Abstract
Background
To compare visual performance between the iris-fixated phakic intraocular len (pIOL) and implantable collamer len (ICL) to correct high myopia.
Methods
Twenty-four eyes underwent iris-fixated pIOL implantation and 24 eyes underwent ICL implantation. At the 6-month follow-up, the best-corrected visual acuity (BCVA) and uncorrected distance visual acuity (UDVA) were compared between the iris-fixated pIOL and ICL groups. The objective scatter index (OSI), modulation transfer function (MTF) cutoff, and ocular aberrations were performed to evaluate postoperative visual quality between the two groups.
Results
No significant difference was found in UDVA, BCVA, and spherical equivalent between the iris-fixated pIOL and ICL groups (P > 0.05). Six months after surgery, the following values were significantly higher in the ICL group than in the iris-fixated pIOL group: MTF cutoff, strehl ratio and optical quality analysis system values at contrasts of 9 %, 20 %, and 100 % (P < 0.01). The OSI in the iris-fixated pIOL group was higher than in the ICL group 6 months after surgery (P < 0.01). All high-order aberrations were slightly more severe in the iris-fixated pIOL group than in the ICL group 6 months after surgery, although only trefoil (P = 0.023) differed significantly in this regard.
Conclusions
Both iris-fixated lenses and ICLs can provide good visual acuity. ICLs confer better visual performance in MTF-associated parameters and induce less intraocular light scattering than iris-fixated pIOLs.
You have requested "on-the-fly" machine translation of selected content from our databases. This functionality is provided solely for your convenience and is in no way intended to replace human translation. Show full disclaimer
Neither ProQuest nor its licensors make any representations or warranties with respect to the translations. The translations are automatically generated "AS IS" and "AS AVAILABLE" and are not retained in our systems. PROQUEST AND ITS LICENSORS SPECIFICALLY DISCLAIM ANY AND ALL EXPRESS OR IMPLIED WARRANTIES, INCLUDING WITHOUT LIMITATION, ANY WARRANTIES FOR AVAILABILITY, ACCURACY, TIMELINESS, COMPLETENESS, NON-INFRINGMENT, MERCHANTABILITY OR FITNESS FOR A PARTICULAR PURPOSE. Your use of the translations is subject to all use restrictions contained in your Electronic Products License Agreement and by using the translation functionality you agree to forgo any and all claims against ProQuest or its licensors for your use of the translation functionality and any output derived there from. Hide full disclaimer