1. Introduction
Influenza and coronaviruses are highly contagious respiratory diseases that cause millions of cases and deaths annually worldwide [1,2]. Influenza typically peaks during winter months and causes approximately 1 billion cases, 3–5 million severe cases, and up to 650,000 deaths [3]. However, public health measures and travel restrictions implemented during the COVID-19 pandemic have significantly curbed global influenza circulation [4]. The influenza B/Yamagata lineage has not been detected since April 2020, and other influenza viruses with considerably lower genetic diversity have circulated [5]. As the COVID-19 pandemic gradually weakens owing to various quarantine and vaccination measures, epidemic prevention, and control, global travel is gradually returning to pre-epidemic levels. Therefore, it is necessary to develop measures to monitor and control the spread of seasonal influenza during the COVID-19 pandemic.
In response to this challenge, various multi-diagnostic real-time reverse transcription polymerase chain reaction (RT-PCR) assays that can simultaneously detect multiple respiratory pathogens, including influenza and COVID-19, have been developed. This is particularly useful during the COVID-19 pandemic because it can simultaneously diagnose influenza and COVID-19, which induce similar symptoms. Bouassa et al. developed an RT-qPCR assay that could simultaneously diagnose influenza, COVID-19, and respiratory syncytial virus (RSV), and this kit exhibited 97.9, 89.5, 97.6, and 100% accuracy for influenza A, influenza B, SARS-CoV-2, and RSV, respectively [6]. Additionally, as another diagnostic method, various rapid antigen tests that are capable of detecting both influenza and COVID-19 antigens in a single test are being developed [7]. Although RT-qPCR is a commonly used gold-standard method of diagnosis, it is relatively slow and requires specialized equipment and skilled personnel, limiting its usefulness to rapid diagnosis in resource-constrained settings [8]. Rapid diagnostic tests, which use antibodies to detect viral antigens or antibodies in patient samples, have the advantage of being quick but have low sensitivity and tend to have a high false-negative rate [9]. Therefore, an alternative diagnostic technique capable of accurately and quickly diagnosing viral infections is required.
Loop-mediated isothermal amplification (LAMP) is a molecular diagnostic technique that enables the rapid and sensitive detection of specific RNA or DNA sequences in target pathogens [10]. This technique amplifies the target nucleic acid sequence at a constant temperature, using a set of four to six primers that recognize distinct regions of the target sequence. Among these, two are loop primers that allow the reaction to occur more efficiently by promoting the formation of stem–loop structures in the amplification product [11]. The LAMP assay has several advantages over other nucleic acid amplification methods, such as being simple, sensitive, and rapid [12]. In addition, LAMP assays can be performed using a variety of detection methods, including colorimetric, lateral flow, fluorescence, and fluorescent probe detection [13,14,15,16]. Among these methods, fluorescent probe detection is particularly useful for the real-time monitoring of the LAMP reaction. Among the various fluorescent probe techniques, the LAMP-assimilating probe consists of a fluorescently-tagged probe and complementary quencher-tagged probe. The assimilating probe was designed to hybridize with the target sequence during the LAMP reaction and was incorporated into the amplification products by Bst DNA polymerase, resulting in an increase in the fluorescent signal [17]. As amplification proceeds, the fluorescence intensity of the probe increases and can be monitored in real-time using a fluorescent detector to determine the presence or absence of the target pathogen. The use of LAMP-assimilating probes enables a highly sensitive and specific simultaneous detection of multiple pathogens in a single reaction, making them a powerful tool for molecular diagnostics in various settings [18].
In this study, we developed a FluA/FluB/SARS-CoV-2 multiplex LAMP assay that can detect influenza A, influenza B, SARS-CoV-2, and an internal control (actin beta) using LAMP primers and assimilating probes that were previously reported [19,20]. Among the five different ratios of LAMP primer sets evaluated, the ratio of 1:0.5:1:0.2 of influenza A, influenza B, SARS-CoV-2, and the internal control primer set showed the best performance in LAMP tests against the clinical samples of influenza A, influenza B, and SARS-CoV-2. Finally, the performance of the FluA/FluB/SARS-CoV-2 multiplex LAMP assay was compared with that of the commercial AllplexTM SARS-CoV-2/FluA/FluB/RSV assay (Seegene, Seoul, Korea) for clinical samples.
2. Materials and Methods
2.1. Clinical Samples and RNA Extraction
The Korea Disease Control and Prevention Agency (KDCA) measured the SARS-CoV-2 titers using the plaque-forming unit (PFU) test and provided 20 strains, including one wild-type sample and 19 mutant samples. Influenza A H1N1, H1N1pdm09, H3N2, and influenza B virus were cultured at the Department of Laboratory Medicine, the Korea University Guro Hospital, of which the viral titers were measured using the TCID50 method. For clinical sensitivity testing, we used clinical SARS-CoV-2 nasopharyngeal swab (NP) (n = 96), influenza A NP (n = 117), and influenza B NP (n = 61) samples collected from SARS-CoV-2-, influenza A-, and influenza B-infected patients (from February 2018 to July 2022) at Korea University Guro Hospital. All clinical samples were confirmed using the AllplexTM SARS-CoV-2 assay (Seegene Inc., Seoul, South Korea) and Anyplex II RV16 Detection Kit (Seegene, Inc., Seoul, South Korea). To assess the specificity and cross-reactivity, 135 NP swab specimens were tested from individuals with (102) and without (33) viral respiratory infections. Respiratory viral infections, as confirmed via PCR using the Anyplex II RV16 detection kit, included nine coronaviruses (KHU1, NL63, and 229E), three RSV A, three RSV B, three adenoviruses (AdV), three parainfluenza virus (PIV) types 1–4, three human bocaviruses (HboV), three human enteroviruses (HEV), three human rhinoviruses (HRV), and three metapneumoviruses (MPV). Nucleic acids were extracted from all samples using Zentrix (Biozenthech, Seoul, Korea), according to the manufacturer’s instructions. Briefly, 200 µL of the sample was dispensed into a 96-well extraction plate and nucleic acid was extracted through the respiratory virus process program. The study was conducted in accordance with the guidelines of the Declaration of Helsinki and approved by the Institutional Review Board of Korea University Guro Hospital (approval number: 2021GR0547).
2.2. Primer Design
The influenza A, influenza B, SARS-CoV-2, and internal control RT-LAMP primer sets used in this study have been previously reported by our study group [19,20]. Primer sets for influenza A, influenza B, and SARS-CoV-2 were designed to target conserved regions of segment 7, the nucleoprotein, and the RdRP gene, respectively. The internal control primer set was designed within a conserved region of human actin beta mRNA (NM_001101.5:c.287-c.498) [19,20]. For the multiplex probe design, two types of additional synthetic oligonucleotide sequences were designed and added to the 5′ end of the LB or LF primer of each LAMP primer set. The 5′ end of the multiplex probe was tagged with a fluorescent marker. To quench the fluorophore multiplex probe, two types of complementary synthetic oligonucleotide sequences tagged with BHQ1 or BHQ2 at the 3′ end were used. All LAMP primers and probes were synthesized by Macrogen Inc. (Seoul, Republic of Korea; Table 1).
2.3. FluA/FluB/SARS-CoV-2 Multiplex LAMP Assay
The RT-LAMP assay was performed using ELPIS RT-LAMP 2X Master Mix (Elpis-Biotech, Daejeon, Republic of Korea). For the FluA/FluB/SARS-CoV-2 multiplex LAMP assay, the reaction mixture was prepared with 12.5 μL of 2X Master Mix, 1 μL of influenza A LAMP primer mix, 0.5 μL of influenza B LAMP primer mix, 1 μL of SARS-CoV-2 RdRP gene LAMP primer mix, 0.2 µL of internal control LAMP primer mix, 540 nM of quencher 1 solution, 30 nM of quencher 2 solution, and 3 μL of sample RNA (final reaction volume: 25 μL). The compositions of influenza A and influenza B LAMP primer mix were 1 μM of two outer primers (F3 and B3) and 32 μM of two inner primers (FIP and BIP), 4 μM of loop LF primer, 6 μM of loop LF probe primer, and 10 μM loop LB primer. The compositions of the SARS-CoV-2 RdRP gene primer mix were 4 μM of two outer primers (F3 and B3) and 32 μM of two inner primers (FIP and BIP), 10 μM of loop LF primer, 4 μM of loop LB probe primer, and 6 μM of loop LB probe primer. The compositions of actin beta gene (internal control) primer mix were 4 μM of two outer primers (F3 and B3) and 32 μM of two inner primers (FIP and BIP), 10 μM of loop LF primer, 4 μM of loop LB probe primer, and 6 μM of loop LB probe primer. The RT-LAMP assay was performed using a CFX 96 Touch Real-Time PCR Detection System (Bio-Rad Laboratories, Hercules, CA, USA) at 62 °C for 40 min. The FAM, Texas Red, Hex, and Cy5 fluorescence channels were used to detect influenza A, influenza B, SARS-CoV-2 RdRP, and internal control, respectively.
2.4. RT-PCR
The performance of the FluA/FluB/SARS-CoV-2 multiplex LAMP assay was compared with that of the AllplexTM SARS-CoV-2/FluA/FluB/RSV assay (Seegene, Seoul, Republic of Korea) using the CFX96 Touch Real-Time PCR Detection System (Bio-Rad, California, USA). The AllplexTM SARS-CoV-2/FluA/FluB/RSV assay was performed according to the manufacturer’s instructions. The PCR cycling conditions of the AllplexTM SARS-CoV-2/FluA/FluB/RSV assay were as follows: reverse transcription at 50 °C for 20 min, inactivation at 95 °C for 15 min, 2 cycles of 95 °C for 10 s, 60 °C for 40 s, and 72 °C for 20 s followed by 41 cycles of 95 °C for 10 s, 60 °C for 15 s, and 72 °C for 10 s with fluorescence detection at 60 °C and 72 °C.
2.5. Limit of Detection (LOD) Tests of the FluA/FluB/SARS-CoV-2 Multiplex LAMP Assay
To determine the detection limit of the FluA/FluB/SARS-CoV-2 multiplex LAMP assay, extracted RNA from culture broth samples of SARS-CoV-2 (1 × 103 PFU/mL), influenza A H1N1 (2 × 106 TCID50/mL), influenza A H1N1pdm09 (2 × 106 TCID50/mL), influenza A H3N2 (2 × 106 TCID50/mL), and influenza B (7 × 106 TCID50/mL) were respectively mixed with RNA isolated from non-infected NP clinical sample. Each sample was serially diluted 10-fold from the original sample to obtain six levels. All tests were repeated three times and determined as the minimum concentration in a 10-fold dilution series, at which three replicates were amplified.
2.6. Statistics
To calculate the diagnostic agreement between the FluA/FluB/SARS-CoV-2 multiplex LAMP assay and AllplexTM SARS-CoV-2/FluA/FluB/RSV assay, positive predictive value (PPV), negative predictive value (NPV), and kappa value, and kappa statistics were used. Statistical analysis was performed using SPSS 18 statistical package (SPSS 18 Inc., Chicago, IL, USA). The kappa index analyzed using crosstabulation was interpreted according to the interpretation suggested by Landis and Koch [21]. A kappa index of <0, 0–0.20, 0.21–0.40, 0.41–0.60, 0.61–0.80, and 0.81–1.00 indicates poor, slight agreement, fair, moderate, substantial, and almost perfect agreement between tests, respectively.
3. Results
3.1. Optimization of the FluA/FluB/SARS-CoV-2 Multiplex LAMP Assay Primer Set
In the multiplex LAMP assay, since there may be interference reactions between each primer set, an important factor in multiplex LAMP development is to obtain an optimal primer set ratio that can detect all target diseases by adjusting the ratio between each primer set. For optimization of the FluA/FluB/SARS-CoV-2 multiplex LAMP assay primer set, five different concentration ratios of the influenza A, influenza B, SARS-CoV-2, and internal control primer sets (No. 1: 1:1:1:0.2, No. 2: 1:0.5:1:0.2, No. 3: 1:0.5:0.5:0.2, No. 4: 0.5:0.5:1:0.2, and No. 5: 1:1:0.5:0.2, respectively) were tested with non-infected NP RNA samples spiked with SARS-CoV-2 (1 × 107 PFU/mL), influenza A H1N1 (5 × 106 TCID50/mL), and influenza B (1 × 106 TCID50/mL) (Table 2, Figure 1). Ratios No. 1 and No. 3 showed fast Ct values for influenza B and A, respectively, but they did not detect influenza A and SARS-CoV-2, respectively. Ratio No. 4 did not detect influenza A, and ratio No. 5 showed all signal detections but slower Ct values in all four signals compared with ratio No. 2. Among five different concentration ratios of primer sets, ratio No. 2 of the FluA/FluB/SARS-CoV-2 multiplex LAMP assay primer set (a ratio of four LAMP primer set of 1:0.5:1:0.2) showed the best performance. Therefore, ratio No. 2 (1:0.5:1:0.2) for influenza A, influenza B, SARS-CoV-2, and the internal control LAMP primer set was determined to be the optimum ratio for the FluA/FluB/SARS-CoV-2 multiplex LAMP assay (Table 2, Figure 1). Figure 1B and Table 3 show the performance of the FluA/FluB/SARS-CoV-2 multiplex LAMP assay against influenza A H1N1, influenza A, H1N1pdm09, influenza A H3N2, influenza B, SARS-CoV-2, and normal NP RNA samples at the optimum conditions (ratio of the influenza A, influenza B, and internal control primer sets was 1:0.5:1:0.2 at 62 °C).
3.2. Detection Limit of the FluA/FluB/SARS-CoV-2 Multiplex LAMP Assay and Allplex™ SARS-CoV-2/FluA/FluB/RSV Assay
Using the detection limit test for influenza A H1N1, H1N1pdm09, H3N2, influenza B, and SARS-CoV-2, the analytical performance of the FluA/FluB/SARS-CoV-2 multiplex LAMP assay was evaluated and compared with that of the Allplex™ SARS-CoV-2/FluA/FluB/RSV assay (Table 4). Each RNA sample, diluted 10-fold in six levels, was used for the detection limit testing of the Allplex™ SARS-CoV-2/FluA/FluB/RSV and FluA/FluB/SARS-CoV-2 multiplex LAMP assays. The detection limits of the Allplex™ SARS-CoV-2/FluA/FluB/RSV and FluA/FluB/SARS-CoV-2 multiplex LAMP assays were 10−1 PFU/mL/7 × 101 TCID50/mL and 100 PFU/mL/7 × 102 TCID50/mL against SARS-CoV-2 and influenza B RNA samples, respectively. For influenza A H1N1, H1N1pdm09, and H3N2 RNA samples, the detection limits of the Allplex™ SARS-CoV-2/FluA/FluB/RSV assay and FluA/FluB/SARS-CoV-2 multiplex LAMP assay were 2 × 102/7 × 108/2 × 102 TCID50/mL and 2 × 104/7 × 1010/2 × 104 TCID50/mL, respectively. In all test samples, the FluA/FluB/SARS-CoV-2 multiplex LAMP assay showed a one- or two-step higher detection limit compared to the Allplex™ SARS-CoV-2/FluA/FluB/RSV assay.
3.3. Comparison of Performance between the FluA/FluB/SARS-CoV-2 Multiplex LAMP and Allplex™ SARS-CoV-2/FluA/FluB/RSV Assays Using Clinical Samples
To confirm the performance of the FluA/FluB/SARS-CoV-2 multiplex LAMP assay, the sensitivities and specificities of the assay were compared with those of the Allplex™ SARS-CoV-2/FluA/FluB/RSV assay (Table 5). Both assays showed 100% specificity for normal NP clinical samples (n = 102). For influenza A NP clinical samples (n = 117), the sensitivities of the FluA/FluB/SARS-CoV-2 multiplex LAMP and Allplex™ SARS-CoV-2/FluA/FluB/RSV assays were 90.60% and 94.87%, respectively. The specificities of the FluA/FluB/SARS-CoV-2 multiplex LAMP assay for influenza A NP clinical samples were 100%, whereas those of the Allplex™ SARS-CoV-2/FluA/FluB/RSV assay were 99.03%. The positive predictive value (PPV) and negative predictive value (NPV) of the two kits were 100/90.27 and 99.10/94.44, respectively. For influenza B NP clinical samples (n = 61), the sensitivities of the FluA/FluB/SARS-CoV-2 multiplex LAMP assay and Allplex™ SARS-CoV-2/FluA/FluB/RSV assay were 86.89% and 88.52%, respectively. The specificities of the FluA/FluB/SARS-CoV-2 multiplex LAMP assay for influenza B NP clinical samples were 100%, whereas those of the Allplex™ SARS-CoV-2/FluA/FluB/RSV assay were 99.03%. The PPV and NPV of the two kits were 100/92.72 and 98.18/93.58, respectively. For SARS-CoV-2 NP clinical samples (n = 96), the sensitivities of the FluA/FluB/SARS-CoV-2 multiplex LAMP and Allplex™ SARS-CoV-2/FluA/FluB/RSV assays were 98.96% and 97.92%, respectively. The specificities of both the FluA/FluB/SARS-CoV-2 multiplex LAMP and AllplexTM SARS-CoV-2/FluA/FluB/RSV assays for SARS-CoV-2 NP clinical samples were 100%. The PPV and NPV of the two kits were 100/99.03 and 100/98.08, respectively. These results indicate that the FluA/FluB/SARS-CoV-2 multiplex LAMP assay possesses comparable sensitivity and specificity to the AllplexTM SARS-CoV-2/FluA/FluB/RSV assay for detecting influenza A/B and SARS-CoV-2 NP in clinical samples.
Next, we calculated the agreement between the FluA/FluB/SARS-CoV-2 multiplex LAMP and Allplex™ SARS-CoV-2/FluA/FluB/RSV assays using SPSS 18 statistical package (SPSS 18 Inc., Chicago, IL, USA) (Table 6). The kappa values for influenza A, influenza B, and SARS-CoV-2 NP clinical samples indicated substantial agreement between the two assays for these viruses. A kappa value of 0.845 for influenza A suggests almost perfect agreement, whereas a kappa value of 0.791 for influenza B suggests substantial agreement. The highest kappa value of 0.970 for SARS-CoV-2 NP clinical samples indicated an almost perfect agreement between the two assays for detecting SARS-CoV-2. Overall, these results suggest that the FluA/FluB/SARS-CoV-2 multiplex LAMP and Allplex™ SARS-CoV-2/FluA/FluB/RSV assays have good agreement for the detection of influenza A, influenza B, and SARS-CoV-2 NP clinical samples.
3.4. Cross-Reactivity Test
To confirm the absence of cross-reactivity of the FluA/FluB/SARS-CoV-2 multiplex LAMP assay with other common respiratory viruses, 33 NP swabs from patients with known infections with nine coronaviruses (229E, NL63, and OC43), three HEV, three AdV, three PIV, three MPV, three HboV, three HRV, and six RSV A/B were tested using the FluA/FluB/SARS-CoV-2 multiplex LAMP assay (Table 7). The FluA/FluB/SARS-CoV-2 multiplex LAMP assay showed negative results for Flu A, Flu B, and SARS-CoV-2 channels against 33 other common respiratory viruses. The internal controls of the kit were positive for all but three RSV B samples. These results suggest that the FluA/FluB/SARS-CoV-2 multiplex LAMP assay did not cross-react with other infectious viruses.
3.5. SARS-CoV-2 Mutant Test
As the coronavirus continues to spread, new strains are emerging. To confirm that the developed FluA/FluB/SARS-CoV-2 multiplex LAMP assay can detect various SARS-CoV-2 mutant strains, one SARS-CoV-2 wild-type and 19 SARS-CoV-2 mutant samples distributed by the KDCA were used to evaluate the performance of the kit (Table 8). The developed FluA/FluB/SARS-CoV-2 multiplex LAMP assay indicated positive reactions to all SARS-CoV-2 wild-type and mutant viruses tested.
4. Discussion
Presently, RT-qPCR is considered the most sensitive and specific molecular diagnostic test for infectious diseases, according to reports [22,23]. When new molecular diagnostic kits are developed, their performance is often evaluated using RT-qPCR as the reference standard [24,25]. Currently, LAMP is the most extensively researched and developed technique among various isothermal amplification methods. However, the detection limit of LAMP is similar or one to two orders of magnitude higher than that of RT-qPCR. Additionally, its sensitivity and specificity for clinical samples are reported to be similar to or slightly lower than those of RT-qPCR, ranging from 0 to 10% lower [23,26,27]. Although LAMP amplification can detect the target gene within 5–10 min when the target gene concentration is high, it does not demonstrate comparable sensitivity to RT-qPCR when the target gene concentration is low.
Despite these drawbacks, isothermal amplification methods remain attractive because of their isothermal amplification, rapidity, and ability to use relatively inexpensive equipment [27,28]. In fact, RT-qPCR kits take approximately 2–3 h to complete, but RT-LAMP analysis can detect several viruses simultaneously within 40 min because it does not require a reverse transcription step. Additionally, although the sensitivity and specificity of LAMP isothermal amplification may be lower than those of qPCR, it has higher sensitivity and specificity than rapid kits that mainly use antigen or antibody tests in the field [29,30]. Therefore, various methods incorporating isothermal amplification technology that can be used in facilities without the necessary equipment, or in the field, are being developed.
In this study, we optimized the primer set and reaction conditions of the FluA/FluB/SARS-CoV-2 multiplex LAMP assay and evaluated its sensitivity and specificity compared to those of the commercially available AllplexTM SARS-CoV-2/FluA/FluB/RSV assay. The sensitivities of the FluA/FluB/SARS-CoV-2 multiplex LAMP assay for influenza A, influenza B, and SARS-CoV-2 NP clinical samples were 90.6%, 86.89%, and 98.96%, respectively. The specificity of the FluA/FluB/SARS-CoV-2 multiplex LAMP assay against non-infected samples was 100%, which was excellent in terms of ruling out false positives. When compared to the commercial Allplex™ SARS-CoV-2/FluA/FluB/RSV assay (94.87%, 88.52%, and 97.92%), sensitivity levels of the FluA/FluB/SARS-CoV-2 multiplex LAMP assay were slightly lower but still quite accurate. In addition, the agreement between the FluA/FluB/SARS-CoV-2 multiplex LAMP assay and the AllplexTM SARS-CoV-2/FluA/FluB/RSV assay revealed high kappa values for influenza A, influenza B, and SARS-CoV-2 NP clinical samples, indicating substantial agreement between the two assays. Moreover, the FluA/FluB/SARS-CoV-2 multiplex LAMP assay showed no cross-reactivity with the other 11 respiratory viruses, which is a significant advantage over some commercial kits that have shown cross-reactivity with other viruses. Moreover, our LAMP kit demonstrated the ability to detect 20 SARS-CoV-2 mutants, including highly transmissible Delta and Omicron variants, which are crucial for identifying and controlling emerging COVID-19 outbreaks. These findings suggest that the multiplex LAMP assay is reliable for detecting influenza A, influenza B, and SARS-CoV-2 and can be used as an alternative to the AllplexTM SARS-CoV-2/FluA/FluB/RSV assay.
Generally, LAMP diagnostic kits are being developed as a middle ground between rapid antigen/antibody kits, which are available for field use but may have limited sensitivity, and the qPCR assays, which have high sensitivity but require complex equipment and trained personnel. In this study, a multiplex LAMP assay was optimized, which showed reliable and efficient diagnostic performance for detecting influenza A, influenza B, and SARS-CoV-2. The assay could diagnose both influenza and SARS-CoV-2 within an hour without requiring complex equipment. Additionally, it showed similar performance to commercial qPCR kits, making it a useful alternative to commercial qPCR kits that may not be suitable for use in the field.
In this study, we have demonstrated that developing a multiplex LAMP assay for detecting influenza and SARS-CoV-2 simultaneously could have some benefits. Our results suggest that the developed LAMP kit could provide a more comprehensive diagnosis of respiratory infections, especially during flu season when both viruses may be circulating. The ability to diagnose both influenza and SARS-CoV-2 using a single test could reduce the need for multiple tests and save time and resources. Additionally, the LAMP kit we developed could be used in resource-limited settings, where access to sophisticated equipment is limited, and could provide faster results compared to traditional PCR assays. This could be particularly useful in outbreak situations where quick and accurate diagnosis is crucial for effective control and management. Overall, our findings suggest that a multiplex LAMP assay has the potential to be a valuable tool for diagnosing respiratory infections, and could complement existing diagnostic methods in both clinical and public health settings.
5. Conclusions
In conclusion, the FluA/FluB/SARS-CoV-2 multiplex LAMP assay shows a similar level of sensitivity to the Allplex™ SARS-CoV-2/FluA/FluB/RSV assay and commercial RT-qPCR in clinical testing, although the FluA/FluB/SARS-CoV-2 multiplex LAMP assay has a higher detection limit than that of the commercial RT-qPCR kit. Furthermore, the FluA/FluB/SARS-CoV-2 multiplex LAMP assay can detect multiple strains of SARS-CoV-2, including Delta and Omicron, without cross-reactivity to other respiratory viruses and demonstrates excellent specificity for uninfected samples. Additionally, the high kappa values obtained in the concordance assays using the AllplexTM SARS-CoV-2/FluA/FluB/RSV assay suggest that the FluA/FluB/SARS-CoV-2 multiplex LAMP assay can serve as an alternative to this commercial kit.
Conceptualization, C.S.L.; funding acquisition, C.S.L. and W.S.J.; investigation, H.J. and S.P.; methodology, H.J. and C.S.L.; resources, S.P. and J.L.; supervision, W.S.J.; validation, S.P. and J.L.; visualization, H.J.; writing—original draft, H.J. and W.S.J.; writing—reviewing and editing, W.S.J. All authors have read and agreed to the published version of the manuscript.
The study was conducted in accordance with the guidelines of the Declaration of Helsinki and was approved by the Institutional Review Board of Korea University Guro Hospital (IRB No. 2021GR0547, approval date: 2021.11.03).
Patient consent was waived by the Institutional Review Board of Korea University Guro Hospital because the identities of the subjects were completely anonymous, and there was minimal risk involved in the study.
The authors declare that all related data are available from the corresponding author upon reasonable request.
The authors declare no conflict of interest.
Footnotes
Disclaimer/Publisher’s Note: The statements, opinions and data contained in all publications are solely those of the individual author(s) and contributor(s) and not of MDPI and/or the editor(s). MDPI and/or the editor(s) disclaim responsibility for any injury to people or property resulting from any ideas, methods, instructions or products referred to in the content.
Figure 1. Optimization of the FluA/FluB/SARS-CoV-2 multiplex LAMP primer set. (A) Different concentration ratios of influenza A, influenza B, SARS-CoV-2, and internal control LAMP primer sets (No. 1–5) for cultured influenza A H1N1-, influenza B-, and SARS-CoV-2 (NCCP43346)-spiked NP RNA sample. (B) Performance of FluA/FluB/SARS-CoV-2 multiplex LAMP assay for influenza A H1N1, H1N1pdm09, H3N2, influenza B, SARS-CoV-2 (NCCP43346), and normal NP RNA.
The FluA/FluB/SARS-CoV-2 multiplex LAMP assay primer sets used in this study.
Target | Name | Sequence (5′-3′) | µM |
---|---|---|---|
Influenza A | IAV F3 | GAC TKG AAR RTG TCT TTG C | 1 |
IAV B3 | TGT TRT TYG GGT CYC CAT T | 1 | |
IAV FIP | TTA GTC AGA GGT GAC ARR ATT GCA GAT CTT GAG GCT CTC | 32 | |
IAV BIP | TTG TKT TCA CGC TCA CCG TGT TTG GAC AAA GCG TCT ACG | 32 | |
IAV BLP | CMA GTG AGC GAG GAC TG | 10 | |
IAV FLP | GTC TTG TCT TTA GCC A | 4 | |
IAV FLP probe 1 | [FAM]-CGG GCC CGT ACA AAG GGA ACA CCC ACA CTC CGG TCT TGT CTT TAG CCA | 6 | |
Influenza B | IBV F3 | GAG CTG CCT ATG AAG ACC | 1 |
IBV B3 | CGT CTC CAC CTA CTT CGT | 1 | |
IBV FIP | GAA CAT GGA AAC CCT TGC ATT TTA AGT TTT GTC TGC ATT AAC AGG C | 32 | |
IBV BIP | GAA CAG RTR GAA GGA ATG GGR GCG ATC TGG TCA TTG GAG CC | 32 | |
IBV BLP | TGC TGA TCT AGG CTT GAA TTC TGT | 10 | |
IBV FLP | AGC TCT GAT GTC CAT CAA GCT CC | 4 | |
IBV FLP probe 1 | [TEX]-CGG GCC CGT ACA AAG GGA ACA CCC ACA CTC CGA GCT CTG ATG TCC ATC AAG CTC C | 6 | |
SARS-CoV-2 |
RdRP F3 | CCG ATA AGT ATG TCC GCA AT | 4 |
RdRP B3 | GCT TCA GAC ATA AAA ACA TTG T | 4 | |
RdRP FIP | ATG CGT AAA ACT CAT TCA CAA AGT CCA ACA CAG ACT TTA TGA GTG TC | 32 | |
RdRP BIP | TGA TAC TCT CTG ACG ATG CTG TTT AAA GTT CTT TAT GCT AGC CAC | 32 | |
RdRP LF | TGT GTC AAC ATC TCT ATT TCT ATA G | 10 | |
RdRP LB | TCA ATA GCA CTT ATG CAT CTC AAG G | 4 | |
RdRP LB probe 1 | [HEX]-CGG GCC CGT ACA AAG GGA ACA CCC ACA CTC CGT CAA TAG CAC TTA TGC ATC TCA AGG | 6 | |
Human |
IC F3 | AGT ACC CCA TCG AGC ACG | 4 |
IC B3 | AGC CTG GAT AGC AAC GTA CA | 4 | |
IB FIP | GAG CCA CAC GCA GCT CAT TGT ATC ACC AAC TGG GAC GAC A | 32 | |
IC BIP | CTG AAC CCC AAG GCC AAC CGG CTG GGG TGT TGA AGG TC | 32 | |
IC LF | TGT GGT GCC AGA TTT TCT CCA | 10 | |
IC LB | CGA GAA GAT GAC CCA GAT CAT GT | 6 | |
IC LB Probe 2 | [CY5]-GTC AGT GCA GGC TCC CGT GTT AGG ACG AGG GTA GGC GAG AAG ATG ACC CAG ATC ATG T | 4 | |
Quencher probe 1 | GAG TGT GGG TGT TCC CTT TGT ACG GGC CCG-BHQ1 | ||
Quencher probe 2 | CCT ACC CTC GTC CTA ACA CGG GAG CCT GCA CTG AC-BHQ2 |
The sensitivity of the FluA/FluB/SARS-CoV-2 multiplex LAMP assay according to different concentration ratios of the influenza A, influenza B, SARS-CoV-2, and internal control LAMP primer sets.
No | Ratio of Primer Set Mixture |
FluA/FluB/SARS-CoV-2 Multiplex LAMP Assay |
||||
---|---|---|---|---|---|---|
Channels | FAM | Texas Red | HEX | Cy5 | ||
Targets | Flu A H1N1 | Flu B | SARS-CoV-2 | IC-ACTB | ||
1 | 1:1:1:0.2 | N/A | 7.69 | 9.90 | N/A | |
2 | 1:0.5:1:0.2 | 9.27 | 9.44 | 9.08 | 24.46 | |
3 | 1:0.5:0.5:0.2 | 7.92 | 8.21 | N/A | 27.03 | |
4 | 0.5:0.5:1:0.2 | N/A | 10.12 | 9.23 | 23.93 | |
5 | 1:1:0.5:0.2 | 10.76 | 10.92 | 10.01 | 23.39 |
N/A, not available.
Performance of FluA/FluB/SARS-CoV-2 multiplex LAMP assay for influenza A (H1N1, H1N1pdm09, H3N2), influenza B, SARS-CoV-2, and NP RNA samples.
Sample | Concentration | FluA/FluB/SARS-CoV-2 Multiplex LAMP Assay |
|||
---|---|---|---|---|---|
FluA | FluB | RdRP | IC-ACTB | ||
Influenza A H1N1 | 2 × 105 TCID50/mL | 8.28 | N/A | N/A | N/A |
Influenza A H1N1pdm09 | 7 × 1012 TCID50/mL | 9.22 | N/A | N/A | N/A |
Influenza A H3N2 | 2 × 105 TCID50/mL | 10.17 | N/A | N/A | N/A |
Influenza B | 7 × 105 TCID50/mL | N/A | 8.52 | N/A | 27.27 |
SARS-CoV-2 | 1 × 102 PFU/mL | N/A | N/A | 11.03 | N/A |
NP RNA sample | N/A | N/A | N/A | 25.47 |
N/A, not available.
Limit of detection test for Allplex™ SARS-CoV-2/FluA/FluB/RSV and FluA/FluB/SARS-CoV-2 multiplex LAMP assays for influenza A H1N1, H1N1pdm09, H3N2, influenza B, and SARS-CoV-2 RNA samples.
Virus | TCID50/mL | Allplex™ SARS-CoV-2/Flua/Flub/RSV Assay | FluA/FluB/SARS-CoV-2 Multiplex LAMP Assay | ||||||
---|---|---|---|---|---|---|---|---|---|
FluA | FluB | CoV-2 | IC | FluA | FluB | CoV-2 | IC | ||
Cycle Threshold Values (Ct Values) | Cycle Threshold Values (Ct Values) | ||||||||
Influenza A H1N1 | 2 × 106 | 18.83 ± 0.17 | N/A | N/A | 26.65 ± 0.24 | 8.55 ± 0.19 | N/A | N/A | N/A |
2 × 105 | 21.05 ± 0.08 | N/A | N/A | 30.99 ± 0.30 | 10.42 ± 0.19 | N/A | N/A | N/A | |
2 × 104 | 24.01 ± 0.19 | N/A | N/A | 34.84 ± 0.04 | 13.15 ± 0.86 | N/A | N/A | N/A | |
2 × 103 | 30.85 ± 0.51 | N/A | N/A | N/A | N/A | N/A | N/A | N/A | |
2 × 102 | 37.32 ± 0.64 | N/A | N/A | N/A | N/A | N/A | N/A | N/A | |
2 × 101 | N/A | N/A | N/A | N/A | N/A | N/A | N/A | N/A | |
2 × 100 | N/A | N/A | N/A | N/A | N/A | N/A | N/A | N/A | |
Influenza A H1N1pdm09 | 7 × 1012 | 18.05 ± 0.07 | N/A | N/A | 26.28 ± 0.07 | 8.95 ± 0.40 | N/A | N/A | N/A |
7 × 1011 | 22.03 ± 0.30 | N/A | N/A | 29.94 ± 1.10 | 10.10 ± 0.61 | N/A | N/A | N/A | |
7 × 1010 | 25.81 ± 0.33 | N/A | N/A | 34.89 ± 1.34 | 13.4 ± 2.0 | N/A | N/A | N/A | |
7 × 109 | 30.75 ± 0.51 | N/A | N/A | N/A | N/A | N/A | N/A | N/A | |
7 × 108 | 36.29 ± 0.55 | N/A | N/A | N/A | N/A | N/A | N/A | N/A | |
7 × 107 | N/A | N/A | N/A | N/A | N/A | N/A | N/A | N/A | |
7 × 106 | N/A | N/A | N/A | N/A | N/A | N/A | N/A | N/A | |
Influenza A H3N2 | 2 × 106 | 18.11 ± 0.02 | N/A | N/A | 27.37 ± 0.29 | 9.29 ± 0.10 | N/A | N/A | N/A |
2 × 105 | 22.52 ± 0.10 | N/A | N/A | 32.31 ± 0.74 | 10.28 ± 0.67 | N/A | N/A | N/A | |
2 × 104 | 26.93 ± 0.33 | N/A | N/A | 35.46 ± 0.65 | 11.92 ± 0.70 | N/A | N/A | N/A | |
2 × 103 | 31.62 ± 0.48 | N/A | N/A | N/A | N/A | N/A | N/A | N/A | |
2 × 102 | 35.19 ± 0.65 | N/A | N/A | N/A | N/A | N/A | N/A | N/A | |
2 × 101 | N/A | N/A | N/A | N/A | N/A | N/A | N/A | N/A | |
2 × 100 | N/A | N/A | N/A | N/A | N/A | N/A | N/A | N/A | |
Influenza B | 7 × 106 | N/A | 17.24 ± 0.14 | N/A | 27.21 ± 0.40 | N/A | 7.21 ± 0.08 | N/A | 21.90 ± 1.94 |
7 × 105 | N/A | 19.36 ± 0.45 | N/A | 31.23 ± 1.08 | N/A | 8.12 ± 0.08 | N/A | 26.13 ± 0.30 | |
7 × 104 | N/A | 23.28 ± 0.65 | N/A | 34.93 ± 0.87 | N/A | 9.48 ± 0.12 | N/A | 28.69 ± 0.46 | |
7 × 103 | N/A | 27.11 ± 1.18 | N/A | N/A | N/A | 10.81 ± 0.18 | N/A | 35.37 ± 1.03 | |
7 × 102 | N/A | 32.49 ± 0.70 | N/A | N/A | N/A | 13.27 ± 1.38 | N/A | N/A | |
7 × 101 | N/A | 36.16 ± 1.08 | N/A | N/A | N/A | N/A | N/A | N/A | |
7 × 100 | N/A | N/A | N/A | N/A | N/A | N/A | N/A | N/A | |
SARS-CoV-2 (NCCP43346) | PFU/mL | ||||||||
1 × 100 | N/A | N/A | 27.10 ± 0.08 | 30.86 ± 0.06 | N/A | N/A | 10.13 ± 0.11 | N/A | |
1 × 10−1 | N/A | N/A | 33.73 ± 0.67 | 34.62 ± 0.25 | N/A | N/A | 11.03 ± 0.04 | N/A | |
1 × 10−2 | N/A | N/A | 37.07 ± 1.03 | N/A | N/A | N/A | N/A | N/A | |
1 × 10−3 | N/A | N/A | N/A | N/A | N/A | N/A | N/A | N/A | |
1 × 10−4 | N/A | N/A | N/A | N/A | N/A | N/A | N/A | N/A | |
1 × 10−5 | N/A | N/A | N/A | N/A | N/A | N/A | N/A | N/A | |
1 × 10−6 | N/A | N/A | N/A | N/A | N/A | N/A | N/A | N/A |
N/A, not available.
Comparison of performance of the FluA/FluB/SARS-CoV-2 multiplex LAMP assay and Allplex™ SARS-CoV-2/FluA/FluB/RSV assay for clinical samples.
Clinical Samples | Assay | P/N | Sensitivity |
Specificity |
PPV |
NPV |
---|---|---|---|---|---|---|
Influenza A |
Allplex™ SARS-CoV-2/FluA/FluB/RSV assay | 111/6 | 94.87 |
99.03 |
99.10 |
94.44 |
FluA/FluB/SARS-CoV-2 multiplex LAMP assay | 106/11 | 90.60 |
100 |
100 |
90.27 |
|
Influenza B |
Allplex™ SARS-CoV-2/FluA/FluB/RSV assay | 54/7 | 88.52 |
99.03 |
98.18 |
93.58 |
FluA/FluB/SARS-CoV-2 multiplex LAMP assay | 53/8 | 86.89 |
100 |
100 |
92.72 |
|
SARS-CoV-2 |
Allplex™ SARS-CoV-2/FluA/FluB/RSV assay | 94/2 | 97.92 |
100 |
100 |
98.08 |
FluA/FluB/SARS-CoV-2 multiplex LAMP assay | 95/1 | 98.96 |
100 |
100 |
99.03 |
“P” and “N” indicate the positive and negative of the reaction, respectively. PPV, positive predictive value, NPV: negative predictive value. CI: confidence interval.
Kappa values for calculating agreement between the FluA/FluB/SARS-CoV-2 multiplex LAMP assay and Allplex™ SARS-CoV-2/FluA/FluB/RSV assay.
Allplex™ SARS-CoV-2/FluA/FluB/RSV Assay | ||||||||||
---|---|---|---|---|---|---|---|---|---|---|
Influenza A | Influenza B | SARS-CoV-2 | ||||||||
P | N | Total | P | N | Total | P | N | Total | ||
FluA/FluB/SARS-CoV-2 multiplex LAMP assay | P | 100 | 6 | 106 | 46 | 7 | 53 | 93 | 2 | 95 |
N | 11 | 102 | 113 | 8 | 102 | 110 | 1 | 102 | 103 | |
Total | 111 | 108 | 219 | 54 | 109 | 163 | 94 | 104 | 198 | |
Cohen’s kappa index |
0.845 (<0.001) | 0.791 (<0.001) | 0.970 (<0.001) | |||||||
Strength of agreement | Almost perfect | Substantial | Almost perfect |
P, positive reaction; N, negative reaction.
Cross-reactivity of the FluA/FluB/SARS-CoV-2 multiplex LAMP assay against other human infectious viruses.
Tested Clinical Samples | FluA/FluB/SARS-CoV-2 Multiplex LAMP Assay (Positive No./Test No.) | |||
---|---|---|---|---|
Flu A | Flu B | SARS-CoV-2 | IC-ACTB | |
CoV 229E | 0/3 | 0/3 | 0/3 | 3/0 |
CoV NL63 | 0/3 | 0/3 | 0/3 | 3/0 |
CoV OC43 | 0/3 | 0/3 | 0/3 | 3/0 |
HEV | 0/3 | 0/3 | 0/3 | 3/0 |
AdV | 0/3 | 0/3 | 0/3 | 3/0 |
PIV | 0/3 | 0/3 | 0/3 | 3/0 |
MPV | 0/3 | 0/3 | 0/3 | 3/0 |
HboV | 0/3 | 0/3 | 0/3 | 3/0 |
HRV | 0/3 | 0/3 | 0/3 | 3/0 |
RSV A | 0/3 | 0/3 | 0/3 | 3/0 |
RSV B | 0/3 | 0/3 | 0/3 | 0/3 |
The performance of FluA/FluB/SARS-CoV-2 multiplex LAMP assay for one SARS-CoV-2 wild-type sample and 19 SARS-CoV-2 mutant samples.
Virus | Genotype |
Lineage | NCCP No. | FluA/FluB/SARS-CoV-2 Multiplex LAMP Assay | |||
---|---|---|---|---|---|---|---|
Flu A | Flu B | RdRP | IC-ACTB | ||||
Cycle Threshold Values (Ct Values) | |||||||
SARS-CoV-2 | GV (wild-type) | B.1.177 | 43346 | N/A | N/A | 7.12 | N/A |
GRY | B.1.1.7 | 43381 | N/A | N/A | 7.29 | N/A | |
GH | B.1.351 | 43382 | N/A | N/A | 8.01 | N/A | |
GR | P.2 | 43383 | N/A | N/A | 7.03 | N/A | |
GH | B.1.427 | 43384 | N/A | N/A | 8.07 | N/A | |
GH | B.1.429 | 43385 | N/A | N/A | 8.35 | N/A | |
G | B.1.525 | 43386 | N/A | N/A | 12.78 | N/A | |
GH | B.1.526 | 43387 | N/A | N/A | 10.89 | N/A | |
GR | P.1 | 43388 | N/A | N/A | 7.12 | N/A | |
G | B.1.617.1 | 43389 | N/A | N/A | 8.06 | N/A | |
G | B.1.620 | 43404 | N/A | N/A | 7.33 | N/A | |
GK | B.1.617.2 | 43405 | N/A | N/A | 8.63 | N/A | |
GH | B.1.621 | 43407 | N/A | N/A | 8.35 | N/A | |
GRA | BA.1 | 43408 | N/A | N/A | 7.15 | N/A | |
GRA | BA.1.1 | 43411 | N/A | N/A | 8.20 | N/A | |
GRA | BA.2 | 43412 | N/A | N/A | 9.08 | N/A | |
GRA | BA.2.12.1 | 43423 | N/A | N/A | 8.19 | N/A | |
GRA | BA.2.3 | 43424 | N/A | N/A | 10.82 | N/A | |
GRA | BA.4 | 43425 | N/A | N/A | 9.75 | N/A | |
GRA | BA.5 | 43426 | N/A | N/A | 10.80 | N/A |
* GSAID: Global Initiative for Sharing All Influenza Data; N/A, not available.
References
1. Javanian, M.; Barary, M.; Ghebrehewet, S.; Koppolu, V.; Vasigala, V.; Ebrahimpour, S. A brief review of influenza virus infection. J. Med. Virol.; 2021; 93, pp. 4638-4646. [DOI: https://dx.doi.org/10.1002/jmv.26990]
2. World Health Organization. COVID-19 Weekly Epidemiological Update; 127th ed. World Health Organization: Geneva, Switzerland, 2023; Available online: https://www.who.int/emergencies/diseases/novel-coronavirus-2019/situation-reports/ (accessed on 25 January 2023).
3. Lampejo, T. The impact of the COVID-19 pandemic on the global burden of influenza. J. Med. Virol.; 2022; 94, pp. 2357-2359. [DOI: https://dx.doi.org/10.1002/jmv.27653] [PubMed: https://www.ncbi.nlm.nih.gov/pubmed/35146774]
4. Heiskanen, A.; Galipeau, Y.; Little, J.; Mortimer, L.; Ramotar, K.; Langlois, M.; Cooper, C.L. Seasonal respiratory virus circulation was diminished during the COVID-19 Ppandemic. Influenza Other Respir. Viruses; 2023; 17, e13065. [DOI: https://dx.doi.org/10.1111/irv.13065]
5. Dhanasekaran, V.; Sullivan, S.; Edwards, K.M.; Xie, R.; Khvorov, A.; Valkenburg, S.A.; Cowling, B.J.; Barr, I.G. Human seasonal influenza under COVID-19 and the potential consequences of influenza lineage elimination. Nat. Commun.; 2022; 13, 1721. [DOI: https://dx.doi.org/10.1038/s41467-022-29402-5] [PubMed: https://www.ncbi.nlm.nih.gov/pubmed/35361789]
6. Mboumba Bouassa, R.-S.; Tonen-Wolyec, S.; Veyer, D.; Péré, H.; Bélec, L. Analytical performances of the AMPLIQUICK® respiratory triplex assay for simultaneous detection and differentiation of SARS-CoV-2, influenza A/B and respiratory syncytial viruses in respiratory specimens. PLoS ONE; 2022; 17, e0262258. [DOI: https://dx.doi.org/10.1371/journal.pone.0262258]
7. Dinç, H.Ö.; Karabulut, N.; Alaçam, S.; Uysal, H.K.; Daşdemir, F.O.; Önel, M.; Tuyji Tok, Y.; Sirekbasan, S.; Agacfidan, A.; Gareayaghi, N. et al. Evaluation of the diagnostic performance of a SARS-CoV-2 and influenza A/B combo rapid antigen test in respiratory samples. Diagnostics; 2023; 13, 972. [DOI: https://dx.doi.org/10.3390/diagnostics13050972]
8. Mautner, L.; Baillie, C.-K.; Herold, H.M.; Volkwein, W.; Guertler, P.; Eberle, U.; Ackermann, N.; Sing, A.; Pavlovic, M.; Goerlich, O. et al. Rapid point-of-care detection of SARS-CoV-2 using reverse transcription loop-mediated isothermal amplification (RT-LAMP). Virol. J.; 2020; 17, 160. [DOI: https://dx.doi.org/10.1186/s12985-020-01435-6] [PubMed: https://www.ncbi.nlm.nih.gov/pubmed/33087160]
9. Mak, G.C.; Cheng, P.K.; Lau, S.S.; Wong, K.K.; Lau, C.; Lam, E.T.; Chan, R.C.; Tsang, D.N. Evaluation of rapid antigen test for detection of SARS-CoV-2 virus. J. Clin. Virol.; 2020; 129, 104500. [DOI: https://dx.doi.org/10.1016/j.jcv.2020.104500] [PubMed: https://www.ncbi.nlm.nih.gov/pubmed/32585619]
10. Notomi, T. Loop-mediated isothermal amplification of DNA. Nucleic. Acids Res.; 2000; 28, e63. [DOI: https://dx.doi.org/10.1093/nar/28.12.e63]
11. Nagamine, K.; Hase, T.; Notomi, T. Accelerated reaction by loop-mediated isothermal amplification using loop primers. Mol. Cell Probes; 2002; 16, pp. 223-229. [DOI: https://dx.doi.org/10.1006/mcpr.2002.0415]
12. Gadkar, V.J.; Goldfarb, D.M.; Gantt, S.; Tilley, P.A.G. Real-time detection and monitoring of loop mediated amplification (LAMP) reaction using self-quenching and de-quenching fluorogenic probes. Sci. Rep.; 2018; 8, 5548. [DOI: https://dx.doi.org/10.1038/s41598-018-23930-1]
13. Londono-Avendano, M.A.; Libreros, G.; Osorio, L.; Parra, B. A rapid RT-LAMP assay for SARS-CoV-2 with colorimetric detection assisted by a mobile application. Diagnostics; 2022; 12, 848. [DOI: https://dx.doi.org/10.3390/diagnostics12040848] [PubMed: https://www.ncbi.nlm.nih.gov/pubmed/35453896]
14. Kim, S.; Kim, J.H.; Kim, S.; Park, J.S.; Cha, B.S.; Lee, E.S.; Han, J.; Shin, J.; Jang, Y.; Park, K.S. Loop-mediated isothermal amplification-based nucleic acid lateral flow assay for the specific and multiplex detection of genetic markers. Anal. Chim. Acta.; 2022; 1205, 339781. [DOI: https://dx.doi.org/10.1016/j.aca.2022.339781] [PubMed: https://www.ncbi.nlm.nih.gov/pubmed/35414391]
15. Cao, D.; Hu, L.; Lin, M.; Li, M.; Ye, Z.; Sun, H.; Huang, J.; Yang, H.; Tian, J. Real-time fluorescence loop-mediated isothermal amplification (LAMP) for rapid and reliable diagnosis of pulmonary tuberculosis. J. Microbiol. Methods; 2015; 109, pp. 74-78. [DOI: https://dx.doi.org/10.1016/j.mimet.2014.12.013] [PubMed: https://www.ncbi.nlm.nih.gov/pubmed/25533219]
16. Shirato, K.; Semba, S.; El-Kafrawy, S.A.; Hassan, A.M.; Tolah, A.M.; Takayama, I.; Kageyama, T.; Notomi, T.; Kamitani, W.; Matsuyama, S. et al. Development of fluorescent reverse transcription loop-mediated isothermal amplification (RT-LAMP) using quenching probes for the detection of the Middle East Respiratory Syndrome Coronavirus. J. Virol. Methods; 2018; 258, pp. 41-48. [DOI: https://dx.doi.org/10.1016/j.jviromet.2018.05.006]
17. Kubota, R.; Alvarez, A.M.; Su, W.W.; Jenkins, D.M. Jenkins FRET-based assimilating probe for sequence-specific real-time monitoring of loop-mediated isothermal amplification (LAMP). Biol. Eng. Trans.; 2011; 4, pp. 81-100. [DOI: https://dx.doi.org/10.13031/2013.38509]
18. Kim, M.-J.; Kim, H.-Y. Direct duplex real-time loop mediated isothermal amplification assay for the simultaneous detection of cow and goat species origin of milk and yogurt products for field use. Food Chem.; 2018; 246, pp. 26-31. [DOI: https://dx.doi.org/10.1016/j.foodchem.2017.11.014]
19. Jang, W.S.; Lim, D.H.; Nam, J.; Mihn, D.-C.; Sung, H.W.; Lim, C.S.; Kim, J. Development of a multiplex isothermal amplification molecular diagnosis method for on-site diagnosis of influenza. PLoS ONE; 2020; 15, e0238615. [DOI: https://dx.doi.org/10.1371/journal.pone.0238615]
20. Jang, W.S.; Lim, D.H.; Yoon, J.; Kim, A.; Lim, M.; Nam, J.; Yanagihara, R.; Ryu, S.-W.; Jung, B.K.; Ryoo, N.-H. et al. Development of a multiplex loop-mediated isothermal amplification (LAMP) assay for on-site diagnosis of SARS CoV-2. PLoS ONE; 2021; 16, e0248042. [DOI: https://dx.doi.org/10.1371/journal.pone.0248042]
21. Landis, J.R.; Koch, G.G. The measurement of observer agreement for categorical data. Biometrics; 1977; 33, pp. 159-174. [DOI: https://dx.doi.org/10.2307/2529310]
22. Lau, Y.L.; Ismail, I.B.; Mustapa, N.I.B.; Lai, M.Y.; Tuan Soh, T.S.; Haji Hassan, A.; Peariasamy, K.M.; Lee, Y.L.; Abdul Kahar, M.K.B.; Chong, J. et al. Development of a reverse transcription recombinase polymerase amplification assay for rapid and direct visual detection of severe acute respiratory syndrome coronavirus 2 (SARS-CoV-2). PLoS ONE; 2021; 16, e0245164. [DOI: https://dx.doi.org/10.1371/journal.pone.0245164]
23. Inaba, M.; Higashimoto, Y.; Toyama, Y.; Horiguchi, T.; Hibino, M.; Iwata, M.; Imaizumi, K.; Doi, Y. Diagnostic accuracy of LAMP versus PCR over the course of SARS-CoV-2 infection. Int. J. Infect. Dis.; 2021; 107, pp. 195-200. [DOI: https://dx.doi.org/10.1016/j.ijid.2021.04.018] [PubMed: https://www.ncbi.nlm.nih.gov/pubmed/33862213]
24. Nandi, S.; Lambe, U.; Sawant, S.; Gohil, T.; Deshpande, J. Development of a RT-LAMP assay for detection of SARS-CoV-2. Indian J. Med. Res.; 2022; 155, 148. [DOI: https://dx.doi.org/10.4103/ijmr.IJMR_713_21] [PubMed: https://www.ncbi.nlm.nih.gov/pubmed/35313427]
25. Brandsma, E.; Verhagen, H.J.M.P.; van de Laar, T.J.W.; Claas, E.C.J.; Cornelissen, M.; van den Akker, E. Rapid, sensitive, and specific severe acute respiratory syndrome coronavirus 2 detection: A multicenter comparison between standard quantitative reverse-transcriptase polymerase chain reaction and CRISPR-based DETECTR. J. Infect. Dis.; 2021; 223, pp. 206-213. [DOI: https://dx.doi.org/10.1093/infdis/jiaa641]
26. Ahn, S.J.; Baek, Y.H.; Lloren, K.K.S.; Choi, W.-S.; Jeong, J.H.; Antigua, K.J.C.; Kwon, H.; Park, S.-J.; Kim, E.-H.; Kim, Y. et al. Rapid and simple colorimetric detection of multiple influenza viruses infecting humans using a reverse transcriptional loop-mediated isothermal amplification (RT-LAMP) diagnostic platform. BMC Infect. Dis.; 2019; 19, 676. [DOI: https://dx.doi.org/10.1186/s12879-019-4277-8]
27. Takayama, I.; Nakauchi, M.; Takahashi, H.; Oba, K.; Semba, S.; Kaida, A.; Kubo, H.; Saito, S.; Nagata, S.; Odagiri, T. et al. Development of real-time fluorescent reverse transcription loop-mediated isothermal amplification assay with quenching primer for influenza virus and respiratory syncytial virus. J. Virol. Methods; 2019; 267, pp. 53-58. [DOI: https://dx.doi.org/10.1016/j.jviromet.2019.02.010] [PubMed: https://www.ncbi.nlm.nih.gov/pubmed/30831121]
28. Zhao, Y.; Chen, F.; Li, Q.; Wang, L.; Fan, C. Isothermal amplification of nucleic acids. Chem. Rev.; 2015; 115, pp. 12491-12545. [DOI: https://dx.doi.org/10.1021/acs.chemrev.5b00428]
29. Heskin, J.; Pallett, S.J.C.; Al-Hindawi, A.; Davies, G.W.; Rayment, M.; Mughal, N.; Randell, P.; Jones, R.; Moore, L.S.P. Evaluating the performance characteristics of five lateral flow assays for the detection of the SARS-CoV-2 nucleocapsid antigen. Sci. Rep.; 2022; 12, 8811. [DOI: https://dx.doi.org/10.1038/s41598-022-12805-1]
30. Irsara, C.; Egger, A.E.; Prokop, W.; Nairz, M.; Loacker, L.; Sahanic, S.; Pizzini, A.; Sonnweber, T.; Mayer, W.; Schennach, H. et al. Evaluation of four commercial, fully automated SARS-CoV-2 antibody tests suggests a revision of the siemens SARS-CoV-2 IgG assay. Clin. Chem. Lab. Med.; 2021; 59, pp. 1143-1154. [DOI: https://dx.doi.org/10.1515/cclm-2020-1758]
You have requested "on-the-fly" machine translation of selected content from our databases. This functionality is provided solely for your convenience and is in no way intended to replace human translation. Show full disclaimer
Neither ProQuest nor its licensors make any representations or warranties with respect to the translations. The translations are automatically generated "AS IS" and "AS AVAILABLE" and are not retained in our systems. PROQUEST AND ITS LICENSORS SPECIFICALLY DISCLAIM ANY AND ALL EXPRESS OR IMPLIED WARRANTIES, INCLUDING WITHOUT LIMITATION, ANY WARRANTIES FOR AVAILABILITY, ACCURACY, TIMELINESS, COMPLETENESS, NON-INFRINGMENT, MERCHANTABILITY OR FITNESS FOR A PARTICULAR PURPOSE. Your use of the translations is subject to all use restrictions contained in your Electronic Products License Agreement and by using the translation functionality you agree to forgo any and all claims against ProQuest or its licensors for your use of the translation functionality and any output derived there from. Hide full disclaimer
© 2023 by the authors. Licensee MDPI, Basel, Switzerland. This article is an open access article distributed under the terms and conditions of the Creative Commons Attribution (CC BY) license (https://creativecommons.org/licenses/by/4.0/). Notwithstanding the ProQuest Terms and Conditions, you may use this content in accordance with the terms of the License.
Abstract
Influenza and coronaviruses cause highly contagious respiratory diseases that cause millions of deaths worldwide. Public health measures implemented during the current coronavirus disease (COVID-19) pandemic have gradually reduced influenza circulation worldwide. As COVID-19 measures have relaxed, it is necessary to monitor and control seasonal influenza during this COVID-19 pandemic. In particular, the development of rapid and accurate diagnostic methods for influenza and COVID-19 is of paramount importance because both diseases have significant public health and economic impacts. To address this, we developed a multi-loop-mediated isothermal amplification (LAMP) kit capable of simultaneously detecting influenza A/B and SARS-CoV-2. The kit was optimized by testing various ratios of primer sets for influenza A/B (FluA/FluB) and SARS-CoV-2 and internal control (IC). The FluA/FluB/SARS-CoV-2 multiplex LAMP assay showed 100% specificity for uninfected clinical samples and sensitivities of 90.6%, 86.89%, and 98.96% for LAMP kits against influenza A, influenza B, and SARS-CoV-2 clinical samples, respectively. Finally, the attribute agreement analysis for clinical tests indicated substantial agreement between the multiplex FluA/FluB/SARS-CoV-2/IC LAMP and commercial AllplexTM SARS-CoV-2/FluA/FluB/RSV assays.
You have requested "on-the-fly" machine translation of selected content from our databases. This functionality is provided solely for your convenience and is in no way intended to replace human translation. Show full disclaimer
Neither ProQuest nor its licensors make any representations or warranties with respect to the translations. The translations are automatically generated "AS IS" and "AS AVAILABLE" and are not retained in our systems. PROQUEST AND ITS LICENSORS SPECIFICALLY DISCLAIM ANY AND ALL EXPRESS OR IMPLIED WARRANTIES, INCLUDING WITHOUT LIMITATION, ANY WARRANTIES FOR AVAILABILITY, ACCURACY, TIMELINESS, COMPLETENESS, NON-INFRINGMENT, MERCHANTABILITY OR FITNESS FOR A PARTICULAR PURPOSE. Your use of the translations is subject to all use restrictions contained in your Electronic Products License Agreement and by using the translation functionality you agree to forgo any and all claims against ProQuest or its licensors for your use of the translation functionality and any output derived there from. Hide full disclaimer
Details



1 BK21 Graduate Program, Department of Biomedical Sciences, College of Medicine, Korea University, Seoul 02841, Republic of Korea;
2 Department of Laboratory Medicine, College of Medicine, Korea University Guro Hospital, Seoul 08308, Republic of Korea;
3 Emergency Medicine, College of Medicine, Korea University Guro Hospital, Seoul 08308, Republic of Korea