Abstract

Background

Phase-contrast magnetic resonance imaging (PC-MRI) quantifies blood flow and velocity noninvasively. Challenges arise in neurovascular disorders due to small vessels. We evaluated the impact of voxel size, number of signal averages (NSA), and velocity encoding (VENC) on PC-MRI measurement accuracy and precision in a small-lumen vessel phantom.

Methods

We constructed an in vitro model with a constant flow rate using a 2.2-mm inner diameter plastic tube. A reservoir with a weight scale and timer was used as standard reference. Gradient-echo T1 weighted PC-MRI sequence was performed on a 3-T scanner with varying voxel size (2.5, 5.0, 7.5 mm3), NSA (1, 2, 3), and VENC (200, 300, 400 cm/s). We repeated measurements nine times per setting, calculating mean flow rate, maximum velocity, and least detectable difference (LDD).

Results

PC-MRI flow measurements were higher than standard reference values (mean ranging from 7.3 to 9.5 mL/s compared with 6.6 mL/s). Decreased voxel size improved accuracy, reducing flow rate measurements from 9.5 to 7.3 mL/s. The LDD for flow rate and velocity varied between 1 and 5%. The LDD for flow rate decreased with increased voxel size and NSA (p = 0.033 and 0.042). The LDD for velocity decreased with increased voxel size (p < 10-16). No change was observed when VENC varied.

Conclusions

PC-MRI overestimated flow. However, it has high precision in a small-vessel phantom with constant flow rate. Improved accuracy was obtained with increasing spatial resolution (smaller voxels). Improved precision was obtained with increasing signal-to-noise ratio (larger voxels and/or higher NSA).

Relevance statement

Phase-contrast MRI is clinically used in large vessels. To further investigate the possibility of using phase-contrast MRI for smaller intracranial vessels in neurovascular disorders, we need to understand how acquisition parameters affect phase-contrast MRI-measured flow rate and velocity in small vessels.

Key points

• PC-MRI measures flow and velocity in a small lumen phantom with high precision but overestimates flow rate.

• The precision of PC-MRI measurements matches the precision of standard reference for flow rate measurements.

• Optimizing PC-MRI settings can enhance accuracy and precision in flow rate and velocity measurements.

Details

Title
Effect of MRI acquisition parameters on accuracy and precision of phase-contrast measurements in a small-lumen vessel phantom
Author
Correia de Verdier, Maria 1   VIAFID ORCID Logo  ; Berglund, Johan 2 ; Wikström, Johan 1 

 Uppsala University, Department of Surgical Sciences, Section of Neuroradiology, Uppsala, Sweden (GRID:grid.8993.b) (ISNI:0000 0004 1936 9457) 
 Uppsala University, Department of Surgical Sciences, Section of Molecular Imaging and Medical Physics, Uppsala, Sweden (GRID:grid.8993.b) (ISNI:0000 0004 1936 9457) 
Pages
45
Publication year
2024
Publication date
Dec 2024
Publisher
Springer Nature B.V.
e-ISSN
25099280
Source type
Scholarly Journal
Language of publication
English
ProQuest document ID
2956005634
Copyright
© The Author(s) 2024. This work is published under http://creativecommons.org/licenses/by/4.0/ (the “License”). Notwithstanding the ProQuest Terms and Conditions, you may use this content in accordance with the terms of the License.