Abstract

Background: Modern surgical staplers should provide precise placement and transection, especially in tight spaces and on thick tissue. Ideally, a stapler would move to accommodate variations in the tissue and anatomy instead of having to move the tissue around to fit the stapler. This study was undertaken to evaluate the performance characteristics of the new Echelon 3000 Stapler (ECH3). Use of the ECH3 was compared to another marketed stapler, including tests for access, seal strength, staple formation in thick tissue, and end effector stability.

Methods: Pelvic anatomy measurements were used to construct a virtual model of a Low Anterior Resection (LAR). Monte Carlo simulations were performed on the staplers to compare the probability of completing a transection with one or two firings. Using water infusion of stapled porcine ileum, pressure at first leak and percentage of leaks at critical pressures were measured. Rate of malformed staples was measured in thick tissue. End effector stability while firing and under moderate pressure were compared between staplers. After use, surgeons were surveyed on the functionality of the device.

Results: ECH3 had a markedly higher probability of completing an LAR transection in one or two firings than the comparator stapler. Median initial leak pressure of stapled ileum was significantly higher, and rate of leaks was lower at 40 and 50 mmHg. ECH3 had fewer malformed staples for both 3.3- and 4.0-mm thick tissue. The end effector exhibited less angular movement during firing, and less deflection under a moderate load. Surgeons agreed the ECH3 provided precise placement and easy one-handed operation.

Conclusion: The Echelon 3000 Stapler demonstrated improved access capability, tighter seals, fewer malformed staples, and greater end effector stability. These advantages were recognized by surgeons who evaluated the use of the device preclinically.

Details

Title
Improving Performance and Access to Difficult-to-Reach Anatomy with a Powered Articulating Stapler
Author
Huang, Z F; Vandewalle JA  VIAFID ORCID Logo  ; Clymer, J W  VIAFID ORCID Logo  ; Ricketts, C D; Petraiuolo, W J
Pages
329-339
Section
Original Research
Publication year
2022
Publication date
2022
Publisher
Taylor & Francis Ltd.
e-ISSN
1179-1470
Source type
Scholarly Journal
Language of publication
English
ProQuest document ID
2715162798
Copyright
© 2022. This work is licensed under https://creativecommons.org/licenses/by-nc/3.0/ (the “License”). Notwithstanding the ProQuest Terms and Conditions, you may use this content in accordance with the terms of the License.