Content area

Abstract

Background

Both outsourcing virtual surgery planning and 3D printed splint fabrication have become the standard in the field of orthognathic surgery. In-house (IH) adaptation of these presurgical operations requires compliance with regulatory bodies when designing and manufacturing medical-grade products. The purpose of this study is to evaluate the dimensional accuracy of IH 3D printed orthognathic surgical splints within a hybrid workflow for externally designed splints.

Materials and methods

An in vitro study was conducted utilizing an outsourced (OS) orthognathic surgical splint file from a previously treated patient. The control group included the splint’s original standard tessellation language (STL) file. Experimental groups included: splint (a) milled with zirconia, (b) 3D printed on a Phrozen 4 K (commercial printer), (c) 3D printed on Formlabs 3B+ (Formlabs 3B + have track record of having FDA cleared materials and workflows in dental applications and have potential materials that could be used for IH manufacturing for orthognathic splints upon further testing and FDA clearance). Surface area analysis was performed using 3-Matic (Materialise© Mimics Software) to generate root mean square (RMS) values between digital copies of the splints and their corresponding original STL files.

Results

The RMS error in equipment processing and analysis was measured at 0.10 mm. Accounting for this error, the 3D-printed splints exhibited an average RMS of 0.20 mm for both the Formlabs 3B + and the Phrozen 4 K. No statistically significant difference was found between splints from different printers or replicas.

Conclusion

This study presents a verification process for providers to verify the geometric stability and reproducibility of their IH-printed orthognathic splints (RMS 0.20 mm). Clinicians may find this study useful when crafting a regulatory-compliant process for the IH manufacturing of OS orthognathic surgery splints.

Details

1009240
Business indexing term
Company / organization
Title
Printing outsourced orthognathic surgical splints in-house: a dimensional verification process for point-of-care printing
Publication title
Volume
11
Issue
1
Pages
24
Publication year
2025
Publication date
Dec 2025
Publisher
Springer Nature B.V.
Place of publication
London
Country of publication
Netherlands
e-ISSN
23656271
Source type
Scholarly Journal
Language of publication
English
Document type
Journal Article
Publication history
 
 
Online publication date
2025-06-06
Milestone dates
2025-05-29 (Registration); 2024-11-13 (Received); 2025-05-29 (Accepted)
Publication history
 
 
   First posting date
06 Jun 2025
ProQuest document ID
3216425017
Document URL
https://www.proquest.com/scholarly-journals/printing-outsourced-orthognathic-surgical-splints/docview/3216425017/se-2?accountid=208611
Copyright
Copyright Springer Nature B.V. Dec 2025
Last updated
2025-06-07
Database
2 databases
  • Coronavirus Research Database
  • ProQuest One Academic