It appears you don't have support to open PDFs in this web browser. To view this file, Open with your PDF reader
Abstract
Background
The aim of this retrospective study was to compare the efficacy and safety profile of a single XEN-microstent in different types of primary and secondary open angle glaucoma.
Methods
A single XEN microstent was implanted in patients with primary open-angle glaucoma (POAG), normal-tension glaucoma (NTG), pseudoexfoliation glaucoma (PEX) and secondary glaucoma (Sec.Gl). The intraocular pressure (IOP), the active substances of the applied IOP-lowering drugs, the best corrected visual acuity (BCVA) and the mean deviation (MD) of the perimetry were measured at baseline and at regular follow-ups, scheduled at 2 days and 1, 3, 6 and 12 months after surgery.
Results
153 eyes were included in this analysis. 113 eyes were affected by POAG (74%), 5 eyes by NTG (3%), 22 eyes by PEX (14%) and 13 eyes by Sec. Gl (9%). Mean IOP decreased in all treatment groups during the 12 months of follow-up (complete group: 23.9 ± 7.4 to 15.4 ± 5.1 mmHg (p < 0.01); POAG: 22.8 ± 6.5 to 15.1 ± 4.6 mmHg (p < 0.01); NTG: 16.6 ± 3.4 to 11.6 ± 2.2 mmHg (p < 0.05); PEX: 28.0 ± 7.9 to 17.1 ± 6.6 mmHg (p < 0.01); Sec.Gl: 28.9 ± 13.9 to 15.5 ± 6.9 mmHg (p < 0.05)). In the 153 eyes the average number of IOP-lowering drugs applied decreased from 2.6 ± 1.2 to 0.8 ± 1.3 12 months after surgery (p < 0.01). BCVA and mean deviation of automated standard perimetry remained stable in all groups during follow-up.
Conclusion
As in eyes suffering from POAG, IOP and number of IOP-lowering drugs applied can be effectively reduced by XEN implantation in eyes suffering from NTG, PEX and secondary glaucoma while leaving BCVA and visual field unchanged.
Trial registration
Trial was registered at DRKS (registration number: DRKS00020800, Registered 25.February 2020 - Retrospectively registered).
You have requested "on-the-fly" machine translation of selected content from our databases. This functionality is provided solely for your convenience and is in no way intended to replace human translation. Show full disclaimer
Neither ProQuest nor its licensors make any representations or warranties with respect to the translations. The translations are automatically generated "AS IS" and "AS AVAILABLE" and are not retained in our systems. PROQUEST AND ITS LICENSORS SPECIFICALLY DISCLAIM ANY AND ALL EXPRESS OR IMPLIED WARRANTIES, INCLUDING WITHOUT LIMITATION, ANY WARRANTIES FOR AVAILABILITY, ACCURACY, TIMELINESS, COMPLETENESS, NON-INFRINGMENT, MERCHANTABILITY OR FITNESS FOR A PARTICULAR PURPOSE. Your use of the translations is subject to all use restrictions contained in your Electronic Products License Agreement and by using the translation functionality you agree to forgo any and all claims against ProQuest or its licensors for your use of the translation functionality and any output derived there from. Hide full disclaimer