Full Text

Turn on search term navigation

© 2021 by the authors. Licensee MDPI, Basel, Switzerland. This article is an open access article distributed under the terms and conditions of the Creative Commons Attribution (CC BY) license (https://creativecommons.org/licenses/by/4.0/). Notwithstanding the ProQuest Terms and Conditions, you may use this content in accordance with the terms of the License.

Abstract

The goal of this study is to evaluate the primary stability of a cementless augment-and-modular-cage system with and without the addition of cranial straps in a standardized in vitro setting. As the surrogate parameter for the evaluation of primary stability, the measurement of relative motion between the implant components themselves and the bone will be used. Acetabular revision components with a trabecular titanium augment in combination with a large fourth-generation composite left hemipelvis were assembled. These constructs were divided into two groups with (S) and without cranial straps (nS). A total of 1000 cycles was applied at each of three load levels. Relative movements (RM) between the components were measured. Load levels display a significant effect on the amount of RM at all interfaces except between shell/augment. The group assignment appears to have an effect on RM due to significantly differing means at all interfaces. Between bone/shell RM increased as load increased. NS displayed significantly more RM than S. Between shell/augment RM remained constant as load increased. Between shell/cup S showed more RM than nS while both groups’ RM increased with load. We conclude a significant increase of primary stability between the shell and the bone through the addition of cranial straps. Relative motion between components (shell/cup) increases through the addition of cranial straps. A clinical impact of this finding is uncertain and requires further investigation. Finally, the cementless fixation of the augment against the rim-portion of the shell appears stable and compares favorably to prior investigation of different fixation techniques.

Details

Title
Primary Stability in Hip Revision Arthroplasty: Comparison of the Stability of Cementless Fixed Augments on a Modular Acetabular Cage System with and without Cranial Straps
Author
Jaenisch, Max 1   VIAFID ORCID Logo  ; Kohlhof, Hendrik 1 ; Wirtz, Dieter Christian 1 ; Schildberg, Frank Alexander 1   VIAFID ORCID Logo  ; Beckmann, Nicholas A 2 ; Kretzer, Jan Philippe 2   VIAFID ORCID Logo  ; Schonhoff, Mareike 2 ; Jäger, Sebastian 2 

 Department for Orthopaedics and Trauma Surgery, University Hospital Bonn, 53127 Bonn, Germany; [email protected] (H.K.); [email protected] (D.C.W.); [email protected] (F.A.S.) 
 Laboratory of Biomechanics and Implant Research, Clinic for Orthopaedics and Trauam Surgery, Heidelberg University Hospital, 69118 Heidelberg, Germany; [email protected] (N.A.B.); [email protected] (J.P.K.); [email protected] (M.S.); [email protected] (S.J.) 
First page
4002
Publication year
2021
Publication date
2021
Publisher
MDPI AG
e-ISSN
20770383
Source type
Scholarly Journal
Language of publication
English
ProQuest document ID
2571317917
Copyright
© 2021 by the authors. Licensee MDPI, Basel, Switzerland. This article is an open access article distributed under the terms and conditions of the Creative Commons Attribution (CC BY) license (https://creativecommons.org/licenses/by/4.0/). Notwithstanding the ProQuest Terms and Conditions, you may use this content in accordance with the terms of the License.