© 2021 Buckenham Boyle et al. This is an open access article distributed under the terms of the Creative Commons Attribution License: http://creativecommons.org/licenses/by/4.0/ (the “License”), which permits unrestricted use, distribution, and reproduction in any medium, provided the original author and source are credited. Notwithstanding the ProQuest Terms and Conditions, you may use this content in accordance with the terms of the License.

Abstract

Purpose

To assess the agreement and repeatability of horizontal visible iris diameter (HVID) or white-to-white (WTW) measurements between four imaging modalities; combination slit scanning elevation/Placido tomography, infrared biometry, dual rotating scheimpflug camera/Placido tomography, and swept source anterior segment optical coherence tomography (AS-OCT).

Methods

A prospective study of 35 right eyes of healthy volunteers were evaluated using the Orbscan IIz, IOL Master 700, Galilei G2, and DRI Triton OCT devices. The inter-device agreement and repeatability of HVID/WTW measurements for each device were analysed.

Results

Mean HVID/WTW values obtained by the Orbscan IIz, IOL Master 700, Galilei G2 and DRI Triton OCT were 11.77 ± 0.40 mm, 12.40 ± 0.43 mm, 12.25 ± 0.42 mm, and 12.42 ± 0.47 mm, respectively. All pairwise comparisons revealed statistically significant differences in mean HVID/WTW measurements (p = <0.01) except for the IOL Master 700—DRI OCT Triton pair (p = 0.56). Mean differences showed that the DRI Triton OCT produced the highest HVID/WTW values, followed by the IOL Master 700, Galilei G2 and Orbscan IIz, respectively. The limits of agreement were large on all device pairs. There was high repeatability for all devices (ICC ≥ 0.980). The highest repeatability was seen in the Galilei G2 (ICC = 0.995) and lowest in the Orbscan IIz (ICC = 0.980).

Conclusions

The four devices exhibit high repeatability, but should not be used interchangeably for HVID/WTW measurements in clinical practice.

Details

Title
Repeatability and agreement of white-to-white measurements between slit-scanning tomography, infrared biometry, dual rotating Scheimpflug camera/Placido disc tomography, and swept source anterior segment optical coherence tomography
Author
Alexander Buckenham Boyle; Namkung, Soobin; Shew, William; Akilesh Gokul; McGhee, Charles N J; Ziaei, Mohammed
First page
e0254832
Section
Research Article
Publication year
2021
Publication date
Jul 2021
Publisher
Public Library of Science
e-ISSN
19326203
Source type
Scholarly Journal
Language of publication
English
ProQuest document ID
2552276890
Copyright
© 2021 Buckenham Boyle et al. This is an open access article distributed under the terms of the Creative Commons Attribution License: http://creativecommons.org/licenses/by/4.0/ (the “License”), which permits unrestricted use, distribution, and reproduction in any medium, provided the original author and source are credited. Notwithstanding the ProQuest Terms and Conditions, you may use this content in accordance with the terms of the License.