Abstract

Abstract

Objectives

To evaluate the impact of the reconstructed field-of-view (FOV) on image quality in computed-tomography angiography (CTA) of the lower extremities.

Methods

A total of 100 CTA examinations of the lower extremities were acquired on a 2 × 192-slice multidetector CT (MDCT) scanner. Three different datasets were reconstructed covering both legs (standard FOV size) as well as each leg separately (reduced FOV size). The subjective image quality was evaluated for the different vessel segments (femoral, popliteal, crural, pedal) by three readers using a semi-quantitative Likert scale. Additionally, objective image quality was assessed using an automated image quality metric on a per-slice basis.

Results

The subjective assessment of the image quality showed an almost perfect interrater agreement. The image quality of the small FOV datasets was rated significantly higher as compared to the large datasets for all patients and vessel segments (p < 0.05) with a tendency towards a higher effect in smaller vessels. The difference of the mean scores between the group with the large FOV and small FOV was 0.68 for the femoral level, 0.83 for the popliteal level, 1.12 for the crural level, and 1.08 for the pedal level. The objective image quality metric also demonstrated a significant improvement of image quality in the small FOV datasets.

Conclusions

Side-separated reconstruction of each leg in CTA of the lower extremities using a small reconstruction FOV significantly improves image quality as compared to a standard reconstruction with a large FOV covering both legs.

Key Points

In CT angiography of the lower legs, the side-separated reconstruction of each leg using a small field-of-views improves image quality as compared to a standard reconstruction covering both legs.

The side-separated reconstruction can be readily implemented at every commercially available CT scanner.

There is no need for additional hardware or software and no additional burden to the patient.

Details

Title
The impact of the field of view (FOV) on image quality in MDCT angiography of the lower extremities
Author
Salimova, Nigar 1 ; Hinrichs, Jan B. 1 ; Gutberlet, Marcel 1 ; Meyer, Bernhard C. 1 ; Wacker, Frank K. 1 ; von Falck, Christian 1 

 Hannover Medical School, Department of Diagnostic and Interventional Radiology, Hannover, Germany (GRID:grid.10423.34) (ISNI:0000 0000 9529 9877) 
Pages
2875-2882
Publication year
2022
Publication date
May 2022
Publisher
Springer Nature B.V.
ISSN
09387994
e-ISSN
14321084
Source type
Scholarly Journal
Language of publication
English