Abstract
We compared the clinical utility of single-shot echo-planar imaging (SS-EPI) using different breathing schemes, readout-segmented EPI and zoomit EPI in the repeatability of apparent diffusion coefficient (ADC) measurements, cortico-medullary contrast to noise ratio (c-mCNR) and image quality. In this institutional review board-approved prospective study, some common clinically applicable diffusion-weighted imaging (b = 50, 400, 800 s/mm2) of kidney on 3.0 T MRI were performed on 22 volunteers using SS-EPI with breath-hold diffusion-weighted imaging (BH-DWI), free-breathing (FB-DWI), navigator-triggered (NT-DWI) and respiratory-triggered (RT-DWI), readout-segmented DWI (RS-DWI), and Zoomit DWI (Z-DWI). ADC and c-mCNR were measured in 12 anatomic locations (the upper, middle, and lower pole of the renal cortex and medulla), and image quality was assessed on these DWI sequences. A DWI with the optimal clinical utility was decided by systematically assessing the ADC repeatability, c-mCNR and image quality among the DWIs. For ADC measurements, Z-DWI had an excellent intra-observer agreement (intra-class correlation coefficients (ICCs): 0.876–0.944) and good inter-observer agreement (inter-class ICCs: 0.798–0.856) in six DWI sequences. Z-DWI had the highest ADC repeatability in most of the 12 anatomic locations of the kidneys (mean ADC absolute difference: 0.070–0.111 × 10−3 mm2/s, limit of agreement: 0.031–0.056 × 10−3 mm2/s). In all DWIs, Z-DWI yielded a slightly higher c-mCNR than other DWIs in most representative locations (P > 0.05), which was significantly higher than BH-DWI and FB-DWI in the middle pole of both kidneys and the upper pole of the left kidney (P < 0.05). In addition, Z-DWI yielded image quality that was similar to RT-DWI and NT-DWI (P > 0.05) and superior to BH-DWI, FB-DWI and RS-DWI (P < 0.05). Our results suggest that Z-DWI provides the highest ADC reproducibility, better c-mCNR and good image quality on 3.0 T MRI, making it the recommended sequence for clinical DWI of the kidney.
You have requested "on-the-fly" machine translation of selected content from our databases. This functionality is provided solely for your convenience and is in no way intended to replace human translation. Show full disclaimer
Neither ProQuest nor its licensors make any representations or warranties with respect to the translations. The translations are automatically generated "AS IS" and "AS AVAILABLE" and are not retained in our systems. PROQUEST AND ITS LICENSORS SPECIFICALLY DISCLAIM ANY AND ALL EXPRESS OR IMPLIED WARRANTIES, INCLUDING WITHOUT LIMITATION, ANY WARRANTIES FOR AVAILABILITY, ACCURACY, TIMELINESS, COMPLETENESS, NON-INFRINGMENT, MERCHANTABILITY OR FITNESS FOR A PARTICULAR PURPOSE. Your use of the translations is subject to all use restrictions contained in your Electronic Products License Agreement and by using the translation functionality you agree to forgo any and all claims against ProQuest or its licensors for your use of the translation functionality and any output derived there from. Hide full disclaimer
Details
1 Xiangya Hospital, Central South University, Department of Radiology, Changsha, People’s Republic of China (GRID:grid.452223.0) (ISNI:0000 0004 1757 7615); Xiangya Hospital, Central South University, National Clinical Research Center for Geriatric Disorders, Changsha, People’s Republic of China (GRID:grid.452223.0) (ISNI:0000 0004 1757 7615)
2 Xiangya Hospital, Central South University, Department of Radiology, Changsha, People’s Republic of China (GRID:grid.452223.0) (ISNI:0000 0004 1757 7615)
3 The Third Xiangya Hospital, Central South University, Department of Radiology, Changsha, People’s Republic of China (GRID:grid.431010.7)