Abstract
Background
Phase-contrast magnetic resonance imaging (PC-MRI) quantifies blood flow and velocity noninvasively. Challenges arise in neurovascular disorders due to small vessels. We evaluated the impact of voxel size, number of signal averages (NSA), and velocity encoding (VENC) on PC-MRI measurement accuracy and precision in a small-lumen vessel phantom.
Methods
We constructed an in vitro model with a constant flow rate using a 2.2-mm inner diameter plastic tube. A reservoir with a weight scale and timer was used as standard reference. Gradient-echo T1 weighted PC-MRI sequence was performed on a 3-T scanner with varying voxel size (2.5, 5.0, 7.5 mm3), NSA (1, 2, 3), and VENC (200, 300, 400 cm/s). We repeated measurements nine times per setting, calculating mean flow rate, maximum velocity, and least detectable difference (LDD).
Results
PC-MRI flow measurements were higher than standard reference values (mean ranging from 7.3 to 9.5 mL/s compared with 6.6 mL/s). Decreased voxel size improved accuracy, reducing flow rate measurements from 9.5 to 7.3 mL/s. The LDD for flow rate and velocity varied between 1 and 5%. The LDD for flow rate decreased with increased voxel size and NSA (p = 0.033 and 0.042). The LDD for velocity decreased with increased voxel size (p < 10-16). No change was observed when VENC varied.
Conclusions
PC-MRI overestimated flow. However, it has high precision in a small-vessel phantom with constant flow rate. Improved accuracy was obtained with increasing spatial resolution (smaller voxels). Improved precision was obtained with increasing signal-to-noise ratio (larger voxels and/or higher NSA).
Relevance statement
Phase-contrast MRI is clinically used in large vessels. To further investigate the possibility of using phase-contrast MRI for smaller intracranial vessels in neurovascular disorders, we need to understand how acquisition parameters affect phase-contrast MRI-measured flow rate and velocity in small vessels.
Key points
• PC-MRI measures flow and velocity in a small lumen phantom with high precision but overestimates flow rate.
• The precision of PC-MRI measurements matches the precision of standard reference for flow rate measurements.
• Optimizing PC-MRI settings can enhance accuracy and precision in flow rate and velocity measurements.
You have requested "on-the-fly" machine translation of selected content from our databases. This functionality is provided solely for your convenience and is in no way intended to replace human translation. Show full disclaimer
Neither ProQuest nor its licensors make any representations or warranties with respect to the translations. The translations are automatically generated "AS IS" and "AS AVAILABLE" and are not retained in our systems. PROQUEST AND ITS LICENSORS SPECIFICALLY DISCLAIM ANY AND ALL EXPRESS OR IMPLIED WARRANTIES, INCLUDING WITHOUT LIMITATION, ANY WARRANTIES FOR AVAILABILITY, ACCURACY, TIMELINESS, COMPLETENESS, NON-INFRINGMENT, MERCHANTABILITY OR FITNESS FOR A PARTICULAR PURPOSE. Your use of the translations is subject to all use restrictions contained in your Electronic Products License Agreement and by using the translation functionality you agree to forgo any and all claims against ProQuest or its licensors for your use of the translation functionality and any output derived there from. Hide full disclaimer
Details

1 Uppsala University, Department of Surgical Sciences, Section of Neuroradiology, Uppsala, Sweden (GRID:grid.8993.b) (ISNI:0000 0004 1936 9457)
2 Uppsala University, Department of Surgical Sciences, Section of Molecular Imaging and Medical Physics, Uppsala, Sweden (GRID:grid.8993.b) (ISNI:0000 0004 1936 9457)