Abstract

Reusable high-resolution 2D dosimeters have a great potential for radiotherapy verification and quality assurance, not least due to the reduced cost and simpler calibration procedures implied by reusability. We have performed proof-of-concept measurements in both photon and proton beams with a reusable 2D dosimeter (5×7 cm2; 1 mm thickness) based on optically stimulated luminescence, using a readout setup scalable to 3D. Our data indicated a strong linear-energy-transfer quenching for this type of dosimeter.

Details

Title
Proof-of-concept of proton and photon measurements for large-area 2D dosimeters utilising optically stimulated luminescence
Author
Nyemann, Jacob S 1 ; Jensen, Mads L 1 ; Valdetaro, Lia B 2 ; Jensen, Morten B 3 ; Turtos, Rosana M 1 ; Julsgaard, Brian 4 ; Petersen, Jørgen B B 5 ; Muren, Ludvig P 3 ; Balling, Peter 4 

 Department of Physics and Astronomy, Aarhus University , Denmark 
 Department of Clinical Medicine, Aarhus University , Denmark; Danish Centre for Particle Therapy, Aarhus University Hospital , Denmark 
 Department of Clinical Medicine, Aarhus University , Denmark; Danish Centre for Particle Therapy, Aarhus University Hospital , Denmark; Department of Medical Physics, Aarhus University and Aarhus University Hospital , Denmark 
 Department of Physics and Astronomy, Aarhus University , Denmark; Interdisciplinary Nanoscience Center (iNANO), Aarhus University , Denmark 
 Department of Medical Physics, Aarhus University and Aarhus University Hospital , Denmark 
First page
012022
Publication year
2023
Publication date
Nov 2023
Publisher
IOP Publishing
ISSN
17426588
e-ISSN
17426596
Source type
Scholarly Journal
Language of publication
English
ProQuest document ID
2890412423
Copyright