Monte Carlo simulation relies on pseudo-random number generators. In general, the quality of these generators can have a direct impact on simulation results. The GATE toolbox, widely adopted in radiotherapy, offers three generators from which users can choose: Mersenne Twister, Ranlux-64, and James-Random. In this study, we used these generators to simulate the head of a medical linear accelerator for 6 MV photons in order to assess their potential impact on the results obtained in radiotherapy simulation. Simulations were conducted for four different field openings. The simulations included a linac head model and a water phantom, all components of the head of the medical linear accelerator, and a water phantom placed at a distance of 100 cm from the electron source. Statistical analysis based on normal probability and Bland–Altman plots were used to compare dose distributions in the voxelized water phantom obtained by each generator. Experimental data (dose profiles, percentage dose at depth, and other dosimetric parameters) were measured using an appropriate quality assurance protocol for comparison with the different simulations. The evaluation of dosimetric criteria shows significant variations, particularly in the physical penumbra of the dose profile for large fields. The gamma index analysis highlights significant distinctions in generator performance. In all simulations, the average time of the primary particle generation rate, number of tracks, and steps in the simulation of different random number generators showed differences. The Mersenne Twister generator was distinguished by high performance in several aspects, particularly in terms of execution time, primary particle production, track and step production flow rate, and coming closer to the experimental results. Regarding computational time, the simulation using the Mersenne Twister generator was about 18% faster than the one using the James-Random generator and 27% faster than the simulation using the Ranlux-64 generator. This suggests that this generator is the most reliable for accurate and fast modeling of the medical linear accelerator head for 6 MV energy.

Details

1009240
Title
Impact of Pseudo-Random Number Generators on Dosimetric Parameters in Validation of Medical Linear Accelerator Head Simulation for 6 MV Photons Using the GATE/GEANT4 Platform
Author
Tantaoui Meriem 1   VIAFID ORCID Logo  ; Krim Mustapha 2   VIAFID ORCID Logo  ; Essaidi El Mehdi 2   VIAFID ORCID Logo  ; Kaanouch Othmane 2 ; Mesradi Mohammed Reda 2 ; Kartouni Abdelkrim 1 ; Sahraoui Souha 3 

 Subatomic Research and Applications Team, Laboratory of the Physics of Condensed Matter (LPMC-ERSA), Faculty of Sciences Ben M’Sick, Hassan II University, Casablanca BP 7955, Morocco 
 Laboratory of Sciences and Health Technologies, High Institute of Health Sciences (ISSS), Hassan I University, Settat BP 555, [email protected] (E.M.E.); 
 Faculty of Medicine and Pharmacy, Hassan II University, Casablanca BP 9154, Morocco; [email protected] 
Publication title
Volume
9
Issue
2
First page
16
Number of pages
14
Publication year
2025
Publication date
2025
Publisher
MDPI AG
Place of publication
Basel
Country of publication
Switzerland
Publication subject
e-ISSN
2412382X
Source type
Scholarly Journal
Language of publication
English
Document type
Journal Article
Publication history
 
 
Online publication date
2025-05-05
Milestone dates
2024-12-07 (Received); 2025-04-21 (Accepted)
Publication history
 
 
   First posting date
05 May 2025
ProQuest document ID
3223939461
Document URL
https://www.proquest.com/scholarly-journals/impact-pseudo-random-number-generators-on/docview/3223939461/se-2?accountid=208611
Copyright
© 2025 by the authors. Licensee MDPI, Basel, Switzerland. This article is an open access article distributed under the terms and conditions of the Creative Commons Attribution (CC BY) license (https://creativecommons.org/licenses/by/4.0/). Notwithstanding the ProQuest Terms and Conditions, you may use this content in accordance with the terms of the License.
Last updated
2025-11-17
Database
ProQuest One Academic