Full text

Turn on search term navigation

© 2025 by the authors. Licensee MDPI, Basel, Switzerland. This article is an open access article distributed under the terms and conditions of the Creative Commons Attribution (CC BY) license (https://creativecommons.org/licenses/by/4.0/). Notwithstanding the ProQuest Terms and Conditions, you may use this content in accordance with the terms of the License.

Abstract

Introduction: Intramedullary nailing is widely used for long bone fractures. Traditional systems are reliable, but they present some complications regarding lack of modularity or possible growth plate damage. Methods: A novel locking rod–screw system featuring a central rod and a grooved screw with a secondary interlocking mechanism was developed and tested. Mechanical testing followed ASTM F543 and ISO 6475 standards using a 3.0 mm steel alloy prototype. Results: The system withstood mechanical testing >200,000 cycles at loads up to 200 N with no rates of failure or loosening, significantly outperforming other implants of the same size (3.0 mm TENS). Conclusions: The proposed implant demonstrates superior biomechanical performance in vitro, enabled by its unique hollow screw and secondary locking configuration. This modular and minimally invasive system shows promise for use in cases of smaller long bones, personalized paediatric fractures, and all types of diaphyseal fractures, but does warrant in vivo validation.

Details

Title
Original Locking Rod System Designed for Diaphyseal Fractures of Long Bones
Author
Liviu-Coriolan, Misca 1   VIAFID ORCID Logo  ; Croicu, Cristian Constantin 2 ; Lazarescu, Adrian Emil 3 ; Mihai-Alexandru, Sandesc 3 ; Patrascu Jenel Marian Jr. 3 ; Florescu Sorin 3 ; Patrascu Jenel Marian Sr. 3 

 Department of Orthopedics and Traumatology, Victor Babes University of Medicine and Pharmacy, 300041 Timisoara, Romania; [email protected] (L.-C.M.); [email protected] (M.-A.S.); [email protected] (J.M.P.J.); [email protected] (S.F.); [email protected] (J.M.P.S.), Department Orthopaedics and Traumatology, Cork University Hospital, T12 DC4A Cork, Ireland 
 Veterinaria Timisoara, 300502 Timisoara, Romania; [email protected] 
 Department of Orthopedics and Traumatology, Victor Babes University of Medicine and Pharmacy, 300041 Timisoara, Romania; [email protected] (L.-C.M.); [email protected] (M.-A.S.); [email protected] (J.M.P.J.); [email protected] (S.F.); [email protected] (J.M.P.S.), Spitalul Clinic Judetean de Urgenta “Pius Brinzeu” Timisoara, 300723 Timisoara, Romania 
First page
348
Publication year
2025
Publication date
2025
Publisher
MDPI AG
e-ISSN
20794983
Source type
Scholarly Journal
Language of publication
English
ProQuest document ID
3254548980
Copyright
© 2025 by the authors. Licensee MDPI, Basel, Switzerland. This article is an open access article distributed under the terms and conditions of the Creative Commons Attribution (CC BY) license (https://creativecommons.org/licenses/by/4.0/). Notwithstanding the ProQuest Terms and Conditions, you may use this content in accordance with the terms of the License.