Abstract

ABSTRACT

Background

Point‐of‐care testing of blood gases plays a critical role in patient management. The aim of this study was to verify the manufacturer's specifications of the Nova Stat Profile Prime Plus Analyser, along with a comparison study with the GEM Premier 4000 Blood Gas Analyser.

Methods

Parameters analysed were pH, pCO2, pO2, Na+, Cl, K+, iCa, lactate, and glucose. Data for the precision and bias study were generated using control samples in a 5 × 5 study design. Linearity was checked using a five‐level Linearity Control Set, while comparison was done between the Nova and GEM analysers using whole blood samples (N = 103). Acceptance was based on the CLIA TEa for all analytes except for lactate, for which the TEa defined by CAP and AAB was used.

Results

The within‐run and between‐run CVR% precision were all lower than the claimed CVs%, except for pCO2 control level 2 within run (CV% 1.5 [claim CV% 1.1]) and iCa control level 5 between run (CV% 1.42 [claim 1.12]). The observed bias for all parameters was within the calculated lower and upper bias limits. Linearity was verified for all analytes except for Na+. Upon comparison of the Nova and GEM analysers, a correlation coefficient above 0.95 was observed for most parameters.

Conclusion

The Nova Stat Profile Prime Plus analyser meets the manufacturer's precision and bias claims. Linearity was confirmed for most analytes. There was a good correlation between the Nova and GEM Blood Gas analyser at concentrations within the reference range intervals for all investigated parameters.

Details

1009240
Title
The Verification Process of a POC Blood Gas Analyser—The Nova Stat Profile Primer Plus Analyser
Author
Costa, Joseph 1 ; Attard, Ritienne 1   VIAFID ORCID Logo 

 GGH Medical Laboratory Services, Gozo General Hospital, Victoria, Gozo, Malta 
Publication title
Volume
39
Issue
7
Publication year
2025
Publication date
Apr 1, 2025
Section
RESEARCH ARTICLE
Publisher
John Wiley & Sons, Inc.
Place of publication
New York
Country of publication
United States
ISSN
08878013
e-ISSN
10982825
Source type
Scholarly Journal
Language of publication
English
Document type
Journal Article
Publication history
 
 
Online publication date
2025-03-25
Milestone dates
2024-12-26 (manuscriptRevised); 2025-04-09 (publishedOnlineFinalForm); 2024-10-28 (manuscriptReceived); 2025-03-25 (publishedOnlineEarlyUnpaginated); 2025-02-07 (manuscriptAccepted)
Publication history
 
 
   First posting date
25 Mar 2025
ProQuest document ID
3188166879
Document URL
https://www.proquest.com/scholarly-journals/verification-process-poc-blood-gas-analyser-nova/docview/3188166879/se-2?accountid=208611
Copyright
© 2025. This work is published under http://creativecommons.org/licenses/by/4.0/ (the "License"). Notwithstanding the ProQuest Terms and Conditions, you may use this content in accordance with the terms of the License.
Last updated
2025-04-28
Database
2 databases
  • Coronavirus Research Database
  • ProQuest One Academic