This is an open access article distributed under the Creative Commons Attribution License, which permits unrestricted use, distribution, and reproduction in any medium, provided the original work is properly cited.
1. Introduction
Type 1 diabetes mellitus (DM1) increases the risk of obstetric and neonatal complications [1]. During gestation, hormonal changes cause a diabetogenic physiological state which can lead to volatile glycemic control in diabetic patients [2]. Previous studies have shown that DM1 women are up to five times more likely to experience severe hypoglycemia during early pregnancy compared to the preconception period [3]; the management of these patients can be challenging, and there is a need for new therapies. Advances in biotechnology seek to achieve glycemic control and minimize the burden of hypoglycemia.
The Medtronic MiniMed 670G is a closed-loop hybrid insulin infusion system approved for patients with diabetes [4, 5]; however, safety in pregnancy has not been established through clinical trials. One of the challenges with this pump is that its algorithm targets fasting glucose of 120 mg/dL, higher than the standard target of <95 mg/dL during pregnancy [6–10]. Herein, we present the case of a female patient with DM1 who achieved satisfactory glycemic control using the MiniMed 670G pump.
2. Methods
A 33-year-old woman with a history of DM1 since 16 years, user of an insulin pump (Paradigm® Veo system) due to high glycemic variability and recurrent hypoglycemia, presented to the outpatient clinic seeking preconception counseling. She reported no hypoglycemic episodes since 2015 and no macrovascular or microvascular complications.
She subsequently reported her pregnancy status at week 10. At the beginning of pregnancy, her weight was 57 Kg, body mass index (BMI) 23 Kg/m2, and glycated hemoglobin (HbA1c) 6.7%. At week 24, laboratories showed HbA1c of 6.3%; however, she developed high glycemic variability and inadvertent hypoglycemias. Enlite Sensor® continuous glucose monitoring (CGM) dysfunction was discovered, and a switch to a MiniMed 670G pump technology was decided.
During the following four weeks, the Paradigm® Veo pump was kept without the use of a sensor, precipitating hypoglycemia with self-reported anxiety and fear of a drop in glucose levels. At week 28, maternal weight was 63 Kg, and the estimated fetal weight was 1230 g for a 67th percentile (P67). At week 30, the training process on MiniMed 670 G began, a protocol that started with two weeks in the manual mode and was then switched to the automatic mode as an off-label indication. This optimized the times in range and reduced the hypoglycemic episodes and hence the related anxiety in the patient.
By week 33, the estimated fetal weight was 2291 g (P88) and at week 35 was 2832 g (P95) with a maternal weight of 65 Kg. The glycemic control ranges remained within goal until the end of the pregnancy, which took place at week 38 via cesarean section, during which the automatic mode and dextrose infusion were maintained (Table 1). The glycemic variability was higher in weeks 30–34 with higher time in range in hyperglycemia and hypoglycemia.
Table 1
Glycemic control monitoring.
Follow-up week | Pump | Mode | Time in range (TIR) | TDD (UI) | Basal/bolus ratio (%) | |||
<54 mg/dl (%) | 54–63 mg/dl (%) | 63–140 mg/dl (%) | >140 mg/dl | |||||
30 | VEO | N/A | 1 | 4 | 87 | 8 | 40 | 38/62 |
32 | 670G | Manual | 1 | 2 | 80 | 17 | 45 | 33/67 |
34 | 670G | Automatic | 0 | 1 | 84 | 15 | 46 | 28/72 |
38 | 670G | Automatic | 0 | 2 | 96 | 2 | 34 | 24/76 |
1 | 670G | Manual | 1 | 2 | 94 | 3 | 29 | 21/79 |
4 | 670G | Manual | 0 | 1 | 92 | 7 | 22 | 45/55 |
6 | 670G | Automatic | 0 | 1 | 87 | 12 | 23 | 35/66 |
There were no obstetric complications, and a healthy female newborn weighting 3635 g, suitable for gestational age (P80), was born. In the immediate postpartum period, the automatic mode was suspended due to recurrent hypoglycemic episodes and the periodic suspension of the system by the patient. The manual mode was continued for four weeks, requiring weekly adjustments due to hypoglycemia and a significant decrease in the total required daily insulin dose. At the end of the fourth postpartum week, the automatic mode was restarted, and good glycemic control was maintained.
3. Discussion
DM1 increases the risk of obstetric and neonatal complications associated with maternal hyperglycemia in pregnancy such as macrosomia, congenital anomalies, shoulder dystocia, stillbirth, and neonatal death [1, 2, 11–13]. The glucose goals of pregnant patients are stricter: fasting and preprandial target glucose is <90 mg/dL. Therefore, various pharmacological management options have been developed with less variable insulin regimens that promote better glycemic control in pregnant women [14]. The insulin infusion pump (IIP) is a treatment strategy used when insulin regimens have not been able to achieve adequate control. This syste m has proven to be effective in reducing HbA1c in the nonpregnant population [15]. Along with the development of IIP systems, continuous glucose monitoring has emerged, and in combination, they improve clinical outcomes related to glycemic control. In pregnancy, the use of an insulin pump over multiple-dose injection therapy has not shown greater benefit [16–19]. However, the CONCEPTT trial demonstrated that the use of CGM helps improve neonatal outcomes and reduce maternal hyperglycemia [20].
Research has focused on the development of an artificial pancreas, which, through prediction algorithms, can increase or decrease the supply of insulin according to glucose levels obtained with CGM [21–24]. Stewart et al. compared a closed-loop system from a DANA Diabecare R insulin pump (SOOIL) and a CGM supported by FreeStyle Navigator II vs. an open-loop system in DM1 patients. Their findings reported that women under the closed-loop system achieved better glycemic control, evidenced by fewer obstetric and neonatal complications [25].
In 2016, the FDA approved the first hybrid system, the MiniMed 670G. This system differs from others since it has an automatic glycemic control algorithm that targets a nonmodifiable glucose goal of 120 mg/dL which is higher than the standard target for pregnant patients. The TIR International Consensus established that the target range for pregnant patients is 63–140 mg/dL and to stay below this range less than 5% of the time and above of it less than 25% of the time [26].
Given the short time, this system has been on the market, and there are few reports of its use during gestation. In the case of our patient, migrating from the VEO system to 670 G reduced hypoglycemia and increased time in the target glucose range at rates higher than what is reported in literature [27].
Additionally, the control of stress, anxiety, fear of hypoglycemia, and the psychological burden of the disease play an important role in the management of the diabetic patient. The use of CGM in the CONCEPTT trial showed a reduction in anxiety related to fear of hypoglycemia [20], improving the quality of life of patients using IIP systems coupled with CGM [28].
4. Conclusion
We report an adequate glycemic control in a pregnant patient with DM1 who was transitioned to a closed-loop system at 30 weeks of gestation. The patient had no macrovascular or microvascular complications. Patient showed decreased episodes of severe hypoglycemia and hyperglycemia with better glycemic control than previously reported cases. Our findings are encouraging and open the door for further clinical studies on the use of closed-loop systems in pregnancy.
Consent
The patient has provided consent to the publication of this case.
[1] J. P. Vandorsten, W. C. Dodson, M. A. Espeland, W. A. Grobman, J. M. Guise, B. M. Mercer, H. L Minkoff, B Poindexter, L. A Prosser, G. F Sawaya, J. R Scott, R. M Silver, L Smith, A Thomas, A. T Tita, "NIH consensus development conference: diagnosing gestational diabetes mellitus," NIH Cconsensus and Sate-of-the-Science Statements, vol. 29 no. 1, 2013.
[2] T. A. Buchanan, A. H. Xiang, "Gestational diabetes mellitus," Journal of Clinical Investigation, vol. 115 no. 3, pp. 485-491, DOI: 10.1172/jci200524531, 2005.
[3] L. R. Nielsen, U. Pedersen-Bjergaard, B. Thorsteinsson, M. Johansen, P. Damm, E. R. Mathiesen, "Hypoglycemia in pregnant women with type 1 diabetes: predictors and role of metabolic control," Diabetes Care, vol. 31 no. 1,DOI: 10.2337/dc07-1066, 2008.
[4] S. K. Garg, S. A. Weinzimer, W. V. Tamborlane, B. A. Buckingham, B. W. Bode, T. S. Bailey, R. L. Brazg, J. Ilany, R. H. Slover, S. M. Anderson, R. M. Bergenstal, B. Grosman, A. Roy, T. L. Cordero, J. Shin, S. W. Lee, F. R. Kaufman, "Glucose outcomes with the in-home use of a hybrid closed-loop insulin delivery system in adolescents and adults with type 1 diabetes," Diabetes Technology & Therapeutics, vol. 19 no. 3, pp. 155-163, DOI: 10.1089/dia.2016.0421, 2017.
[5] M. de Bock, J. Dart, A. Roy, R. Davey, W. Soon, C. Berthold, A. Retterath, B. Grosman, N. Kurtz, E. Davis, T. Jones, "Exploration of the performance of a hybrid closed loop insulin delivery algorithm that includes insulin delivery limits designed to protect against hypoglycemia," Journal of Diabetes Science and Technology, vol. 11 no. 1, pp. 68-73, DOI: 10.1177/1932296816668876, 2016.
[6] J. L. Ruiz, J. L. Sherr, E. Cengiz, L. Carria, A. Roy, G. Voskanyan, W. V. Tamborlane, S. A. Weinzimer, "Effect of insulin feedback on closed-loop glucose control: a crossover study," Journal of Diabetes Science and Technology, vol. 6 no. 5, pp. 1123-1130, DOI: 10.1177/193229681200600517, 2012.
[7] G. M. Steil, "Algorithms for a closed-loop artificial pancreas: the case for proportional-integral-derivative control," Journal of Diabetes Science and Technology, vol. 7 no. 6, pp. 1621-1631, DOI: 10.1177/193229681300700623, 2013.
[8] S. A. Weinzimer, G. M. Steil, K. L. Swan, J. Dziura, N. Kurtz, W. V. Tamborlane, "Fully automated closed-loop insulin delivery versus semiautomated hybrid control in pediatric patients with type 1 diabetes using an artificial pancreas," Diabetes Care, vol. 31 no. 5, pp. 934-939, DOI: 10.2337/dc07-1967, 2008.
[9] Panel, "International association of diabetes and pregnancy study groups recommendations on the diagnosis and classification of hyperglycemia in pregnancy," Diabetes Care, vol. 33 no. 3, pp. 676-682, 2010.
[10] A. D. Association, "Management of diabetes in pregnancy: standards of medical care in diabetes—2020," Diabetes Care, vol. 43 no. Supplement 1, pp. S183-S192, DOI: 10.2337/dc20-s014, 2020.
[11] S. Handwerger, M. Freemark, "The roles of placental growth hormone and placental lactogen in the regulation of human fetal growth and development," Journal of Pediatric Endocrinology and Metabolism, vol. 13 no. 4, pp. 343-356, DOI: 10.1515/jpem.2000.13.4.343, 2000.
[12] L. Barbour, "New concepts in insulin resistance of pregnancy and gestational diabetes: long-term implications for mother and offspring," Journal of Obstetrics and Gynaecology, vol. 23 no. 5, pp. 545-549, DOI: 10.1080/0144361031000156500, 2003.
[13] H. Yamashita, J. Shao, J. E. Friedman, "Physiologic and molecular alterations in carbohydrate metabolism during pregnancy and gestational diabetes mellitus," Clinical Obstetrics and Gynecology, vol. 43 no. 1, pp. 87-98, DOI: 10.1097/00003081-200003000-00009, 2000.
[14] L. R. Mack, P. G. Tomich, "Gestational diabetes: diagnosis, classification, and clinical care gestational diabetes," Obstetrics and Gynecology Clinics of North America, vol. 44 no. 2, pp. 207-217, DOI: 10.1016/j.ogc.2017.02.002, 2017.
[15] G. Guzmán, V. Martínez, J. D. Yara, M. A. Mina, J. S. Solarte, A. M. Victoria, K Fériz, "Glycemic control and hypoglycemia in patients treated with insulin pump therapy: an observational study," Journal of Diabetes Research, vol. 2020,DOI: 10.1155/2020/1581726, 2020.
[16] O. Cohen, N. Keidar, M. Simchen, B. Weisz, M. Dolitsky, E. Sivan, "Macrosomia in well controlled CSII treated type I diabetic pregnancy," Gynecological Endocrinology, vol. 24 no. 11, pp. 611-613, DOI: 10.1080/09513590802531062, 2008.
[17] D. Bruttomesso, M. Bonomo, S. Costa, M. Dal Pos, G. Di Cianni, F. Pellicano, E. Vitacolonna, A. R. Dodesini, L. Tonutti, A. Lapolla, A. Di Benedetto, E. Torlone, "Type 1 diabetes control and pregnancy outcomes in women treated with continuous subcutaneous insulin infusion (CSII) or with insulin glargine and multiple daily injections of rapid-acting insulin analogues (glargine-MDI)," Diabetes and Metabolism, vol. 37 no. 5, pp. 426-431, DOI: 10.1016/j.diabet.2011.02.002, 2011.
[18] D. Farrar, D. J. Tuffnell, J. West, H. M. West, "Continuous subcutaneous insulin infusion versus multiple daily injections of insulin for pregnant women with diabetes," Cochrane Database of Systematic Reviews, vol. 6,DOI: 10.1002/14651858.cd005542.pub3, 2016.
[19] G. Petrovski, C. Dimitrovski, M. Bogoev, T. Milenkovic, I. Ahmeti, I. Bitovska, "Is there a difference in pregnancy and glycemic outcome in patients with type 1 diabetes on insulin pump with constant or intermittent glucose monitoring? A pilot study," Diabetes Technology & Therapeutics, vol. 13 no. 11, pp. 1109-1113, DOI: 10.1089/dia.2011.0081, 2011.
[20] D. S. Feig, L. E. Donovan, R. Corcoy, K. E. Murphy, S. A. Amiel, K. F. Hunt, E Asztalos, J. F. R Barrett, J. J Sanchez, A de Leiva, M Hod, L Jovanovic, E Keely, R McManus, E. K Hutton, C. L Meek, Z. A Stewart, T Wysocki, R O’Brien, K Ruedy, C Kollman, G Tomlinson, H. R Murphy, "Continuous glucose monitoring in pregnant women with type 1 diabetes (CONCEPTT): a multicentre international randomised controlled trial," Lancet (London, England), vol. 390 no. 10110, pp. 2347-2359, DOI: 10.1016/S0140-6736(17)32400-5, 2017.
[21] C. A. Combs, "Continuous glucose monitoring and insulin pump therapy for diabetes in pregnancy," Journal of Maternal-Fetal and Neonatal Medicine, vol. 25 no. 10, pp. 2025-2027, DOI: 10.3109/14767058.2012.670409, 2012.
[22] J. Kropff, J. H. DeVries, "Continuous glucose monitoring, future products, and update on worldwide artificial pancreas projects," Diabetes Technology & Therapeutics, vol. 18 no. Suppl 2, pp. S253-S263, DOI: 10.1089/dia.2015.0345, 2016.
[23] M. Breton, A. Farret, D. Bruttomesso, S. Anderson, L. Magni, S. Patek, C. Dalla Man, J. Place, S. Demartini, S. Del Favero, C. Toffanin, C. Hughes-Karvetski, E. Dassau, H. Zisser, F. J. Doyle, G. De Nicolao, A. Avogaro, C. Cobelli, E. Renard, B. Kovatchev, "Fully integrated artificial pancreas in type 1 diabetes: modular closed-loop glucose control maintains near normoglycemia," Diabetes, vol. 61 no. 9, pp. 2230-2237, DOI: 10.2337/db11-1445, 2012.
[24] R. Hovorka, K. Kumareswaran, J. Harris, J. M. Allen, D. Elleri, D. Xing, C. Kollman, M. Nodale, H. R. Murphy, D. B. Dunger, S. A. Amiel, S. R. Heller, M. E. Wilinska, M. L. Evans, "Overnight closed loop insulin delivery (artificial pancreas) in adults with type 1 diabetes: crossover randomised controlled studies," BMJ, vol. 342,DOI: 10.1136/bmj.d1855, 2011.
[25] Z. A. Stewart, M. E. Wilinska, S. Hartnell, R. C. Temple, G. Rayman, K. P. Stanley, D. Simmons, G. R. Law, E. M. Scott, R. Hovorka, H. R. Murphy, "Closed-loop insulin delivery during pregnancy in women with type 1 diabetes," New England Journal of Medicine, vol. 375 no. 7, pp. 644-654, DOI: 10.1056/nejmoa1602494, 2016.
[26] M. Kalligeros, F. Shehadeh, E. K. Mylona, G. Benitez, C. G. Beckwith, P. A. Chan, E. Mylonakis, "Association of obesity with disease severity among patients with coronavirus disease 2019," Obesity (Silver Spring), vol. 28 no. 7, pp. 1200-1204, DOI: 10.1002/oby.22859, 2020.
[27] S. Polsky, H. K. Akturk, "Case series of a hybrid closed-loop system used in pregnancies in clinical practice," Diabetes/Metabolism Research and Reviews, vol. 36 no. 3,DOI: 10.1002/dmrr.3248, 2020.
[28] H. Ghazanfar, S. Rizvi, A. Khurram, F. Orooj, I. Qaiser, "Impact of insulin pump on quality of life of diabetic patients," Indian Journal of Endocrinology and Metabolism, vol. 20 no. 4,DOI: 10.4103/2230-8210.183472, 2016.
You have requested "on-the-fly" machine translation of selected content from our databases. This functionality is provided solely for your convenience and is in no way intended to replace human translation. Show full disclaimer
Neither ProQuest nor its licensors make any representations or warranties with respect to the translations. The translations are automatically generated "AS IS" and "AS AVAILABLE" and are not retained in our systems. PROQUEST AND ITS LICENSORS SPECIFICALLY DISCLAIM ANY AND ALL EXPRESS OR IMPLIED WARRANTIES, INCLUDING WITHOUT LIMITATION, ANY WARRANTIES FOR AVAILABILITY, ACCURACY, TIMELINESS, COMPLETENESS, NON-INFRINGMENT, MERCHANTABILITY OR FITNESS FOR A PARTICULAR PURPOSE. Your use of the translations is subject to all use restrictions contained in your Electronic Products License Agreement and by using the translation functionality you agree to forgo any and all claims against ProQuest or its licensors for your use of the translation functionality and any output derived there from. Hide full disclaimer
Copyright © 2021 Guillermo E. Guzmán Gómez et al. This is an open access article distributed under the Creative Commons Attribution License (the “License”), which permits unrestricted use, distribution, and reproduction in any medium, provided the original work is properly cited. Notwithstanding the ProQuest Terms and Conditions, you may use this content in accordance with the terms of the License. https://creativecommons.org/licenses/by/4.0/
Abstract
Objective. Closed-loop insulin systems represent a technological advance in diabetes management but have rarely been studied in pregnancy. We report a case of a patient with type 1 diabetes mellitus who was previously a user of the Paradigm VEO pump and then migrated to Medtronic 670G. Research Design and Methods. We reviewed the case of a G1P0 patient with type 1 diabetes, treated with the Medtronic 670G system during pregnancy; a comparison with current literature was done. Results. The patient achieved improved glycemic control as measured by time spent in different ranges as follows: <70 mg/dL, 8–4% and 70–180 mg/dL, 83–94%. Secondary outcomes included reduction of stress, anxiety, fear of hypoglycemia, and the psychological burden related to the disease. There were no obstetric or neonatal complications. Conclusion. The Medtronic 670G closed-loop system was used safely in a pregnant woman; nevertheless, further research is needed to validate this system in this patient population.
You have requested "on-the-fly" machine translation of selected content from our databases. This functionality is provided solely for your convenience and is in no way intended to replace human translation. Show full disclaimer
Neither ProQuest nor its licensors make any representations or warranties with respect to the translations. The translations are automatically generated "AS IS" and "AS AVAILABLE" and are not retained in our systems. PROQUEST AND ITS LICENSORS SPECIFICALLY DISCLAIM ANY AND ALL EXPRESS OR IMPLIED WARRANTIES, INCLUDING WITHOUT LIMITATION, ANY WARRANTIES FOR AVAILABILITY, ACCURACY, TIMELINESS, COMPLETENESS, NON-INFRINGMENT, MERCHANTABILITY OR FITNESS FOR A PARTICULAR PURPOSE. Your use of the translations is subject to all use restrictions contained in your Electronic Products License Agreement and by using the translation functionality you agree to forgo any and all claims against ProQuest or its licensors for your use of the translation functionality and any output derived there from. Hide full disclaimer
Details

1 Fundación Clínica Valle del Lili, Department of Endocrinology, Universidad Icesi, Cali, Colombia
2 PGY3 Internal Medicine Resident, Fundación Clínica Valle del Lili, Universidad Icesi, Cali, Colombia
3 Universidad Icesi, Cali, Colombia
4 Fundación Clínica Valle del Lili, Department of Internal Medicine, Universidad Icesi, Cali, Colombia
5 Fundación Clínica Valle del Lili, Clinical Investigations Center, Cali, Colombia