1. Introduction
Radiotherapy is the main treatment method for nasopharyngeal carcinoma (NPC). However, the presence of several important organs adjacent to the nasopharynx, such as the optic chiasma, brainstem, spinal cord, and parotid gland, increases the difficulty of radiotherapy. Conventional external beam radiotherapy often leads to side effects and has an adverse impact on the patient’s quality of life. In recent years, several new treatment techniques have been developed, including intensity-modulated radiotherapy (IMRT), volumetric-modulated arc therapy (VMAT), and tomotherapy. These techniques mainly involve stacking a large number of small fields to form a large field with different dose intensities, which allows for a more conformal dose curve surrounding the tumor, thereby lowering the dose to normal tissues and organs [1, 2]. This greatly reduces the probability of side effects in NPC patients.
Many studies have shown that tissue inhomogeneities, due to air cavities, lead to dose reduction at the air-tissue interface [3–8]. Because the scope of NPC treatment involves the nasopharyngeal cavity, this would be expected to affect the accuracy of dose calculations in the treatment planning system (TPS), with the magnitude of the dose reduction dependent on the cavity volume, field size, and photon energy. This effect is greater with a smaller field size, higher energy, and larger cavity volume [3, 5–7]. This is primarily due to the electronic disequilibrium effect at the air-tissue interface in the presence of air cavities, leading to a dose reduction at the air-tissue interface. For NPC patients, such interface regions may contain tumor cells, which could increase the probability of future tumor recurrence. Therefore, understanding the effect of the nasopharyngeal cavity on the accuracy dose calculations in a clinical TPS is necessary.
Monte Carlo simulation is one of the reliable methods to evaluate the dose of build-up area and heterogeneous substance. In order to simulate the effect of simple irradiation conditions (single field or rectangular field) on dose prescription, most of the studies have generally involved the measurement of simple fields or an applied Monte Carlo simulation. Zhao et al. further evaluated the accuracy of clinical tomotherapy treatment plans in the heterogeneous phantom against Monte Carlo simulation and measurement [9]. Haga et al. utilized XVMC algorithm that is a fast Monte Carlo calculation algorithm to compare with the dose of clinical VMAT treatment plans in the homogeneous and heterogeneous phantom [10]. Their works revealed that the commercial Pinnacle and TomoTherapy TPS provide acceptable dose accuracy in most condition and the Monte Carlo simulation is a trustworthy and feasible method to verify clinical dose, especially for inhomogeneous and low-dose region.
Our main concern was regarding the effect of the nasopharyngeal cavity on the actual dose delivery during the treatment of early stage NPC in a clinical setting. The nasopharyngeal cavity in this stage would not like the advanced cases to be filled with tumor tissue, and the maximum cavity volume could be maintained. Hence, in this study, ultrathin dosimeters, such as GR-200F, TLD-100, and EBT3 radiochromic film, were used to conduct dosimetric measurements in order to evaluate the accuracy of dose calculations for the nasopharyngeal cavity during clinical treatment, using the commercial Pinnacle and TomoTherapy TPS, which utilize collapsed cone convolution superposition (CCCS) algorithms. The aim of this study was to understand the effect of the nasopharynx on TPS dose calculations.
2. Methods and Materials
2.1. Phantom Design
A custom-made anthropomorphic acrylic phantom of the head was created by milling machine according to coronal computed-tomography (CT) images of patients. The phantom contained an air cavity similar in shape to the nasopharyngeal cavity of the patient. This custom-made anthropomorphic phantom consisted of 47 coronal sections in slices, which were fixed in place using five acrylic cylinders. EBT3 radiochromic film and thermoluminescent dosimeters (TLD) could be placed on these coronal slices in order to measure the radiation dose and dose distribution. The slice thickness at the tumor site was 5 mm, as shown in Figure 1.
[figure(s) omitted; refer to PDF]
2.2. Treatment Plan
CT (LightSpeed GE, USA) was used to acquire images of the phantom using a CT slice thickness of 1.25 mm. The images were transferred to Pinnacle TPS (Pinnacle3® version 9.8 Philips, USA) and TomoTherapy TPS (TomoTherapy Planning Station Hi-ART ® version 4.3.2 Accuray, USA). Pinnacle TPS and Tomotherapy TPS based on CCCS are the commonly used commercial treatment planning system for linear accelerator and tomotherapy to treat NPC disease, respectively. A clinical radiation oncologist defined the position and treatment range of the NPC with tumor stage T1N0M0, including the gross target volume (GTV), clinical target volume 1 (CTV1), and clinical target volume 2 (CTV2). A 3-mm margin was used to form the planning target volume (PTV). The positions of the normal tissues and organs were also specified (Figure 2). The GTV, PTV1, and PTV2 were prescribed doses of 7200 cGy, 6480 cGy, and 5940 cGy, respectively. Pinnacle TPS was used to create the treatment plan required to implement VMAT and IMRT. The Elekta Axesse (Sweden) linear accelerator was employed, together with the Agility 5 mm leaf width multileaf collimator (MLC), using 6 MV photon beams. TomoTherapy TPS was used to calculate the treatment plan required for tomotherapy using 6 MV photon beams.
[figure(s) omitted; refer to PDF]
In order to evaluate the effect of actual nasopharyngeal cavity on the simple irradiation conditions, the beam with
The nine beam angles applied in IMRT were 165°, 130°, 80°, 40°, 0°, 320°, 280°, 230°, and 195°. The minimum segment area was set at 4 cm2, the minimum segment monitor unit (MU) was set at 3, and the dose rate was 600 MU/min. The beams applied in VMAT were counterclockwise from 176° to 184° with a collimator angle of 350°, and clockwise from 184° to 176° with a collimator angle of 10°. The maximum dose rate was 600 MU/min and the dose grid of the three axes set in the TPS was 3 mm. Furthermore, 6 MV photon beams were used for tomotherapy, and the parameters were as follows: modulation factor, 3.5; field width, 2.5 cm; pitch, 0.215; maximum dose rate, 900 MU/min; and the fine dose grid setting in the TPS.
2.3. Dose Algorithm
Pinnacle and TomoTherapy TPS both employ CCCS algorithms for dose calculations. This algorithm is able to correct for the influence of tissue inhomogeneity, patient surface contours, and beam modifiers based on correcting the energy fluence. The system calculates the total energy released per unit mass (TERMA) of each monoenergetic beam after entering the tissue. The TERMA is then convolved with the dose kernel to calculate the dose deposit caused by the beams. The dose kernel spread function of various monoenergetic beams that the TPS used to calculate dose is generated in water by Monte Carlo method with the assumption of the existence of electronic equilibrium. Finally, total incident energy in the energy spectrum is summarized to obtain the dose [12–14]. The following equation describes the dose distribution:
2.4. Radiation Dose Measurements
This study used GR-200F (LiF:Mg,Cu,P, Hangzhou Freq-Control Electronic Technology Ltd, China) and Harshaw TLD-100 (LiF:Mg,Ti, Thermo, USA) to measure the surface dose at the interface between the nasopharyngeal cavity and tissues. The measurement points are shown in Figure 2. The five-time measurements were performed at each point. GR-200F is a highly sensitive, ultrathin TLD with a diameter of 0.5 cm. The TLD powder with thickness of about 0.1 mm (effective thickness: 5 mg·cm-2) and density of 2.675 g·cm-3 was coated and adhered to one side of PE supported layer and is suitable for measuring surface doses [16]. The reader used was the Rexon UL-320 (Rexon, USA). TLD-100 has dimensions of
EBT3 radiochromic film (Ashland, USA) was also used in this study to measure the dose distribution on the coronal plane and the dose at the interface of the tumor site. An Epson Expression 10000XL scanner was used to scan the film using a red scanning light source. FilmQA™ software (version 2.2) was then used to analyze the coronal dose, followed by gamma evaluation to determine the difference between the two-dimensional doses measured using the EBT3 film and the treatment plan. Since the purpose of this study was to investigate the dose perturbance of nasopharyngeal cavity, the range of gamma analysis was focused on the high-dose area,
Setup of all measurement was accurately aligned by an alignment laser. IMRT and VMAT measurements were accompanied by alignment using the six degree-of-freedom image-guided cone-beam CT to confirm the position. MVCT was utilized to align the tomotherapy measurements and correct the position shifts and roll-angle deviations in the rotating gantry.
3. Results
3.1. Calculation of Results of Treatment Plans
The
Table 1
Dose summary of target and organ at risk calculated by IMRT, VMAT, and tomotherapy.
IMRT | VMAT | Tomotherapy | |
100 | 100 | 100 | |
96.0 | 96.0 | 100 | |
95.0 | 97.0 | 99.9 | |
3899 | 3806 | 3574 | |
3597 | 3228 | 3257 | |
2209 | 2174 | 2387 | |
2101 | 2057 | 2064 |
3.2. Dose Measurements of Dosimeters
3.2.1. Dose Measurements of Single Field and Parallel Opposed Fields
Figure 3 shows the dose-linearity curves of EBT3, TLD-100, and GR-200F for 6 MV photon beams (0–300 cGy); the
[figure(s) omitted; refer to PDF]
When the irradiation was a
When the irradiation was a
Figure 5 shows the relative dose of the depth behind the air-tissue interface of
[figure(s) omitted; refer to PDF]
Figure 6 shows two
[figure(s) omitted; refer to PDF]
3.2.2. Dose Measurements of Multiple Fields in Clinical Treatment Plans
Table 2 shows the results measured using TLD-100 and GR-200F at five dose points on the tissue interface, the difference between these measurements, and the calculated TPS values. The quadratic means (QM) of the percentage errors determined using TLD-100 and GR-200F values for the IMRT treatment plan were 2.3% and 2.1%, respectively; those for the VMAT treatment plan were 2.3% and 4.0%, respectively, while those for the tomotherapy treatment plan were both 3.1%. For IMRT, the GRF-200F measurements indicated that the TPS at point 5 and tomotherapy at point 1 were overestimated by 4.2% and 4.3%, respectively, while the remaining points were not significantly underdosed. Overall, TLD-100 and GR-200F did not show significant differences in the measurement of treatment plans using three clinical treatment techniques. This implies that these clinical treatment techniques did not form significant rebuild-up regions at the nasopharyngeal air-tissue interface. Moreover, the measured results of TLD with different thickness under clinical applications of multi-angle irradiation did not significantly different from the TPS calculations.
Table 2
Dosage measured by TLD-100 and GR-200F for IMRT, VMAT, and tomotherapy at the air-tissue interface.
Position | TPS calculation (cGy) | TLD-100 measurement (cGy) | GR-200F measurement (cGy) | TLD-100 difference (%) | GR-200 difference (%) | |
IMRT | 1 | 167.8 | −3.9 | 0.7 | ||
2 | 208.0 | −2.8 | 0.4 | |||
3 | 184.6 | 1.8 | 1.5 | |||
4 | 167.1 | −1.0 | 1.1 | |||
5 | 198.7 | −0.5 | −4.2 | |||
QM | 2.3 | 2.1 | ||||
VMAT | 1 | 169.1 | 2.5 | 4.4 | ||
2 | 207.3 | 3.2 | 5.8 | |||
3 | 190.0 | 1.2 | 4.1 | |||
4 | 169.2 | −0.5 | 0.7 | |||
5 | 202.2 | 2.9 | 3.1 | |||
QM | 2.3 | 4.0 | ||||
Tomotherapy | 1 | 171.6 | −3.3 | −4.3 | ||
2 | 204.1 | −1.4 | 2.8 | |||
3 | 188.4 | −2.3 | 1.0 | |||
4 | 156.03 | −3.8 | 0.6 | |||
5 | 185.6 | −3.9 | 4.6 | |||
QM | 3.1 | 3.1 |
The dose profiles of the phantom cavity on the coronal plane, measured using EBT3 with the beam direction along the left-right directions of the body (
[figure(s) omitted; refer to PDF]
Figure 8 shows the gamma evaluation maps for nasopharyngeal cavity on the coronal plane, measured within the phantom by EBT3; the gamma criteria were 3% and 3 mm. The gamma pass rates measured for IMRT, VMAT, and tomotherapy were 98.5%, 94.1%, and 97.3%, respectively. These results indicate that, despite the presence of a cavity, EBT3 measurements still showed high pass rates. However, certain fail regions were observed in the nasopharyngeal cavity for VMAT, as indicated by higher measurement dose compared to the calculated values. The higher measurement dose might be due to the error of MLC position rather than electronic disequilibrium effect. Previous study had revealed that it will cause significant impact on the dose even the error of MLC position less than 1 mm [18, 19]. The results measured for all three clinical techniques did not show evidence of fail regions that would be caused by significant dose overestimation at the nasopharyngeal air-tissue interface.
[figure(s) omitted; refer to PDF]
4. Discussion
The differences in the measured and calculated values at the proximal interface were smaller than 3.5% with regard to the
Kan et al. used TLD to measure the BUR-1 of a phantom containing a
The primary reason for the dose overestimation of TPS calculation values at the air-tissue interface is that the CCCS algorithms used in the TPS perform distribution expansion of the cavity kernel based on density scaling. However, the magnitude of expansion may be underestimated, which leads to an underestimation of the lateral electronic disequilibrium effect at the air-tissue interface. This will in turn cause the TPS to overestimate the dose deposit at the interface under a small field. Due to the differences in dose contributions by lateral electronic scattering, the electronic disequilibrium effect at an interface with perpendicular beam incidence will be more severe than an interface parallel to the beam. Kan et al. showed that the rebuild-up effects at the interface due to electronic disequilibrium caused by the lack of lateral electronic scatter are negligible when the field size is greater than
Wang et al. employed the Monte Carlo method to calculate the treatment plan doses of head-and-neck cancer patients receiving 3D conformal radiotherapy [23]. These calculated doses were then compared to the original treatment plan calculation based on the pencil beam algorithm. Their results showed that there were no significant differences between the two algorithms in clinical 3D conformal radiotherapy. Our results also showed that the interface doses for IMRT, VMAT, and tomotherapy were not overestimated due to electronic disequilibrium, and the measured differences in dose were all within the clinical uncertainty range of dose measurements (±5% recommended by AAPM TG-119 and TG-142) [24, 25]. This was primarily because the beam angles specified in the clinical treatment plans were multiangle radiation beams. In terms of the air-tissue interface, radiation doses that passed through the cavity and were incident perpendicularly on the tissue interface, with field sizes smaller than
Chen et al. showed that the outcomes of IMRT and 3DCRT were similar considering locoregional control (LRC) for early-stage patients [26]. IMRT was associated with better LRC compared with 3DCRT for advanced-stage patients. The result of clinical follow-up also revealed that the nasopharyngeal cavity has no significant effect on the outcome of the IMRT treatment.
5. Conclusions
This study examined the accuracy of CCCS algorithms in calculating the interface dose under a simple small field in a simulated model that closely aligned to the actual human nasopharyngeal profile. In addition, this study also measured the accuracy of the algorithm at the nasopharyngeal cavity and soft tissue interface, when applied to IMRT, VMAT, and tomotherapy in the clinical treatment of early stage NPC that the air cavity would not be filled up by tumor and the maximum cavity volume could be maintained. Under single small fields and parallel opposed fields, the measurement results indicated that the TPS overestimated the interface dose. However, measurements of IMRT, VMAT, and tomotherapy revealed that the measurement values at the air-tissue interface were not significantly lower than the calculated values. This indicates that under clinical multiangle irradiation, the dose errors from these calculations, due to the small-field electronic disequilibrium effect, are negligible when compared to the total prescribed dose. Nevertheless, to further reduce the impact of electronic disequilibrium caused by small-field size at the air-tissue interface, the following measures can be taken: (a) limit the minimum area of subfields when formulating treatment plans, (b) prescribe the dose using multiangle beams whenever possible, (c) reduce the dose weighting of beams entering the concerned interface through the cavity, and (d) use beam eye view to examine the area of each subfield; if the subfield area of beams entering the tissue interface through the cavity is too small or unilaterally too narrow, then these can be removed manually or the subfield area can be enlarged.
Disclosure
The preliminary results of this research have been published on the 33nd Joint Annual Conference of Biomedical Science (2018) in the form of a poster (poster number: MI0120).
Authors’ Contributions
Yuan-Chun Lai and Li-Chung Hung contributed equally to this work.
Acknowledgments
This study was supported in part by the Ministry of Science and Technology and Atomic Energy Council of Taiwan (MOST 109-2314-B-010-026 and AEC 11012048 L).
[1] M. A. Hunt, M. J. Zelefsky, S. Wolden, C. S. Chui, T. LoSasso, K. Rosenzweig, L. Chong, S. V. Spirou, L. Fromme, M. Lumley, H. A. Amols, C. C. Ling, S. A. Leibel, "Treatment planning and delivery of intensity-modulated radiation therapy for primary nasopharynx cancer," International Journal of Radiation Oncology • Biology • Physics, vol. 49 no. 3, pp. 623-632, DOI: 10.1016/S0360-3016(00)01389-4, 2001.
[2] S. Qu, Z. G. Liang, X. D. Zhu, "Advances and challenges in intensity-modulated radiotherapy for nasopharyngeal carcinoma," Asian Pacific Journal of Cancer Prevention, vol. 16 no. 5, pp. 1687-1692, DOI: 10.7314/APJCP.2015.16.5.1687, 2015.
[3] E. E. Klein, L. M. Chin, R. K. Rice, B. J. Mijnheer, "The influence of air cavities on interface doses for photon beams," International Journal of Radiation Oncology • Biology • Physics, vol. 27 no. 2, pp. 419-427, DOI: 10.1016/0360-3016(93)90255-T, 1993.
[4] T. D. Solberg, F. E. Holly, A. A. De Salles, R. E. Wallace, J. B. Smathers, "Implications of tissue heterogeneity for radiosurgery in head and neck tumors," International Journal of Radiation Oncology • Biology • Physics, vol. 32 no. 1, pp. 235-239, DOI: 10.1016/0360-3016(94)00495-7, 1995.
[5] X. A. Li, C. Yu, T. Holmes, "A systematic evaluation of air cavity dose perturbation in megavoltage x-ray beams," Medical Physics, vol. 27 no. 5, pp. 1011-1017, DOI: 10.1118/1.598966, 2000.
[6] C. F. Behrens, "Dose build-up behind air cavities for Co-60, 4, 6 and 8 MV. Measurements and Monte Carlo simulations," Physics in Medicine and Biology, vol. 51 no. 22, pp. 5937-5950, DOI: 10.1088/0031-9155/51/22/015, 2006.
[7] C. P. Joshi, J. Darko, P. B. Vidyasagar, L. J. Schreiner, "Dosimetry of interface region near closed air cavities for Co-60, 6 MV and 15 MV photon beams using Monte Carlo simulations," Journal of Medical Physics/Association of Medical Physicists of India, vol. 35 no. 2, pp. 73-80, DOI: 10.4103/0971-6203.62197, 2010.
[8] P. Haraldsson, T. Knöös, H. Nyström, P. Engström, "Monte Carlo study of TLD measurements in air cavities," Physics in Medicine and Biology, vol. 48 no. 18, pp. N253-N259, DOI: 10.1088/0031-9155/48/18/401, 2003.
[9] Y. L. Zhao, M. Mackenzie, C. Kirkby, B. G. Fallone, "Monte Carlo evaluation of a treatment planning system for helical tomotherapy in an anthropomorphic heterogeneous phantom and for clinical treatment plans," Medical Physics, vol. 35 no. 12, pp. 5366-5374, DOI: 10.1118/1.3002316, 2008.
[10] A. Haga, T. Magome, S. Takenaka, T. Imae, A. Sakumi, A. Nomoto, H. Igaki, K. Shiraishi, H. Yamashita, K. Ohtomo, K. Nakagawa, "Independent absorbed-dose calculation using the Monte Carlo algorithm in volumetric modulated arc therapy," Radiation Oncology, vol. 9 no. 1, article 75,DOI: 10.1186/1748-717X-9-75, 2014.
[11] E. R. Epp, A. L. Boyer, K. P. Doppke, "Underdosing of lesions resulting from lack of electronic equilibrium in upper respiratory air cavities irradiated by 10 MV X-ray beams," International Journal of Radiation Oncology • Biology • Physics, vol. 2 no. 7-8, pp. 613-619, DOI: 10.1016/0360-3016(77)90040-2, 1977.
[12] T. R. Mackie, J. W. Springer, J. J. Battista, "A convolution method of calculating dose for 15-MV X rays," Medical Physics, vol. 12 no. 2, pp. 188-196, DOI: 10.1118/1.595774, 1985.
[13] A. Ahnesjö, P. Andreo, A. Brahme, "Calculation and application of point spread functions for treatment planning with high energy photon beams," Acta Oncologica, vol. 26 no. 1, pp. 49-56, DOI: 10.3109/02841868709092978, 1987.
[14] N. Papanikolaou, T. R. Mackie, C. Meger-Wells, M. Gehring, P. Peckwerdt, "Investigation of the convolution method for polyenergetic spectra," Medical Physics, vol. 20 no. 5, pp. 1327-1336, DOI: 10.1118/1.597154, 1993.
[15] C. Y. Huang, T. C. Chu, S. Y. Lin, J. P. Lin, C. Y. Hsieh, "Accuracy of the convolution/superposition dose calculation algorithm at the condition of electron disequilibrium," Applied Radiation and Isotopes, vol. 57 no. 6, pp. 825-830, DOI: 10.1016/S0969-8043(02)00228-2, 2002.
[16] J. P. Lin, T. C. Chu, S. Y. Lin, M. T. Liu, "Skin dose measurement by using ultra-thin TLDs," Applied Radiation and Isotopes, vol. 55 no. 3, pp. 383-391, DOI: 10.1016/S0969-8043(01)00082-3, 2001.
[17] N. Lee, J. Harris, A. S. Garden, W. Straube, B. Glisson, P. Xia, W. Bosch, W. H. Morrison, J. Quivey, W. Thorstad, C. Jones, K. K. Ang, "Intensity-modulated radiation therapy with or without chemotherapy for nasopharyngeal carcinoma: radiation therapy oncology group phase II trial 0225," Journal of Clinical Oncology, vol. 27 no. 22, pp. 3684-3690, DOI: 10.1200/JCO.2008.19.9109, 2009.
[18] K. Nithiyanantham, G. K. Mani, V. Subramani, L. Mueller, K. K. Palaniappan, T. Kataria, "Analysis of direct clinical consequences of MLC positional errors in volumetric-modulated arc therapy using 3D dosimetry system," Journal of Applied Clinical Medical Physics, vol. 16 no. 5, pp. 296-305, DOI: 10.1120/jacmp.v16i5.5515, 2015.
[19] F. Chen, M. Rao, J. S. Ye, D. M. Shepard, D. Cao, "Impact of leaf motion constraints on IMAT plan quality, deliver accuracy, and efficiency," Medical Physics, vol. 38 no. 11, pp. 6106-6118, DOI: 10.1118/1.3651698, 2011.
[20] M. W. Kan, J. Y. Cheung, L. H. Leung, B. M. Lau, P. K. Yu, "The accuracy of dose calculations by anisotropic analytical algorithms for stereotactic radiotherapy in nasopharyngeal carcinoma," Physics in Medicine and Biology, vol. 56 no. 2, pp. 397-413, DOI: 10.1088/0031-9155/56/2/008, 2011.
[21] C. Martens, N. Reynaert, C. De Wagter, P. Nilsson, M. A. Coghe, H. Palmans, H. Thierens, W. De Neve, "Underdosage of the upper-airway mucosa for small fields as used in intensity-modulated radiation therapy: a comparison between radiochromic film measurements, Monte Carlo simulations, and collapsed cone convolution calculations," Medical Physics, vol. 29 no. 7, pp. 1528-1535, DOI: 10.1118/1.1487421, 2002.
[22] W. K. Kan, P. M. Wu, H. T. Leung, T. C. Lo, C. W. Chung, D. L. W. Kwong, S. T. Sham, "The effect of the nasopharyngeal air cavity on x-ray interface doses," Physics in Medicine and Biology, vol. 43 no. 3, pp. 529-537, DOI: 10.1088/0031-9155/43/3/005, 1998.
[23] L. Wang, E. Yorke, C. S. Chui, "Monte Carlo evaluation of tissue inhomogeneity effects in the treatment of the head and neck," International Journal of Radiation Oncology • Biology • Physics, vol. 50 no. 5, pp. 1339-1349, DOI: 10.1016/S0360-3016(01)01614-5, 2001.
[24] E. E. Klein, J. Hanley, J. Bayouth, F. F. Yin, W. Simon, S. Dresser, C. Serago, F. Aguirre, L. Ma, B. Arjomandy, C. Liu, C. Sandin, T. Holmes, Task Group 142, American Association of Physicists in Medicine, "Task group 142 report: quality assurance of medical acceleratorsa," Medical Physics, vol. 36, 9 Part 1, pp. 4197-4212, DOI: 10.1118/1.3190392, 2009.
[25] G. A. Ezzell, J. W. Burmeister, N. Dogan, T. J. LoSasso, J. G. Mechalakos, D. Mihailidis, A. Molineu, J. R. Palta, C. R. Ramsey, B. J. Salter, J. Shi, P. Xia, N. J. Yue, Y. Xiao, "IMRT commissioning: multiple institution planning and dosimetry comparisons, a report from AAPM Task Group 119," Medical Physics, vol. 36 no. 11, pp. 5359-5373, DOI: 10.1118/1.3238104, 2009.
[26] J. L. Chen, Y. S. Huang, S. H. Kuo, R. L. Hong, J. Y. Ko, P. J. Lou, C. W. Wang, "Intensity-modulated radiation therapy achieves better local control compared to three-dimensional conformal radiation therapy for T4-stage nasopharyngeal carcinoma," Oncotarget, vol. 8 no. 8, pp. 14068-14077, DOI: 10.18632/oncotarget.12736, 2017.
You have requested "on-the-fly" machine translation of selected content from our databases. This functionality is provided solely for your convenience and is in no way intended to replace human translation. Show full disclaimer
Neither ProQuest nor its licensors make any representations or warranties with respect to the translations. The translations are automatically generated "AS IS" and "AS AVAILABLE" and are not retained in our systems. PROQUEST AND ITS LICENSORS SPECIFICALLY DISCLAIM ANY AND ALL EXPRESS OR IMPLIED WARRANTIES, INCLUDING WITHOUT LIMITATION, ANY WARRANTIES FOR AVAILABILITY, ACCURACY, TIMELINESS, COMPLETENESS, NON-INFRINGMENT, MERCHANTABILITY OR FITNESS FOR A PARTICULAR PURPOSE. Your use of the translations is subject to all use restrictions contained in your Electronic Products License Agreement and by using the translation functionality you agree to forgo any and all claims against ProQuest or its licensors for your use of the translation functionality and any output derived there from. Hide full disclaimer
Copyright © 2022 Yuan-Chun Lai et al. This is an open access article distributed under the Creative Commons Attribution License (the “License”), which permits unrestricted use, distribution, and reproduction in any medium, provided the original work is properly cited. Notwithstanding the ProQuest Terms and Conditions, you may use this content in accordance with the terms of the License. https://creativecommons.org/licenses/by/4.0/
Abstract
This study combined the use of radiation dosimeteric measurements and a custom-made anthropomorphic phantom in order to evaluate the accuracy of therapeutic dose calculations at the nasopharyngeal air-tissue interface. The doses at the nasopharyngeal air-tissue interface obtained utilizing the Pinnacle and TomoTherapy TPS, which are based on collapsed cone convolution superposition (CCCS) algorithms, were evaluated and measured under single
You have requested "on-the-fly" machine translation of selected content from our databases. This functionality is provided solely for your convenience and is in no way intended to replace human translation. Show full disclaimer
Neither ProQuest nor its licensors make any representations or warranties with respect to the translations. The translations are automatically generated "AS IS" and "AS AVAILABLE" and are not retained in our systems. PROQUEST AND ITS LICENSORS SPECIFICALLY DISCLAIM ANY AND ALL EXPRESS OR IMPLIED WARRANTIES, INCLUDING WITHOUT LIMITATION, ANY WARRANTIES FOR AVAILABILITY, ACCURACY, TIMELINESS, COMPLETENESS, NON-INFRINGMENT, MERCHANTABILITY OR FITNESS FOR A PARTICULAR PURPOSE. Your use of the translations is subject to all use restrictions contained in your Electronic Products License Agreement and by using the translation functionality you agree to forgo any and all claims against ProQuest or its licensors for your use of the translation functionality and any output derived there from. Hide full disclaimer
Details

1 Department of Radiation Oncology, Changhua Christian Hospital, Changhua, Taiwan; Department of Medical Imaging and Radiological Sciences, Central Taiwan University of Science and Technology, Taichung, Taiwan
2 Department of Radiation Oncology, Changhua Christian Hospital, Changhua, Taiwan
3 Department of Physics, National Chung Hsing University, Taichung, Taiwan
4 Department of Biomedical Imaging and Radiological Sciences, National Yang Ming Chiao Tung University, Taipei, Taiwan
5 Department of Radiation Oncology, Changhua Christian Hospital, Changhua, Taiwan; Department of Medical Imaging and Radiological Sciences, Central Taiwan University of Science and Technology, Taichung, Taiwan; Department of Medical Imaging and Radiological Technology, Yuanpei University of Medical Technology, Hsinchu, Taiwan