Copyright © 2022 Mariia Smolinska et al. This is an open access article distributed under the Creative Commons Attribution License (the “License”), which permits unrestricted use, distribution, and reproduction in any medium, provided the original work is properly cited. Notwithstanding the ProQuest Terms and Conditions, you may use this content in accordance with the terms of the License. https://creativecommons.org/licenses/by/4.0/

Abstract

Simple, fast, and validated UV-spectrophotometric, HPLC, and GC methods for the analysis of benzalkonium chloride in a disinfectant were developed. UV-spectrophotometric determination was based on measuring the amount of light absorption of aqueous solutions of benzalkonium chloride at 268 nm. HPLC determination was achieved with a 150 mm × 4.6 mm, 5.0 μm C18 column. The mobile phase consisted of a 0.01% water solution of triethylamine (with pH 2.5) and acetonitrile in the ratio of 40 : 60 v/v. The column temperature was kept at 30°C, and the injection volume was 10 μL. The flow rate was 1.0 mL/min, and the diode array detector was set at 215 nm. GC determination was performed using a flame ionization detector on a glass capillary column ZB-WAX plus 30 m × 0.25 mm with an inner coating thickness of 0.25 μm. It creates a gradient increase in the temperature of the furnace to the maximum of 200°C. The temperature of the injector was 250°C, 1 μl was injected, and the separation of the sample was 1 : 100. Helium was used as a carrier gas, and the gas flow rate was constant, equaling 1.6 ml/min. The temperature of the detector was 250°C, and a mixture of hydrogen, helium, and air in a ratio of 30 : 8.5 : 350 was used in the split of the detector. All proposed methods were validated according to ICH guidelines with respect to the accuracy, precision (interday, intraday, and reproducibility), linearity, the limit of detection, the limit of quantitation, and robustness. All three methods were linear (R2 = 0.997–0.999) over a concentration range of 400–600 μg/ml for UV and 80–120 μg/ml for HPLC and GC, accurate (recovery was 98.4–101.7% for all methods), precise (RSD <2%), and robust. The cost of the analysis of one sample of the disinfectant for the content of benzalkonium chloride by three methods was calculated. The comparison of the obtained results facilitates a more efficient allocation of laboratory resources, depending on the goal.

Details

Title
Determination of Benzalkonium Chloride in a Disinfectant by UV Spectrophotometry and Gas and High-Performance Liquid Chromatography: Validation, Comparison of Characteristics, and Economic Feasibility
Author
Smolinska, Mariia 1   VIAFID ORCID Logo  ; Ostapiv, Roman 1   VIAFID ORCID Logo  ; Yurkevych, Mariia 1   VIAFID ORCID Logo  ; Poliuzhyn, Liubov 1 ; Korobova, Olha 1 ; Kotsiumbas, Ihor 1 ; Tesliar, Hryhorii 1 

 State Scientific Research Control Institute of Veterinary Preparations and Fodder Additives, Donetska Str 11, Lviv 79019, Ukraine 
Editor
Neil D Danielson
Publication year
2022
Publication date
2022
Publisher
John Wiley & Sons, Inc.
ISSN
16878760
e-ISSN
16878779
Source type
Scholarly Journal
Language of publication
English
ProQuest document ID
2720247947
Copyright
Copyright © 2022 Mariia Smolinska et al. This is an open access article distributed under the Creative Commons Attribution License (the “License”), which permits unrestricted use, distribution, and reproduction in any medium, provided the original work is properly cited. Notwithstanding the ProQuest Terms and Conditions, you may use this content in accordance with the terms of the License. https://creativecommons.org/licenses/by/4.0/