Background
The ClearSight system measures blood pressure non-invasively and determines cardiac output by analyzing the continuous pressure waveform. We performed a multi-center clinical study in China to test the equivalence of cardiac output measured with the ClearSight system (CSCO) and cardiac output measured with the pulmonary artery catheter bolus thermodilution (TDCO) method.
Methods
We included adult patients undergoing cardiac surgery in three Chinese hospitals and measured TDCO and CSCO simultaneously after induction of anesthesia. Hemodynamic stability was required during measurement of TDCO and CSCO. At least four TDCO determinations were performed. The corresponding CSCO was determined as the average over a 30-s period following the injection of each bolus. A data pair for the comparison included the average of three or four accepted TDCO values and the average of the matching CSCO values. Main outcomes included Bland-Altman analysis of bias and standard deviation (SD) and the percentage error (PE).
Results
One hundred twenty-five subjects were enrolled, and 122 TDCO and CSCO data pairs were available for analysis. Ninety-five (75.4%) data pairs were collected in hemodynamically stable conditions, mean (SD) CSCO was 4.21 (0.78) l/min, and mean TDCO was 3.90 (0.67) l/min. Bias was 0.32 (0.51) l/min, and PE was 25.2%. Analyzing all 122 data pairs resulted in a mean CSCO of 4.19 (0.82) l/min and a mean TDCO of 3.83 (0.71) l/min. Resulting bias was 0.36 (0.53) l/min, and PE was 26.4%.
Conclusions
CSCO and TDCO agreed with a low systematic bias. Besides, mean PE was well below the pre-defined 30%. Hemodynamic stability only had a small impact on the analysis. We conclude that CSCO is equivalent to TDCO in cardiac surgery patients.
The trial was retrospectively registered in ClinicalTrials.gov, identifier NCT03807622; January 17, 2019
You have requested "on-the-fly" machine translation of selected content from our databases. This functionality is provided solely for your convenience and is in no way intended to replace human translation. Show full disclaimer
Neither ProQuest nor its licensors make any representations or warranties with respect to the translations. The translations are automatically generated "AS IS" and "AS AVAILABLE" and are not retained in our systems. PROQUEST AND ITS LICENSORS SPECIFICALLY DISCLAIM ANY AND ALL EXPRESS OR IMPLIED WARRANTIES, INCLUDING WITHOUT LIMITATION, ANY WARRANTIES FOR AVAILABILITY, ACCURACY, TIMELINESS, COMPLETENESS, NON-INFRINGMENT, MERCHANTABILITY OR FITNESS FOR A PARTICULAR PURPOSE. Your use of the translations is subject to all use restrictions contained in your Electronic Products License Agreement and by using the translation functionality you agree to forgo any and all claims against ProQuest or its licensors for your use of the translation functionality and any output derived there from. Hide full disclaimer