1. Introduction
Successful surgical cancer treatment requires radical surgical resection of the malignant tumor. Positive or close pathology tumor margins will significantly decrease survival and necessitate adjuvant treatments, increasing comorbidity risk and decreasing quality of life. Ex vivo imaging of the resected cancer tissue could, therefore, be useful for the cancer surgeon to assess the tumor margin immediately perioperatively. Different modalities to scan the ex vivo specimen have been proposed, such as magnetic resonance imaging (MRI) [1,2,3], computed tomography (CT) [3,4], and ultrasound [3,5].
CT is an imaging modality that can provide 3D images of the sample, enabling the margin assessment in all directions. Intraoperative imaging of the surgical specimen using CT has been previously proposed as a tool to reduce the number of reoperations [6]. High-resolution images of the ex vivo tissue have been obtained by micro-CT, which enabled the discrimination of cancer area from the normal tissue [7,8].
Ultrasound is a portable and cheap imaging modality that can be used to provide high-resolution visualization of surgical specimens [9]. It is possible to image the border between the tumor and healthy tissue accurately using ultrasound [10,11,12]. The limitation with 2D ultrasound is the user-dependency and generation of dynamic, but only two-dimensional, image “slices” of the tissue. However, new technology improvements have made it possible to obtain immediate 3D ultrasound imaging of the tumor, which, therefore, overcome some of the problems with B-mode ultrasound [13]. While pathology slicing provides interpretations based on the 2D observations, a 3D imaging model holds the potential of a better 3D description of the tumor. Several previous studies have reported a higher diagnostic accuracy using 3D ultrasound compared to the conventional 2D B-mode imaging [3,14,15,16,17].
In this work, we investigate the accuracy of 3D ultrasound, as a portable and low-cost modality, compared to CT to evaluate the tumor volume in an animal specimen.
2. Materials and Methods
We conducted an experimental study to compare the tumor volume measurement obtained using 3D ultrasound and CT to gross pathology. We used an animal model to imitate a surgical specimen with a soft tissue tumor. The animal model was made from a small chicken particle (tumor) placed in a piece of calf liver (normal tissue) with dimensions of about 1.5 cm × 2 cm × 3.5 cm. The phantom was wrapped tightly using cling wrap and kept in the fridge overnight. Then, the sample was placed in formalin for 24 h to fixate. The specimen was scanned using 3D ultrasound and CT modalities before it was sliced and evaluated by a pathologist. The tumor and normal tissue had very similar densities.
2.1. 3D Ultrasound and Computed Tomography Imaging
Ultrasound 3D imaging was performed using a SAMSUNG RS85 Prestige ultrasound machine, Copenhagen, Denmark, and a 3D linear ultrasound probe (LV3-14A) with a center frequency of 6.8 MHz. The specimen was placed in a box filled with isotone saline to perform the 3D ultrasound scan. A probe holder and a pair of elongators were designed and 3D-printed to keep the probe at the desired position and depth related to the specimen (see Figure 1).
The depth adjustment was optimized to preserve the highest possible resolution. To detect the curved edges in the axial direction, the compounding feature of the machine was employed. The image was optimized by gain adjustment and by setting the focus point at the center of the specimen.
The animal specimen was scanned with the CT scanner (Canon medical systems, Aquilion ONE, Copenhagen, Denmark) set at 120 kVp, 100 mA, and 0.5 mm of voxel dimension. The images were reconstructed with a 0.5 mm slice thickness and exported as DICOM file series.
2.2. Specimen Slicing
To validate the measurements performed by the ultrasound and CT data, a correlation with the gross pathology results was necessary. The formalin-fixated sample was embedded in 4% agar and sliced systematically into 2.0 mm thick slaps with a random starting point within the slab thickness (see Figure 2). Then, all the slaps were photographed for estimating the tumor area of each slide as well as the entire volume of the tumor in the specimen.
The boundaries of the tumor were delineated on the digital images of the slices by a pathologist. The pathological assessment was considered as the gold standard and used to evaluate the results from 3D ultrasound and CT.
2.3. Three-Dimensional Ultrasound and CT Data Processing
The acquired volumetric ultrasound images and CT scan data were imported in ITK-SNAP segmentation software, which is a free software application used to segment structures in 3D medical images (
Tumor area measurement on gross pathology was performed by an integration function in MATLAB (
(1)
(2)
where n is the slice number, L is the number of the tumor delineation rounds, and is the measured area of slice number n at the lth delineation session.Tumor volume () was calculated by multiplication of the summation of the mean slice areas ( by the slice intervals (d), which is 2 mm in this work.
(3)
The error percentage in the volume measurement could be calculated using the following equation:
(4)
where is the gold-standard gross pathology volume.We also measured the root-mean-square error (RMSE) of the results by the following equation:
(5)
where is the gross pathology tumor area measurement on the nth slide.3. Results
By the slicing technique described above, the model was sliced into 18 parallel 2 mm thick sections. A total of 11 of the slices included the tumor. When slicing the model, two of the slices were damaged, so we did not include them in the experiment. An example of the correlation between slices for gross pathology assessment and the corresponding 3D ultrasound and CT is presented in Figure 3. Figure 3a shows the digital image from a 2 mm thick slice. The corresponding planes from the 3D ultrasound volume and CT are presented in Figure 3b,c, respectively. Figure 3d shows the segmentation result in which two different labels are allocated to the tumor and surrounding healthy tissue region. A 3D image of the segmentation result is presented in Figure 3e.
The mean tumor area (Equation (1)) from the CT and 3D ultrasound at parallel, equally distanced planes corresponding to the pathological slices were measured and are reported in Figure 4 and Table 1.
The tumor volume (Equation (3)) measured by gross pathological findings was 980.2 mm3. The measured volume using CT was 890.4 ± 90 mm3, and 3D ultrasound resulted in a tumor volume of 924.2 ± 96 mm3. By using Equation (4), the volume measurement errors using CT and 3D ultrasound were 9.1% and 5.7%, respectively. To investigate how close the area measurements by 3D ultrasound/CT are to the gross pathology areas, we used Equation (5) for RMSE calculation. RMSE was 6.3 for the area measurements from 3D ultrasound and 8.2 for the results from CT.
According to the results presented in Figure 4, a higher overlap between the standard deviation of the 3D ultrasound measurements and the gross pathology was observed compared to the results from CT.
To statistically validate the hypothesis against measured data, we applied the paired t-test to test for significant differences between gross pathology and CT/3D ultrasound. As seen in Figure 5, the p-value for CT and 3D ultrasound was 0.045 and 0.121, respectively, showing a difference between gross pathology and CT, but not 3D ultrasound. However, the p-value for CT was still very close to 0.05. The correlation coefficient for CT was 0.91 and that for the 3D ultrasound was 0.96.
4. Discussion
In this experimental study, we found that 3D ultrasound is a feasible and accurate imaging method to assess the tumor volume in an animal specimen. The method is comparable to CT. The results were correlated to the gross pathological findings, which were considered the gold standard. An almost consistent standard deviation was observed in the results of 3D ultrasound (Figure 4a), with a higher variation present in the results obtained by CT (Figure 4b). These results suggest higher predictability of the accuracy of the results obtained from 3D ultrasound compared to CT. In addition, the standard deviation of the 3D ultrasound technique covered most of the data from gross pathology, while this was not the case in CT. Together, this points to the possibility of obtaining results closer to the pathological assessment by the ultrasound method compared to CT. Another advantage was the higher resolution of the ultrasound in the imaging of superficial soft tissue compared to CT, as can be seen in Figure 3b,c. The higher resolution in ultrasound volume led to a higher differentiation between the tumor and healthy tissue. Therefore, it was easier to delineate the tumor borders.
The strength of the current experiment is the precise slicing technique. The slicing device provided equal thickness and similar cutting directions for all the slices. This significantly reduces the error due to miscorrelation of the equal slices in different methods.
Our study finds that 3D ultrasound could be a promising modality to evaluate surgical specimens. The low cost and portability of ultrasound make it very suitable to implement in the operation room, while the specimen would need to be taken to the radiology department to be scanned by CT. Therefore, further studies are needed to explore the possibility of transferring the findings from current work to surgical specimen margin analysis.
Limitations
In this work, we used an animal model to conduct a proof-of-concept experiment to examine if 3D ultrasound leads to results comparable to CT when correlating to the gold standard of gross pathology. However, we do not know if the animal model is comparable to a real cancer model where the tumor cells are invading normal tissue. Moreover, we have only analyzed one animal specimen and the results, as well as discussions, are inferred from a limited experiment. Therefore, we are not able to compare the variability between different cases of tumors.
The three-dimensional ultrasound scan by the technique employed in the current work suffers from a limited field of view (37.4 mm × 29°), which could be a challenge for a larger specimen. One solution to this challenge could be freehand 3D ultrasound techniques. Freehand scanners enable the examination of the region of interest (ROI) in arbitrary directions and positions. However, positions and orientations of 2D B-scans are required for reconstructing 3D images, which adds up to the complexity of the method, and the accuracy of the results needs to be explored further.
Our animal model had the disadvantage that the density of the “tumor-tissue” and the background tissue was very similar, resulting in very low contrast between the tissues on CT. In patients, this is overcome by including a contrast agent in the examination, but this is not feasible ex vivo. The delineation of the tumor would probably have been more accurate with a larger difference in densities.
5. Conclusions
Three-dimensional ultrasound is a feasible imaging method for the measurement of soft tissue volume in an animal model comparable to CT. However, this is a pilot study of the concept and further studies on animal models as well as clinical trials on different tumor types are necessary to confirm the result. The low cost and portability of ultrasound make this a promising imaging modality that may be useful in the operating room for tumor assessment.
Conceptualization, F.M. and T.T.; methodology, F.M., T.K.A., C.E. and M.V.O.; software, F.M.; validation, F.M. and T.T.; investigation, F.M., C.E. and T.K.A.; resources, T.T. and B.P.; writing—original draft preparation, F.M.; writing—review and editing, C.E., T.K.A., M.V.O. and T.T.; visualization, F.M.; supervision, T.T.; project administration, T.T. All authors have read and agreed to the published version of the manuscript.
Not applicable.
Not applicable.
The experimental data in this study are available per request.
The software used in this work are ITK-SNAP—3.8.0 (
The authors declare no conflict of interest.
Publisher’s Note: MDPI stays neutral with regard to jurisdictional claims in published maps and institutional affiliations.
Figure 1. Three-dimensional ultrasound imaging of the specimen. A holder is used to keep the probe at the desired position. The height of the probe front-face in relation to the specimen surface is adjusted using the yellow elongators.
Figure 2. Slicing of the specimen using a specific device that allows for thin and parallel cutting.
Figure 3. An example of the correlation between slices for gross pathology assessment and corresponding imaging. (a) Specimen slice, (b) corresponding slice on 3D ultrasound image, (c) corresponding slice on CT image, (d) corresponding slice on segmentation, and (e) 3D segmentation of the animal model.
Figure 3. An example of the correlation between slices for gross pathology assessment and corresponding imaging. (a) Specimen slice, (b) corresponding slice on 3D ultrasound image, (c) corresponding slice on CT image, (d) corresponding slice on segmentation, and (e) 3D segmentation of the animal model.
Figure 4. Area measurement at parallel equally distanced planes corresponding to the pathological slices. The dashed blue line is the result of the pathological assessment, and the shaded gray area is the standard deviation of the three measurements. (a) 3D ultrasound results compared to the pathological assessment. The solid red line is the mean area at each slice obtained by 3D ultrasound. (b) CT results compared to the pathological assessment. The solid purple line is the mean area at each slice measured by CT.
Figure 5. Paired t-test for statistical evaluation of the CT and 3D ultrasound compared to gross pathology. ns stands for non-significant, and * stands for significant.
Mean area (A_mean ± S) in mm2 measured on each slice by CT, 3D ultrasound, and gross pathology. The mean results are calculated by averaging over three different delineation sessions.
Slice Number | CT | 3D Ultrasound | Gross Pathology |
---|---|---|---|
1 | 22.0 ± 6.3 | 19.6 ± 8.8 | 29.3 |
2 | 39.1 ± 5.1 | 38.7 ± 11.4 | 51.5 |
3 | 53.8 ± 2.8 | 56.9 ± 5.3 | 58.8 |
4 | 51.6 ± 10.2 | 64.4 ± 5.6 | 67.1 |
5 | 65.4 ± 2.3 | 62.2 ± 6.0 | 63.6 |
6 | 69.1 ± 7.0 | 65.9 ± 1.7 | 66.9 |
7 | 67.4 ± 14.2 | 64.5 ± 2.4 | 64.9 |
8 | 54.6 ± 17.7 | 64.6 ± 4.6 | 58.1 |
9 | 45.6 ± 3.0 | 49.6 ± 13.4 | 55.7 |
References
1. Heidkamp, J.; Weijs, W.L.J.; Grunsven, A.C.H.V.E.; Vries, I.D.L.; Maas, M.C.; Rovers, M.M.; Fütterer, J.J.; Steens, S.C.A.; Takes, R.P. Assessment of surgical tumor-free resection margins in fresh squamous-cell carcinoma resection specimens of the tongue using a clinical MRI system. Head Neck; 2020; 42, pp. 2039-2049. [DOI: https://dx.doi.org/10.1002/hed.26125] [PubMed: https://www.ncbi.nlm.nih.gov/pubmed/32119170]
2. Steens, S.C.A.; Bekers, E.M.; Weijs, W.L.J.; Litjens, G.J.S.; Veltien, A.; Maat, A.; Broek, G.B.V.D.; van der Laak, J.A.W.M.; Fütterer, J.J.; van der Kaa, C.A.H. et al. Evaluation of tongue squamous cell carcinoma resection margins using ex-vivo MR. Int. J. Comput. Assist. Radiol. Surg.; 2017; 12, pp. 821-828. [DOI: https://dx.doi.org/10.1007/s11548-017-1524-6] [PubMed: https://www.ncbi.nlm.nih.gov/pubmed/28130702]
3. Rogers, S.; Carreira, J.; Thompson, R.; Morais, A.; Miller, C.; Wein, W.; Ghosh, J.; McCollum, C. An Ex Vivo Evaluation of Tomographic 3-D Ultrasound, B-Mode Ultrasound, CT And MR Imaging to Measure Artery Diameter, Length and Wall Volume. Ultrasound Med. Biol.; 2019; 45, pp. 2819-2829. [DOI: https://dx.doi.org/10.1016/j.ultrasmedbio.2019.07.002] [PubMed: https://www.ncbi.nlm.nih.gov/pubmed/31375217]
4. Stadler, A.; Schima, W.; Prager, G.; Homolka, P.; Heinz, G.; Saini, S.; Eisenhuber, E.; Niederle, B. CT Density Measurements for Characterization of Adrenal Tumors Ex Vivo: Variability Among Three CT Scanners. Am. J. Roentgenol.; 2004; 182, pp. 671-675. [DOI: https://dx.doi.org/10.2214/ajr.182.3.1820671] [PubMed: https://www.ncbi.nlm.nih.gov/pubmed/14975968]
5. de Koning, K.J.; Koppes, S.A.; de Bree, R.; Dankbaar, J.W.; Willems, S.M.; van Es, R.J.; Noorlag, R. Feasibility study of ultrasound-guided resection of tongue cancer with immediate specimen examination to improve margin control—Comparison with conventional treatment. Oral Oncol.; 2021; 116, 105249. [DOI: https://dx.doi.org/10.1016/j.oraloncology.2021.105249] [PubMed: https://www.ncbi.nlm.nih.gov/pubmed/33774501]
6. Tang, R.; Buckley, J.M.; Fernandez, L.; Coopey, S.; Aftreth, O.; Michaelson, J.; Saksena, M.; Lei, L.; Specht, M.; Gadd, M. et al. Micro-computed tomography (Micro-CT): A novel approach for intraoperative breast cancer specimen imaging. Breast Cancer Res. Treat.; 2013; 139, pp. 311-316. [DOI: https://dx.doi.org/10.1007/s10549-013-2554-6]
7. Nakamura, S.; Mori, K.; Iwano, S.; Kawaguchi, K.; Fukui, T.; Hakiri, S.; Ozeki, N.; Oda, M.; Yokoi, K. Micro-Computed tomography images of lung adenocarcinoma: Detection of lepidic growth patterns. Nagoya J. Med. Sci.; 2020; 82, pp. 25-31. [DOI: https://dx.doi.org/10.18999/nagjms.82.1.25] [PubMed: https://www.ncbi.nlm.nih.gov/pubmed/32273629]
8. Neubauer, C.; Yilmaz, J.S.; Bronsert, P.; Pichotka, M.; Bamberg, F.; Windfuhr-Blum, M.; Erbes, T.; Neubauer, J. Accuracy of cone-beam computed tomography, digital mammography and digital breast tomosynthesis for microcalcifications and margins to microcalcifications in breast specimens. Sci. Rep.; 2022; 12, 17639. [DOI: https://dx.doi.org/10.1038/s41598-022-21616-3]
9. de Koning, K.J.; van Es, R.J.; Klijn, R.J.; Breimer, G.E.; Dankbaar, J.W.; Braunius, W.W.; van Cann, E.M.; Dieleman, F.J.; Rijken, J.A.; Tijink, B.M. et al. Application and accuracy of ultrasound-guided resections of tongue cancer. Oral Oncol.; 2022; 133, 106023. [DOI: https://dx.doi.org/10.1016/j.oraloncology.2022.106023]
10. de Koning, S.B.; Karakullukcu, M.; Lange, C.; Schreuder, W.; Karssemakers, L.; Ruers, T. Ultrasound aids in intraoperative assessment of deep resection margins of squamous cell carcinoma of the tongue. Br. J. Oral Maxillofac. Surg.; 2020; 58, pp. 285-290. [DOI: https://dx.doi.org/10.1016/j.bjoms.2019.11.013] [PubMed: https://www.ncbi.nlm.nih.gov/pubmed/32044145]
11. Nulent, T.J.K.; Noorlag, R.; Van Cann, E.M.; Pameijer, F.A.; Willems, S.M.; Yesuratnam, A.; Rosenberg, A.J.; de Bree, R.; van Es, R.J. Intraoral ultrasonography to measure tumor thickness of oral cancer: A systematic review and meta-analysis. Oral Oncol.; 2018; 77, pp. 29-36. [DOI: https://dx.doi.org/10.1016/j.oraloncology.2017.12.007] [PubMed: https://www.ncbi.nlm.nih.gov/pubmed/29362123]
12. Pradipta, A.R.; Tanei, T.; Morimoto, K.; Shimazu, K.; Noguchi, S.; Tanaka, K. Emerging Technologies for Real-Time Intraoperative Margin Assessment in Future Breast-Conserving Surgery. Adv. Sci.; 2020; 7, 1901519. [DOI: https://dx.doi.org/10.1002/advs.201901519] [PubMed: https://www.ncbi.nlm.nih.gov/pubmed/32382473]
13. Huang, Q.; Zeng, Z. A Review on Real-Time 3D Ultrasound Imaging Technology. BioMed Res. Int.; 2017; 2017, 6027029. [DOI: https://dx.doi.org/10.1155/2017/6027029] [PubMed: https://www.ncbi.nlm.nih.gov/pubmed/28459067]
14. Kim, S.C.; Kim, J.H.; Choi, S.H.; Yun, T.J.; Wi, J.Y.; Kim, S.A.; Sun, H.Y.; Ryoo, I.; Park, S.-W.; Sohn, C.-H. Off-Site evaluation of three-dimensional ultrasound for the diagnosis of thyroid nodules: Comparison with two-dimensional ultrasound. Eur. Radiol.; 2016; 26, pp. 3353-3360. [DOI: https://dx.doi.org/10.1007/s00330-015-4193-2] [PubMed: https://www.ncbi.nlm.nih.gov/pubmed/26795614]
15. Chung, D.; Bandarkar, A.; Rana, S.; Tabrizi, P.R.; Preciado, D.; Jago, J.; Linguraru, M.G.; Reilly, B.K. Pilot study of the potential of 3D ultrasound to measure tonsillar volume and hypertrophy. Int. J. Pediatr. Otorhinolaryngol.; 2019; 126, 109612. [DOI: https://dx.doi.org/10.1016/j.ijporl.2019.109612] [PubMed: https://www.ncbi.nlm.nih.gov/pubmed/31408743]
16. Malago, R.; D’Onofrio, M.; Ferdeghini, M.; Mantovani, W.; Colato, C.; Brazzarola, P.; Motton, M.; Mucelli, R.P. Thyroid volumetric quantification: Comparative evaluation between conventional and volumetric ultrasonography. J. Ultrasound Med.; 2008; 27, pp. 1727-1733. [DOI: https://dx.doi.org/10.7863/jum.2008.27.12.1727] [PubMed: https://www.ncbi.nlm.nih.gov/pubmed/19022998]
17. Inoue, T.; Tamaki, Y.; Sato, Y.; Nakamoto, M.; Tamura, S.; Tanji, Y.; Taguchi, T.; Noguchi, S. Three-Dimensional ultrasound imaging of breast cancer by a real-time intraoperative navigation system. Breast Cancer; 2005; 12, pp. 122-129. [DOI: https://dx.doi.org/10.2325/jbcs.12.122] [PubMed: https://www.ncbi.nlm.nih.gov/pubmed/15858443]
18. Yushkevich, P.A.; Piven, J.; Hazlett, H.C.; Smith, R.G.; Ho, S.; Gee, J.C.; Gerig, G. User-Guided 3D active contour segmentation of anatomical structures: Significantly improved efficiency and reliability. Neuroimage; 2006; 31, pp. 1116-1128. [DOI: https://dx.doi.org/10.1016/j.neuroimage.2006.01.015] [PubMed: https://www.ncbi.nlm.nih.gov/pubmed/16545965]
19. Henderson, A. ParaView Guide, A Parallel Visualization Application; Kitware Inc.: New York, NY, USA, 2007.
20.
You have requested "on-the-fly" machine translation of selected content from our databases. This functionality is provided solely for your convenience and is in no way intended to replace human translation. Show full disclaimer
Neither ProQuest nor its licensors make any representations or warranties with respect to the translations. The translations are automatically generated "AS IS" and "AS AVAILABLE" and are not retained in our systems. PROQUEST AND ITS LICENSORS SPECIFICALLY DISCLAIM ANY AND ALL EXPRESS OR IMPLIED WARRANTIES, INCLUDING WITHOUT LIMITATION, ANY WARRANTIES FOR AVAILABILITY, ACCURACY, TIMELINESS, COMPLETENESS, NON-INFRINGMENT, MERCHANTABILITY OR FITNESS FOR A PARTICULAR PURPOSE. Your use of the translations is subject to all use restrictions contained in your Electronic Products License Agreement and by using the translation functionality you agree to forgo any and all claims against ProQuest or its licensors for your use of the translation functionality and any output derived there from. Hide full disclaimer
© 2022 by the authors. Licensee MDPI, Basel, Switzerland. This article is an open access article distributed under the terms and conditions of the Creative Commons Attribution (CC BY) license (https://creativecommons.org/licenses/by/4.0/). Notwithstanding the ProQuest Terms and Conditions, you may use this content in accordance with the terms of the License.
Abstract
The margin of the removed tumor in cancer surgery has an important influence on survival. Adjuvant treatments, prognostic complications, and financial costs are required when the pathologist observes a close/positive surgical margin. Ex vivo imaging of resected cancer tissue has been suggested for margin assessment, but traditional cross-sectional imaging is not optimal in a surgical setting. Instead, three-dimensional (3D) ultrasound is a portable, high-resolution, and low-cost method to use in the operation room. In this study, we aimed to investigate the accuracy of 3D ultrasound versus computed tomography (CT) to measure the tumor volume in an animal model compared to gross pathology assessment. The specimen was formalin fixated before systematic slicing. A slice-by-slice area measurement was performed to compare the accuracy of the 3D ultrasound and CT techniques. The tumor volume measured by pathological assessment was 980.2 mm3. The measured volume using CT was 890.4 ± 90 mm3, and the volume using 3D ultrasound was 924.2 ± 96 mm3. The correlation coefficient for CT was 0.91 and that for 3D ultrasound was 0.96. Three-dimensional ultrasound is a feasible and accurate modality to measure the tumor volume in an animal model. The accuracy of tumor delineation on CT depends on the soft tissue contrast.
You have requested "on-the-fly" machine translation of selected content from our databases. This functionality is provided solely for your convenience and is in no way intended to replace human translation. Show full disclaimer
Neither ProQuest nor its licensors make any representations or warranties with respect to the translations. The translations are automatically generated "AS IS" and "AS AVAILABLE" and are not retained in our systems. PROQUEST AND ITS LICENSORS SPECIFICALLY DISCLAIM ANY AND ALL EXPRESS OR IMPLIED WARRANTIES, INCLUDING WITHOUT LIMITATION, ANY WARRANTIES FOR AVAILABILITY, ACCURACY, TIMELINESS, COMPLETENESS, NON-INFRINGMENT, MERCHANTABILITY OR FITNESS FOR A PARTICULAR PURPOSE. Your use of the translations is subject to all use restrictions contained in your Electronic Products License Agreement and by using the translation functionality you agree to forgo any and all claims against ProQuest or its licensors for your use of the translation functionality and any output derived there from. Hide full disclaimer
Details

1 Department of Otorhinolaryngology, Head and Neck Surgery and Audiology, Rigshospitalet, Copenhagen University Hospital, DK-2100 Copenhagen, Denmark
2 Department of Radiology, Rigshospitalet, Copenhagen University Hospital, DK-2100 Copenhagen, Denmark; Institute of Clinical Medicine, Faculty of Health and Medical Sciences, University of Copenhagen, DK-2200 Copenhagen, Denmark
3 Institute of Clinical Medicine, Faculty of Health and Medical Sciences, University of Copenhagen, DK-2200 Copenhagen, Denmark; Department of Pathology, Rigshospitalet, Copenhagen University Hospital, DK-2100 Copenhagen, Denmark
4 Centre for Neuroscience and Stereology, Department of Neurology, Bispebjerg-Frederiksberg Hospital, DK-2400 Copenhagen, Denmark
5 Institute of Clinical Medicine, Faculty of Health and Medical Sciences, University of Copenhagen, DK-2200 Copenhagen, Denmark; Centre for Neuroscience and Stereology, Department of Neurology, Bispebjerg-Frederiksberg Hospital, DK-2400 Copenhagen, Denmark
6 Department of Otorhinolaryngology, Head and Neck Surgery and Audiology, Rigshospitalet, Copenhagen University Hospital, DK-2100 Copenhagen, Denmark; Institute of Clinical Medicine, Faculty of Health and Medical Sciences, University of Copenhagen, DK-2200 Copenhagen, Denmark; Copenhagen Academy for Medical Education and Simulation, DK-2100 Copenhagen, Denmark