene][NO], a straight line with the slope corresponding to is obtained. It is important to note that because this absolute kinetic method consists of measuring the decay of trans-2-butene for a known concentration of NO, and not the decay of NO radicals for a known concentration of the VOC, it is not affected by NO additional loss processes (e.g., wall losses, reactions with NO or with peroxy radicals). Only additional losses of the VOC would lead to an overestimation of the rate constant. So, it was checked prior to the experiments (i.e., before the injection of NO) that no significant loss of the VOC was observed in the timescale of the experiment (see below).
Six kinetic experiments have been conducted in the dark, at room temperature (292–294 K) and atmospheric pressure in synthetic air. The initial conditions of reactants (trans-2-butene, NO, NO) are listed in Table 2. The reflectivity was measured prior to each experiment by introducing NO into the chamber. When present, NO initial concentrations were used also to slow down the reaction by shifting the NO equilibrium. Trans-2-butene (Air Liquide, purity 99 %) was then introduced, and it was checked that no significant loss was observed in the timescale of the experiment. Nitrate radicals were generated into the simulation chamber from the thermal decomposition of dinitrogen pentoxide. NO was injected stepwise in order to assure a consumption of trans-2-butene on a proper timescale to satisfactorily monitor the reactants. Time-resolved concentrations of trans-2-butene, NO and NO were monitored from their infrared absorption spectra every 2 min. The integrated band intensity used to quantify the VOC is IBI (870–1100 cm) (2.8 0.3) 10 cm molecule (base , measured by previous internal work). NO was monitored with the IBBCEAS technique with an acquisition time of 30 s.
Table 2Injected mixing ratios for the kinetic study of the reaction trans-2-butene NO.
| Experiment | [NO] | [NO] (ppb) | [trans-2-butene] |
|---|---|---|---|
| (ppb) | number of | (ppb) | |
| injections | |||
| 1 | / | 2500 | 1920 |
| 2 | / | 300 2; 150 2 | 750 |
| 3 | 920 | 1000 3 | 990 |
| 4 | 950 | 1500 2 | 1110 |
| 5 | 750 | 300 3 | 1110 |
| 6 | / | 300 3 | 1030 |
Figure 7 shows time profiles of reactants during a typical experiment. At the moment when NO is injected, a rapid decrease of trans-2-butene and NO concentrations is observed, together with large production of NO due to NO decomposition.
Figure 7
Concentrations of trans-2-butene and NO measured by FTIR (left axis) and NO measured by IBBCEAS vs. time (right axis) during Experiment 1.
[Figure omitted. See PDF]
Figure 8
Absolute kinetic plot for the reaction of trans-2-butene with NO radicals, showing the loss of trans-2-butene vs. [VOC] [NO] . The inset on the right shows a close-up of the points close to the origin.
[Figure omitted. See PDF]
Table 3Comparison of the rate constant obtained for the reaction of trans-2-butene with NO with previous determinations.
| Technique | Reference | ||
|---|---|---|---|
| (cm molecule s) | (K) | ||
| (4.13 0.78) 10 | 293 | (NO/CEAS) | This study |
| (3.90 0.78) 10 | 298 | recommendation | IUPAC |
| ( 0.08) | |||
| (3.78 0.17) 10 | 298 | (AR/CEAS) | Kasyutich et al. (2002) |
| (3.74 0.45) 10 | 298 | (AR/LIF) | Berndt et al. (1998) |
| (4.06 0.36) 10 | 298 | (AR/LIF) | Rudich et al. (1996) |
| (3.88 0.30) 10 | 298 | (AR/MS) | Benter et al. (1992) |
| (3.96 0.48) 10 | 298 | (AR/LIF) | Dlugokencky and Howard (1989) |
| (3.78 0.17) 10 | 298 | (AR/LIF) | Ravishankara and Mauldin (1985) |
| (3.09 0.27) 10 | 298 | (RR) | Atkinson et al. (1984) |
| (2.31 0.17) 10 | 300 | (RR) | Japar and Niki (1975) |
Indicates kinetic method (AR absolute rate, RR relative rate) and NO measurement technique used: CEAS, LIF (laser-induced fluorescence) or MS (mass spectrometry). Relative rate determinations are relative to the equilibrium constant ().
The kinetic plot ([trans-2-butene] vs. the product [trans-2-butene] [NO] ) gathering data from all experiments is presented in Fig. 8. The uncertainty on each experimental point was calculated as the sum of the relative uncertainties on [trans-2-butene] and [NO] for the abscissa scale (the uncertainty on the time was considered to be negligible) and as twice the uncertainty on the [trans-2-butene] for the ordinate scale. From this figure, it can be observed that all experiments are in good agreement. In consequence, a linear regression was performed on all data points, leading to a rate constant of (4.13 0.78) 10 cm molecule s. The uncertainty on the rate constant was estimated as the sum of the relative uncertainties on NO concentrations and twice the standard deviation on the linear regression. The obtained rate constant has been compared to the values from previous determinations and to the value recommended by IUPAC in Table 3.
This new determination appears to be in very good agreement with the IUPAC recommendation and with previous absolute determinations, within the uncertainty. However this work does not agree with the two relative determinations which are up to 45 % lower than our determination. Nevertheless, these two values are in disagreement with all of the previous absolute determinations. These relative rate determinations are relative to the equilibrium constant (). A possible explanation for this disagreement would be that NO concentration is overestimated because it was considered that the equilibrium is reached. However, the reaction with trans-2-butene is fast enough to significantly disturb the equilibrium and prevent it from being established. An overestimation of NO concentrations would hence lead to an underestimation of the rate constant.
In conclusion, this agreement shows that the determination made with the IBBCEAS technique presented in this paper is correct, allowing reliable measurement of NO at low concentration with good sensitivity and time resolution. This technique is now operational for application to other absolute kinetic studies.
6 ConclusionsAn IBBCEAS technique has been developed and coupled to the CSA simulation chamber for the in situ measurement of NO radicals at the parts per trillion level. This instrument also allows the monitoring of NO in the parts per billion range. Thanks to various tests, the instrument has been carefully characterized in order to identify potential bias and to define the optimal operation conditions. The performance of the instrument in terms of detection limit and uncertainties was also determined. The instrument exhibits very good detection limit for NO radicals (6 ppt) for 10 s of integration time. This detection limit fully complies with our needs for kinetic applications.
The instrument was also validated thanks to an intercomparison experiment with the in situ FTIR technique. With this technique, NO concentration was indirectly obtained by monitoring NO and NO concentrations and by using the well-known equilibrium constant (). The concentrations measured by the two techniques were shown to be in very good agreement (better than 10 %) for both NO and NO, over a wide range of concentrations: from parts per trillion to parts per billion for NO radicals and from parts per billion to hundreds of parts per billion for NO.
Finally, this technique was used for the absolute rate determination of a well-documented reaction, trans-2-butene NO. The value of (4.13 0.45) 10 cm molecule s found in this study is in very good agreement with the previous absolute determinations. Moreover, the good sensitivity and the good time resolution represent excellent performances, allowing the use of this technique for monitoring NO radicals when involved in fast reactions.
The IBBCEAS technique is now operational and will be used in further works, particularly to monitor NO concentrations for the absolute rate determinations of fast reactions of volatile organic compounds with NO radicals.
Data availability
The rate constant for the NO oxidation of trans-2-butene is available in Table 3.
It is also available through the Library of Advanced Data Products (LADP) of
the EUROCHAMP Data Center (
Author contributions
MCi, BPV and AF designed the IBBCEAS technique with the help of GM and DR. AF installed and did characterization experiments with the technical support of XL (optical development), MCa and EP (experiments) and PZ and GS (data treatment). MCi, BPV and AF wrote the article, and AF was responsible for the final version. All coauthors revised the content of the original manuscript and approved the final version of the paper.
Competing interests
The authors declare that they have no conflict of interest.
Special issue statement
This article is part of the special issue “Simulation chambers as tools in atmospheric research (AMT/ACP/GMD inter-journal SI)”. It is not associated with a conference.
Financial support
This research has been supported by the French national programme LEFE/INSU and by the European Commission's Seventh Framework Programme (EUROCHAMP-2; grant no. 228335) and H2020 Research Infrastructures (EUROCHAMP-2020; grant no. 730997).
Review statement
This paper was edited by Hartmut Herrmann and reviewed by two anonymous referees.
You have requested "on-the-fly" machine translation of selected content from our databases. This functionality is provided solely for your convenience and is in no way intended to replace human translation. Show full disclaimer
Neither ProQuest nor its licensors make any representations or warranties with respect to the translations. The translations are automatically generated "AS IS" and "AS AVAILABLE" and are not retained in our systems. PROQUEST AND ITS LICENSORS SPECIFICALLY DISCLAIM ANY AND ALL EXPRESS OR IMPLIED WARRANTIES, INCLUDING WITHOUT LIMITATION, ANY WARRANTIES FOR AVAILABILITY, ACCURACY, TIMELINESS, COMPLETENESS, NON-INFRINGMENT, MERCHANTABILITY OR FITNESS FOR A PARTICULAR PURPOSE. Your use of the translations is subject to all use restrictions contained in your Electronic Products License Agreement and by using the translation functionality you agree to forgo any and all claims against ProQuest or its licensors for your use of the translation functionality and any output derived there from. Hide full disclaimer
© 2020. This work is published under https://creativecommons.org/licenses/by/4.0/ (the “License”). Notwithstanding the ProQuest Terms and Conditions, you may use this content in accordance with the terms of the License.
Abstract
An incoherent broadband cavity-enhanced absorption spectroscopy (IBBCEAS) technique has been developed for the in situ monitoring of NO
You have requested "on-the-fly" machine translation of selected content from our databases. This functionality is provided solely for your convenience and is in no way intended to replace human translation. Show full disclaimer
Neither ProQuest nor its licensors make any representations or warranties with respect to the translations. The translations are automatically generated "AS IS" and "AS AVAILABLE" and are not retained in our systems. PROQUEST AND ITS LICENSORS SPECIFICALLY DISCLAIM ANY AND ALL EXPRESS OR IMPLIED WARRANTIES, INCLUDING WITHOUT LIMITATION, ANY WARRANTIES FOR AVAILABILITY, ACCURACY, TIMELINESS, COMPLETENESS, NON-INFRINGMENT, MERCHANTABILITY OR FITNESS FOR A PARTICULAR PURPOSE. Your use of the translations is subject to all use restrictions contained in your Electronic Products License Agreement and by using the translation functionality you agree to forgo any and all claims against ProQuest or its licensors for your use of the translation functionality and any output derived there from. Hide full disclaimer
Details
; Romanini, Daniele 2 ; Picquet-Varrault, Bénédicte 1
1 LISA, UMR CNRS 7583, Université Paris-Est Créteil, Université de Paris, Institut Pierre Simon Laplace (IPSL), Créteil, France
2 LIPHY, UMR CNRS 5588, Université Grenoble Alpes, Grenoble, France





