Abstract

This thesis focuses on the research of multi-spectrum thermometry image distortion and non-uniformity correction based on the characteristic of single CCD transient imaging multi-spectrum thermometry. A distortion correction based on Zhang calibration is used to rectify the lens distortion. A non-uniformity correction combined two-point calibration with neural network is carried out to rectify the non-uniform images. Compare the images before and after correction, the distortion calibration method used in this thesis can decrease the error between the field image and the target location. The comparison of the images before and after non-uniformity correction illustrates that the non-uniformity correction used in this thesis can reduce the temperature measurement error caused by CCD pixel conformance differences observably.

Details

Title
Image Distortion and Non-Uniformity Correction for Transient Imaging Multi-spectrum Thermometry
Author
Cai, Jing 1 ; Zhang, Xuecong 1 ; Yang, Yongjun 1 ; Su, Meng 1 

 Changcheng Institute of Metrology & Measurement, Beijing, China 
Publication year
2018
Publication date
Aug 2018
Publisher
IOP Publishing
ISSN
17426588
e-ISSN
17426596
Source type
Scholarly Journal
Language of publication
English
ProQuest document ID
2572720542
Copyright
© 2018. This work is published under http://creativecommons.org/licenses/by/3.0/ (the “License”). Notwithstanding the ProQuest Terms and Conditions, you may use this content in accordance with the terms of the License.