Introduction
Soil-borne radon (Rn) is widely used as a tracer for atmospheric mixing and transport model validation (e.g. Jacob and Prather, 1990; Taguchi et al., 2011), as its sources are relatively well known (e.g. Nazaroff, 1992; Karstens et al., 2015), while its only sink is radioactive decay. Continuous atmospheric Rn is normally measured via the activity of its short-lived progeny, such as polonium (Po) and polonium (Po), which are attached to aerosols shortly after generation in the atmosphere (Porstendörfer, 1994). In so-called two-filter systems, filtered air that only contains Rn (with ambient progeny removed) is flushed through a large delay chamber and the new Rn progeny produced in situ inside the tank is then collected on a second internal filter where their decay is monitored (e.g. Whittlestone and Zahorowski, 1998; Chambers et al., 2011). Alternative Rn monitors, the so-called one-filter systems, collect Rn progeny directly from ambient air on a static filter or automatically changing filters and measure its decay rates on the filter. In this style of detector, atmospheric Rn activity concentration is then deduced from the filter activity of its progeny, making assumptions about the disequilibrium between Rn and its progeny on the filter and in the atmosphere (e.g. Levin et al., 2002; Schmithüsen et al., 2017). Disequilibrium between Rn and its daughters depends on height above ground level, on meteorological conditions and on the influence of processes causing aerosol loss by dry or wet deposition (see e.g. Jacobi and André, 1963; Schery and Wasiolek, 1993; Porstendörfer, 1994). These non-negligible biases can be investigated by comparison with two-filter detectors (Xia et al., 2010).
In addition, potential artefacts from air intake and sampling lines of one-filter systems need to be investigated before one-filter monitors could be implemented at locations (e.g. at tower stations), where long intake tubing is required. To minimize loss of aerosol-attached Rn daughters in the intake systems, one-filter monitors are normally used only at ground level stations, where short intake lines are possible. This principally disqualifies one-filter systems to be applied at tall tower stations, unless the filter head can be installed directly at the level where the Rn daughter measurements will be performed (e.g. at an accessible platform). Alternatively, corrections must be applied for loss of aerosol-attached Po or other progeny in the tubing. This, however, requires full quantitative understanding of all aerosol transport loss and deposition processes in the intake system, such as diffusion to the walls, sedimentation or inertial deposition (see e.g. Willeke and Baron, 2005, for a comprehensive discussion of these processes).
In the present work, we present the results from dedicated laboratory and
field experiments, which had the aim to determine Rn progeny loss in
long tubing. The study was motivated by the requirement to accompany
greenhouse gases measurements with continuous Rn observations within
the new European ICOS (Integrated Carbon Observation System) atmospheric
station network
(
Methods
Heidelberg static (one) filter Rn monitor (HRM)
In Heidelberg, a medium-sized city located in the Upper Rhine Valley in south-western Germany (4925 N, 841 E; 116 m a.s.l.), atmospheric Rn daughter activity concentrations have been continuously monitored since 1996 with a static filter system. Measurements were conducted at the Institut für Umweltphysik, Heidelberg University, located on the university campus on the outskirts of the city. Since 1999, ambient air is collected at about 30 m above local ground. These Heidelberg Rn data have been used as a tracer for boundary layer mixing in a number of studies (e.g. Hammer and Levin, 2009; Levin et al., 2011) or for regional transport model comparison exercises (Taguchi et al., 2011).
The principal set-up of the HRM was designed in the 1990s (Levin et al., 2002). The monitor was modernized in 2010 by implementing state-of-the-art electronics as well as data acquisition and evaluation hardware and software (Rosenfeld, 2010). The system consists of a homemade filter holder, carrying a Whatman filter (QMA Ø 47 mm), where all aerosols from an ambient airflow of ca. 1 m h (measured with a Bronkhorst mass flow meter, model F-112AC-AAD-22-V) are actively deposited. A surface barrier detector (Canberra CAM 900 mm active surface) mounted in the filter holder at about 5 mm distance from the loaded filter surface measures the particles from the decaying Rn and Rn daughters. Hourly or half-hourly integrated spectra are stored for later separation of polonium (Po) counts from the spectrum. Levin et al. (2002) explain evaluation of the spectra in detail. Taking into account the airflow rate through the filter, the filter efficiency (assumed as 100 %) and the solid angle of the detector, which depends on the distance of the detector from the filter, finally allow estimation of the atmospheric Po activity concentration. For calculation of ambient Rn activity concentration, a correction for disequilibrium of Po has to be made (Levin et al., 2002; Schmithüsen et al., 2017).
Set-up of the line tests at the Heidelberg measurement site
To minimize aerosol loss in the sample intake, the standard intake line for the HRM is a ca. 0.5 m long 8 mm inner diameter (ID) Teflon tubing. In Heidelberg, this tubing is connected directly to outside air through a hole in the laboratory window. For investigating the effect of different intake lines on the measured Po activity concentration, we used standard Decabon (SertoFlex) tubing with 12 mm outer and 8.2 mm inner diameter, which was varied in length between 16 and 200 m. Before each test, the monitor with the long intake tube was run with a standard intake line for more than 1 week in parallel to our routine Heidelberg radon monitor. This allowed us to determine a “calibration” of the so-called line test monitor (LTM) relative to the Heidelberg routine instrument (HD-R). Small systematic differences of typically less than 5 % can occur between monitors, which may be due to slightly different solid angles or efficiencies of the surface barrier detectors or of the mass flow meter calibrations. From this comparison we estimate a statistical measurement uncertainty of activity ratios for the two Heidelberg monitors of about 25 % for ambient activity concentrations below 1 Bq m and of 10 % or less for concentrations above 1 Bq m After the initial calibration period, each line test was run for at least 3 weeks to make sure that the typical activity concentration range in Heidelberg of 1–15 Bq m was covered by the measurements. For these laboratory tests, the extended intake lines were rolled up to a diameter of approximately 0.5 m; they were always kept horizontally and at laboratory temperature of 19–22 C. For all laboratory tests with line lengths 100 m, flow rates were around 1 m h (10 %), keeping the contact time per length of tubing constant within about 10 %. For the tests with shorter tubing, flow rates were up to 70 % higher.
Set-up of the Cabauw experiment (NL, 5158 N 456 E; m a.s.l.)
Cabauw tower is a 213 m high tower, specifically built for meteorological research to establish relations between the states of the atmospheric boundary layer, land surface conditions and the general weather situation for all seasons. Cabauw tower is located in the western part of the Netherlands in a polder 0.7 m below average sea level. The North Sea is more than 50 km away to the WNW.
Rn is continuously measured at Cabauw station from two different heights, 20 and 200 m above ground level; air is drawn at approximately 100 L min through 7 cm outer diameter terylene fibre water pipes through stack blowers into the measurement devices. These ANSTO radon monitoring devices are based on the two-filter technique (Whittlestone and Zahorowski, 1998; Zahorowski et al., 2004). The details of the Cabauw monitors and their calibration are described in Popa et al. (2011) as well as in the companion paper by Schmithüsen et al. (2017).
Our first set of measurements with a HRM at Cabauw tower (15 December 2011 to 21 February 2012) was through a pre-installed Decabon tubing (in total 200 m) from the 180 m level down to the basement, where the HRM filter head was installed. This tubing had a slightly larger inner diameter than the tubing tested in the laboratory, namely 10 mm, and the flow rate was 1.1 m h (10 %). A second set of measurements was performed with the HRM standard intake line (0.5 m) also from the 180 m platform (10 July–26 August 2012). For these latter measurements we assume no aerosol loss during air intake of the HRM; the respective data thus serve as reference for the compatibility of the two measurement systems (ANSTO and HRM) located at slightly different heights above ground (ANSTO: 200 m and HRM: 180 m). They allow the determination of possible combined calibration biases or mean differences due to disequilibrium of the HRM-measured Po to the ANSTO-measured Rn activity concentrations (see accompanying paper by Schmithüsen et al., 2017).
(a) Correlation of the Po activity concentration measured with the line test monitor (LTM) sampling via a 200 m intake line vs. that measured with the Heidelberg routine monitor (HD-R). The red line is a least squares fit (York et al., 2004) through the data yielding a slope of 0.66; the dashed blue line shows the 1 : 1 relation. (b) Activity concentration ratio of the LTM and the HD-R monitor plotted vs. the activity concentration measured with the LTM. The red line is an exponential fit through the data according to Eq. (1).
[Figure omitted. See PDF]
(a) Length-dependent saturation values as determined from individual line tests (compare also Figs. S1–S9 in the Supplement). The solid red line is an exponential fit calculated through all experimental data according to Eq. (2). Black open symbols show experimental values adjusted to a standard flow rate of 1 m h, according to Eqs. (3)–(5), with the dashed black line being the exponential fit through these adjusted data. (b) Length-corrected line test results (i.e. normalized to ) together with the saturation fit curve calculated according to Eq. (). (c) Activity concentration ratio of corrected LTM data and HD-R data without tubing. Panel (d) gives the standard deviations of ratios displayed in (c) binned into 0.5 Bq m ranges.
[Figure omitted. See PDF]
Results
Results of the line tests performed in Heidelberg
Figure 1 shows an example of results from a line test with 8.2 mm inner diameter Decabon tubing in the Heidelberg laboratory. Here we plot the Po activity concentration measured on ambient air collected via a 200 m intake line (LTM) vs. that measured with the HD-R monitor sampling the same air via a 0.5 m intake tubing. The correlation between the two data sets is very good; however, the slope of the regression line is only 0.66, indicating that we lose more than 30 % of Rn progeny activity concentration in the tubing. When plotting the ratio of the respective data vs. the activity measured with the LTM we can further see that at activity concentrations smaller than 2 Bq m, the relative activity loss for the LTM appears to be even larger than at high activities. Line tests with shorter tubing than 200 m show a similar pattern; however, the shorter the tubing is, the higher the “saturation ratio” at high atmospheric activity concentrations. The activity dependency of the ratios LTM/HD-R (e.g. Fig. 1b) can be approximated by a saturation curve according to the following function: where the free parameters and are constants, and is the “saturation” activity ratio for line length . The activity ratios from the 200 m line test have been fitted accordingly in Fig. 1b with the fit curve plotted as solid red line with . Similar fit curves as shown in Fig. 1 have been calculated through all laboratory line test results with tubing between 16 and 200 m. This process yielded saturation ratios decreasing with increasing line length ; respective results are displayed in Fig. 2a (red squares), where the saturation ratios are plotted vs. the length of the tubing of the LTM. An exponentially decreasing saturation value is observed with increasing length of the tubing and a respective exponential fit is plotted through the data (solid red line in Fig. 2a): where is a constant parameter. Note that a straight line would fit our data equally well, but we expect to approach zero at infinite line length; therefore we used an exponential fit. The error bars in Fig. 2a correspond to the standard deviation of all measurements from each line test with activity concentrations 7 Bq m. Equation (2) could therefore in principle be used to correct measurements for the loss of Rn progeny activity concentration at activity concentrations 7 Bq m in 8.2 mm Decabon tubing.
In order to additionally account for the activity-dependent loss (i.e. at activity values 7 Bq m, we corrected all our line test data with different line lengths for the general “saturation” loss by dividing all values by their respective saturation value and combined them into one single “normalized” data set of corrected activity ratios (Fig. 2b). We then applied a joint fit through all data according to Eq. (1). This fit using the length-corrected data – by definition – yields a mean saturation value . The maximum activity concentration loss at very low activities () approaches 40 % (Fig. 2b). The parameters of the joint fit as well as those of the fit for the general length-dependent loss correction (as shown in Fig. 2a) are m; ; Bq m.
Evaluation of the empirical activity loss correction
In order to test if our empirical correction method yields reliable results, we have applied Eqs. (1) and (2), using the fit parameters from Fig. 2 to all individual line tests and plotted the results in Figs. S1–S9 in the Supplement. These figures show in their upper left panels the original ratios, in their lower left panels the ratios of the LTM-corrected data, in their upper right panels the linear correlation of corrected with and in their lower right panels the regular residuum from the fit line in the lower left panel. For all line lengths the slopes lie between 0.997 and 1.064 and the scatter of the corrected ratios for activities 2 Bq m is less than 20 % in all cases, but with tubing longer than 50 m it may increase to more than 50 % for activity concentrations lower than 1 Bq m. For a quantitative estimate of the average uncertainty of our correction of the laboratory results, we display in Fig. 2c the ratios between normalized and corrected LTM data and the routine HRM data. In Fig. 2d, we calculated standard deviations for binned activity concentration ranges of 0.5 Bq m each. The standard deviation is of order 60 % for the lowest activity concentration range of 0–0.5 Bq m, but it quickly decreases to less than 20 % above 1 Bq m. Taking into account the uncertainty of activity ratios from two HRMs without tubing of ca. 25 % below 1 Bq m (see Sect. 2.2) and about 10 % above 1 Bq m, the additional uncertainty of the correction for the effects of long tubing is of the order of 40 % below 1 Bq m, decreases to about 17 % for 1.5 Bq m and to less than 10 % above 2 Bq m. We find a small bias in the corrected mean ratios of ca. 10 % below 1.5 Bq m. A more sophisticated correction may allow for adjusting for that bias; however, we decided to keep the correction simple and valid for all tube lengths. In view of the relatively large statistical uncertainty of the measurement through long tubing ( 50 m) at activity concentrations below 1.5 Bq m and in view of the fact that we cannot fully transfer our laboratory experiments to real field measurements (see below), we refrain here from further second-order adjustments.
Rn progeny loss during ambient air sampling at Cabauw
In addition to the line tests made with rolled-up 8.2 mm ID Decabon tubing in the Heidelberg laboratory and to evaluate the potential correction for radon progeny loss in vertical tubing, a line test has been made during the comparison campaign at Cabauw (Schmithüsen et al., 2017). From a direct comparison between the local ANSTO system, measuring Rn at 200 m and the HRM, sampling air through a 0.5 m tubing at the 180 m platform, we obtained a mean difference between the ANSTO Rn data and the HRM Po data of ANSTO HRM 1.11. The 11 % bias between the two systems may be due to calibration and/or disequilibrium differences between the two systems but is, most probably, due to calibration differences (Schmithüsen et al., 2017). For the line test we sampled air for the HRM over 2 months through 200 m and 10 mm ID Decabon tubing from the 180 m level of the Cabauw tower at a flow rate of 1.1 m h (10 %), while the routine ANSTO instrumentation was also measuring Rn in air collected at 200 m. In the comparison shown in Fig. 3, we have corrected the ANSTO data with the above mentioned calibration factor of , but there is still a significant difference visible between the two data sets. The slope of the regression line shown in Fig. 3b is HRM/ANSTO_corr 0.73, indicating a mean loss of 27 % of aerosol-bound Rn progeny in the 200 m tubing of the HRM. From our laboratory tests we would have expected a larger loss, i.e. 34 % for a 200 m tubing. Note, however, that a tubing with slightly smaller inner diameter of 8.2 mm was used in the laboratory tests, while at Cabauw tower the tubing had an inner diameter of 10 mm, which can explain most of the difference. Moreover, in the laboratory tests the tubing was rolled up and lying on the floor, with potential loss due to inertial deposition on the bent tube walls as well as gravitational settling of aerosol and associated loss (see Sect. 4).
Discussion
The relative loss of Po activity concentrations (see Eq. 2: when sampling through a long tubing shows the expected correlation with tubing length (assumed here to be exponential, but a linear correlation would also fit the measurement data) and a rather subtle dependency on the atmospheric activity concentration itself (Fig. 2b). Intake lines are known to be a potential sink for aerosols (see e.g. Willeke and Baron, 2005; von der Weiden et al., 2009), and the majority of radon daughters measured with the one-filter HRM are attached to aerosols. The exact dependency of on the intake line length depends on the size of the aerosols carrying the radon daughter activity and on the flow regime in the tubing as well as its shape and orientation. Reineking et al. (1988) and Porstendörfer et al. (1994) report that most radon progeny activity in open air is carried by particles in the accumulation mode, with an average median aerodynamic diameter of 369 nm (range of 173–645 nm). Depending on meteorological conditions, a small fraction could also be found in the nucleation mode, with a diameter size range of 10–100 nm (Reineking et al., 1988). The flow regime in our Decabon tubing with or 10 mm ID, a flow rate of about 1 m h and air (and particle) velocities between m s cannot be characterized as purely laminar (Reynolds number ) or as purely turbulent (). With , where is the kinematic viscosity of the medium (in our case air at approximately 20 C, i.e. m s, our Reynold numbers lie between 2350 and 2900, i.e. in the transitional regime, where for many loss processes no formula is available (von der Weiden et al., 2009).
Results from the line test experiment at Cabauw tower. (a) shows Rn activity concentrations measured with the ANSTO system on air collected from the 200 m platform (red line), but corrected by a factor of (see text) in comparison to the HRM Po measurements on air from the 180 m level of the tower collected through a 200 m Decabon tubing (black line). (b) shows the scatter plot of the two records with a geometric mean regression line (York et al., 2004) yielding a slope of 0.73.
[Figure omitted. See PDF]
For the laboratory experiment, four transport loss processes are potentially relevant to explain the observed radon progeny loss: (1) diffusion by Brownian motion towards the walls; (2) turbulent inertial deposition, when the (larger) particles cannot follow turbulent streamlines and thus stick to the walls; (3) gravitational settling in the horizontal tube; and (4) inertial deposition on the bent tube walls. For the Cabauw experiment, with vertical tubing over most of the intake line, only the first two processes are potentially of relevance.
-
Loss by molecular diffusion following a gradient towards the walls is stated by von der Weiden et al. (2009) as being relevant only for very small particles ( 100 nm aerodynamic diameter). With aerodynamic diameters between 173 and 645 nm, this process is probably negligible for the Heidelberg lab experiments and at Cabauw tower, too. However, if a significant fraction of radon progeny would be unattached, this loss process may be significant (see below).
-
Although our flow is in the transitional regime between laminar and turbulent, loss through turbulent inertial deposition may still be relevant. Von der Weiden et al. (2009) give an equation to estimate the turbulent inertial deposition efficiency (their Eq. 28), based on an experimentally determined turbulent inertial deposition velocity (Eq. 29) in tubing of length with diameter can be estimated with Eq. (29) of von der Weiden et al. (2009), based on the Stokes number Stk and the Reynolds Re number according toThe Stokes number is calculated according to Brockmann (2005, his Eqs. 6-1 and 6-2) , where is the gravitational settling velocity (in our case of aerosol with 369 nm median aerodynamic diameter estimated to about 2 cm h; Seinfeld and Pandis, 2006), and g is the gravitational acceleration. With a Stokes number of for the lab experiments, Eq. (3) yields a transport efficiency of . For the Cabauw tower tubing of 10 mm ID and flow rate of 1.1 m h the Stokes number is smaller; i.e. and the aerosol loss by turbulent deposition is also smaller and estimated by Eqs. (3) and (4) to .
-
In the laboratory tests, loss may also have been due to gravitational settling in the rolled-up tubes kept horizontally, although this effect is more important for larger particles 500 nm (von der Weiden et al., 2009). We can estimate the respective loss for laminar flow according to Eq. (24) of von der Weiden et al. (2009):where and . For turbulent flow von der Weiden et al. (2009) give a different equation, i.e. their Eq. (26) (Eq. 6-40 in Brockmann, 2005):However, for a settling velocity cm h (as estimated for our median diameter aerosol of 369 nm, see above) both Eq. (5a) and (5b) yield the same loss, i.e. for a 200 m horizontal tube of mm ID, a loss of about 3 %. Loss due to gravitational settling would double for a 645 nm median diameter aerosol and be only about 1 % for 175 nm aerosol.
-
The fourth process to be considered for loss in the laboratory experiment is due to inertial deposition at the walls of the bent tubing. Calculations of this lost process are commonly made following the data-based empirical correction functions given by Pui et al. (1987) (see e.g. Brockmann, 2005, and Von der Weiden et al., 2009). For laminar flow, Pui et al. (1987) show in their Figs. 7 and 8 that aerosol loss in bent tubing is negligible for Stokes numbers smaller than 0.01, which is the case in our laboratory experiments (see above). For turbulent flow, the latter authors developed a correction function independent of the Reynolds number and the curvature of the bend, i.e.where is the angle of the bend in radians and Stk is the Stokes number. Although the Stokes number in our experimental set-up for aerosol with median aerodynamic diameter of about 370 nm ( cm h is small (ca. , in the laboratory experiments with up to 200 m tubing rolled up, gets very large. With Eq. (6) we would thus obtain an efficiency of , only for this loss process. The data set used by Pui et al. (1987) to establish Eq. (6) had, however, Re numbers of 6000 and 10 000, and a curvature ratio (i.e. the radius of the bend divided by the radius of the tube, Brockmann, 2005) of 6. Thus flow and bending conditions under which Eq. (6) was obtained were quite a lot more turbulent and more sharply bent, respectively, compared to the conditions encountered within our experiment ( and ). However, Pui et al. (1987) affirm the insignificance of in Eq. (6) only for a range of . Considering this, we have to conclude that the conditions of our experiment might be only partially or not at all suitable to be evaluated with Eq. (6), leaving a quantitative estimate of this loss process open. Therefore, although we cannot fully rule out in our laboratory experiments aerosol loss due to inertial deposition in the bent tubing, it is likely that this effect is only small. One strong argument for this assumption is that we would have observed much larger differences in aerosol loss between laboratory experiments (with the rolled-up tubing) and the field experiment at Cabauw tower, where Rn progeny was sampled through an almost straight vertical tubing.
In conclusion, we assume that the most relevant process for aerosol loss in both our experimental settings was most probably through turbulent inertial deposition on the walls of the tubing. We favour this loss process also because, contrary to the other loss processes, it does not depend much on the median aerodynamic diameter of the aerosol. In fact, we cannot expect the size distributions of the Rn daughter-carrying aerosol to be identical at Cabauw tower located close to the sea and in Heidelberg, a rather polluted and more continental station. A few percent of loss may have additionally occurred through gravitational settling in the laboratory experiments. As we used higher flow rates in the laboratory for the shorter line lengths and as aerosol loss in the tubing depends on the flow rate (see Eqs. 3–5), we made an attempt to correct the laboratory results by normalising them all to a flow rate of 1 m h (see open squares in Fig. 2a). This yielded slightly higher values for the shorter line lengths and an exponential fit with a slightly smaller decrease rate with length . The difference to the original fit curve through the uncorrected experimental data increases with line lengths to 4.5 % for a 200 m tubing, which is well within the uncertainty of our measurements.
The activity dependency of aerosol loss, which was found in the laboratory (see Fig. 2b) as well as in the field experiment at Cabauw (not shown), is more difficult to explain. One hypothesis is that at low activity concentration also the aerosol concentration is lower, with a potentially shifted size distribution of the progeny-carrying aerosol towards smaller aerodynamic diameters. A significant fraction of progeny may also be unattached during these clean situations (Schery and Wasiolek, 1993). Small particles have a higher diffusion coefficient than the median diameter aerosol and have thus a higher probability to be lost by Brownian diffusion to the walls of the tubing. An even larger loss may be expected for charged (unattached) progeny. A more sophisticated loss calculation would, however, be required to explain the behaviour of progeny loss during these circumstances. This is outside the scope of our study, which was only looking for semi-quantitative explanations of our experimental findings.
Conclusions
Our laboratory and field experiments provided first results on Rn progeny loss in long Decabon tubing of 8.2 and 10 mm ID. As a likely loss mechanism we identified turbulent inertial deposition on the walls of the tubing, because the theoretical estimates of this loss process would equally well explain our findings in the two different experimental settings, i.e. in the laboratory with horizontally kept rolled-up 8.2 mm ID tubing and in the field with 10 mm ID straight vertical tubing. In these two cases with flow rates of 1–1.1 m h, we estimate sampling efficiencies for 200 m long tubing between 0.66 (8.2 mm ID) and 0.73 (10 mm ID). When also considering additional loss due to gravitational settling for the rolled-up tubing kept horizontally in the laboratory, the total sampling efficiency is reduced to 0.64. These theoretical estimates are in surprisingly good agreement with our experimental data that yielded efficiencies of 0.66 for the rolled-up tubing in the laboratory and 0.73 at the Cabauw tower.
Based on this agreement, the correction function, which was empirically derived from our laboratory experiments, showing an exponential increase of loss with tube length seems adequate and applicable for real field measurements with similar tubing and flow rates. One may consider taking a theoretical approach (Eqs. 3 and 4) to correct for Rn progeny loss in the tubing. However, we do not recommend this approach as our theoretical estimates were too simplified to be transferred to other tubing and flow conditions.
The observed activity dependency of the loss, when sampling through long tubing, could possibly be explained by decreasing aerodynamic diameter of the radon progeny-carrying aerosol. In order to correct for this effect we suggest applying our empirical observation-based function given by Eq. (1). However, one should keep in mind that the uncertainty of this correction is strongly increasing with decreasing atmospheric activity concentration. An additional uncertainty of 10–20 % will be associated with the corrected data, if Po activity concentrations lie between 1 and 2 Bq m. For activity concentrations above 2 Bq m, the additional uncertainty of the correction is smaller than 5 %, which seems acceptable in view of other uncertainties associated with the Rn progeny measurement. However, for activity concentrations below 1 Bq m, which very often occur at coastal stations or at sites not too far from the sea, like Cabauw (see Fig. 3), Rn progeny sampling through long tubing (i.e. longer than 50 m) is not recommended. Here our proposed correction will have huge uncertainty ( 50 %, see also figures in the Supplement). For more continental stations, i.e. more than a few hundred kilometres from the coastline, Rn activity concentrations drop below 1 Bq m only during intensive vertical mixing during summer days or when the air masses have stayed over the continent for less than a day and the continental Rn accumulation is still low. Therefore, at these stations one may well accept additional statistical uncertainty due to line loss corrections, also of order 10 %, which would be in the same range as the uncertainty of disequilibrium between Rn and its measured progeny (see accompanying paper by Schmithüsen et al., 2017).
Data tables are going to be published from the University of Heidelberg online at:
The Supplement related to this article is available online at
Acknowledgements
The research leading to these results has received funding from the European Community's Seventh Framework Program (FP7/2007-2013) in the InGOS project under grant agreement no. 284274. We wish to thank the technicians at Cabauw station, Pim van den Bulk and Mark Blom, for their support during the comparison study and Susanne Preunkert, LGGE, Grenoble for helpful discussions. Two anonymous reviewers are gratefully acknowledged for their insightful comments, which helped to improve the manuscript. We acknowledge financial support by Deutsche Forschungsgemeinschaft and Ruprecht-Karls-Universität Heidelberg within the funding programme Open Access Publishing. Edited by: C. Brümmer Reviewed by: two anonymous referees
You have requested "on-the-fly" machine translation of selected content from our databases. This functionality is provided solely for your convenience and is in no way intended to replace human translation. Show full disclaimer
Neither ProQuest nor its licensors make any representations or warranties with respect to the translations. The translations are automatically generated "AS IS" and "AS AVAILABLE" and are not retained in our systems. PROQUEST AND ITS LICENSORS SPECIFICALLY DISCLAIM ANY AND ALL EXPRESS OR IMPLIED WARRANTIES, INCLUDING WITHOUT LIMITATION, ANY WARRANTIES FOR AVAILABILITY, ACCURACY, TIMELINESS, COMPLETENESS, NON-INFRINGMENT, MERCHANTABILITY OR FITNESS FOR A PARTICULAR PURPOSE. Your use of the translations is subject to all use restrictions contained in your Electronic Products License Agreement and by using the translation functionality you agree to forgo any and all claims against ProQuest or its licensors for your use of the translation functionality and any output derived there from. Hide full disclaimer
© 2017. This work is published under http://creativecommons.org/licenses/by/3.0/ (the “License”). Notwithstanding the ProQuest Terms and Conditions, you may use this content in accordance with the terms of the License.
Abstract
Aerosol loss in air intake systems potentially hampers the application of one-filter systems for progeny-based atmospheric
You have requested "on-the-fly" machine translation of selected content from our databases. This functionality is provided solely for your convenience and is in no way intended to replace human translation. Show full disclaimer
Neither ProQuest nor its licensors make any representations or warranties with respect to the translations. The translations are automatically generated "AS IS" and "AS AVAILABLE" and are not retained in our systems. PROQUEST AND ITS LICENSORS SPECIFICALLY DISCLAIM ANY AND ALL EXPRESS OR IMPLIED WARRANTIES, INCLUDING WITHOUT LIMITATION, ANY WARRANTIES FOR AVAILABILITY, ACCURACY, TIMELINESS, COMPLETENESS, NON-INFRINGMENT, MERCHANTABILITY OR FITNESS FOR A PARTICULAR PURPOSE. Your use of the translations is subject to all use restrictions contained in your Electronic Products License Agreement and by using the translation functionality you agree to forgo any and all claims against ProQuest or its licensors for your use of the translation functionality and any output derived there from. Hide full disclaimer