1. Introduction
Despite significant advancements in chemotherapy-based conditioning regimens for haematological malignancies, Total Body Irradiation (TBI) has remained, over recent decades, a cornerstone for haematopoietic stem cell transplantation (HSCT), particularly in clinical situations where penetration of radiation in sanctuary organs confers advantages over systemic therapy alone [1].
The foundations of modern HSCT date back to the 1950s, when E. Donnall Thomas and colleagues first combined TBI with bone marrow infusion, a pioneering approach that ultimately earned a Nobel Prize and established the basis for contemporary bone marrow transplantation [2].
Early TBI regimens employed single-fraction radiation using cobalt machines, which posed several challenges, including complex patient positioning and notable side effects [3,4]. The introduction of novel technologies led to the development of linear accelerators and refined fractionation schedules, markedly improving delivery time and treatment tolerability [5,6]. Today, TBI continues to evolve, with the emergence of reduced-intensity regimens and targeted approaches like total marrow irradiation (TMI) and total marrow and lymphoid irradiation (TMLI), which aim to preserve efficacy while minimising toxicity [7].
This work aims to present a comprehensive overview of the evolution of TBI, with a particular focus on fractionation schedules, delivery techniques, and ongoing challenges. To this end, we performed a PubMed search on TBI and HSCT, prioritising randomised trials, prospective cohort studies, and comparative analyses. In parallel, we report over one decade of institutional experience at the University Hospital Policlinico of Catania, to highlight the development and adaptation of local protocols in response to technological advancements and shifting clinical indications.
2. TBI Indications and Integration with Systemic Therapy
TBI is a specialised radiotherapy technique that delivers a uniform dose of ionising radiation to the entire body [4]. Initially used as a standalone treatment, TBI has since become an integral part of conditioning regimens for HSCT for different haematological malignancies, with both myeloablative and immunosuppressive intents [8,9].
In allogeneic HSCT, especially for acute leukaemia, TBI facilitates the eradication of residual haematopoietic cells and helps create immunologic space for donor cell engraftment [10].
Historically, one of the earliest and most widely used radiotherapy schedules for TBI was a single 10 Gy fraction. However, since the early 1980s, fractionated TBI has demonstrated superior outcomes, including improved overall survival and lower rates of graft-versus-host disease (GVHD) [11,12].
Nowadays, the typical myeloablative TBI regimens include: 12 Gy delivered as 2 Gy twice daily over 3 days (6 fractions) [13]; 12–13.5 Gy delivered as 1.5 Gy twice daily over 4–4.5 days (8–9 fractions) [13]; 12–13.2 Gy delivered as 1.2 Gy three times daily over 4 days (10–11 fractions) [13]; 12 Gy delivered as 3 Gy once daily over 4 days (4 fractions) [13];
In the paediatric population, the typical myeloablative doses range between 12 and 14.4 Gy (1.6–2 Gy per fraction) [14]. Additionally, in acute lymphoblastic leukaemia, a testicular boost of 4 Gy (2 Gy × 2 fractions) is often administered to reduce relapse in this sanctuary site, although this practice varies by institution and is under continued evaluation [15].
Regarding chemotherapy regimens, the most widely adopted agent is cyclophosphamide; however, high-dose etoposide can be considered an alternative for high-risk or refractory cases where deeper cytoreduction is needed [16].
In efforts to replace TBI in the conditioning of acute leukaemia, several studies have investigated non-TBI containing regimens, including intravenous busulfan and cyclophosphamide [17,18]. Despite these efforts, TBI has demonstrated to improve event-free survival rates as well as overall survival and is therefore still a key component for allogeneic HSCT. In particular, the superiority of fractionated TBI over chemotherapy-only conditioning has been confirmed by the phase-III FORUM trial, which significantly improved the 2-year overall survival rates (0.91; 95% CI, 0.86 to 0.95; p < 0.0001 versus 0.75; 95% CI, 0.67 to 0.81) and cumulative incidence of relapse (0.12; 95% CI, 0.08 to 0.17; p < 0.0001 versus 0.33; 95% CI, 0.25 to 0.40) in paediatric and adolescents and young adults (AYA) affected by acute lymphoblastic leukaemia [19].
For elderly patients or those with comorbidities, reduced-intensity conditioning (RIC) regimens are increasingly adopted, with total doses ranging from 1 to 8 Gy [13]. The Quesenberry group demonstrated that even 1 Gy of TBI could already support engraftment in a series of 25 patients with refractory cancer [20,21]. Similarly, the Seattle group showed that 2 Gy TBI with fludarabine is sufficient to achieve reliable engraftment in 45 patients with haematologic malignancies, ineligible for conventional HSCT [22].
Nowadays, RIC regimens of 2 Gy TBI, alone or in association with fludarabine or fludarabine and melphalan, offer sufficient immunosuppression for full donor chimerism while minimising regimen-related toxicity [23,24]. In cases requiring intensified immunosuppression, such as with HLA-mismatched donors or heavily pretreated recipients, the TBI dose may be escalated to 4 Gy (2 Gy × 2 fractions) [25].
The development of RIC regimens, including thiotepa-based conditioning, has allowed for the expansion of transplant eligibility over the years [26]. However, the potential risks of graft rejection or disease relapse must still be considered.
In the autologous HSCT setting, particularly for multiple myeloma, relapsed Hodgkin lymphoma, and non-Hodgkin lymphoma, TBI was historically incorporated into conditioning regimens due to its cytoreductive properties [9]. However, its role has diminished over time and is now largely reserved for selected high-risk cases.
An overview of fractionation schedules based on their indications is reported in Table 1.
3. Technological Advancements in TBI
3.1. From Isotopes to Linear Accelerators
Some of the earliest experiences reported in the literature on TBI date back to the late 1950s, when multiple radiation sources were used to achieve adequate dose distribution.
In 1961, Jacobs and Pape described one of the first dedicated TBI chambers based on cesium-137 sources. The sources were contained in rods placed at the extremity of the treatment bed, at 2 m, and were controlled by lead filters and by an on/off system [27].
From cesium-137 to cobalt-60, in 1959, Sahler developed one of the first dual-source cobalt-based radiation units for TBI, which allowed the delivery of a parallel opposed field at about 3 m from the patient, with a low dose rate irradiation (<10 cGy/min) [28].
The following innovation utilised a single cobalt-60 source and was reported in 1981 by investigators from the Princess Margaret Hospital in Toronto. At that time in the facility, the radiation unit was engineered to deliver a large radiation field at a relatively short source-to-skin distance (SSD) of 90 cm, with high dose rates (100–200 rad/min). Despite the challenges of dosimetry for such a large field, this advance represented a practical and high-output solution for that time [29].
Subsequently, several innovations were made, including vertical-beam TBI and translation-irradiation technique, until the implementation of photon beams linear accelerators, paving the way for conventional methods [30,31,32]
Conventional TBI refers to a specific delivery approach where the patient lies or stands at an extended SSD of around 3–4 m and is treated with a single radiation beam and with the largest possible field size. With the aim of obtaining a uniform dose rate along the patient’s body, compensating filters or boluses are adopted, together with a Plexiglas plate, usually placed between the radiation source and the patient’s body, to counteract the skin-sparing effect. Regarding positioning verifications, a light field simulating the irradiation field is normally used. Anterior and posterior fields are adopted, with the dose prescribed to the mid-plane [33].
To reduce the dose to the lungs, organ shielding is performed through customised blocks of high-density material, accurately positioned and verified with the use of radiographic films [34]. However, attention should be paid to reach an adequate beam transmission, balancing the risk of interstitial pneumonia and of leukaemia recurrence. Nowadays, Computed Tomography (CT) simulation-based solutions are adopted for accurate shielding in a minimal treatment time [35].
3.2. Intensity Modulated Total Body Irradiation
Starting from the early 2000s, the most recent technological advancements in radiotherapy have enabled the delivery of highly conformal radiation to even large anatomical volumes. Techniques such as intensity-modulated radiotherapy (IMRT), volumetric modulated arc therapy (VMAT), Helical Tomotherapy (HT), and Image-Guided Radiotherapy (IGRT) allow for more precise targeting of the bone marrow, lymphatic structures, and circulating blood, while offering improved results in distribution. These innovations have led to the development of TMI and TMLI, which are organ-sparing approaches that refine traditional TBI by selectively irradiating haematopoietic and lymphoid tissue [36].
Early experiences with TMI with HT already showed the technical feasibility, dosimetric precision, and clinical benefits of this approach over traditional methods, although associated with longer treatment planning and delivery times [37].
Compared to conventional TBI, VMAT and HT allow for a more comfortable patient position and better sparing of organs at risk with fewer acute toxicities. However, the dual treatment plan requires meticulous planning to avoid over- and under-dosing in the junctions; with an overall treatment duration that can extend up to 30–90 min, due to the necessity to irradiate the entire body in sequential sections [38,39]
Overall, excellent results have been obtained with conventional linear accelerators, although highlighting the necessity to develop novel planning techniques to account for large treatment fields, as well as the integration of daily cone-beam CT to verify patient position [40].
Nowadays, TMI and TMLI are normally delivered through static or dynamic HT or VMAT and involve a meticulous patient setup, immobilisation, and imaging [41,42,43].
In the simulation phase, the patient lies in the supine position and is normally immobilised with the help of a thermoplastic mask and a personalised vacuum pillow. Based on the patient’s size, two separate CT scans (head-first and feet-first) and a dual treatment plan might be necessary to cover the entire body of the patient [41]. To facilitate changing positions, a rotating couch top is often required.
Unlike conventional TBI, an accurate definition of the clinical target volume (CTV) is needed, including the skeletal bone, the entire lymphatic system when indicated, as well as other sanctuary sites such as the testicles and the brain [41]. CTV is then expanded to the planning target volume (PTV), with a margin that is usually based on the institutional protocol.
Organs at risk typically include critical structures such as the brain (if not part of the target volume), eyes, lenses, optic nerves and chiasm, oral cavity, major salivary glands, thyroid, oesophagus, lungs, heart, breasts, stomach, liver, spleen (if excluded from the target), kidneys, small and large intestines, rectum, bladder, and genital organs [41].
Despite these benefits, the most advanced irradiation techniques have not demonstrated superiority, compared to conventional TBI, in terms of immunosuppressive effect on bone marrow cells and tumoricidal effect on circulating leukaemic cells. Furthermore, no substantial difference between modern techniques and conventional TBI has been shown in the occurrence of relapses [44].
4. Ongoing Challenges and Future Directions
4.1. Cellular Hallmarks: Key Considerations
As early as the 1990s, experiments conducted on leukaemic cell cultures revealed a heterogeneous response to radiation, with some cell lines exhibiting marked radiosensitivity and others demonstrating significant radioresistance [45,46].
Building on this work, Wheldon discovered that a TBI regimen of 7 fractions of 2 Gy could result in a wide spectrum of treatment responses in leukaemic cells, ranging from easily curable to entirely unresponsive cases [46].
Although the mechanisms driving this variability were not fully understood then, and still today remain incompletely defined, early work referred to dysregulation of apoptosis-related genes (bcl2, p53, c-myc) and cellular differentiation pathways as possible key contributors [47,48,49].
In a more recent study, for example, Li et al. demonstrated that activation of the Hedgehog signalling pathway mediates resistance to radiation in acute myeloid leukaemia, with cell lines exhibiting enhanced DNA repair and suppressed apoptosis. Importantly, the pharmacologic inhibition of this pathway with LDE225 restored radiosensitivity both in vitro and in vivo, providing proof-of-concept that molecular targeting can be used to reverse resistance and personalise TBI strategies [49,50].
Besides genetic alteration, differences in radiobiological parameters were also reported as potential contributors to radiation response heterogeneity, namely the α/β ratio [46,50]. The α/β ratio defines the fraction dose where both α (non-reparable DNA damage) and β (reparable DNA damage) components cause the same value of cell death, and is therefore higher in early responding tissues (less sensitive to fraction size or dose rate variations), while is lower in late-responding tissues (more susceptible to fraction size or dose rate increments) [51]. Cosset et al. reported high variability in cell survival curves within haematological malignancies, with low fraction size sensitivity for acute non-lymphoblastic leukaemia, but observed significant fractionation sensitivity in chronic myeloid leukaemia, along with notable interindividual variability in acute lymphoblastic leukaemia [52].
Collectively, these genetic and radiobiological considerations paint a picture of highly individual radiation sensitivity. To translate that bench-to-bedside decision-making, modern molecular response tools have also recently been developed to identify individualised metrics. Specifically, advanced techniques such as PCR-based minimal residual disease (MRD) monitoring and next-generation sequencing (NGS) are increasingly used to assess patient-specific tumour burden and treatment response [53,54].
Furthermore, early-phase clinical trials have successfully incorporated radiolabelled anti-CD20, -CD45, and -CD66 antibodies into standard TBI-based conditioning regimens for HSCT [55,56,57]. These approaches represent a major step toward Wheldon’s vision of biologically targeted, spatially precise irradiation, enhancing tumoricidal effect where needed while potentially reducing systemic toxicity.
4.2. Ongoing Clinical Trials
Over the past decades, several prospective studies have confirmed TBI’s role as part of conditioning for allogeneic HSCT. In parallel, the introduction of haploidentical HSCT has broadened donor availability and led to the evaluation of TBI-containing regimens also in this setting [10,58]. Accordingly, research focus has shifted from whether to irradiate to how best to do so, with a special focus on reducing long-term sequelae, especially in paediatric and AYA survivors (endocrine and metabolic disfunctions, neurocognitive effects, secondary malignancies, cataracts, etc.) [14].
Treatment-modality optimisation is now being investigated by a multi-institutional phase-I study (NCT04281199), which is testing whether highly conformal IMRT-TBI (12 Gy in 6 fractions) can keep the mean lung dose below 8 Gy while still covering ≥ 85% of the target volume. If successful, it will provide the first prospective evidence that modern IMRT can cut the historical pulmonary cost of myeloablative TBI without losing antileukaemic potency [56].
Complementing this dosimetric approach, a phase-II escalation trial at City of Hope (NCT04262843) is pushing IMRT TMI/TMLI to 20 Gy while enforcing the same < 8 Gy lung- and kidney-dose threshold, seeking to determine whether selectively intensifying marrow/lymphoid dose can further improve disease control in high-risk acute myeloid leukaemia/myelodysplastic syndrome [57].
Dose de-escalation is being pursued in two parallel “mini-TBI” studies that give only 4 Gy in two fractions: City of Hope combines low-dose TBI with fludarabine-melphalan for haplo-identical grafts (NCT03118492), while Roswell Park adds post-transplant cyclophosphamide to curb GVHD (NCT03192397) [58,59].
Biology-guided omission of irradiation is under investigation in the phase-II EndRAD trial (NCT03509961), which assigns children, adolescents, and young adults with B-cell acute lymphoblastic leukaemia who are negative for minimal residual disease as shown by next-generation sequencing to a fully myeloablative, radiation-free conditioning regimen [60]. Patients who remain positive for minimal residual disease continue on the conventional observational arm that delivers 12 Gy total-body irradiation in 6 fractions. The study will clarify whether these excellent-risk patients can forego total-body irradiation and its long-term pulmonary, endocrine, and neuro-cognitive complications without sacrificing disease control [60].
At the opposite end of the spectrum, dose-intensification strategies are also emerging. The single-arm phase-II studies by the University of Chicago (NCT06802315, NCT04187105) are testing 9–6 Gy IM-TMI boosts within fludarabine–busulfan or fludarabine–cyclophosphamide–based conditioning to high-risk acute myeloid leukaemia/myelodysplastic syndrome or haplo-identical recipients, respectively, aiming to expand transplant eligibility while capping critical-organ doses [61,62].
In parallel, a first-in-human phase I/II study at Fred Hutchinson combines an ultra-low-dose (≤2 Gy) TBI with an astatine-211-labelled anti-CD45 monoclonal antibody (NCT04083183) to test “α-conditioning” as a highly targeted, organ-sparing alternative for non-malignant indications [63].
Ongoing trials are listed in Table 2.
Taken together, these complementary trials, spanning conformal delivery, de-escalation, complete omission and selective intensification, promise to delineate a data-driven therapeutic window in which irradiation can be tailored to each patient’s disease biology and tolerance, thereby maximising cure while minimising late toxicities.
5. Experience at the University Hospital Policlinico G. Rodolico San Marco
To give a complementary clinical insight on local protocols and their development in response to technological advancements, here we report over one decade of institutional experience at the University Hospital Policlinico “G. Rodolico-S. Marco” of Catania.
Between January 2015 and July 2025, we delivered conventional TBI to 77 patients, including both adults and paediatric patients (median age 28 years old). The underlying diagnoses were B- or T-cell acute lymphoblastic leukaemia (n = 52), acute myeloid leukaemia (n = 5), Hodgkin or non-Hodgkin lymphoma (n = 13), aplastic anaemia (n = 3), mycosis fungoides (n = 3), and treatment-refractory chronic myeloid leukaemia (n = 1).
Most patients (n = 51) received an immunosuppressive schedule of 12 Gy in 2 Gy fractions, administered twice daily with a 4–6 h interfraction interval. Four patients followed an intermediate regimen of 8 Gy (2 Gy × 2 daily) combined with intensified systemic therapy, while 17 patients, mainly with relapsed Hodgkin or non-Hodgkin lymphoma, aplastic anaemia, or mycosis fungoides, underwent single-fraction 2 Gy.
For all 8–12 Gy courses, a non-contrast chest simulation CT scan was performed (Figure 1) to quantify lung density and to enable fabrication of individual blocks (Figure 2) that limited pulmonary exposure to ≈8–10 Gy.
Treatment was delivered in a semi-standing position with the gantry at 90° and the collimator at 45°, placing the whole body along the field diagonally (Figure 3).
A Plexiglas spoiler countered the skin-sparing effect of 6 MV photons (Figure 4). Alternating AP and PA fields ensured dose homogeneity; this was verified each fraction by in vivo dosimetry; orthogonal X-rays confirmed block placement before irradiation (Figure 5). All patients received anti-emetic and corticosteroid pre-medication.
Despite the demanding schedule, 71 of 77 patients (92%) completed TBI, and 68/77 (95.8%) proceeded with allogeneic haematopoietic cell transplantation. Of these, 18/68 patients (26.5%) received haploidentical grafts. Acute toxicities were monitored during the entire treatment course and included nausea and vomiting, syncope, and severe fatigue in up to 80% of patients. Daily full blood counts were used to monitor leucopenia and neutropenia.
Although long-term toxicity data was not available at the time of analysis, results confirm the feasibility of conventional TBI in the context of HSCT, highlighting, however, the need for optimal interdisciplinary collaboration and appropriate patient monitoring and support.
To keep pace with recent technological advances, our centre is now developing intensity-modulated protocols for TBI and will possibly start including TMI and TMLI in clinical practice.
6. Conclusions
Over seven decades, total body irradiation has evolved from single-shot cobalt treatments to sophisticated intensity-modulated platforms, yet its biological rationale, such as uniform myeloablation and potent immunosuppression, remains unchanged. The literature consistently shows that fractionated myeloablative schedules (e.g., 12 Gy in 6 fractions) improve survival and reduce GVHD with historical single-dose regimens, while reduced-intensity protocols (2–4 Gy) safely extend transplant eligibility to older or medically fragile patients. Parallel advances in treatment delivery, including VMAT, HT, and emerging total-marrow/lymphoid irradiation, now allow centimetre-level conformality and organ sparing without compromising engraftment or disease control.
Our 10-year series of 77 paediatric and adult recipients treated with conventional extended-SSD TBI corroborates these broader observations. A 92% completion rate, acceptable acute toxicity, and successful engraftment in nearly all eligible patients affirm the robustness of traditional techniques even in the era of precision radiotherapy. Nonetheless, the dosimetric discipline required—individual lung blocks, in vivo dosimetry, and labour-intensive setup—highlights a pressing need to streamline workflows. Commissioning of IMRT-based TMI or TMLI at our centre is therefore a logical next step, promising shorter sessions, lower pulmonary doses, and a platform for future biologically guided dose painting.
Looking ahead, the convergence of molecular disease monitoring, and targeted radionuclide conjugates foreshadows a shift from “one-size-fits-all” irradiation to truly personalised conditioning. Ongoing phase-I/II trials testing conformal 12 Gy regimens, 4 Gy de-escalation, and MRD-stratified omission of TBI will clarify how best to balance efficacy against long-term sequelae. Multi-institutional cooperation and prospective data collection—including late toxicity, quality-of-life, and health-economic endpoints—will be essential to translate these innovations into new standards of care.
In summary, total body irradiation continues to be an indispensable, versatile component of haematopoietic stem-cell transplantation. By marrying the proven curative potential of TBI with modern imaging, dosimetry, and biologic insight, the field is poised to deliver safer, more effective, and more patient-tailored conditioning in the decade to come.
Conceptualization, M.C.L.G., R.M., G.A., R.L.E.L., M.L.R., G.M. and C.S.; methodology, M.L.R. and G.M.; investigation, A.B., P.V.F., S.P. (Stefano Palmucci), E.D., C.I., and L.A.; resources, R.M., G.A., R.L.E.L., V.S. and V.A.L.M.; data curation, M.L.R., G.M., A.B., P.V.F., S.P. (Stefano Palmucci), E.D., C.I., and L.A.; writing—original draft preparation, M.C.L.G., and R.M.; writing—review and editing, M.C.L.G., R.M., G.A., R.L.E.L., S.P. (Stefano Pergolizzi) and C.S.; visualisation, A.B., P.V.F., S.P. (Stefano Palmucci), E.D., C.I., L.A., S.P. (Stefano Pergolizzi) and C.S.; supervision, C.S. and S.P. (Stefano Pergolizzi). All authors have read and agreed to the published version of the manuscript.
Not applicable.
Written informed consent has been obtained from the patients to publish this paper.
The data presented in this study are available on request from the corresponding author. The data are not publicly available due to privacy restrictions.
The authors declare no conflicts of interest.
Footnotes
Disclaimer/Publisher’s Note: The statements, opinions and data contained in all publications are solely those of the individual author(s) and contributor(s) and not of MDPI and/or the editor(s). MDPI and/or the editor(s) disclaim responsibility for any injury to people or property resulting from any ideas, methods, instructions or products referred to in the content.
Figure 1 Simulation CT scan (axial, sagittal, and coronal axes) visualised on our institutional treatment planning system.
Figure 2 Individual blocks for lung shielding.
Figure 3 Visual representation of a conventional TBI treatment session.
Figure 4 Plexiglas spoiler for skin-sparing effect.
Figure 5 Lung shielding placement verification.
Summary of fractionation schedules per conditioning regime and corresponding indications.
| Schedule | Administration | Conditioning | Clinical Indications | Key Notes |
|---|---|---|---|---|
| 10 Gy/1 fraction [ | QD over 1 day | Historical myeloablative | Early AML/ALL transplants (pre-1980) | Excess toxicity; largely obsolete |
| 12 Gy/6 fractions [ | BID over 3 days | Standard myeloablative | Front-line paediatric and adult allo-HSCT | Benchmark OS & GVHD profile |
| 12–13.5 Gy/8–9 fractions [ | BID over 4–4.5 days | Standard myeloablative | Centres favouring smaller fraction sizes | Slightly longer course |
| 12–13.2 Gy/10–11 fractions [ | TID over 4 days | Hyper-fractionated myeloablative | Centres favouring smaller fraction sizes | Lowest per-fraction dose |
| 12 Gy/4 fractions [ | QD over 4 days | Condensed myeloablative | Logistical constraints or BID-intolerant patients | Requires meticulous lung shielding |
| 4 Gy testicular boost/2 fractions [ | QD over 2 days | Sanctuary boost | Boys with ALL | Practice varies; omission under study |
| 1 Gy/1 fraction [ | QD over 1 day | Minimal-intensity/experimental | Older/comorbid adults; | Very low toxicity |
| 2 Gy/1 fraction [ | QD over 1 day | Standard RIC | Older/comorbid adults | Very low toxicity |
| 4 Gy/2 fractions [ | BID or QD over 1–2 days | Augmented RIC | Mismatched donors or high-risk disease | Balances better engraftment vs. added toxicity |
QD = once daily; BID = twice daily; TID = three times daily; RIC = reduced-intensity conditioning; AML = acute myeloid leukaemia; ALL = acute lymphoblastic leukaemia; HSCT = haematopoietic stem-cell transplantation; OS = overall survival; GVHD = graft-versus-host disease.
Ongoing trials details.
| Trial | Aim | Investigational Arm | Status |
|---|---|---|---|
| NCT04281199 (Phase I) [ | IMRT-TBI dosimetric optimisation | Conformal IMRT-TBI 12 Gy/6 fx | Active, not recruiting |
| NCT04262843 (Phase II) [ | Escalated TMI/TMLI to 20 Gy | TMLI 20 Gy (lung < 8 Gy) | Recruiting at the main site, but several satellites are now “Active-not-recruiting” |
| NCT03118492 (Phase I) [ | 4 Gy mini-TBI + Flu-Mel in haplo HSCT | 4 Gy (2 × 2 Gy) + Flu-Mel | Active, not-recruiting |
| NCT03192397 (Phase Ib/II) [ | 4 Gy mini-TBI + PTCy | 4 Gy (2 × 2 Gy) + PTCy | Active, not recruiting |
| NCT03509961 (EndRAD, Phase II) [ | MRD-guided omission of TBI | Chemo vs. 12 Gy TBI | Recruiting |
| NCT06802315 (BMT-13, Phase II) [ | High-risk AML/CML/MDS | IM-TMI 9 Gy + Flu-Bu | Active, not recruiting |
| NCT04187105 (BMT-06, Phase II) [ | Haplo-HSCT dose-escalated TMI | IM-TMI 6 Gy + Flu-Cy + PTCy | Recruiting |
| NCT04083183 (Phase I/II) [ | 211At-anti-CD45 α-conditioning | ≤2 Gy TBI + 211At-BC8-B10 | Recruiting |
AML = acute myeloid leukaemia; At = astatine-211; α-conditioning = targeted astatine-211–anti-CD45 alpha-particle conditioning; Bu = busulfan; CML = chronic myeloid leukaemia; Cy = cyclophosphamide; Flu = fludarabine; fx = fraction (s); Gy = grey; HSCT = haematopoietic stem-cell transplantation; IM-TMI = intensity-modulated total marrow irradiation; IMRT = intensity-modulated radiotherapy; MDS = myelodysplastic syndrome; Mel = melphalan; MRD = measurable residual disease; PTCy = post-transplant cyclophosphamide; TBI = total body irradiation; TMI = total marrow irradiation; TMLI = total marrow and lymphoid irradiation.
1. Paix, A.; Antoni, D.; Waissi, W.; Ledoux, M.P.; Bilger, K.; Fornecker, L.; Noel, G. Total body irradiation in allogeneic bone marrow transplantation conditioning regimens: A review. Crit. Rev. Oncol./Hematol.; 2018; 123, pp. 138-148. [DOI: https://dx.doi.org/10.1016/j.critrevonc.2018.01.011]
2. Blume, K.G.; Weissman, I.L. E. Donnall Thomas (1920–2012). Proc. Natl. Acad. Sci. USA; 2012; 109, pp. 20777-20778. [DOI: https://dx.doi.org/10.1073/pnas.1218913109] [PubMed: https://www.ncbi.nlm.nih.gov/pubmed/23197829]
3. Johns, H.E.; Bates, L.M.; Epp, E.R.; Cormack, D.V.; Fedorux, S.O.; Morrison, A.; Dixon, W.R.; Garrett, C. 1,000-Curie Cobalt 60 Units for Radiation Therapy. Nature; 1951; 168, pp. 1035-1036. [DOI: https://dx.doi.org/10.1038/1681035a0]
4. Sabloff, M.; Tisseverasinghe, S.; Babadagli, M.E.; Samant, R. Total Body Irradiation for Hematopoietic Stem Cell Transplantation: What Can We Agree on?. Curr. Oncol.; 2021; 28, pp. 903-917. [DOI: https://dx.doi.org/10.3390/curroncol28010089]
5. Malicki, J.; Kosicka, G.; Stryczyńska, G.; Wachowiak, J. Cobalt 60 versus 15 MeV photons during total body irradiation: Doses in the critical organs and complexicity of the procedure. Ann. Transplant.; 2001; 6, pp. 18-22.
6. Chiang, Y.; Tsai, C.H.; Kuo, S.H.; Liu, C.Y.; Yao, M.; Li, C.C.; Huang, S.Y.; Ko, B.S.; Lin, C.T.; Hou, H.A.
7. Chilukuri, S.; Sundar, S.; Thiyagarajan, R.; Easow, J.; Sawant, M.; Krishanan, G.; Panda, P.K.; Sharma, D.; Jalali, R. Total marrow and lymphoid irradiation with helical tomotherapy: A practical implementation report. Radiat. Oncol. J.; 2020; 38, pp. 207-216. [DOI: https://dx.doi.org/10.3857/roj.2020.00528] [PubMed: https://www.ncbi.nlm.nih.gov/pubmed/33012149]
8. Hoeben, B.A.W.; Wong, J.Y.C.; Fog, L.S.; Losert, C.; Filippi, A.R.; Bentzen, S.M.; Balduzzi, A.; Specht, L. Total Body Irradiation in Haematopoietic Stem Cell Transplantation for Paediatric Acute Lymphoblastic Leukaemia: Review of the Literature and Future Directions. Front. Pediatr.; 2021; 9, 774348. [DOI: https://dx.doi.org/10.3389/fped.2021.774348] [PubMed: https://www.ncbi.nlm.nih.gov/pubmed/34926349]
9. Heinzelmann, F.; Ottinger, H.; Müller, C.H.; Allgaier, S.; Faul, C.; Bamberg, M.; Belka, C. Total-body irradiation–Role and indications: Results from the German Registry for Stem Cell Transplantation (DRST). Strahlenther. Onkol.; 2006; 182, pp. 222-230. [DOI: https://dx.doi.org/10.1007/s00066-006-1468-1]
10. Jung, J.; Lee, H.; Suh, Y.G.; Eom, H.S.; Lee, E. Current Use of Total Body Irradiation in Haploidentical Allogeneic Hematopoietic Stem Cell Transplantation. J. Korean Med. Sci.; 2021; 36, e55. [DOI: https://dx.doi.org/10.3346/jkms.2021.36.e55]
11. Thomas, E.D.; Clift, R.A.; Hersman, J.; Sanders, J.E.; Stewart, P.; Buckner, C.D.; Fefer, A.; McGuffin, R.; Smith, J.W.; Storb, R. Marrow transplantation for acute nonlymphoblastic leukemic in first remission using fractionated or single-dose irradiation. Int. J. Radiat. Oncol. Biol. Phys.; 1982; 8, pp. 817-821. [DOI: https://dx.doi.org/10.1016/0360-3016(82)90083-9] [PubMed: https://www.ncbi.nlm.nih.gov/pubmed/7050046]
12. Shank, B.; Chu, F.C.; Dinsmore, R.; Kapoor, N.; Kirkpatrick, D.; Teitelbaum, H.; Reid, A.; Bonfiglio, P.; Simpson, L.; O’Reilly, R.J. Hyperfractionated total body irradiation for bone marrow transplantation. Results in seventy leukemia patients with allogeneic transplants. Int. J. Radiat. Oncol. Biol. Phys.; 1983; 9, pp. 1607-1611. [DOI: https://dx.doi.org/10.1016/0360-3016(83)90412-1] [PubMed: https://www.ncbi.nlm.nih.gov/pubmed/6358154]
13. Wong, J.Y.C.; Filippi, A.R.; Dabaja, B.S.; Yahalom, J.; Specht, L. Total Body Irradiation: Guidelines from the International Lymphoma Radiation Oncology Group (ILROG). Int. J. Radiat. Oncol. Biol. Phys.; 2018; 101, pp. 521-529. [DOI: https://dx.doi.org/10.1016/j.ijrobp.2018.04.071]
14. Blomain, E.S.; Jiang, A.; Donaldson, S.S.; Agarwal, R.; Bertaina, A.; Shyr, D.; Eisenberg, M.L.; Hoppe, R.T.; Hiniker, S.M.; Oh, J. Effect of Testicular Boost in Children with Leukemia Receiving Total Body Irradiation and Stem Cell Transplant: A Single-Institution Experience. Adv. Radiat. Oncol.; 2022; 8, 101071. [DOI: https://dx.doi.org/10.1016/j.adro.2022.101071]
15. Bruserud, O.; Reikvam, H.; Kittang, A.O.; Ahmed, A.B.; Tvedt, T.H.; Sjo, M.; Hatfield, K.J. High-dose etoposide in allogeneic stem cell transplantation. Cancer Chemother. Pharmacol.; 2012; 70, pp. 765-782. [DOI: https://dx.doi.org/10.1007/s00280-012-1990-z] [PubMed: https://www.ncbi.nlm.nih.gov/pubmed/23053272]
16. Bunin, N.; Aplenc, R.; Kamani, N.; Shaw, K.; Cnaan, A.; Simms, S. Randomized trial of busulfan vs total body irradiation containing conditioning regimens for children with acute lymphoblastic leukemia: A Pediatric Blood and Marrow Transplant Consortium study. Bone Marrow Transplant.; 2003; 32, pp. 543-548. [DOI: https://dx.doi.org/10.1038/sj.bmt.1704198]
17. Davies, S.M.; Ramsay, N.K.; Klein, J.P.; Weisdorf, D.J.; Bolwell, B.; Cahn, J.Y.; Camitta, B.M.; Gale, R.P.; Giralt, S.; Heilmann, C.
18. Peters, C.; Dalle, J.H.; Locatelli, F.; Poetschger, U.; Sedlacek, P.; Buechner, J.; Shaw, P.J.; Staciuk, R.; Ifversen, M.; Pichler, H.
19. Ballen, K.K.; Becker, P.S.; Emmons, R.V.; Fitzgerald, T.J.; Hsieh, C.C.; Liu, Q.; Heyes, C.; Clark, Y.; Levy, W.; Lambert, J.F.
20. Ballen, K.K.; Colvin, G.; Porter, D.; Quesenberry, P.J. Low dose total body irradiation followed by allogeneic lymphocyte infusion for refractory hematologic malignancy—An updated review. Leuk. Lymphoma; 2004; 45, pp. 905-910. [DOI: https://dx.doi.org/10.1080/10428190310001628167]
21. McSweeney, P.A.; Niederwieser, D.; Shizuru, J.A.; Sandmaier, B.M.; Molina, A.J.; Maloney, D.G.; Chauncey, T.R.; Gooley, T.A.; Hegenbart, U.; Nash, R.A.
22. Kornblit, B.; Maloney, D.G.; Storb, R.; Storek, J.; Hari, P.; Vucinic, V.; Maziarz, R.T.; Chauncey, T.R.; Pulsipher, M.A.; Bruno, B.
23. Modi, D.; Chi, J.; Kim, S.; Deol, A.; Ayash, L.; Ratanatharathorn, V.; Uberti, J.P. Outcomes of Fludarabine, Melphalan and Total Body Irradiation as a Reduced Intensity Conditioning Regimen in Matched Donor Allogeneic Peripheral Blood Stem Cell Transplantation. Transplant. Cell. Ther.; 2021; 27, pp. 665.e1-665.e7. [DOI: https://dx.doi.org/10.1016/j.jtct.2021.04.029]
24. Nakamae, H.; Storer, B.E.; Storb, R.; Storek, J.; Chauncey, T.R.; Pulsipher, M.A.; Petersen, F.B.; Wade, J.C.; Maris, M.B.; Bruno, B.
25. Jacobs, M.L.; Pape, L. Dosimetry for a total-body irradiation chamber. Radiology; 1961; 77, pp. 788-792. [DOI: https://dx.doi.org/10.1148/77.5.788] [PubMed: https://www.ncbi.nlm.nih.gov/pubmed/14450690]
26. Sahler, O.D. Development of a room specifically designed for total-body irradiation. Radiology; 1959; 72, pp. 266-267. [DOI: https://dx.doi.org/10.1148/72.2.266]
27. Leung, P.M.; Rider, W.D.; Webb, H.P.; Aget, H.; Johns, H.E. Cobalt-60 therapy unit for large field irradiation. Int. J. Radiat. Oncol. Biol. Phys.; 1981; 7, pp. 705-712. [DOI: https://dx.doi.org/10.1016/0360-3016(81)90461-2] [PubMed: https://www.ncbi.nlm.nih.gov/pubmed/7287530]
28. Peters, V.G.; Herer, A.S. Modification of a standard cobalt-60 unit for total body irradiation at 150 cm SSD. Int. J. Radiat. Oncol. Biol. Phys.; 1984; 10, pp. 927-932. [DOI: https://dx.doi.org/10.1016/0360-3016(84)90397-3]
29. Quast, U. Physical treatment planning of total-body irradiation: Patient translation and beam-zone method. Med. Phys.; 1985; 12, pp. 567-574. [DOI: https://dx.doi.org/10.1118/1.595677]
30. Pla, M.; Chenery, S.G.; Podgorsak, E.B. Total body irradiation with a sweeping beam. Int. J. Radiat. Oncol. Biol. Phys.; 1983; 9, pp. 83-89. [DOI: https://dx.doi.org/10.1016/0360-3016(83)90214-6]
31. Litoborska, J.; Piotrowski, T.; Jodda, A.; Malicki, J. Evolution of treatment planning and dose delivery methods during radiotherapy for patients undergoing bone marrow transplantation: A review. Nukleonika; 2020; 65, pp. 19-30. [DOI: https://dx.doi.org/10.2478/nuka-2020-0003]
32. Girinsky, T.; Socie, G.; Ammarguellat, H.; Cosset, J.M.; Briot, E.; Bridier, A.; Gluckman, E. Consequences of two different doses to the lungs during a single dose of total body irradiation: Results of a randomized study on 85 patients. Int. J. Radiat. Oncol. Biol. Phys.; 1994; 30, pp. 821-824. [DOI: https://dx.doi.org/10.1016/0360-3016(94)90355-7]
33. Mekdash, H.; Shahine, B.; Jalbout, W.; Chehab, C.; Abdel Khalek, H.; Youssef, B. A simple technique for an accurate shielding of the lungs during total body irradiation. Tech. Innov. Patient Support Radiat. Oncol.; 2017; 3–4, pp. 13-18. [DOI: https://dx.doi.org/10.1016/j.tipsro.2017.07.001]
34. Zeverino, M.; Agostinelli, S.; Taccini, G.; Cavagnetto, F.; Garelli, S.; Gusinu, M.; Vagge, S.; Barra, S.; Corvò, R. Advances in the implementation of helical tomotherapy-based total marrow irradiation with a novel field junction technique. Med. Dosim. Off. J. Am. Assoc. Med. Dosim.; 2012; 37, pp. 314-320. [DOI: https://dx.doi.org/10.1016/j.meddos.2011.12.001]
35. Hui, S.K.; Kapatoes, J.; Fowler, J.; Henderson, D.; Olivera, G.; Manon, R.R.; Gerbi, B.; Mackie, T.R.; Welsh, J.S. Feasibility study of helical tomotherapy for total body or total marrow irradiation. Med. Phys.; 2005; 32, pp. 3214-3224. [DOI: https://dx.doi.org/10.1118/1.2044428]
36. Hui, C.; Simiele, E.; Lozko, Y.; Romero, I.; Skinner, L.; Binkley, M.S.; Hoppe, R.; Kovalchuk, N.; Hiniker, S.M. Volumetric modulated arc therapy total body irradiation improves toxicity outcomes compared to 2D total body irradiation. Front. Oncol.; 2024; 14, 1459287. [DOI: https://dx.doi.org/10.3389/fonc.2024.1459287] [PubMed: https://www.ncbi.nlm.nih.gov/pubmed/39351359]
37. Wong, J.Y.C.; Filippi, A.R.; Scorsetti, M.; Hui, S.; Muren, L.P.; Mancosu, P. Total marrow and total lymphoid irradiation in bone marrow transplantation for acute leukaemia. Lancet. Oncol.; 2020; 21, pp. e477-e487. [DOI: https://dx.doi.org/10.1016/S1470-2045(20)30342-9] [PubMed: https://www.ncbi.nlm.nih.gov/pubmed/33002443]
38. Aydogan, B.; Mundt, A.J.; Roeske, J.C. Linac-based intensity modulated total marrow irradiation (IM-TMI). Technol. Cancer Res. Treat.; 2006; 5, pp. 513-519. [DOI: https://dx.doi.org/10.1177/153303460600500508] [PubMed: https://www.ncbi.nlm.nih.gov/pubmed/16981794]
39. Dogliotti, I.; Levis, M.; Martin, A.; Bartoncini, S.; Felicetti, F.; Cavallin, C.; Maffini, E.; Cerrano, M.; Bruno, B.; Ricardi, U.
40. Loginova, A.A.; Tovmasian, D.A.; Lisovskaya, A.O.; Kobyzeva, D.A.; Maschan, M.A.; Chernyaev, A.P.; Egorov, O.B.; Nechesnyuk, A.V. Optimized Conformal Total Body Irradiation methods with Helical TomoTherapy and Elekta VMAT: Implementation, Imaging, Planning and Dose Delivery for Pediatric Patients. Front. Oncol.; 2022; 12, 785917. [DOI: https://dx.doi.org/10.3389/fonc.2022.785917]
41. Salz, H.; Bohrisch, B.; Howitz, S.; Banz, N.; Weibert, K.; Wiezorek, T.; Wendt, T.G. Intensity-modulated Total Body Irradiation (TBI) with TomoDirect™. Radiat. Oncol.; 2015; 10, 58. [DOI: https://dx.doi.org/10.1186/s13014-015-0362-3] [PubMed: https://www.ncbi.nlm.nih.gov/pubmed/25889227]
42. Hoeben, B.A.W.; Saldi, S.; Aristei, C.; Engellau, J.; Ocanto, A.; Hiniker, S.M.; Misson-Yates, S.; Kobyzeva, D.A.; Pazos, M.; George Mikhaeel, N.
43. Amin, A.E.; Wheldon, T.E.; O’Donoghue, J.A.; Barrett, A. Radiobiological modeling of combined targeted 131I therapy and total body irradiation for treatment of disseminated tumors of differing radiosensitivity. Int. J. Radiat. Oncol. Biol. Phys.; 1993; 27, pp. 323-330. [DOI: https://dx.doi.org/10.1016/0360-3016(93)90244-P]
44. Wheldon, T.E. The radiobiological basis of total body irradiation. Br. J. Radiol.; 1997; 70, pp. 1204-1207. [DOI: https://dx.doi.org/10.1259/bjr.70.840.9505837]
45. Bosman, F.T.; Visser, B.C.; van Oeveren, J. Apoptosis: Pathophysiology of programmed cell death. Pathol. Res. Pract.; 1996; 192, pp. 676-683. [DOI: https://dx.doi.org/10.1016/S0344-0338(96)80089-6]
46. Chen, X.; Zhang, T.; Su, W.; Dou, Z.; Zhao, D.; Jin, X.; Lei, H.; Wang, J.; Xie, X.; Cheng, B.
47. Li, X.; Chen, F.; Zhu, Q.; Ding, B.; Zhong, Q.; Huang, K.; Jiang, X.; Wang, Z.; Yin, C.; Zhu, Y.
48. Uckun, F.M.; Chandan-Langlie, M.; Jaszcz, W.; Obuz, V.; Waddick, K.; Song, C.W. Radiation damage repair capacity of primary clonogenic blasts in acute lymphoblastic leukemia. Cancer Res.; 1993; 53, pp. 1431-1436.
49. Lo Greco, M.C.; Marletta, G.; Marano, G.; Fazio, A.; Buffettino, E.; Iudica, A.; Liardo, R.L.E.; Milazzotto, R.; Foti, P.V.; Palmucci, S.
50. Cosset, J.M.; Socie, G.; Dubray, B.; Girinsky, T.; Fourquet, A.; Gluckman, E. Single dose versus fractionated total body irradiation before bone marrow transplantation: Radiobiological and clinical considerations. Int. J. Radiat. Oncol. Biol. Phys.; 1994; 30, pp. 477-492. [DOI: https://dx.doi.org/10.1016/0360-3016(94)90031-0]
51. Shaffer, B.C.; Kebriaei, P.; de Lima, M.; Jimenez Jimenez, A.M. Measurable residual disease testing and allogeneic hematopoietic cell transplantation for AML: Adapting Pre-MEASURE to clinical practice. Bone Marrow Transplant.; 2025; 60, pp. 128-134. [DOI: https://dx.doi.org/10.1038/s41409-024-02481-2] [PubMed: https://www.ncbi.nlm.nih.gov/pubmed/39562717]
52. Muffly, L.; Liang, E.C.; Dolan, J.G.; Pulsipher, M.A. How I use next-generation sequencing-MRD to plan approach and prevent relapse after HCT for children and adults with ALL. Blood; 2024; 144, pp. 253-261. [DOI: https://dx.doi.org/10.1182/blood.2023023699] [PubMed: https://www.ncbi.nlm.nih.gov/pubmed/38728375]
53. Chen, R.; Palmer, J.M.; Martin, P.; Tsai, N.; Kim, Y.; Chen, B.T.; Popplewell, L.; Siddiqi, T.; Thomas, S.H.; Mott, M.
54. Mawad, R.; Gooley, T.A.; Rajendran, J.G.; Fisher, D.R.; Gopal, A.K.; Shields, A.T.; Sandmaier, B.M.; Sorror, M.L.; Deeg, H.J.; Storb, R.
55. Orchard, K.; Langford, J.; Guy, M.; Lewis, G.; Michopoulou, S.; Cooper, M.; Zvavamwe, C.; Richardson, D.; Lewington, V. Efficient bone marrow irradiation and low uptake by non-haematological organs with an yttrium-90-anti-CD66 antibody prior to haematopoietic stem cell transplantation. Bone Marrow Transplant.; 2024; 59, pp. 1247-1257. [DOI: https://dx.doi.org/10.1038/s41409-024-02317-z] [PubMed: https://www.ncbi.nlm.nih.gov/pubmed/38867006]
56. ClinicalTrials.gov. Study Details|TBI Using IMRT (VMAT or Tomotherapy) for the Prevention of Pul Toxicities in Patients Undergoing Donor SCT. Available online: https://www.clinicaltrials.gov/study/NCT04281199 (accessed on 19 June 2025).
57. ClinicalTrials.gov. Total Marrow and Lymphoid Irradiation as Conditioning Regimen Before Hematopoietic Cell Transplantation in Patients with Myelodysplastic Syndrome or Acute Leukemia. Available online: https://clinicaltrials.gov/study/NCT04262843?cond=%22Myelodysplastic%20Syndrome%22&intr=%22Vidarabine%22&viewType=Table&rank=5 (accessed on 19 June 2025).
58. ClinicalTrials.gov. Combination Chemotherapy, Total Body Irradiation, and Donor Blood Stem Cell Transplant in Treating Patients with Secondary Myelofibrosis. Available online: https://clinicaltrials.gov/study/NCT03118492 (accessed on 19 June 2025).
59. Home|ClinicalTrials.gov. Chemotherapy, Total Body Irradiation, and Post-Transplant Cyclophosphamide in Reducing Rates of Graft Versus Host Disease in Patients with Hematologic Malignancies Undergoing Donor Stem Cell Transplant. Available online: https://www.clinicaltrials.gov/study/NCT03192397 (accessed on 19 June 2025).
60. ClinicalTrials.gov. The EndRAD Trial: Eliminating Total Body Irradiation (TBI) for NGS-MRD Negative Children, Adolescents, and Young Adults with B-ALL. Available online: https://clinicaltrials.gov/study/NCT03509961 (accessed on 19 June 2025).
61. ClinicalTrials.gov. Study Details|Intensity Modulated Total Marrow Irradiation in Fully Human Leukocyte Antigen (HLA)-Matched and Partially-HLA Mismatched Allogeneic Transplantation Patients with High-Risk Acute Myeloid Leukemia (AML), Chronic Myeloid Leukemia (CML), and Myelodysplastic Syndrome (MDS). Available online: https://clinicaltrials.gov/study/NCT06802315?cond=acute%20leukemia&aggFilters=status:,phase:2 (accessed on 19 June 2025).
62. ClinicalTrials.gov. BMT-06: Study of Intensity Modulated Total Marrow Irradiation (IM-TMI). Available online: https://clinicaltrials.gov/study/NCT04187105 (accessed on 19 June 2025).
63. ClinicalTrials.gov. Study Details|Total Body Irradiation and Astatine-211-Labeled BC8-B10 Monoclonal Antibody for the Treatment of Nonmalignant Diseases. Available online: https://clinicaltrials.gov/study/NCT04083183?term=astatine&rank=3 (accessed on 19 June 2025).
You have requested "on-the-fly" machine translation of selected content from our databases. This functionality is provided solely for your convenience and is in no way intended to replace human translation. Show full disclaimer
Neither ProQuest nor its licensors make any representations or warranties with respect to the translations. The translations are automatically generated "AS IS" and "AS AVAILABLE" and are not retained in our systems. PROQUEST AND ITS LICENSORS SPECIFICALLY DISCLAIM ANY AND ALL EXPRESS OR IMPLIED WARRANTIES, INCLUDING WITHOUT LIMITATION, ANY WARRANTIES FOR AVAILABILITY, ACCURACY, TIMELINESS, COMPLETENESS, NON-INFRINGMENT, MERCHANTABILITY OR FITNESS FOR A PARTICULAR PURPOSE. Your use of the translations is subject to all use restrictions contained in your Electronic Products License Agreement and by using the translation functionality you agree to forgo any and all claims against ProQuest or its licensors for your use of the translation functionality and any output derived there from. Hide full disclaimer
© 2025 by the authors. Published by MDPI on behalf of the Lithuanian University of Health Sciences. Licensee MDPI, Basel, Switzerland. This article is an open access article distributed under the terms and conditions of the Creative Commons Attribution (CC BY) license (https://creativecommons.org/licenses/by/4.0/). Notwithstanding the ProQuest Terms and Conditions, you may use this content in accordance with the terms of the License.
Abstract
Background and Objectives: Total body irradiation (TBI) remains a cornerstone of conditioning for allogeneic haematopoietic stem-cell transplantation (HSCT). Whereas early research debated the need for irradiation, contemporary investigations focus on optimising dose, fractionation and delivery techniques. Material and Methods: We synthesised six decades of evidence, spanning from single-fraction cobalt treatments to modern helical tomotherapy and intensity-modulated total-marrow/lymphoid irradiation (TMI/TMLI). To complement the literature, we reported our institutional experience on 77 paediatric and adult recipients treated with conventional extended-source-to-skin-distance TBI at the University Hospital Policlinico “G. Rodolico–San Marco” between 2015 and 2025. Results: According to literature data, fractionated myeloablative schedules, typically 12 Gy in 6 fractions, provide superior overall survival and lower rates of severe graft-versus-host disease (GVHD) compared with historical single-dose regimens. Conversely, reduced-intensity protocols of 2–4 Gy broaden HSCT eligibility for older or comorbid patients with acceptable toxicity. Conformal planning reliably decreases mean lung dose without compromising engraftment, and early-phase trials are testing selective escalation to 16–20 Gy or omission of TBI in molecularly favourable cases. With regard to our institutional retrospective series, 92% of patients completed a 12-Gy regimen with only transient grade 1–2 nausea, fatigue or hypotension; all transplanted patients engrafted, and no grade ≥ 3 radiation pneumonitis occurred. Conclusions: Collectively, the published evidence and our experience support TBI as an irreplaceable component of HSCT conditioning and suggest that coupling it with advanced imaging, organ-sparing dosimetry and molecular response monitoring can deliver safer, more personalised therapy in the coming decade.
You have requested "on-the-fly" machine translation of selected content from our databases. This functionality is provided solely for your convenience and is in no way intended to replace human translation. Show full disclaimer
Neither ProQuest nor its licensors make any representations or warranties with respect to the translations. The translations are automatically generated "AS IS" and "AS AVAILABLE" and are not retained in our systems. PROQUEST AND ITS LICENSORS SPECIFICALLY DISCLAIM ANY AND ALL EXPRESS OR IMPLIED WARRANTIES, INCLUDING WITHOUT LIMITATION, ANY WARRANTIES FOR AVAILABILITY, ACCURACY, TIMELINESS, COMPLETENESS, NON-INFRINGMENT, MERCHANTABILITY OR FITNESS FOR A PARTICULAR PURPOSE. Your use of the translations is subject to all use restrictions contained in your Electronic Products License Agreement and by using the translation functionality you agree to forgo any and all claims against ProQuest or its licensors for your use of the translation functionality and any output derived there from. Hide full disclaimer
Details
; Milazzotto Roberto 2
; Acquaviva Grazia 2 ; Liardo Rocco Luca Emanuele 2
; Marano Giorgia 1
; La Rocca Madalina 1
; Basile, Antonio 3 ; Foti, Pietro Valerio 3
; Palmucci Stefano 3
; Emanuele, David 3 ; Iní Corrado 3
; Aliotta Lorenzo 4
; Salamone Vincenzo 5 ; La Monaca Viviana Anna 5 ; Pergolizzi Stefano 1
; Spatola Corrado 6
1 Radiation Oncology Unit, Department of Biomedical, Dental and Morphological and Functional Imaging Sciences, University of Messina, 98122 Messina, Italy; [email protected] (M.C.L.G.); [email protected] (G.M.); [email protected] (M.L.R.); [email protected] (S.P.)
2 Radiation Oncology Unit, University Hospital Policlinico “G. Rodolico-San Marco”, 95123 Catania, Italy; [email protected] (G.A.); [email protected] (R.L.E.L.)
3 Department of Medical Surgical Sciences and Advanced Technologies “G.F. Ingrassia”, University of Catania, 95123 Catania, Italy; [email protected] (A.B.); [email protected] (P.V.F.); [email protected] (S.P.); [email protected] (E.D.); [email protected] (C.I.); [email protected] (L.A.), Radiology I Unit, University Hospital Policlinico “G. Rodolico-San Marco”, 95123 Catania, Italy
4 Department of Medical Surgical Sciences and Advanced Technologies “G.F. Ingrassia”, University of Catania, 95123 Catania, Italy; [email protected] (A.B.); [email protected] (P.V.F.); [email protected] (S.P.); [email protected] (E.D.); [email protected] (C.I.); [email protected] (L.A.)
5 U.O.S.D. Fisica Sanitaria, University Hospital Policlinico “G. Rodolico-San Marco”, 95123 Catania, Italy; [email protected] (V.S.); [email protected] (V.A.L.M.)
6 Radiation Oncology Unit, University Hospital Policlinico “G. Rodolico-San Marco”, 95123 Catania, Italy; [email protected] (G.A.); [email protected] (R.L.E.L.), Department of Medical Surgical Sciences and Advanced Technologies “G.F. Ingrassia”, University of Catania, 95123 Catania, Italy; [email protected] (A.B.); [email protected] (P.V.F.); [email protected] (S.P.); [email protected] (E.D.); [email protected] (C.I.); [email protected] (L.A.)