ARTICLE
Received 13 Mar 2012 | Accepted 30 May 2012 | Published 3 Jul 2012 DOI: 10.1038/ncomms1935
S.P. Giblin1, M. Kataoka1, J.D. Fletcher1, P. See1, T.J.B.M. Janssen1, J.P. Grifths2, G.A.C. Jones2, I. Farrer2 & D.A. Ritchie2
Electron pumps generate a macroscopic electric current by controlled manipulation of single electrons. Despite intensive research towards a quantum current standard over the last 25 years, making a fast and accurate quantized electron pump has proved extremely difcult. Here we demonstrate that the accuracy of a semiconductor quantum dot pump can be dramatically improved by using specially designed gate drive waveforms. Our pump can generate a current of up to 150 pA, corresponding to almost a billion electrons per second, with an experimentally demonstrated current accuracy better than 1.2 parts per million (p.p.m.) and strong evidence, based on tting data to a model, that the true accuracy is approaching 0.01 p.p.m. This type of pump is a promising candidate for further development as a realization of the SI base unit ampere, following a redenition of the ampere in terms of a xed value of the elementary charge.
Towards a quantum representation of the ampere using single electron pumps
1 National Physical Laboratory, Hampton Road, Teddington, Middlesex TW11 0LW, UK. 2 Cavendish Laboratory, University of Cambridge, J J Thomson Avenue, Cambridge CB3 0HE, UK. Correspondence and requests for materials should be addressed to S.P.G. (email: [email protected]).
NATURE COMMUNICATIONS | 3:930 | DOI: 10.1038/ncomms1935 | www.nature.com/naturecommunications
2012 Macmillan Publishers Limited. All rights reserved.
ARTICLE
NATURE COMMUNICATIONS | DOI: 10.1038/ncomms1935
The AC Josephson eect and quantum Hall eect have revolutionized electrical metrology by providing stable reference standards linked to fundamental physics rather than partic
ular artefacts1. The resulting system of electrical units is coupled to the SI only through very difficult and time-consuming electromechanical experiments2, and recently a debate has ignited about redening the SI ampere in terms of the fundamental elementary charge e3. A redened ampere could be realised directly using a quantized electron pump, which generates a current by moving single electrons rapidly and in a controlled way. The electron pump accepts a periodic input signal at a repetition frequency f, and transports an integer number n of electrons between source and drain leads for each cycle of the input to yield a current IP = nef. Application as a quantum standard of electric current requires a pump having a combination of accuracy, simplicity of operation and ability to generate a reasonably high current IP100 pA4. In the past two decades, many pump technologies have been investigated in pursuit of this goal. These include chains of submicron normal-metal tunnel junctions5,6, quantum dots driven by a surface acoustic wave7 and normal metal/superconductor turnstiles8,9. Although 0.015 parts per million (p.p.m.) accuracy has been demonstrated for IP1 pA5, and IP500 pA with around 100 p.p.m. accuracy10, none of these
technologies have yet demonstrated the required combination of accuracy and large enough current output.
Our tunable-barrier electron pump (scanning electron microscope (SEM) image in Fig. 1a, with associated biasing voltages also shown) consists of a conducting channel etched in a two-dimensional electron gas (2-DEG), crossed by two metallic gates, which we denote the entrance and exit gates. Potential barriers are created in the 2-DEG below the gates by applying negative voltages VG1 and VG2 to these gates. An isolated quantum dot, holding
a small number of electrons, forms between the gates. A drive signal VRF(t) superimposed on VG1 periodically lowers the entrance barrier below the Fermi level of the source lead, picking
up electrons and liing them over the exit barrier into the drain lead. The exit barrier is kept high ( > 10 meV above the Fermi level) to suppress unwanted thermally activated and cotunnel transport11. A series of schematic potential diagrams in Fig. 1b illustrate the pump cycle for the case of n = 1. The pump operates at zero source-drain bias voltage; the direction of the pump current is determined by which gate is driven with the AC signal. If the amplitude of VRF
is large enough, the probability of the trapped electron being ejected into the drain lead (frame 4 of Fig. 1b) is unity, and the current quantization mainly depends on the initial part of the pump cycle (frame 2). Here, the relatively large number of electrons initially trapped by the rising entrance barrier is reduced to just one by a cascade of back-tunnel events whereby excess electrons return to the source lead (red arrow)12,13. Owing to the Coulomb charging energy of the dot, the back-tunnel probability for the last electron is many orders of magnitude less than for the second electron, so the convergence on the 1-electron state is robust and insensitive to details of the potential barrier shape. The GaAs pump has an important advantage compared with other types of electron pump, which is that the quantization accuracy can be improved with an external tuning parameter, the perpendicular magnetic eld B1418. The mechanism for this improvement is not fully understood, but the additional connement of electrons within the dot owing to the magnetic eld has an important role18.
Previous high-resolution measurements19,20 used sine wave drive for VRF(t). The sine wave cycle is shown in Fig. 1c with numbered points approximately indicating the stages of the pump cycle shown in Fig. 1b. These measurements could not be extended above f350 MHz because the quantized plateau degraded with further increase in f 20. The general mechanism for this is not currently understood, although clear signatures of back-tunneling due to nonadiabatic excitation at high f have been seen in some samples21.
To extend the pump operation range to higher frequencies, we developed a technique using an arbitrary waveform generator (AWG) to generate a custom VG1(t) waveform that takes into account the time-dependent electron tunelling dynamics. To design the waveform, rst we note, as shown schematically in Fig. 1c, that, for sine wave drive, most of the cycle time is used to raise the trapped electron over the exit barrier, and following ejection, lower the empty dot in readiness for the next cycle. This is a consequence of the height of the exit barrier and the large amplitude (0.5 V) of the entrance barrier modulation. Second, we postulate that the increased error rate at high frequency is related to the rate at which the entrance barrier is raised during the early part of the pump cycle, when back- tunneling occurs. Our AWG waveform, shown in Fig. 1d, was designed with a small initial dVG1/dt, speeding up for subsequent parts of the pump cycle. Our technique has some similarity to earlier work employing trapezoidal gate pulses with variable rise time12, and we will return to this point later in the Discussion section. The output of the AWG measured through a similar transmission line to the one used in the experiment showed a slight smearing of the actual VG1(t) applied to the entrance gate (red line in Fig. 1d), but this does not aect the important part of the cycle, which
Entrancegate Exit
gate
a
VG2
b
e
VG1
4
3
2
1
VRF(t)
EF
c
1 2 3 4
1
VRF
VRF
d
2 3 4
0 1 2 2ft
Figure 1 | Sample, pump mechanism and gate drive waveforms. (a) Scanning electron microscope image of a pump similar to the one used in this study. The etched 2-DEG channel runs from left to right, and the two metallic gates (light grey in the image) cross the channel. Scale bar indicates 1 m. (b) Schematic diagrams of the potential along the channel during four phases of the pump cycle: (1) loading, (2) back-tunneling, (3) trapping and (4) ejection. One cycle transports an electron from the left (source) to the right (drain) lead. (c,d) The two types of gate drive waveforms VRF(t) used in this study: sine waves (c) and shaped pulses (d). Numbered points approximately indicate the corresponding stagesin the pump cycle shown in panel (b). In panel (d), the black line shows the pulse shape programmed into the AWG, and the red line shows the actual waveform measured with a fast sampling oscilloscope after passing through a coaxial line similar to the one in the cryogenic measurement probe. The dotted line shows a sine wave of frequency f/5.
NATURE COMMUNICATIONS | 3:930 | DOI: 10.1038/ncomms1935 | www.nature.com/naturecommunications
2012 Macmillan Publishers Limited. All rights reserved.
NATURE COMMUNICATIONS | DOI: 10.1038/ncomms1935
ARTICLE
151.407
a
151.406
I P(pA)
ef
~/ p (p.p.m.)
~/ p (p.p.m.)
151.405
0.45 VG2 (V )
0.44
151.404 0.435 0.430 0.425 0.420
VG2 (V )
b c
Figure 2 | High-accuracy data. High-resolution measurement of the pump current as a function of exit gate voltage on the n = 1 plateau withf = 945 MHz and B = 14 T obtained by averaging over 17 (small grey points) and 96 (large red points) onoff cycles (Fig. 7 and Methods section). Error bars show the systematic unceratinty US. The horizontal dotted
line shows the expected current corresponding to exactly one electron pumped for each pump cycle, and the horizonal solid line shows the mean of the nine red points, with error bar indicating the systematic uncertainty UR.
20
0
20
40
0.46
0.02 0.00
100
0
100
0.02 0.00~VG2 (V )
~VG2 (V )
Figure 3 | Effect of AWG pulse waveform on pump performance for sample 1. (a) Fractional deviation of pump current from ef in p.p.m., as a function of exit gate voltage, using sine wave drive at 400 MHz (circles), 630 MHz (squares) and 945 MHz (triangles). (b,c) Pump current as a function of offset exit gate voltage VG2 = VG2 VG2,H,
where VG2,H is the high-voltage edge of the n = 1 plateau, at 630 MHz (panel b) and 945 MHz (panel c). Open (solid) points indicate sine
wave (AWG pulse) drive.
contains only low-frequency components. The slow part of the waveform with repetition frequency f is approximately a segment of sine wave of frequency f/5 (dotted line), enabling us to slow down the electron capture process by a factor of 5. Thus, we expect the pump driven with the AWG pulse waveform at frequency f to have the same characteristics as it would have with sine wave drive at frequency f/5.
Using the AWG waveform, and employing a more accurate current measurement system (described in detail in the Methods section), we demonstrate a semiconductor pump generating currents up to 150 pA accurate to better than 1.2 p.p.m., the limit set by our measurement uncertainty. Theoretical analysis predicts that the true pump accuracy is between one and two orders of magnitude better than our experimental uncertainty, making this type of pump an extremely promising candidate for development as a future quantum representation of the ampere.
ResultsAccuracy of the pump. We cooled our pumps to 300 mK and measured the pump current IP using a technique described in more detail in the Methods section. All high-resolution data in Figs 2, 3, 4 and 6 were obtained in a perpendicular magnetic eld B = 14 T. Most measurements reported in this paper were obtained on a single sample, denoted sample 1. Some measurements were made on an additional sample, denoted sample 2. Data are for sample 1 unless stated otherwise. Figure 2 shows a measurement of the pump current on the n = 1 plateau as the exit gate voltage is varied. The 9 red data points were obtained in three experimental runs spread over 4 dierent days, and their mean and s.d. are shown by the horizontal pink line (with error bar indicating the 1.2 p.p.m. systematic uncertainty) and grey box, respectively. This data demonstrate the stability, plateau atness and absolute accuracy of the pump at the 1 p.p.m. level. The accuracy can be seen by comparing the mean, with its associated uncertainty, with the value of ef indicated by the blue dotted line. We used the 2006 CODATA value of e, which has an uncertainty, insignicant for this measurement, of 0.025 p.p.m.22. On the plateau, the dierence IP ef is ( 0.077 0.18) fA or ( 0.51 1.2) p.p.m.
Quantization improvement using AWG drive. The dramatic eect of using AWG pulse drive on sample 1 is shown in more detail in Fig. 3. First, we made high-resolution measurements of the pump current as a function of VG2, using conventional sine wave drive,
shown in Fig. 3a. The quantity plotted is the fractional deviation of pump current from ef in p.p.m., dened as IP106(IP ef)/ef.
At 400 MHz, a broad plateau is seen, accurately quantized within the uncertainty of ~3 p.p.m., which is dominated by the random uncertainty UR for this combination of f and averaging time. When the frequency is increased to 630 MHz, the plateau becomes narrower, although it remains quantized over a nite range of VG2.
Further increase in f results in a catastrophic loss of quantization. Figure 3b and c illustrate how the quantization can be recovered by switching to the AWG pulse waveform illustrated in Fig. 1d. At 630 MHz, the pulse waveform results in a 2 increase in the width of the plateau compared with the sine wave drive, but at 945 MHz, where no plateau was present on this expanded current scale with sine wave drive, the pulse waveform completely restores accurate quantized pumping. Data for sample 2, at f = 630 MHz, is shown in Fig. 4a. This sample did not exhibit as good a quantized plateau as sample 1, illustrated by the lack of a plateau using sine wave drive. However, the plateau could be progressively recovered using AWG waveforms with slower rising edges, illustrated in Fig. 4b. Waveform AWG1 (AWG2) had a rising edge corresponding to a sine wave of frequency 315 (218) MHz.
Fitting to the decay cascade model. The improvement using the AWG pulse drive can be demonstrated more quantitatively by considering theoretical predictions for the overall shape of the IP(VG2) plateau. In Fig. 5a, we plot the measured average
number of electrons pumped per cycle nIP/ef, for four combinations of magnetic eld and pumping frequency, using sine wave drive. In contrast to the high-resolution data of Figs 2, 3 and 4, these data were taken over a wide range of VG2, with relatively low current resolution, to show the full transition n = 01. As B
is increased, the n = 1 plateau becomes wider and the transitions between plateaux become sharper, resulting in improved current accuracy on the plateau, while increasing f has the opposite eect. The n(VG2) data are tted to an analytical formula derived from
NATURE COMMUNICATIONS | 3:930 | DOI: 10.1038/ncomms1935 | www.nature.com/naturecommunications
2012 Macmillan Publishers Limited. All rights reserved.
ARTICLE
NATURE COMMUNICATIONS | DOI: 10.1038/ncomms1935
a
b
a b
14 T1 GHz 150 MHz 150 MHz 150 MHz
14 T 5 T 0 T
~I P(p.p.m.)
0
1.5
V RF
1.0
0.5
0.0
1
0
200
n
n
n FIT
0.9999
100
0
100
200
400
0.02 0.01 0.00
Time
0.454
0.02
0.452
VG2 (V )
~VG2 (V )
0.5 0.4
1.0000
P( 2)
VG2, MIN
VG2 (V )
Figure 4 | Effect of AWG pulse waveform on pump performance for sample 2. (a) Fractional deviation of pump current from ef in p.p.m., as a function of normalized exit gate voltage at 630 MHz, using sine wave drive (open circles), waveform AWG1 (closed triangles) and waveform AWG2 (closed circles). (b) The three drive waveforms used to obtain the data in panel (a). Dotted line: sine wave, dashed line: AWG1 and solid line: AWG 2. The overall amplitude of the waveforms is ~0.8 Vpp.
c d
/ P (p.p.m.)
0.48 0.44 0.01 0.00
VG2 (V )
VG2 (V )
e
the decay cascade model12,13:
= + =
20 106107
108
2
n aV m m
FIT G
2 exp( exp( ))
2
1
(1)(1)
0
0
10 1,000
0
B (T) f (MHz)
where a, 1 and 2 are tting parameters. The exponential approach to the n = 1 quantized plateau, as VG2 is made more positive, is understood as an exponentially decreasing probability of the electron
tunnelling back out of the dot into the source lead, while excess electrons tunnel back out with essentially unit probability (Fig. 1b). The parameter 2 = 2 1 can be used as a practical gure of merit for evaluating the accuracy of a given pump, because there is a simple relationship between 2 and the predicted pump error P1 nFIT
at the point of inection in IP(VG2)13. The point of inection can be found from a relatively fast, low-resolution measurement. The t line and its derivative for the 1-GHz data are shown in Fig. 5b with axes expanded to show P.
Comparison of data with decay cascade model. Our high-resolution measurements enable us to investigate the agreement between efnFIT, obtained from low-resolution data, and the actual value of the pump current on the plateau. Figure 5c and d show a t to low-resolution data, and the comparison with the high-resolution data, using AWG drive at 945 MHz. It can be seen that the t slightly underestimates the width of the plateau, and consequently P is an overestimate of the error at VG2 = VG2,MIN, consistent with earlier, lower-resolution, measurements20. The other high-resolution data sets contributing to our assessment of the pump accuracy posessed similar properties; the t always underestimated the true plateau atness. Consequently, we consider P as a lower limit (in other words, a pessimistic estimate) to the predicted pump accuracy, which for the 945 MHz data is approaching 0.01 p.p.m.
In Fig. 5e, we show 2 as a function of B deduced from the experimental data measured with sine wave drive at f = 150 MHz (le panel) and as a function of f measured at B = 14 T with sine drive (right panel, open circles) and AWG pulse drive (right panel, lled circle). Horizontal lines show the thresholds for obtaining predicted accuracies of 1, 0.1 and 0.01 p.p.m. For f1 GHz using pulse AWG drive and B = 14 T, we predict an accuracy better than 0.1 p.p.m. The value of 2 extracted from the AWG pulse data at f = 945 MHz roughly matches that for f~200 MHz using sine drive. This is in good agreement with the 5 scaling of the time scale expected from the design of the pulse waveform. The data of Fig. 5e together with Figs 3 and 4 support our assumption that the pump errors at high frequency originate during the rst part of the pump cycle, and can be minimized by slowing down this part of the cycle at the expense of other parts.
Figure 5 | Comparison of data with the decay cascade model. (a) Points: average number of pumped electrons nIP/ef as a function of exit gate voltage for four combinations of magnetic eld and pump frequency, using sine wave drive. Solid lines: nFIT obtained by tting the data to equation (1).
Values of the tting parameter 2 are (leftright) 12.2, 20.7, 9.1 and 4.6. (b) Solid line: close-up of the nFIT(VG2) line for the 1-GHz data in (a).
Dotted line: derivative dnFIT/dVG2 (arbitrary units) that is minimized at gate voltage VG2,MIN. The theoretically predicted pump error P1 nFIT is dened for VG2 = VG2,MIN. (c) Points: average number of pumped electrons as a function of exit gate voltage obtained from a quick low-resolution measurement (0.4 s/data point) for AWG drive at 945 MHz. Line: nFIT
obtained by tting the data to equation (1). (d) Comparison between the t from (c) (line) and the high-resolution data using AWG drive at 945 MHz from Fig. 3c (points) plotted on an offset gate voltage scale. (e) Fitting parameter 2 extracted from ts similar to those shown in (a). Horizontal dotted lines show the values of 2 corresponding to predicted pumping errors P of 1, 0.1 and 0.01 p.p.m. The left panel shows 2 as a function of B
for f = 150 MHz, sine wave drive, and the right panel shows 2 as a function of f for B = 14 T. Open circles: sine wave drive; closed circle: AWG drive.
Summary of pump accuracy. In Fig. 6, we summarize the main experimental nding of this study. This gure shows the pump current averaged over the VG2 plateau for seven experimental runs at three frequencies for sample 1, and a single run for sample 2.
The total measurement time is around 10 h per data point. We see that any deviation of the pump current from ef is less than our total measurement uncertainty U U U
T s R
= +
2 2 , which is as low as 1.2 p. p.m. for the highest pump frequencies where the larger signal-to-noise ratio reduces the contribution of UR to a negligible level. A full investigation of the invariance of the pump current as a function of the additional control parameters VG1, the source-drain bias voltage and the amplitude of VRF is beyond the scope of this study, but an initial investigation into the invariance of IP(VG1) was undertaken.
The data point in Fig. 6 marked with a star was obtained with VG1 shied by 10 mV away from the plateau centre. No signicant deviation from ef was seen following this adjustment.
Discussion
We have demonstrated that the pumping errors at high frequency can be reduced by many orders of magnitude by the use of specially
NATURE COMMUNICATIONS | 3:930 | DOI: 10.1038/ncomms1935 | www.nature.com/naturecommunications
2012 Macmillan Publishers Limited. All rights reserved.
NATURE COMMUNICATIONS | DOI: 10.1038/ncomms1935
ARTICLE
a b
200
2
IP e
Vc
I(fA)
~I P(p.p.m.)
A
V
*
3000.000
0.001
151.431 151.432
0.0 100.0 200.0 300.0 Time (s)
On Off On Off
0
R
IR
V(mV)
~V
2
Figure 6 | Summary of experimentally demonstrated pump accuracy. Fractional deviation of pump current from ef in p.p.m., averaged from seven experimental runs on sample 1 at 400 MHz, sine (open circles), 630 MHz, sine (open squares) and 945 MHz, AWG (closed circles), and one experimental run on sample 2 at 630 MHz, AWG (closed diamond). Error bars show the total uncertainty UT. The data point marked with a * was measured with VG1 shifted by 10 mV away from the plateau centre.
Figure 7 | Measurement circuit and raw data. (a) Circuit used for accurate measurement of the pump current. The pump is depicted as a current source. (b) A short section of raw ammeter (upper panel, red points)and voltmeter (lower panel, blue points) data from a measurement run pumping with f = 945 MHz, showing two onoff cycles. Horizonal lines show the means of each data segment, ignoring the rst 20 readings to reject transient effects. The offset current 250 fA results from the small stray bias voltage present at the pre-amp input, driving a current in R.
designed gate drive waveforms with a slow initial rising edge. The data of Figs 3 and 4 support our postulate that errors at high pumping frequency arise during the initial phase of the pump cycle, when the number of electrons in the dot is reduced by the back-tunnel cascade. However, the reason for this is unclear. Indeed, the back-tunnel model12,13 does not predict any characteristic time-scale for the decay cascade as a consequence of assuming a strictly exponential dependence for the tunnel rates on VG1. Instead, it predicts a shi of the plateau to more negative VG2 as the rise time of VG1 is decreased. As already noted, one experiment on silicon MOSFET pumps provides support for the simple back-tunnel model: as the rise time of trapezoidal gate drive pulses was decreased from 20 ns to 2 ns, the plateaus shied to more negative values of the top gate that controlled the dot depth, but the plateau shape did not change12. We do not believe that this result contradicts our own ndings. We have observed the eect of decreasing rise time through the degredation of initially very at plateaus measured to high resolution. In contrast, the plateaus presented in the variable rise-time experiment of ref. 12 were clearly not at even at the longest rise time investigated. It is likely that the plateau atness in this case was not limited by the rise time, and if much shorter rise times than 2 ns had been employed, a further reduction in the plateau atness would have been seen, in agreement with our experiments. Understanding the upper frequency limit of accurate pumping in these devices requires more accurate physical models of the pumping mechanism, for example by incorporating a physically realistic model of the time dependence of tunnel rates into the back-tunnel model, or including the eect of nonadiabatic excitations21.
This is the rst time an electron pump current has been directly compared with a reference current generated outside the pump cryostat with an uncertainty of order 1 p.p.m. In experiments using metallic pumps, a small pump current IP
< 0 5
. pA was compared with units of voltage, capacitance and time by charging a capacitor located in the same cryostat as the pump, and measuring the resulting voltage change across the capacitor6,23. The lowest uncertainty in these experiments was 0.92 p.p.m.24, similar to that in the present work, but with a current 300 times smaller. At the 100-pA current level of the semiconductor pump, a direct realization of the quantum metrological triangle25 at the 10 7 uncertainty level is feasible within an averaging time of several hours. Direct calibrations of a pico-ammeter or a micro-ammeter using a suitable high-gain current comparator2628 are also possible. This makes our result a signicant step towards the application of electron pumps in primary electrical metrology. We have shown that the pump current is
invariant in one control paramter (VG2) and equal to ef within 1.2 p.p.m. Future experiments will investigate in detail the invariance of the current across the additional control parameters and over a range of samples.
Another class of experiment uses mesoscopic charge detectors capactively coupled to part of the pump circuit to count transport errors at the single electron level5. These experiments enable measurements of much smaller error rates than are possible by averaged current measurements such as those reported in this paper. Furthermore, they provide a direct measure of the pump error that is competely independent of the performance of any reference standard. If the electron pump is to be used as the current leg of a metrological triangle experiment, charge-counting accuracy tests are important because they allow the experiment to distinguish between errors in the number of transported charge quanta, and a more fundamental correction to the value of the charge on each quantum29. Error-counting experiments at the GHz frequencies employed in semiconductor pumps are technically challenging owing to the limited bandwidth of detectors, but some initial progress has already been made30,31, and these experiments will continue to be persued.
In summary, we have demonstrated a milestone in single- electron control, by transferring 109 electrons per second through a semiconductor quantum dot with a resulting current accuracy better than 1.2 p.p.m. By using shaped pulses to drive the control gate, we are able to limit errors and operate the pump accurately at higher frequency than was previously possible. These results are extremely encouraging for the development of a quantum current standard, redenition of the unit ampere and more generally for precise high-speed control of electrons in semiconductor devices.
Methods
Sample and cryostat. We fabricated our pumps on GaAs/AlxGa1 xAs wafers using standard techniques: wet etching to dene the channel, and electron-beam lithography for the metallic gates19. The samples were cooled in a sorbtion-pumped helium-3 refrigerator with a base temperature of 300 mK. The special gate pulses were generated using a 12 GS/s Tektronix 7122B AWG.
Current measurement system. Conventional electronic pico-ammeters are limited in practice to 100 p.p.m. accuracy, even following a calibration, owing to dri in the gain of the pre-amp stage. To achieve p.p.m. accuracy, we generated a reference current IR with opposite polarity to IP by applying a voltage VC across a temperature-controlled (26.57 0.02) 1-G standard resistor R (circuit in Fig. 7a).
A room-temperature current pre-amplier, with transimpedance gain 1010 V/A calibrated to 0.1% accuracy, measured the small dierence between the pump and reference currents. To remove oset currents and thermoelectric potentials, VC and
NATURE COMMUNICATIONS | 3:930 | DOI: 10.1038/ncomms1935 | www.nature.com/naturecommunications
2012 Macmillan Publishers Limited. All rights reserved.
ARTICLE
NATURE COMMUNICATIONS | DOI: 10.1038/ncomms1935
VRF were synchronously switched on and o with a cycle time of 150 s. The dierences in current and voltage, I and V, respectively, were recorded for each cycle as indicated in the section of raw data shown in Fig. 7b. The pump current is given by IP = V/R + I + (r/R)I, where r10 k is the input resistance of the pre-amp.
As | |/ , /
<
I I r R
P 10
4, the nal term can be neglected and the pump current is determined by the voltmeter reading, the value of R and the small residual current I with the pre-amp gain calibration contributing a small ( < 0.1 p.p.m.) correction. The voltmeter and resistor were calibrated, via some intermediate steps, againstthe AC Josephson eect and quantum Hall resistance, respectively. Formally, we have measured current in 1990 units, using the agreed values of the von Klitzing and Josephson constants RK 90 and KJ 90. The uncertainties of RK and KJ in SI units are sufficiently small, less than 0.1 p.p.m., that we can quote our result in SI units without a signicant increase in uncertainty. The resistor was calibrated at a voltage of 10 V, with additional calibrations performed at higher voltages to check for possible power-coefficient eects as shown in Supplementary Fig. S1. No power coefficient was found in the voltage range 10 V V100 V. Below 10 V, the resistor dissipates < 100 nW and power coeffiecient eects are expected to be insignicant.
Uncertainty. The total systematic uncertainty US in our measurement of the pump current is 1.2 p.p.m., dominated by the 0.8-p.p.m. systematic uncertainty in the calibration of the resistor. The 2 calibration uncertainty quoted for 1 G resistors at NPL is 1.6 p.p.m. The relevent entry on the peer-reviewed calibration and measurement capability data base is available at http://kcdb.bipm.org/AppendixC/ country_list_search.asp? CountSelected = GB&iservice = EM/DC.2.1.3. All unceratinties quoted in this paper are 1 and have been rounded to the nearest 0.1 p.p.m. A full break down of the uncertainty is shown in Supplementary Table S1. Calibrations of R were performed at regular intervals over several years to characterize the small dri, which is common in artefact standards: R = 1000051260 800 on 27/4/2011, with a dri correction of 11 per day. The maximum dri correction applied for pump measurements before and aer the calibration date was 0.5 p.p.m. The random uncertainty UR is dominated by the Johnson current noise in R4 fA/Hz. Thus, 1 p.p.m. resolution of IP100 pA requires averaging times of the order of 1 h.
References
1. Zimmerman, N. M. A primer on electrical units in the systeme international. Am. J. Phys. 66, 324331 (1998).
2. Kibble, B. P. Present state of the electrical units In Science, Measurement and Technology, IEE Proceedings A Vol 138 (IET, 1991) pp. 187197.
3. Milton, M. J. T., Williams, J. M. & Forbes, A. B. The quantum metrology triangle and the redenition of the SI ampere and kilogram; analysis of a reduced set of observational equations. Metrologia 47, 279286 (2010).
4. Zimmerman, N. M. & Keller, M. W. Electrical metrology with single electrons. Meas. Sci. Technol. 14, 12371242 (2003).
5. Keller, M. W., Martinis, J. M., Zimmerman, N. M. & Steinbach, A. H. Accuracy of electron counting using a 7-junction electron pump. Appl. Phys. Lett. 69, 18041806 (1996).
6. Keller, M. W., Eichenberger, A. L., Martinis, J. M. & Zimmerman, N. M. A capacitance standard based on counting electrons. Science 285, 17061709 (1999).
7. Shilton, J. M. et al. High-frequency single-electron transport in a quasi- one-dimensional GaAs channel induced by surface acoustic waves. J. Phys. 8, L531L539 (1996).
8. Pekola, J. P. et al. Hybrid single-electron transistor as a source of quantized electric current. Nat. Phys. 4, 120124 (2007).
9. Maisi, V. F., Pashkin, Y. A., Kafanov, S., Tsai, J. S. & Pekola, J. P. Parallel pumping of electrons. N. J. Phys. 11, 113057 (2009).
10. Janssen, T.J.B.M. & Hartland, A. Accuracy of quantized single-electron current in a one-dimensional channel. Physica B 284, 17901791 (2000).
11. Zimmerman, N. M., Hourdakis, E., Ono, Y., Fujiwara, A. & Takahashi, Y. Error mechanisms and rates in tunable-barrier single-electron turnstiles and charge-coupled devices. J. Appl. Phys. 96, 52545266 (2004).
12. Fujiwara, A., Nishiguchi, K. & Ono, Y. Nanoampere charge pump by single-electron ratchet using silicon nanowire metal-oxide-semiconductor eld-eect transistor. Appl. Phys. Lett. 92, 042102 (2008).
13. Kashcheyevs, V. & Kaestner, B. Universal decay cascade model for dynamic quantum dot initialization. Phys. Rev. Lett. 104, 186805 (2010).
14. Wright, S. J. et al. Enhanced current quantization in high-frequency electron pumps in a perpendicular magnetic eld. Phys. Rev. B 78, 233311 (2008).
15. Kaestner, B. et al. Single-parameter quantized charge pumping in high magnetic elds. Appl. Phys. Lett. 94, 012106 (2009).
16. Leicht, C. et al. Generation of energy selective excitations in quantum hall edge states. Semiconductor Sci. Technol. 26, 055010 (2011).
17. Wright, S. J. et al. Single- and few-electron dynamic quantum dots in a perpendicular magnetic eld. J. Appl. Phys. 109, 102422 (2011).
18. Fletcher, J. D. et al. Stabilisation of single-electron pumps by high magnetic elds (2011)Preprint at arXiv:1107.4560.
19. Blumenthal, M. D. et al. Gigahertz quantized charge pumping. Nat. Phys. 3, 343347 (2007).
20. Giblin, S. P. et al. An accurate high-speed single-electron quantum dot pump. N. J. Phys. 12, 073013 (2010).21. Kataoka, M. et al. Tunable nonadiabatic excitation in a single-electron quantum dot. Phys. Rev. Lett. 106, 126801 (2011).
22. Mohr, P. J., Taylor, B. N. & Newell, D. B. CODATA recommended values of the fundamental physical constants: 2006. Rev. Mod. Phys. 80, 633730 (2008).
23. Camarota, B. et al. Electron counting capacitance standard with an improved ve-junction Rpump. Metrologia 49, 814 (2012).
24. Keller, M. W., Zimmerman, N. M. & Eichenberger, A. L. Uncertainty budget for the NIST electron counting capacitance standard, ECCS-1. Metrologia 44, 505512 (2007).
25. Piquemal, F. & Geneves, G. Argument for a direct realization of the quantum metrological triangle. Metrologia 37, 207211 (2000).
26. Rietveld, G. et al. 1: 30000 cryogenic current comparator with optimum SQUID readout. Instrum. Meas., IEEE Trans. 52, 621625 (2003).
27. Feltin, N., Devoille, L., Piquemal, F., Lotkhov, S. V. & Zorin, A. B. Progress in measurements of a single-electron pump by means of a CCC. Instrum. Meas., IEEE Trans. 52, 599603 (2003).
28. Steck, B. et al. Characterization and metrological investigation of an R-pump with driving frequencies up to 100 Mhz. Metrologia 45, 482491 (2008).
29. Keller, M. W. Current status of the quantum metrology triangle. Metrologia 45, 102109 (2008).
30. Fricke, L. et al. Quantized current source with mesoscopic feedback. Phys. Rev. B 83, 193306 (2011).
31. Yamahata, G., Nishiguchi, K. & Fujiwara, A. Accuracy evaluation of single-electron shuttle transfer in si nanowire metal-oxide-semiconductor eld-eect transistors. Appl. Phys. Lett. 98, 222104 (2011).
Acknowledgements
This research was supported by the UK Department for Business, Innovation and Skills, the European Metrology Research Programme (grant no. 217257) and the UK EPSRC.
Author contributions
S.P.G. designed and calibrated the measurement system, M.K. designed the pulsedrive technique, J.D.F. designed the measurement probe, S.P.G. and M.K. performed experiments and analysed data, M.K. designed the sample, P.S. performed and supervised sample fabrication, D.A.R. and I.F. provided GaAs wafers, J.P.G. and G.A.C.J. performed electron beam lithography, T.J.B.M.J. contributed project leadership and supervision, and S.P.G. wrote the paper with extensive comments/revisions contributed by M.K. and J.D.F.
Additional information
Supplementary Information accompanies this paper at http://www.nature.com/ naturecommunications
Competing nancial interests: The authors declare no competing nancial interests.
Reprints and permission information is available online at http://npg.nature.com/ reprintsandpermissions/
How to cite this article: Giblin, S.P. et al. Towards a quantum representation of the ampere using single electron pumps. Nat. Commun. 3:930 doi: 10.1038/ncomms1935 (2012).
NATURE COMMUNICATIONS | 3:930 | DOI: 10.1038/ncomms1935 | www.nature.com/naturecommunications
2012 Macmillan Publishers Limited. All rights reserved.
You have requested "on-the-fly" machine translation of selected content from our databases. This functionality is provided solely for your convenience and is in no way intended to replace human translation. Show full disclaimer
Neither ProQuest nor its licensors make any representations or warranties with respect to the translations. The translations are automatically generated "AS IS" and "AS AVAILABLE" and are not retained in our systems. PROQUEST AND ITS LICENSORS SPECIFICALLY DISCLAIM ANY AND ALL EXPRESS OR IMPLIED WARRANTIES, INCLUDING WITHOUT LIMITATION, ANY WARRANTIES FOR AVAILABILITY, ACCURACY, TIMELINESS, COMPLETENESS, NON-INFRINGMENT, MERCHANTABILITY OR FITNESS FOR A PARTICULAR PURPOSE. Your use of the translations is subject to all use restrictions contained in your Electronic Products License Agreement and by using the translation functionality you agree to forgo any and all claims against ProQuest or its licensors for your use of the translation functionality and any output derived there from. Hide full disclaimer
Copyright Nature Publishing Group Jul 2012
Abstract
Electron pumps generate a macroscopic electric current by controlled manipulation of single electrons. Despite intensive research towards a quantum current standard over the last 25 years, making a fast and accurate quantized electron pump has proved extremely difficult. Here we demonstrate that the accuracy of a semiconductor quantum dot pump can be dramatically improved by using specially designed gate drive waveforms. Our pump can generate a current of up to 150 pA, corresponding to almost a billion electrons per second, with an experimentally demonstrated current accuracy better than 1.2 parts per million (p.p.m.) and strong evidence, based on fitting data to a model, that the true accuracy is approaching 0.01 p.p.m. This type of pump is a promising candidate for further development as a realization of the SI base unit ampere, following a redefinition of the ampere in terms of a fixed value of the elementary charge.
You have requested "on-the-fly" machine translation of selected content from our databases. This functionality is provided solely for your convenience and is in no way intended to replace human translation. Show full disclaimer
Neither ProQuest nor its licensors make any representations or warranties with respect to the translations. The translations are automatically generated "AS IS" and "AS AVAILABLE" and are not retained in our systems. PROQUEST AND ITS LICENSORS SPECIFICALLY DISCLAIM ANY AND ALL EXPRESS OR IMPLIED WARRANTIES, INCLUDING WITHOUT LIMITATION, ANY WARRANTIES FOR AVAILABILITY, ACCURACY, TIMELINESS, COMPLETENESS, NON-INFRINGMENT, MERCHANTABILITY OR FITNESS FOR A PARTICULAR PURPOSE. Your use of the translations is subject to all use restrictions contained in your Electronic Products License Agreement and by using the translation functionality you agree to forgo any and all claims against ProQuest or its licensors for your use of the translation functionality and any output derived there from. Hide full disclaimer