1. Introduction
A magnet is a material that can produce a magnetic field based on the orientation of its atomic magnetic dipole moments. This magnetic field attracts unlike poles while repelling like poles. The resulting electromagnetic field and attractive force inherent in magnetic materials are useful properties in the development of novel surgical devices. Ever since the use of electromagnetic bougienage by Drs. Hendren and Hale for the lengthening of disconnected esophageal pouches in children with esophageal atresia, magnetic devices have been explored across the breadth of modern surgical procedures in almost every organ system (Table 1) [1].
As minimally invasive surgical (MIS) techniques, including endoscopy, laparoscopy, and robotic surgery, have allowed surgical procedures to be performed via smaller incisions (or no incisions) with improved patient outcomes (e.g., decreased hospital length of stay, faster return to activity), magnetic devices have been explored to augment existing MIS techniques and instruments [2]. With this aim, multidisciplinary and international collaboration have led to the rise in magnetic surgical device development from preclinical animal studies to the first in-human trials studying safety and feasibility [3]. This article serves to provide an overview of key biomedical considerations during device development and discuss the core functions of existing magnetic surgical devices.
Table 1Historical in-human application of magnets in surgery.
Year | Organ System | Medical |
Use of Magnets | Citation |
---|---|---|---|---|
1975 | Esophagus | Esophageal atresia | Intermittent application of electromagnetic field to metal bougies that lengthen atretic esophageal pouches | Hendren and Hale, 1975 [1] |
1980 | Vascular | Need for |
Intravascular paired ring magnets for side-to-side anastomosis between two blood vessels | Obora et al., 1980 [4] |
1981 | Colorectal | Presence of |
Internal–external paired magnetic ring system for colostomy closure | Jansen et al., 1981 [5] |
1989 | Urology | Urethral |
Intraurethral miniaturized magnets for stricturoplasty | Isakov et al., 1989 [6] |
1992 | Small intestine | Small bowel fistula | Intraluminal paired magnets to form compression anastomosis bypassing intestinal fistula | Stepanov et al., 1992 [7] |
1993 | Hepatobiliary | Biliary |
Paired magnets to form compression anastomosis between proximal common bile duct and stomach or duodenum | Saveliev et al., 1993 [8] |
2. Biomedical Considerations for Magnetic Surgical Devices
The utility of a magnetic material in surgical device development is dependent on its strength, durability, and mitigation of potential toxicity in the human body. The strength of a material is based on the elements that comprise it and is determined by the product of a material’s strength as well as the force required to demagnetize it (i.e., coercivity). This is known as the energy product (Gauss Oersted or Joules/m3) which is graded on the N grading system—increasing strength correlating with a higher grade [9]. Currently, rare-earth elements, such as N52-grade neodymium-iron-boron (Nd-Fe-B), are commonly used for industrial magnetic devices as they have the highest recorded maximum energy product (474 kJ/m3) [10]. This corresponds to a magnetic field of approximately 10,000 Gauss, which is 100 times stronger than a household refrigerator magnet at 100 Gauss [11]. One limitation of Nd-Fe-B magnets for industrial use is their relatively low Curie temperature, or the temperature at which a material loses its dipole alignment and subsequently its magnetism; however, the Curie temperature for Nd-Fe-B at 310 °C is well above environmental temperatures in medical use [12].
Another potential limitation is the relatively poor durability and potential toxicity of Nd-Fe-B magnets to human tissues. Nd and Fe are known to oxidize rapidly in air, leading not only to the corrosion and weakening of the magnet but also to the formation of reactive oxygen species (ROS) that have been shown to be cytotoxic to human cells [13,14]. Additionally, Nd is a brittle material that can break easily [15]. These limitations are mitigated by hermetically sealing the Nd-Fe-B magnetic core in a durable biocompatible coating. Considerations when selecting a coating include adhesion of the coating to the magnet, avoidance of magnet oxidation during the coating process, and the coating’s durability, biocompatibility, and cost [16]. Examples of biocompatible medical-grade coating materials include titanium, parylene, and gold [17,18]. During device development, the stringent testing of physical and chemical properties is regulated by the United States Food and Drug Administration (FDA) to ensure the mitigation of these risks prior to clinical implementation. Key considerations during testing include the following: type of device, duration of contact with the human body, and nature of contact with the human body (Table 2).
In addition to potential toxicity from the physical properties of Nd-Fe-B, one should consider the potential risks from exposure to a static magnetic field. Medical implants such as cardiac pacemakers or implantable cardiac defibrillators (ICD) may be at risk of interference from magnetic fields. Although the operation of implanted devices has not been shown to be adversely affected by static magnetic fields below 0.5 mT and has been proven to withstand magnetic resonance imaging (MRI), the proximity to medical implants—along with the size and strength of a magnetic surgical device—warrants consideration [19,20,21]. Theoretical carcinogenic risks from exposure to a long-term magnetic implant remain unproven given the current data on long-term MRI exposure [22].
In addition to the strength, durability, and potential toxicity of a magnetic device, a device’s intended use and location in the human body, shape, thickness, and cost are all key factors to consider during device development. More specifically, the type of tissue that the device will be in contact with (e.g., intestinal tract, biliary tract, blood vessel wall, or bone) and the method of device placement/retrieval must be considered. For example, magnetic devices that are deployed intraluminally in the foregut may be more susceptible to degradation due to their increased exposure to gastric acid and digestive enzymes, relative to devices that are implanted in bone or deployed within blood vessels. Thus, these devices may warrant additional layers of hermetic sealing and more stringent durability testing. Additionally, devices that are placed within blood vessels must consider the size, shape, and coating to facilitate feasible placement into the blood vessel—as well as the prevention of device-associated thrombus formation [23]. Thus, the intended tissue type and device environment strongly influence device design. This highlights the importance of early collaboration between clinicians and bioengineers.
The risks of magnetic surgical devices during use must also be considered during device development. Although rare, potential off-target events associated with an in situ magnetic device can lead to catastrophic morbidity and even death. This is evident in reported cases of ingested rare-earth magnets that have led to bowel fistula, perforation, sepsis, and death [24]. In order to mitigate these off-target events, devices that utilize paired systems should consider the risk of injury to the intervening tissue and the use of unidirectional ferromagnetic backplates to minimize off-target effects [25]. Considering these risks, paired systems for magnetic compression anastomosis (e.g., intestinal, biliary, vascular) also require target tissues to be apposed, without any intervening tissue/fluid, to facilitate magnet mating, avoid magnet migration, and mitigate the risk of an off-target serious adverse event. Thus, while patient selection is ultimately key in device success, optimizing tissue-specific device specifications for its intended use is also paramount.
3. Core Functions of Magnetic Surgical Devices
The following sections serve to discuss the core functions of existing magnetic surgical devices on the market, in addition to those of novel devices currently under investigation. They also briefly discuss the clinical uses of the individual devices introduced, but are by no means comprehensive discussions of each device (Table 3).
3.1. Guidance System
The ability of magnets to transmit a detectable electromagnetic field is of particular interest in the development of guidance systems for the placement of devices in the human body. This is evident in current devices that aid the placement of nasoenteric feeding tubes and peripherally inserted central venous catheters (PICCs) (Table 1).
In order to provide nutrition to critically ill patients who may be incapable of eating by mouth, small-bore nasoenteric feeding tubes are frequently placed at the bedside. Conventional placement includes blind insertion, followed by radiograph confirmation prior to the use of the tube; however, this approach can place patients at risk of the tube being placed into the airways and causing pulmonary trauma (e.g., pneumothorax), as well as cause a delay in the use of the tube for hydration, nutrition, or vital medications while the radiograph is obtained and interpreted. In addition, if the radiograph is misinterpreted, tube feeding into the lungs can lead to catastrophic morbidity and even death. An alternative approach uses a stylet with an electromagnetic tip inside of the feeding tube during placement that transmits a signal to an external receiver unit placed on the patient’s epigastric region [26]. This allows for the real-time visualization of the path of the tip of the tube to avoid placement into the lung, as well as for the confirmation of the correct placement in the stomach or small intestine without requiring radiograph confirmation. This has been shown to reduce the risk of pulmonary complications, such as feeding into the lungs or pneumothorax, while decreasing the delay in tube feeding initiation [27]. Additionally, this has reduced the burden on radiology resources and has the potential to lead to a cost avoidance of USD 346,000 over a 2-year period [28].
Similar to the placement of feeding tubes, the detection of an internal magnet for real-time guidance is also being used for PICC placement. PICCs are catheters that provide central venous access for medications, parenteral nutrition, repeated blood sampling, and invasive hemodynamic monitoring. These specialized catheters are commonly placed at the bedside under blind insertion with radiographic confirmation or in the radiology suite under live fluoroscopic guidance. While the former approach introduces an increased risk of malposition or cardiovascular injury that can result in arrhythmia or cardiac tamponade, the latter introduces increased radiation exposure, higher cost, and limited availability in resource-limited settings [31]. The magnetic approach uses an inner stylet with a magnetic tip that is detected by an external sensor placed on a patient’s sternum, providing real-time tracking of the catheter’s path, which has decreased rates of malposition and radiation exposure [29,30].
The potential of intravascular magnets in guiding the formation of arteriovenous fistulas (AVFs) in patients requiring long-term hemodialysis is being explored. AVFs continue to be created via conventional open surgery connecting an extremity artery to a vein; however, specialized centers are now exploring the ability to create AVFs in the radiology suite via percutaneous means. Specialized catheters with magnetic tips are inserted into the artery and vein of interest, which guide the catheters to the area of interest and align the vessels together, allowing a radiofrequency electrode to weld the vessels together to form a fistula [32]. Although not yet widely available, this introduces a novel use of paired magnets to guide a novel minimally invasive approach to AVF creation. Thus, existing magnetic surgical devices have improved conventional approaches by minimizing radiation exposure as well as the use of radiology resources, improving procedural accuracy, and creating a novel minimally invasive approach for a common vascular operation.
3.2. Magnetic Compression Anastomosis to Create New Connections
New connections can be created between two hollow organs (e.g., the esophagus, stomach, small intestine, colon, and bile duct) to bypass an obstructed segment of the gastrointestinal (GI) tract due to atresia, stricture, or mass—or to bypass the large absorptive surface of the small intestine (e.g., bariatric bypass surgery). Currently, open or laparoscopic approaches to the creation of an anastomosis are performed with either surgical staplers or hand-sutured methods; however, the use of compression to create an equally robust anastomosis resistant to anastomotic leak would create an approach that could further minimize the invasiveness of this procedure by decreasing or eliminating the need for incisions. This approach also has the potential to create an anastomosis in patients that require one to restore GI tract continuity, but that have medical or anatomic comorbidities that may prohibit conventional surgical methods [56]. For example, in patients with an obstructing gastric cancer and a history of multiple intra-abdominal operations or radiation therapy, subsequent dense adhesions may prohibit a safe palliative gastrointestinal bypass. Thus, an endoluminal magnetic approach would allow these high-risk surgical patients to receive treatment and avoid a permanent proximal gastrointestinal diversion.
The modern use of compressive force to create an anastomosis re-emerged in the 1980s with the creation of the biofragmentable bowel anastomosis ring (BAR) and compression anastomotic clips [110,111,112,113,114]. More recently, the use of paired magnets to create magnetic compression anastomoses (MCAs) has also been shown to form robust anastomoses [53]. Magnets are placed in the organs of desired connection via endoscopic- or laparoscopic-assisted approaches. Magnets are guided to mate, and once the anastomosis is fully formed the magnets unite and are excreted or retrieved endoscopically [36,39,50].
Various paired magnet systems have been described in the literature (Table 3). Magnet configurations include spherical, circular discoid (flat or concave–convex), circular-ring-shaped, with varying inner diameter sizes, cylindrical, and bullet-shaped [50]. The magnetic cores of Nd-Fe-B or samarium-cobalt are commonly described, with coating methods varying between being single-layer (e.g., titanium oxide, polycarbonate) and multi-layer (e.g., inner nickel-copper-nickel coating, outer gold-parylene C coating for biocompatibility) [53]. Additional considerations during the development of MCA devices include how to best place the magnets into organs of interest (e.g., endoscopy, laparoscopy, and percutaneous) and the effects of magnet size/shape on magnet delivery as well as apposition, mating force, anastomosis formation, and the risk of postoperative complications (e.g., anastomotic leak, stricture). Prior to in-human use, extensive preclinical testing on device placement, biocompatibility, compressive force, and burst pressure performance is often performed [34,36,37,38,39,40,43,115]. Burst pressure is used to test anastomotic resistance to leakage with maximal observed burst pressures in MCAs consistently being > 100 mmHg, which is above the physiologic intraluminal pressure [53]. Continued research aims to optimize magnet delivery via completely endoscopic means (i.e., incisionless surgery) (Figure 1) while improving anastomotic outcomes by reducing stricture rates (Figure 2) [45,55,56].
The miniaturization of MCA device systems has also broadened their applicability to treat many pediatric conditions. A key example of this is in infants with esophageal atresia (EA)—a congenital malformation resulting in two separate non-communicating pouches that require a surgical anastomosis in order to achieve esophageal continuity and allow for nutrition via the mouth [116]; however, the current surgical options are sometimes associated with significant morbidity (e.g., musculoskeletal deformity, anastomotic leak or stricture, or vocal cord paresis)—the rates for which have remained relatively constant over the past 80 years [117]. Thus, the MCA devices currently used for EA repair utilize paired magnets that are placed into apposing pouches—one via the mouth and one via the stomach—and mated to create a completely endoscopic anastomosis. While the widespread adoption of this technique has been limited by early experience, observing the prohibitive rates of anastomotic strictures [54], newer devices with wider mating surfaces and unique mating geometries aim to solve this problem [53,55,56,118] (Figure 2).
Examples of in-human use of MCA include the following:
Esophago-esophageal anastomoses of two disconnected esophageal ends in children with esophageal atresia [49,51,52,54,55,56];
Gastroenteric or intestinal anastomoses to restore gastrointestinal (GI) tract continuity [35,42];
Uretero-ileal anastomoses for urinary diversions in patients with neurogenic bladders [42];
Colorectal anastomoses of disconnected colons and rectums in children with rectal atresia [41];
Intestinal anastomoses for enteral bypasses/diversions with comparable weight loss and decrease in hemoglobin A1c, fasting blood glucose, and use of medications/insulin for diabetes [44,45];
Stricturoplasties, or intraluminal resections of obstructing strictures, to restore GI tract continuity;
Small bowel non-anastomotic strictures [63];
Anastomotic strictures after esophagectomy for esophageal cancer [64], esophageal atresia repair [61,62], and colon resections for colorectal cancer [66];
Bilioenteric (the bile duct to the small intestine) anastomotic ischemic strictures after liver transplant or a bile duct injury [57,58,59,60].
Cardiac and vascular surgeons are also utilizing paired magnet systems to create new connections between two arteries. Currently, the majority of vascular anastomoses are hand-sewn; however, in contrast to MCAs, magnetic vascular ports (MVPs) are permanent devices that are deployed within arteriotomies to form a vascular port opening on each vessel of interest. Ports are then mated to form anastomoses [69]. This method has been utilized most frequently for the minimally invasive direct coronary artery bypass (MIDCAB) procedure to connect the left internal mammary artery (LIMA) on the chest wall to the left anterior descending (LAD) artery on the heart [69]. In-human trials have demonstrated decreased anastomotic and total procedure times, with no device-related adverse events and favorable patency rates at 6-month follow-ups [68,70,72,73,74,76,119]. Preclinical animal studies are also being used to explore the use of MCAs in the creation of vascular anastomoses, but these have not progressed to in-human trials at this point [120,121,122,123].
3.3. Recreating a Physiologic Function
The dysfunction of muscular sphincter complexes can lead to gastroesophageal reflux disease (GERD; lower esophageal sphincter) or incontinence (fecal; anal sphincter complex). Anti-reflux procedures for the treatment of medically refractory GERD, such as a Nissen fundoplication, can lead to abdominal bloating, inability to vomit, and persistent dysphagia [124]. Additionally, there are multiple surgical options for fecal incontinence, such as sacral nerve stimulation and artificial sphincter balloon implants, but the adoption of one has been limited by suboptimal outcomes [125]. Magnetic sphincter augmentation has the potential to recreate physiologic sphincter function. Current magnetic devices consist of small titanium-coated beads with Nd-Fe-B magnetic cores interlinked with titanium wires to form a flexible expandable ring (Table 3) [87]. This configuration allows for opening with increased dynamic pressure, but maintains a closed position at lower passive pressures, preventing reflux or incontinence. Devices are sized to the external diameter of the esophagus (e.g., GERD) or the anal canal (e.g., fecal incontinence) without compressing the underlying muscle [79,87]. Magnetic sphincter devices for GERD have demonstrated safety and the normalization of esophageal acid exposure (i.e., reflux), reduced or discontinued reflux medication use (i.e., proton pump inhibitors), and improved quality of life [78,80,81,83,84,85]. Magnetic devices for fecal incontinence have also shown improvements in incontinence severity and quality of life, with comparable outcomes to sacral nerve stimulation, but their adoption has been limited due to reports of device-related infection and perineal pain [88,89,126,127,128,129]. Magnetic sphincter augmentation is also being explored for urinary incontinence in the preclinical phase [90]. Thus, magnetically augmented sphincters are currently being used to recreate physiologic sphincter function, while also having the potential to be reversible (with device explantation).
3.4. Use of Internal–External Paired Magnet Systems
Systems that use an internal magnet guided by or acted upon by an external magnet have a variety of surgical applications, ranging from the surgical treatment of congenital disorders (e.g., scoliosis, pectus excavatum) to improving organ retraction during laparoscopic surgery (Table 3). The surgical management of early onset scoliosis employs dynamic instrumentation to allow for continued longitudinal growth. Traditional growing rods require repeated surgeries to lengthen the rods as a child grows. Magnetic growing rod (MGR) systems use a similar system with single or dual implanted titanium spinal rods fixed to the spine cranially and caudally [130]; however, as a child grows, an external magnetic device can be used to rotate the rod’s internal actuator, which lengthens the rod without the need for repeated surgeries [103]. In addition to a reduction in operations, the MGR system has also led to subsequent long-term cost savings, decreased infection rates, and similar rates of implant failure (e.g., rod or foundation failure) [104,106,107,108,109,131].
The attractive magnetic force of an internal–external paired magnet system is also being employed to gradually remodel musculoskeletal deformities. Pectus excavatum is a congenital deformation of the cartilages that connect the ribs to the sternum that pushes the sternum inward and can cause the compression of the heart. Conventional repairs involve major surgical reconstruction. The modified Ravitch procedure requires the open removal of the abnormal cartilages, the fracturing of the sternum, and fixation into a satisfactory alignment. The less invasive Nuss procedure uses smaller incisions to place a titanium metal bar within the chest and behind the sternum to gradually remodel the chest wall over a 2-year period; however, both procedures are associated with a serious risk of injury to the heart, lungs, or major blood vessels, as well as significant postoperative pain. The Nuss procedure also requires a second procedure to remove the titanium bar after te remodeling of the chest wall is complete. The magnetic approach implants an internal Nd-Fe-B magnet with a ferromagnetic focusing plate encapsulated in a low-profile titanium shell to the anterior sternum [25,99]. The internal magnet is mated with a second magnet housed in an external polypropylene custom-fitted anterior chest wall brace [25]. This approach has been shown to decrease surgical risk while improving chest wall deformation and decreasing overall cost [100,101]; however, this technology has seen limited adoption due to major improvements in postoperative pain control with the advent of intercostal nerve cryoablation [132]. A similar use of an internal–external paired magnet system is also being studied in obstructive sleep apnea treatment—as an alternative to continuous positive airway pressure (CPAP)—by advancing the hyoid bone forward to maintain airway patency during sleep (Figure 3) [102,133]. Other surgical applications of an internal–external paired magnetic system include the use of internal detachable magnetic graspers manipulated by external magnets to aid in organ retraction during laparoscopic gallbladder surgery and bariatric surgery to decrease the number of incisions (Table 3) [91,92,93,95,96,134]. These multiple uses highlight the potential and versatility of magnetic surgical systems that can manipulate internal tissues via an external source.
4. Conclusions
The use of magnetic devices in surgery is an actively growing field with various systems on the market or in preclinical development. Multidisciplinary collaboration during device development is recommended due to the combination of biomedical engineering and surgical considerations. Existing magnetic surgical devices can be used to guide the placement of catheters and tubes, create new connections in the GI tract as well as blood vessels, recreate physiologic sphincter function, and manipulate internal tissues via an external magnet. This versatility highlights the potential of magnetic devices to improve existing surgical techniques and minimally invasive approaches.
Conceptualization, methodology, formal analysis, investigation, resources, data curation, writing (original draft preparation, review and editing), supervision, and project administration—W.G.L., L.L.E., S.M.J. and R.K.W. All authors have read and agreed to the published version of the manuscript.
Not applicable.
Not applicable.
No new data were created or analyzed in this study. Data sharing is not applicable to this article.
The authors report no conflicts of interest.
Footnotes
Disclaimer/Publisher’s Note: The statements, opinions and data contained in all publications are solely those of the individual author(s) and contributor(s) and not of MDPI and/or the editor(s). MDPI and/or the editor(s) disclaim responsibility for any injury to people or property resulting from any ideas, methods, instructions or products referred to in the content.
Figure 1. GI Windows SurgicalTM self-forming magnetic rings that can be delivered through an endoscope into the target segments of the bowel. Magnetics rings auto-align and mate—leading to the formation of a magnetic compression anastomosis. (Permission for use granted by GI Windows Surgical, West Bridgewater, MA, USA.)
Figure 2. Connect-EATM device composed of two N52-grade magnets with a unique concave–convex surface that produces a pressure gradient in the intervening tissue compressed between the magnets. The maximal pressure in the center creates tissue necrosis, while allowing for tissue healing and epithelialization at the periphery, which forms a robust anastomosis with decreased stricture formation [53]. (Permission for use granted by Myka Labs, San Francisco, CA, USA.)
Figure 3. The magnetic apnea prevention (Mag-Nap) system utilizes an N52-grade neodymium-iron-boron magnet encased in titanium with a ferromagnetic directional back plate that is implanted onto the hyoid bone (left). The custom-fitted removable external neck accessory (right) to be worn during sleep contains a second magnet that attracts the implanted hyoid magnet forward to keep the airway open [102]. (Permission for use granted by Mag-Nap Inc., San Francisco, CA, USA).
FDA framework for the evaluation of biocompatibility in device development.
Contact |
Type of Device | Nature of Body Contact | Recommended Endpoint Testing |
---|---|---|---|
Limited contact (≤24 h) | Surface device | Intact skin | C, S, and I |
Mucosal membrane | C, S, and I | ||
Breached or compromised surface | C, S, I, Sy, and P | ||
External |
Blood path, indirect | C, S, I, Sy, P, and H | |
Tissue/bone/dentin | C, S, I, Sy, and P | ||
Circulating blood | C, S, I, Sy, P, G, and H | ||
Implant device | Tissue/bone | C, S, I, Sy, and P | |
Blood | C, S, I, Sy, P, G, Im, and H | ||
Prolonged |
Surface device | Intact skin | C, S, and I |
Mucosal membrane | C, S, I, Sy, P, Sub, and Im | ||
Breached or compromised surface | C, S, I, Sy, P, Sub, and Im | ||
External |
Blood path, indirect | C, S, I, Sy, P, Sub, and H | |
Tissue/bone/dentin | C, S, I, Sy, P, Sub, G, and Im | ||
Circulating blood | C, S, I, Sy, P, Sub, G, Im, and H | ||
Implant device | Tissue/bone | C, S, I, Sy, P, Sub, G, and Im | |
Blood | C, S, I, Sy, P, Sub, G, Im, and H | ||
Long-term/ |
Surface device | Intact skin | C, S, and I |
Mucosal membrane | C, S, I, Sy, P, Sub, G, Im, and Ct | ||
Breached or compromised surface | C, S, I, Sy, P, Sub, G, Im, Ct, and Car | ||
External |
Blood path, indirect | C, S, I, Sy, P, Sub, G, Im, Ct, and Car | |
Tissue/bone/dentin | C, S, I, Sy, P, Sub, G, Im, Ct, Car, and H | ||
Circulating blood | C, S, I, Sy, P, Sub, G, Im, Ct, and Car | ||
Implant device | Tissue/bone | C, S, I, Sy, P, Sub, G, Im, Ct, and Car | |
Blood | C, S, I, Sy, P, Sub, G, Im, Ct, Car, and H |
C (cytotoxicity), S (sensitization), I (irritation or intracutaneous reactivity), Sy (acute systemic toxicity), P (material-mediated pyrogenicity), H (hemocompatibility), G (genotoxicity), Im (implantation), Sub (subacute/subchronic toxicity), Ct (chronic toxicity), and Car (carcinogenicity).
Published surgical applications of magnetic devices.
Core Function | Type of Magnet |
Use of Magnet System | Example of Devices Approved for Humanitarian or Commercial Use | Published |
---|---|---|---|---|
GS | Electromagnetic tube/catheter tip and |
Real-time transmission of nasoenteric tube location during placement | CORTRAK Enteral Access System (Avanos Medical, Inc., Alpharetta, Georgia, USA) | Mathus-Vliegen 2010 [ |
Real-time transmission of peripherally inserted |
Sherlock 3CG Tip Confirmation |
Tomaszewski 2017 [ |
||
Paired intravascular |
Mating of catheter tips aligns as well as holds an artery and vein together for percutaneous arteriovenous fistula creation | everlinQ endoAVF System |
Lok 2017 [ |
|
NC | Magnetic compression anastomosis between two paired intraluminal magnets (spherical, discoid, ring, and |
Connecting two small |
Xu 2015 [ |
|
Connecting two enteric segments (e.g., stomach, small intestine, and colon) | Magnamosis Magnetic Compression Anastomosis Device (Myka Labs, UCSF Surgical Innovations, San Francisco, CA, USA) | Cope 1995 [ |
||
Connecting the proximal intestine to the distal intestine to create a bypass channel |
Self-Forming Magnetic Anastomosis Device (GI Windows Surgical, West Bridgewater, MA, USA) | Ryou 2016 [ |
||
Connecting the proximal and distal esophageal pouches in esophageal atresia |
Magnamosis Connect-EA (Myka Labs, UCSF Surgical Innovations, San Francisco, CA, USA); |
Zaritzky 2009 [ |
||
Connecting the bile duct to the stomach or small intestine to bypass bile duct stricture | Mimuro 2003 [ |
|||
Resecting a strictured esophagus, small intestine, bile duct, or colon to allow luminal contents to pass through | Takamizawa 2007 [ |
|||
Paired intravascular |
Creation of anastomosis |
Magnetic Vascular Positioner (MVP) Series 6000 Distal Anastomosis System (Ventrica, Inc., Fremont, CA, USA) | Falk 2003 [ |
|
PHYS | Magnetic beads |
Placed around the distal esophagus to recreate a physiologic lower |
LINX (Torax Medical, Inc., Shoreview, MN, USA) | Lipham 2012 [ |
Placed around the external anal sphincter to recreate physiologic sphincter function in fecal incontinence | FENIX Continence Restoration System (Torax Medical, Inc., Shoreview, MN, USA) | Bortolotti 2008 [ |
||
Paired flat magnets | Placed anterior and posterior to the urethra to recreate urethral resistance in |
Ali-El-Dein 2000 [ |
||
IE | Detachable internal |
Retraction of the gallbladder during cholecystectomy | Levita Magnetic Surgical System (Levita Magnetics, Inc., Menlo Park, CA, USA) | Dominguez 2009 [ |
Retraction of the liver, |
Morales-Conde 2013 [ |
|||
Internal ureteral stent with a distal magnetic tip | Bedside removal of a stent by mating an internal stent tip with an introduced urethral magnetic catheter retrieval device | Magnetic Blackstar |
Rassweiler 2017 [ |
|
Implanted magnet and custom-fitted external brace with a paired magnet | Gradual chest wall remodeling in the pectus excavatum (congenital disorder) | Magnetic Mini-Mover (3MP): |
Harrison 2007 [ |
|
Hyoid bone advancement to maintain airway patency during sleep in obstructive sleep apnea (OSA) | Magnetic Apnea Prevention Device (Mag-Nap) (Mag-Nap, Inc., UCSF Surgical Innovations, San Francisco, CA, USA) | Rosenbluth 2011 [ |
||
Implanted distractable |
Non-invasive spinal |
MAGEC System (NuVasive, Inc., San Diego, CA, USA) | Cheung 2012 [ |
GS (guidance system), NC (creating new connection), PHYS (recreating physiologic function), IE (internal–external paired system).
References
1. Hendren, W.H.; Hale, J.R. Electromagnetic bougienage to lengthen esophageal segments in congenital esophageal atresia. N. Engl. J. Med.; 1975; 293, pp. 428-432. [DOI: https://dx.doi.org/10.1056/NEJM197508282930905] [PubMed: https://www.ncbi.nlm.nih.gov/pubmed/1152954]
2. Trondsen, E.; Reiertsen, O.; Andersen, O.K.; Kjaersgaard, P. Laparoscopic and open cholecystectomy. A prospective, randomized study. Eur. J. Surg.; 1993; 159, pp. 217-221. [PubMed: https://www.ncbi.nlm.nih.gov/pubmed/8102542]
3. Gagner, M. Magnetic Surgery; 1st ed. Springer: Cham, Switzerland, 2022; 254. [DOI: https://dx.doi.org/10.1007/978-3-030-73947-8]
4. Obora, Y.; Tamaki, N.; Matsumoto, S. Nonsuture microvascular anastomosis using magnet rings. Neurol. Med. Chir.; 1980; 20, pp. 497-505. [DOI: https://dx.doi.org/10.2176/nmc.20.497] [PubMed: https://www.ncbi.nlm.nih.gov/pubmed/6157124]
5. Jansen, A.; Brummelkamp, W.H.; Davies, G.A.; Klopper, P.J.; Keeman, J.N. Clinical applications of magnetic rings in colorectal anastomosis. Surg. Gynecol. Obstet.; 1981; 153, pp. 537-545. [PubMed: https://www.ncbi.nlm.nih.gov/pubmed/7280943]
6. Isakov, I.A.S.E.; Erokhin, A.P.; Nikolaev, V.V.; Vasi’lev, G.S. Surgical treatment of urethral strictures in children. Vestn. Khir. Im. II Grek.; 1989; 142, pp. 61-66. [PubMed: https://www.ncbi.nlm.nih.gov/pubmed/2800174]
7. Stepanov, E.A.V.e.G.; Sharipov, N.A.; Rudakov, S.S.; Romakhin, I.A. Use of permanent magnets in digestive tract surgery in children. Vestn. Akad. Med. Nauk. SSSR; 1992; 9, pp. 6-11.
8. Saveliev, V.S.; Avaliani, M.V.; Bashirov, A.D. Endoscopic magnetic cholecystodigestive anastomoses: Personal technique for palliative treatment of distal bile duct obstruction. J. Laparoendosc. Surg.; 1993; 3, pp. 99-112.
9. Robinson, A.L. Powerful New Magnet Material Found: Several groups in Japan and the United States independently hit upon the same iron-based compound, which is already going to market. Science; 1984; 223, pp. 920-922. [PubMed: https://www.ncbi.nlm.nih.gov/pubmed/17781617][DOI: https://dx.doi.org/10.1126/science.223.4639.920]
10. Horikawa, T.; Yamazaki, M.; Matsuura, M.; Sugimoto, S. Recent progress in the development of high-performance bonded magnets using rare earth-Fe compounds. Sci. Technol. Adv. Mater.; 2021; 22, pp. 729-747. [DOI: https://dx.doi.org/10.1080/14686996.2021.1944780]
11. Castle, E.; Sheridan, R.; Zhou, W.; Grasso, S.; Walton, A.; Reece, M.J. High coercivity, anisotropic, heavy rare earth-free Nd-Fe-B by Flash Spark Plasma Sintering. Sci. Rep.; 2017; 7, 11134. [DOI: https://dx.doi.org/10.1038/s41598-017-11660-9]
12. Kimura, T.; Goto, T.; Shintani, H.; Ishizaka, K.; Arima, T.; Tokura, Y. Magnetic control of ferroelectric polarization. Nature; 2003; 426, pp. 55-58. [DOI: https://dx.doi.org/10.1038/nature02018] [PubMed: https://www.ncbi.nlm.nih.gov/pubmed/14603314]
13. Rim, K.T.; Koo, K.H.; Park, J.S. Toxicological evaluations of rare earths and their health impacts to workers: A literature review. Saf. Health Work; 2013; 4, pp. 12-26. [DOI: https://dx.doi.org/10.5491/SHAW.2013.4.1.12] [PubMed: https://www.ncbi.nlm.nih.gov/pubmed/23516020]
14. Donohue, V.E.; McDonald, F.; Evans, R. In vitro cytotoxicity testing of neodymium-iron-boron magnets. J. Appl. Biomater.; 1995; 6, pp. 69-74. [PubMed: https://www.ncbi.nlm.nih.gov/pubmed/7703540][DOI: https://dx.doi.org/10.1002/jab.770060110]
15. Croat, J.J. Current status and future outlook for bonded neodymium permanent magnets (invited). J. Appl. Phys.; 1997; 81, pp. 4804-4809. [DOI: https://dx.doi.org/10.1063/1.365469]
16. Williams, D.F. On the mechanisms of biocompatibility. Biomaterials; 2008; 29, pp. 2941-2953. [DOI: https://dx.doi.org/10.1016/j.biomaterials.2008.04.023]
17. Markides, H.; Rotherham, M.; El Haj, A.J. Biocompatibility and Toxicity of Magnetic Nanoparticles in Regenerative Medicine. J. Nanomater.; 2012; 2012, 614094. [DOI: https://dx.doi.org/10.1155/2012/614094]
18. Geetha, M.; Singh, A.K.; Asokamani, R.; Gogia, A.K. Ti based biomaterials, the ultimate choice for orthopaedic implants—A review. Prog. Mater. Sci.; 2009; 54, pp. 397-425. [DOI: https://dx.doi.org/10.1016/j.pmatsci.2008.06.004]
19. Martin, E.T.; Coman, J.A.; Shellock, F.G.; Pulling, C.C.; Fair, R.; Jenkins, K. Magnetic resonance imaging and cardiac pacemaker safety at 1.5-Tesla. J. Am. Coll. Cardiol.; 2004; 43, pp. 1315-1324. [DOI: https://dx.doi.org/10.1016/j.jacc.2003.12.016]
20. International Commission on Non-Ionizing Radiation, P. Guidelines on limits of exposure to static magnetic fields. Health Phys.; 2009; 96, pp. 504-514. [DOI: https://dx.doi.org/10.1097/01.HP.0000343164.27920.4a]
21. Ryf, S.; Wolber, T.; Duru, F.; Luechinger, R. Interference of neodymium magnets with cardiac pacemakers and implantable cardioverter-defibrillators: An in vitro study. Technol. Health Care; 2008; 16, pp. 13-18. [DOI: https://dx.doi.org/10.3233/THC-2008-16102]
22. Zamanian, A.; Hardiman, C.J. Electromagnetic radiation and human health: A review of sources and effects. High Freq. Electron.; 20 July 2005; 4, pp. 16-26.
23. Timsit, J.F.; Rupp, M.; Bouza, E.; Chopra, V.; Kärpänen, T.; Laupland, K.; Lisboa, T.; Mermel, L.; Mimoz, O.; Parienti, J.J. et al. A state of the art review on optimal practices to prevent, recognize, and manage complications associated with intravascular devices in the critically ill. Intensive Care Med.; 2018; 44, pp. 742-759. [DOI: https://dx.doi.org/10.1007/s00134-018-5212-y] [PubMed: https://www.ncbi.nlm.nih.gov/pubmed/29754308]
24. Hodges, N.L.; Denny, S.A.; Smith, G.A. Rare-Earth Magnet Ingestion-Related Injuries in the Pediatric Population: A Review. Am. J. Lifestyle Med.; 2017; 11, pp. 259-263. [DOI: https://dx.doi.org/10.1177/1559827615594336] [PubMed: https://www.ncbi.nlm.nih.gov/pubmed/30202341]
25. Harrison, M.R.; Estefan-Ventura, D.; Fechter, R.; Moran, A.M., Jr.; Christensen, D. Magnetic Mini-Mover Procedure for pectus excavatum: I. Development, design, and simulations for feasibility and safety. J. Pediatr. Surg.; 2007; 42, pp. 81–85; discussion 85–86. [DOI: https://dx.doi.org/10.1016/j.jpedsurg.2006.09.042] [PubMed: https://www.ncbi.nlm.nih.gov/pubmed/17208545]
26. Mathus-Vliegen, E.M.; Duflou, A.; Spanier, M.B.; Fockens, P. Nasoenteral feeding tube placement by nurses using an electromagnetic guidance system (with video). Gastrointest. Endosc.; 2010; 71, pp. 728-736. [PubMed: https://www.ncbi.nlm.nih.gov/pubmed/20170911][DOI: https://dx.doi.org/10.1016/j.gie.2009.10.046]
27. Smithard, D.; Barrett, N.A.; Hargroves, D.; Elliot, S. Electromagnetic Sensor-Guided Enteral Access Systems: A Literature Review. Dysphagia; 2015; 30, pp. 275-285. [DOI: https://dx.doi.org/10.1007/s00455-015-9607-4]
28. McCutcheon, K.P.; Whittet, W.L.; Kirsten, J.L.; Fuchs, J.L. Feeding Tube Insertion and Placement Confirmation Using Electromagnetic Guidance: A Team Review. JPEN J. Parenter. Enter. Nutr.; 2018; 42, pp. 247-254. [DOI: https://dx.doi.org/10.1002/jpen.1015]
29. Tomaszewski, K.J.; Ferko, N.; Hollmann, S.S.; Eng, S.C.; Richard, H.M.; Rowe, L.; Sproule, S. Time and resources of peripherally inserted central catheter insertion procedures: A comparison between blind insertion/chest X-ray and a real time tip navigation and confirmation system. Clin. Outcomes Res.; 2017; 9, pp. 115-125. [DOI: https://dx.doi.org/10.2147/CEOR.S121230]
30. Mack, V.; Nißler, D.; Kasikci, D.; Malouhi, A.; Aschenbach, R.; Teichgräber, U. Magnetic Tracking and Electrocardiography-Guided Tip Confirmation System Versus Fluoroscopy for Placement of Peripherally Inserted Central Catheters: A Randomized, Noninferiority Comparison. Cardiovasc. Intervent. Radiol.; 2020; 43, pp. 1891-1897. [DOI: https://dx.doi.org/10.1007/s00270-020-02551-0]
31. Sone, M. Image Guidance for PICC Insertion: Commentary to “Magnetic Tracking and Electrocardiography-Guided Tip Confirmation System Versus Fluoroscopy for Placement of Peripherally Inserted Central Catheters: A Randomized, Noninferiority Comparison”. Cardiovasc. Intervent. Radiol.; 2020; 43, pp. 1898-1899. [DOI: https://dx.doi.org/10.1007/s00270-020-02629-9] [PubMed: https://www.ncbi.nlm.nih.gov/pubmed/32978670]
32. Lok, C.E.; Rajan, D.K.; Clement, J.; Kiaii, M.; Sidhu, R.; Thomson, K.; Buldo, G.; Dipchand, C.; Moist, L.; Sasal, J. et al. Endovascular Proximal Forearm Arteriovenous Fistula for Hemodialysis Access: Results of the Prospective, Multicenter Novel Endovascular Access Trial (NEAT). Am. J. Kidney Dis.; 2017; 70, pp. 486-497. [DOI: https://dx.doi.org/10.1053/j.ajkd.2017.03.026] [PubMed: https://www.ncbi.nlm.nih.gov/pubmed/28624422]
33. Xu, Z.F.; Li, Z.J.; Sun, Y.L.; Zheng, Y.; Shi, H.Y.; Zhen, Y.N.; Wang, B.; Niu, H.X.; Wang, R.G. A novel spherical magnetic compression device for colorectal anastomosis in a Swine model. J. Laparoendosc. Adv. Surg. Tech. A; 2015; 25, pp. 323-329. [DOI: https://dx.doi.org/10.1089/lap.2014.0404] [PubMed: https://www.ncbi.nlm.nih.gov/pubmed/25853183]
34. Cope, C. Creation of compression gastroenterostomy by means of the oral, percutaneous, or surgical introduction of magnets: Feasibility study in swine. J. Vasc. Interv. Radiol.; 1995; 6, pp. 539-545. [PubMed: https://www.ncbi.nlm.nih.gov/pubmed/7579861][DOI: https://dx.doi.org/10.1016/S1051-0443(95)71131-9]
35. Chopita, N.; Vaillaverde, A.; Cope, C.; Bernedo, A.; Martinez, H.; Landoni, N.; Jmelnitzky, A.; Burgos, H. Endoscopic gastroenteric anastomosis using magnets. Endoscopy; 2005; 37, pp. 313-317. [DOI: https://dx.doi.org/10.1055/s-2005-861358] [PubMed: https://www.ncbi.nlm.nih.gov/pubmed/15824939]
36. Jamshidi, R.; Stephenson, J.T.; Clay, J.G.; Pichakron, K.O.; Harrison, M.R. Magnamosis: Magnetic compression anastomosis with comparison to suture and staple techniques. J. Pediatr. Surg.; 2009; 44, pp. 222-228. [PubMed: https://www.ncbi.nlm.nih.gov/pubmed/19159747][DOI: https://dx.doi.org/10.1016/j.jpedsurg.2008.10.044]
37. Myers, C.; Yellen, B.; Evans, J.; DeMaria, E.; Pryor, A. Using external magnet guidance and endoscopically placed magnets to create suture-free gastro-enteral anastomoses. Surg. Endosc.; 2010; 24, pp. 1104-1109. [PubMed: https://www.ncbi.nlm.nih.gov/pubmed/20033734][DOI: https://dx.doi.org/10.1007/s00464-009-0735-5]
38. Pichakron, K.O.; Jelin, E.B.; Hirose, S.; Curran, P.F.; Jamshidi, R.; Stephenson, J.T.; Fechter, R.; Strange, M.; Harrison, M.R. Magnamosis II: Magnetic compression anastomosis for minimally invasive gastrojejunostomy and jejunojejunostomy. J. Am. Coll. Surg.; 2011; 212, pp. 42-49. [PubMed: https://www.ncbi.nlm.nih.gov/pubmed/21184956][DOI: https://dx.doi.org/10.1016/j.jamcollsurg.2010.09.031]
39. Gonzales, K.D.; Douglas, G.; Pichakron, K.O.; Kwiat, D.A.; Gallardo, S.G.; Encinas, J.L.; Hirose, S.; Harrison, M.R. Magnamosis III: Delivery of a magnetic compression anastomosis device using minimally invasive endoscopic techniques. J. Pediatr. Surg.; 2012; 47, pp. 1291-1295. [DOI: https://dx.doi.org/10.1016/j.jpedsurg.2012.03.042]
40. Wall, J.; Diana, M.; Leroy, J.; Deruijter, V.; Gonzales, K.D.; Lindner, V.; Harrison, M.; Marescaux, J. MAGNAMOSIS IV: Magnetic compression anastomosis for minimally invasive colorectal surgery. Endoscopy; 2013; 45, pp. 643-648. [PubMed: https://www.ncbi.nlm.nih.gov/pubmed/23807805][DOI: https://dx.doi.org/10.1055/s-0033-1344119]
41. Russell, K.W.; Rollins, M.D.; Feola, G.P.; Scaife, E.R. Magnamosis: A novel technique for the management of rectal atresia. BMJ Case Rep.; 2014; 2014, bcr2013201330. [DOI: https://dx.doi.org/10.1136/bcr-2013-201330]
42. Graves, C.E.; Co, C.; Hsi, R.S.; Kwiat, D.; Imamura-Ching, J.; Harrison, M.R.; Stoller, M.L. Magnetic Compression Anastomosis (Magnamosis): First-In-Human Trial. J. Am. Coll. Surg.; 2017; 225, pp. 676-681 e671. [PubMed: https://www.ncbi.nlm.nih.gov/pubmed/28843832][DOI: https://dx.doi.org/10.1016/j.jamcollsurg.2017.07.1062]
43. Ryou, M.; Aihara, H.; Thompson, C.C. Minimally invasive entero-enteral dual-path bypass using self-assembling magnets. Surg. Endosc.; 2016; 30, pp. 4533-4538. [DOI: https://dx.doi.org/10.1007/s00464-016-4789-x] [PubMed: https://www.ncbi.nlm.nih.gov/pubmed/26895911]
44. Machytka, E.; Buzga, M.; Zonca, P.; Lautz, D.B.; Ryou, M.; Simonson, D.C.; Thompson, C.C. Partial jejunal diversion using an incisionless magnetic anastomosis system: 1-year interim results in patients with obesity and diabetes. Gastrointest. Endosc.; 2017; 86, pp. 904-912. [DOI: https://dx.doi.org/10.1016/j.gie.2017.07.009] [PubMed: https://www.ncbi.nlm.nih.gov/pubmed/28716404]
45. Schlottmann, F.; Ryou, M.; Lautz, D.; Thompson, C.C.; Buxhoeveden, R. Sutureless Duodeno-Ileal Anastomosis with Self-Assembling Magnets: Safety and Feasibility of a Novel Metabolic Procedure. Obes. Surg.; 2021; 31, pp. 4195-4202. [DOI: https://dx.doi.org/10.1007/s11695-021-05554-z] [PubMed: https://www.ncbi.nlm.nih.gov/pubmed/34181211]
46. Gumustop, D.; Seddon, D.; Gumustop, B.; Wang, J. A comparison of anastomosis strength between sutures, staples, and self-forming magnets. iGIE; 2022; 1, pp. 11-14. [DOI: https://dx.doi.org/10.1016/j.igie.2022.10.007]
47. Ore, A.S.; Althoff, A.; Kull, D.R.; Baldwin, T.J.; Van Eps, J.L.; Messaris, E. Comparative early histologic healing quality of magnetic versus stapled small bowel anastomosis. Surgery; 2022; 17, pp. 1060-1065. [PubMed: https://www.ncbi.nlm.nih.gov/pubmed/36566103][DOI: https://dx.doi.org/10.1016/j.surg.2022.11.030]
48. Ore, A.S.; Askenasy, E.; Ryou, M.; Baldwin, T.; Thompson, C.C.; Messaris, E. Evaluation of sutureless anastomosis after ileostomy takedown using the self-forming magnet anastomosis system in a porcine model. Surg. Endosc.; 2022; 36, pp. 7664-7672. [DOI: https://dx.doi.org/10.1007/s00464-022-09082-4]
49. Zaritzky, M.; Ben, R.; Zylberg, G.I.; Yampolsky, B. Magnetic compression anastomosis as a nonsurgical treatment for esophageal atresia. Pediatr. Radiol.; 2009; 39, pp. 945-949. [DOI: https://dx.doi.org/10.1007/s00247-009-1305-7]
50. Zaritzky, M.; Ben, R.; Johnston, K. Magnetic gastrointestinal anastomosis in pediatric patients. J. Pediatr. Surg.; 2014; 49, pp. 1131-1137. [DOI: https://dx.doi.org/10.1016/j.jpedsurg.2013.11.002]
51. Dorman, R.M.; Vali, K.; Harmon, C.M.; Zaritzky, M.; Bass, K.D. Repair of esophageal atresia with proximal fistula using endoscopic magnetic compression anastomosis (magnamosis) after staged lengthening. Pediatr. Surg. Int.; 2016; 32, pp. 525-528. [PubMed: https://www.ncbi.nlm.nih.gov/pubmed/27012861][DOI: https://dx.doi.org/10.1007/s00383-016-3889-y]
52. Slater, B.J.; Borobia, P.; Lovvorn, H.N.; Raees, M.A.; Bass, K.D.; Almond, S.; Hoover, J.D.; Kumar, T.; Zaritzky, M. Use of Magnets as a Minimally Invasive Approach for Anastomosis in Esophageal Atresia: Long-Term Outcomes. J. Laparoendosc. Adv. Surg. Tech. A; 2019; 29, pp. 1202-1206. [DOI: https://dx.doi.org/10.1089/lap.2019.0199]
53. Muensterer, O.J.; Sterlin, A.; Oetzmann von Sochaczewski, C.; Lindner, A.; Heimann, A.; Balus, A.; Dickmann, J.; Nuber, M.; Patel, V.H.; Manfredi, M.A. et al. An experimental study on magnetic esophageal compression anastomosis in piglets. J. Pediatr. Surg.; 2020; 55, pp. 425-432. [DOI: https://dx.doi.org/10.1016/j.jpedsurg.2019.04.029] [PubMed: https://www.ncbi.nlm.nih.gov/pubmed/31128845]
54. Wolfe, E.; Zidane, M.; Hancock, B.J.; Lum Min, S.A.; Zaritzky, M.; Keijzer, R. Magnamosis for esophageal atresia is associated with anastomotic strictures requiring an increased number of dilatations. J. Pediatr. Surg.; 2020; 55, pp. 821-823. [DOI: https://dx.doi.org/10.1016/j.jpedsurg.2020.01.022] [PubMed: https://www.ncbi.nlm.nih.gov/pubmed/32061365]
55. Muensterer, O.J.; Evans, L.L.; Sterlin, A.; Sahlabadi, M.; Aribindi, V.; Lindner, A.; Konig, T.; Harrison, M.R. Novel Device for Endoluminal Esophageal Atresia Repair: First-in-Human Experience. Pediatrics; 2021; 148, e2020049627. [DOI: https://dx.doi.org/10.1542/peds.2020-049627] [PubMed: https://www.ncbi.nlm.nih.gov/pubmed/34615695]
56. Evans, L.L.; Chen, C.S.; Muensterer, O.J.; Sahlabadi, M.; Lovvorn, H.N.; Novotny, N.M.; Upperman, J.S.; Martinez, J.A.; Bruzoni, M.; Dunn, J.C.Y. et al. The novel application of an emerging device for salvage of primary repair in high-risk complex esophageal atresia. J. Pediatr. Surg.; 2022; 57, pp. 810-818. [DOI: https://dx.doi.org/10.1016/j.jpedsurg.2022.05.018] [PubMed: https://www.ncbi.nlm.nih.gov/pubmed/35760639]
57. Mimuro, A.; Tsuchida, A.; Yamanouchi, E.; Itoi, T.; Ozawa, T.; Ikeda, T.; Nakamura, R.; Koyanagi, Y.; Nakamura, K. A novel technique of magnetic compression anastomosis for severe biliary stenosis. Gastrointest. Endosc.; 2003; 58, pp. 283-287. [DOI: https://dx.doi.org/10.1067/mge.2003.354]
58. Muraoka, N.; Uematsu, H.; Yamanouchi, E.; Kinoshita, K.; Takeda, T.; Ihara, N.; Matsunami, H.; Itoh, H. Yamanouchi magnetic compression anastomosis for bilioenteric anastomotic stricture after living-donor liver transplantation. J. Vasc. Interv. Radiol.; 2005; 16, pp. 1263-1267. [PubMed: https://www.ncbi.nlm.nih.gov/pubmed/16151070][DOI: https://dx.doi.org/10.1097/01.RVI.0000173280.56442.9E]
59. Matsuno, N.; Uchiyama, M.; Nakamura, Y.; Iwamoto, H.; Hama, K.; Ashizawa, T.; Nagao, T.; Yamanouchi, E. A nonsuture anastomosis using magnetic compression for biliary stricture after living donor liver transplantation. Hepatogastroenterology; 2009; 56, pp. 47-49.
60. Jang, S.I.; Cho, J.H.; Lee, D.K. Magnetic Compression Anastomosis for the Treatment of Post-Transplant Biliary Stricture. Clin. Endosc.; 2020; 53, pp. 266-275. [DOI: https://dx.doi.org/10.5946/ce.2020.095]
61. Takamizawa, S.; Yamanouchi, E.; Muraji, T.; Nishijima, E.; Satoh, S.; Tsugawa, J. MCRA of an anastomotic stenosis after esophagoesophagostomy for long gap esophageal atresia: A case report. J. Pediatr. Surg.; 2007; 42, pp. 769-772. [DOI: https://dx.doi.org/10.1016/j.jpedsurg.2006.12.042]
62. Woo, R.; Wong, C.M.; Trimble, Z.; Puapong, D.; Koehler, S.; Miller, S.; Johnson, S. Magnetic Compression Stricturoplasty For Treatment of Refractory Esophageal Strictures in Children: Technique and Lessons Learned. Surg. Innov.; 2017; 24, pp. 432-439. [PubMed: https://www.ncbi.nlm.nih.gov/pubmed/28745145][DOI: https://dx.doi.org/10.1177/1553350617720994]
63. Kamada, T.; Ohdaira, H.; Takeuchi, H.; Takahashi, J.; Marukuchi, R.; Suzuki, N.; Narihiro, S.; Hoshimoto, S.; Yoshida, M.; Yamanouchi, E. et al. Magnetic compression anastomosis for non-anastomotic stenosis of the proximal jejunum after total gastrectomy with Roux-en-Y reconstruction: A case report. Surg. Case Rep.; 2020; 6, 167. [DOI: https://dx.doi.org/10.1186/s40792-020-00932-8] [PubMed: https://www.ncbi.nlm.nih.gov/pubmed/32648159]
64. Isozaki, T.; Murakami, K.; Yamanouchi, E.; Uesato, M.; Toyozumi, T.; Koide, Y.; Tsukamoto, S.; Sakata, H.; Hayano, K.; Kano, M. et al. Magnetic compression anastomosis is effective in treating stenosis after esophageal cancer surgery: A case report. Surg. Case Rep.; 2020; 6, 213. [DOI: https://dx.doi.org/10.1186/s40792-020-00974-y] [PubMed: https://www.ncbi.nlm.nih.gov/pubmed/32804348]
65. Liu, H.; Yang, Z.; Lu, W.; Chen, Z.; Chen, L.; Han, S.; Wu, X.; Cai, T.; Cai, Y. Chemokines and chemokine receptors: A new strategy for breast cancer therapy. Cancer Med.; 2020; 9, pp. 3786-3799. [DOI: https://dx.doi.org/10.1002/cam4.3014]
66. Kılıç, A.; Şişik, A. A novel method for treatment of persistent colorectal anastomotic strictures: Magnetic compression strictureplasty. J. Surg. Med.; 2020; 4, pp. 331-333. [DOI: https://dx.doi.org/10.28982/josam.737762]
67. Liu, S.; Fang, Y.; Lv, Y.; Zhao, J.; Luo, R.; Luo, R.; Cheng, J.; Yang, H.; Zhang, A.; Shen, Y. et al. Magnetic compression stricturoplasty in patients with severe stricture after simultaneous esophageal atresia and duodenal obstruction repair: A case report. Exp. Ther. Med.; 2022; 23, 93. [DOI: https://dx.doi.org/10.3892/etm.2021.11016] [PubMed: https://www.ncbi.nlm.nih.gov/pubmed/34976135]
68. Falk, V.; Walther, T.; Stein, H.; Jacobs, S.; Walther, C.; Rastan, A.; Wimmer-Greinecker, G.; Mohr, F.W. Facilitated endoscopic beating heart coronary artery bypass grafting using a magnetic coupling device. J. Thorac. Cardiovasc. Surg.; 2003; 126, pp. 1575-1579. [DOI: https://dx.doi.org/10.1016/S0022-5223(03)00793-1]
69. Klima, U.; Falk, V.; Maringka, M.; Bargenda, S.; Badack, S.; Moritz, A.; Mohr, F.; Haverich, A.; Wimmer-Greinecker, G. Magnetic vascular coupling for distal anastomosis in coronary artery bypass grafting: A multicenter trial. J. Thorac. Cardiovasc. Surg.; 2003; 126, pp. 1568-1574. [DOI: https://dx.doi.org/10.1016/S0022-5223(03)01314-X]
70. Klima, U.; Maringka, M.; Bagaev, E.; Kirschner, S.; Haverich, A. Total magnetic vascular coupling for arterial revascularization. J. Thorac. Cardiovasc. Surg.; 2004; 127, pp. 602-603. [DOI: https://dx.doi.org/10.1016/j.jtcvs.2003.09.050]
71. Wong, T.; Mayet, J.; Casula, R.P. Minimal invasive direct revascularisation of the left anterior descending artery using a novel magnetic vascular anastomotic device. Heart; 2004; 90, 632. [DOI: https://dx.doi.org/10.1136/hrt.2003.023929]
72. Athanasiou, T.; Ashrafian, H.; Glenville, B.; Casula, R. Coronary artery bypass with the use of a magnetic distal anastomotic device: Surgical technique and preliminary experience. Heart. Surg. Forum; 2004; 7, pp. 356-359. [DOI: https://dx.doi.org/10.1532/HSF98.2004-1024] [PubMed: https://www.ncbi.nlm.nih.gov/pubmed/15769704]
73. Falk, V.; Walther, T.; Jacobs, S.; Wolf, R.K.; Mohr, F.W. Facilitated MIDCAB using a magnetic coupling device. Ann. Thorac. Surg.; 2005; 79, pp. 691-693. [DOI: https://dx.doi.org/10.1016/S0003-4975(03)01403-6] [PubMed: https://www.ncbi.nlm.nih.gov/pubmed/15680865]
74. Vicol, C.; Eifert, S.; Oberhoffer, M.; Boekstegers, P.; Knez, A.; Christ, F.; Reichart, B. Early clinical results with a magnetic connector for distal coronary artery anastomoses. Ann. Thorac. Surg.; 2005; 79, pp. 1738–1742; discussion 1742–1743. [DOI: https://dx.doi.org/10.1016/j.athoracsur.2004.04.096] [PubMed: https://www.ncbi.nlm.nih.gov/pubmed/15854967]
75. Klima, U.; Kofidis, T.; Beilner, J.; Maringka, M.; Kirschner, S.; Haverich, A. Anastomotic obstruction after magnetic vascular coupling after minimally invasive direct coronary artery bypass grafting. Ann. Thorac. Surg.; 2006; 81, pp. 728-729. [DOI: https://dx.doi.org/10.1016/j.athoracsur.2004.11.019]
76. Vicol, C.; Eifert, S.; Oberhoffer, M.; Boekstegers, P.; Reichart, B. Mid-term patency after magnetic coupling for distal bypass anastomosis in coronary surgery. Ann. Thorac. Surg.; 2006; 82, pp. 1452-1456. [DOI: https://dx.doi.org/10.1016/j.athoracsur.2006.04.090] [PubMed: https://www.ncbi.nlm.nih.gov/pubmed/16996952]
77. Charitou, A.; Panesar, S.S.; De, L.S.R.; Athanasiou, T. Novel use of a magnetic coupling device to repair damage of the internal thoracic artery. J. Card. Surg.; 2006; 21, pp. 89-91. [DOI: https://dx.doi.org/10.1111/j.1540-8191.2006.00177.x] [PubMed: https://www.ncbi.nlm.nih.gov/pubmed/16426358]
78. Lipham, J.C.; DeMeester, T.R.; Ganz, R.A.; Bonavina, L.; Saino, G.; Dunn, D.H.; Fockens, P.; Bemelman, W. The LINX(R) reflux management system: Confirmed safety and efficacy now at 4 years. Surg. Endosc.; 2012; 26, pp. 2944-2949. [PubMed: https://www.ncbi.nlm.nih.gov/pubmed/22538694][DOI: https://dx.doi.org/10.1007/s00464-012-2289-1]
79. Ganz, R.A.; Peters, J.H.; Horgan, S.; Bemelman, W.A.; Dunst, C.M.; Edmundowicz, S.A.; Lipham, J.C.; Luketich, J.D.; Melvin, W.S.; Oelschlager, B.K. et al. Esophageal sphincter device for gastroesophageal reflux disease. N. Engl. J. Med.; 2013; 368, pp. 719-727. [DOI: https://dx.doi.org/10.1056/NEJMoa1205544]
80. Bonavina, L.; Saino, G.; Bona, D.; Sironi, A.; Lazzari, V. One hundred consecutive patients treated with magnetic sphincter augmentation for gastroesophageal reflux disease: 6 years of clinical experience from a single center. J. Am. Coll. Surg.; 2013; 217, pp. 577-585. [DOI: https://dx.doi.org/10.1016/j.jamcollsurg.2013.04.039]
81. Smith, C.D.; DeVault, K.R.; Buchanan, M. Introduction of mechanical sphincter augmentation for gastroesophageal reflux disease into practice: Early clinical outcomes and keys to successful adoption. J. Am. Coll. Surg.; 2014; 218, pp. 776-781. [DOI: https://dx.doi.org/10.1016/j.jamcollsurg.2013.12.034]
82. Bauer, M.; Meining, A.; Kranzfelder, M.; Jell, A.; Schirren, R.; Wilhelm, D.; Friess, H.; Feussner, H. Endoluminal perforation of a magnetic antireflux device. Surg. Endosc.; 2015; 29, pp. 3806-3810. [DOI: https://dx.doi.org/10.1007/s00464-015-4145-6] [PubMed: https://www.ncbi.nlm.nih.gov/pubmed/25877789]
83. Saino, G.; Bonavina, L.; Lipham, J.C.; Dunn, D.; Ganz, R.A. Magnetic Sphincter Augmentation for Gastroesophageal Reflux at 5 Years: Final Results of a Pilot Study Show Long-Term Acid Reduction and Symptom Improvement. J. Laparoendosc. Adv. Surg. Tech. A; 2015; 25, pp. 787-792. [DOI: https://dx.doi.org/10.1089/lap.2015.0394] [PubMed: https://www.ncbi.nlm.nih.gov/pubmed/26437027]
84. Aiolfi, A.; Asti, E.; Bernardi, D.; Bonitta, G.; Rausa, E.; Siboni, S.; Bonavina, L. Early results of magnetic sphincter augmentation versus fundoplication for gastroesophageal reflux disease: Systematic review and meta-analysis. Int. J. Surg.; 2018; 52, pp. 82-88. [DOI: https://dx.doi.org/10.1016/j.ijsu.2018.02.041] [PubMed: https://www.ncbi.nlm.nih.gov/pubmed/29471155]
85. Bell, R.; Lipham, J.; Louie, B.E.; Williams, V.; Luketich, J.; Hill, M.; Richards, W.; Dunst, C.; Lister, D.; McDowell-Jacobs, L. et al. Magnetic Sphincter Augmentation Superior to Proton Pump Inhibitors for Regurgitation in a 1-Year Randomized Trial. Clin. Gastroenterol. Hepatol.; 2020; 18, pp. 1736-1743 e1732. [DOI: https://dx.doi.org/10.1016/j.cgh.2019.08.056] [PubMed: https://www.ncbi.nlm.nih.gov/pubmed/31518717]
86. Bortolotti, M.; Ugolini, G.; Grandis, A.; Montroni, I.; Mazzero, G. A novel magnetic device to prevent fecal incontinence (preliminary study). Int. J. Color. Dis.; 2008; 23, pp. 499-501. [DOI: https://dx.doi.org/10.1007/s00384-008-0437-9]
87. Lehur, P.A.; McNevin, S.; Buntzen, S.; Mellgren, A.F.; Laurberg, S.; Madoff, R.D. Magnetic anal sphincter augmentation for the treatment of fecal incontinence: A preliminary report from a feasibility study. Dis. Colon. Rectum.; 2010; 53, pp. 1604-1610. [DOI: https://dx.doi.org/10.1007/DCR.0b013e3181f5d5f7]
88. Barussaud, M.L.; Mantoo, S.; Wyart, V.; Meurette, G.; Lehur, P.A. The magnetic anal sphincter in faecal incontinence: Is initial success sustained over time?. Color. Dis.; 2013; 15, pp. 1499-1503. [PubMed: https://www.ncbi.nlm.nih.gov/pubmed/24103055][DOI: https://dx.doi.org/10.1111/codi.12423]
89. Jayne, D.G.; Williams, A.E.; Corrigan, N.; Croft, J.; Pullan, A.; Napp, V.; Kelly, R.; Meads, D.; Vargas-Palacios, A.; Martin, A. et al. Sacral nerve stimulation versus the magnetic sphincter augmentation device for adult faecal incontinence: The SaFaRI RCT. Health Technol. Assess.; 2021; 25, pp. 1-96. [DOI: https://dx.doi.org/10.3310/hta25180]
90. Ali-El-Dein, B.; El-Demerdash, R.; Kock, N.G.; Ghoneim, M.A. A magnetic device for increasing the urethral resistance to flow: An experimental study in female dogs. BJU Int.; 2000; 85, pp. 150-154. [DOI: https://dx.doi.org/10.1046/j.1464-410x.2000.00391.x]
91. Dominguez, G.; Durand, L.; De Rosa, J.; Danguise, E.; Arozamena, C.; Ferraina, P.A. Retraction and triangulation with neodymium magnetic forceps for single-port laparoscopic cholecystectomy. Surg. Endosc.; 2009; 23, pp. 1660-1666. [DOI: https://dx.doi.org/10.1007/s00464-009-0504-5]
92. Rivas, H.; Robles, I.; Riquelme, F.; Vivanco, M.; Jimenez, J.; Marinkovic, B.; Uribe, M. Magnetic Surgery: Results From First Prospective Clinical Trial in 50 Patients. Ann. Surg.; 2018; 267, pp. 88-93. [PubMed: https://www.ncbi.nlm.nih.gov/pubmed/27759614][DOI: https://dx.doi.org/10.1097/SLA.0000000000002045]
93. Haskins, I.N.; Strong, A.T.; Allemang, M.T.; Bencsath, K.P.; Rodriguez, J.H.; Kroh, M.D. Magnetic surgery: First U.S. experience with a novel device. Surg. Endosc.; 2018; 32, pp. 895-899. [DOI: https://dx.doi.org/10.1007/s00464-017-5762-z] [PubMed: https://www.ncbi.nlm.nih.gov/pubmed/28733750]
94. Morales-Conde, S.; Dominguez, G.; Gomez, J.C.; Socas, M.; Barranco, A.; Moreno, J.G.; Padillo, F.J. Magnetic-assisted single-port sleeve gastrectomy. Surg. Innov.; 2013; 20, pp. NP9-NP11. [DOI: https://dx.doi.org/10.1177/1553350611427548] [PubMed: https://www.ncbi.nlm.nih.gov/pubmed/22143751]
95. Rahman, R.; Azagury, D.E. Novel technologies and techniques in bariatric surgery. Minerva Chir.; 2017; 72, pp. 125-139. [PubMed: https://www.ncbi.nlm.nih.gov/pubmed/27981827][DOI: https://dx.doi.org/10.23736/S0026-4733.16.07265-5]
96. Davis, M.; Davalos, G.; Ortega, C.; Chen, S.; Schimpke, S.; Jain-Spangler, K.; Yoo, J.; Seymour, K.; Sudan, R.; Portenier, D. et al. Magnetic Liver Retraction: An Incision-Less Approach for Less Invasive Bariatric Surgery. Obes. Surg.; 2019; 29, pp. 1068-1073. [DOI: https://dx.doi.org/10.1007/s11695-018-03655-w]
97. Rassweiler, M.C.; Michel, M.S.; Ritter, M.; Honeck, P. Magnetic Ureteral Stent Removal Without Cystoscopy: A Randomized Controlled Trial. J. Endourol.; 2017; 31, pp. 762-766. [PubMed: https://www.ncbi.nlm.nih.gov/pubmed/28478732][DOI: https://dx.doi.org/10.1089/end.2017.0051]
98. Sevcenco, S.; Eredics, K.; Lusuardi, L.; Klingler, H.C. Evaluation of pain perception associated with use of the magnetic-end ureteric double-J stent for short-term ureteric stenting. World J. Urol.; 2018; 36, pp. 475-479. [DOI: https://dx.doi.org/10.1007/s00345-017-2146-x] [PubMed: https://www.ncbi.nlm.nih.gov/pubmed/29236163]
99. Harrison, M.R.; Curran, P.F.; Jamshidi, R.; Christensen, D.; Bratton, B.J.; Fechter, R.; Hirose, S. Magnetic mini-mover procedure for pectus excavatum II: Initial findings of a Food and Drug Administration-sponsored trial. J. Pediatr. Surg.; 2010; 45, pp. 185–191; discussion 191–192. [DOI: https://dx.doi.org/10.1016/j.jpedsurg.2009.10.032] [PubMed: https://www.ncbi.nlm.nih.gov/pubmed/20105602]
100. Harrison, M.R.; Gonzales, K.D.; Bratton, B.J.; Christensen, D.; Curran, P.F.; Fechter, R.; Hirose, S. Magnetic mini-mover procedure for pectus excavatum III: Safety and efficacy in a Food and Drug Administration-sponsored clinical trial. J. Pediatr. Surg.; 2012; 47, pp. 154-159. [DOI: https://dx.doi.org/10.1016/j.jpedsurg.2011.10.039]
101. Graves, C.E.; Hirose, S.; Raff, G.W.; Iqbal, C.W.; Imamura-Ching, J.; Christensen, D.; Fechter, R.; Kwiat, D.; Harrison, M.R. Magnetic Mini-Mover Procedure for pectus excavatum IV: FDA sponsored multicenter trial. J. Pediatr. Surg.; 2017; 52, pp. 913-919. [DOI: https://dx.doi.org/10.1016/j.jpedsurg.2017.03.009]
102. Rosenbluth, K.H.; Kwiat, D.A.; Harrison, M.R.; Kezirian, E.J. Hyoid bone advancement for improving airway patency: Cadaver study of a magnet-based system. Otolaryngol. Head Neck Surg.; 2012; 146, pp. 491-496. [DOI: https://dx.doi.org/10.1177/0194599811429522] [PubMed: https://www.ncbi.nlm.nih.gov/pubmed/22114308]
103. Cheung, K.M.; Cheung, J.P.; Samartzis, D.; Mak, K.C.; Wong, Y.W.; Cheung, W.Y.; Akbarnia, B.A.; Luk, K.D. Magnetically controlled growing rods for severe spinal curvature in young children: A prospective case series. Lancet; 2012; 379, pp. 1967-1974. [DOI: https://dx.doi.org/10.1016/S0140-6736(12)60112-3] [PubMed: https://www.ncbi.nlm.nih.gov/pubmed/22520264]
104. Jenks, M.; Craig, J.; Higgins, J.; Willits, I.; Barata, T.; Wood, H.; Kimpton, C.; Sims, A. The MAGEC system for spinal lengthening in children with scoliosis: A NICE Medical Technology Guidance. Appl. Health Econ. Health Policy; 2014; 12, pp. 587-599. [DOI: https://dx.doi.org/10.1007/s40258-014-0127-4] [PubMed: https://www.ncbi.nlm.nih.gov/pubmed/25172432]
105. Lorenz, H.M.; Badwan, B.; Hecker, M.M.; Tsaknakis, K.; Groenefeld, K.; Braunschweig, L.; Hell, A.K. Magnetically Controlled Devices Parallel to the Spine in Children with Spinal Muscular Atrophy. JB JS Open Access; 2017; 2, e0036. [DOI: https://dx.doi.org/10.2106/JBJS.OA.17.00036]
106. Subramanian, T.; Ahmad, A.; Mardare, D.M.; Kieser, D.C.; Mayers, D.; Nnadi, C. A six-year observational study of 31 children with early-onset scoliosis treated using magnetically controlled growing rods with a minimum follow-up of two years. Bone Jt. J.; 2018; 100-B, pp. 1187-1200. [DOI: https://dx.doi.org/10.1302/0301-620X.100B9.BJJ-2018-0031.R2] [PubMed: https://www.ncbi.nlm.nih.gov/pubmed/30168755]
107. Oetgen, M.E.; McNulty, E.M.; Matthews, A.L. Cost-Effectiveness of Magnetically Controlled Growing Rods: Who Really Benefits?. Spine Deform.; 2019; 7, pp. 501-504. [DOI: https://dx.doi.org/10.1016/j.jspd.2018.09.066] [PubMed: https://www.ncbi.nlm.nih.gov/pubmed/31053322]
108. Harshavardhana, N.S.; Noordeen, M.H.H.; Dormans, J.P. Cost Analysis of Magnet-driven Growing Rods for Early-onset Scoliosis at 5 Years. Spine; 2019; 44, pp. 60-67. [PubMed: https://www.ncbi.nlm.nih.gov/pubmed/29894450][DOI: https://dx.doi.org/10.1097/BRS.0000000000002738]
109. Guan, D.; Zhang, Y.; Xu, J. Clinical Outcome of Magnetically Controlled Growing Rod in Early-onset Scoliosis: A Systematic Review. Clin. Spine Surg.; 2020; 33, pp. 150-155. [DOI: https://dx.doi.org/10.1097/BSD.0000000000000907]
110. Steichen, F.M.; Ravitch, M.M. History of mechanical devices and instruments for suturing. Curr. Probl. Surg.; 1982; 19, pp. 3-52. [DOI: https://dx.doi.org/10.1016/0011-3840(82)90004-1] [PubMed: https://www.ncbi.nlm.nih.gov/pubmed/7039981]
111. Hardy, T.G., Jr.; Pace, W.G.; Maney, J.W.; Katz, A.R.; Kaganov, A.L. A biofragmentable ring for sutureless bowel anastomosis. An experimental study. Dis. Colon. Rectum.; 1985; 28, pp. 484-490. [DOI: https://dx.doi.org/10.1007/BF02554090]
112. Wullstein, C.; Gross, E. Compression anastomosis (AKA-2) in colorectal surgery: Results in 442 consecutive patients. Br. J. Surg.; 2000; 87, pp. 1071-1075. [DOI: https://dx.doi.org/10.1046/j.1365-2168.2000.01489.x] [PubMed: https://www.ncbi.nlm.nih.gov/pubmed/10931053]
113. Zhang, D.; Zheng, C.; Song, C. A modified method of using laparoscopic compression anastomosis clip for gastrointestinal anastomosis. Minim. Invasive Ther. Allied Technol.; 2010; 19, pp. 350-354. [DOI: https://dx.doi.org/10.3109/13645706.2010.527777] [PubMed: https://www.ncbi.nlm.nih.gov/pubmed/21091069]
114. Tabola, R.; Cirocchi, R.; Fingerhut, A.; Arezzo, A.; Randolph, J.; Grassi, V.; Binda, G.A.; D’Andrea, V.; Abraha, I.; Popivanov, G. et al. A systematic analysis of controlled clinical trials using the NiTi CAR compression ring in colorectal anastomoses. Tech. Coloproctol.; 2017; 21, pp. 177-184. [DOI: https://dx.doi.org/10.1007/s10151-017-1583-2] [PubMed: https://www.ncbi.nlm.nih.gov/pubmed/28132113]
115. Ryou, M.; Agoston, A.T.; Thompson, C.C. Endoscopic intestinal bypass creation by using self-assembling magnets in a porcine model. Gastrointest. Endosc.; 2016; 83, pp. 821-825. [PubMed: https://www.ncbi.nlm.nih.gov/pubmed/26522371][DOI: https://dx.doi.org/10.1016/j.gie.2015.10.023]
116. Pedersen, R.N.; Calzolari, E.; Husby, S.; Garne, E. Oesophageal atresia: Prevalence, prenatal diagnosis and associated anomalies in 23 European regions. Arch. Dis. Child.; 2012; 97, pp. 227-232. [DOI: https://dx.doi.org/10.1136/archdischild-2011-300597]
117. Zimmer, J.; Eaton, S.; Murchison, L.E.; De Coppi, P.; Ure, B.M.; Dingemann, C. State of Play: Eight Decades of Surgery for Esophageal Atresia. Eur. J. Pediatr. Surg.; 2019; 29, pp. 39-48. [DOI: https://dx.doi.org/10.1055/s-0038-1668150]
118. Sterlin, A.; Evans, L.; Mahler, S.; Lindner, A.; Dickmann, J.; Heimann, A.; Sahlabadi, M.; Aribindi, V.; Harrison, M.R.; Muensterer, O.J. An experimental study on long term outcomes after magnetic esophageal compression anastomosis in piglets. J. Pediatr. Surg.; 2022; 57, pp. 34-40. [DOI: https://dx.doi.org/10.1016/j.jpedsurg.2021.09.032]
119. Morbiducci, U.; Lemma, M.; Ponzini, R.; Boi, A.; Bondavalli, L.; Antona, C.; Montevecchi, F.M.; Redaelli, A. Does the Ventrica magnetic vascular positioner (MVP) for coronary artery bypass grafting significantly alter local fluid dynamics? A numeric study. Int. J. Artif. Organs; 2007; 30, pp. 628-639. [DOI: https://dx.doi.org/10.1177/039139880703000711]
120. Liu, S.Q.; Lei, P.; Cui, X.H.; Lv, Y.; Li, J.H.; Song, Y.L.; Zhao, G. Sutureless anastomoses using magnetic rings in canine liver transplantation model. J. Surg. Res.; 2013; 185, pp. 923-933. [DOI: https://dx.doi.org/10.1016/j.jss.2013.07.025]
121. Wang, H.H.; Ma, J.; Wang, S.P.; Ma, F.; Lu, J.W.; Xu, X.H.; Lv, Y.; Yan, X.P. Magnetic Anastomosis Rings to Create Portacaval Shunt in a Canine Model of Portal Hypertension. J. Gastrointest. Surg.; 2019; 23, pp. 2184-2192. [PubMed: https://www.ncbi.nlm.nih.gov/pubmed/30132290][DOI: https://dx.doi.org/10.1007/s11605-018-3888-5]
122. Yan, X.; Fan, C.; Ma, J.; Li, J.; Dong, D.; Wang, H.; Ma, F.; Zheng, X.; Lv, Y. Portacaval shunt established in six dogs using magnetic compression technique. PLoS ONE; 2013; 8, e76873. [DOI: https://dx.doi.org/10.1371/journal.pone.0076873] [PubMed: https://www.ncbi.nlm.nih.gov/pubmed/24098809]
123. Yan, X.P.; Liu, W.Y.; Ma, J.; Li, J.P.; Lv, Y. Extrahepatic portacaval shunt via a magnetic compression technique: A cadaveric feasibility study. World J. Gastroenterol.; 2015; 21, pp. 8073-8080. [DOI: https://dx.doi.org/10.3748/wjg.v21.i26.8073] [PubMed: https://www.ncbi.nlm.nih.gov/pubmed/26185378]
124. Wang, Y.R.; Dempsey, D.T.; Richter, J.E. Trends and perioperative outcomes of inpatient antireflux surgery in the United States, 1993-2006. Dis. Esophagus; 2011; 24, pp. 215-223. [DOI: https://dx.doi.org/10.1111/j.1442-2050.2010.01123.x] [PubMed: https://www.ncbi.nlm.nih.gov/pubmed/21073616]
125. Wong, W.D.; Congliosi, S.M.; Spencer, M.P.; Corman, M.L.; Tan, P.; Opelka, F.G.; Burnstein, M.; Nogueras, J.J.; Bailey, H.R.; Devesa, J.M. et al. The safety and efficacy of the artificial bowel sphincter for fecal incontinence: Results from a multicenter cohort study. Dis. Colon. Rectum.; 2002; 45, pp. 1139-1153. [DOI: https://dx.doi.org/10.1007/s10350-004-6381-z]
126. Bridoux, V.; Gourcerol, G.; Leroi, A.M.; Ducrotte, P.; Michot, F.; Tuech, J.J. Response to Barussaud et al.: The magnetic anal sphincter in faecal incontinence, is initial success sustained over time?. Color. Dis.; 2014; 16, pp. 145-146. [DOI: https://dx.doi.org/10.1111/codi.12519] [PubMed: https://www.ncbi.nlm.nih.gov/pubmed/24330499]
127. Wong, M.T.; Meurette, G.; Stangherlin, P.; Lehur, P.A. The magnetic anal sphincter versus the artificial bowel sphincter: A comparison of 2 treatments for fecal incontinence. Dis. Colon. Rectum.; 2011; 54, pp. 773-779. [DOI: https://dx.doi.org/10.1007/DCR.0b013e3182182689]
128. Wong, M.T.; Meurette, G.; Wyart, V.; Lehur, P.A. Does the magnetic anal sphincter device compare favourably with sacral nerve stimulation in the management of faecal incontinence?. Color. Dis.; 2012; 14, pp. e323-e329. [PubMed: https://www.ncbi.nlm.nih.gov/pubmed/22339789][DOI: https://dx.doi.org/10.1111/j.1463-1318.2012.02995.x]
129. Lehur, P.A.; Christoforidis, D.; Meurette, G. Artificial Sphincters to Treat Severe Fecal Incontinence: Currently in a Deadlock. Dis. Colon. Rectum.; 2020; 63, pp. 1017-1019. [DOI: https://dx.doi.org/10.1097/DCR.0000000000001708]
130. Block, A.M.; Tamburini, L.M.; Zeng, F.; Mancini, M.R.; Jackson, C.A.; Antonacci, C.L.; Karsmarski, O.P.; Stelzer, J.W.; Wellington, I.J.; Lee, M.C. Surgical Treatment of Pediatric Scoliosis: Historical Origins and Review of Current Techniques. Bioengineering; 2022; 9, 600. [PubMed: https://www.ncbi.nlm.nih.gov/pubmed/36290568][DOI: https://dx.doi.org/10.3390/bioengineering9100600]
131. Choi, E.; Yaszay, B.; Mundis, G.; Hosseini, P.; Pawelek, J.; Alanay, A.; Berk, H.; Cheung, K.; Demirkiran, G.; Ferguson, J. et al. Implant Complications After Magnetically Controlled Growing Rods for Early Onset Scoliosis: A Multicenter Retrospective Review. J. Pediatr. Orthop.; 2017; 37, pp. e588-e592. [DOI: https://dx.doi.org/10.1097/BPO.0000000000000803]
132. Graves, C.E.; Moyer, J.; Zobel, M.J.; Mora, R.; Smith, D.; O’Day, M.; Padilla, B.E. Intraoperative intercostal nerve cryoablation During the Nuss procedure reduces length of stay and opioid requirement: A randomized clinical trial. J. Pediatr. Surg.; 2019; 54, pp. 2250-2256. [DOI: https://dx.doi.org/10.1016/j.jpedsurg.2019.02.057] [PubMed: https://www.ncbi.nlm.nih.gov/pubmed/30935731]
133. Brana, I.; Calles, A.; LoRusso, P.M.; Yee, L.K.; Puchalski, T.A.; Seetharam, S.; Zhong, B.; de Boer, C.J.; Tabernero, J.; Calvo, E. Carlumab, an anti-C-C chemokine ligand 2 monoclonal antibody, in combination with four chemotherapy regimens for the treatment of patients with solid tumors: An open-label, multicenter phase 1b study. Target. Oncol.; 2015; 10, pp. 111-123. [DOI: https://dx.doi.org/10.1007/s11523-014-0320-2] [PubMed: https://www.ncbi.nlm.nih.gov/pubmed/24928772]
134. Lurje, G.; Raptis, D.A.; Steinemann, D.C.; Amygdalos, I.; Kambakamba, P.; Petrowsky, H.; Lesurtel, M.; Zehnder, A.; Wyss, R.; Clavien, P.A. et al. Cosmesis and Body Image in Patients Undergoing Single-port Versus Conventional Laparoscopic Cholecystectomy: A Multicenter Double-blinded Randomized Controlled Trial (SPOCC-trial). Ann. Surg.; 2015; 262, pp. 728–734; discussion 734–735. [DOI: https://dx.doi.org/10.1097/SLA.0000000000001474] [PubMed: https://www.ncbi.nlm.nih.gov/pubmed/26583659]
You have requested "on-the-fly" machine translation of selected content from our databases. This functionality is provided solely for your convenience and is in no way intended to replace human translation. Show full disclaimer
Neither ProQuest nor its licensors make any representations or warranties with respect to the translations. The translations are automatically generated "AS IS" and "AS AVAILABLE" and are not retained in our systems. PROQUEST AND ITS LICENSORS SPECIFICALLY DISCLAIM ANY AND ALL EXPRESS OR IMPLIED WARRANTIES, INCLUDING WITHOUT LIMITATION, ANY WARRANTIES FOR AVAILABILITY, ACCURACY, TIMELINESS, COMPLETENESS, NON-INFRINGMENT, MERCHANTABILITY OR FITNESS FOR A PARTICULAR PURPOSE. Your use of the translations is subject to all use restrictions contained in your Electronic Products License Agreement and by using the translation functionality you agree to forgo any and all claims against ProQuest or its licensors for your use of the translation functionality and any output derived there from. Hide full disclaimer
© 2023 by the authors. Licensee MDPI, Basel, Switzerland. This article is an open access article distributed under the terms and conditions of the Creative Commons Attribution (CC BY) license (https://creativecommons.org/licenses/by/4.0/). Notwithstanding the ProQuest Terms and Conditions, you may use this content in accordance with the terms of the License.
Abstract
The novel use of magnetic force to optimize modern surgical techniques originated in the 1970s. Since then, magnets have been utilized as an adjunct or alternative to a wide array of existing surgical procedures, ranging from gastrointestinal to vascular surgery. As the use of magnets in surgery continues to grow, the body of knowledge on magnetic surgical devices from preclinical development to clinical implementation has expanded significantly; however, the current magnetic surgical devices can be organized based on their core function: serving as a guidance system, creating a new connection, recreating a physiologic function, or utilization of an internal–external paired magnet system. The purpose of this article is to discuss the biomedical considerations during magnetic device development and review the current surgical applications of magnetic devices.
You have requested "on-the-fly" machine translation of selected content from our databases. This functionality is provided solely for your convenience and is in no way intended to replace human translation. Show full disclaimer
Neither ProQuest nor its licensors make any representations or warranties with respect to the translations. The translations are automatically generated "AS IS" and "AS AVAILABLE" and are not retained in our systems. PROQUEST AND ITS LICENSORS SPECIFICALLY DISCLAIM ANY AND ALL EXPRESS OR IMPLIED WARRANTIES, INCLUDING WITHOUT LIMITATION, ANY WARRANTIES FOR AVAILABILITY, ACCURACY, TIMELINESS, COMPLETENESS, NON-INFRINGMENT, MERCHANTABILITY OR FITNESS FOR A PARTICULAR PURPOSE. Your use of the translations is subject to all use restrictions contained in your Electronic Products License Agreement and by using the translation functionality you agree to forgo any and all claims against ProQuest or its licensors for your use of the translation functionality and any output derived there from. Hide full disclaimer