Full Text

Turn on search term navigation

© 2020 Author(s) (or their employer(s)) 2020. Re-use permitted under CC BY-NC. No commercial re-use. See rights and permissions. Published by BMJ. http://creativecommons.org/licenses/by-nc/4.0/ This is an open access article distributed in accordance with the Creative Commons Attribution Non Commercial (CC BY-NC 4.0) license, which permits others to distribute, remix, adapt, build upon this work non-commercially, and license their derivative works on different terms, provided the original work is properly cited, appropriate credit is given, any changes made indicated, and the use is non-commercial. See:  http://creativecommons.org/licenses/by-nc/4.0/ . Notwithstanding the ProQuest Terms and Conditions, you may use this content in accordance with the terms of the License.

Abstract

Objectives

Versius is a teleoperated surgical robotic system intended for use in minimal access surgery. This study aimed to validate the safety and effectiveness of the Versius user interface in the hands of trained users and identify and address the causes of any user errors.

Design

Surgical teams completed a commercially representative training program over 3.5 days. After training was completed, the usability-related aspects of the system were assessed.

Setting

A simulated operating room using a cadaveric model.

Participants

Surgical teams consisting of a lead surgeon, assistant surgeon, scrub nurse and circulating nurse.

Main outcome measures

Usability-related aspects of the system were assessed through the completion of critical and non-critical tasks. A critical task was defined according to the Food and Drug Administration’s definition, as a user task which, if performed incorrectly or not performed at all, would or could cause serious harm to the patient or user, where harm is defined to include compromised medical care.

Results

In total, 17 surgical teams participated in the study and all were experienced in laparoscopic surgery. The number of robotic surgeries performed by the participants per month ranged from 0 to 100. Surgical specialties were similarly represented from obstetrics and gynecology, colorectal, urology and upper gastrointestinal. No critical task failures were observed. Of all the tasks completed, 98% were recorded as a pass or a pass with difficulty.

Conclusions

These results demonstrate that in a simulated clinical setting, Versius can be safely used by both laparoscopically and robotically trained healthcare professionals. These results support the progression to assessment of Versius in preclinical studies.

Details

Title
Usability assessment of Versius, a new robot-assisted surgical device for use in minimal access surgery
Author
Haig, Fiona 1 ; Barbosa Medeiros, Ana Cristina 1 ; Chitty, Karen 1 ; Slack, Mark 1   VIAFID ORCID Logo 

 CMR Surgical, Cambridge, Cambridgeshire, UK 
First page
e000028
Section
Original research
Publication year
2020
Publication date
May 2020
Publisher
BMJ Publishing Group LTD
e-ISSN
26314940
Source type
Scholarly Journal
Language of publication
English
ProQuest document ID
2552744942
Copyright
© 2020 Author(s) (or their employer(s)) 2020. Re-use permitted under CC BY-NC. No commercial re-use. See rights and permissions. Published by BMJ. http://creativecommons.org/licenses/by-nc/4.0/ This is an open access article distributed in accordance with the Creative Commons Attribution Non Commercial (CC BY-NC 4.0) license, which permits others to distribute, remix, adapt, build upon this work non-commercially, and license their derivative works on different terms, provided the original work is properly cited, appropriate credit is given, any changes made indicated, and the use is non-commercial. See:  http://creativecommons.org/licenses/by-nc/4.0/ . Notwithstanding the ProQuest Terms and Conditions, you may use this content in accordance with the terms of the License.