Content area

Abstract

During adaptive laboratory evolution experiments, any unexpected interruption in data monitoring or control could lead to the loss of valuable experimental data and compromise the integrity of the entire experiment. Most homemade mini-bioreactors are built employing microcontrollers such as Arduino. Although affordable, these platforms lack the robustness of the programmable logic controller (PLC), which enhances the safety and robustness of the control process. Here, we describe the design and validation of a PLC-controlled morbidostat, an innovative automated continuous-culture mini-bioreactor specifically created to study the evolutionary pathways to drug resistance in microorganisms. This morbidostat includes several improvements, both at the hardware and software level, for better online monitoring and a more robust operation. The device was validated employing Escherichia coli, exploring its adaptive evolution in the presence of didecyldimethylammonium chloride (DDAC), a quaternary ammonium compound widely used for its antimicrobial properties. E. coli was subjected to increasing concentrations of DDAC over 3 days. Our results demonstrated a significant increase in DDAC susceptibility, with evolved populations exhibiting substantial changes in their growth after exposure.

Details

1009240
Business indexing term
Location
Title
Design and Validation of a PLC-Controlled Morbidostat for Investigating Bacterial Drug Resistance
Author
Pedreira, Adrián 1   VIAFID ORCID Logo  ; Vázquez, José A 2   VIAFID ORCID Logo  ; Romanenko, Andrey 3   VIAFID ORCID Logo  ; García, Míriam R 4   VIAFID ORCID Logo 

 Biosystems and Bioprocess Engineering Group (Bio2Eng), Spanish National Research Council (IIM-CSIC), Rúa Eduardo Cabello 6, 36208 Vigo, Spain; [email protected]; Group of Recycling and Valorization of Waste Materials (REVAL), Spanish National Research Council (IIM-CSIC), Rúa Eduardo Cabello 6, 36208 Vigo, Spain; [email protected] 
 Group of Recycling and Valorization of Waste Materials (REVAL), Spanish National Research Council (IIM-CSIC), Rúa Eduardo Cabello 6, 36208 Vigo, Spain; [email protected] 
 CIENGIS Lda, Rua Pedro Nunes, 3030-199 Coimbra, Portugal; [email protected] 
 Biosystems and Bioprocess Engineering Group (Bio2Eng), Spanish National Research Council (IIM-CSIC), Rúa Eduardo Cabello 6, 36208 Vigo, Spain; [email protected] 
Publication title
Volume
11
Issue
8
First page
815
Publication year
2024
Publication date
2024
Publisher
MDPI AG
Place of publication
Basel
Country of publication
Switzerland
Publication subject
e-ISSN
23065354
Source type
Scholarly Journal
Language of publication
English
Document type
Journal Article
Publication history
 
 
Online publication date
2024-08-10
Milestone dates
2024-07-01 (Received); 2024-08-05 (Accepted)
Publication history
 
 
   First posting date
10 Aug 2024
ProQuest document ID
3097834359
Document URL
https://www.proquest.com/scholarly-journals/design-validation-plc-controlled-morbidostat/docview/3097834359/se-2?accountid=208611
Copyright
© 2024 by the authors. Licensee MDPI, Basel, Switzerland. This article is an open access article distributed under the terms and conditions of the Creative Commons Attribution (CC BY) license (https://creativecommons.org/licenses/by/4.0/). Notwithstanding the ProQuest Terms and Conditions, you may use this content in accordance with the terms of the License.
Last updated
2025-01-21
Database
ProQuest One Academic