1. Introduction
Seasonal frozen areas are widely distributed, covering 53% of the world’s land surface [1,2,3]. The frost heave and thaw settlement of frozen soil caused by seasonal temperature fluctuations endanger the safety of various engineering facilities in seasonally frozen areas. Current studies have found that phase transitions and the migration of internal moisture during the freeze–thaw process, along with the movement of the freezing–melting boundary (the ice front), are the main triggers of engineering failures in seasonally frozen regions [4]. Thus, accurately identifying the location of the ice front in frozen soil and understanding the internal moisture distribution is crucial for disaster prevention and reduction in seasonally frozen areas.
Currently, various measurement techniques, including ground-penetrating radar [5], CT techniques [6,7], the electric resistance measuring method [8], and portable nuclear magnetic resonance detectors [9], can be used to measure the composition distribution of frozen soil and monitor the movement of the freezing boundary. Bittelli et al. used dielectric spectroscopy to estimate the ice content in frozen porous media and derived the volume fractions of different phases in the medium based on the volume and dielectric constant of each phase [10]. Kunio et al. simultaneously measured the liquid water content and relative permittivity of various unsaturated soils at above-zero and subzero temperatures by using pulsed nuclear magnetic resonance (NMR) and time-domain reflectometry (TDR) [11]. Wen et al. used an NMR technique and thermal resistor temperature probe to characterize unfrozen water content and soil matric potential [12]. Generally, it is hard to achieve real-time and non-invasive measurements on the freezing interface of the frozen soil. Given the complexity and nature of the issue, finding a reliable approach to measure the composition distribution and monitor the movement of the freezing interface is still a critical challenge.
Electrical capacitance tomography (ECT) is one of the techniques employed in industrial process tomography (IPT), which can be used to evaluate the spatial distribution of permittivity based on current and voltage conditions along the boundaries of a specified region [13]. The measurement principle of ECT is based on the fact that the capacitance of a capacitor is a function of the permittivity (ε) of the medium between the electrode plates over the entire sensing space [14,15]. To inversely calculate the relative permittivity, the measured capacitance values are processed using an appropriate image reconstruction algorithm [16,17]. Electrical capacitance tomography has the advantages of a rapid response, low cost, non-intrusiveness, and robustness in harsh environments [18,19]. Compared to previous moisture detection methods in frozen soil, the ECT has great potential for evaluating moisture distribution and identifying ice fronts inside frozen soil due to its high sensitivity to water [20,21,22].
The objective of this paper is to design and optimize a mobile circular capacitance sensor (MCCS). This is employed in frozen soil testing to obtain the moisture distribution and the position of the ice front. Finally, the MCCS’s measurement indexes are validated based on the finite element simulation results. The research is divided into four goals: (a) to optimize MCCS’s structural parameters based on fuzzy optimization design; (b) to evaluate the moisture distribution along the height of the specimen based on the two proposed indexes; (c) to reconstruct the 3D ice front image via trilinear interpolation; (d) to obtain the simulated unfrozen moisture content to verify the MCCS’s evaluation index.
2. ECT Principle
The basic principle of the CCS used in this study is based on the fringe effect [23]. One of the electrodes of the CCS is applied to an AC or DC voltage as the excitation electrode, generating an electric field between the other electrodes, which are kept at zero potential and serve as the detectors. The capacitance between the excitation and the detection electrodes varies with the permittivity changes due to the different material distributions inside the sensor [24].
The essence of the ECT imaging problem is the need to solve two major mathematical problems, forward and inverse [16], as shown in Figure 1. The forward problem can be explained as follows: the sensor structure, medium distribution, and electrode measurement strategy are set up to solve the electrical field and potential distribution in the measurement field and obtain the sensitivity matrix. The inverse problem inversely calculates the medium distribution using image reconstruction algorithms based on the measured capacitance data and forward solution. Thus, the key to the forward problem is to solve the capacitance value of the medium distribution based on the set condition.
The following Maxwell equations represent the macroscopic electromagnetic field in the array capacitive sensor [25]:
(1)
where is the magnetic field intensity, is the electric field intensity, is the magnetic induction intensity, is the electric flux density, is the current density, and is the charge density.The operating frequency of the excitation electrode is in the low-frequency range, which corresponds to the electrostatic field condition, and the basic equation of the electrostatic field is obtained:
(2)
where is the permittivity and is the electric potential.According to the above equations, the mathematical model of the ECT system is derived as the Poisson equation:
(3)
Based on the Gauss formula, the capacitance between the and electrode pairs of the forward problem is as follows:
(4)
where is the permittivity distribution in the sensing field; V is the potential difference between the two electrodes forming the capacitance; is the potential distribution; Γ is the electrode surface. Equation (4) can be written in the following form:(5)
where is the sensitivity matrix. The sensitivity matrix reflects the sensitivity of the capacitance values between electrodes to changes in the permittivity in the sensing field. The permittivity distribution is inverted by the capacitance value and the sensitivity matrix using an image reconstruction algorithm to characterize the moisture distribution of frozen soil.3. Material and Methods
3.1. Fuzzy Satisfaction Evaluation Method
The evaluation indexes for circular capacitive sensors are typically associated with the capacitance measurements and sensitivity field distribution. This paper chooses capacitive dynamic range and the uniformity of sensitivity distribution as the two indicators for assessing MCCS performance, both of which can be calculated using COMSOL Multiphysics. The capacitive dynamic range () is defined as the ratio of the maximum capacitance () measured under full-field conditions to the minimum capacitance () measured in a null field, as indicated in Equation (6). The capacitive dynamic range should not be excessively high to minimize noise interference.
(6)
Non-uniform sensitivity can severely impact image quality. In the MCCS system, the arrangement of electrode plates is symmetrical, and the sensitivity field distribution () between any two plates remains consistent upon rotation. Therefore, only half of the number of electrode plates needs to be considered in the sensitivity distribution, with the calculation formula as shown in Equation (7).
(7)
where is the average sensitivity; is the standard deviation of micro-elements in the measured area; is defined as the sensitivity distribution coefficient; is the number of the electrode plates. A smaller value of indicates a more uniform distribution of sensitivity.Due to the numerous optimization indicators involved in ECT sensors, there may be conflicts among these indicators. When one indicator is optimal, the others are not necessarily optimal. To alleviate the conflicts between indicators, a reasonable combination of multiple indicators is necessary, followed by an experimental evaluation of the influencing factors. This paper employs the multi-index orthogonal experimental fuzzy analysis method proposed by Ji [26], establishing satisfaction levels regarding the indicators and their corresponding membership functions based on the concepts of fuzzy mathematics. The satisfaction function is defined to characterize the degree of acceptance, utilizing the existing normal distribution function to create the satisfaction function for evaluation indicators. These can be categorized into large-type, small-type, and intermediate-type indicators, with each influencing the overall evaluation results in distinct directions. Both and are small-type indicators, and Equation (8) is used for calculation:
(8)
where relates the index value obtained from the experiment and the satisfaction level. In this paper, is determined using an empirical method. In order to enhance the rationality and effectiveness of the comprehensive satisfaction function for the indicators, this paper employs a weighted arithmetic mean approach to develop the comprehensive satisfaction function, as shown in Equation (9).(9)
where is defined as the fuzzy satisfaction index.3.2. Specimen Preparation and Test Method
3.2.1. Material and Sample Preparation
The loess used was sourced from a subgrade construction project in a seasonally frozen region of China. The compaction testing was conducted according to ASTM D1557 [27]. Figure 2 presents the results of the compaction test. The maximum dry density is 1.746 g/cm3, with an optimal moisture content of 16.28%. Based on the optimal moisture content, the initial moisture content of the samples was set at 10%, 15%, and 20%. The prepared soil sample measured 10 cm in diameter and 16 cm in height. After preparation, the surface of the sample was wrapped with waterproof and insulating layers, made of polyethylene and polyester foam, to prevent moisture and heat exchange between the sample and its surroundings. The samples were subjected to a constant temperature of +2 °C for 24 h prior to the initiation of freezing.
3.2.2. Testing Methods
After the sample was prepared, capacitance values were collected every 1 cm along the depth using optimized MCCS, and these values were regarded as the full field. The control samples were fully dried, and capacitance values were also collected to represent the empty field. A multifunctional freeze–thaw testing machine was used for unidirectional freezing of the soil samples. The top temperature was set to −10 °C and the bottom to 2 °C, simulating unidirectional freezing in natural soil conditions. After freezing for 24 h, the sample was removed, and measurements were taken along the depth using MCCS, with the results considered as the measured field. After capacitance testing, the sample could be sliced at 1 cm intervals along the height, and the moisture content of each slice could be measured using the oven-drying method. For capacitance measurement, a high-sensitivity ECT system based on a commercial precision LCR meter (VICTOR 4090A) (Shenzhen Yisheng Victory Technology Co., Ltd., Shen Zhen, China) was employed. The excitation voltage was set at 2V while the frequency of data acquisition was 100 kHz. Figure 3 displays the whole experimental procedure.
3.2.3. Characterization Parameters Based on Capacitance Data
The sum of capacitance values () can be used to visually represent the moisture content across different layers of the frozen soil sample. However, the total capacitance value does not reflect the distribution of dielectric material across the cross-section, leading to considerable error. The average dielectric constant distribution () is introduced as an indicator to evaluate moisture migration in frozen soil due to the substantial difference between the dielectric constants of frozen and unfrozen soil. Using the sensitivity matrix derived from prior simulations and the Landweber algorithm, the normalized dielectric constant distribution of a specific cross-section of emulsified asphalt can be calculated.
(10)
where represents the normalized dielectric constant of the i-th pixel.3.3. Finite Element Method
Since MCCS primarily functions to detect the ice front in frozen soil, a two-dimensional thermal–hydraulic coupling model was established in COMSOL Multiphysics.
Step 1. Derivation of the mathematical model. According to Fourier’s law, latent heat from phase changes is treated as a heat source, and the differential equation for thermal conduction in frozen soil is given by Equation (11) [28,29,30].
(11)
where is the transient temperature of the soil (°C); is time (s); is water content (%); C(q) represents the volume heat capacity of soil (J/m3); is thermal conductivity (W·m−1 K−1); is ice content (%); and are the density of water and ice (kg/m3), respectively; is the latent heat of the phase change (kJ/kg). Using Darcy’s law and the principle of mass conservation as the foundation, a water migration control equation is established [30,31,32].(12)
where is the volumetric content of unfrozen water (%); represents the permeability coefficient of unsaturated soil in the gravity direction (m/s); is the water diffusion coefficient (m2/s). the solid–liquid ratio () is introduced to solve the hydrothermal coupling equations simultaneously [21]. The volumetric ice content of ice is expressed as follows:(13)
Step 2. Establishment of the numerical simulation model. The geometric model was constructed, and Equations (11) and (12) were transformed into the coefficient form of partial differential equations in COMSOL Multiphysics.
(14)
Step 3. Numerical calculation and export of results. Boundary conditions were set to match the experimental conditions, and the calculated moisture distribution results were exported.
4. Results and Discussion
4.1. Optimization Results of MCCS
Due to the large number of parameters involved in sensor optimization and the complex influence of external factors on certain parameters, it is essential to select relatively independent parameters for structural optimization. The structural parameters of MCCS primarily include the number of electrode plates, electrode opening angle coverage, plate width, and grounding shield radius. For a problem with a factor of 4 and a level of 3, all experiments must be performed 81 times. The orthogonal design method was employed to solve this problem, reducing the number of experiments required for the MCCS to nine using a three-level orthogonal table. Table 1 presents the orthogonal design layout.
The capacitive dynamic range (), sensitivity distribution coefficient (), and fuzzy satisfaction index () were selected as optimization indexes. Orthogonal experiment results are shown in Table 2. It was found that and cannot be optimized simultaneously, indicating that a conflict will occur when they are both chosen as optimization objectives.
The range analysis of the orthogonal test results is shown in Table 3, where represents the range of the optimization indicator for , represents the range for , and represents the range for the . A can be used to indicate the number of electrode plates, B to indicate the electrode opening angle coverage, C to indicate the plate width, and D to indicate the grounding shield radius. For example, A1B2C3D4 indicates the optimal combination, where the first set of electrode plate numbers, the second set of electrode opening angle coverages, the third set of plate widths, and the fourth set of the grounding shield radius are selected for a specific function. When the optimization indicator is , range analysis shows that the order of influence of the parameters on is A > D > B > C, and the optimal result is A2B1C3D3. For the optimization indicator , range analysis determines the influence order on as A > D > B > C, resulting in the optimal combination A2B3C3D1. The influence order of factors in the two optimization results is the same, but the corresponding level of each factor cannot be identical. With as the optimization indicator, range analysis reveals that the order of influence of the parameters on is A > D > B > C and the optimal result is A2B1C3D1.
As the optimal combination derived from the orthogonal experiment was not incorporated into the simulation test plan, a simulation test was performed on the selected parameters. The optimized parameter results are shown in Table 4. Comparing the results with those of with and , using to evaluate the structural parameters of MCCS can simultaneously obtain a low-capacitance dynamic range and a uniformly sensitive field distribution.
4.2. Characterization of Moisture Distribution
The frozen soil samples with different initial moisture contents were tested using MCCS to analyze the feasibility of using measurement indexes to assess moisture distribution. The test results are shown in Figure 4 and Figure 5. Figure 4 illustrates the distribution of the sum of capacitance values and the total moisture content along the depth of the specimen, while Figure 5 shows the average dielectric constant distribution and the total moisture content along the same depth direction. It was found that the moisture content in the unfrozen area decreased while the moisture content in the frozen area increased. As the experiment was conducted in a closed system with no external water supply, it is assumed that moisture migrated from the unfrozen area to the frozen area. As the initial moisture content increased, the slope of the curves in the figure increased, indicating more significant variations in moisture content along the height direction. Based on the increase in the order of initial moisture content, the peak moisture content positions for the 10%, 15%, and 20% samples were 8 cm, 10 cm, and 12 cm, respectively, with peak moisture content increasing by 9.5%, 11.3%, and 5.3% compared to the initial values.
As the initial moisture content increased, more water accumulated in the frozen region, and less water remained in the unfrozen region, showing a trend of increasing moisture migration. This is because, under the same cold junction temperature conditions, a higher initial moisture content in the samples meant that more time was required for freezing, providing more time for moisture migration and leading to a greater migration volume. Conversely, with a low initial moisture content in the sample, the freezing speed increased, resulting in a shorter migration time and a smaller migration volume.
It can be seen from the curves of the three specimens that there was a uniform decreasing trend from top to bottom, attributed to the much lower relative dielectric constant of ice compared to water. However, the slope variations showed no discernible pattern. The curves also showed a monotonically decreasing trend from bottom to top, but the greatest slope change occurred at the peak moisture content, indicating the most significant change in the dielectric properties of the material, with clear ice crystal formation at this location. If the value is less than this point, it is considered to be a frozen region, and if the value exceeds this threshold, it is considered to be an unfrozen region. Notably, the moisture content here was measured using the oven-drying method, which accounted for the migrated and frozen water. Therefore, both curves exhibited a monotonically decreasing trend.
A clear boundary between the frozen and unfrozen zones appeared near the peak moisture content, as illustrated in Figure 6, indicating that the freezing front occurred at the moisture content peak. Due to the characteristics of the material, the point with the minimum slope in the curve can be considered the ice front. It was found that the freezing fronts identified by the two methods were consistent, which preliminarily verified the accuracy of using to assess moisture distribution. Since the measurement step was 1 cm, the freezing front obtained using this indicator could only be taken as the intermediate value between two measured heights; an error analysis will be performed in the following finite element simulations.
4.3. Three-Dimensional Imaging of Ice Fronts
Before calculating the average dielectric constant distribution, a 2D image of the material distribution within the measured cross-section can be obtained. As shown in Figure 7, the dielectric constant distribution matrix was converted into a 2D image. Due to the presence of waterproof and insulation layers on all sides, the moisture distribution in the radial direction of the specimen showed little variation. Using trilinear interpolation algorithm, the eight 2D images were indirectly reconstructed into a 3D image, providing an intuitive visualization of the moisture distribution and the precise location of the ice front, as illustrated in Figure 8.
4.4. Finite Element Simulation Verification
The temperature field distribution of the three models with different initial moisture contents after 24 h is shown in Figure 9. Theoretically, the freezing temperature is 0 °C, but in real conditions, the freezing point may vary due to factors like water salinity and soil composition, typically being below 0 °C. In this finite element simulation, the ice front is defined as the 0 °C surface. The position of the ice front obtained from the curves is compared with the simulation results, as illustrated in Figure 10.
The simulated unfrozen water content along the depth of the specimen was compared to the after normalization, with the red solid line representing the model’s solution results and the blue solid dots indicating the results of the laboratory freezing test. The simulation results indicate a downward gradient in moisture content from bottom to top, with the greatest variation near the ice front. Above the ice front, the unfrozen water content approaches zero, and at this point, value reflects the dielectric properties of the ice–soil mixture. The index proposed in this paper aligns well with the simulation results, with minimal deviation, as shown in Table 5. This verifies that the index can accurately represent the moisture distribution in frozen soil and the position of the ice front, whereas can only provide a qualitative analysis of moisture distribution without pinpointing the ice front.
5. Conclusions
In this study, a self-designed MCCS was employed to evaluate moisture distribution and identify the ice front position in frozen soil. The following conclusions can be summarized from this paper:
-
The fuzzy optimization design method was employed to optimize the structural parameters. Comparing the results with and , using FSI to evaluate the structural parameters of MCCS can simultaneously obtain a low-capacitance dynamic range and a uniformly sensitive field distribution.
-
The sum of the capacitance values () and the average dielectric constant distribution () were introduced to evaluate moisture migration. The point with the minimum slope in the curve can be considered as the ice front, which was consistent with the moisture content.
-
The trilinear interpolation algorithm was used to reconstruct a 3D image of the ice front. The precise location of the ice front could be identified.
-
The temperature field and unfrozen moisture distribution were obtained via finite element simulation. The index proposed in this paper aligns well with the simulation results, with minimal deviation.
Future investigations should be implemented to evaluate the accuracy and sensitivity of the method, to develop appropriate reconstruction models and effective numerical methods, and to discover optimal sensor structures and discretization methods for the reconstruction domain for better measurement results.
Conceptualization, Q.D.; methodology, X.H. and Q.D.; software, B.S.; validation, X.H. and K.Y.; formal analysis, X.H. and X.C.; investigation, X.H., B.S., X.C., K.Y. and X.Y.; resources, Q.D. and X.C.; data curation, X.H., B.S., K.Y. and X.Y.; writing—original draft preparation, X.H. and Q.D.; writing—review and editing, X.H.; visualization, X.H. and K.Y.; supervision, Q.D. and X.C.; project administration, X.H. and X.Y.; funding acquisition, Q.D. and X.C. All authors have read and agreed to the published version of the manuscript.
Not applicable.
Not applicable.
Data are contained within the article.
This paper was significantly improved with the aid of anonymous reviewers and the editor. All support is gratefully acknowledged.
The authors declare no conflicts of interest.
Footnotes
Disclaimer/Publisher’s Note: The statements, opinions and data contained in all publications are solely those of the individual author(s) and contributor(s) and not of MDPI and/or the editor(s). MDPI and/or the editor(s) disclaim responsibility for any injury to people or property resulting from any ideas, methods, instructions or products referred to in the content.
Figure 4. [Forumla omitted. See PDF.] and moisture content variation in different specimens: (a) 10% initial moisture content; (b) 15% initial moisture content; (c) 20% initial moisture content.
Figure 5. [Forumla omitted. See PDF.] and moisture content variation in different specimens: (a) 10% initial moisture content; (b) 15% initial moisture content; (c) 20% initial moisture content.
Figure 6. Ice fronts of different specimens: (a) 10% initial moisture content; (b) 15% initial moisture content; (c) 20% initial moisture content.
Figure 7. 2D image of the relative permittivity distribution at different specimen heights with 10% initial moisture content.
Figure 8. Three-dimensional interpolated cloud plot of relative permittivity: (a) 10% initial moisture content; (b) 15% initial moisture content; (c) 20% initial moisture content.
Figure 9. Simulation results of temperature field after 24 h of freezing: (a) 10% initial moisture content; (b) 15% initial moisture content; (c) 20% initial moisture content.
Figure 10. Normalized simulation results of moisture content compared with measured [Forumla omitted. See PDF.] after 24 h of freezing: (a) 10% initial moisture content; (b) 15% initial moisture content; (c) 20% initial moisture content.
Orthogonal experimental design.
| Level | A. Electrode Plates Number | B. Electrode Opening Angle Coverage (%) | C. Plate Width (mm) | D. Grounding Shield Radius (mm) |
|---|---|---|---|---|
| 1 | 8 | 60 | 6 | 58 |
| 2 | 12 | 70 | 8 | 63 |
| 3 | 16 | 80 | 10 | 68 |
Optimization indexes results of orthogonal tests.
| Experiment | Structural Parameters | Optimization Indexes | |||||
|---|---|---|---|---|---|---|---|
| Group | A. Electrode Plates Number | B. Electrode Opening Angle Coverage (%) | C. Plate Width (mm) | D. Grounding Shield Radius (mm) | | | |
| 1 | 8 | 60 | 0.6 | 58 | 12.453 | 4.645 | 0.886 |
| 2 | 8 | 70 | 1.0 | 68 | 13.282 | 5.498 | 0.857 |
| 3 | 8 | 80 | 0.8 | 63 | 16.907 | 5.994 | 0.808 |
| 4 | 12 | 60 | 1.0 | 68 | 12.758 | 4.542 | 0.886 |
| 5 | 12 | 70 | 0.8 | 63 | 14.089 | 4.676 | 0.871 |
| 6 | 12 | 80 | 0.6 | 58 | 15.330 | 4.765 | 0.856 |
| 7 | 16 | 60 | 0.8 | 63 | 21.668 | 5.335 | 0.774 |
| 8 | 16 | 70 | 0.6 | 58 | 29.150 | 5.953 | 0.670 |
| 9 | 16 | 80 | 1.0 | 68 | 18.876 | 4.752 | 0.820 |
Range analysis of orthogonal test results.
| Level | A. Electrode Plates Number | B. Electrode Opening Angle Coverage | C. Plate Width | D. Grounding Shield Radius |
|---|---|---|---|---|
| | 42.176 | 43.880 | 53.933 | 58.815 |
| | 39.641 | 56.521 | 52.664 | 50.280 |
| | 69.694 | 51.112 | 44.916 | 42.418 |
| | 30.053 | 12.641 | 9.017 | 16.397 |
| | 16.137 | 15.511 | 15.363 | 14.073 |
| | 13.983 | 16.127 | 16.005 | 15.598 |
| | 16.041 | 14.523 | 14.793 | 16.490 |
| | 2.155 | 1.604 | 1.212 | 2.417 |
| | 2.264 | 2.568 | 2.435 | 2.599 |
| | 2.613 | 2.397 | 2.453 | 2.488 |
| | 2.574 | 2.486 | 2.563 | 2.364 |
| | 0.348 | 0.171 | 0.128 | 0.235 |
Optimized parameter results.
| Optimal Combination | A2B1C3D3 | A2B3C3D1 | A2B1C3D1 |
|---|---|---|---|
| | 12.758 | 12.812 | 12.741 |
| | 4.542 | 4.537 | 4.528 |
Measurement deviation compared with simulation.
| Initial moisture content | 10% | 15% | 20% |
| Measurement deviation | 0.9% | 1.5% | 2.8% |
References
1. Zhan, Y.; Lu, Z.; Yao, H.; Xian, S. A Coupled Thermo-Hydromechanical Model of Soil Slope in Seasonally Frozen Regions under Freeze-Thaw Action. Adv. Civ. Eng.; 2018; 2018, 7219826. [DOI: https://dx.doi.org/10.1155/2018/7219826]
2. Xinchun, L.; Yongde, K.; Hongna, C.; Hui, L. Hydrothermal Effects of Freeze-Thaw in the Taklimakan Desert. Sustainability; 2021; 13, 1292. [DOI: https://dx.doi.org/10.3390/su13031292]
3. Tai, B.; Yue, Z.; Sun, T.; Qi, S.; Li, L.; Yang, Z. Novel anti-frost subgrade bed structures a high speed railways in deep seasonally frozen ground regions: Experimental and numerical studies. Constr. Build. Mater.; 2021; 269, 121266. [DOI: https://dx.doi.org/10.1016/j.conbuildmat.2020.121266]
4. Li, N.; Xu, B. A new type of pile used in frozen soil foundation. Cold Reg. Sci. Technol.; 2008; 53, pp. 355-368. [DOI: https://dx.doi.org/10.1016/j.coldregions.2007.10.005]
5. Lei, S. Research on Ground Penetrating Radar (GPR) Exploration on Artificial Frozen Wall Development. J. China Univ. Min. Technol.; 2005; 34, pp. 143-147.
6. Zhu, Z.; Jia, J.; Zhang, F. A damage and elastic-viscoplastic constitutive model of frozen soil under uniaxial impact loading and its numerical implementation. Cold Reg. Sci. Technol.; 2020; 175, 103081. [DOI: https://dx.doi.org/10.1016/j.coldregions.2020.103081]
7. Zhang, L.; Pu, Y.; Liao, Q.; Gu, T. Dynamic investigation on the coupled changing process of moisture and density fields in freezing soil. Sci. China Ser. D Earth Sci.; 1999; 42, pp. 141-145. [DOI: https://dx.doi.org/10.1007/BF02878512]
8. Shan, W.; Liu, Y.; Hu, Z.; Xiao, J. A Model for the Electrical Resistivity of Frozen Soils and an Experimental Verification of the Model. Cold Reg. Sci. Technol.; 2015; 119, pp. 75-83. [DOI: https://dx.doi.org/10.1016/j.coldregions.2015.07.010]
9. Mercier, O.R.; Hunter, M.W.; Callaghan, P.T. Brine diffusion in first-year sea ice measured by Earth’s field PGSE-NMR. Cold Reg. Sci. Technol.; 2005; 42, pp. 96-105. [DOI: https://dx.doi.org/10.1016/j.coldregions.2004.12.004]
10. Bittelli, M.; Flury, M.; Roth, K. Use of dielectric spectroscopy to estimate ice content in frozen porous media. Water Resour. Res.; 2004; 40, W04212. [DOI: https://dx.doi.org/10.1029/2003WR002343]
11. Watanabe, K.; Wake, T. Measurement of unfrozen water content and relative permittivity of frozen unsaturated soil using NMR and TDR. Cold Reg. Sci. Technol.; 2009; 59, pp. 34-41. [DOI: https://dx.doi.org/10.1016/j.coldregions.2009.05.011]
12. Wen, Z.; Ma, W.; Feng, W.; Deng, Y.; Wang, D.; Fan, Z.; Zhou, C. Experimental study on unfrozen water content and soil matric potential of Qinghai-Tibetan silty clay. Environ. Earth Sci.; 2012; 66, pp. 1467-1476. [DOI: https://dx.doi.org/10.1007/s12665-011-1386-0]
13. Mühlbacher-Karrer, S.; Zangl, H. Object detection based on electrical capacitance tomography. Proceedings of the 2015 IEEE Sensors Applications Symposium (SAS); Zadar, Croatia, 13–15 April 2015; pp. 1-5.
14. Hosani, E.A.; Zhang, M.; Soleimani, M. A Limited Region Electrical Capacitance Tomography for Detection of Deposits in Pipelines. IEEE Sens. J.; 2015; 15, pp. 6089-6099. [DOI: https://dx.doi.org/10.1109/JSEN.2015.2453361]
15. Voss, A.; Pour-Ghaz, M.; Vauhkonen, M.; Seppänen, A. Electrical capacitance tomography to monitor unsaturated moisture ingress in cement-based materials. Cem. Concr. Res.; 2016; 89, pp. 158-167. [DOI: https://dx.doi.org/10.1016/j.cemconres.2016.07.011]
16. Voss, A.; Pour-Ghaz, M.; Vauhkonen, M.; Seppänen, A. Difference reconstruction methods for electrical capacitance tomography imaging of two-dimensional moisture flow in concrete. Proceedings of the 9th International Conference on Inverse Problems in Engineering (ICIPE); Waterloo, ON, Canada, 23–26 May 2017.
17. Wang, W.; Zhao, K.; Zhang, P.; Bao, J.; Xue, S. Application of three self-developed ECT sensors for monitoring the moisture content in sand and mortar. Constr. Build. Mater.; 2020; 267, 121008. [DOI: https://dx.doi.org/10.1016/j.conbuildmat.2020.121008]
18. Liu, S.; Wang, H.; Li, Y. Level measurement of separator by using electrical tomography sensor. Proceedings of the 2018 IEEE International Conference on Imaging Systems and Techniques (IST); Krakow, Poland, 16–18 October 2018; pp. 1-6.
19. Krupa, A.; Lackowski, M.; Jaworek, A. Capacitance sensor for measuring void fraction in small channels. Measurement; 2021; 175, 109046. [DOI: https://dx.doi.org/10.1016/j.measurement.2021.109046]
20. Rimpiläinen, V.; Heikkinen, L.M.; Vauhkonen, M. Moisture distribution and hydrodynamics of wet granules during fluidized-bed drying characterized with volumetric electrical capacitance tomography. Chem. Eng. Sci.; 2012; 75, pp. 220-234. [DOI: https://dx.doi.org/10.1016/j.ces.2012.03.028]
21. Qing-bo, B.A.I.; Xu, L.I.; Ya-hu, T.; Jian-hong, F. Equations and numerical simulation for coupled water and heat transfer in frozen soil. Chin. J. Geotech. Eng.; 2015; 37, pp. 131-136. [DOI: https://dx.doi.org/10.11779/CJGE2015S2026]
22. Nandi, R.; Shrestha, D. Assessment of Low-Cost and Higher-End Soil Moisture Sensors across Various Moisture Ranges and Soil Textures. Sensors; 2024; 24, 5886. [DOI: https://dx.doi.org/10.3390/s24185886]
23. Eren, H.; Sandor, L.D. Fringe-Effect Capacitive Proximity Sensors for Tamper Proof Enclosures. Proceedings of the 2005 Sensors for Industry Conference; Houston, TX, USA, 8–10 February 2005; pp. 22-26.
24. Li, N.; Guo, B.l.; Huang, C. Characterisation of Liquid Properties by Electrical Capacitance Tomography Sensor for Security Applications. Proceedings of the 2009 Second International Symposium on Electronic Commerce and Security; Nanchang, China, 22–24 May 2009; pp. 305-308.
25. Lei, J.; Liu, S.; Wang, X.; Liu, Q. An Image Reconstruction Algorithm for Electrical Capacitance Tomography Based on Robust Principle Component Analysis. Sensors; 2013; 13, pp. 2076-2092. [DOI: https://dx.doi.org/10.3390/s130202076]
26. Ji, Z.; Liu, J.; Tian, H.; Zhang, W. ECT Sensor Simulation and Fuzzy Optimization Design Based on Multi Index Orthogonal Experiment. IEEE Access; 2020; 8, pp. 190039-190048. [DOI: https://dx.doi.org/10.1109/ACCESS.2020.3029839]
27.
28. Deng, Q.; Liu, X.; Zeng, C.; He, X.; Chen, F.; Zhang, S. A Freezing-Thawing Damage Characterization Method for Highway Subgrade in Seasonally Frozen Regions Based on Thermal-Hydraulic-Mechanical Coupling Model. Sensors; 2021; 21, 6251. [DOI: https://dx.doi.org/10.3390/s21186251] [PubMed: https://www.ncbi.nlm.nih.gov/pubmed/34577458]
29. Harlan, R.L. Analysis of coupled heat-fluid transport in partially frozen soil. Water Resour. Res.; 1973; 9, pp. 1314-1323. [DOI: https://dx.doi.org/10.1029/WR009i005p01314]
30. Yu, Z.; Yang, L.; Zhou, S. Comparative study of relating equations in coupled thermal-hydraulic finite element analyses. Cold Reg. Sci. Technol.; 2019; 161, pp. 150-158. [DOI: https://dx.doi.org/10.1016/j.coldregions.2019.03.005]
31. Ranz, J.; Aparicio, S.; Romero, H.; Casati, M.J.; Molero, M.; González, M. Monitoring of Freeze-Thaw Cycles in Concrete Using Embedded Sensors and Ultrasonic Imaging. Sensors; 2014; 14, pp. 2280-2304. [DOI: https://dx.doi.org/10.3390/s140202280]
32. Li, Z.; Chen, J.; Sugimoto, M. Pulsed NMR Measurements of Unfrozen Water Content in Partially Frozen Soil. J. Cold Reg. Eng.; 2020; 34, 04020013. [DOI: https://dx.doi.org/10.1061/(ASCE)CR.1943-5495.0000220]
You have requested "on-the-fly" machine translation of selected content from our databases. This functionality is provided solely for your convenience and is in no way intended to replace human translation. Show full disclaimer
Neither ProQuest nor its licensors make any representations or warranties with respect to the translations. The translations are automatically generated "AS IS" and "AS AVAILABLE" and are not retained in our systems. PROQUEST AND ITS LICENSORS SPECIFICALLY DISCLAIM ANY AND ALL EXPRESS OR IMPLIED WARRANTIES, INCLUDING WITHOUT LIMITATION, ANY WARRANTIES FOR AVAILABILITY, ACCURACY, TIMELINESS, COMPLETENESS, NON-INFRINGMENT, MERCHANTABILITY OR FITNESS FOR A PARTICULAR PURPOSE. Your use of the translations is subject to all use restrictions contained in your Electronic Products License Agreement and by using the translation functionality you agree to forgo any and all claims against ProQuest or its licensors for your use of the translation functionality and any output derived there from. Hide full disclaimer
© 2024 by the authors. Licensee MDPI, Basel, Switzerland. This article is an open access article distributed under the terms and conditions of the Creative Commons Attribution (CC BY) license (https://creativecommons.org/licenses/by/4.0/). Notwithstanding the ProQuest Terms and Conditions, you may use this content in accordance with the terms of the License.
Abstract
As the interface between frozen and unfrozen soil, the ice front is not only a spatial location concept, but also a potentially dangerous interface where the mechanical properties of soil could change abruptly. Accurately identifying its spatial position is essential for the safe and efficient execution of large-scale frozen soil engineering projects. Electrical capacitance tomography (ECT) is a promising method for the visualization of frozen soil due to its non-invasive nature, low cast, and rapid response. This paper presents the design and optimization of a mobile circular capacitance sensor (MCCS). The MCCS was used to measure frozen soil samples along the depth direction to obtain moisture distribution and three-dimensional images of the ice front. Finally, the experimental results were compared with the simulation results from COMSOL Multiphysics to analyze the deviations. It was found that the fuzzy optimization design based on multi-criteria orthogonal experiments makes the MCCS meet various performance requirements. The average permittivity distribution was proposed to reflect moisture distribution along the depth direction and showed good correlation. Three-dimensional reconstructed images could provide the precise position of the ice front. The simulation results indicate that the MCCS has a low deviation margin in identifying the position of the ice front.
You have requested "on-the-fly" machine translation of selected content from our databases. This functionality is provided solely for your convenience and is in no way intended to replace human translation. Show full disclaimer
Neither ProQuest nor its licensors make any representations or warranties with respect to the translations. The translations are automatically generated "AS IS" and "AS AVAILABLE" and are not retained in our systems. PROQUEST AND ITS LICENSORS SPECIFICALLY DISCLAIM ANY AND ALL EXPRESS OR IMPLIED WARRANTIES, INCLUDING WITHOUT LIMITATION, ANY WARRANTIES FOR AVAILABILITY, ACCURACY, TIMELINESS, COMPLETENESS, NON-INFRINGMENT, MERCHANTABILITY OR FITNESS FOR A PARTICULAR PURPOSE. Your use of the translations is subject to all use restrictions contained in your Electronic Products License Agreement and by using the translation functionality you agree to forgo any and all claims against ProQuest or its licensors for your use of the translation functionality and any output derived there from. Hide full disclaimer
Details
; Qiao Dong 1
; Shi, Bin 1
; Kang, Yao 1
; Chen, Xueqin 2
; Yuan, Xin 1
1 Department of Roadway Engineering, Southeast University, Nanjing 211189, China;
2 Department of Civil Engineering, Nanjing University of Science and Technology, Nanjing 210094, China;