Highlights
Integrative Co-Design Framework: We synthesize current advances in sensing, models, accuracy/reliability assessment, and hardware into a sensor–model–deployment–assessment framework that organizes evidence and design trade-offs for cuffless blood pressure monitoring. The framework seeks to balance precision and efficiency by jointly considering low-power edge AI, streamlined sensor architectures, and adaptive computational models, providing a structured basis for reproducible and clinically meaningful wearable solutions.
Pathways to Clinical Translation: We critically assess barriers to real-world deployment, offering actionable strategies to bridge the translational gap between laboratory innovations and scalable implementation in low-resource regions with minimal healthcare infrastructure.
Interdisciplinary Synthesis: By integrating cutting-edge advances in materials science, digital health, and embedded AI, we provide evidence-based recommendations to empower biomedical researchers, engineers, and data scientists in advancing equitable diagnostic solutions.
Accurate blood pressure (BP) monitoring is essential for preventing and managing cardiovascular disease. Advancements in materials science, medicine, flexible electronic, and artificial intelligence (AI) have enabled cuffless, unobtrusive BP monitoring systems, offering an alternative to traditional sphygmomanometers. However, extending these advances to real-world cardiovascular care particularly in resource-limited settings remains challenging due to constraints in computational resources, power efficiency, and deployment scalability. This review presents a comprehensive synthesis of AI-enhanced wearable BP monitoring, emphasizing its potential for personalized, scalable, and accessible healthcare. We systematically analyze the end-to-end system architecture, from mechano-electric sensing principles and AI-based estimation models to edge-aware deployment strategies tailored for low-resource environments. We further discuss clinical validation metrics and implementation barriers and prospective strategies. To bridge lab-to-field translation, we propose an innovative "sensor-model-deployment-assessment" co-design framework. This roadmap highlights how AI-enhanced BP technologies can support proactive hypertension control and promote cardiovascular health equity on a global scale.
Introduction
Hypertension is a major risk factor for cardiovascular diseases (CVDs), contributing significantly to global morbidity and mortality [1]. Accurate and continuous blood pressure (BP) monitoring is thus essential for early diagnosis, preventive care, and personalized intervention [2], particularly in resource-limited settings where access to episodic clinical measurement is limited. Traditional cuff-based BP measurement [3], despite its clinical acceptance, remains inherently episodic, cumbersome, and ill-suited for unobtrusive long-term monitoring [4]. Cuffless BP estimation represents a paradigm shift in non-invasive monitoring by eliminating the need for traditional cuffs, supporting cost-effective, continuous BP monitoring during daily life and holds potentials for personalized, proactive hypertension management [5].
Recent advances in sensing technologies have further empowered this field, enabling the acquisition of high-quality physiological data through increasingly miniaturized and affordable wearable devices [6, 7, 8–9]. Concurrently, artificial intelligence (AI) has emerged as a transformative tool for analyzing these complex signals, significantly enhancing the accuracy and robustness of cuffless BP estimation [8]. These trends have created new opportunities for deploying AI-driven BP monitoring beyond traditional healthcare settings. In particular, resource-limited settings—including low- and middle-income countries (LMICs), remote communities, and underserved populations in high-income countries—represent environments where the potential impact of wearable BP monitoring is especially high [6, 10, 11]. These settings are often characterized by limited healthcare infrastructure, insufficient access to trained personnel, and high unmet needs for hypertension screening and management. Yet, deploying state-of-the-art AI-based BP estimation methods in such settings requires a fundamental rethinking of system design. Existing approaches [12, 13] typically assume ample computational resources and high-quality signals, whereas real-world deployments in resource-constrained settings must navigate variable signal quality, limited model capacity, strict energy budgets, intermittent connectivity, and fragmented data ecosystems [14]. Scalable and reliable BP monitoring in such environments demands a delicate balance between model accuracy, algorithmic complexity, and hardware efficiency [15, 16–17]. It requires coordinated advances in sensor technology, learning frameworks, and edge-aware system deployment tailored to the realities of diverse populations and care infrastructures.
In this review, we provide an integrated perspective on the co-design of sensing, modeling, deployment and assessment, which is critical in real-world deployments but often has been neglected in prior reviews [7, 9, 18]. To be specific, we will systematically analyze the infrastructure, model, and deployment challenges of AI-based blood pressure estimation in resource-limited settings and summarize promising solutions and emerging directions for scalable and accessible healthcare. Figure 1 shows a system view that spans from hardware-proximal sensing, model design to execution across device–edge–cloud framework and comprehensive assessment. First, advanced wearable sensing technologies such as optical [19], electrical [20], mechanical [21], acoustic [22], and electromagnetic [23] enable various physiological signals acquisition. Second, the captured signals are further processed using physiological-based models, physics-based models, and data-driven machine learning models. These models are employed to estimate BP in three clinical scenarios, high-demand: beat-to-beat or BP waveform [24] for stress tests and acute monitoring; moderate-demand: intermittent BP for ambulatory follow-up and therapy titration; and low-demand: snapshot BP for spot checks. During the hardware deployment phase, challenges related to model optimization, compilation, scheduling, and adaptability across heterogeneous hardware platforms (microcontrollers, mobile devices, edge servers, and cloud platforms) will be discussed. Finally, we provide a comprehensive, system-level evaluation from both the model and device perspectives.
[See PDF for image]
Fig. 1
The sensor–model–deployment–assessment framework for AI-based cuffless BP estimation in real-world resource constraints
Wearable Sensing Technology for BP Measurement
We start with a concise overview of wearable sensing technologies for BP measurement, including photoplethysmography, electrical, acoustic, mechano-electric, and radar methods, focusing on their material innovation, sensor design, principles, performance, and limitations.
Photoplethysmography (PPG) is a cost-effective and non-invasive optical technique that measures changes in peripheral blood volume to monitor cardiovascular parameters (Fig. 2a). The fundamental operation involves light-emitting diodes (LEDs) illuminating tissue and photodetectors capturing backscattered or transmitted light, the intensity of which is modulated by cardiac cycle-induced blood flow variations. It is noteworthy that over 95% of the total detected light intensity originates from static tissue compartments (e.g., dermis, subcutaneous fat, muscle), with only a small variable component attributable to pulsatile blood volume changes, underscoring the challenge of extracting clean hemodynamic signals [19]. The interaction between light and biological tissue is highly wavelength-dependent. Ultraviolet light (10–380 nm) is predominantly absorbed by epidermal proteins, while visible (380–760 nm) and near-infrared (760–1300 nm) light penetrates several millimeters into tissue, allowing interrogation of deeper vasculature, with hemoglobin and water being the primary absorbers [25]. Widely adopted in wearable devices, PPG detects blood volume modulations driven by the cardiac cycle, enabling continuous cardiovascular monitoring (e.g., BP estimation) under appropriate calibration and constraints [19]. Recent innovations in flexible electronics have significantly improved PPG performance. Organic light-emitting diodes (OLEDs) [26], polymer LEDs [27], and hybrid inorganic–organic devices [28, 29] offer superior mechanical conformity to the skin, significantly improving signal acquisition stability and reducing motion-induced artifacts [26, 27]. Despite its advantages, the accuracy of PPG-based BP monitoring is compromised by motion artifacts, skin pigmentation, tissue thickness, ambient light interference, and even vasomotor activity [30, 31]. Additionally, PPG measurements are typically taken at peripheral sites (e.g., wrist or finger), which may not fully reflect central blood pressure, potentially limiting clinical precision.
[See PDF for image]
Fig. 2
Working principle of wearable BP sensing. a Optical sensing (PPG). Adapted from [40], Electronics, 2014, published by MDPI, under the terms of the Creative Commons CC-BY license. Adapted with permission from [41], Copyright Springer Nature, 2019. b ECG sensing.
Adapted from https://www.capitalheart.sg/what-does-an-abnormal-ecg-mean/. c Bioimpedance sensing. Adapted with permission from [32], Copyright Springer Nature, 2022. d Korotkoff sounds. Adapted from [22], Frontiers in Cardiovascular Medicine, 2022, under the terms of the Creative Commons CC-BY license. e Mechano-electric sensing, including piezoresistive, capacitive, piezoelectric, and triboelectric sensing. i), iii), and iv) Adapted with permission from [7], Copyright Springer Nature, 2025. ii) Adapted from [42], with permission from the Royal Society of Chemistry. f Ultrasound sensing. Adapted with permission from [7], Copyright Springer Nature, 2025. g Radar sensing. Adapted with permission from [23], Copyright Elsevier, 2023
Electrocardiography (ECG) measures the heart's electrical activity through skin electrodes on the chest or limbs (Fig. 2b), capturing characteristic waveforms including the P wave (atrial depolarization), QRS complex (ventricular depolarization), and T wave (ventricular repolarization) [20]. Besides, bioimpedance (BioZ) sensing (Fig. 2c) complements ECG by analyzing tissue electrical properties through applied high-frequency currents to detect arterial pulse-induced impedance variations, blood flow dynamics, and tissue dielectric properties [32, 33]. The performance of these electrical modalities is heavily dependent on the electrode–skin interface. Three primary electrode types are utilized: wet, dry, and non-contact. Wet electrodes (Ag/AgCl with hydrogel) [34] provide excellent initial signal quality and low impedance but suffer from long-term drying, irritancy, and performance degradation [35]. Dry electrodes, fabricated from conductive polymers or metal coatings, offer improved durability and comfort for sustained use, though they can be more prone to motion artifact without hydrogel [36]. Besides, non-contact electrodes, employing capacitive coupling through an insulating layer or elastic electrolytes, represent a significant advancement by eliminating direct skin contact, thereby maximizing user comfort and minimizing irritation and motion artifacts, making them suitable for wearable health monitoring applications [37].
Auscultatory methods based on Korotkoff sounds (K-sounds) represent a classical approach reinvented for modern wearables. These acoustic signals are produced by turbulent blood flow when an artery is partially constricted. Figure 2d reproduces the “core” theory of the mechanism and compares the changes in K-sounds produced by normal and abnormal cardiac function. These sounds are categorized into five phases, with the first (K1) marking the onset of SBP and the fifth (K5) denoting DBP as flow stabilizes [22]. While traditionally detected via stethoscope, modern approaches leverage acoustic sensors and signal processing techniques to capture and analyze these sounds. Recent advances incorporate deep learning to model the complex temporal and spectral patterns of K-sounds [22], demonstrating promise for automated, auscultatory-based BP assessment. Another application of acoustic sensing is the emerging flexible ultrasound technology (Fig. 2f). Leveraging the deep penetration and high spatiotemporal resolution of ultrasound waves, flexible ultrasound sensors have emerged as a powerful tool for non-invasively monitoring central blood pressure, which carries greater clinical significance than peripheral pressure [38, 39]. Acoustic sensors provide more direct physiological measures of pressure and flow, though often at increased cost and system complexity.
Mechano-electric sensors, which transduce mechanical pressure or vibration into quantifiable electrical signals through physical contact, encompass a diverse family including piezoresistive, capacitive, piezoelectric, and triboelectric types (Fig. 2e). Piezoresistive sensors operate on the principle of pressure-induced resistance change. Their performance is enhanced through material innovation (e.g., graphene porous networks [43], carbon nanotube/PDMS composites [44]) and microstructure design (e.g., micropillars, honeycombs), achieving high sensitivity (> 1 kPa⁻1), wide dynamic range, and excellent cyclic stability (> 8000 cycles) [45]. Capacitive sensors measure pressure via capacitance changes [42]. Performance optimization focuses on microstructured dielectrics (e.g., micro-pyramids [46] and others [47]) to concentrate stress and reduce modulus and optimized electrode materials (e.g., CNTs, ITO, metal coatings [48]) for flexibility and conductivity. This yields devices with exceptional sensitivity (down to 0.1 Pa), rapid response (~ 10 ms), and low hysteresis, helpful for high-fidelity pulse waveform acquisition [46, 49]. Piezoelectric sensors convert mechanical deformation into electrical charge through intrinsic material polarization [50]. Strategies to boost sensitivity include nanoparticle doping (e.g., BaTiO₃ in PVDF [51, 52]) and microstructuring (pyramids, waves [53]), achieving outputs exceeding 685 mV N−1. Ultra-flexible, skin-conformable patches fabricated via techniques like laser lift-off enable stable long-term monitoring [54, 55]. Dynamic analyses have further shown that piezoelectric sensors can faithfully capture arterial pulse waveforms, providing a mechanophysiological link to blood pressure [56]. Triboelectric sensors use contact electrification for self-powered sensing [7]. Nanostructured surfaces (e.g., nanogratings [57, 58]) and textile integration [59] have led to sensitive, comfortable, and robust devices [60, 61]. Moreover, system-level integration with wireless modules and low-power circuits has been achieved, allowing continuous, real-time hemodynamic monitoring in wearable form factors [62, 63], but these sensors still face challenges in static pressure detection and long-term stability.
Radar-based systems, particularly millimeter-wave radar (30–300 GHz), operate by emitting electromagnetic waves and analyzing the phase or frequency shift of signals reflected from the body surface, which vibrates minutely with each cardiac cycle (Fig. 2g). Systems like mmBP [64] employ advanced signal processing and neural networks to extract pulse signals and achieve accurate estimation, reporting deviations of 9.00% for SBP and 3.69% for DBP. Other systems integrate continuous-wave radar with BioZ and ECG to derive pulse arrival time (PAT) or pulse transit time (PTT) for BP estimation, showing strong statistical correlations with reference methods [23, 64]. Radar methods enable unique non-contact operation but are still evolving in terms of accuracy and robustness.
Nevertheless, accurate BP estimation using wearable sensor modalities, such as PPG, ECG, bioimpedance, and tonometry, is challenged by multiple physiological and environmental factors in practice [32, 65], including vasomotor activity (e.g., vasodilation or vasoconstriction), motion artifacts, skin tone variations, temperature and respiratory influences, arterial stiffness, sensor placement inconsistencies, contact-pressure drift, and physiological variability (e.g., heart rate, autonomic activity, blood viscosity) [31]. These factors distort sensor signals, complicating reliable BP measurement, particularly as vasomotor activity can counteract BP-related changes in PPG signals. Independent evaluations have underscored these challenges for commercial cuffless systems in real-world settings [66, 67–68]. Further comprehensive evaluation of continuous BP monitoring sensors and their model will be even more necessary and critical which will be discussed later in this review.
AI-Based BP Estimation Model
Advancements in wearable sensing technologies have enabled the acquisition of high-quality and varied physiological signals, which has spurred the development of AI-driven models for accurate, non-invasive BP estimation. This section, we will introduce these AI-driven models and their limitations, which are essential for their effective implementation in continuous physiological monitoring.
Current research on AI-based BP estimation models encompasses diverse methodological paradigms, each contributing to distinct aspects of accuracy, interpretability, and adaptability. Broadly, these methods can be categorized into two primary technical routes: physics- or physiology-informed models and machine learning models. Both approaches leverage observable hemodynamic signals and derived parameters such as PPG, ECG, pulse wave velocity (PWV), and pulse transit time (PTT) to estimate BP. However, these surrogate parameters do not directly reflect absolute BP values, necessitating calibration to establish a reliable mapping between the measured parameters and BP [69]. Mathematically, the cuffless BP estimation problem can be formulated as:
1
where x denotes the input vector derived from physiological measurements, Φ represents subject-specific physiological parameters, fθ is the mapping function parameterized by θ, the includes both beat-to-beat BP values (e.g., SBP and DBP) and the continuous BP waveform, reflecting dynamic hemodynamic changes over time. In the following, we will introduce the state-of-the-art BP methods, and the details of corresponding calibration strategies are referred to Note S1.Physics or Physiology Informed Network
Physics- or physiology-informed network (PPIN) incorporates cardiovascular and hemodynamic principles to model the relationship between physiological signals and BP. These models define the BP estimation mechanism through mathematical and biophysical equations rooted in domain knowledge. In this context, fθ is a predefined function derived from hemodynamic principles or physics laws; θ represents universal constants (e.g., blood density, geometric ratios, or fluid constants [70]) that are assumed to be invariant across individuals. In contrast, Φ represents subject-specific calibration parameters (e.g., baseline SBP₀ and DBP₀, reference PTT₀, and vascular elasticity coefficients [71, 72]). Unlike purely black-box AI methods, PPINs offer interpretable, knowledge-driven insights. The foundation of PPINs in blood pressure estimation is related to existing analytical techniques [7, 73], including the arterial BP physiological regulation, the arterial wall mechanics, and the arterial pulse wave propagation model.
The physiological regulation of arterial BP is influenced by arterial compliance, cardiac output (CO), systemic vascular resistance (SVR), and blood volume, Fig. 3a-i&ii. According to the Windkessel model, mean BP (MBP) = CO × SVR [74]. While CO is measurable, SVR is not, complicating BP modeling. Multi-wavelength pulse transit time (MWPTT [75]) and cardiovascular coupling models with heart rate and systolic time interval [76] improve BP estimation accuracy by correlating these parameters to SVR. Typically, these related parameters are determined by physiological signals like ECG and PPG [77], Fig. 3a-iii. In addition, factors such as vascular resistance, the renin–angiotensin–aldosterone system [78], arterial diameter, skin temperature [79], and blood viscosity [80] influence SVR, requiring further quantitative research. Besides, a more complicated cardiovascular hybrid modeling was developed by Shi et al. [81] to directly reconstruct arterial BP waveforms from PPG signals. In practice, the Winkessel model estimates systemic arterial compliance and total peripheral resistance from pulse pressure data, commonly used in cardiovascular research [74]. It uses calibrated arterial pressure waveforms, suitable for real-time monitoring. However, it assumes a lumped parameter system, limiting accuracy for localized pressure dynamics and complex vascular geometries, and requires precise calibration.
[See PDF for image]
Fig. 3
Physiology informed models for cuffless BP estimation. a Physiological regulation of arterial BP, influenced by CO, SVR, arterial compliance, and blood volume, modeled through the Windkessel framework. Adapted with permission from [76], Copyright IEEE, 2022. b Arterial wall mechanics-based models estimate BP from vessel elasticity or diameter changes, incorporating flexible ultrasound and resonance sonomanometry. i) Adapted with permission from the authors of [102]. ii) Adapted with permission from [84], Copyright Oxford University Press, 2024. c Pulse wave propagation models based on PTT, PWV, MWPTT, and vascular control to infer BP with improved robustness under physiological variability. i) Adapted with permission from [89], Copyright IEEE, 2016. ii) Adapted from [83], PNAS, 2019, under the terms of the Creative Commons CC-BY license. iii) Adapted from [94], Nat. Commun., 2021, under the terms of the Creative Commons CC-BY license
Arterial BP can also be estimated from arterial wall mechanics, i.e., arterial elasticity or distensibility, using local PWV or diameter variation [82]. The Hughes equation [83] provides an empirical arterial elasticity-BP link, while the distension-based BP model proposes an exponential relationship, unproven in microcirculation. Flexible ultrasound [34] enables direct vessel diameter measurement (Fig. 3b-i) that posits an exponential relationship between BP and arterial cross-sectional area: , where α is the vessel rigidity coefficient. However, α may vary with daily activities or physiological changes, posing challenges for long-term tracking accuracy without frequent recalibration. Alternatively, resonance sonomanometry [84] (Fig. 3b-ii) offers a calibration-free approach, capturing audio-induced arterial resonance. However, its application requires accurate arterial geometry measurement and has limited validation. These models leverage arterial wall biomechanics (e.g., elasticity, stiffness) to estimate BP via vessel deformation and stress–strain analysis. They are suited for detailed arterial property analysis, such as age-related vascular stiffening or hypertension-induced remodeling studies, relying on imaging modalities (e.g., ultrasound, MRI). While demanding high computational resources and expertise, they are limited by the need for high-resolution imaging and patient-specific data and are sensitive to assumptions about arterial properties.
Arterial pulse wave propagation models are well-known and fundamental to cuffless BP estimation, particularly those using PTT, PAT, or PWV (Fig. 3c). Classical models like Moens–Korteweg and Bramwell–Hill equations [70, 71], which link PWV with BP, are limited by idealized thin shell assumptions [85]. Ma et al. [83] introduced an analytical alternative based on the Fung hyperelastic model (Fig. 3c-i), expressing BP as , avoiding reliance on empirical assumptions, though requiring further validation. Recent studies have improved PTT measurement using signals such as ballistocardiography (BCG) [86], impedance cardiography (ICG) [87], seismocardiography (SCG), phonocardiography (PCG) [88], and multi-wavelength photoplethysmography (MWPPG, Fig. 3c-ii) [75, 89]. Modern extensions incorporate multimodal signal features to increase robustness [72, 90, 91]. Xiang et al. [79, 92] proposed multimodal physiological models integrating temperature, PPG, ECG, and IPG. These methods are suited to non-invasive BP estimation in clinical and wearable settings, especially for PTT or PWV techniques. It uses sensors to capture pulse wave signals (e.g., photoplethysmography, ECG) at multiple arterial locations. While adaptable for continuous monitoring, it needs robust signal processing to mitigate noise. Accuracy relies on precise transit time and distance measurements, which can be compromised by motion artifacts or anatomical differences. Furthermore, it may face challenges with complex wave reflections in impaired arteries.
In contrast to previous pulse analysis techniques, the volume clamp method, also known as the Penaz method [93, 94] (see Fig. 3c-iii), employs advanced vascular control strategies for fingertip pulse monitoring. This approach uses a high-precision controller to apply targeted pressure, maintaining constant vessel volume at the monitoring site to capture an optimal PPG signal. However, the method requires expensive, high-precision controllers and small cuffs to ensure precise pressure regulation and intimate contact with the finger. Additionally, it relies on initial calibration using an oscillometric method to ensure accuracy.
Recently, physics-informed neural networks (PINNs) embed physical laws such as continuity equations and nonlinear partial differential equations (PDEs) into the neural network's training process were developed [95, 96, 97, 98, 99–100], enabling models to learn from data while simultaneously respecting known physiological principles. Originally demonstrated in domains such as fluid mechanics [95] and power systems [96], PINNs are increasingly being applied to cardiovascular modeling. Sel et al. [97] employed a PINN architecture combining a two-layer CNN with bioimpedance signals, incorporating impedance-derived hemodynamic features (e.g., pulse wave velocity, arterial volume) into the model, reducing the requirement for ground-truth training data by a factor of ~ 15. Building upon this, a physics-informed temporal networks (PITN) with temporal blocks and adversarial contrastive learning [98], a DeepONet constrained by the Navier–Stokes equation with time-periodic conditions and Windkessel-type boundary conditions [99], and meta-learning with physics-driven modeling [100], were developed, respectively, to mitigate the interpretation and accuracy of the AI-based BP estimation. These examples illustrate the diversity of physical priors that can be embedded within PINNs. Some models enforce relatively simple hemodynamic relations (e.g., continuity of blood flow or Windkessel-type pressure–flow coupling), while others incorporate more complex formulations such as Navier–Stokes fluid dynamics or pulse wave propagation. Given the complexity of blood pressure regulation, it is imperative to add physiologically informed constraints. Since no single PINN formulation can comprehensively encapsulate the full regulatory spectrum, existing approaches prioritize core equations that are both mechanistically grounded and generalizable across individuals. Thanks to PDEs detailed physiological process, these methods are best suited for advanced research or precision medicine, where complex, patient-specific blood pressure dynamics are modeled using sparse or noisy data while enhancing interpretability [97]. The deployment of PINN requires significant computational resources, large datasets for training, and expertise in machine learning and hemodynamics. It is typically implemented in research-grade systems rather than real-time clinical settings. Most importantly, PINNs also depend on the quality of input data and may overfit if physical constraints are not well-defined [101].
A comparative analysis of their advantages and trade-offs would enhance its utility for guiding model selection and implementations, summarized in Table S1.
Machine Learning Model
Machine learning (ML)-based models have contributed significantly to the early development of cuffless BP estimation by capturing complex, nonlinear relationships between physiological signals and BP values without relying on explicit physiological equations. Figure 4 illustrates the flowchart of the process. In this context, fθ is a trainable function that maps inputs to BP estimations, θ denotes model hyperparameters (e.g., layer numbers, neurons count, and activation functions), Φ represents individual-specific or context-related variables (e.g., demographic attributes [103, 104–105]). Through joint optimization of fθ and Φ, ML models enhance both adaptability and accuracy across heterogeneous populations.
[See PDF for image]
Fig. 4
A schematic overview of machine learning-based blood pressure estimation method. Traditional ML methods rely on handcrafted features, whereas DL methods enable end-to-end estimation of beat-to-beat BP values, continuous BP waveforms, and 24-h dynamic BP values from multimodal physiological signals
Traditional ML methods typically utilize handcrafted features extracted from signals such as bioimpedance, PPG, or ECG, followed by regression-based models to estimate BP. Representative models include linear regression (LR), multi-instance regression, support vector machines (SVM), ridge regression, random forests (RF), AdaBoost, and artificial neural networks (ANN) [106, 107–108]. While computationally efficient and interpretable, their performance is often limited by feature quality and poor generalizability across subjects and conditions, though they remain useful in data scarcerios.
Deep learning (DL) methods eliminate the need for manual feature engineering by automatically learning spatiotemporal representations from raw physiological signals. CNNs and their variants [109] such as AlexNet [110], MobileNet, EfficientNet, and ResNet capture spatial dependencies and hierarchical features from physiological signals, while Recurrent Neural Networks (RNNs) [111] and their advanced forms including Long Short-Term Memory (LSTM) [112] and Gated Recurrent Units (GRU) [113], are effective in modeling the temporal dependencies of BP-related signals. Hybrid architectures, such as CNN-LSTM [114], CNN-GRU, CNN/RNN-attention [115, 116–117], and Transformer-based models [13, 118], combine the strengths of spatial feature extraction and sequential learning, enhancing model robustness. Temporal Convolutional Networks (TCNs) [119] further improve long-range temporal learning without the need for recurrent connections, making them well-suited for real-time BP monitoring on edge devices.
The emergence of foundation models has introduced a paradigm shift in AI, with growing potential in physiological signal analysis and cuffless BP estimation [12, 120, 121]. These large-scale models leverage massive pretraining data and self-supervised learning to capture generalizable representations across tasks and individuals. PaPaGei [121], the first open-source foundation model for PPG signal analysis, was pre-trained on 57,000 h of data from three public dataset. Its variants, PaPaGei-P and PaPaGei-S, target intra-subject and morphology-based consistency, with PaPaGei-S showing strong performance on cardiovascular tasks including blood pressure, hypertension, and heart rate estimation. SiamQuality [120] uses CNN-based self-supervised architecture to learn quality-invariant representations from over 36 million ICU signal pairs. By pairing high- and low-quality signals, it demonstrates robustness to noise and outperforms conventional baselines on BP estimation and atrial fibrillation detection. Liu et al. [12] explored instruction-tuned Large Language Models (LLMs) for cuffless BP estimation, using ECG and PPG-derived features with domain-specific prompts. Fine-tuned LLMs outperformed task-specific baselines with an estimation error of 0.00 ± 9.25 mmHg for SBP and 1.29 ± 6.37 mmHg for DBP. While foundation models offer notable generalization and adaptability, they still lag behind task-specific models in BP estimation accuracy and impose substantial computational demands.
Table 1 provides a comparative summary of representative studies on BP estimation using PPIN-, ML-, and DL-based methods, detailing input signal types, algorithmic approaches, number of clinically recruited subjects, BP estimation errors, and calibration strategies employed, compensating recent works in standardized benchmarks [122, 123]. These efforts report statistically grounded, cross-dataset results and enable fairer, repeatable comparisons. Nevertheless, the coverage remains incomplete, particularly lacking representativeness among older adults, hypertensive populations, diverse skin tones, and different types of devices. And physics-informed models additionally require standardized evaluations of constraint validity and failure modes. Additional model-level limitations are discussed in Sect. 4.3.
Table 1. Comparison of BP estimation methods using physics or physiology informed network (PPIN), machine learning (ML), and deep learning (DL) models
Input | Method | Datasets (N*) | Evaluation data split* | SBP (mmHg) | DBP (mmHg) | Algorithmic calibration | Test scenario | References | |||
|---|---|---|---|---|---|---|---|---|---|---|---|
MAE | ME ± SD | MAE | ME ± SD | ||||||||
PPIN | MWPTT features | Arteriolar PTT | Private: N = 20 | 1.86 | ± 2.85 | 1.49 | ± 1.75 | Personalized | Lab-stress | [75] | |
ECG, PPG features | Cardiac-vascular coupling | Private: N = 9 | 3.7 | 0.07 ± 4.9 | 4.6 | 1.4 ± 5.6 | Population | Lab-stress, ambulatory | [76] | ||
MIMIC: N = 53 | 6.8 | 0.2 ± 8.8 | 3.8 | − 0.1 ± 5.0 | |||||||
JOCOC-36 days: N = 23 | 6.2 | 0.3 ± 8.0 | 3.7 | 0.2 ± 4.7 | |||||||
Arterial dimensions, artery wall resonances | Resonance sonomanometry | Private: N = 6 | − 22.7 | − 2.1 | Calibration-free | Lab-static | [84] | ||||
PTT, PIR features | PIR | Private: N = 27 | 4.09 | − 0.37 ± 5.21 | 3.18 | − 0.1 ± 4.1 | Personalized | Lab-static | [72] | ||
PTT features | PTT-IPG | Private: N = 15 | 0.31 ± 8.55 | − 0.5 ± 5.1 | Personalized | Lab-static | [87] | ||||
ECG, PPG, IPG, temperature features | Multimodal | Private: N = 23 | 5.78 | − 1.4 ± 8.0 | 4.15 | − 1.0 ± 5.6 | Personalized | Lab-stress | [79] | ||
IPG signal and features | PINN | Graphene-HGCPT: N = 6, calfree-HGCPT: N = 5, Ring-CPT: N = 4 | LOSO* | 1.3 ± 7.6 | 0.6 ± 6.4 | Personalized | Lab-static | [97] | |||
IPG, PPG, millimeter wave signal | PITN | Graphene-HGCPT: N = 6, Ring-CPT: N = 4, Blumio: N = 115 | LOSO* | − 0.05 ± 6.1 | 0.5 ± 5.9 | Personalized | Lab-static /stress | [98] | |||
ML | ECG, PPG, PPW features | AdaBoost, SVM | CAS-BP: N = 3077 | 5-folds CV* | 7.38 | 2.3 ± 9.6 | 5.13 | 1.3 ± 6.4 | Hybrid | Lab-static, ambulatory | [132] |
ECG, PPG | RF | MIMIC-II: N = 1212 | 10-folds CV* | 5.2 | 0.1 ± 6.3 | 3.8 | 0.2 ± 4.5 | Population | Ambulatory | [133] | |
HRV, PPG, PTT features | Adaboost | MIMIC + VitalDB: N = 3337 | 60:20%:20% | 7.73 | − 0.16 ± 7.96 | 4.3 | − 0.13 ± 4.50 | Population | Lab-static | [91] | |
ECG, PPW features | Multi-instance regression | Private: N = 85 | – | 6.13 | 1.62 ± 7.76 | 4.54 | 1.49 ± 5.52 | Personalized | Ambulatory | [107] | |
TAG, PPG, ECG features | Ridge regression | Aurora-BP: N = 1125 | – | 0.32 ± 9.8 | 0.54 ± 7.7 | Population | Ambulatory | [134] | |||
DL | PPG signal | CNN Siamese | MIMIC II: N = 304 | 60%:20%:20% | 5.95 | ± 6.69 | 3.41 | ± 3.91 | Population | Lab-static | [109] |
PPG signal | CNN-GRU | MIMIC-III: N = 50 | 80%:10%:10% | 3.52 | 0.11 ± 4.56 | 2.20 | 0.05 ± 2.82 | Hybrid | Lab-static | [135] | |
ECG signal | ResNet-LSTM | MIMIC-III: N = 1711 | 65%:10%:25% | 7.1 | − 0.11 ± 9.99 | 4.61 | 0.01 ± 6.29 | Population | Lab-static | [114] | |
PPG signal | BiLSTM-attention | MIMIC-III: N = 225 | 60%:20%:20% | 2.82 | − 0.01 ± 4.04 | 1.88 | − 0.13 ± 2.98 | Hybrid | Lab-static | [116] | |
ECG signal | BiLSTM-attention | MIMIC II: 21,442 records | 80%:10%:10% | 7.16 | 0.18 ± 10.8 | 3.89 | 1.24 ± 5.90 | Hybrid | Lab-static | [117] | |
PPG signal | TCN | multi-datasets: N = 134 | 19%:81% | 8.9 | 0.99 ± 7.91 | 5.8 | 0.36 ± 5.43 | Personalized | Lab-static, ambulatory | [119] | |
ECG and PPG features | ANN + RNN | VitalDB: N = 1376 | 70%:30% | 5.07 | 0.05 ± 6.92 | 2.86 | − 0.05 ± 3.99 | Population | Lab-static | [111] | |
Features and ECG, PPG, PPW signals | CNN-Transformer | CAS-BP: N = 1272 | 60%:20%:20% | 6.3 | 0.7 ± 8.3 | 5.1 | 0.9 ± 6.5 | Population | Ambulatory | [118] | |
Aurora-BP: N = 1125 | 6.1 | − 0.4 ± 8.6 | 5.25 | − 0.4 ± 7.0 | Population | ||||||
ECG, PPG features | LLaMA3-8B | CAS-BP: N = 1272 | – | 7.08 | 0.00 ± 9.25 | 5.31 | 1.29 ± 6.37 | Hybrid | Lab-static | [12] | |
*N: subject numbers
*LOSO: leave-one-subject-out
*CV: cross-validation
*Data split: the percentage of participants in train:validation:test/train:test dataset
Challenges and Solutions of BP Deployment in Resource-Limited Settings
The deployment of AI-enhanced BP monitoring in resource-limited settings introduces multifaceted challenges that extend beyond algorithmic accuracy. Limitations in computational hardware, power supply, data availability, and healthcare infrastructure demand a holistic rethinking of system architectures and deployment paradigms. To achieve clinically viable, scalable, and accessible implementation, it is essential to integrate model efficiency with hardware feasibility and real-world usability. This section systematically reviews representative computing architectures, resource-aware optimization strategies and analyzes persistent barriers and emerging solutions shaping the field.
Resource-Aware Computing Architectures
Deployment of AI models under resource constraints requires computing architectures that align model complexity with hardware capabilities [124, 125]. Table 2 summarizes representative academic efforts in deploying AI-based BP estimation, detailing input modalities, optimization techniques, hardware platforms, and performance metrics (e.g., latency, memory, energy, and clinical accuracy). All methods in Table 2 were validated in laboratory scenarios [68], including lab–static (controlled resting), lab–stress (acute perturbations such as cold pressor, posture/handgrip, pharmacologic stimuli). In parallel, Table 3 catalogs commercial cuffless BP systems (e.g., Aktiia, Omron HeartGuide, Valencell) from peer-reviewed publications, FDA filings, and manufacturer whitepapers. This consolidated comparison facilitates cross-method evaluation and informs practical design decisions under real-world constraints.
Table 2. Overview of the representative studies on the deployment of cuffless BP estimation model
Input | Sampling rate and windows | Models | SBP (mmHg) | DBP (mmHg) | Deployment device | Run time | Parameters | Memory | Power consumption | References | ||
|---|---|---|---|---|---|---|---|---|---|---|---|---|
MAE | ME ± SD | MAE | ME ± SD | |||||||||
4 ECG, PPG features | -, 8 s | LR | 16.70 | 7.59 | Arduino uno, ESP32, PyBadge | µs level | < 10 KB | [136] | ||||
SVM | 16.92 | 7.38 | µs level | < 10 KB | ||||||||
DT | 14.49 | 6.98 | ~ 15–40 µs | 10 ~ 50 KB | ||||||||
RF | 14.08 | 6.85 | > 50 µs | > 100 KB | ||||||||
rPPG | 25 Hz, 12 s | ResNet-attention | 9.27 | 5.84 | Intel i7 | 1.616 s | ~ 2.4 M | [137] | ||||
PPG features | 125 Hz, beats | XGBoost | 7.27 | ± 9.5 | 3.33 | ± 4.55 | 24-cores CPU | ~ 2.36 ms | 49.55 KB | [138] | ||
PPG signal | 125 Hz, 2.1 s/5 s | ResNet, UNet | 17.2 | 8.08 | GAP8 | 7.04–8.91 ms | 23.4 ~ 156.3 k | < 512 KB | 0.36 ~ 0.45 mJ | [139] | ||
PPG signal | 125 Hz, beats | ANN | 3.42 | ± 5.42 | 1.92 | ± 3.30 | EFM32 | 42.2 ms | < 25 KB | 2.1 mJ | [140] | |
PPG signal | 125 Hz, 10 s | UNet | 5.16 | 2.89 | Raspberry Pi 4 | 42.53 ms | [141] | |||||
ECG signal | 125 Hz, 4 s | ANN | 5.98 | 0.83 ± 9.13 | 3.53 | 0.1 ± 6.2 | Intel i7 | 0.898 s | 7 KB | [142] | ||
82 PPG features | 125 Hz, 8 s | CNN-BiLSTM | 1.38 | 0.95 | Jetson Nano, AGX Xavier | [143] | ||||||
IPG, BCG signals | 500 Hz, 2.048 s | XGBoost | 8.58 | 2.2 ± 10.9 | 5.27 | 1.9 ± 6.8 | stm32F756ZG | ~ ms | 339.77 ~ 350.3 KB | [144] | ||
ECG | 128 Hz, 0.78 s | NARX-ANN | Jetson Nano | 81.8 ms | 2000 mw | [128] | ||||||
PPG signal | LeNet | 11.27 | 2.5 ± 16.9 | 5.95 | 1.9 ± 9.6 | Raspberry Pi | 8 ms | 0.30 M | 70 KB | [145] | ||
ResNet | 0.6 ± 16.5 | 2.3 ± 8.7 | 186 ms | 0.27 M | ||||||||
125 Hz, 4 s | SqueezeNet | 2.5 ± 17.2 | 2.7 ± 9.0 | 25 ms | 0.80 M | |||||||
AlexNet | − 2.8 ± 14.5 | − 2.7 ± 7.8 | 27 ms | 0.94 M | ||||||||
MobileNet | − 2.2 ± 17.6 | 2.2 ± 9.7 | 12 ms | 0.82 M | ||||||||
PPG features | 125 Hz, 3 s | TCN | 2.38 | 0.07 ± 3.2 | 1.23 | 0.1 ± 1.7 | Raspberry Pi zero w | 2.5 s | 32.22 KB | [146] | ||
PPG signal | 100 Hz, 10 s | SCIGTCN | 8.9 | 0.99 ± 7.9 | 5.8 | 0.4 ± 5.4 | Jetson Nano | 1.92 s | [119] | |||
4 PPG features | 125 Hz, beats | CAE | 2.25 | ± 2.82 | 5.01 | ± 2.1 | Arduino nano 33 BLE | [147] | ||||
PPG signal | 125 Hz, 0.8 s | ANN | 3.85 | ± 4.29 | stm32L4 + | ~ 25 k | [148] | |||||
ECG, PPG features | 125 Hz, − | LSTM | MSE = 0.0151 | ASIC | 0.215 ms | ~ 270 k | 3.27 mW | [149] | ||||
PPG features | −, 0.4 ~ 3 s | DNN | 4.9 | ± 6.2 | 3.4 | ± 4.4 | ZYNQ7020 | 0.3 ~ 4 s | ~ 1 k | 4.4 uW | [150] | |
PPG features | −, beats | ANN | 2.47 | ± 3.48 | 1.45 | ± 1.88 | ASIC | 19.72 uW, 0.2 mJ | [151] | |||
Table 3. Overview of the representative commercial cuffless BP systems
Device | Input | Method | SBP errors* | DBP errors* | Clearance | Calibration | References |
|---|---|---|---|---|---|---|---|
Aktiia | PPG signal | PWA | 1.3 ± 7.11 | − 0.2 ± 5.46 | FDA/CE | Monthly cuff | [152] |
Omron HeartGuide | Oscillometric | Oscillometry | − 0.9 ± 7.6 | − 1.1 ± 6.1 | FDA | No | [153] |
Valencell | PPG, demographics | ML | 0.0 ± 7.9 | 0.4 ± 7.4 | – | No | [154] |
Healthstats | Radial pressure | Tonometry | < 5 ± 8 | < 5 ± 8 | FDA | Weekly cuff | |
Sotera Visi Mobile | ECG, PPG | PAT/PTT | − -1.88 ± 6.17 | − 1.65 ± 3.62 | FDA | Initial cuff BP, recalibrate ≥ 24 h | |
Biobeat BB-613WP | PPG | PWTT | − 0.1 ± 3.6 | 0 ± 3.5 | FDA | Periodic cuff | [155] |
LiveOne | Pressure/tonometry | ML | 0.0 ± 6.9 | 1.2 ± 5.7 | FDA | Demographic | [156] |
NanoWear SimpleSense | PPG, ECG, heart sound, thoracic impedance, activity, demographics | Ensemble ML | − 2.94 ± 4.82 | − 0.77 ± 3.75 | FDA | Initial cuff BP, periodic update | [157, 158] |
PyrAmes/Boppli | Capacitive sensor array | ML | − 0.7 ± 7.7 | 1.4 ± 4.7 | FDA | Demographic | [159] |
Biospectal/OptiBP | camera PPG | ML | 1.5 ± 6.7 | − 0.2 ± 4.1 | CE | Cuff | [160] |
Samsung Galaxywatch | PPG | PWA | − 2.05 ± 15.5 | − 5.58 ± 22.5 | CE | Every 28 days cuff | [161] |
CART-I | PPG | PWA | 1.74 ± 6.69 | − 3.24 ± 6.51 | Korea MFDS | Two-step cuff (+ periodic) | [162] |
*Accuracy metric: ME ± SD (mmHg)
Based on Tables 2 and S2, the deployability can be directly shaped by model-side factors: input configuration (sampling rate, window length, channels), model architecture (depth, width, parameters), target metrics (accuracy, latency, robustness) and optimizations; hardware-side constraints: on-chip SRAM/Flash, compute throughput, and accelerator availability and AFE design.
As illustrated in Fig. 5, such deployments are typically structured into these tiers: on-device, edge, and cloud, facilitating adaptive distribution of computational tasks across hardware platforms with differing resource constraints [17, 126]. Accordingly, low-complexity models (e.g., physiology-informed and traditional ML models) align with an on-device tier where sensing, preprocessing, and inference run locally on MCUs or ultra-low-power SoCs (e.g., Nordic nRF52840, ESP32-S3, and STM32N6) for maximal privacy and minimal latency/energy [124, 127]. Mid-complexity models (e.g., compact deep networks) align with a device–edge tier where sensing and light preprocessing remain on the device, while the main inference runs on nearby NPUs/DSPs (smartphones/gateways), balancing responsiveness, and compute [125]. High complexity (e.g., transformers and foundation models) align with a device–edge–cloud tier that keeps local preprocessing and lightweight inference while offloading heavy inference or training to the cloud, trading connectivity, and latency for scalability [128]. Detailed deployment factors and the model–hardware selection strategy are provided in Note S2.
[See PDF for image]
Fig. 5
Scalable resource-aware computing architectures for wearable BP monitoring. Three computing paradigms aligned with model complexity and system capabilities: on-device, edge, and cloud
Together, these deployment strategies define a scalable design spectrum for BP monitoring from ultra-low-power real-time inference to cloud-based precision analytics. Robust deployment in resource-limited settings requires coordinated optimization across the system stack, including AFE design, digital processing platforms, and AI model architecture [127, 129, 130–131]. Representative processor platforms and associated trade-offs are summarized in Table S2, with detailed implementation examples provided in Note S3.
Optimization and Deployment Strategies for BP Estimation
Building upon the computing architectures described above, effective deployment of AI-driven BP estimation models in resource-limited settings requires a structured optimization pipeline that spans both algorithm-level design and graph-level compilation (Fig. 6). This dual-layer pipeline enables scalable execution across heterogeneous platforms from ultra-low-power wearables to edge and cloud infrastructures.
[See PDF for image]
Fig. 6
End-to-end optimization pipeline for deploying AI-driven BP estimation on resource-limited devices. The process spans algorithm-level design, compression and personalization, followed by graph-level compilation, system integration, and runtime scheduling
Algorithm-Level Optimization
At the algorithm level, model architecture and parameter design must account for memory, energy, and latency constraints while preserving estimation accuracy. Lightweight manually designed models [141, 145, 149] (e.g., MobileNet, SqueezeNet) and hardware-aware Neural Architecture Search (NAS)[139, 163] methods can generate compact models that fit on constrained devices. A typical NAS framework comprises three core components [163]: a search space defining possible architectures and hyperparameters (e.g., layer type, kernel size, depth, width); a search algorithm (e.g., reinforcement learning, evolutionary methods); and a performance estimator that evaluates candidates using full training or proxy methods like weight sharing. The models can be prepared through dynamic graph frameworks (e.g., PyTorch, TensorFlow, PaddlePaddle); they are commonly pretrained on large-scale datasets such as MIMIC-II/III/IV [140, 141–142, 145, 146, 148, 151] and fine-tuned on edge-deployable datasets [119, 136]. Signal-level optimization such as adaptive windowing, downsampling, filtering, and dimensionality reduction (e.g., PCA) further reduce computational load. This stage defines the foundational neural architecture and pretrained weights to improve convergence and generalization. These efforts ensure reliable signal-to-model mappings as a basis for further optimization.
Compression techniques further reduce model size and computational cost. These strategies can be applied synergistically [127, 136] and tailored to target hardware constraints via automated frameworks such as AdaDeep [131].
Pruning reduces model complexity by eliminating redundant weights or structural components (e.g., neurons, channels) [138]. Unstructured pruning (e.g., magnitude-based) yields sparse weight matrices, while structured pruning targets high-level computational units (e.g., neurons, channels, kernels, or entire blocks), yielding a regular architecture optimized for hardware acceleration and parallelism [164]. Energy-aware pruning [165] selectively removes components based on energy cost metrics, enhancing overall power efficiency.
Quantization reduces numerical precision (e.g., FP32 to INT8) for compatibility with integer-only compute engines such as MCUs or NPUs [147, 151]. The widely supported frameworks include TFLite, CMSIS-NN, TensorRT, and QNNPACK.
Knowledge Distillation (KD) transfers knowledge from large “teacher” models to smaller “student” models through various strategies [166]: response-based, where the student mimics the teacher’s soft outputs; feature-based, which transfers intermediate representations (e.g., hidden states); relation-based, capturing inter-sample or inter-layer dependencies.
Low-Rank Factorization approximates large weight matrices with rank-decomposed components, enabling faster inference and model simplification with minimal retraining [167]. It is particularly effective when layers exhibit strong linear dependencies. Common techniques include canonical polyadic decomposition and tucker decomposition [168], where convolutional layers were successfully factorized and fine-tuned to maintain model performance.
Model Personalization enhances real-world robustness by adapting models to individual variability [76, 138]. Personalized adaptation strategies such as transfer learning, feature sharing, and parameter-efficient fine-tuning (e.g., Low-Rank Adaptation (LoRA), QLoRA) were widely used [116, 119, 169]. Continual and incremental learning [149, 170] further support long-term adaptation by updating model parameters in real time, mitigating performance drift without full retraining. In multi-user scenarios, federated learning enables decentralized model updates without raw data exchange, preserving privacy [171]. While not yet widely adopted for BP estimation, split learning offers promising potential for ultra-constrained settings by enabling partial computation offloading to edge servers [172], balancing local adaptation and efficiency. These strategies allow the same model backbone to generalize across users while adapting to temporal or population-specific physiological differences.
Graph-Level Compilation Optimization
Once model architecture and parameters are finalized, the trained models are compiled into static computation graphs (e.g., ONNX, TFLite, PaddlePaddle IR), which are then lowered into hardware-executable code through compilation toolchains (e.g., TVM, CMSIS-NN, or TensorRT) [173] with graph- and tensor-level optimizations [130]. Core graph-level optimizations include graph simplification (e.g., constant folding, operator fusion, layout transformation), tensor scheduling (e.g., tiling, unrolling, vectorization) [174], memory planning for buffer reuse and alignment, and auto-tuning for device-specific kernel scheduling. These optimizations maximize efficiency across diverse hardware in BP estimation field, including MCUs, NPUs, and AI ASICs [140, 144, 150].
Compiled binaries are integrated into firmware stacks alongside drivers and real-time operating systems (RTOS) kernels, enabling efficient on-device inference. At runtime, lightweight execution engines (e.g., TFLite Micro, CMSIS-NN) manage memory and task scheduling under RTOSs or bare-metal conditions [175]. Dynamic scheduling techniques [176] have been introduced to enable dynamic adaptation to context, improving responsiveness and energy efficiency in daily use. For instance, dynamic routing selectively activates sub-networks based on input complexity [126], while early exit [125] architectures terminate inference once confidence thresholds are met. Intermittent scheduling enables periodic or event-driven inference aligned with physiological rhythms[146, 150]. In addition to purely on-device execution, hybrid and offload scheduling schemes [17, 125] can alleviate local computational burdens by distributing inference across edge servers or the cloud. In such designs, some researches [128, 137, 177] conduct lightweight signal preprocessing on-device, while offloaded heavier computation tasks.
Infrastructure, Model, and Deployment Barriers
AI-driven wearable BP monitoring in resource-limited settings faces complex and interdependent barriers spanning infrastructure, model design, practical deployment, and ethical considerations, which limit clinical translation and real-world reliability.
Infrastructure barriers are foundational. Low-cost, energy-constrained wearable devices, and basic smartphones severely limit computational capacity, memory, and data storage [16, 178]. These limitations hinder real-time inference, long-term logging, and scalable data integration. Unreliable power supply and intermittent network connectivity further complicate deployment, making edge-only, or edge-first AI processing essential [179]. Moreover, the lack of standardized data ecosystems and interoperability across heterogeneous devices further hinders scalable and coordinated BP monitoring efforts [14].
Model-level barriers present a critical bottleneck. First, physiological non-specificity remains a foundational barrier: Hemodynamic features extracted from non-invasive signals (e.g., PPG, mechano-electric) are modulated by vasomotor tone, autonomic state, contact pressure, motion, temperature, and device/subject identifiers, leading to feature shifts that are weakly or non-uniquely associated with BP [11]. Second, the burden of calibration and drift is substantial: Many systems require frequent recalibration, and calibration-heavy designs often perform similarly to strong non-physiological baselines (e.g., calibration BP and time). Third, evaluation and implementation flaws, over-optimistic results due to data leakage, hyperparameter tuning on test sets, calibration leakage, and selective metrics. Fourth, subgroup fairness and robustness, small, cohort-biased datasets, and weak generalization protocols limit robustness across real-world conditions such as ambulatory, exercise, and thermal variability [6]. Performance disparities across subgroups (e.g., elderly, hypertensive, different skin tones, or arterial stiffness levels) remain under-explored, posing fairness concerns. Lastly, the absence of prospective, real-time clinical trials prevents regulatory adoption and obscures the actual benefit of cuffless BP systems in long-term management. Early solutions using hardware-aware neural architecture search, model quantization, federated learning, and lightweight personalization show promise [178].
Deployment barriers include energy constraints, maintenance challenges, user trust, and the absence of validation in daily life. In LMICs deployment scenarios, power and hardware constraints must be explicitly aligned with reporting requirements. Clinical applications range from beat-to-beat estimation in critical care, to intermittent 30–60-min monitoring in ambulatory management, to daily or weekly tracking for lifestyle support. These scenarios entail distinct trade-offs in sensing complexity, model latency, and energy consumption. Table S3 contextualizes these trade-offs, linking use cases to deployment constraints and guiding resource-aware system design. Continuous health monitoring imposes substantial energy demands, while reliable charging remains impractical in many resource-limited settings due to unstable electricity infrastructure [14]. Devices must be sturdy, intuitive, and require minimal maintenance, as high device loss and limited local support can critically limit adoption [11]. Acceptance studies have highlighted user concerns around data accuracy, trust, and loss of human interaction, emphasizing the need for transparency and human–AI collaboration [180].
Beyond technical constraints, ethical and regulatory barriers remain. Insufficient transparency, algorithmic bias, data privacy issues, and lack of governance infrastructure continue to limit trust and equitable access in many regions [6, 180]. Integrative deployment strategies combining edge inference with human-in-the-loop oversight and community health can alleviate trust barriers and enhance usability [10].
Despite these challenges, encouraging precedents from adjacent domains demonstrate that meaningful clinical impact is possible through thoughtful system-level design. Wearable systems have shown real-world utility in fetal and maternal monitoring, arrhythmia detection, and hearing screening when paired with frugal hardware and edge-optimized AI [15, 16, 179]. For instance, Ryu et al. developed a wireless network of soft, flexible sensors capable of comprehensive maternal and fetal monitoring including HR, uterine activity, and fetal movements across both high- and low-resource settings [15]. Their system was validated in clinical environments ranging from tertiary hospitals to rural clinics, demonstrating the feasibility of low-cost, wearable-based monitoring even in infrastructure-limited environments. Similarly, Chan et al. presented an off-the-shelf otoacoustic emission probe using low-cost hardware and AI-based signal interpretation to enable newborn hearing screening in LMICs [16]. This study illustrates how combining frugal hardware design with intelligent signal processing can make clinical-quality screening accessible at scale. Recent work demonstrated that AI models optimized for consumer smartwatches can achieve clinical-grade performance in detecting critical physiological events such as loss of pulse, with fully edge-based, energy-efficient inference [179]. These successes may offer valuable design patterns translatable to cuffless BP monitoring. At the infrastructure level, low-cost sensing and edge AI reduce reliance on cloud connectivity and power-hungry processing.
Evaluation and Validation
Model Evaluation Metrics
Accurate evaluation is critical to ensure that AI-based cuffless BP estimation systems are clinically valid and practically deployable [40]. Researchers usually use a variety of metrics to evaluate model performance against reference methods (e.g., intra-arterial or auscultatory cuff-based measurements), including mean error (ME), standard deviation of error (SDE), mean absolute error (MAE), root mean squared error (RMSE), and mean absolute percentage error (MAPE) [181] for waveform-based methods. Tracking metrics such as time-series RMSE or mean tracking error are employed to assess a model’s responsiveness to BP fluctuations in dynamic monitoring scenarios. Correlation coefficient (e.g., Pearson’s r) and Bland–Altman plots evaluate trend alignment and agreement with reference values.
In addition, recent studies have advocated more rigorous and reproducible practices that directly probe physiological discriminability, calibration dependence, and real-world robustness: (1) Leakage-free evaluation, using subject-wise splits to prevent information leakage from the same individual. (2) Baseline comparisons against non-physiological models (e.g., models rely solely on non-measurement features such as calibration BP, population-average BP, demographics, or time of day) [12, 182] to assess physiological signal utility. In the Microsoft Research Aurora Project [67], all models based on waveform features produced errors comparable to those of a baseline model using only calibration BP and time. In parallel, include cuff-anchored baselines (e.g., periodic upper-arm cuff recalibration) to assess any added benefit beyond calibration. (3) Feature attribution (e.g., explainable methods such as SHAP [65]) and ablation studies are recommended to quantify the contribution of input features to the final estimations. (4) BP changes (ΔBP) tracking under induced BP changes (e.g., physical exercise, vasomotor provocation tasks such as cold pressor tests, mental stress tasks, or pharmacological interventions) and over long time to assess dynamic adaptability. (5) Individual-level evaluation, identifying error distributions across subgroups (e.g., elderly, hypertensive, different skin tones/arterial stiffness), addressing fairness and robustness. (6) Lab-to-ambulatory with concurrent logging of motion, skin temperature, contact pressure, posture, and other contextual variables, to assess degradation in non-ideal settings. (7) Reporting and metrics, explicitly document calibration burden and schedules, and release protocols/device settings/code where possible to enable reproducibility.
On-Device Evaluation
In addition to model robustness and fairness evaluations, on-device evaluation focuses on assessing the system-level performance and robustness of BP estimation models when deployed on resource-limited wearable platforms. Key system-level metrics [17, 127, 136, 139, 145] include model size, inference latency (ms), floating point of operations (FLOPs), memory footprint (KB), energy consumption (mJ), and clinical accuracy which are assessed to ensure that the system meets real-world requirements under diverse operating conditions. Comparisons between edge and server inference typically show tolerable accuracy degradation (~ 8%–10%), confirming the feasibility of wearable BP estimation. Beyond these core metrics, robust on-device evaluation should further encompass assessments of system robustness and long-term usability. This includes: (1) Robustness to motion artifacts and environmental noise [7], by testing model performance under controlled motion scenarios (e.g., walking, wrist rotation) and across varying ambient conditions (light, temperature, humidity), using both synthetic and real-world datasets. (2) Battery impact analysis [183], quantifying the additional power consumption introduced by BP estimation tasks, and evaluating its effect on overall device battery life under typical usage patterns. 3) OTA update robustness, validating the integrity and consistency of model performance following over-the-air updates, ensuring clinical reliability is maintained post-update.
Finally, the lack of standardized benchmark datasets and testing protocols for on-device BP estimation poses a challenge to cross-study comparisons. Establishing such benchmarks—including standardized motion protocols, battery stress tests, and runtime performance evaluation guidelines—would greatly enhance the comparability of published results and accelerate progress toward clinically robust wearable BP monitoring systems.
Standard Requirements
To enable clinical translation, cuffless BP estimation methods must align with internationally recognized standards, many of which were initially developed for cuff-based systems but are now widely referenced for wearable and cuffless technologies. These standards define acceptable error thresholds, data distribution requirements, validation protocols, and reference measurement methods. The AAMI/ESH/ISO standard [184] mandates static testing with normal data distribution and fixed thresholds, along with a demographically diverse (e.g., age, gender, arm circumference) subject pool. The British Hypertension Society (BHS) standard [185] provides a grading system (A–D) based on the cumulative percentage of errors within 5, 10, and 15 mmHg (CP5, CP10, CP15), requiring independent accuracy for both SBP and DBP. The IEEE 1708-2014/2019 standards were the first tailored for ‘cuffless wearable BP devices’ [42, 46], which introduce the use of mean absolute difference (MAD) and mean absolute percentage difference (MAPD) as key metrics and emphasized dynamic testing (e.g., postural and motion). The ISO 81060-3:2022 standard, targeting continuous automated sphygmomanometers, adopts a relaxed criterion with simplified requirements [186]. More recently, the European Society of Hypertension (ESH) [103] released an application-driven protocol specifically for cuffless BP systems, incorporating six evaluation scenarios and advanced statistical modeling (e.g., mixture of multivariate normal distributions). These standards provide essential benchmarks for assessing model performance, guiding validation efforts, and determining clinical acceptability. The specific metrics and performance thresholds summarized in Table 4 and the grading framework [92] illustrated in Fig. 7 offer a comprehensive overview of the prevailing regulatory landscape.
Table 4. Summary of BP validation standards
Standard | Sample size | Reference method | Acceptance criteria |
|---|---|---|---|
AAMI/ESH/ISO | ≥ 85 | Sphygmomanometer/ invasive arterial line | ME ≤ ± 5 mmHg, SD ≤ 8 mmHg |
BHS | 85 | Sphygmomanometer | Grade A: CP5 ≥ 60%, CP10 ≥ 85%, CP15 ≥ 95% Grade B: CP5 ≥ 50%, CP10 ≥ 75%, CP15 ≥ 90% Grade C: CP5 ≥ 40%, CP10 ≥ 65%, CP15 ≥ 85% Grade D: Worth than Grade C |
IEEE 1708-2014/2019 | ≥ 85 | Sphygmomanometer/ invasive arterial line | Phase 1: MAD ≤ 7 mmHg Phase 2: Grade A: MAD ≤ 5 mmHg Grade B: 5 < MAD ≤ 6 mmHg Grade C: 6 < MAD ≤ 7 mmHg Grade D: MAD > 7 mmHg |
ESH 2023 | 85–175 | Auscultatory/24-h oscillometric | ME ≤ ± 5 mmHg, SD ≤ 8 mmHg |
ISO | 85 | Invasive arterial line | |ME|≤ 6 mmHg, SD ≤ 10 mmHg |
[See PDF for image]
Fig. 7
Accuracy grading criteria for different standards. Adapted with permission from [92], Copyright IEEE, 2025. The x-axis represents the standard deviation of error, and the y-axis denotes the mean error. The shaded regions indicate the accuracy grading criteria defined by the IEEE 1708-2014/2019 standard. The solid line corresponds to the clinical thresholds specified by the AAMI/ESH/ISO standard, while the dashed lines represent the grading levels defined by the BHS protocol
Summary
The integration of non-invasive sensing, artificial intelligence, and resource-efficient deployment offers transformative potential for cardiovascular health monitoring, particularly in blood pressure management in low- and middle-income countries. Wearable AI-driven BP monitoring systems show promise for hypertension screening and long-term care, though current cuffless wearables are best suited as trend-tracking companions. Key challenges include developing high-fidelity, low-cost, and energy-efficient sensors, lightweight, low-power models that can accurately distinguish BP changes from confounding factors, and addressing the issue of frequent calibration. Furthermore, leveraging regional resources to enable scalable deployment of BP models is crucial. Data scarcity, lack of standardized datasets, and the need for robust evaluation batteries also impede progress. Clinical deployment faces challenges related to regulation, ethics, and trust, necessitating standardized validation frameworks and independent assessments. Future solutions should be designed for global impact, prioritizing frugality, resilience, interpretability, and clinical meaningfulness to democratize cardiovascular diagnostics and support accessible care.
Acknowledgements
The authors would like to thank the Chinese University of Hong Kong for providing research resources and institutional support.
Author Contributions
Yiming Zhang and Shirong Qiu helped in conceptualization, investigation, analysis, figure design, original draft writing and editing. Kai Du, Shun Wu, Ting Xiang contributed to analysis, original draft writing and editing. Kenghao Zheng helped in figure design, original draft writing. Zijun Liu, Hanjie Chen, Nan Ji contributed to original draft writing. Fa Wang reviewed the manuscript. Yuanting Zhang helped in review, supervision.
Declarations
Conflict of Interest
The authors declare no interest conflict. They have no known competing financial interests or personal relationships that could have appeared to influence the work reported in this paper.
Publisher's Note
Springer Nature remains neutral with regard to jurisdictional claims in published maps and institutional affiliations.
References
1. Haldar, RN. Global brief on hypertension: silent killer, global public health crisis. Indian J. Phys. Med. Rehabil.; 2013; 24,
2. Ohkubo, T; Imai, Y; Tsuji, I; Nagai, K; Kato, J; Kikuchi, N; Nishiyama, A; Aihara, A; Sekino, M; Kikuya, M; Ito, S; Satoh, H; Hisamichi, S. Home blood pressure measurement has a stronger predictive power for mortality than does screening blood pressure measurement: a population-based observation in Ohasama, Japan. J. Hypertens.; 1998; 16,
3. Levy, D. The progression from hypertension to congestive heart failure. JAMA; 1996; 275,
4. Ding, X-R; Zhao, N; Yang, G-Z; Pettigrew, RI; Lo, B et al. Continuous blood pressure measurement from invasive to unobtrusive: celebration of 200th birth anniversary of Carl Ludwig. IEEE J. Biomed. Health Inform.; 2016; 20,
5. Chan, G; Cooper, R; Hosanee, M; Welykholowa, K; Kyriacou, PA; Zheng, D; Allen, J; Abbott, D; Lovell, NH; Fletcher, R; Elgendi, M. Multi-site photoplethysmography technology for blood pressure assessment: challenges and recommendations. J. Clin. Med.; 2019; 8,
6. Ciecierski-Holmes, T; Singh, R; Axt, M; Brenner, S; Barteit, S. Artificial intelligence for strengthening healthcare systems in low- and middle-income countries: a systematic scoping review. NPJ Digit. Med.; 2022; 5,
7. Min, S; An, J; Lee, JH; Kim, JH; Joe, DJ et al. Wearable blood pressure sensors for cardiovascular monitoring and machine learning algorithms for blood pressure estimation. Nat. Rev. Cardiol.; 2025; 22,
8. Li, J; Chu, H; Chen, Z; Yiu, CK; Qu, Q et al. Recent advances in materials, devices and algorithms toward wearable continuous blood pressure monitoring. ACS Nano; 2024; 18,
9. Zhao, L; Liang, C; Huang, Y; Zhou, G; Xiao, Y et al. Emerging sensing and modeling technologies for wearable and cuffless blood pressure monitoring. NPJ Digit. Med.; 2023; 6,
10. R.C. Zhao, X. Yuan, AI in healthcare for resource limited settings: an exploration and ethical evaluation. in Companion Proceedings of the ACM on Web Conference 2025. (ACM, Sydney, 2025), pp. 1953–1960. https://doi.org/10.1145/3701716.3717747
11. Dangi, RR; Sharma, A; Vageriya, V. Transforming healthcare in low-resource settings with artificial intelligence: recent developments and outcomes. Public Health Nurs.; 2025; 42,
12. Z. Liu, C. Chen, J. Cao, M. Pan, J. Liu et al., Large language models for cuffless blood pressure measurement from wearable biosignals. in Proceedings of the 15th ACM International Conference on Bioinformatics, Computational Biology and Health Informatics. (ACM, Shenzhen, 2024), pp. 1–11. https://doi.org/10.1145/3698587.3701447
13. Ma, C; Zhang, P; Song, F; Sun, Y; Fan, G et al. KD-informer: a cuff-less continuous blood pressure waveform estimation approach based on single photoplethysmography. IEEE J. Biomed. Health Inform.; 2023; 27,
14. L. Hamamoto, O. Meyer, P. Natarajan, M. Young, A. John et al., Reimagining wearables to bolster sustainable development in low-resource settings. in 2024 IEEE Global Humanitarian Technology Conference (GHTC). (IEEE, 2024), pp. 9–16. https://doi.org/10.1109/GHTC62424.2024.10771560
15. Ryu, D; Kim, DH; Price, JT; Lee, JY; Chung, HU et al. Comprehensive pregnancy monitoring with a network of wireless, soft, and flexible sensors in high- and low-resource health settings. Proc. Natl. Acad. Sci. U. S. A.; 2021; 118,
16. Chan, J; Ali, N; Najafi, A; Meehan, A; Mancl, LR et al. An off-the-shelf otoacoustic-emission probe for hearing screening via a smartphone. Nat. Biomed. Eng.; 2022; 6,
17. Zhou, Z; Chen, X; Li, E; Zeng, L; Luo, K et al. Edge intelligence: paving the last mile of artificial intelligence with edge computing. Proc. IEEE; 2019; 107,
18. Schutte, AE; Kollias, A; Stergiou, GS. Blood pressure and its variability: classic and novel measurement techniques. Nat. Rev. Cardiol.; 2022; 19,
19. Allen, J. Photoplethysmography and its application in clinical physiological measurement. Physiol. Meas.; 2007; 28,
20. J. Homsy, P.J. Podrid, P.J. Podrid, P.J. Podrid, Electrocardiography. in MGH Cardiology Board Review, Springer London (2013), pp, 580–622. https://doi.org/10.1007/978-1-4471-4483-0_36
21. Ao, Y; Jin, L; Wang, S; Lan, B; Tian, G et al. Dual structure reinforces interfacial polarized MXene/PVDF-TrFE piezoelectric nanocomposite for pressure monitoring. Nano-Micro Lett.; 2025; 17,
22. Lin, W; Jia, S; Chen, Y; Shi, H; Zhao, J; Li, Z; Wu, Y; Jiang, H; Zhang, Qi; Wang, W; Feng, C; Xia, S. Korotkoff sounds dynamically reflect changes in cardiac function based on deep learning methods. Front. Cardiovasc. Med.; 2022; 9, [DOI: https://dx.doi.org/10.3389/fcvm.2022.940615] 940615.
23. Geng, F; Bai, Z; Zhang, H; Yao, Y; Liu, C; Wang, P; Chen, X; Du, L; Li, X; Han, B; Fang, Z. Contactless and continuous blood pressure measurement according to caPTT obtained from millimeter wave radar. Measurement; 2023; 218, [DOI: https://dx.doi.org/10.1016/j.measurement.2023.113151] 113151.
24. X. Ding, W. Dai, N. Luo, J. Liu, N. Zhao et al., A flexible tonoarteriography-based body sensor network for cuffless measurement of arterial blood pressure. in 2015 IEEE 12th International Conference on Wearable and Implantable Body Sensor Networks (BSN). (IEEE, 2015), pp. 1–4. https://doi.org/10.1109/BSN.2015.7299405
25. Bashkatov, AN; Genina, EA; Kochubey, VI; Tuchin, VV. Optical properties of human skin, subcutaneous and mucous tissues in the wavelength range from 400 to 2000 nm. J. Phys. D Appl. Phys.; 2005; 38,
26. Lochner, CM; Khan, Y; Pierre, A; Arias, AC. All-organic optoelectronic sensor for pulse oximetry. Nat. Commun.; 2014; 5, 5745. [DOI: https://dx.doi.org/10.1038/ncomms6745]
27. Yokota, T; Zalar, P; Kaltenbrunner, M; Jinno, H; Matsuhisa, N; Kitanosako, H; Tachibana, Y; Yukita, W; Koizumi, M; Someya, T. Ultraflexible organic photonic skin. Sci. Adv.; 2016; 2,
28. Xu, H; Liu, J; Zhang, J; Zhou, G; Luo, N et al. Flexible organic/inorganic hybrid near-infrared photoplethysmogram sensor for cardiovascular monitoring. Adv. Mater.; 2017; 29,
29. Zhao, Y; Sun, Y; Pei, C; Yin, X; Li, X; Hao, Yi; Zhang, M; Yuan, M; Zhou, J; Chen, Yu; Song, Y. Low-temperature fabrication of stable black-phase CsPbI(3) perovskite flexible photodetectors toward wearable health monitoring. Nano-Micro Lett.; 2024; 17,
30. Jeong, IC; Yoon, H; Kang, H; Yeom, H. Effects of skin surface temperature on photoplethysmograph. J. Healthcare Eng.; 2014; 5,
31. Fine, J; Branan, KL; Rodriguez, AJ; Boonya-Ananta, T; Ajmal, S et al. Sources of inaccuracy in photoplethysmography for continuous cardiovascular monitoring. Biosensors; 2021; 11,
32. Kireev, D; Sel, K; Ibrahim, B; Kumar, N; Akbari, A et al. Continuous cuffless monitoring of arterial blood pressure via graphene bioimpedance tattoos. Nat. Nanotechnol.; 2022; 17,
33. H.P. Schwan, Electrical properties of tissue and cell suspensions. in Advances in Biological and Medical Physics. (Elsevier, 1957), pp. 147–209. https://doi.org/10.1016/b978-1-4832-3111-2.50008-0
34. Zheng, K; Zheng, C; Zhu, L; Yang, B; Jin, X et al. Machine learning enabled reusable adhesion, entangled network-based hydrogel for long-term, high-fidelity EEG recording and attention assessment. Nano-Micro Lett.; 2025; 17,
35. Lim, C; Hong, YJ; Jung, J; Shin, Y; Sunwoo, S-H et al. Tissue-like skin-device interface for wearable bioelectronics by using ultrasoft, mass-permeable, and low-impedance hydrogels. Sci. Adv.; 2021; 7,
36. Gao, W; Ota, H; Kiriya, D; Takei, K; Javey, A. Flexible electronics toward wearable sensing. Acc. Chem. Res.; 2019; 52,
37. Chen, X; Gao, X; Nomoto, A; Shi, K; Lin, M et al. Fabric-substrated capacitive biopotential sensors enhanced by dielectric nanoparticles. Nano Res.; 2021; 14,
38. Gill, RW. Measurement of blood flow by ultrasound: accuracy and sources of error. Ultrasound Med. Biol.; 1985; 11,
39. Wang, C; Li, X; Hu, H; Zhang, L; Huang, Z et al. Monitoring of the central blood pressure waveform via a conformal ultrasonic device. Nat. Biomed. Eng.; 2018; 2,
40. Tamura, T; Maeda, Y; Sekine, M; Yoshida, M. Wearable photoplethysmographic sensors: past and present. Electronics; 2014; 3,
41. Lee, G-H; Moon, H; Kim, H; Lee, GH; Kwon, W et al. Multifunctional materials for implantable and wearable photonic healthcare devices. Nat. Rev. Mater.; 2020; 5,
42. Zang, Y; Zhang, F; Di, C-A; Zhu, D. Advances of flexible pressure sensors toward artificial intelligence and health care applications. Mater. Horiz.; 2015; 2,
43. Pang, Y; Tian, H; Tao, L; Li, Y; Wang, X et al. Flexible, highly sensitive, and wearable pressure and strain sensors with graphene porous network structure. ACS Appl. Mater. Interfaces; 2016; 8,
44. Yu, G; Hu, J; Tan, J; Gao, Y; Lu, Y; Xuan, F. A wearable pressure sensor based on ultra-violet/ozone microstructured carbon nanotube/polydimethylsiloxane arrays for electronic skins. Nanotechnology; 2018; 29,
45. Choong, C-L; Shim, M-B; Lee, B-S; Jeon, S; Ko, D-S; Choong, C-L; Shim, M-B; Lee, B-S; Ko, D-S; Kang, T-H; Bae, J; Lee, SH; Byun, K-E; Im, J; Jeong, YJ; Park, CE; Park, J-J; Chung, U-I. Highly stretchable resistive pressure sensors using a conductive elastomeric composite on a micropyramid array. Adv. Mater.; 2014; 26,
46. Schwartz, G; Tee, BC; Mei, J; Appleton, AL; Kim, DH et al. Flexible polymer transistors with high pressure sensitivity for application in electronic skin and health monitoring. Nat. Commun.; 2013; 4, 1859. [DOI: https://dx.doi.org/10.1038/ncomms2832]
47. Tang, X; Wu, C; Gan, L; Zhang, T; Zhou, T et al. Multilevel microstructured flexible pressure sensors with ultrahigh sensitivity and ultrawide pressure range for versatile electronic skins. Small; 2019; 15,
48. Pang, C; Koo, JH; Nguyen, A; Caves, JM; Kim, M-G; Chortos, A; Kim, K; Wang, PJ; Tok, JB; Bao, Z. Highly skin-conformal microhairy sensor for pulse signal amplification. Adv. Mater.; 2015; 27,
49. Wang, H; Li, Z; Liu, Z; Fu, J; Shan, T et al. Flexible capacitive pressure sensors for wearable electronics. J. Mater. Chem. C; 2022; 10,
50. Wang, Z; Wang, S; Lan, B; Sun, Y; Huang, L et al. Piezotronic sensor for bimodal monitoring of Achilles tendon behavior. Nano-Micro Lett.; 2025; 17,
51. Nie, Z; Kwak, JW; Han, M; Rogers, JA. Mechanically active materials and devices for bio-interfaced pressure sensors-a review. Adv. Mater.; 2024; 36,
52. Chang, J; Li, J; Ye, J; Zhang, B; Chen, J et al. AI-enabled piezoelectric wearable for joint torque monitoring. Nano-Micro Lett.; 2025; 17,
53. Kim, Y; Lee, J; Hong, H; Park, S; Ryu, W. Self-powered wearable micropyramid piezoelectric film sensor for real-time monitoring of blood pressure. Adv. Eng. Mater.; 2023; 25,
54. Petritz, A; Karner-Petritz, E; Uemura, T; Schäffner, P; Araki, T et al. Imperceptible energy harvesting device and biomedical sensor based on ultraflexible ferroelectric transducers and organic diodes. Nat. Commun.; 2021; 12,
55. Yin, H; Li, Y; Tian, Z; Li, Q; Jiang, C; Liang, E; Guo, Y. Ultra-high sensitivity anisotropic piezoelectric sensors for structural health monitoring and robotic perception. Nano-Micro Lett.; 2024; 17,
56. Yi, Z; Liu, Z; Li, W; Ruan, T; Chen, X; Liu, J; Yang, B; Zhang, W. Piezoelectric dynamics of arterial pulse for wearable continuous blood pressure monitoring. Adv. Mater.; 2022; 34,
57. Lee, B-Y; Kim, S-U; Kang, S; Lee, S-D. Transparent and flexible high power triboelectric nanogenerator with metallic nanowire-embedded tribonegative conducting polymer. Nano Energy; 2018; 53, pp. 152-159. [DOI: https://dx.doi.org/10.1016/j.nanoen.2018.08.048]
58. Kim, D; Tcho, I-W; Jin, IK; Park, S-J; Jeon, S-B et al. Direct-laser-patterned friction layer for the output enhancement of a triboelectric nanogenerator. Nano Energy; 2017; 35, pp. 379-386. [DOI: https://dx.doi.org/10.1016/j.nanoen.2017.04.013]
59. Xu, Z; Zhang, C; Wang, F; Yu, J; Yang, G et al. Smart textiles for personalized sports and healthcare. Nano-Micro Lett.; 2025; 17,
60. Dong, K; Wu, Z; Deng, J; Wang, AC; Zou, H; Chen, C; Hu, D; Gu, B; Sun, B; Wang, ZL. A stretchable yarn embedded triboelectric nanogenerator as electronic skin for biomechanical energy harvesting and multifunctional pressure sensing. Adv. Mater.; 2018; 30,
61. Lei, H; Ji, H; Liu, X; Lu, B; Xie, L et al. Self-assembled porous-reinforcement microstructure-based flexible triboelectric patch for remote healthcare. Nano-Micro Lett.; 2023; 15,
62. Li, J; Jia, H; Zhou, J; Huang, X; Xu, L et al. Thin, soft, wearable system for continuous wireless monitoring of artery blood pressure. Nat. Commun.; 2023; 14,
63. Kong, L; Li, W; Zhang, T; Ma, H; Cao, Y; Wang, K; Zhou, Y; Shamim, A; Zheng, Lu; Wang, X; Huang, W. Wireless technologies in flexible and wearable sensing: from materials design, system integration to applications. Adv. Mater.; 2024; 36,
64. Y. Ran, D. Zhang, J. Chen, Y. Hu, Y. Chen, Contactless blood pressure monitoring with mmWave radar. in GLOBECOM 2022—2022 IEEE Global Communications Conference. (IEEE, 2023), pp. 541–546. https://doi.org/10.1109/GLOBECOM48099.2022.10001592
65. Franklin, D; Tzavelis, A; Lee, JY; Chung, HU; Trueb, J et al. Synchronized wearables for the detection of haemodynamic states via electrocardiography and multispectral photoplethysmography. Nat. Biomed. Eng.; 2023; 7,
66. Tan, I; Gnanenthiran, SR; Chan, J; Kyriakoulis, KG; Schlaich, MP; Rodgers, A; Stergiou, GS; Schutte, AE. Evaluation of the ability of a commercially available cuffless wearable device to track blood pressure changes. J. Hypertens.; 2023; 41,
67. Mukkamala, R; Shroff, SG; Landry, C; Kyriakoulis, KG; Avolio, AP et al. The microsoft research aurora project: important findings on cuffless blood pressure measurement. Hypertension; 2023; 80,
68. Mukkamala, R; Shroff, SG; Kyriakoulis, KG; Avolio, AP; Stergiou, GS. Cuffless blood pressure measurement: where do we actually stand?. Hypertension; 2025; 82,
69. Mukkamala, R; Hahn, J-O; Inan, OT; Mestha, LK; Kim, C-S et al. Toward ubiquitous blood pressure monitoring via pulse transit time: theory and practice. IEEE Trans. Biomed. Eng.; 2015; 62,
70. Chen, W; Kobayashi, T; Ichikawa, S; Takeuchi, Y; Togawa, T. Continuous estimation of systolic blood pressure using the pulse arrival time and intermittent calibration. Med. Biol. Eng. Comput.; 2000; 38,
71. C.C.Y. Poon, Y.T. Zhang, Cuff-less and noninvasive measurements of arterial blood pressure by pulse transit time. in 2005 IEEE Engineering in Medicine and Biology 27th Annual Conference. (IEEE, 2006), pp. 5877–5880.
72. Ding, X-R; Zhang, Y-T; Liu, J; Dai, W-X; Tsang, HK. Continuous cuffless blood pressure estimation using pulse transit time and photoplethysmogram intensity ratio. IEEE Trans. Biomed. Eng.; 2016; 63,
73. Pilz, N; Picone, DS; Patzak, A; Opatz, OS; Lindner, T; Fesseler, L; Heinz, V; Bothe, TL. Cuff-based blood pressure measurement: challenges and solutions. Blood Press.; 2024; 33,
74. P. Salvi, Pulse waves: How vascular hemodynamics affects blood pressure 2012. Epub ahead of print. (2012). https://doi.org/10.1007/978-88-470-2439-7
75. Liu, J; Yan, BP; Zhang, Y-T; Ding, X-R; Su, P et al. Multi-wavelength photoplethysmography enabling continuous blood pressure measurement with compact wearable electronics. IEEE Trans. Biomed. Eng.; 2019; 66,
76. Qiu, S; Zhang, Y-T; Lau, S-K; Zhao, N. Scenario adaptive cuffless blood pressure estimation by integrating cardiovascular coupling effects. IEEE J. Biomed. Health Inform.; 2023; 27,
77. Qiu, S; Yan, BPY; Zhao, N. Stroke-volume-allocation model enabling wearable sensors for vascular age and cardiovascular disease assessment. NPJ Flex. Electron.; 2024; 8, [DOI: https://dx.doi.org/10.1038/s41528-024-00307-1] 24.
78. Forrester, SJ; Booz, GW; Sigmund, CD; Coffman, TM; Kawai, T; Rizzo, V; Scalia, R; Eguchi, S. Angiotensin II signal transduction: an update on mechanisms of physiology and pathophysiology. Physiol. Rev.; 2018; 98,
79. T. Xiang, Y. Zhang, Y. Zhang, A novel multimodal physiological model for the noninvasive and continuous measurements of arterial blood pressure. in 2024 46th Annual International Conference of the IEEE Engineering in Medicine and Biology Society (EMBC). (IEEE, 2024), pp. 1–6. https://doi.org/10.1109/EMBC53108.2024.10782499
80. Song, SH; Kim, JH; Lee, JH; Yun, Y-M; Choi, D-H et al. Elevated blood viscosity is associated with cerebral small vessel disease in patients with acute ischemic stroke. BMC Neurol.; 2017; 17,
81. Shi, W; Zhou, C; Zhang, Y; Li, K; Ren, X et al. Hybrid modeling on reconstitution of continuous arterial blood pressure using finger photoplethysmography. Biomed. Signal Process. Control; 2023; 85, [DOI: https://dx.doi.org/10.1016/j.bspc.2023.104972] 104972.
82. J. Joseph, P.M. Nabeel, M.I. Shah, M. Sivaprakasam, Arterial compliance probe for calibration free pulse pressure measurement. in 2016 IEEE International Symposium on Medical Measurements and Applications (MeMeA). (IEEE, 2016), pp. 1–6. https://doi.org/10.1109/MeMeA.2016.7533810
83. Ma, Y; Choi, J; Hourlier-Fargette, A; Xue, Y; Chung, HU et al. Relation between blood pressure and pulse wave velocity for human arteries. Proc. Natl. Acad. Sci. U.S.A.; 2018; 115,
84. Jimenez, R; Yurk, D; Dell, S; Rutledge, AC; Fu, MK et al. Resonance sonomanometry for noninvasive, continuous monitoring of blood pressure. PNAS Nexus; 2024; 3,
85. Fung, Y-c. Biomechanics: Circulation; 2013; Springer:
86. Solà, J; Proença, M; Ferrario, D; Porchet, J-A; Falhi, A et al. Noninvasive and nonocclusive blood pressure estimation via a chest sensor. IEEE Trans. Biomed. Eng.; 2013; 60,
87. Huynh, TH; Jafari, R; Chung, W-Y. Noninvasive cuffless blood pressure estimation using pulse transit time and impedance plethysmography. IEEE Trans. Biomed. Eng.; 2019; 66,
88. H. Truong, A. Montanari, F. Kawsar, Non-invasive blood pressure monitoring with multi-modal in-ear sensing. in ICASSP 2022—2022 IEEE International Conference on Acoustics, Speech and Signal Processing (ICASSP). (IEEE, 2022), pp. 6–10. https://doi.org/10.1109/ICASSP43922.2022.9747661
89. J. Liu, Y.-T. Zhang, X.-R. Ding, W.-X. Dai, N. Zhao, A preliminary study on multi-wavelength PPG based pulse transit time detection for cuffless blood pressure measurement. in 2016 38th Annual International Conference of the IEEE Engineering in Medicine and Biology Society (EMBC). (IEEE, 2016), pp. 615–618. https://doi.org/10.1109/EMBC.2016.7590777
90. Chen, Y; Shi, S; Liu, Y-K; Huang, S-L; Ma, T. Cuffless blood-pressure estimation method using a heart-rate variability-derived parameter. Physiol. Meas.; 2018; 39,
91. Zhang, Y; Zhou, C; Huang, Z; Ye, X. Study of cuffless blood pressure estimation method based on multiple physiological parameters. Physiol. Meas.; 2021; 42,
92. Xiang, T; Jin, Y; Liu, Z; Clifton, L; Clifton, DA et al. Dynamic beat-to-beat measurements of blood pressure using multimodal physiological signals and a hybrid CNN-LSTM model. IEEE J. Biomed. Health Inform.; 2025; 29,
93. J. Penaz, Photoelectric measurement of blood pressure, volume and flow in the finger. in Digest of the 10th International Conference on Medical and Biological Engineering-Dresden, 1973, vol. 104 (1973)
94. Fortin, J; Rogge, DE; Fellner, C; Flotzinger, D; Grond, J et al. A novel art of continuous noninvasive blood pressure measurement. Nat. Commun.; 2021; 12, 1387. [DOI: https://dx.doi.org/10.1038/s41467-021-21271-8]
95. Cai, S; Mao, Z; Wang, Z; Yin, M; Karniadakis, GE. Physics-informed neural networks (PINNs) for fluid mechanics: a review. Acta Mech. Sin.; 2021; 37,
96. Zhang, L; Wang, G; Giannakis, GB. Real-time power system state estimation and forecasting via deep unrolled neural networks. IEEE Trans. Signal Process.; 2019; 67,
97. Sel, K; Mohammadi, A; Pettigrew, RI; Jafari, R. Physics-informed neural networks for modeling physiological time series: a case study with continuous blood pressure. Res. Sq.; 2023; 3, 2423200. [DOI: https://dx.doi.org/10.21203/rs.3.rs-2423200/v1]
98. R. Wang, M. Qi, Y. Shao, A. Zhou, H. Ma, Adversarial contrastive learning based physics-informed temporal networks for cuffless blood pressure estimation (2024). arXiv preprint arXiv:240808488
99. L. Li, X.-C. Tai, R. Chan, BP-DeepONet: a new method for cuffless blood pressure estimation using the physcis-informed DeepONet (2024). arXiv E Prints, arXiv: 2402.18886. https://doi.org/10.48550/arXiv.2402.18886
100. Sun, H; Ma, J; Li, B; Liu, Y; Liu, J et al. Estimation of central aortic pressure waveforms by combination of a meta-learning neural network and a physics-driven method. Int. J. Numer. Meth. Biomed. Eng.; 2025; 41,
101. Chen, Z; Liu, Y; Sun, H. Physics-informed learning of governing equations from scarce data. Nat. Commun.; 2021; 12,
102. Ting, X. A Novel Multimodal Physiological-Informed Model (mpim) for the Unobtrusive Estimation of Dynamic Beat-to-Beat Arterial Blood Pressure; 2025; City University of Hong Kong:
103. Stergiou, GS; Avolio, AP; Palatini, P; Kyriakoulis, KG; Schutte, AE et al. European Society of Hypertension recommendations for the validation of cuffless blood pressure measuring devices: European Society of Hypertension Working Group on Blood Pressure Monitoring and Cardiovascular Variability. J. Hypertens.; 2023; 41,
104. Chowdhury, MH; Shuzan, MNI; Chowdhury, MEH; Mahbub, ZB; Uddin, MM et al. Estimating blood pressure from the photoplethysmogram signal and demographic features using machine learning techniques. Sensors; 2020; 20,
105. Elgendi, M. On the analysis of fingertip photoplethysmogram signals. Curr. Cardiol. Rev.; 2012; 8,
106. Haddad, S; Boukhayma, A; Caizzone, A. Continuous PPG-based blood pressure monitoring using multi-linear regression. IEEE J. Biomed. Health Inform.; 2022; 26,
107. Miao, F; Liu, Z-D; Liu, J-K; Wen, B; He, Q-Y et al. Multi-sensor fusion approach for cuff-less blood pressure measurement. IEEE J. Biomed. Health Inform.; 2020; 24,
108. Hasanzadeh, N; Ahmadi, MM; Mohammadzade, H. Blood pressure estimation using photoplethysmogram signal and its morphological features. IEEE Sens. J.; 2020; 20,
109. O. Schlesinger, N. Vigderhouse, D. Eytan, Y. Moshe, Blood pressure estimation from PPG signals using convolutional neural networks and Siamese network. in ICASSP 2020 - 2020 IEEE International Conference on Acoustics, Speech and Signal Processing (ICASSP). May 4–8, 2020. (IEEE, Barcelona, 2020), pp. 1135–1139. https://doi.org/10.1109/icassp40776.2020.9053446
110. Wang, W; Mohseni, P; Kilgore, KL; Najafizadeh, L. Cuff-less blood pressure estimation from photoplethysmography via visibility graph and transfer learning. IEEE J. Biomed. Health Inform.; 2022; 26,
111. Yang, S; Sohn, J; Lee, S; Lee, J; Kim, HC. Estimation and validation of arterial blood pressure using photoplethysmogram morphology features in conjunction with pulse arrival time in large open databases. IEEE J. Biomed. Health Inform.; 2021; 25,
112. P. Su, X.-R. Ding, Y.-T. Zhang, J. Liu, F. Miao et al., Long-term blood pressure prediction with deep recurrent neural networks. in 2018 IEEE EMBS International Conference on Biomedical & Health Informatics (BHI). (IEEE, 2018), pp. 323–328. https://doi.org/10.1109/BHI.2018.8333434
113. Liu, Z; Zhang, Y; Zhou, C. BiGRU-attention for Continuous blood pressure trends estimation through single channel PPG. Comput. Biol. Med.; 2024; 168, [DOI: https://dx.doi.org/10.1016/j.compbiomed.2023.107795] 107795.
114. Miao, F; Wen, B; Hu, Z; Fortino, G; Wang, X-P; Liu, Z-D; Tang, M; Li, Ye. Continuous blood pressure measurement from one-channel electrocardiogram signal using deep-learning techniques. Artif. Intell. Med.; 2020; 108, [DOI: https://dx.doi.org/10.1016/j.artmed.2020.101919] 101919.
115. Kim, D-K; Kim, Y-T; Kim, H; Kim, D-J. DeepCNAP: a deep learning approach for continuous noninvasive arterial blood pressure monitoring using photoplethysmography. IEEE J. Biomed. Health Inform.; 2022; 26,
116. Zhang, Y; Ren, X; Liang, X; Ye, X; Zhou, C. A refined blood pressure estimation model based on single channel photoplethysmography. IEEE J. Biomed. Health Inform.; 2022; 26,
117. Fan, X; Wang, H; Xu, F; Zhao, Y; Tsui, K-L. Homecare-oriented intelligent long-term monitoring of blood pressure using electrocardiogram signals. IEEE Trans. Ind. Inform.; 2020; 16,
118. Liu, Z-D; Li, Y; Zhang, Y-T; Zeng, J; Chen, Z-X et al. HGCTNet: handcrafted feature-guided CNN and transformer network for wearable cuffless blood pressure measurement. IEEE J. Biomed. Health Inform.; 2024; 28,
119. Zhang, Y; Zhou, C; Ren, X; Wang, Q; Wang, H et al. Personalized continuous blood pressure tracking through single channel PPG in wearable scenarios. IEEE J. Biomed. Health Inform.; 2025; 29,
120. Ding, C; Guo, Z; Chen, Z; Lee, RJ; Rudin, C et al. SiamQuality: a ConvNet-based foundation model for photoplethysmography signals. Physiol. Meas.; 2024; 45,
121. A. Pillai, D. Spathis, F. Kawsar, M. Malekzadeh, PaPaGei: open foundation models for optical physiological signals (2024). arXiv: 2410.20542. https://doi.org/10.48550/arXiv.2410.20542
122. Qin, K; Huang, W; Zhang, T; Tang, S. Machine learning and deep learning for blood pressure prediction: a methodological review from multiple perspectives. Artif. Intell. Rev.; 2022; 56,
123. Cheung, MY; Sabharwal, A; Coté, GL; Veeraraghavan, A. Wearable blood pressure monitoring devices: understanding heterogeneity in design and evaluation. IEEE Trans. Biomed. Eng.; 2024; 71,
124. Alajlan, NN; Ibrahim, DM. TinyML: enabling of inference deep learning models on ultra-low-power IoT edge devices for AI applications. Micromachines; 2022; 13,
125. Wang, X; Han, Y; Leung, VCM; Niyato, D; Yan, X; Chen, Xu. Convergence of edge computing and deep learning: a comprehensive survey. IEEE Commun. Surv. Tutor.; 2020; 22,
126. Su, W; Li, L; Liu, F; He, M; Liang, X. AI on the edge: a comprehensive review. Artif. Intell. Rev.; 2022; 55,
127. Durmaz Incel, O; Bursa, SÖ. On-device deep learning for mobile and wearable sensing applications: a review. IEEE Sens. J.; 2023; 23,
128. Goossens, W; Mustefa, D; Scholle, D; Fotouhi, H; Denil, J. Evaluating edge computing and compression for remote cuff-less blood pressure monitoring. J. Sens. Actuator Netw.; 2023; 12,
129. Gao, Y-H; Li, J; Fan, C-H; Leung, KN; Zhang, Y-T et al. A 9.84-μW 148.9-dB total DR light-to-digital converter with current-integration SAR quantizer for multi-wavelength PPG applications. IEEE Trans. Circuits Syst. I Regul. Pap.; 2025; 72,
130. T. Sipola, J. Alatalo, T. Kokkonen, M. Rantonen, Artificial intelligence in the IoT era: a review of edge AI hardware and software. in 2022 31st Conference of Open Innovations Association (FRUCT). (IEEE, 2022), pp. 320–331. https://doi.org/10.23919/FRUCT54823.2022.9770931
131. S. Liu, Y. Lin, Z. Zhou, K. Nan, H. Liu et al., On-demand deep model compression for mobile devices: a usage-driven model selection framework. in Proceedings of the 16th Annual International Conference on Mobile Systems, Applications, and Services. (ACM, Munich, 2018), pp. 389–400. https://doi.org/10.1145/3210240.3210337
132. Liu, Z-D; Li, Y; Zhang, Y-T; Zeng, J; Chen, Z-X et al. Cuffless blood pressure measurement using smartwatches: a large-scale validation study. IEEE J. Biomed. Health Inform.; 2023; 27,
133. R. He, Z.-P. Huang, L.-Y. Ji, J.-K. Wu, H. Li et al., Beat-to-beat ambulatory blood pressure estimation based on random forest. in 2016 IEEE 13th International Conference on Wearable and Implantable Body Sensor Networks (BSN). (IEEE, 2016), pp. 194–198. https://doi.org/10.1109/BSN.2016.7516258
134. Mieloszyk, R; Twede, H; Lester, J; Wander, J; Basu, S et al. A comparison of wearable tonometry, photoplethysmography, and electrocardiography for cuffless measurement of blood pressure in an ambulatory setting. IEEE J. Biomed. Health Inform.; 2022; 26,
135. Leitner, J; Chiang, P-H; Dey, S. Personalized blood pressure estimation using photoplethysmography: a transfer learning approach. IEEE J. Biomed. Health Inform.; 2022; 26,
136. K. Ahmed, M. Hassan, tinyCare: a tinyML-based low-cost continuous blood pressure estimation on the extreme edge. in 2022 IEEE 10th International Conference on Healthcare Informatics (ICHI). (IEEE, 2022), pp. 264–275. https://doi.org/10.1109/ICHI54592.2022.00047
137. Jia, M; Qin, Y; Song, C; Yue, Z; Ding, S. CEBPM: a cloud-edge collaborative noncontact blood pressure estimation model. IEEE Trans. Instrum. Meas.; 2022; 71, [DOI: https://dx.doi.org/10.1109/TIM.2022.3205679] 5022712.
138. González-Nóvoa, JA; Busto, L; Campanioni, S; Martínez, C; Fariña, J; Rodríguez-Andina, JJ; Juan-Salvadores, P; Jiménez, V; Íñiguez, A; Veiga, C. Advancing cuffless arterial blood pressure estimation: a patient-specific optimized approach reducing computational requirements. Future Gener. Comput. Syst.; 2025; 166, [DOI: https://dx.doi.org/10.1016/j.future.2024.107689] 107689.
139. A. Burrello, F. Carlucci, G. Pollo, X. Wang, M. Poncino et al., Optimization and deployment of deep neural networks for PPG-based blood pressure estimation targeting low-power wearables. In 2024 IEEE Biomedical Circuits and Systems Conference (BioCAS). (IEEE, 2024), pp. 1–5. https://doi.org/10.1109/BioCAS61083.2024.10798404
140. W. Lin, B.U. Demirel, M.A. Al Faruque, G.P. Li, Energy-efficient blood pressure monitoring based on single-site photoplethysmogram on wearable devices. in 2021 43rd Annual International Conference of the IEEE Engineering in Medicine & Biology Society (EMBC). November 1–5, 2021. (IEEE, Mexico, 2021), pp. 504–507. https://doi.org/10.1109/embc46164.2021.9630488
141. K. Rishi Vardhan, S. Vedanth, G. Poojah, K. Abhishek, M. Nitish Kumar et al., BP-net: efficient deep learning for continuous arterial blood pressure estimation using photoplethysmogram. in 2021 20th IEEE International Conference on Machine Learning and Applications (ICMLA). December 13–16, 2021. (IEEE, Pasadena, 2021), pp. 1495–1500. https://doi.org/10.1109/icmla52953.2021.00241
142. S. Banerjee, B. Kumar, A.P. James, J.N. Tripathi, Blood pressure estimation from ECG data using XGBoost and ANN for wearable devices. in 2022 29th IEEE International Conference on Electronics, Circuits and Systems (ICECS). (IEEE, 2022), pp. 1–4. https://doi.org/10.1109/ICECS202256217.2022.9970924
143. Bernard, D; Msigwa, C; Yun, J. Toward IoT-based medical edge devices: PPG-based blood pressure estimation application. IEEE Internet Things J.; 2023; 10,
144. Liu, S-H; Wu, B-Y; Zhu, X; Chin, C-L. Using a bodily weight-fat scale for cuffless blood pressure measurement based on the edge computing system. Sensors; 2024; 24,
145. B. Sun, S. Bayes, A.M. Abotaleb, M. Hassan, The case for tinyML in healthcare: CNNs for real-time on-edge blood pressure estimation. in Proceedings of the 38th ACM/SIGAPP Symposium on Applied Computing. (ACM, Tallinn Estonia, 2023), pp. 629–638. https://doi.org/10.1145/3555776.3577747
146. Roy, MS; Gupta, R; Das Sharma, K. Bepcon: a photoplethysmography-based quality-aware continuous beat-to-beat blood pressure measurement technique using deep learning. IEEE Trans. Instrum. Meas.; 2022; 71, [DOI: https://dx.doi.org/10.1109/TIM.2022.3212750] 2519709.
147. N.F. Ali, M. Hussein, F. Awwad, M. Atef, Convolutional autoencoder for real-time PPG based blood pressure monitoring using TinyML. in 2023 International Conference on Microelectronics (ICM). (IEEE, 2024), pp. 41–45. https://doi.org/10.1109/ICM60448.2023.10378901
148. Casadei, BC; Gumiero, A; Tantillo, G; Della Torre, L; Olmo, G. Systolic blood pressure estimation from PPG signal using ANN. Electronics; 2022; 11,
149. Joseph, T. Real-time blood pressure prediction on wearables with edge-based DNNs: a co-design approach. ACM Trans. Des. Autom. Electron. Syst.; 2025; 30,
150. M. Sheng, R. Xing, Y. Xin, B. Zhang, Z. Guo et al., A 4.4 μW cuffless blood pressure measurement processor based on event-driven and module-level asynchronous scheme. in 2024 IEEE Biomedical Circuits and Systems Conference (BioCAS). (IEEE, 2024), pp. 1–5. https://doi.org/10.1109/BioCAS61083.2024.10798356
151. J. Zhang, J. Li, Y. Jiang, K. Wang, R. Guo et al., A hardware-based lightweight ANN for real-time wearable blood pressure estimation. in 2022 44th Annual International Conference of the IEEE Engineering in Medicine & Biology Society (EMBC). (IEEE, 2022), pp. 4295–4298. https://doi.org/10.1109/EMBC48229.2022.9871215
152. Almeida, TP; Cortés, M; Perruchoud, D; Alexandre, J; Vermare, P et al. Aktiia cuffless blood pressure monitor yields equivalent daytime blood pressure measurements compared to a 24-h ambulatory blood pressure monitor: preliminary results from a prospective single-center study. Hypertens. Res.; 2023; 46,
153. Vaseekaran, M; Kaese, S; Görlich, D; Wiemer, M; Samol, A. WATCH-BPM: comparison of a WATCH-type blood pressure monitor with a conventional ambulatory blood pressure monitor and auscultatory sphygmomanometry. Sensors; 2023; 23,
154. Valencell. Valencell unveils calibration free cuffless blood pressure monitoring solution targeting over the counter use (2023)
155. Nachman, D; Gepner, Y; Goldstein, N; Kabakov, E; Ben Ishay, A et al. Comparing blood pressure measurements between a photoplethysmography-based and a standard cuff-based manometry device. Sci. Rep.; 2020; 10,
156. Sayer, G; Piper, G; Vorovich, E; Raikhelkar, J; Kim, GH et al. Continuous monitoring of blood pressure using a wrist-worn cuffless device. Am. J. Hypertens.; 2022; 35,
157. Kumar, PS; Rai, P; Ramasamy, M; Varadan, VK; Varadan, VK. Multiparametric cloth-based wearable, simplesense, estimates blood pressure. Sci. Rep.; 2022; 12,
158. Food, A. Drug. Nanowear inc. 510(k) submission summary (k232053) (2025)
159. Food, A. Drug; 2025; Food and Drug Administration: Pyrames Inc:
160. Festo, C; Vannevel, V; Ali, H; Tamrat, T; Mollel, GJ et al. Accuracy of a smartphone application for blood pressure estimation in Bangladesh, South Africa, and Tanzania. npj Digit. Med.; 2023; 6,
161. Falter, M; Scherrenberg, M; Driesen, K; Pieters, Z; Kaihara, T; Xu, L; Caiani, EG; Castiglioni, P; Faini, A; Parati, G; Dendale, P. Smartwatch-based blood pressure measurement demonstrates insufficient accuracy. Front. Cardiovasc. Med.; 2022; 9, [DOI: https://dx.doi.org/10.3389/fcvm.2022.958212] 958212.
162. Lee, H; Park, S; Kwon, H; Cho, B; Park, JH; Lee, H-Y. Feasibility and effectiveness of a ring-type blood pressure measurement device compared with 24-hour ambulatory blood pressure monitoring device. Korean Circ. J.; 2024; 54,
163. H. Benmeziane, K. El Maghraoui, H. Ouarnoughi, S. Niar, M. Wistuba et al., Hardware-aware neural architecture search: survey and taxonomy. in Proceedings of the Thirtieth International Joint Conference on Artificial Intelligence. August 19–27, 2021. Montreal, Canada. International Joint Conferences on Artificial Intelligence Organization (2021), pp. 4322–4329. https://doi.org/10.24963/ijcai.2021/592
164. Y. He, X. Zhang, J. Sun, Channel pruning for accelerating very deep neural networks. in 2017 IEEE International Conference on Computer Vision (ICCV). (IEEE, 2017), pp. 1398–1406. https://doi.org/10.1109/ICCV.2017.155
165. T.-J. Yang, Y.-H. Chen, V. Sze, Designing energy-efficient convolutional neural networks using energy-aware pruning. in 2017 IEEE Conference on Computer Vision and Pattern Recognition (CVPR). (IEEE, 2017), pp. 6071–6079
166. Gou, J; Yu, B; Maybank, SJ; Tao, D. Knowledge distillation: a survey. Int. J. Comput. Vis.; 2021; 129,
167. Zhang, X; Zou, J; He, K; Sun, J. Accelerating very deep convolutional networks for classification and detection. IEEE Trans. Pattern Anal. Mach. Intell.; 2016; 38,
168. V. Lebedev, Y. Ganin, M. Rakhuba, I. Oseledets, V. Lempitsky, Speeding-up convolutional neural networks using fine-tuned cp-decomposition (2014). arXiv preprint arXiv:14126553
169. D. Gupta, A. Bhatti, S. Parmar, C. Dan, Y. Liu et al., Low-rank adaptation of time series foundational models for out-of-domain modality forecasting. in International Conference on Multimodel Interaction. (ACM, San Jose Costa Rica, 2024), pp. 382–386. https://doi.org/10.1145/3678957.3685724
170. Y.D. Kwon, J. Chauhan, A. Kumar, P.H. Hkust, C. Mascolo, Exploring system performance of continual learning for mobile and embedded sensing applications. in 2021 IEEE/ACM Symposium on Edge Computing (SEC) (2021), pp. 319–332
171. Chen, Y; Qin, X; Wang, J; Yu, C; Gao, W. Fedhealth: a federated transfer learning framework for wearable healthcare. IEEE Intell. Syst.; 2020; 35,
172. Ni, W; Ao, H; Tian, H; Eldar, YC; Niyato, D. Fedsl: federated split learning for collaborative healthcare analytics on resource-constrained wearable iomt devices. IEEE Internet Things J.; 2024; [DOI: https://dx.doi.org/10.1109/JIOT.2024.3370985]
173. T. Chen, T. Moreau, Z. Jiang, L. Zheng, E. Yan et al., Tvm: an automated end-to-end optimizing compiler for deep learning. in 13th USENIX Symposium on Operating Systems Design and Implementation (OSDI 18). (2018), pp. 578–594
174. M. Sponner, B. Waschneck, A. Kumar, Compiler toolchains for deep learning workloads on embedded platforms. arXiv preprint arXiv:210404576 (2021)
175. Cai, H; Lin, J; Lin, Y; Liu, Z; Tang, H et al. Enable deep learning on mobile devices: methods, systems, and applications. ACM Trans. Des. Autom. Electron. Syst.; 2022; 27,
176. Shuvo, MMH; Islam, SK; Cheng, J; Morshed, BI. Efficient acceleration of deep learning inference on resource-constrained edge devices: a review. Proc. IEEE; 2023; 111,
177. B. Liang, K. Duan, Q. Xie, M. Atef, Z. Qian et al., Live demonstration: a support vector machine based hardware platform for blood pressure prediction. in 2016 IEEE Biomedical Circuits and Systems Conference (BioCAS). (IEEE, 2017), pp. 130–130. https://doi.org/10.1109/BioCAS.2016.7833744
178. Mahajan, A; Heydari, K; Powell, D. Wearable AI to enhance patient safety and clinical decision-making. NPJ Digit. Med.; 2025; 8,
179. Shah, K; Wang, A; Chen, Y; Munjal, J; Chhabra, S; Stange, A; Wei, E; Phan, T; Giest, T; Hawkins, B; Puppala, D; Silver, E; Cai, L; Rajagopalan, S; Shi, E; Lee, Y-L; Wimmer, M; Rudrapatna, P; Rea, T; Yuen, S; Pathak, A; Patel, S; Malhotra, M; Stogaitis, M; Phan, J; Patel, B; Vasquez, A; Fox, C; Connell, A; Taylor, J; Shreibati, J; Miller, D; McDuff, D; Kohli, P; Gadh, T; Sunshine, J. Automated loss of pulse detection on a consumer smartwatch. Nature; 2025; 642,
180. Mennella, C; Maniscalco, U; De Pietro, G; Esposito, M. Ethical and regulatory challenges of AI technologies in healthcare: a narrative review. Heliyon; 2024; 10,
181. Wołos, K; Pstras, L; Debowska, M; Dabrowski, W; Siwicka-Gieroba, D; Poleszczuk, J. Non-invasive assessment of stroke volume and cardiovascular parameters based on peripheral pressure waveform. PLoS Comput. Biol.; 2024; 20,
182. Mukkamala, R; Yavarimanesh, M; Natarajan, K; Hahn, J-O; Kyriakoulis, KG et al. Evaluation of the accuracy of cuffless blood pressure measurement devices: challenges and proposals. Hypertension; 2021; 78,
183. A. Burrello, M. Risso, N. Tomasello, Y. Chen, L. Benini et al., Energy-efficient wearable-to-mobile offload of ML inference for PPG-based heart-rate estimation. in 2023 Design, Automation & Test in Europe Conference & Exhibition (DATE). (IEEE, 2023), pp: 1–6. https://doi.org/10.23919/DATE56975.2023.10137129
184. Stergiou, GS; Alpert, B; Mieke, S; Asmar, R; Atkins, N et al. A universal standard for the validation of blood pressure measuring devices: Association for the Advancement of Medical Instrumentation/European Society of Hypertension/International Organization for Standardization (AAMI/ESH/ISO) collaboration statement. J. Hypertens.; 2018; 36,
185. O’Brien, E; Petrie, J; Littler, W; de Swiet, M; Padfield, PL et al. An outline of the revised British Hypertension Society protocol for the evaluation of blood pressure measuring devices. J. Hypertens.; 1993; 11,
186. Iso 81060-3:2022. Non-invasive sphygmomanometers—part 3: clinical investigation of continuous automated measurement type (2022)
© The Author(s) 2026. This work is published under http://creativecommons.org/licenses/by/4.0/ (the “License”). Notwithstanding the ProQuest Terms and Conditions, you may use this content in accordance with the terms of the License.